CardioGen-82 (Rubidium Rb 82 Generator) INFUSION SYSTEM ... · date of issue: february - 2015 page...
Transcript of CardioGen-82 (Rubidium Rb 82 Generator) INFUSION SYSTEM ... · date of issue: february - 2015 page...
Date of Issue: February - 2015 PATENT NUMBERS US 8,058,632,
Rev. 23 US 7,504,646, US 7,394,074,
US 8,216,181, US 8,216,184
PATENT PENDING
Rx Only (USA)
CardioGen-82®
(Rubidium Rb 82 Generator)
INFUSION SYSTEM USER GUIDE
Manufactured by:
RbM Services, Inc.
For:
Bracco Diagnostics Inc.
259 Prospect Plains Road
Monroe Township, NJ 08831 USA
1-800-447-6883
EU Manufacturer:
Bracco Diagnostics Inc.
259 Prospect Plains Road
Monroe Township, NJ 08831 USA
1-800-447-6883
1-609-514-2200
Authorized Representative
Bracco UK Ltd
Wooburn Green
Bucks, HP10 0HH
United Kingdom
+44 (0) 1628 851500
9300003-00
Date of Issue: February - 2015 Page 2 of 88
CardioGen-82 (Rubidium Rb 82 Generator) Infusion System
Limited Warranty
Bracco Diagnostics Inc. (“BRACCO DIAGNOSTICS”) warrants the CardioGen-82
Infusion System (the “Infusion System”) against any defects in materials and
workmanship for a period of one year from the date of installation. BRACCO
DIAGNOSTICS’ warranty covers all parts, repair labor and its associated expenses for
failures of the Infusion System to perform to its specifications during the warranty period,
subject to the following exceptions: (i) misuse, (ii) abuse, or (iii) alteration (without
BRACCO DIAGNOSTICS’ express written consent).
Any part or component of the CardioGen-82 Infusion System that is judged to be
defective by BRACCO DIAGNOSTICS in material or workmanship during the warranty
period will be repaired or replaced by BRACCO DIAGNOSTICS at its sole option and its
expense. Remedies available under this warranty are limited to repair or replacement of
malfunctioning parts, system replacement, or refund of the purchase price with the
specific remedy subject to election by BRACCO DIAGNOSTICS in its sole judgment.
Application for a warranty remedy must be made to BRACCO DIAGNOSTICS within
(30) days of the apparent malfunction.
EXCEPT AS EXPRESSLY PROVIDED HEREIN, BRACCO DIAGNOSTICS MAKES
NO ADDITIONAL WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, WITH
RESPECT TO THE INFUSION SYSTEM, INCLUDING, BUT NOT LIMITED TO,
ANY IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR USE OR PURPOSE.
BRACCO DIAGNOSTICS SHALL UNDER NO CIRCUMSTANCES BE LIABLE FOR
SPECIAL, INCIDENTAL, CONSEQUENTIAL OR EXEMPLARY DAMAGES OF
ANY NATURE, WHATSOEVER, INCLUDING BUT NOT LIMITED TO,
COMMERCIAL LOSS FROM ANY CAUSE, BUSINESS INTERRUPTION OF ANY
NATURE, LOSS OF PROFITS OR REVENUE, REAL OR PERCEIVED LOSS OF
USE, LOSS ARISING FROM A DEFECT IN DESIGN, MATERIAL AND/OR
MANUFACTURE OR WORKMANSHIP, OR ARISING OUT OF THE
PURCHASER’S FAILURE TO COMPLY WITH ALL OR ANY OF THE
PROVISIONS OF THE INFUSION SYSTEM MANUAL AND/OR THE FAILURE OF
THE INFUSION SYSTEM TO PERFORM AS SPECIFIED, EVEN IF BRACCO
DIAGNOSTICS SHALL HAVE BEEN ADVISED TO THE POSSIBILITY OF SUCH
DAMAGES.
The CardioGen-82 Infusion System should only be serviced by personnel authorized by
BRACCO DIAGNOSTICS. Any service performed by other than BRACCO
DIAGNOSTICS-authorized personnel will void this warranty.
For product complaints or questions regarding the operation and service of the system,
please contact your assigned Bracco Representative.
Date of Issue: February - 2015 Page 3 of 88
TABLE OF CONTENTS
I. SAFETY SUMMARY & GENERAL PRECAUTIONS 6
II. DAILY QUALITY CONTROL 14
III. ELECTROMAGNETIC/ELECTROSTATIC INTERFERENCE 16
IV. LABELS AND SYMBOLS 20
1. INTRODUCTION 23
2. GENERAL DESCRIPTION 23 2.1 CARDIOGEN-82 INFUSION SYSTEM – SYSTEM OVERVIEW 23 2.2 CARDIOGEN-82 INFUSION SYSTEM - DISPOSABLES OVERVIEW 23 2.3 AUTOMATIC INFUSION MODE - OVERVIEW 27
3. SYSTEM SPECIFICATIONS 28
4. SYSTEM OPERATION 32 4.1 INTRODUCTION 32 4.2 THE DISPLAY/CONTROL PANEL 32 4.2.1 POWER ON/OFF SWITCH 32 4.2.2 EMERGENCY STOP/ABORT SWITCH 32 4.2.3 MODE SWITCH 33 4.2.4 MODE DISPLAY 33 4.2.5 INJECT START/STOP SWITCH 35 4.2.6 PURGE SWITCH 35 4.2.7 REFILL SWITCH 35 4.2.8 FLOW RATE SWITCH 35 4.2.9 ELUTION VOLUME DISPLAY AND LIMIT SWITCH 35 4.2.10 PATIENT VOLUME DISPLAY AND LIMIT SWITCH 36 4.2.11 PATIENT DOSE DISPLAY AND LIMIT SWITCH 36 4.2.12 DOSE RATE DISPLAY AND THRESHOLD SWITCH 36 4.2.13 PUMP LIMIT INDICATOR LIGHT 36 4.2.14 HIGH PRESSURE INDICATOR LIGHT 36 4.2.15 VALVE FAILURE INDICATOR LIGHT 37 4.2.16 AC MAIN INDICATOR LIGHT 37 4.3 THE CALIBRATION FACTOR SWITCH 37 4.4 SYSTEM ACTIVATION 37 4.5 PUMP REFILL OPERATION 38 4.6 PURGE MODES 38 4.7 AUTOMATIC INFUSION MODE 40 SUMMARY OF THE AUTOMATIC INFUSION OPERATION 40 4.7.1 DESCRIPTION 43 4.7.2 REPORT PRINTOUT 44
5. OPERATING PROCEDURE 45 5.1 GENERAL NOTES 45
Date of Issue: February - 2015 Page 4 of 88
5.2 CARDIOGEN-82 GENERATOR AND STERILE COMPONENTS 45 5.3 CARDIOGEN-82 INFUSION SYSTEM PREPARATION 46 5.4 INSTALLATION OF THE SYRINGE PUMP COMPONENTS 46 5.4.1 PUMP SYRINGE INSTALLATION (PUMP SYRINGE PKG. #1) 46 5.4.2 PRESSURE SENSOR SYRINGE INSTALLATION (INLET ASSEMBLY PKG. #2) 46 5.4.3 PUMP COVER INSTALLATION 47 5.5 CARDIOGEN-82 GENERATOR INLET ASSEMBLY PACKAGE 47 5.5.1 PRESSURE SENSOR LINE INSTALLATION 47 5.5.2 GENERATOR INLET LINE INSTALLATION 47 5.5.3 GENERATOR INLET LINE STERILIZING FILTER INSTALLATION 47 5.6 CARDIOGEN-82 GENERATOR OUTLET ASSEMBLY (PKG. #3) 48 5.6.1 DIVERGENCE VALVE INSTALLATION 48 5.6.2 GENERATOR OUTLET LINE INSTALLATION 48 5.6.3 GENERATOR WASTE LINE INSTALLATION 48 5.6.4 GENERATOR WASTE LINE TO WASTE SHIELD 48 5.6.5 WASTE BOTTLE PREPARATION AND STERILIZING FILTER INSTALLATION 49 5.7 PATIENT ADMINISTRATION SET INSTALLATION 51 5.8 SALINE SUPPLY INSTALLATION 51 5.9 CARDIOGEN-82 GENERATOR INSTALLATION 51 5.9.1. INSTALLATION PREPARATION 51 5.9.2. CARDIOGEN-82 GENERATOR INSTALLATION 51 5.10 PURGE OPERATION 52 5.10.1 SYRINGE FILLING 52 5.10.2 WASTE LINE PURGING 52 5.10.3 PRESSURE SENSOR LINE 52 5.10.4 PATIENT ADMINISTRATION SET PURGING 52 5.10.5 VOLUMETRIC FLOW RATE VERIFICATION 53 5.11 PATIENT ADMINISTRATION 53
CARDIOGEN-82 (RUBIDIUM 82) GENERATOR - SR-82/85 TESTING
INSTRUCTIONAL WORKSHEET 57
CARDIOGEN-82 INFUSION SYSTEM-CALIBRATION DATA SHEET 61
6. TROUBLESHOOTING GUIDE 65
7. CONTACT INFORMATION 88
8. CARDIOGEN-82® INFUSION SYSTEM SUPPLIES 88
Date of Issue: February - 2015 Page 5 of 88
LIST OF FIGURES
Figure 2. 1 CardioGen-82 Infusion System Diagram 25 Figure 2. 2 CardioGen-82 Infusion System Photograph 25 Figure 2. 3 CardioGen-82 Infusion System Fluid-System Diagram 26
Figure 4.1 Image of the Display/Control Panel 34
Figure 4.2 a Purge Generator to Waste 39
Figure 4.2 b Purge Generator to Patient Administration Set 39
Figure 4.3 a Activity Build Up Cycle 40
Figure 4.3 b Patient Infusion Cycle 41
Figure 4.3 c Infusion Ending Cycle 42
Figure 4.4 Sample of Infusion Report Printout 44
Figure 5.1 Tubing Diagram 56
Figure 5.2 SR-82 / SR-85 Testing Worksheet 62
Figure 5.3 Rb-82 Infusion System Calibration Log Sheet 63
Date of Issue: February - 2015 Page 6 of 88
I. SAFETY SUMMARY & GENERAL PRECAUTIONS
BEFORE OPERATING THE CARDIOGEN-82 INFUSION SYSTEM:
READ THE ENTIRE SAFETY SUMMARY & GENERAL PRECAUTIONS
SECTION
CONSULT THE CARDIOGEN-82 (RUBIDIUM-82 GENERATOR),
PACKAGE INSERT, FOR PROPER DOSAGE AND ADMINISTRATION
The CardioGen-82 Infusion System MUST only be eluted with 0.9% additive-free
sodium chloride for injection, USP (saline).
Do not use any other type of solution with the system.
BEFORE DAILY USE, OPERATORS MUST ENSURE:
ALL SYSTEM CONNECTIONS ARE IN PLACE, SECURE, AND FUNCTIONAL.
PROPER GROUNDING AND ISOLATION STANDARDS ARE MAINTAINED.
OPERATIONAL AND CALIBRATION CHECKS ARE PERFORMED DAILY
PRIOR TO USE OF THE SYSTEM/PATIENT STUDIES.
READ BELOW FOR INFORMATION CONSISTENT WITH CARDIOGEN-82
PACKAGE INSERT
PLEASE SEE CARDIOGEN-82 (RUBIDIUM RB 82 GENERATOR) FULL
PRESCRIBING INFORMATION FOR ADDITIONAL DETAILS
IMPORTANT SAFETY INFORMATION
WARNING: UNINTENDED STRONTIUM-82 (Sr-82) AND STRONTIUM-85 (Sr-85)
RADIATION EXPOSURE
Unintended radiation exposure occurs when the levels of Sr-82 or Sr-85 in the rubidium Rb 82
chloride injection exceed specified limits
Perform generator eluate tests:
1) Record each generator eluate volume, including waste and test volumes, and
keep a record of the cumulative eluate volume.
2) Determine Rb-82, Sr-82, Sr-85 in the generator eluate:
Once a day, prior to any drug administrations, and
At additional daily tests after detection of an Alert Limit. Alert Limits are:
o 14 L for the generator’s cumulative eluate volume, or
o An eluate Sr-82 level of 0.002 μCi/ mCi Rb-82, or
o An eluate Sr-85 level of 0.02 Sr-85 μCi/ mCi Rb-82.
o Perform the additional daily tests at time points determined by the day’s elution
volume; tests are performed every 750 mL.
3) Stop use of a generator at an Expiration Limit of:
o 17 L for the generator’s cumulative eluate volume, or
o 42 days post generator calibration date, or
o An eluate Sr-82 level of 0.01 μCi /mCi Rb-82, or
o An eluate Sr-85 level of 0.1 μCi /mCi Rb-82
Date of Issue: February - 2015 Page 7 of 88
Intended Use
The CardioGen-82 Infusion System is intended to be used in a healthcare facility to deliver the
radiopharmaceutical eluate rubidium-82 chloride to a patient in a controlled manner.
Principles of Operation
A. Type of protection against electric shock:
Class 1
B. Degree of protection against electric shock:
Type B applied part
C. Methods of sterilization or disinfection: Disposable parts used, no sterilized parts
D. Equipment not suitable for use in the presence of flammable mixtures
E. Mode of operation: Continuous
Date of Issue: February - 2015 Page 8 of 88
Electrical Safety
To reduce the risk of ELECTRICAL SHOCK, do not move or operate this system without
all the appropriate doors and panels in place.
HAZARDOUS VOLTAGE exists within the CardioGen-82 Infusion System. To avoid
shock, only trained, qualified service personnel should service the system. Always disconnect
the system from line power before attempting to perform any maintenance. Never touch any
pins on connectors or cables that have become disconnected from a live system.
This system can be set for 100/120/220/240V~ operation via the power entry module. For
230V~ OPERATION USE THE 240V~ SETTING. To change this setting, ensure that the
system is unplugged from its power source, and use a small flat screwdriver to remove the
fuse drawer of the power entry module. Remove the small voltage selector card and rotate the
card until the desired voltage is shown. Then replace the card and fuse drawer. Only use the
fuse type and rating as indicated on the system label located near the power entry module.
System power is 2.5A. The protective ground connection via the grounding conductor in the
power cord is essential for safe operation.
Connections to the patient are PHYSICALLY ISOLATED from all CardioGen-82 Infusion
System power sources. Follow standard health care facility procedures to ensure that there is
no degradation of system electrical performance.
This system is equipped with a three-conductor POWER CORD marked “Hospital Grade.”
The power cord must be plugged into an approved three contact electrical outlet marked
“Hospital Only” or “Hospital Grade” to ensure a reliable ground. Use only a power cord that is
in good condition.
To avoid fire hazard, use a FUSE of the correct type, voltage rating, and current rating as
specified:
Power entry module: Use only the fuse type and rating as indicated on the system
label located near the power entry module. Internal fuses should only be replaced by
qualified personnel.
F1, Power chassis assembly (rear panel), ¼ x 1 ¼ T2, 5AL250V~ (2.5A, 250V~ time
delay)
F2, Power chassis assembly, ¼ x 1 ¼ T1, 5AL250V~ (1.5A, 250V~ time delay)
Refer all service to authorized personnel.
To prevent equipment damage, ACCESSORY EQUIPMENT connected to the
CardioGen-82 Infusion System must be certified to the respective IEC standard (i.e., UL/IEC
60601-1 for medical equipment). Any person who connects equipment to the input or output
signal part configures a medical system and is, therefore, responsible that the system complies
with the requirements of the system standard UL/IEC 60601-1. For questions regarding the
operation and service of the system, please contact your assigned Bracco Representative.
Date of Issue: February - 2015 Page 9 of 88
Explosion Hazard
Do not use the CardioGen-82 Infusion System in the presence of flammable anesthetics or
other flammable substances in combination with air, oxygen-enriched environments, or
nitrous oxide.
Cleaning & Maintenance
Before performing any cleaning operations, turn off the system and unplug the power cord. The
system should only be cleaned with an alcohol-dampened cloth. Be very careful when wiping
the Mylar windows that cover the seven segments LED displays, so as not to puncture them.
The interior of the system should only be cleaned by trained and qualified radiation and
safety personnel. Observe all safety precautions noted in this manual.
To ensure all safety features are maintained to system specifications and that your
CardioGen-82 Infusion system is in optimal working condition, annual preventative
maintenance is recommended. A preventive maintenance program is available through Bracco
Diagnostics Inc. Refer all service to authorized personnel.
Training
In accordance with the CardioGen-82 (Rubidium Rb 82 Generator) handling instructions, all
authorized CardioGen-82 users must be certified in product safety and trained on quality
control procedures.
Aseptic Techniques
Since the eluate obtained from the CardioGen-82 Generator may be intended for intravenous
administration to a human patient, aseptic technique must be strictly observed in all handling
of the eluate and tubing set.
Each user is responsible for maintaining the generator and associated tubing in aseptic
condition. The user must strictly observe aseptic technique following all internal facility
procedures and guidelines.
Air Embolism
An air embolism can cause patient injury or death. Operator vigilance and care, combined
with a set procedure, are essential to avoid injecting air and causing an air embolism:
Before patient infusions, visually inspect tubing for air bubbles.
Before patient infusions, purge air from the patient administration set.
Do not connect the patient until after the purge.
Date of Issue: February - 2015 Page 10 of 88
Radiation Safety
The CardioGen-82 Generator supplied by Bracco Diagnostics Inc. for use with the
CardioGen-82 Infusion System, emits radiation. All applicable radiation safety regulations
should be followed by the user. Inadvertent radiation exposure has been documented with the
misuse of the CardioGen-82 Generator due to Strontium-82/Strontium-85 breakthrough.
When working with any radioactive material, care should be taken to minimize radiation
exposure to the patient, consistent with proper patient management and following ALARA
(As low as reasonably achievable) principles. Additionally, care should be taken to minimize
the radiation exposure to attending personnel.
To limit the exposure of personnel, the CardioGen-82 Generator must be installed in the lead
shield provided in the CardioGen-82 Infusion System as a radiation safety precaution.
Gloves should be worn when handling any of the generator, generator tubing, or waste
components.
All components that may contain residual radioactivity must be stored and disposed of in
accordance with the facility’s radioactive materials license.
Radiopharmaceuticals should be used only by physicians who are qualified through training
and experience in the safe use and handling of radionuclides and whose experience and
training have been approved by the appropriate government agency authorized to license the
use of radionuclides.
If assistance is needed for the proper disposal of this product (including accessories and
components), please contact Bracco Diagnostics Inc.
Date of Issue: February - 2015 Page 11 of 88
Date of Issue: February - 2015 Page 12 of 88
Generator and Disposable Components
Only use and install the CardioGen-82 Generator into the CardioGen-82 Infusion System.
Only use and install the CardioGen-82 compatible sterile tubing set provided by Bracco
Diagnostics Inc. with each generator into the CardioGen-82 Infusion system.
Verify installation is correct before using.
Do not use any sterile tubing sets after the expiration date shown on the labeling.
Do not use sterile tubing sets if the package is opened or damaged.
The Patient Administration Set is single use only. Re-use carries increased risk of cross-
contamination.
For locations where the CardioGen-82 generator and the CardioGen-82 Infusion System
remain in one location (i.e., hospital, physician’s office, or imaging center): the tubing set
connected to the generator and the Infusion System must remain in place until a new generator
is installed. New tubing sets are provided by Bracco Diagnostics Inc. with each new
generator.
Due to the short half-life of CardioGen-82 (Rubidium-82, Rb-82), a time period of 10 minutes
is sufficient to permit Rb-82 to decay before handling Rb-82 eluate. Hospital personnel should
wait at least 10 minutes before handling Rb-82 eluate. Gloves should be worn when handling
any of the generator, tubing, or waste bottle components.
Used sterile tubing sets are biohazardous waste and should be handled in accordance with
institutional procedures and disposed of in accordance with local and national laws.
Please contact Bracco Diagnostics Inc. for ordering additional CardioGen-82 Accessory
supplies.
Date of Issue: February - 2015 Page 13 of 88
Transportation and Storage
The system should not be stacked with other equipment.
Do not attempt to lift this cart. The system is very heavy.
To avoid personal injury, do not move or operate this system without all appropriate lids and
panels in place.
Ensure that the power cord is placed in a manner that it does not get run over. The power cord
can be detached from the system. Reattach the power cord, if necessary, and/or ensure the
power cord is fully engaged in the power entrance module when arriving at the final
destination.
The saline bag hanger can be removed by pulling straight up if removal is necessary for height
clearance during movement. Reinsert when at final destination.
See the information under Generator and Disposable Components for specific details on
transportation requirements regarding the generator and disposable components.
Contact Bracco Diagnostics Inc. for additional instructions about your specific transportation
situation.
Date of Issue: February - 2015 Page 14 of 88
II. DAILY QUALITY CONTROL
Consistent with the CardioGen-82 Generator instructions for use, all quality control procedures
must be performed and recorded each day before patient studies are performed. These procedures
include:
1. Daily generator column wash
2. QUALITY CONTROL TESTING REQUIREMENTS AND EXPIRY LIMITS
PLEASE SEE CARDIOGEN-82 (RUBIDIUM RB 82 GENERATOR) FULL
PRESCRIBING INFORMATION FOR ADDITIONAL DETAILS
IMPORTANT SAFETY INFORMATION
WARNING: UNINTENDED STRONTIUM-82 (Sr-82) AND STRONTIUM-85 (Sr-85)
RADIATION EXPOSURE
Unintended radiation exposure occurs when the levels of Sr-82 or Sr-85 in the rubidium Rb 82
chloride injection exceed specified limits
Perform generator eluate tests:
1) Record each generator eluate volume, including waste and test volumes, and
keep a record of the cumulative eluate volume.
2) Determine Rb-82, Sr-82, Sr-85 in the generator eluate:
Once a day, prior to any drug administrations, and
At additional daily tests after detection of an Alert Limit. Alert Limits are:
o 14 L for the generator’s cumulative eluate volume, or
o An eluate Sr-82 level of 0.002 μCi/ mCi Rb-82, or
o An eluate Sr-85 level of 0.02 Sr-85 μCi/ mCi Rb-82.
o Perform the additional daily tests at time points determined by the day’s elution
volume; tests are performed every 750 mL.
3) Stop use of a generator at an Expiration Limit of:
o 17 L for the generator’s cumulative eluate volume, or
o 42 days post generator calibration date, or
o An eluate Sr-82 level of 0.01 μCi /mCi Rb-82, or
o An eluate Sr-85 level of 0.1 μCi /mCi Rb-82
Date of Issue: February - 2015 Page 15 of 88
3. Daily calibration test
Results must be within the specified limits:
Initial calibration of a new generator should be within 0.95 and 1.05.
Daily calibration should be within 0.90 and 1.10.
If calibration exceeds limits, calculate a new calibration factor. Repeat calibration procedure
until the ratio is within the specified limits.
THE CARDIOGEN-82 INFUSION SYSTEM MUST ONLY BE ELUTED WITH 0.9%
ADDITIVE-FREE SODIUM CHLORIDE FOR INJECTION, USP (SALINE).
Do not use any other type of solution with the system.
Date of Issue: February - 2015 Page 16 of 88
III. ELECTROMAGNETIC/ELECTROSTATIC INTERFERENCE
The CardioGen-82 Infusion system may fail to operate appropriately if exposed to high electromagnetic
fields (which may be generated by sources such as radio transmitters and cellular phones), or to high levels
of electrostatic discharge.
The CardioGen-82 Infusion System is compliant with EN/IEC 60601-1-2 Medical electrical equipment-
Part 1-2: General requirements for safety – Collateral standard: Electromagnetic compatibility-
Requirements and tests. Medical Electrical Equipment needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information provided in this manual.
CardioGen-82 Infusion System should be observed for normal operation when used with other electronic
equipment. The Use of cables other than those specified may result in increased emissions and decreased
immunity of this system. Portable and mobile communications equipment can and may affect Medical
Electrical Equipment.
Guidance and manufacturer’s declaration –electromagnetic emissions
The CardioGen-82 Infusion System is intended for use in the electromagnetic environment specified below. The customer or the user of the CardioGen-82 Infusion System should ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The CardioGen-82 Infusion System uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class A
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations / flicker emissions IEC 61000-3-3
Complies
The CardioGen-82 Infusion System is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies power used for domestic purposes.
Date of Issue: February - 2015 Page 17 of 88
Guidance and manufacturer’s declaration –electromagnetic immunity
The CardioGen-82 Infusion System is intended for use in the electromagnetic environment specified below. The customer or the user of the CardioGen-82 Infusion System should ensure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment –
guidance
Electrostatic discharge (ESD) IEC 61000-4-2
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Electrical fast transient/burst IEC 61000-4-4
±2 kV for power supply lines ±1 kV for input/output lines
±2 kV for power supply lines ±1 kV for input/output lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
±1 kV differential mode ±2 kV common mode
±1 kV differential mode ±2 kV common mode
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
<5 % UT (>95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles <5 % UT (>95 % dip in UT) for 5 sec
<5 % UT (>95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles <5 % UT (>95 % dip in UT) for 5 sec
Mains power quality should be that of a typical commercial or hospital environment. If the user of the [EQUIPMENT or SYSTEM] requires continued operation during power mains interruptions, it is recommended that the [EQUIPMENT or SYSTEM] be powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
3 A / m 3 A / m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Date of Issue: February - 2015 Page 18 of 88
Guidance and manufacturer’s declaration –electromagnetic immunity
The CardioGen-82 Infusion System is intended for use in the electromagnetic environment specified below. The customer or the user of the CardioGen-82 Infusion System should assure that it is used in such an environment.
Immunity test IEC 60601 test
level Compliance
level Electromagnetic environment – guidance
Conducted RF IEC 61000-4-6 Radiated RF IEC 61000-4-3
3 Vrms 150 kHz to 80 MHz 3 V/m 80 MHz to 2,5 GHz
3 Vrms 150 kHz to 80 MHz
3 V/m 80 MHz to 2,5
GHz
Portable and mobile RF communications equipment should be used no closer to any part of the [EQUIPMENT or SYSTEM], including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
Pd
3
5,3
Pd
3
5,3 80 MHz to 800 MHz
Pd
3
7 800 MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey
a, should be less than the
compliance level in each frequency range. b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the CardioGen-82 Infusion System is used exceeds the applicable RF compliance level above, the CardioGen-82 Infusion System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the CardioGen-82 Infusion System.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Date of Issue: February - 2015 Page 19 of 88
Recommended separation distances
between portable and mobile RF communications equipment and the CardioGen-82 Infusion System
The CardioGen-82 Infusion System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the CardioGen-82 Infusion System can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the CardioGen-82 Infusion System as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
Pd
3
5,3 Pd
3
5,3 Pd
3
7
0,01 0.12 0.12 0.23
0,1 0.37 0.37 0.74
1 2.17 2.17 2.33
10 3.69 3.69 7.37
100 11.66 11.66 23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Date of Issue: February - 2015 Page 20 of 88
IV. LABELS AND SYMBOLS
Symbol Definition Symbol Definition
Nameplate/Serial
Number
Type and Rating of Fusing
Manufacturer
Symbol
Operating Temperature
Type B Applied
Part
Attention: Consult
Accompanying Documents
100/120/220/240V~, 50/60Hz, 6.3A
Operating Supply
Voltage and
Frequency (Hz)
XXYYZZZ Serial number.
Read User Manual
Before Operation
Alternating current.
Protective earth
ground
Consult instructions for
use
Date of
manufacture Catalog number
Date of Issue: February - 2015 Page 21 of 88
Symbol Definition Symbol Definition
CAUTION: Federal
law (USA) restricts
this device to sale,
distribution, and use
by, or on the order of,
a physician
With respect to electric
shock, fire, mechanical,
and other specified
hazards, only in
accordance with UL
60601-1 and CAN/CSA
C22.2 No. 601.1 medical
equipment
Dangerous voltage
Explosion hazard
European
representative Contact for service
Single Use Only
Sterile
This product should
be recycled and not
disposed of as
general waste
(subject to WEEE
annex IV resp. EN
50419).
In accordance with
European Union
WEEE Directive
2002/96/EC, Bracco
UK will be fully
responsible for the
coordination,
logistics, and costs
of the WEEE
process
Pyrogen Free /
Non-Pyrogenic
Do not use if package
is damaged
Radioactive Hazard:
WASTE SHIELD
Date of Issue: February - 2015 Page 22 of 88
Radioactive Hazard:
GENERATOR
SHIELD
Radioactive Hazard:
VALVE SHIELD
Radioactive Hazard:
TUBING SHIELD
Date of Issue: February - 2015 Page 23 of 88
1. INTRODUCTION
The CardioGen-82 Infusion System is a complete system for the generation and
delivery of Rubidium-82 (Rb-82) from a CardioGen-82 Generator to a patient for
cardiovascular nuclear medicine procedures. Rb-82 is a short half-life (75 seconds)
positron emitter, highly extracted by the myocardium and is a potassium analogue.
For these reasons, Rb-82 is very useful for myocardial perfusion studies. The Rb-82
obtained from the CardioGen-82 Infusion System is eluted in sterile normal saline
(0.9% additive-free sodium chloride injection, USP) for direct injection into a
patient.
2. GENERAL DESCRIPTION
2.1 CardioGen-82 Infusion System – System Overview
The CardioGen-82 Infusion System is a mobile, self-contained cart complete with a
shield for a CardioGen-82 Generator, a waste bottle shield, a saline syringe pump,
sterile tubing and valve components, a positron detector, and all of the support
electronics necessary to administer controlled levels of Rb-82 activity to a patient.
The CardioGen-82 Infusion System delivers Rb-82 chloride by pumping saline (0.9%
additive-free sodium chloride injection, USP) (which acts as an eluant) through a
CardioGen-82 Generator to produce the Rb-82 chloride eluate. A diagram of the
fluid system for the CardioGen-82 Infusion System is shown in Figure 2.3. Note that
the CardioGen-82 Generator eluate is assayed by a positron (beta) probe which
consists of a plastic scintillator and a photomultiplier tube. This detector and its
associated electronics are designed to reject the normally-occurring 511-keV gamma
rays associated with positron annihilations while detecting the interaction of
positrons in the detector's thin scintillator. The positron detector cannot detect the
presence of Sr-85 (which does not emit positrons), and cannot distinguish between
Sr-82 and Rb-82, in eluate. The system is illustrated in Figure 2.1.
In addition to the syringe pump, the CardioGen-82 Generator, and the positron
detector, the fluid system contains a divergence valve for directing fluid flow in the
CardioGen-82 Infusion System. This divergence valve (see Figure 2.3) is used to
direct the low-level Rb-82 activity that initially leaves the Sr-82/Rb-82 generator to a
shielded waste bottle. Once the Rb-82 activity leaving the CardioGen-82 Generator
reaches levels sufficient for patient injection, this valve directs the Rb-82 eluate to
the patient administration set.
2.2 CardioGen-82 Infusion System - Disposables Overview
The CardioGen-82 sterile disposable tubing set consists of two (2) primary
components: (1) the generator tubing set and (2) the patient administration set which
is replaced for each patient.
Date of Issue: February - 2015 Page 24 of 88
The CardioGen-82 Accessory Package connects the saline bag to the patient
administration set. It consists of tubing, a syringe piston, a high pressure syringe,
filters, valves, and various other fittings necessary to achieve the eluate and bypass
fluid pathways.
The patient administration set connects the generator tubing set to the delivery site
which typically consists of another tubing set, not provided with the CardioGen-82
Infusion System, which provides access to the patient’s vascular system.
THE PATIENT ADMINISTRATION SET IS SINGLE USE ONLY. Re-use
carries increased risk of cross-contamination.
The CardioGen-82 sterile disposable tubing set is Ethylene Oxide (EO) sterilized and
is provided sterile to the user.
CardioGen-82 elution vials are also provided.
Product Description Catalog # Usability
CardioGen-82 Patient Administration Set,
X25 001506 Single-Use Only
CardioGen-82 Accessory Package, X1 001510
Multi-Use
(Must be replaced when
the generator is replaced)
CardioGen-82 Elution Vials, X25 001530 Single-Use Only
CardioGen-82 Elution System:
Accessory Package, X1
Patient Administration Line, X25
Elution Vials, 2 X25
001536
Refer to above for usability of
specific components of the Elution
system
CardioGen-82 Waste Bottle, 2X500ml 001520
Multi-Use
CardioGen-82 Vented Cap for Waste Bottle,
X12 001511
Multi-Use
(Must be replaced when
the generator is replaced)
Date of Issue: February - 2015 Page 25 of 88
Figure 2. 1 CardioGen-82 Infusion System Diagram
Figure 2. 2 CardioGen-82 Infusion System Photograph
Date of Issue: February - 2015 Page 26 of 88
Figure 2. 3 CardioGen-82 Infusion System Fluid-System Diagram
Date of Issue: February - 2015 Page 27 of 88
2.3 Automatic Infusion Mode - Overview
The CardioGen-82 Infusion System provides an AUTOMATIC INFUSION mode for
automatic delivery of Rb-82 chloride to a patient. Additionally, the system provides
two purge modes, PURGE-GENERATOR-TO-WASTE and PURGE-
GENERATOR-TO-PATIENT, for system setup and maintenance. The two purge
modes are used to purge air out of the system tubing following the installation of a
CardioGen-82 Generator and the associated sterile tubing-component set. Detailed
descriptions for all three operating modes are included in Section 4, and an overview
of the AUTOMATIC INFUSION mode is contained in the following section.
The AUTOMATIC INFUSION Mode is used for delivering a pre-selected quantity of
Rb-82 eluate into the patient for myocardial perfusion studies.
To perform an automatic infusion operation, the operator selects the desired patient
dose in mCi and a flow rate of 50 mL/min in accordance with the CardioGen-82
package insert. (NOTE: The term “dose” or “patient dose” is used throughout to refer
to the activity (mCi) of Rb-82 delivered or to be delivered to a patient.)
Additionally, the operator selects a patient volume (mL) and elution volume (mL),
which are used as backup limits to ensure patient safety. Note that patient volume is
the volume administered to a patient, and elution volume is the total volume pumped
through the generator during an infusion. Finally, the operator sets the dose rate
threshold, to 1.0 mCi/sec, which controls when the system will direct eluate (which
initially is routed to the waste bottle) to the patient.
Once an automatic infusion is started, saline (0.9% additive-free sodium chloride
injection, USP) is pumped through the generator and the resulting eluate is routed to
the waste bottle until its dose rate (mCi/sec) exceeds the pre-selected dose rate
threshold. It typically takes 10 to 18 seconds before eluate leaving the generator
becomes sufficiently concentrated to reach the required dose rate threshold of
1.0 mCi/sec. Once the dose rate threshold is exceeded, the Rb-82 eluate is directed
to the patient administration set and both the patient dose (mCi) and patient volume
(mL) are measured and displayed. The infusion continues and stops on whichever
limit is reached first: patient dose, patient volume, or elution volume. As mentioned,
the normal stopping limit is the patient dose.
Once the infusion is complete, the pump stops and the generator eluate is directed
back to the waste bottle to vent any residual generator pressure. At this time, a
complete report of the Rb-82 infusion is printed on the system printer and this same
data is echoed out a RS-232C serial data port. The RS-232C port can be connected
to the customer's computer for analysis or storage of Rb-82 infusion data.
Date of Issue: February - 2015 Page 28 of 88
3. SYSTEM SPECIFICATIONS
OPERATING
MODES
AUTOMATIC INFUSION
PURGE-GENERATOR-TO-WASTE
PURGE-GENERATOR-TO-PATIENT
AUTOMATIC INFUSION MODE
Preset Elution Volume 0-99 mL in 1 mL increments
Preset Patient Volume 0-99 mL in 1 mL increments
Preset Patient Dose 0-99 mCi in 1 mCi increments
Preset Dose Rate
Threshold
0.0 - 9.9 mCi/sec in 0.1 mCi/sec increments.
The proper preset dose rate threshold for Rb-82 is
1.0 mCi/sec.
Flow Rate 20, 35, 50, 65, and 80 mL/min
(50 mL/min must be used for the CardioGen-82 Generator)
Rb-82 Delivery
Accuracy
(Generator Sr-82 Activity 30-120 mCi, Flow Rate 50mL/min) mCi, Flow Rate 50 mL/min)
Generator Activity Accuracy
>40 mCi +/- 10%
20 - 40 mCi +/- 10%
10 - 20 mCi +/- 15%
5 - 10 mCi +/- 20%
Patient Volume
Delivery Accuracy
+0 mL / -5 mL
Automatic Infusion
Report
The system printer prepares a detailed report following every
automatic infusion. This report lists all system settings,
measured infusion volumes and activities, and a history (in one-
second intervals) of activity seen at the detector and patient port.
Additionally, any error conditions are listed if an automatic
infusion is unable to start or is terminated due to an error
condition. Data identical to the system printout is provided on
the RS-232C port.
Date of Issue: February - 2015 Page 29 of 88
SPECIAL SYSTEM MONITORING FUNCTIONS
System Pressure 90 psig (+/- 5 psig) pressure limit sensor is provided
along with a second, redundant pressure limit of 100
psig (+/- 10 psig) provided by the syringe pump.
Valve Operation Valve position is continuously monitored.
Additionally, the valve must transition between
waste and patient positions in less than 2 seconds or
an error will be detected.
Pump Limit Syringe pump limits, both refill (nom. 128 mL) and
pump limits (nom. 15 mL) are monitored.
Valve Shield The valve shield is monitored to detect if it is open.
Additionally, the valve shield remains locked for
any purge or infusion operation and does not unlock
until the detector measures a count rate less than
50 cps.
Monitoring Function Automatic infusion operations will not start if
operation shield-open error is detected. Automatic
infusion operations are terminated (if in progress)
for all of the preceding conditions.
Purge-generator-to-waste and purge-generator-to-
patient operations will not start and will be
terminated if in progress if a high-pressure error,
valve error, pump limit, or shield-open error is
detected.
Syringe Pump Capacity Mechanically limited to ~130mL full capacity.
Date of Issue: February - 2015 Page 30 of 88
RS-232C Port Characteristics
Description:
The RS-232C port echoes out infusion report
printout data for access by a remote computer.
Although this port has a receive line, the
CardioGen-82 Infusion System does not recognize
incoming data.
RS-232C Port Interface Details:
The recommended RS-232 interface cable to be
attached to the CardioGen-82 Infusion System
should be a shielded type cable less than or equal to
100 feet long. If an RS-232 cable is attached to the
CardioGen-82 Infusion System that exceeds the 100
foot length, the system should be observed to verify
normal operation as well as other electronic
equipment in the vicinity of the CardioGen-82
Infusion System. If abnormal operation is
observed, it may be necessary to re-orient the
RS-232 cable to the CardioGen-82 Infusion System.
Signal DB25P Connector Pin
TX (transmit data) (pin 2)
RX (receive data) (pin 3)
This signal is not used.
GND (signal ground) (pin 7)
Data Format Baud Rate
1 Start bit
1 Stop bit
8 Data bits
No Parity
2400 bit/sec
Date of Issue: February - 2015 Page 31 of 88
System Power Requirements
System power is 2.5A max. This system can be set
for 100/120/220/240V~ operation via the power
entry module. To change this setting, ensure that the
system is unplugged from its power source, and use
a small flat screwdriver to remove the fuse drawer
of the power entry module. Remove the small
voltage selector card and rotate the card until the
desired voltage is shown. Then replace the card and
fuse drawer. Use only the fuse type and rating as
indicated on the system label located near the power
entry module.
System Weight: 650 lbs.
System Size: See Fig. 2.1
Shipping Weight: 875 lbs.
Shipping Size: 48"H x 31"W x 41"L
Life Time of Device 10 years, provided all maintenance and service is
performed as required.
Environmental Conditions For
Operation, Transporting And
Storage of System
The system should be operated in a normal office
environment, but can be safely operated, stored, and
transported between temperatures of 40°F (4.4°C) -
110°F (43.3°C). This is primarily due to
characteristics of the detector. Although no known
problems have been encountered with storage and
transport of the system below or above these
temperatures, it would be best not to subject it to
extreme high or low temperatures beyond this point
for an extended period. Avoid condensing humidity
conditions.
Date of Issue: February - 2015 Page 32 of 88
4. SYSTEM OPERATION
4.1 Introduction
This section of the CardioGen-82 Infusion System manual describes the
display/control panel features and operating modes of the infusion system. The user,
however, must refer to the specific operating procedures contained in Section 5 of
this manual for the CardioGen-82 Generator when performing system preparations,
calibrations, breakthrough strontium-82/strontium-85 measurements, and infusions.
4.2 The Display/Control Panel
All displays and controls for the CardioGen-82 Infusion System, except for the
CALIBRATION FACTOR switch, are located on the Display/Control Panel which is
illustrated in Figure 4.1. This Figure should be referred to during the following
discussion of the Display/Control Panel features.
4.2.1 POWER ON/OFF Switch
This momentary-contact, push-button-indicator switch is used to apply power and
turn off the infusion system. When the system is powered, this switch glows green.
If AC power to the infusion system is interrupted, it is necessary to depress this
switch to repower the system. When the system is turned off by the E-STOP /
ABORT switch, the system can only be turned back on by rotating the E-
STOP/ABORT switch clock-wise, until the E-STOP/ABORT switch “resets” by
popping up. If this E-STOP/ABORT switch is not “reset,” the Start switch is
rendered disabled.
4.2.2 EMERGENCY STOP/ABORT Switch
If the Infusion System Model has a red E-STOP /ABORT Switch: Please note that
this switch does not turn off AC power to the system, and as long as the system’s
power cord is plugged into a live AC electrical outlet, there will be AC power to the
system as indicated by the lighted AC Main LED indicator on the Display Control
Panel. This switch works as an emergency cutoff to stop the system in case of
emergencies. In order to turn the system back on, the switch must be rotated clock-
wise, until the switch “resets” by popping up. If this E-STOP/ABORT switch is not
“reset” the Start switch is rendered disabled.
If the Infusion System Model has a momentary-contact, push-button-indicator Power
Off/Abort Switch: Depressing this switch turns off the system.
Date of Issue: February - 2015 Page 33 of 88
4.2.3 MODE Switch
This rotary switch is used to select the operating mode of the infusion system. The
modes available are:
AUTOMATIC INFUSION,
PURGE-GENERATOR-TO-WASTE, and
PURGE-GENERATOR-TO-PATIENT.
4.2.4 MODE Display
The mode display is an illuminated flow diagram showing the routing of saline in the
infusion system. The mode display shows the saline flow path for all operating
modes of the infusion system.
Date of Issue: February - 2015 Page 34 of 88
Figure 4.1 Image of the Display/Control Panel
Date of Issue: February - 2015 Page 35 of 88
4.2.5 INJECT START/STOP Switch
This momentary-contact, indicator-push-button switch is used to start an automatic
infusion. This switch illuminates during an infusion and is extinguished once the
infusion pumping operation is complete. This switch may be depressed during an
infusion to stop the infusion.
4.2.6 PURGE Switch
This momentary-contact, push-button switch is used to purge the infusion system in
the PURGE-GENERATOR-TO-WASTE or PURGE-GENERATOR-TO-PATIENT
mode. It is necessary to continuously depress this switch during a purge operation.
4.2.7 REFILL Switch
This momentary-contact, push-button switch is used to refill the saline pump. To
start a refill operation, depress the REFILL switch once. The pump will begin
refilling and will stop when either the pump limit is reached or the operator depresses
the REFILL switch a second time.
4.2.8 FLOW RATE Switch
This rotary switch is used to select the saline-pump flow rate used during infusion
and purge operations. The available flow rates are:
20 mL/min
35 mL/min
50 mL/min (50 mL/min must be used for the CardioGen-82 Generator) 65 mL/min
80 mL/min
4.2.9 ELUTION VOLUME Display and Limit Switch
This display and limit switch combination is used to display the total generator
elution volume during an infusion operation and to set the elution volume limit for
the infusion. Note that the ELUTION VOLUME display provides a running display
of generator elution volume during an infusion and holds the total elution volume
after the infusion operation is completed. Both the ELUTION VOLUME display and
limit switch operate over the range of 0-99 mL.
Date of Issue: February - 2015 Page 36 of 88
4.2.10 PATIENT VOLUME Display and Limit Switch
This display and limit switch combination is used to display the patient elution
volume during an infusion operation and to set the patient volume limit for the
infusion. Note that the PATIENT VOLUME display provides a running display
of patient elution volume during an infusion and holds the total patient elution
volume after the infusion operation is completed. Both the PATIENT VOLUME
display and limit switch operate over the range of 0-99 mL.
4.2.11 PATIENT DOSE Display and Limit Switch
This display and limit switch is used to display the patient dose (mCi) during an
infusion operation and to set the patient dose limit for the infusion. Note that the
PATIENT DOSE display provides a running display of patient dose during an
infusion and holds the total patient dose after the infusion operation is completed.
Both the PATIENT DOSE display and limit switch operate over the range of 0-99
mCi.
4.2.12 DOSE RATE Display and Threshold Switch
This display and threshold switch combination is used to display the infusion
system dose rate (mCi) during an infusion operation and to set the dose rate
threshold for the infusion. Both the DOSE RATE display and threshold switch
operate over the range of 0.0-9.9 mCi/sec, where mCi/sec represents the amount
of activity produced by the infusion system in a one-second time interval. For
Rb-82 the proper setting is 1.0 mCi/sec.
4.2.13 PUMP LIMIT Indicator Light
This light glows yellow whenever the saline pump is at either the refill limit
(pump fully extended) or at the pump limit (pump fully contracted). No infusion
or purge operation can be started with the pump at the fully contracted pump limit
position. Additionally, if the pump reaches the fully contracted pump limit
position during an infusion or purge operation, that operation will be terminated
automatically.
4.2.14 HIGH PRESSURE Indicator Light
This light glows red whenever the generator inlet pressure exceeds the preset
high-pressure threshold. Infusion and purge operations cannot be started if there
is a high-pressure error. Additionally, if a high-pressure error occurs during an
infusion or purge operation, that operation will be terminated automatically.
Date of Issue: February - 2015 Page 37 of 88
4.2.15 VALVE FAILURE Indicator Light
This light glows red whenever the system has detected a valve error. The system
detects valve positioning errors by measuring the time required for the valve to
move from one position to the next position. Additionally, the system
continuously tests the valve to ensure that it is in the correct position. Infusion
and purge operations cannot be started if there is a valve error. Additionally, if a
valve error occurs during an infusion or purge operation, that operation will be
terminated automatically.
4.2.16 AC MAIN Indicator Light
If equipped, this light glows green as long as the system’s power cord is plugged
into a live AC electrical outlet. This light indicates there is live AC power inside
the system, even though the system is not “powered on.”
4.3 The CALIBRATION FACTOR Switch
The CALIBRATION FACTOR switch contains a four-digit number which
controls the calibration of Rb-82 delivery. This switch is adjusted only during
calibration and is located away from the Display/Control panel to prevent
accidental changing of its setting. The CALIBRATION FACTOR switch is
located on the processing-electronics chassis front panel.
The CALIBRATION FACTOR switch is set by depressing the small buttons
associated with each digit using a pointed object such as a writing pen.
4.4 System Activation
Depressing the POWER ON/OFF switch activates the infusion system and causes
this switch to glow green indicating that the infusion system is powered. Note that
the POWER ON/OFF switch is used in a power latching-relay system to prevent the
infusion system from restarting following an AC power failure. If AC power to the
infusion system is interrupted, or if the E-STOP/ABORT switch is depressed, it is
necessary to depress the POWER ON/OFF switch again to reactivate the infusion
system.
Depressing the POWER ON/OFF switch shuts the infusion system off. This switch
allows the operator to easily shut the system off during normal shutdown. In the
event a problem is experienced, the E-STOP/ABORT switch must be depressed to
shut down the system.
Date of Issue: February - 2015 Page 38 of 88
4.5 Pump Refill Operation
The saline pump must be filled with saline (0.9% additive-free sodium chloride
injection, USP) solution before the CardioGen-82 Infusion System can be purged of
air or before an infusion can be performed. In order to refill the saline pump, the
REFILL switch is depressed once and the pump will begin refilling. The refill
operation can be stopped by the operator by depressing the REFILL switch a second
time or by letting the pump reach its refill limit. It is, of course, important to ensure
that the saline supply valve is open to permit the pump to accept saline from the
saline supply.
4.6 Purge Modes
The CardioGen-82 Infusion System must be purged of air before an infusion can be
performed. There are two purge modes used for purging the system: PURGE-
GENERATOR-TO-WASTE, and PURGE-GENERATOR-TO-PATIENT. In the
PURGE-GENERATOR-TO-WASTE mode, saline (0.9% additive-free sodium
chloride injection, USP) is pumped through the generator and is routed to the waste
bottle by the automatic flow-control valve. In the PURGE-GENERATOR-TO-
PATIENT mode, saline (0.9% additive-free sodium chloride injection, USP) is also
pumped through the generator, but is routed to the patient administration set instead
of the waste bottle. The PURGE-GENERATOR-TO-PATIENT mode SHOULD
NOT be used to purge the infusion system into a patient. Instead this mode is used
to clear the patient administration set using a collection bottle to collect the purged
eluate.
The operator activates the PURGE-GENERATOR-TO-WASTE mode by selecting
PURGE-GENERATOR-TO-WASTE on the MODE switch and by continuously
depressing the PURGE switch. The mode display on the Display/Control panel will
show the routing of the saline (0.9% additive-free sodium chloride injection, USP)
solution through the CardioGen-82 Infusion System during the purge operation. The
PURGE-GENERATOR-TO-WASTE operation is summarized in Figure 4.2(a)
which includes an illustration of the Display/Control mode display.
The operator activates the PURGE-GENERATOR-TO-PATIENT mode by selecting
PURGE-GENERATOR-TO-PATIENT on the MODE switch and by continuously
depressing the PURGE switch. The mode display on the Display/Control panel will
show the routing of the saline (0.9% additive-free sodium chloride injection, USP)
solution through the CardioGen-82 Infusion System during the purge operation. The
PURGE-GENERATOR-TO-PATIENT operation is summarized in Figure 4.2(b)
which includes an illustration of the Display/Control mode display.
Date of Issue: February - 2015 Page 39 of 88
Purge switch is depressed continuously during purge operation
Generator eluate is purged into waste bottle.
Figure 4.2 a Purge Generator to Waste
Purge switch is depressed continuously during purge operation
Generator eluate is purged into the patient administration set.
Figure 4.2 b Purge Generator to Patient Administration Set
Date of Issue: February - 2015 Page 40 of 88
4.7 Automatic Infusion Mode
SUMMARY OF THE AUTOMATIC INFUSION OPERATION
Figure 4.3 a Activity Build Up Cycle
Date of Issue: February - 2015 Page 41 of 88
Figure 4.3 b Patient Infusion Cycle
Date of Issue: February - 2015 Page 42 of 88
Figure 4.3 c Infusion Ending Cycle
Date of Issue: February - 2015 Page 43 of 88 Page 43 of 88
4.7.1 Description The AUTOMATIC INFUSION mode is the normal infusion system mode for delivering a
set dose of Rb-82 (mCi) to the patient, a set patient volume limit (mL), or a set total elution volume limit (mL), whichever comes first. The operator initiates an automatic infusion by selecting AUTOMATIC INFUSION on the MODE switch and then depressing the INJECT START/STOP push-button switch. The infusion system then begins pumping saline (0.9% additive-free sodium chloride injection, USP) through the generator at the flow rate selected by the FLOW RATE switch and the running generator elution volume is displayed on the ELUTION VOLUME display.
The radioactive saline eluate from the generator is first directed to the waste bottle.
However, once the measured infusion dose rate (mCi/sec of infused activity as measured by the detector) exceeds the preset dose rate threshold (selected by the DOSE RATE THRESHOLD switch), the radioactive eluate from the generator is directed to the patient administration set. At this time, both the patient dose (mCi) and the patient volume (mL) are measured and displayed on the PATIENT DOSE and PATIENT VOLUME displays.
The radioactive eluate continues to be infused into the patient administration set until either the patient dose limit (mCi), patient volume limit (mL), or elution volume limit (mL) is reached. These limits are selected by controls on the Display/Control Panel. With a new generator, for the first patient infusion, the patient volume limit may reasonably be set at 30 mL, and the elution volume limit at 40 mL, as is seen at the top of the sample infusion report printout on the next page (Figure 4.4). The patient volume limit and the elution volume limit for each succeeding patient infusion may reasonably be set using as a guide the actual infused patient volume and elution volume, as reported in the infusion report printout, from the previous patient infusion.
Once one of the preset infusion limits is reached, the saline pump is stopped and the
INJECT START/STOP switch stops glowing red to indicate the conclusion of the infusion. At this time, the generator outlet is directed back to the waste bottle and the system printer begins printing out a report of the infusion. The infusion report includes setup parameters, measured infusion data, and a history of activity (mCi) passing the detector port and the patient port in one-second time intervals. The printout data is also sent out the RS-232C port for access by a remote computer.
Note that reaching any of three independent parameter limits (the patient dose limit (mCi),
the patient volume limit (mL), or the elution volume limit (mL)) will stop an infusion. The AUTOMATIC INFUSION mode is summarized in Figure 4.3. Note that the Figure
shows the Display/Control Panel mode display for the activity buildup cycle, the patient infusion cycle, and the infusion end cycle. Additionally, each cycle is described briefly in the Figure.
Date of Issue: February - 2015 Page 44 of 88 Page 44 of 88
4.7.2 Report Printout At the end of an infusion operation, a complete report of the
Rb-82 infusion is printed on the system printer and this same
data is echoed out the RS-232C serial data port. The RS-
232C port can be connected to the facility’s computer for
analysis or storage of Rb-82 infusion data.
A sample automatic infusion printout is shown in Figure 4.4.
Note that the first entry of the report contains status
information. A normal infusion is indicated by the statement
"INFUSION TERMINATED NORMALLY." If the
infusion is unable to start or is interrupted in process, the
detected error condition(s) is (are) printed out. Infusion
error conditions are described in detail in the Section 6 of
this manual.
The next entries in the infusion report are the set-points
selected by the operator for the infusion. Elution volume,
patient volume, patient dose, and dose rate threshold set-
points are listed in the report. Additionally, the selected
flow rate and the selected calibration factor are listed.
The next entry in the infusion report is actual measured
infusion data. The actual elution volume, patient volume,
and patient dose are listed. Additionally, the activity present
at the end of infusion is listed. This activity is the activity
that would be present in an infusion collection bottle at the
end of an infusion. It is this activity value that is used for
calibrations since on-going decay of the collected eluate is
considered. Note that patient dose is the direct sum of
activity leaving the patient port and does not consider on-
going decay of the collected eluate. Finally, the activity
present between detector and waste valve and the activity
present between waste valve and patient are listed along
with the corresponding saline (0.9% additive-free sodium
chloride injection, USP) volumes. This data is informational
only and requires no user action as the infusion system
automatically considers the effects of radioactivity stored in
the tubing lines.
Finally, the infusion report contains an activity profile listing
in one-second intervals. Both the activity measured at the
detector and the activity calculated at the patient port are
displayed for each one-second interval. This data should be
interpreted as the amount of activity passing the detector and
patient port in successive one-second time intervals. Note
that the transit-time delay between the detector and the
patient port can be readily observed in the sample infusion
report. NOTE: THE POSITRON DETECTOR DOES NOT
DETECT THE PRESENCE OF STRONTIUM ISOTOPES
(Sr-82 and/or Sr-85) IN THE ELUATE.
Figure 4.4 Sample of Infusion Report Printout
INFUSION TERMINATED NORMALLY
SET-POINT VALUES:
Elution Volume = 40 mL
Patient Volume = 30 mL
Patient Dose = 50.0 mCi
Dose Rate Threshold. = 1.0 mCi/sec
Flow Rate = 50 mL/min
Calibration Factor = 926
ACTUAL INFUSION DATA:
Elution Volume = 26 mL
Patient Volume = 19 mL
Patient Dose = 50.1 mCi
Infused Activity Present
At End Of Infusion = 45.4 mCi
Activity Present Between
Detector and Waste Valve = 0.674 mCi
in Volume of 0.592 mL
Activity Present Between
Waste Valve and Patient = 3.23 mCi
in Volume of 2.75 mL
ACTIVITY PROFILE Time sec
Detector mCi/sec
Patient mCi/sec
1 0.000 0.000
2 0.000 0.000
3 0.000 0.000
4 0.000 0.000
5 0.000 0.000
6 0.000 0.000
7 0.003 0.000
8 0.056 0.000
9 0.377 0.000
10 1.149 0.000
11 1.937 0.016
12 3.026 0.422
13 3.558 1.191
14 3.648 1.996
15 3.624 3.023
16 3.652 3.487
17 3.650 3.548
18 3.564 3.528
19 3.385 3.558
20 3.134 3.549
21 2.848 3.457
22 2.556 3.270
23 2.277 3.024
24 2.023 2.743
25 1.797 2.462
26 1.611 2.188
27 1.450 1.946
28 1.314 1.729
29 1.204 1.552
30 1.103 1.398
31 1.020 1.268
32 0.578 0.709
Date of Issue: February - 2015 Page 45 of 88 Page 45 of 88
5. OPERATING PROCEDURE
5.1 General Notes
The sterility of each component must be preserved. Be sure to wear protective gloves.
Aseptic techniques must be strictly observed in all handling. Do not remove any luer
lock protective caps until instructed to do so. Additionally, do not allow any
unprotected luer lock fittings to make contact with anything other than its mating luer
lock fitting.
As system pressure can approach 100 PSI, it is important that luer lock fittings be
connected tightly without stripping or otherwise damaging the fittings. Improper luer
lock connections will result in saline (0.9% additive-free sodium chloride injection,
USP) leaks that may be difficult to find and correct.
5.2 CardioGen-82 Generator and Sterile Components
The CardioGen-82 Infusion System sterile disposable tubing set consists of two (2)
primary components: (1) the outlet generator assembly extension tubing which is for
use on multiple patients and is replaced with each generator, and (2) the patient
administration set which is replaced for each patient.
The generator tubing set connects the saline (0.9% additive-free sodium chloride
injection, USP) bag to the patient administration set. It consists of tubing, a syringe
piston, a high pressure syringe, filters, valves and various other fittings necessary to
achieve the eluate and bypass fluid pathways.
The patient administration set connects the generator tubing set to the delivery site
which typically consists of another tubing set, not provided with the CardioGen-82
System, which provides access to the patient’s vascular system.
The CardioGen-82 Infusion System sterile disposable tubing set is Ethylene Oxide
(EO) sterilized, and is provided sterile to the user.
Product Description Catalog # Usability
CardioGen-82 Patient
Administration Set 001506 Single-Use
CardioGen-82 Accessory Package,
X1 001510 Multiple Use
Compare the CardioGen-82 Accessory Package sterile disposable tray
label's parts list, list 001510, with the actual components in the tray. If not
correct, contact BDI Nuclear Medicine Customer Service at 1-800-447-6883.
Date of Issue: February - 2015 Page 46 of 88 Page 46 of 88
5.3 CardioGen-82 Infusion System Preparation
Remove the syringe pump cover by loosening the three mounting thumb screws and
carefully lifting the cover off the pump assembly. Activate the CardioGen-82
Infusion System by depressing the power switch and moving the mode switch to the
Purge Generator to Waste position.
Raise the lid of the hinged valve shield cover to expose the inside of the valve shield
assembly. Visually make sure there are no used disposables present in the system. If
any of the used disposables are present they must be disposed of to prevent possible
confusion with the new set-up. Close the valve cover.
Move the syringe pump to its upper end of the travel position by continuously
depressing the purge switch while the system is in the Purge Generator to Waste
mode. The pump will automatically stop when it reaches the upper end of the travel
position. The LED pump limit light will come on.
NOTE: The contents of the Accessory Pack are packaged in 3 plastic pouches, labeled
Package 1, Package 2, and Package 3. Only open and install the components in the
sequence in which they are packaged, before opening the next subsequent package in the
sequence.
5.4 Installation of the Syringe Pump Components
5.4.1 Pump Syringe Installation (Pump Syringe Pkg. #1)
Loosen the 4 thumb screws, upper and lower, on the pump syringe carriage. Install
the 140 mL pump syringe with its luer lock fitting facing upward into its carriage in
the pump assembly. Be sure to rotate the syringe so that the graduations face toward
the operator. Tighten the upper and lower thumbscrews to secure both the syringe
body and plunger into the carriage. Do not remove the protective cap on this syringe
until directed to do so.
5.4.2 Pressure Sensor Syringe Installation (Inlet Assembly Pkg. #2)
Install the 10 mL pressure sensor syringe by placing it with the luer lock fitting facing
upward into the carriage in the pump assembly. This is done by tilting the syringe
into place with the plunger contacting the bottom of the carriage followed by pushing
the top of the syringe back until it's locked in place. Be sure to rotate the syringe so
the graduations face toward the operator. Do not remove the protective cap on this
syringe until directed to do so.
Date of Issue: February - 2015 Page 47 of 88 Page 47 of 88
5.4.3 Pump Cover Installation
Re-install the pump cover by carefully sliding it over the top of the pump assembly
with the pump syringe and pressure sensor syringe protruding through the pump
cover. Be sure the protective caps on the syringes protect the syringe luer lock fittings
from contacting the pump cover. Verify that the pump cover is fully seated on the
pump assembly.
5.5 CardioGen-82 Generator Inlet Assembly Package
Remove protective cap from 140 mL pump syringe. Remove the tubing assembly
from its package and connect the check valve to the pump syringe. Tighten the luer
lock fittings, but be careful not to damage the pump syringe connection.
5.5.1 Pressure Sensor Line Installation
Remove protective cap from 10 mL pressure sensor syringe. Connect the shortest line
of the inlet tubing assembly to the 10 mL sensor syringe. Tighten the syringe to the
luer lock connection, but be careful not to damage.
5.5.2 Generator Inlet Line Installation
Open the valve shield. Take the longest line of the generator inlet tubing assembly
and insert it down through the tubing shield so that it appears in the generator shield.
Remove the generator shield lead lid and retrieve the free end of the generator inlet
line. Approximately three to six inches of this line will now be in the generator shield
with the luer lock fitting and the cap on. Place the saline supply hook assembly back
onto the pump assembly.
5.5.3 Generator Inlet Line Sterilizing Filter Installation
The Generator Inlet line filter and Valve Assembly is approximately 11 inches long
and has a female luer lock at the proximal end and a purple check valve with a male
luer lock and a blue cap at the distal end. Take the inlet filter and valve assembly
from the generator inlet assembly package 2, remove the protective caps and connect
the female luer lock fitting of the Generator Inlet Filter and Valve Assembly tubing
to the male luer lock fitting on the free end of the generator inlet line that is now
located in the generator shield. Make sure that the luer lock fitting with the blue cap
on the other end of this filter remains in place until directed to do otherwise.
Date of Issue: February - 2015 Page 48 of 88 Page 48 of 88
5.6 CardioGen-82 Generator Outlet Assembly (Pkg. #3)
5.6.1 Divergence Valve Installation
If it is not already open, raise the lid on the valve shield assembly located on the top
of the CardioGen-82 Infusion System and remove the valve retainer. Note: The lid
cannot be raised if the CardioGen-82 Infusion System is unpowered or if the system
detects activity within this shield. Take the divergence valve, (the white valve with
the three arrows and the off sign) and orient it with the handle facing down toward
the bottom of the CardioGen-82 Infusion System. Note: The shortest line of this
connection will be facing towards the left end of the infusion system as you are
looking at it. Finally, without changing the orientation of the valve, rotate the valve
handle so that the valve can be dropped into its actuation carriage, to prevent it from
being moved, with the "valve" arm fitting into the groove of the actuation carriage.
5.6.2 Generator Outlet Line Installation
Take the line with the red cap end and feed it through the hole in the valve shield
assembly. Verify that the end of this line has entered the generator shield. There
should now be two lines in the generator shield; the generator inlet line with its
associated filter and the generator outlet line. Secure the end of the generator outlet
line (located in the valve shield assembly) by inserting into the tubing slot near the
detector and route the line past the detector secure with detector cover.
5.6.3 Generator Waste Line Installation
Take the line from the outlet tubing assembly with the clear end and feed it down
the opening of the valve shield assembly. There should now be three sets of tubing
in the generator shield.
5.6.4 Generator Waste Line To Waste Shield
Take the line in the generator shield with the clear cap and feed it through the hole
inside the generator shield into the waste shield. Tape the tubing in the generator
shield so that the generator will not crimp the tubing.
Date of Issue: February - 2015 Page 49 of 88 Page 49 of 88
5.6.5 Waste Bottle Preparation and Sterilizing Filter Installation
The Waste Line Filter set is approximately 15 inches long and has a female luer lock
at the proximal end and a male luer lock with a clear cap at the distal end. From the
outlet assembly package take the short line with the filter in the middle, remove the
protective caps and connect the female luer lock fitting of the Waste Line Filter
tubing to the male luer lock fitting on the free end of the generator waste line that is
now located in the waste shield. Once this connection has been made, take the waste
bottle and connect the waste line to the top of the waste bottle cap. Make sure you
remove the cover from the needle under the bottle cap cover. Verify that the waste
bottle connections are correct. Take the waste filter and hold it flat against the waste
bottle about half way down the waste bottle and carefully lower the waste bottle into
the waste shield. Make sure no tubing is under the waste bottle as you place it back
in the waste shield and route the extra waste line tubing to prevent any kinking or
obstruction in the tubing. Place the lid on top of the waste shield.
Bracco Diagnostics recommends that a new waste bottle cap be installed with each
new generator installation. The waste bottle should be emptied every morning
prior to system use.
5.6.5.1 Removing Waste Bottle from Waste Shield
The Waste Bottle should be emptied every morning prior to system use.
1. Precautions:
a. The CardioGen-82 generator eluate is radioactive and should be
handled with proper radiation safety precautions.
b. Wait at least 10 minutes after any purge or infusion operation
before lifting the waste bottle shield cover. This will allow any
Rb-82 eluate to decay to a safe level.
c. Always wear gloves before touching any of the tubing, syringe
pump, valve, or collection bottle system components.
2. All components that may contain residual radioactivity must be stored and
disposed of in accordance with the facility’s radioactive materials license.
3. Preparations: Place an absorbent pad on the top of the generator well lid. Be
sure you have prepared the saline disposal location prior to removing the
waste bottle from the system. It is recommended that you use absorbent
pads (see absorbent Pad Kit, sku 001575) designed for placement at the
bottom of the waste bottle well.
4. Open system cabinet door and lift off waste shield lid. Place lid on the
infusion system counter top.
5. Grasp the waste bottle cap and pull the bottle straight out of the waste
shield. DO NOT USE THE WASTE BOTTLE TUBING TO PULL THE
WASTE BOTTLE OUT OF THE WASTE BOTTLE WELL.
Date of Issue: February - 2015 Page 50 of 88 Page 50 of 88
6. Unscrew the cap from the bottle and tap gently over the opening of the
waste bottle to dislodge any drops of waste solution from cap spike back
into the waste bottle. Place the cap on the absorbent pad on top of the
generator well lid.
5.6.5.2 Emptying Waste Bottle
1. Empty the waste bottle into the appropriate receptacle and dispose of in
accordance with the facility’s radioactive materials license.
2. Inspect the waste bottle and waste bottle cap for obvious signs of wear or
any cracks or holes.
3. If the bottle is acceptable, screw the cap back on the bottle. If not, take a
new bottle and a new cap from your inventory.
4. Note: If you use a new cap, make sure you remove the cover from the needle
under the bottle cap cover.
5.6.5.3 Reinstalling Waste Bottle
1. Take the waste filter and hold it flat against the waste bottle about half way
down the waste bottle.
2. Make sure no tubing is under the waste bottle as you place it back in the
waste shield.
3. Carefully place the waste bottle and waste line filter back into the waste
shield. They should slide easily back into the waste shield. If they don’t fit
all the way to the bottom easily, remove and inspect the tubing position.
4. Do not force the waste bottle or waste line filter into position. Doing so may
break or kink part of the system.
5. Pick up waste shield lid and place it on the waste shield making sure no
tubing is caught in between it and the waste shield.
6. Collect the absorbent pad from the top of the generator well lid and dispose
of in accordance with the facility’s radioactive materials license.
Date of Issue: February - 2015 Page 51 of 88 Page 51 of 88
5.7 Patient Administration Set Installation
Remove a patient administration set from its protective wrapping. Connect the
female side of the luer lock fitting to the line on top of the infusion cart system which
comes out the left side of the valve. This patient administration set contains its own
sterilizing filter. This is the only part of the system that you must change as a new
patient is being imaged. Make sure that the valve retainer is now over the valve and
close the lid.
5.8 Saline Supply Installation
0.9% additive-free sodium chloride injection, USP
Take the remaining free tubing of the generator inlet line, which was the first kit you
hooked up, pinch off valve and then carefully insert the spike into your bottle or bag
of preservative-free, normal saline (0.9% additive-free sodium chloride injection,
USP). Make sure the spike is fully inserted. Hang the saline supply on the saline
supply hook located on top of the pump assembly.
5.9 CardioGen-82 Generator Installation
5.9.1.Installation Preparation
Look inside the generator shield and verify that there are two lines available for
connection:
Generator inlet line and its attached filter with a male luer lock fitting.
Generator outlet line with a female luer lock fitting
Make sure the waste line that passes through the generator shield chamber is against
the wall of the generator shield so that it does not interfere with the generator
installation.
5.9.2.CardioGen-82 Generator Installation
Remove the generator from its shipping container and untie the outlet and inlet
tubing. Carefully lower it into the generator shield. Make sure that the generator does
not interfere with the waste line that passes through the generator shield. Remove the
protective caps (generator pigtail line cap is white, and inlet filter and valve cap is
blue) and connect the inlet line on the generator (marked "inlet," and has a female
luer lock fitting on it) to the remaining fitting on the Generator Inlet line filter and
Valve Assembly which has a male luer lock fitting on it.
Remove the protective caps (generator pigtail line cap is white and outlet line cap is
red) and connect the outlet line of the generator (marked outlet and has a male luer
lock fitting on it) to the remaining line in the generator shield which has a female luer
lock fitting on it. Carefully lower the generator shield lid making sure that none of
the lines are restricted.
Date of Issue: February - 2015 Page 52 of 88 Page 52 of 88
5.10 Purge Operation
5.10.1 Syringe Filling
Position the syringe pump to its limit, if not already there, by continuously
depressing the PURGE switch (in the PURGE GENERATOR TO WASTE mode)
until the pump automatically stops. Open the saline supply pinch valve and
depress the REFILL switch once. The syringe pump should begin drawing in
saline (0.9% additive-free sodium chloride injection, USP) and a stream of air
bubbles should appear in the saline supply bottle. The refill operation can be
stopped at any time by depressing the REFILL switch a second time. Otherwise,
the pump will continue refilling until it reaches its refill limit and automatically
stops. Unless problems are encountered, allow the pump to refill until it
automatically stops on its refill limit.
5.10.2 Waste Line Purging
With the MODE switch in the PURGE GENERATOR TO WASTE position
continually depress the PURGE switch to pump saline (0.9% additive-free sodium
chloride injection, USP) through the generator and into the waste bottle. A volume
of 50 mL of saline (0.9% additive-free sodium chloride injection, USP) should be
sufficient to guarantee purging the waste lines of air. Note: The purge volume can
be checked by observing the markings on the pump syringe. Lift the waste bottle
shield cover and look for any signs of leakage. Correct any leaking connections
before continuing to use the CardioGen-82 infusion system.
Note: It is recommended that leak testing be carried out by wipe test and radiation
detection survey.
5.10.3 Pressure Sensor Line
Pinch off the saline bag valve, disconnect the pressure sensor line from the pressure
sensor syringe, and place it in a container. Depress the purge switch until liquid
flows into the beaker. Reconnect the pressure sensor line to the pressure sensor
syringe. Now unpinch the saline valve.
5.10.4 Patient Administration Set Purging
Place the MODE switch in the PURGE GENERATOR TO PATIENT position.
Connect the patient administration set to a shielded 50 mL vial and insert a venting
needle. Continuously depress the PURGE switch until all the air has been expelled
from the patient administration set into the vial. A volume of 20 mL should be
sufficient to purge the patient line. The purge volume can be determined by
observing the pump syringe markings.
Lift the generator shield cover and look for any signs of leakage. Correct any
leaking connections before continuing to use the CardioGen-82 Infusion System.
Date of Issue: February - 2015 Page 53 of 88 Page 53 of 88
Note: It is recommended that leak testing be carried out by wipe test and radiation
detection survey.
5.10.5 Volumetric Flow Rate Verification
This is performed with the installation of a new generator.
Set the controls on the Display/Control Panel as follows:
• Mode Switch = Automatic Infusion
• Elution Volume Limit = 99 mL
• Patient Volume Limit = 50 mL
• Patient Dose Limit = 99 mCi
• Dose Rate Threshold = 1.0 mCi/sec.
• Flow Rate = 50 mL/min.
Check for a 50 mL/min. flow rate using a stop watch.
5.11 Patient Administration
PATIENT ADMINISTRATION MAY BE PERFORMED ONLY AFTER
SUCCESSFUL COMPLETION OF DAILY QUALITY CONTROL
PROCEDURES ( COLUMN WASH, SR-82/85 TESTING, AND
CALIBRATION).
THE INFUSION SYSTEM’S DAILY CALIBRATION MUST BE
PERFORMED USING THE SAME DOSE/ACTIVITY (mCi Rb-82) THAT
WILL BE USED FOR A PATIENT ADMINISTRATION DOSE..
IN THE EVENT OF A POWER-FAILURE OR THE SYSTEM IS
INADVERTENTLY SHUT DOWN, CALIBRATION SHOULD BE
RECONFIRMED.
5.11.1. Replace the patient administration set for each new patient.
5.11.2. Purge all air out of the patient administration set and verify that all air is purged
from the system. Refer to Installation Instructions, section 5.
5.11.3. Set the controls on the Display/Control panel as prescribed by the administering
physician.
Note: Patient dose must be at the same level of activity (mCi Rb-82) as was
used in the most recent calibration. If the level of activity of the patient dose
is different than the level of activity used for calibration, calibration must be
performed again at the new activity level.
Date of Issue: February - 2015 Page 54 of 88 Page 54 of 88
5.11.4. Verify that the syringe pump has been filled with saline (0.9% additive-free
sodium chloride injection, USP). The syringe pump volume must at least equal the
selected elution volume-set point plus 20 mL. This volume will cover the 15 mL
nominal dead volume in the syringe.
5.11.5. Verify that the system printer is on and that approximately 1 inch of paper extends
out of the printer. If the RS-232C port is being used for infusion data acquisition,
the system printer can be turned off. Infusion report data is transmitted out the
RS-232C port whether the system printer is powered or not.
5.11.6. Attach the CardioGen-82 Infusion System patient administration set to the
patient’s intravenous line.
5.11.7. Make sure that at least 10 minutes has elapsed following any purge or infusion
operation. Start the infusion by depressing the INJECT START/STOP switch.
NOTE: The infusion can be terminated at any time by depressing the INJECT
START/STOP switch a second time.
The running elution volume (mL), patient volume (mL), patient dose (mCi), and
dose rate (mCi/sec) will be registered on the control/display panel. Once the
infusion operation is completed the INJECT START/STOP switch will stop
glowing red and the infusion report data will be printed on the system printer and
echoed out the RS-232C port.
5.11.8. Unless manually interrupted, the infusion will terminate when either a) the patient
dose limit (mCi), b) the patient volume limit (mL), or c) the total elution volume
limit (mL) has been reached, whichever comes first.
5.11.9. If the infusion system repetitively terminates patient infusions based on the
reaching of patient or elution volume limits, particularly if the upper limit of 99
mL is being reached, then the user should verify that the generator has not reached
Expiration Limits.
NOTE: THE CardioGen-82 GENERATOR ELUATE IS RADIOACTIVE
AND SHOULD BE HANDLED WITH PROPER RADIATION
SAFETY PRECAUTIONS.
NOTE: WAIT AT LEAST 10 MINUTES AFTER ANY PURGE OR
INFUSION OPERATION BEFORE LIFTING THE GENERATOR
SHIELD COVER, WASTE BOTTLE SHIELD COVER, OR THE
VALVE-SHIELD COVER. THIS WILL ALLOW ANY RB-82
ELUATE TO DECAY TO A SAFE LEVEL.
NOTE: THE VALVE-SHIELD COVER AUTOMATICALLY LOCKS IN
THE CLOSED POSITION DURING ANY PURGE OR INFUSION
OPERATION. THE COVER REMAINS LOCKED UNTIL THE
DETECTED RADIOACTIVITY DECAYS TO A SAFE LEVEL. NO
PURGE OR INFUSION OPERATION CAN BE STARTED UNLESS
THE COVER IS CLOSED. ADDITIONALLY, THE COVER
Date of Issue: February - 2015 Page 55 of 88 Page 55 of 88
CANNOT BE RAISED UNLESS THE CARDIOGEN-82 INFUSION
SYSTEM IS POWERED.
NOTE: ALWAYS WEAR GLOVES BEFORE TOUCHING ANY OF THE
TUBING, SYRINGE PUMP, VALVE, OR COLLECTION BOTTLE
SYSTEM COMPONENTS.
Date of Issue: February - 2015 Page 56 of 88 Page 56 of 88
Figure 5.1 Tubing Diagram
Date of Issue: February - 2015 Page 57 of 88 Page 57 of 88
CardioGen-82 (Rubidium 82) Generator - Sr-82/85 Testing Instructional Worksheet
Use an ionization chamber-type dose calibrator for eluate testing.
Daily, before administering rubidium Rb 82 chloride injection to any patient,
perform an eluate testing to determine Rb-82, Sr-82, and Sr-85 levels
Perform additional daily eluate tests after detecting any of the following Alert
Limits:
o 14 L total elution volume has passed through the generator column, or
o Sr-82 level reaches 0.002 µCi per mCi Rb-82, or
o Sr-85 level reaches 0.02 µCi per mCi Rb-82.
Perform the additional daily eluate tests at time points determined by the day’s
elution volume; tests are performed every 750 mL.
o For example, if an Alert Limit were reached and the clinical site eluted less than
750 mL from the generator during the day, then no additional eluate tests would
have been performed that day.
o If the same clinical site the next day eluted 1,500 mL from the generator, then the
site would have performed three tests that day: 1) the required daily test that
precedes any patient dosing, 2) a test at the 750 mL elution point, and 3) a test at
the 1,500 mL elution point.
o If a generator’s Alert Limit is reached, the clinical site performs the additional
daily tests (at intervals of 750 mL) each subsequent day the generator is used. The
additional tests are necessary to promptly detect excessive Sr-82 and/or Sr-85 in
eluates.
Date _________________
Infusion System Control Panel Settings
Mode Switch: Automatic Infusion
Elution Volume: 99 mL
Patient Volume: 50 mL
Patient Dose: 99 mCi
Dose Rate: 1 mCi/sec.
Flow Rate: 50 mL/min.
Dose Calibrator Setting
504 (Capintec only) or
Co-60 setting then divide reading obtained by 0.548
_______________________________________________________
Date of Issue: February - 2015 Page 58 of 88 Page 58 of 88
1. Elute the generator with 50 mL of Sodium Chloride Injection USP at a rate of 50
mL/min and collect the eluate in a 50 mL stoppered glass vial using the Infusion
System Control Panel Settings as above. Note the exact time at end of elution (EOE).
Time at end of elution (EOE)____________________________________.
2. Set dose calibrator as above. Setting used: ________________________.
3. Measure Rb-82 activity in the dose calibrator. Note the exact time (minutes/seconds)
when the measurement is made.
Rb-82 activity ________________________________________(mCi).
Time of measurement ______________________________________.
4. To account for radioactive decay, correct the Rb-82 measurement back to the EOE
using the appropriate decay correction factor (called “fraction remaining” in the
table) obtained from Table 1 below:
TABLE 1
Physical Decay Chart: Rb-82 half-life 75 seconds
Seconds* Fraction
Remaining
Seconds* Fraction
Remaining
0 1.000 165 0.218
15 0.871 180 0.190
30 0.758 195 0.165
45 0.660 210 0.144
60 0.574 225 0.125
75 0.500 240 0.109
90 0.435 255 0.095
105 0.379 270 0.083
120 0.330 285 0.072
135 0.287 300 0.063
150 0.250
To calculate the Rb-82 activity at EOE, divide the measurement of Rb-82 activity by
the decay correction factor corresponding to the time delay (in seconds) between
EOE and measurement, as found in Table 1 above.
Decay correction factor_________________________________
Rb-82 activity at EOE (mCi)_____________________________
5. Using the same vial, allow the sample to stand for at least one hour to allow for the
complete decay of Rb-82.
6. Measure the residual activity of the sample in the dose calibrator. Set the instrument
to read in microcuries (μCi) and record the measurement.
Residual activity (μCi)_________________________________
Date of Issue: February - 2015 Page 59 of 88 Page 59 of 88
7. Calculate the ratio (R) of Sr-85/Sr-82 on the day (post-calibration) of measurement
using the ratio of Sr-85/Sr-82 on the day of calibration provided on the generator
label and the Sr-85/Sr-82 Ratio Factor from Table 2 below.
TABLE 2
Sr-85/Sr-82 Ratio Chart
Days Ratio Factor Days Ratio Factor Days Ratio Factor
0* 1.00 16 1.31 32 1.73
1 1.02 17 1.34 33 1.76
2 1.03 18 1.36 34 1.79
3 1.05 19 1.38 35 1.82
4 1.07 20 1.41 36 1.85
5 1.09 21 1.43 37 1.88
6 1.11 22 1.46 38 1.91
7 1.13 23 1.48 39 1.95
8 1.15 24 1.51 40 1.98
9 1.17 25 1.53 41 2.01
10 1.19 26 1.56 42 2.05
11 1.21 27 1.58
12 1.23 28 1.61
13 1.25 29 1.64
14 1.27 30 1.67
15 1.29 31 1.70
*Day of calibration
Determine R using the following equation:
[Sr-85]
R = ———— on calibration date X Ratio Factor on the day
[Sr-82] (post-calibration) of measurement
8. Use a correction factor (F) of 0.478 to compensate for the contribution of Sr-85 to
the reading.
9. Calculate the amount of Sr-82 in the sample using the following equation:
dose calibrator reading (μCi)
Sr-82 (μCi) = —————————————
[1 + (R) (F)]
Example: dose calibrator reading (μCi) = 0.8; Sr85/Sr82 ratio (R) =1.48; correction
factor (F) = 0.478.
Date of Issue: February - 2015 Page 60 of 88 Page 60 of 88
0.8
Sr-82 (μCi) = ———————— = 0.47
[1 + (1.48)(0.478)]
Alternatively, calculate the amount of Sr-82 in the sample using the divisor provided
by the manufacturer, as per the following equation:
Sr-82 (μCi) = ______dose calibrator reading (µCi)______
divisor (from sheet supplied with Generator)
Example: dose calibrator reading (μCi) = 0.8; divisor = 1.7
Sr-82 (μCi) = ___0.8___ = 0.47
1.7
10. Determine if Sr-82 level in the eluate exceeds an Alert or Expiration Limit by
dividing the µCi of Sr-82 by the mCi of Rb-82 at EOE. (See below for further
instructions based on the Sr-82 level.)
Example:
0.47 µCi Sr-82 = 0.0094 µCi Sr-82/mCi Rb-82
50 mCi Rb-82
(test result is above Alert limit of 0.002 µCi Sr-82/mCi Rb-82; additional daily eluate
testing must be performed)
11. Determine if Sr-85 level in the eluate exceeds an Alert or Expiration Limit by
multiplying the result obtained in step 10 by the Sr-85/Sr-82 ratio (R), either as
calculated in step 7 above, or as provided in the data sheet supplied with the
generator.
Example:
0.0094 x 1.48 = 0.014 µCi Sr-85/mCi Rb-82
(test result is below the Alert limit of 0.02 µCi Sr-85/mCi Rb-82; result is within
specification for once daily testing)
Stop use of the CardioGen-82 generator once any one of the following Expiration
Limits is reached.
A total elution volume of 17 L has passed through the generator column, or
42 days post calibration date, or
An eluate Sr-82 level of 0.01 μCi /mCi Rb-82, or
An eluate Sr-85 level of 0.1 μCi /mCi Rb-82.
Date of Issue: February - 2015 Page 61 of 88 Page 61 of 88
CardioGen-82 INFUSION SYSTEM-CALIBRATION DATA SHEET
Infusion System Control Panel Settings Mode Switch: Automatic infusion Elution Volume: 99 mL Patient Volume: 20-50 mL
Patient Dose: Calibration activities must be performed with the SAME
LEVEL OF ACTIVITY (mCi Rb-82) that will be used for
patient doses. Dose Rate: 1 mCi/sec. Flow Rate: 50 mL/min. 1. Date of Calibration or Calibration Verification: ______________________ 2. Initial Generator Sr-82 Potency on calibration date: _______________mCi 3. Generator Sr-82 Potency on present date: ______________________mCi 4. Present Calibration Factor: _____________________________________ 5. Measured Rb-82 Activity from dose calibrator (corrected for decay):
___________________mCi 6. Printed Rb-82 Activity Present at the end of infusion: _______________mCi 7. Ratio of Measured Rb-82 Activity and Printed Rb-82: Measured Rb-82 Activity Printed Rb-82 Activity = Ratio 8. Is the difference between Measured Rb-82 Activity and Printed Rb-82:
+/-5% (Ratio=0.95 to 1.05) for initial calibration (generator installation) or +-/10% (Ratio=0.9 to 1.1) daily calibration? ____________________ (If no, unit requires recalibration).
9. Calculated New Calibration Factor (if needed): ______________________ Measured Rb-82 Activity at End of New Initial Infusion ___________________ Cal. = Cal. ________x __________________________ Factor Factor Printed Rb-82 Activity Present at
End of Infusion ______________ .
Date of Issue: February - 2015 Page 62 of 88
Figure 5.2 SR-82 / SR-85 Testing Worksheet
Date of Issue: February - 2015 Page 63 of 88
Figure 5.3 Rb-82 Infusion System Calibration Log Sheet
Date of Issue: February - 2015 Page 64 of 88
Date of Issue: February - 2015 Page 65 of 88
6. TROUBLESHOOTING GUIDE
The table below describes possible purge and infusion error conditions that are
displayed on the Control/Display panel. These errors are also listed in the infusion
report printout. Suggested corrective action is provided in the table for each error
condition.
For product complaints or questions regarding the operation and service of the
system, please contact your assigned Bracco Representative.
This device has been tested for protection against electro-magnetic interference
according to IEC 60601-1-2. However, the proximity of other devices that give off
electro-magnetic radiation, such as X-ray machines, can interfere with the operation
of this system.
Date of Issue: February - 2015 Page 66 of 88
CardioGen-82®
(Rubidium Rb 82 Generator) Operator’s Manual Trouble Shooting Guide Index
Errors / Illustrations Page
Infusion System Diagram 67
Automatic Infusion Errors 68
Elution Volume (Illustration) 69
Patient Volume (Illustration) 70
Patient Dose (Illustration) 71
Pump Limit (Illustration) 72-73
Thumbwheel (Illustration) 74-75
Automatic Infusion and Purge Errors 76
Valve Failure (Illustration) 77-78
Valve Shield Open (Illustration) 79-80
High Pressure Errors 81
High Pressure Errors (Illustration) 82-83
Infusion Calibration Issues 84-85
Thumb Screws (Illustration) 86-87
Date of Issue: February - 2015 Page 67 of 88
CardioGen-82 Infusion System Diagram
Date of Issue: February - 2015 Page 68 of 88
Automatic Infusion Errors
Error Condition
Potential Root
Cause Indicative Signs of Error Suggested Action Illustration
Elution Volume
Elution Volume Set-
Point = 0
Print out will read "INFUSION NOT STARTED INFUSION PARAMETER ERROR Elution Vol. setpt. =
0"
Select non-zero infusion set points
See "Elution Volume" pg.
69
Patient Volume
Patient Volume Set-
Point = 0
Print out will read "INFUSION NOT STARTED INFUSION PARAMETER ERROR Pt. Vol. setpt. = 0"
Select non-zero infusion set points
See "Patient Volume" pg.
70
Patient Dose Patient Dose Set-Point =
0
Print out will read "INFUSION NOT STARTED INFUSION PARAMETER ERROR Pt.Dose setpt. = 0"
Select non-zero infusion set points
See "Patient Dose" pg. 71
Pump Limit
Pump At Extreme Forward
Limit Position
With Minimal Saline in Syringe
Print out will read "INFUSION UNABLE TO START Pump at Limit POS" and yellow pump limit light is
illuminated Refill syringe pump
See "Pump Limit" pg. 72-
73
Thumbwheel Oxidation on thumbwheel connections
Elution Volume, Patient Volume, Patient Dose, and/or Dose Rate on control panel will not match up with the
tape print out.
Cycle through a full range of numbers to clear
contacts of oxidation. Perform two elutions, one at a 35 ml and one at 50 ml for the parameter with
the issue, i.e. elution volume, and compare to
tape print out for accuracy. Contact Clinical
Applications Specialist (CAS).
See “Thumbwheel”
pg. 74-75
Date of Issue: February - 2015 Page 69 of 88
Elution Volume Print Out
Date of Issue: February - 2015 Page 70 of 88
Patient Volume Print Out
Date of Issue: February - 2015 Page 71 of 88
Patient Dose Print Out
Date of Issue: February - 2015 Page 72 of 88
Pump Limit Control Panel
NOTE: Pump limit light will be illuminated when the syringe is full or nearly empty. When the system is set up correctly for an infusion, the pump limit light will be illuminated. This error will occur when there is not enough saline in the syringe to complete an infusion.
Date of Issue: February - 2015 Page 73 of 88
gggg
Pump Limit Print Out
Date of Issue: February - 2015 Page 74 of 88
Thumbwheel Error Control Panel
NOTE: In this example the Patient Volume Limit is set for 35 mL but the printout on the next page shows 15 mL is the set point value and what the system actually eluted at patient volume
Date of Issue: February - 2015 Page 75 of 88
Thumbwheel Error Print Out
Date of Issue: February - 2015 Page 76 of 88
Automatic Infusion and Purge Errors
Error Condition
Potential Root Cause
Indicative Signs of Error Suggested Action Illustration
Valve Failure Tight Divergence Valve or Loss of
Mechanism
Print out will read "INFUSION UNABLE TO START Valve Error" and Valve Failure light on
the control panel display is illuminated
Remove divergence valve and gently rotate lever-arm on tubing set before placing back in original position. If this process fails, contact
CAS.
See "Valve Failure" pg. 77-78
Valve Shield Open
Valve Shield Open Print out will read "INFUSION UNABLE TO
START Pt. Valve Shield Open"
Close the valve shield for all purge and
infusion operations. Make sure the inlet and patient port are properly inserted into the tubing
channel and not preventing the shield from completely
closing.
See "Valve Shield Open" pg. 79-80
Date of Issue: February - 2015 Page 77 of 88
Valve Failure Control Panel
Date of Issue: February - 2015 Page 78 of 88
Valve Failure Print Out
Date of Issue: February - 2015 Page 79 of 88
Valve Shield Open
NOTE: This error is activated even if valve shield is slightly open and not depressing the red switch button.
Date of Issue: February - 2015 Page 80 of 88
Valve Shield Open Print Out
Date of Issue: February - 2015 Page 81 of 88
High Pressure Errors
Error Condition
Potential Root Cause
Indicative Signs of Error Suggested Action Illustration
High Pressure
Restricted tubing
Print out will read "INFUSION ABORTED BY ERROR High Pressure Error" and Red High Pressure light on the control panel display is illuminated.
Ensure all tubing is free of kinks and is not pinched in the system
See "High Pressure" pg.
82-83
High Pressure
Inadequate patient IV access
See Reference 6
Print out will read "INFUSION ABORTED BY ERROR High Pressure Error" and Red High Pressure light on the control panel display is illuminated.
Make sure gauge of needle is equal to or greater than 20 gauge.
Make sure stopcock or tubing clip is not locked. Position IV, check for flow in catheter.
Make sure patient’s arm is not bent in such a way as to restrict flow.
See "High Pressure" pg.
82-83
High Pressure
Faulty tubing, filter, or other
accessory pack or PAS component
Print out will read "INFUSION ABORTED BY ERROR High Pressure Error" and Red High Pressure light on the control panel display is illuminated.
Replace accessories as deemed necessary. The CAS or Field Service Engineer will decide if
customer should bypass the generator
See "High Pressure" pg.
82-83
High Pressure
Vent needle not fully pierced
through the elution test vial
septum during testing
Print out will read "INFUSION ABORTED BY ERROR High Pressure Error" and Red High Pressure light on the control panel display is illuminated.
Make sure proper positioning of vent needle and the patient administration set (PAS) needles are
inserted fully into elution test vial septum
See "High Pressure" pg.
82-83
High Pressure
Waste bottle cap filter failure or
clogging
Print out will read "INFUSION ABORTED BY ERROR High Pressure Error" and Red High Pressure light on the control panel display is illuminated.
Replace waste bottle cap NOTE: New waste bottle cap should be installed
with each new generator
See "High Pressure" pg.
82-83
High Pressure
Improper flow rate setting
(80ml/min instead of 50ml/min)
Print out will read "INFUSION ABORTED BY ERROR High Pressure Error" and Red High Pressure light on the control panel display is illuminated.
Flow rate setting should always be set to 50ml/min in accordance with Prescribing
Information (PI)
See "High Pressure" pg.
82-83
High Pressure
Generator column failure/blockage
Print out will read "INFUSION ABORTED BY ERROR High Pressure Error" and Red High Pressure light on the control panel display is illuminated.
Run through all other troubleshooting potential causes before bypassing generator. NOTE:
Customers must contact CAS prior to bypassing the generator.
See "High Pressure" pg.
82-83
Date of Issue: February - 2015 Page 82 of 88
High Pressure Control Panel
Date of Issue: February - 2015 Page 83 of 88
High Pressure Print Out
Date of Issue: February - 2015 Page 84 of 88
Infusion System Calibration Issues NOTE: Normal calibration QC may take several infusions before being within acceptable limits when a new generator is installed.
Error Condition
Potential Root Cause
Indicative Signs of Error Suggested Action Illustration
Infusion System Calibration
Issues
Pump syringe is not secure to pump
syringe tower (chatter)
Multiple attempts to calibrate the system and/or large swings in the calibration factor
Ensure all four (4) thumb screws are secure See "Thumb
Screws" pg. 86 & 87
Infusion System Calibration
Issues System Leak
Multiple attempts to calibrate the system and/or large swings in the calibration factor
Lower than normal calibration factor: observed leak on the inlet side of the
generator Higher than normal calibration factor: observed leak on the outlet side of the
generator
Infusion System Calibration
Issues
Timing inconsistency in
dose assay
Multiple attempts to calibrate the system and/or large swings in the calibration factor
Make sure assays are performed consistently at 3 min. 45 sec. by all end-
users. Ensure the timer is started immediately at the end of infusion (when the light goes
out).
Infusion System Calibration
Issues
Incorrect channel selected on dose calibrator when assaying Rb-82
eluate for CardioGen-82 QC
Multiple attempts to calibrate the system and/or large swings in the calibration factor
Determine if the dose calibrator has an Rb-82 pre-set channel: Program dose
calibrator for Rb-82 as recommended by the manufacturer, Biodex: Atomlab 100 =
8.8 ; Atomlab 400/500 = 8.7, Capintec: 504 Check user manual for setting specifics,
The facility must have a dose calibrator capable of reading microcuries (uCi) out to
2 decimal places
Date of Issue: February - 2015 Page 85 of 88
Infusion System Calibration Issues Continued NOTE: Normal calibration QC may take several infusions before being within acceptable limits when a new generator is installed.
Error Condition
Potential Root Cause Indicative Signs of
Error Suggested Action Illustration
Infusion System Calibration
Issues
Dose calibrator non-linearity: Assaying the dose at 2min 30sec on a Capintec dose calibrator when the
dose is between 12-15mCi
Multiple attempts to calibrate the system and/or
large swings in the calibration factor
Measure assays consistently at 3 min 45 sec
Infusion System Calibration
Issues Equipment power surge
Multiple attempts to calibrate the system and/or
large swings in the calibration factor
Equipment power surge can effect dose accuracy of dose calibrator and infusion system
Notify CAS.
Infusion System Calibration
Issues
High background in hot lab when assaying Rb-82 elution vials for QC
Multiple attempts to calibrate the system and/or
large swings in the calibration factor
Reduce background to a minimum. Always make sure
background is subtracted when performing level testing
Infusion System Calibration
Issues
Possible control panel toggle switch setting contact issues: Oxidation of contacts, especially when the
infusion system is not used for an extended period of time
Multiple attempts to calibrate the system and/or
large swings in the calibration factor
After observing differences between the Infusion System Print-out Set Point Values and actual Control Panel Setting
values, notify CAS.
Infusion System Calibration
Issues
A site unknowingly tries to be within +/- 5% threshold for daily calibration, instead of the +/- 10% threshold
Multiple attempts to calibrate the system and/or
large swings in the calibration factor
Instruct the user that they only have to achieve +/- 5%
threshold on day one and +/-10% thereafter
Date of Issue: February - 2015 Page 86 of 88
Thumb Screws
Date of Issue: February - 2015 Page 87 of 88
Thumb Screws
Date of Issue: February - 2015 Page 88 of 88
7. CONTACT INFORMATION
Bracco Diagnostics Inc. 259 Prospect Plains Road, Monroe Township, NJ 08831 USA
1-609-514-2200
1-800-631-5245
Bracco Nuclear Medicine Customer Service 1-800-447-6883
Option 1
Technical Support
Service, Parts, and Technical Support; Bracco Professional
Services
1-800-257-5181
Option 2
8. CARDIOGEN-82® Infusion System Supplies
Catalog #
Description: CARDIOGEN-82®
0270-0015-00
CARDIOGEN-82® (Rubidium Rb 82 Generator)
001575
Absorbent Pad
001510
Accessory Package
001536
Elution System
001530
Elution Test Vial
001550
Elution Vial Shield
001537
Infusion System
001565
Infusion System Thermo Paper (Case of 12)
001506
Patient Administration Sets (25 sets)
001511
Vented Cap for Waste Bottle
001520
Waste Bottle
ECN 14-1212-1 Rev. 23