D4 Instruction Manual - strain gages

download D4 Instruction Manual - strain gages

of 48

Transcript of D4 Instruction Manual - strain gages

  • 8/18/2019 D4 Instruction Manual - strain gages

    1/48

     

    Model D4 Data Acquisition 

    Conditioner  

    Instruction ManualVersion 1.34

    Micro-Measurements

    P.O. Box 27777Raleigh, NC 27611

    USA

    Phone: (919) 365-3800

    Email: [email protected]

    Website: www.micro-measurements.com 

    January 2013  130-000208

    mailto:[email protected]:[email protected]://www.micro-measurements.com/http://www.micro-measurements.com/http://www.micro-measurements.com/http://www.micro-measurements.com/mailto:[email protected]

  • 8/18/2019 D4 Instruction Manual - strain gages

    2/48

     

    This page intentionally left blank.

  • 8/18/2019 D4 Instruction Manual - strain gages

    3/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    TABLE OF CONTENTS 1  DESCRIPTION ....................................................................... 1 

    1.1  General .......................................................................... 1 

    1.2  Functional Features ...................................................... 2 

    2  SPECIFICATIONS ................................................................. 3 

    2.1  Input Connections ......................................................... 3 

    2.2  Bridge Configurations ................................................... 3 2.3  Data Conversion ............................................................ 3 

    2.4  Measurement Range/Resolution .................................. 4 

    2.5  Measurement Accuracy ................................................ 4 

    2.6  Gage Factor Control ...................................................... 4 

    2.7  Balance Control ............................................................. 4 

    2.8  Bridge Excitation ........................................................... 4 

    2.9  Communication Interface ............................................. 4 2.10  Shunt Calibration .......................................................... 5 

    2.11  Power ............................................................................ 5 

    2.12  Operational Environment ............................................. 5 

    2.13  Case ............................................................................... 5 

    2.14  Size and Weight ............................................................ 5 

    3  FRONT AND REAR PANELS .............................................. 6 

    3.1  USB Interface ................................................................ 7 3.2  Input Connectors .......................................................... 7 

    3.3  Status LED ..................................................................... 7 

    3.4  Software ........................................................................ 7 

    4  HARDWARE OPERATION ................................................. 8 

    4.1  Power Up....................................................................... 8 

    4.2  Input Connections ......................................................... 8 4.2.1  Pin Functions ......................................................................... 8 4.2.2  Strain Gage and Transducer Input Circuits ......................... 11 

    5  SOFTWARE OPERATION ................................................ 14 

    5.1  Overview ..................................................................... 14 

  • 8/18/2019 D4 Instruction Manual - strain gages

    4/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    5.2  Connecting the D4(s) to the PC ................................... 14 5.3  Application Startup ..................................................... 15 

    5.5  Screen Management ................................................... 16 

    5.6  Control Screen............................................................. 16 

    5.7  Monitor Screen ........................................................... 16 5.7.1  Identify a D4 ........................................................................ 17 5.7.2  Peak Values ......................................................................... 17 

    5.8  Channel Configuration ................................................ 19 5.8.1  Channel Name ..................................................................... 20 5.8.2  Bridge Type ......................................................................... 20 5.8.3  Number of Decimal Places to Display ................................. 23 5.8.4  Scaling ................................................................................. 23 5.8.5  Balance (Zero) ..................................................................... 24 

    5.9  Saving and Recalling Settings (PC) .............................. 26 

    5.10  Saving Settings to the D4 ............................................ 27 

    5.10.1  Reset (Set to Default) .......................................................... 27 5.10.2  Excitation Output and Noise Rejection Settings ................. 27 5.10.3  Format Offscale Value ......................................................... 28 5.10.4  Poisson’s Ratio .................................................................... 28 

    5.11  Charting ....................................................................... 29 5.11.1  Charting Controls ................................................................ 29 5.11.2  Charting Colors .................................................................... 30 

    5.12  Recording .................................................................... 30 

    5.12.1  Setting up the Data File ....................................................... 30 5.12.2  Data File Format .................................................................. 31 

    5.13  Calibration ................................................................... 33 5.13.1  Shunt Calibration ................................................................ 33 5.13.2  System Calibration Software ............................................... 34 5.13.3  Required Instruments ......................................................... 34 

    5.14  Program Messages ...................................................... 40 

    5.15  Upgrading Firmware ................................................... 40 6  WARRANTY ........................................................................ 42 

  • 8/18/2019 D4 Instruction Manual - strain gages

    5/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    Model D4 Data Acquisition Conditioner 

  • 8/18/2019 D4 Instruction Manual - strain gages

    6/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    This page intentionally left blank.

  • 8/18/2019 D4 Instruction Manual - strain gages

    7/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    -1-

    1 DESCRIPTION 

    1.1 General

    The Model D4 Data Acquisition Conditioner is a portable,

    USB-powered, precision instrument for use with resistive strain

    gages and strain gage –  based transducers.

    The Model D4 has four channels of data acquisition.Connection to each channel is via an RJ-45 connector. Each

    channel of input accepts full-, half-, or quarter-bridge

    configuration. All required bridge completion components for

    120-, 350-, and 1000-ohm bridges are supplied.

    Operation of the Model D4 is performed with commands sent

    via the USB connection. Application software is provided tocontrol the D4 with an MS Windows –  based personal computer.

    A programmer’s reference kit is available to simplify writing

    custom applications.

    The Model D4 uses modern digital signal processing

    technology to provide excellent noise rejection and stability.

    Proprietary scaling and linearization algorithms provideunsurpassed measurement accuracy for strain-gage bridge

    measurements.

  • 8/18/2019 D4 Instruction Manual - strain gages

    8/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    -2-

    1.2 Functional FeaturesThe Model D4 Data Acquisition Conditioner has the following

    features:

      Four input channels.

      Hardware and software support for quarter-, half-and full-bridge circuits.

      Built-in precision bridge completion for 120-, 350-

    and 1000-ohm half and quarter bridges.  8Hz sampling rate.

      Automatic and manual zero-balance and calibration.

      Full control of all functions via the USB Interface.

      Portable, lightweight, rugged design.

      Powered via the USB interface.

      Intuitive, user-friendly software.

      Programmability.

  • 8/18/2019 D4 Instruction Manual - strain gages

    9/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 3 -

    2 SPECIFICATIONS    Note: Performance may be degraded at high levels of

    repetitive electrostatic discharge; however, no damage to

    the unit will occur.

    2.1 Input Connections

    Type:  RJ-45 Modular

    Quantity: Four

    2.2 Bridge Configurations

    Types: Quarter, half, and full bridges

     Bridge Impedance: 60Ω to 2000Ω 

     Internal Bridge Completion:

    Quarter bridge: 120Ω, 350Ω, and 1000Ω ±0.01%

     Half bridge: 1000Ω ±0.01%2.3 Data Conversion

     A/D Converter:  Delta-sigma with integral chopper-stabilized,

     programmable gain instrumentation amplifier.

     Resolution: 24 bits. Noise-free resolution: 18 bits typ.

     Filter : Integrated linear phase FIR Sinc5  filter followed by a

    Sinc3  filter with a programmable decimation rate. Softwareselectable output rate provides >120 dB rejection of 50Hz,

    60Hz and higher level harmonics.

  • 8/18/2019 D4 Instruction Manual - strain gages

    10/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 4 -

    2.4 Measurement Range/ResolutionStrain Range: ±31,000με at GF = 2.000. (±15.5 mV/V)

     Resolution: ±1με at GF = 2.000 (±0.0005 mV/V)

    2.5 Measurement Accuracy

    ±0.1% of reading ±3 counts.

    (Instrument Gage Factor = 2.000)

    2.6 Gage Factor Control Range: 0.500 to 9.900

    2.7 Balance Control

    Type: Software

    Control: Manual or automatic

    2.8 Bridge ExcitationValue: 1.5 VDC nominal

    Control: Software enable/disable

    Measurements are fully ratiometric and not degraded by

    variations in excitation voltage.

    2.9 Communication Interface

    Universal Serial Bus (USB). Cable included.

  • 8/18/2019 D4 Instruction Manual - strain gages

    11/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 5 -

    2.10 Shunt Calibration Location: Across each quarter-bridge completion resistor

    Control: Software

    Values:

    P- to D120: 11.94KΩ ±0.1% (5000με at GF = 2.00) 

    P- to D350: 34.8KΩ ±0.1% (5000με at GF = 2.00) 

    P- to D1000: 99.5KΩ ±0.1% (5000με at GF = 2.00) 

    2.11 Power

    USB: 5V 100mA

    2.12 Operational Environment

    Temperature: 0° to +50°C

     Humidity: Up to 90% RH. Non-condensing. 

    2.13 Case Material : Aluminum

    2.14 Size and Weight

    Size: 4.3W x 1.4H x 5.7L inches (110 x 36 x 145 mm)

    Weight : 0.8 lb. (0.36 kg)

  • 8/18/2019 D4 Instruction Manual - strain gages

    12/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 6 -

    3 FRONT AND REAR PANELS

    Status LED

    Input

    Connectors

    USB Port

  • 8/18/2019 D4 Instruction Manual - strain gages

    13/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 7 -

    3.1 USB InterfaceThe USB interface is the communication channel between the

    Model D4 and a host PC.

    When connected, a PC (running MS Windows XP, Vista, or 7)

    identifies the D4 as a standard USB Human Interface Device;

    therefore, no special device drivers are required.

    3.2 Input Connectors

    The input connectors connect the strain gage or transducer to

    the D4.

    Mating connectors for the D4:

    Tyco 554739 or equivalent (MM Part No. D4-A106)

    Recommended crimping tool for the mating connectors:

    Tyco 3-231652-0 (MM Part No. D4-A108)

     Note: Crimping tool can be used on both shielded and

    unshielded connectors.

    3.3 Status LED

    The Status LED illuminates when the D4 is powered up and

    ready for operation.

    3.4 Software

    The D4 is shipped with user-friendly application software that

     provides data recording and control as well as calibration. A

    Programmer’s Reference Kit is included that contains a

    Programmer’s Reference Manual, an NI LabVIEW instrumentdriver, and programming examples. 

  • 8/18/2019 D4 Instruction Manual - strain gages

    14/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 8 -

    4 HARDWARE OPERATION 4.1 Power Up

    The D4 receives its power via the USB connection. The Status

    LED illuminates when the D4 is ready to begin operation.

    4.2 Input Connections

    Strain gages and strain gage –  based transducers are connectedto the D4 through the four input connectors labeled 1 through

    4.

    4.2.1 Pin Functions

    Each channel of the D4 accommodates an eight-pin modular

     plug (MM Part No. D4-A106). The pin arrangement and

    functions are shown in Table 1 and in Figure 1. Input Shielding: To reduce noise and protect the integrity of

    the bridge signal from electrostatic discharge (ESD) and radio

    frequency (RF) noise, a shielded cable connection is

    recommended. The shield or drain wire should be connected to

    the shield of the input connector.

    Symbol Strain Gage Card Function Pin

    P- Negative Excitation 1

    P+ Positive Excitation 2

    S- Negative signal amplifier input 3

    D120 120 ohm Dummy Resistor 4

    D350 350 ohm Dummy Resistor 5

    S+ Positive signal amplifier input 6

    D1000 1000 ohm Dummy Resistor 7

     N/C Unused 8

    SHIELD Shield S (Shield)

    Table 1 - Model D4 Individual Pin Functions

  • 8/18/2019 D4 Instruction Manual - strain gages

    15/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 9 -

    Figure 1 - Model D4 Input Plug Orientation

  • 8/18/2019 D4 Instruction Manual - strain gages

    16/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 10 -

    2

    6

    1

    3

    4

    5

    7

    S

    EXCITATION +

    SIGNAL +

    EXCITATION  –

    SIGNAL -

    D120

    D350

    SHIELD

    Connector on D4

    D1000

     

    Figure 2 - Model D4 Input Plug Connections 

    Half-Bridge and Quarter-Bridge Inputs: In addition to full-

     bridge operation, the D4 can accept half- and quarter-bridge

    configurations. Internal completion resistors are provided for

    120Ω, 350Ω, and 1000Ω quarter - bridges. A 1000Ω internal

    half-bridge circuit is available for quarter- and half-bridgeconfigurations.

    The connection to S+ for the half-bridge completion resistors is

    under firmware control.

  • 8/18/2019 D4 Instruction Manual - strain gages

    17/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 11 -

    4.2.2 Strain Gage and Transducer Input Circuits

    The basic input circuits for strain gage and strain gage –  based

    transducers are shown in Figure 3 through 9.

    2

    4

    3

    S

    EXCITATION +

    D120

    SIGNAL -

    SHIELD

     Figure 3 –  120 Ohm Quarter Bridge / Three-Wire Connection

    2

    5

    3

    S

    EXCITATION +

    D350

    SIGNAL -

    SHIELD

     

    Figure 4 –  350 Ohm Quarter Bridge / Three-Wire Connection

  • 8/18/2019 D4 Instruction Manual - strain gages

    18/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 12 -

    2

    7

    3

    S

    EXCITATION +

    D1000

    SIGNAL -

    SHIELD

     Figure 5 –  1000 Ohm Quarter Bridge / Three-Wire Connection

    2

    3

    1

    S

    EXCITATION +

    SIGNAL -

    EXCITATION -

    SHIELD

    REXT

     

    Figure 6 –  Quarter Bridge / External Dummy

    2

    3

    1

    S

    EXCITATION +

    SIGNAL -

    EXCITATION -

    SHIELD

     

    Figure 7 –  Half Bridge / Adjacent Gages

  • 8/18/2019 D4 Instruction Manual - strain gages

    19/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 13 -

    2

    3

    1

    S

    EXCITATION +

    SIGNAL -

    EXCITATION -

    SHIELD

     Figure 8 –  Half Bridge / Separate Gages

    2

    6

    1

    3

    S

    EXCITATION +

    SIGNAL +

    EXCITATION -

    SIGNAL -

    SHIELD

    i

     

    Figure 9 –  Full Bridge / Four-Wire plus shield

  • 8/18/2019 D4 Instruction Manual - strain gages

    20/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 14 -

    5 SOFTWARE OPERATION 5.1 Overview

    The Model D4 is supplied with application software to

    configure the D4 and collect data via the USB Interface. The

    software supports a combination of up to six Model D4 or

    Model P3 devices, differentiated by serial number. The

    firmware version of your D4 (or P3) must be 2.50 or greater towork with this application. Use the upgrade firmware feature

    of your previous application to update the firmware to the latest

    level.

    5.2 Connecting the D4(s) to the PC

    Connect the USB interface on the D4 to an open USB port on

    the PC using the supplied cable. After the Status LED on the

    D4 illuminates, the PC should recognize the D4 automatically.

    When multiple D4s or P3s are connected at the same time, it

    may take the PC several minutes to recognize all units.

    You may monitor the PC’s

    recognition of each device

    in the Windows DeviceManager. A USB Human

    Interface Device will appear

    in the Device Manager for

    each D4 and P3 connected.

  • 8/18/2019 D4 Instruction Manual - strain gages

    21/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 15 -

    5.3 Application StartupOnce the devices are connected to your PC, start the

    application from Windows by navigating to the following item

    in the startup menu:

    Start>>Programs>>

     Micro-Measurements P3-D4>> P3-D4 Application

    The application will automatically connect to the devices. Depending on the speed of the PC and the number of devices,

    this process can take 1-5 minutes.

    When the software has recognized the devices, the software

    will display a Monitor Screen for each device connected and

    one Control Screen that contains the menu bar.

  • 8/18/2019 D4 Instruction Manual - strain gages

    22/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 16 -

     Note: Additional devices should not be added while theapplication software is running. Exit the software before

    adding or removing devices from the system.

    5.5 Screen Management

    The Control Screen and Monitor Screens may be moved by

    clicking and dragging the title bar. A Monitor Screen may be

    re-sized by clicking and dragging the lower right corner of the

    screen.

    5.6 Control Screen

    The control screen gives you access to the main program menu

    and the recording controls.

    5.7 Monitor ScreenThe Monitor Screen shows the channel readings. By default,

    the channel readings are active when you start the program. To

    turn the readings display on or off, on the Control Screen

    menu, select Hardware>>Real Time Display.

  • 8/18/2019 D4 Instruction Manual - strain gages

    23/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 17 -

    The screen on the left shows a Monitor Screen when readingsare inactive. The screen on the right shows the screen when

    readings are active. Note that readings are grayed out when the

    display is inactive.

    5.7.1 Identify a D4The serial number of each D4 displays in the title bar of the

    window. If you need to identify which D4 is associated with a

    serial number, you may select Hardware>>Identify and then

    select that serial number. The LED light will blink on the D4

    that has that serial number. 

    5.7.2 Peak ValuesThe display of peak values (“Min peak ” and “Max peak ”) may

     be turned on or off with the menu item  File>>Display Peak

    Values. Peak values for all channels may be reset with the

  • 8/18/2019 D4 Instruction Manual - strain gages

    24/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 18 -

    menu item Channels>>Reset Peak Values. Peak values for aspecific channel may be reset by right-clicking in the channel

    reading segment and selecting Reset Peak.

    No peak valuesdisplayed 

    Peak values

    reset 

    Peak valuesdisplayed 

  • 8/18/2019 D4 Instruction Manual - strain gages

    25/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 19 -

    5.8 Channel ConfigurationTo access the configuration values for a channel, select

    Channels>>Settings  on the Control Screen. (The real-time

    display readings  must be disabled to access this screen.) If

    multiple devices are connected, click on the serial number of

    the desired D4 under Devices. The LED light on that D4 will

     blink to confirm the selection.

    Click on the channel to be configured.  Note: Each of the four

    channels on a D4 must be set independently.

  • 8/18/2019 D4 Instruction Manual - strain gages

    26/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 20 -

    At the bottom of the screen are three buttons:  Click Cancel to exit the screen without saving changes.

     Note: Cancel will not undo any changes that have been

    applied. 

      Click Apply to apply the configuration changesimmediately. Results may be viewed on the Monitor

    Screen.

      Click OK to exit the channel configuration screen andsave changes.

    5.8.1 Channel Name

    The channel name entered will identify the channel on the

    Monitor Screen and in the recorded data file. The default is:

    Channel N, where N is the channel number 1 through 4.

    5.8.2 Bridge Type

    The bridge type determines whether the internal half bridge is

    used and determines the appropriate scaling/linearization

    algorithm. The following table summarizes the bridge types.

    For more information on the calculations used by the D4,

     please contact the Micro-Measurements application

    engineering department.

    The following bridge configuration examples assume that the

    raw strain data is to be reduced to equivalent uniaxial strain

    with the nonlinearity correction applied where appropriate.

    Select the "Undef" bridge, either Half-Bridge/Quarter-Bridge

    or Full-Bridge when you require the net output of the active

    strain gages without mathematical correction for either bridge.

    mailto:[email protected]?subject=D4%20Channel%20Configurationmailto:[email protected]?subject=D4%20Channel%20Configurationmailto:[email protected]?subject=D4%20Channel%20Configurationmailto:[email protected]?subject=D4%20Channel%20Configurationmailto:[email protected]?subject=D4%20Channel%20Configuration

  • 8/18/2019 D4 Instruction Manual - strain gages

    27/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 21 -

    Selection Bridge Diagram Description

    Quarter

    Single active gage in

    uniaxial tension or

    compression.

    HB adj

    ,-  

    Two active gages with

    equal and opposite

    strains — typical of

     bending-beam

    arrangement.

    HB opp

    ,-  

    Two active gages with

    equal strains of the same

    sign — used on opposite

    sides of column with

    low thermal gradient

    (bending cancellation,

    for instance.)

    HB shear

    Two active gages with

    equal and opposite

    strains aligned with the

    maximum and minimum

     principal strains to

    measure shear strain.

    HB adj

    ,  

    Two active gages in

    uniaxial stress field — 

    one aligned with

    maximum principal

    strain, the other with

    transverse "Poisson"strain. Default Poisson’s

    Ratio = 0.3.

  • 8/18/2019 D4 Instruction Manual - strain gages

    28/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 22 -

    Selection Bridge Diagram Description

    FB 4

    active

    Four active gages with

     pairs subjected to equal

    and opposite strains(beam in bending).

    FB shear

    Four active gages with

     pairs subjected to equal

    and opposite strains

    (shaft in torsion).

    FB  opp

    Four active gages in

    uniaxial stress field — 

    two aligned withmaximum principal

    strain, the other two with

    transverse "Poisson"

    strain (column). Default

    Poisson’s Ratio = 0.3.

    FB  adj

    Four active gages inuniaxial stress field — 

    two aligned withmaximum principal

    strain, the other two with

    transverse "Poisson"

    strain (beam). Default

    Poisson’s Ratio = 0.3. 

    Undef FB

    Undefined full bridge

    circuit. No correction for

    Wheatstone Bridge

    nonlinearity or activearms.

  • 8/18/2019 D4 Instruction Manual - strain gages

    29/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 23 -

    Selection Bridge Diagram Description

    Undef

    HB/QB

    Undefined half- or

    quarter-bridge circuit.

     No correction for

    Wheatstone Bridge

    nonlinearity or active

    arms.

    Table 2 Bridge Types 

    5.8.3 Number of Decimal Places to Display

    Enter the number of values you would like to have after the

    decimal place on the displayed data.

    Value Description

    0 Display no decimal places; e.g. 123

    1 Display 1 decimal place; e.g. 123.12 Display 2 decimal places; e.g. 123.12

    n  Where n is the number of decimal places.

    5.8.4 Scaling

    The engineering units for a channel may be microstrain (ue),

    millivolts per volt (mV/V), or may be user defined. Scaling

    options on the D4 are determined by the units selected.

    5.8.4.1 Strain Gage Inputs

     Microstrain: If the selected units are ue (microstrain), scaling

    is determined by the gage factor of the sensor as follows:

    Range: 0.500 to 9.900

    Default: 2.000

     Millivolts per volt: If the selected units are millivolts per volt

    (mV/V), no scaling is necessary.

  • 8/18/2019 D4 Instruction Manual - strain gages

    30/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 24 -

    5.8.4.2 Transducer InputsCustom: User-defined units (such as psi or cm) assume linear

    scaling. To properly scale the data, enter the following

     parameters:

      Engineering Units

    Enter up to four characters. This will be displayed in

    the channel segment on the Monitor Screen.

      Full Scale in mV/V

    Enter the millivolt per volt value at full scale. This

    value is used in conjunction with the Full Scale in

    Engineering Units value. It is used with engineering

    units other than microstrain (ue) or mV/V. During

    manual shunt calibration, fix the full-scale value in

    engineering units and change the full-scale mV/Vvalue. It is recommended to set this value at or near the

    expected full-scale value when manually scaling the

    channel.

    Range: -15.625 to +15.625 (excluding 0)

    Default: 2.000

      Full Scale in Engineering UnitsDefines the full-scale value for engineering units other

    than ue or mV/V. All scaling is linear.

    Range: -99999 to 99999 (excluding a value of 0)

    Default: 1

    5.8.5 Balance (Zero)

    Balancing specifies that the current value of a sensor is the zerooffset reading. Because this zero offset reading is rarely 0.00,

    it is typically subtracted from all subsequent readings. Channel

    readings returned from the D4 have the zero offset removed.

  • 8/18/2019 D4 Instruction Manual - strain gages

    31/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 25 -

    You may auto-balance a channel, manually balance the

    channel, or disable the removal of the zero offset. The default

     balance value is 0.

    5.8.5.1 Auto Balance (Zero)

    The steps in performing an auto-balance operation for a single

    channel are:

    1.  Enter all of the other channel settings before youauto balance.

    2.  Adjust the sensor to the correct setting for a zeroreading.

    3.  Set the Balance Mode of the channel to Auto.4.  Click the Zero button.5.  The new balance value is reflected in the slide bar

    and applied to the channel reading.

    A zero (auto-balance) also may be applied to an individual

    channel on the Monitor Screen by right-clicking in the channel

    segment and selecting Zero.

    You may also perform a zero (auto-balance) of all channels on

    all D4s by the menu item Channels>>Zero All .

    5.8.5.2 Manual Balance

    Before performing a manual balance operation, it is

    recommended that the real time display is active. The steps in

    fine-tuning the balance value after an auto-balance operation

    are:

    1.  Perform the auto-balance operation; make sure thevalue is applied. (If you don’t wish to auto-balance

    first, then configure the channel and adjust your

    sensor to the correct setting for a zero reading.)

  • 8/18/2019 D4 Instruction Manual - strain gages

    32/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 26 -

    2.  Set the Balance Mode to Manual.3.  The balance value from the auto-balance operationis reflected in the slide bar. The balance value is

    displayed in raw counts.

    4.  Depending on the current reading and the desiredreading, adjust the balance value on the slide bareither up or down.

    The slide bar has the range selections of coarse, medium, fine

    and ultra-fine. The coarse selection has the widest range of

     balance settings and a coarser step size. Ultra-fine has the

    narrowest range of balance settings and a small step size. The

    medium and fine selections fall in between.

    5.8.5.3 Balance Mode Disabled

    Channels with the Balance Mode disabled will not remove any

    zero offset from the reading; however, the initial amplifier zero

    is removed. This mode allows you to evaluate initial bridge

    offsets. Click the Apply button to see the changes reflected in

    the channel reading on the Monitor Screen.

    5.9 Saving and Recalling Settings (PC)

    You may save the application’s current settings to a file onyour PC with the menu item File>>Save Configuration. The

    settings may be recalled with the menu item File>>LoadConfiguration.

    In addition to the channel settings, the Poisson’s Ratio, Record

    Interval, Peak Value setting and Offscale Values are saved.

  • 8/18/2019 D4 Instruction Manual - strain gages

    33/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 27 -

    5.10 Saving Settings to the D4The menu item Hardware>>Save Configuration to Flash lets

    you select a device and store the current device settings to the

    device’s internal flash memory. This will become the default

    configuration for the device after a power-cycle. If you wish to

    return the factory default settings, use the Channels>>Reset

     Al l menu item to reset the settings and then store those settings

    to flash. Channel Readings Adjustments

    5.10.1 Reset (Set to Default)

    The menu item Channels>>Reset All  resets the channel

    configuration values on all devices to their default settings.

    5.10.2 Excitation Output and Noise Rejection Settings

    The Hardware>>Other Device Settings menu item allows youto access the Excitation Output setting and the Noise Rejection

    frequency setting. These fields are only available if the D4

    firmware level is 2.42 or greater.

    The bridge excitation

    output should be enabled

    during normal operation.

    You may disable the

    excitation output to

    identify self-generating

    noise sources.

    The digital filters in the D4

    may be tuned to optimize

    the noise rejection from power line frequencies.

  • 8/18/2019 D4 Instruction Manual - strain gages

    34/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 28 -

    5.10.3 Format Offscale ValueYou may specify the value that is used to indicate positive

    offscale and negative

    offscale readings

    using the menu item

     File>>ConfigureOffscale Display. This

    value is used on theMonitor Screen and in

    recorded data.

    If you are planning on charting your recorded data with a

    spreadsheet, you may wish to change these values to a numeric

    (for example, 99999 and -99999).

    5.10.4 Poisson’s Ratio 

    You may define the Poisson’s ratio of the material to which the

    strain gage is bonded using the menu item

     Hardware>>Pois son’s Ratio. Poisson’s ratio is used to

    calculate strain in bridge configurations that are dependent

    upon Poisson’s ratio. This value is shared among all channels 

    on the selected D4.

    Range: 0.0 to 0.5

    Default: 0.3

  • 8/18/2019 D4 Instruction Manual - strain gages

    35/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 29 -

    5.11 ChartingCharting maps the real time display readings. To begin

    charting, select Channels>>Chart.  The chart window will

    display and charting will begin.

    5.11.1 Charting ControlsThe following fields displayed at the bottom of the screen may

     be used to control the charting:

      Pause Display checkbox –  Suspends/resumes thecharting.

      Update Interval –  Enter the desired interval for howoften the chart is updated in milliseconds. The

    minimum interval is 125 milliseconds.   Number of points –  Enter the desired number of

    data points (readings) to be displayed.

  • 8/18/2019 D4 Instruction Manual - strain gages

    36/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 30 -

      Min Y Axis –  Two options exist for charting theminimum Y axis value:o  One is to check the Auto (min) box. This will

    automatically adjust the min value based on the

    min value of the plotted data.

    o  Or, uncheck the Auto checkbox. This will allowentry of a fixed value in Min Y Axis.

      Max Y Axis –  As with the minimum value, eithercheck the Auto checkbox, or uncheck it and enter avalue in Max Y Axis.

      Channels - Select the channels to be charted.

    5.11.2 Charting Colors

    Colors may be adjusted by selecting File>>Configure Chart

    Trace Colors.  Select and save the colors prior to launching thechart.

    5.12 Recording

    The D4 application software supports both manual and time-

     based recording.

    5.12.1 Setting up the Data File

    To specify the data file name and location, click Select File onthe Control Screen.

    Browse to the path of the desired

    location, enter the file name,

    then click Save.

  • 8/18/2019 D4 Instruction Manual - strain gages

    37/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 31 -

    5.12.1.1 Time-based RecordingThe application records data on a periodic basis. To set the

    interval, select File>>Record Interval . The minimum

    recording interval for a scan is 125 milliseconds. A scan is

    defined as one reading from all active channels. Active

    channels are displayed on the Monitor Screen(s).

     Note: P3s cannot acquire data faster than 2Hz.

    Time-based recording can be started, paused, or stopped from

    the Control Screen.

    5.12.1.2 Manual Recording

    Click the manual recording button Manual Record (Single

    Shot) to record one scan to the data file. Manual recording can be performed when time-based recording is active. The

    maximum recording rate is 8Hz.

    5.12.2 Data File Format

    The data is stored in a standard, ASCII-format text file. This

    format may be viewed with MS Notepad or with a spreadsheet

     program such as MS Excel.

    The first line is the data header. All columns are separated by

    the ASCII tab character 09 (09h). The line is terminated with a

    carriage return and line feed.

    Subsequent lines contain the tab-delimited data. All lines are

    terminated with a carriage return (ASCII character 13 (0Dh)

    and line feed (ASCII character 10 (0Ah).

  • 8/18/2019 D4 Instruction Manual - strain gages

    38/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 32 -

    Text file shown in Notepad 

    File shown after import into MS Excel 

  • 8/18/2019 D4 Instruction Manual - strain gages

    39/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 33 -

    5.13 Calibration5.13.1 Shunt Calibration

    Shunt Calibration functionality allows you to adjust the

    sensitivity of the selected channel (gage factor or full-scale

    value) while a calibration signal is applied. The shunt

    calibration factor is primarily used to remove any offsets due to

    lead-wire resistance.

    When you enable shunt calibration, a precision shunt resistor is

     placed between the P- and the appropriate dummy (D120, D350,

    D1000) resistor. The internal shunt calibration resistors simulate

    an equivalent strain of 5000με  at a gage factor of 2.000. The

    shunt resistor is applied to all channels simultaneously.

    The steps in performing a shunt calibration are:

    1.  Configure the channel.2.  Balance (zero) the channel.3.  Enable the shunt calibration by selecting

     Hardware>>Enable Shunt Calibration on the

    Control Screen.

    4.  View the adjusted channel reading on the MonitorScreen.

    5.  If the channel’s engineering unit is set to ue(microstrain), adjust the gage factor. If the

    engineering units are user-defined, adjust the full-

    scale in millivolts/volt setting. Click the Apply

     button to see the changes reflected in the channel

    reading on the main screen.

    6.  Loop through steps 4 and 5 until the desired readingis obtained.

    7.  Unselect Hardware>>Enable Shunt Calibration onthe Control Screen.

  • 8/18/2019 D4 Instruction Manual - strain gages

    40/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 34 -

    5.13.2 System Calibration SoftwareTo calibrate a D4, first stop the channel readings by

    unselecting  Hardware>>Real Time Display  on the Control

    Screen. Then select  Hardware>>Calibrate  on the Control

    Screen and select the serial number of the D4 to be calibrated.

     Note: Performing an improper calibration may render your D4inoperable.

    5.13.3 Required Instruments

    The calibration process requires a Micro-Measurements Model

    1550A Strain Indicator Calibrator, or another device with

    equivalent features and specifications. The D4 is a ratiometric

    measuring device, so measurement is not affected by

    fluctuations in the excitation voltage. It is therefore important

    to use a Wheatstone Bridge –  based calibrator. Other sources,such as a precision voltage reference, are not suitable for

    calibration purposes. This section assumes that you are using a

    Model 1550A.

    The system calibration process takes four measurements on

    each channel: zero, positive and negative full scale, and amp

    zero. After the measurements are taken, calibration values foramplifier zero and software gain are stored in the internal flash

    memory.

    Following are the steps involved in performing a system

    calibration. These steps must be performed in the order shown

    and all channels must be calibrated. To abort the process after

     beginning calibration, select the Abort Calibration button and power-cycle the D4. To abort before beginning calibration,

    select Exit Calibration.

  • 8/18/2019 D4 Instruction Manual - strain gages

    41/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 35 -

     Note: Each time Take Reading is selected, in-process andcompletion messages are displayed in the status bar at the

     bottom left of the screen.

    Step 1: Get previous calibration values (optional)

    Before calibrating the system, you may want to report the lastcalibration date and the previous (on-card) calibration settings

    for each channel. Select the Get Previous Cal Values button to

    retrieve the data from the D4. Enter any desired text into the

    Additional Report Data text box; this text will be included in

    saved/printed calibration reports.

    To save and print a calibration report, select the Save PreviousCal Values button. When the Save As window displays.

    navigate to the desired location in which to save the file.

    Retain the default file name or change as desired.

    The default file name for the calibration report, whether created

     before or after calibration, is as follows:

    SSSSSS_YYYY_MM_DD_M-M_Device_Cal_Report.txtwhere:

    SSSSSS is the serial number of the D4,

    YYYY_MM_DD is the date, and

    M-M_Device_Cal_Report.txt is the terminator.

  • 8/18/2019 D4 Instruction Manual - strain gages

    42/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 36 -

    When Save is clicked, your default text file viewer, e.g., MS Notepad, will be launched, displaying the calibration values.

    The file may be printed using the viewer.

    Step 2: Calibrate each channel

    All of step 2 must be repeated for each of the four channels on

    the D4. The channels may be calibrated in any order, but steps

    2A through 2E must be performed sequentially. Begin by

    selecting a channel (1 through 4) from the drop-down list.

  • 8/18/2019 D4 Instruction Manual - strain gages

    43/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 37 -

    Step 2A: Install calibrator on selected channelConfigure the Model 1550A Strain Indicator Calibrator as a

    full-bridge circuit and install on the selected channel. See

    Figure 9 for connection details.

    Step 2B: Take a zero readingSet the 1550A calibrator to 0 µε, positive polarity and click

    the Take Zero Reading button.

    Step 2C: Take a positive, full-scale reading

    Set the 1550A calibrator to +30,000 µε  (positive polarity)

    and click the Take Reading button.

    Step 2D: Take a negative, full-scale reading

    Set the 1550A calibrator to -30,000 µε (negative polarity)

    and the click the Take Reading button.

    Step 2E: Take an amp zero reading

    Finally, the system needs to establish a true electrical zero

    (amp zero). Set the 1550A calibrator to 0 µε, positive

     polarity, short the S+ and S- terminals on the 1550A

    calibrator and click the Take Reading button below step

    2E. Values will appear in the Amp Zero and Gain indicatorfields.

     Note: After the first channel has been completed, the Abort

    Calibration is available.

    Step 3: Finish the system calibration process

    After measurements for all four channels have been taken,select the Finish Calibration button to complete the system

    calibration process.

  • 8/18/2019 D4 Instruction Manual - strain gages

    44/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 38 -

    When the calibration process is complete, “Calibration

    Complete” displays in the status bar at the bottom of the

    screen. The system will store the new calibration values in the

    D4’s internal flash memory, along with the calibration date.

    Step 4: Retrieve the new calibration data and date

    (optional)

    After calibration, you may view, save and print the new

    calibration date and calibration settings. To view the data,

    select the Get Current Cal button.

  • 8/18/2019 D4 Instruction Manual - strain gages

    45/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 39 -

    To save and print a calibration report, select the Save Current

    Cal Values button. When the Save As window displays,

    navigate to the desired location in which to save the file.

    Retain the default file name or change as desired. The default

    file name for the calibration report is as follows:

    SSSSSS_YYYY_MM_DD_M-M_Device_Cal_Report.txt, where:

    SSSSSS is the serial number of the D4,

    YYYY_MM_DD is the date, and

    M-M_Device_Cal_Report.txt is the terminator.

    When Save is clicked, your default text file viewer, e.g., MS

     Notepad, will be launched, displaying the calibration report.

    Both the previous and current calibration data will be included.

    The file may be printed using the viewer.

    Click the Exit Calibration button to end the calibration process.

  • 8/18/2019 D4 Instruction Manual - strain gages

    46/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 40 -

    5.14 Program MessagesProgram status and informational messages may be displayed

     by selecting the File>>Show Activity Log  menu item.

    Messages are added to this log only when it is opened.

    5.15 Upgrading Firmware

     Note: Before upgrading, disable the real-time display. Go to

     Hardware>>Real Time Display and unselect it if it is checked.

    To upgrade to a newer version of firmware on the D4, select

     Hardware>>Update Firmware on the menu. Select the serial

    number of the D4 to receive the update. (Recording and

    displays will be halted when selection is made.)

  • 8/18/2019 D4 Instruction Manual - strain gages

    47/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    - 41 -

    A file browser appears that lets you select the firmwareupgrade file. By default this file is named

    “158000134 _D4.hex”. Typically you will navigate to the latest

    file provided by Micro-Measurements. After you select OK, a

    confirmation prompt will appear showing the current firmware

    version of the D4 and the version that you are attempting toinstall. If the versions are correct, select OK to continue. A

     progress bar will display to track the upgrade.

    While the upgrade is occurring, do not touch the D4 hardware

    or the software application. You will be requested to power-

    cycle the D4 and restart the application when the upgrade is

    complete.

    If you need to upgrade several D4’s, it is okay to choose

    Restart Later. That way you only need to restart theapplication once after you have upgraded all of the D4’s.

    However, power-cycle the D4 immediately. Note: You will be

    unable to communicate with the D4 until the application has

     been restarted.

    If you need to stop the upgrade, it’s okay to pull the plug from

    the D4. Just do it early on or it will be too late. You will seesome error messages; then just close the application, and

    restart.

  • 8/18/2019 D4 Instruction Manual - strain gages

    48/48

    Model D4 Data Acquisition ConditionerInstruction Manual

    6 Warranty Vishay Measurements Group, Inc. warrants all instruments it manufactures to be free from defect in materials and

    factory workmanship, and agrees to repair or replace any instrument that fails to perform as specified within threeyears after date of shipment. Coverage of computers, cameras, rechargeable batteries, and similar items, sold in

    conjunction with equipment manufactured by Vishay Measurements Group, Inc. and bearing the identifying name ofanother company, is limited under this warranty to one year after the date of shipment. The warranty on non-

    rechargeable batteries and similar consumable items is limited to the delivery of goods free from defects in materials

    and factory workmanship. This warranty shall not apply to any instrument that has been:

    i.  repaired, worked on or altered by persons unauthorized by Vishay Measurements Group, Inc. in such amanner as to injure, in our sole judgment, the performance, stability, or reliability of the instrument;

    ii.  subject to misuse, negligence, or accident; oriii.  connected, installed, adjusted, or used otherwise than in accordance with the instructions furnished by

    us.

    At no charge, we will repair, at our plant, or an authorized repair station, or at our option, replace any of our products

    found to be defective under this warranty.

    This warranty is in lieu of any other warranties, expressed or implied, including any implied warranties of

    merchantability or fitness for a particular purpose. There are no warranties, which extend beyond the

    description on the face hereof. Purchaser acknowledges that all goods purchased from Vishay Measurements

    Group, Inc. are purchased as is, and buyer states that no salesman, agent, employee or other person has made

    any such representations or warranties or otherwise assumed for Vishay Measurements Group, Inc. any

    liability in connection with the sale of any goods to the purchaser. Buyer hereby waives all rights buyer may

    have arising out of any breach of contract or breach of warranty on the part of Vishay Measurements Group,

    Inc., to any incidental or consequential damages, including but not limited to damages to property, damages

    for injury to the person, damages for loss of use, loss of time, loss of profits or income, or loss resulting frompersonal injury.

    Some states do not allow the exclusion or limitation of incidental or consequential damages for consumer products,

    so the above limitations or exclusions may not apply to you.

    The purchaser agrees that the Purchaser is responsible for notifying any subsequent buyer of goods manufactured by

    Vishay Measurements Group, Inc. of the warranty provisions, limitations, exclusions and disclaimers stated herein, prior to the time any such goods are purchased by such buyer, and the Purchaser hereby agrees to indemnify and hold

    Vishay Measurements Group, Inc. harmless from any claim asserted against or liability imposed on Vishay

    Measurements Group, Inc. occasioned by the failure of the Purchaser to so notify such buyer. This provision is not

    intended to afford subsequent purchasers any warranties or rights not expressly granted to such subsequent

     purchasers under the law.

    Vishay Measurements Group, Inc. reserves the right to make any changes in the design or construction of its

    instruments at any time, without incurring any obligation to make any change whatever in units previously delivered.

    Vishay Measurements Group, Inc.’ sole liabilities, and buyer’s sole remedies, under this agreement shall be limited

    to the purchase price, or at our sole discretion, to the repair or replacement of any instrument that proves, upon

    examination, to be defective, when returned to our factory, transportation prepaid by the buyer, within the applicable

     period of time from the date of original shipment.

    Return transportation charges of repaired or replacement instruments under warranty will be prepaid by Micro-

    Measurements.

    Vishay Measurements Group, Inc. is solely a manufacturer and assumes no responsibility of any form for the

    accuracy or adequacy of any test results, data, or conclusions, which may result from the use of its equipment.

    The manner in which the equipment is employed and the uses to which the data and test results may be put are

    completely in the hands of the Purchaser. Vishay Measurements Group, Inc. shall in no way be liable for damagesconsequential or incidental to defects in any of its products.