Page 1 of 15
CHECKLIST OF EQUPMENTS
Sl. NoTS
EQUIPMENT DESCRIPTIONAs per contract Available Calibration Due date of Shortfall/
Remarks
S/No requirement in lab done calibration excess
1 1/d Pan balance with full set of weight capacity 5Kg 4 Set 1 Set Yes 2/4/2010 3 Set
2 1/b 600 gm capacity semi- indicating electronic type ,2 Set 1 Set Yes 2/4/2009 1 Set
2Kg/ electronic avail
accuracy:0.01gms acc:0.01gms
3 1/a 7 to 10 Kg capacitty semi- indicating electronic type2 Set 1 Set Yes 2/4/2009 1 Set
30Kg/ electronic avail
accuracy:1 gm acc:0.1gms
4 1/c Chemical balance 100 gms capacity1 Set _ _ _ 1 Set
Not available in
accuracy:0.0001gms laboratory
5 2/a Ovens electrically operated themostatically controlled
1 Set 1 Set Yes – –
Spot calibration done
including thermometer stainless steel interior during inspection
range ambient to 300ºCsensitivity 1ºC,capacity:120 ltr
6 3 Sieves as per IS 2386/PART1
Not specified 1 Set Yes – –
Spot calibration done
3/a Standard test size 450 mm dia complete with lid and pan during inspection
of sieve size 40,25,20,16,12.5,10,4.75,2.36.mm
3/c Standard test size 200 mm dia complete with lid and pan1 Set Yes – _
Spot calibration done
during inspection
7 4 Complete of taking 200 mm and 450 mm dia sieves
1 Nos 1 Nos – – –
Spot inspection done
electrically operated with time switch assembly as per IS 1440 RPM
specification
8 5 200 tonnes compressive testing machine electrically cum
1 Nos 1 Nos Yes 19/12/2008 _
Calibration report
manually operated fitted with 3 dial gauges attached
0-2000X10 KN;0-1000X5KN;0-500X2.5KN
7 8/d Circular plates of 200 mm dia 6 Nos 8 Nos Yes – – Spot calib done
8 52 Vicat's needle apparatus for setting time with plunger 1 Set 1Set Yes _ _
Spot calibration done
as per IS 269-1968 during inspection
9 6/a Soundness testing apparatus for cement 1 Set 3 Set Yes – 2 Set Spot calib done
10 53/b 150 mm cube mould and 100 mm (each size) as per IS Spot calibration done
150mm cube mould with base plate 36 Nos 36 Nos Yes – during inspection
100mm cube mould with base plate 36 Nos 06 Nos Yes – 30 Nos do
70.6mm cube mould with base plate – 12 Nos Yes – 12 Nos do
50mm cube mould with base plate – 09 Nos Yes – 09 Nos do
11 54 High frequency mortar cube vibrator for cement testing 1 Nos 1 Nos – – – Spot inspection done
12 60 LOS ANGELES apparatus as per IS 2386/PART IV 1 Nos – – – – Not available in lab
13 62 Equipment for slump test (C-143) compacting factor 4 Set 4 Set Yes – –
Slump cone avail
apparatus complete Spot inspection done
14 Electrically operated and thermostatically controlled1 Nos 1 Nos No – –
Calibration not
waterbath, sensitivity 1ºC done
15Steel trowels
Not specified2 Nos – – –
Not RequiredSmall
Medium 5 Nos – – –
Big 1 Nos – – –
of sieve size 4.75,2.36,1.18,1mm,600,300,250,150,125,90, 75mic.
Page 2 of 15
Sl. NoTS
EQUIPMENT DESCRIPTIONAs per contract Available Calibration Due date of Shortfall/
Remarks
S/No requirement in lab done calibration excess
Contractor's Representative Consultant's Representative
16 12 Spatula set of 100mm and 200 mm long
2 Nos – – – 2 Nos Not Required100mm
200mm 2 Nos 2 Nos – – –
17 40/b Glass type (mercury in glass) themometer in range of 2 Nos 1 Nos – – 1 Nos Not Required
0-110ºC
18 Digital thermometer Not specified 1 Nos Yes – 1 Nos Not Required
19 Stop watch Not specified 1 Nos – – 1 Nos Not Required
20 55 Concrete mixer power driven ,capacity 1cft(tilting) 1 Nos 1 Nos Yes – – Not Required
21 33 Aggregate impact value testing apparatus as per1 Nos 1 Nos Yes – –
Spot calibration done
IS :2386/PART IV-1964 during inspection
22 31 Flakiness and elongation test gauges as per IS 23862 Nos 1 Nos Yes – –
Spot calibration done
PART 1 -1963 during inspection
23 34/a Specific gravity determination test apparatus 2 Set 2 Set – – – Not Required
with pycnometer bottle
34/b A wire busket of not more than 6.3 mm mesh 1 Nos 1 Nos – – – Spot inspection done
24 7 Hot plate 200 mm dia-1500 watt 1 Nos 1 Nos Not Required – – Spot inspection done
2532 Standard measure of 30,15,3 ltr capacity
2 Sets 2 Sets Yes
– –
3ltr-150mm dia/170mm ht/3.15mm thk
15ltr-250mm dia/300 mm ht/4.00mm thk
30 ltr-350mm dia/300mm ht/5.00mm thk
26 6 Glass measuring cylinder-BOROSIL
Not specified Not Required – Not Required
100ml 1 Nos
250ml 1 Nos
500ml 1 Nos
1000ml 1 Nos
Glass beaker
Not specified 2 Nos – Not Required200ml
250ml –
Plastic measuring cylinder
Not specified Not Required Not Required
100ml 1 Nos –
200ml –
250ml 1 Nos –
500ml 2 Nos –
1000ml 1 Nos –
27 8 Enamel tray
Not specified Not Required Not Requireda)600x450x50mm 1 Nos –
b)450x300x40mm 1 Nos –
c)300x250x40mm 1 Nos –
28 Aluminium tray
Not specified Not Required Not Required600x600x50mm 1 Nos –
900x600x50mm 1 Nos –
300x250x40mm 1 Nos –
29 Hydrometer(sp gr)- range 1000 to 1200 Not specified 1 Nos Not Required – Not Required
Calibration done in Laboratory
Page 3 of 15
Sl. NoTS
EQUIPMENT DESCRIPTIONAs per contract Available Calibration Due date of Shortfall/
Remarks
S/No requirement in lab done calibration excess
Contractor's Representative Consultant's Representative
CALIBRATION OF BULK DANSITY MEASURING CYLINDER
Lab. Calibration No : Date: 16/04/2008
Testing Location : Ch: 1030+700
A. Capacity of Measuring Cylinder : 3 ltr.
A(a) Cylinder Nos : 01
Sl. No Description Unit Trial-01 Trial-02 Trial-03 Remarks
1 Cylinder Weight Kg 2.728 2.729 2.729
2 Cylinder Weight+Water Kg 5.731 5.731 5.73
3 Weight of Water Kg 3.002 3.002 3.001
4 Mean Kg 3.002
A(b) Cylinder Nos : 02
Sl. No Description Unit Trial-01 Trial-02 Trial-03 Remarks
1 Cylinder Weight Kg 2.701 2.701 2.7
2 Cylinder Weight+Water Kg 5.69 5.691 5.689
3 Weight of Water Kg 2.989 2.99 2.989
4 Mean Kg 2.989
B. Capacity of Measuring Cylinder : 15 ltr.
B(a) Cylinder Nos : 01
Sl. No Description Unit Trial-01 Trial-02 Trial-03 Remarks
1 Cylinder Weight Kg 8.736 8.736 8.736
2 Cylinder Weight+Water Kg 23.388 23.389 23.388
3 Weight of Water Kg 14.652 14.653 14.652
4 Mean Kg 14.652
B(b) Cylinder Nos : 01
Sl. No Description Unit Trial-01 Trial-02 Trial-03 Remarks
1 Cylinder Weight Kg 9.362 9.362 9.362
2 Cylinder Weight+Water Kg 24.057 24.058 24.058
3 Weight of Water Kg 14.696 14.696 14.696
4 Mean Kg 14.696
Contractor's Representative Consultant's Representative
I.D.=0.150 mtr. HT.=0.170 mtr.
I.D.=0.150 mtr. HT.=0.170 mtr.
I.D.=0.250 mtr. HT.=0.300 mtr.
I.D.=0.250 mtr. HT.=0.300 mtr.
SPOT CALIBRATION OF CUBE MOULD (SIZE-150x150x150)
Lab. Calibration No : Date: 16/04/2008Testing Location : Ch: 1030+700A . Quantity = 36 nos
Sl. No Description UnitDimension
RemarksL B H
1 Cube Mould MM 150 150 150
2 -do- MM 150 150 150
3 -do- MM 150 150 150
4 -do- MM 150 150 150
5 -do- MM 150 150 150
6 -do- MM 150 150 150
7 -do- MM 150 150 150
8 -do- MM 150 150 150
9 -do- MM 150 150 150
10 -do- MM 150 150 150
11 -do- MM 150 150 150
12 -do- MM 150 150 150
13 -do- MM 150 150 150
14 -do- MM 150 150 150
15 -do- MM 150 150 150
16 -do- MM 150 150 150
17 -do- MM 150 150 150
18 -do- MM 150 150 150
19 -do- MM 150 150 150
20 -do- MM 150 150 150
21 -do- MM 150 150 150
22 -do- MM 150 150 150
23 -do- MM 150 150 150
24 -do- MM 150 150 150
25 -do- MM 150 150 150
26 -do- MM 150 150 150
27 -do- MM 150 150 150
28 -do- MM 150 150 150
29 -do- MM 150 150 150
30 -do- MM 150 150 150
31 -do- MM 150 150 150
32 -do- MM 150 150 150
33 -do- MM 150 150 150
34 -do- MM 150 150 150
35 -do- MM 150 150 150
36 -do- MM 150 150 150
Contractor's Representative Consultant's Representative
SPOT CALIBRATION OF CUBE MOULD
Lab. Calibration No : Date: 16/04/2008Testing Location : Ch: 1030+700
A . CUBE MOULD SIZE- 100x100x100Quantity = 6 nos
Sl. No Description UnitDimension
RemarksL B H
1 Cube Mould MM 100 100 100
2 -do- MM 100 100 100
3 -do- MM 100 100 100
4 -do- MM 100 100 100
5 -do- MM 100 100 100
6 -do- MM 100 100 100
B CUBE MOULD SIZE- 70.6x70.6x70.6Quantity = 12 nos
Sl. No Description UnitDimension Remarks
L B H
1 Cube Mould MM 70.6 70.6 70.6
2 -do- MM 70.6 70.6 70.6
3 -do- MM 70.6 70.6 70.6
4 -do- MM 70.6 70.6 70.6
5 -do- MM 70.6 70.6 70.6
6 -do- MM 70.6 70.6 70.6
7 -do- MM 70.6 70.6 70.6
8 -do- MM 70.6 70.6 70.6
9 -do- MM 70.6 70.6 70.6
10 -do- MM 70.6 70.6 70.6
11 -do- MM 70.6 70.6 70.6
12 -do- MM 70.6 70.6 70.6
Contractor's Representative Consultant's Representative
SPOT CALIBRATION OF CUBE MOULD
Lab. Calibration No : Date: 16/04/2008Testing Location : Ch: 1030+700
A . CUBE MOULD SIZE- 50x50x50
Quantity = 9 nos
Sl. No Description UnitDimension
RemarksL B H
1 Cube Mould MM 50 50 50
2 -do- MM 50 50 50
3 -do- MM 50 50 50
4 -do- MM 50 50 50
5 -do- MM 50 50 50
6 -do- MM 50 50 50
7 -do- MM 50 50 50
8 -do- MM 50 50 50
9 -do- MM 50 50 50
Contractor's Representative Consultant's Representative
SPOT CALIBRATION OF SIEVE
Lab. Calibration No : Date: 16/04/2008Testing Location : Ch: 1030+700
A SIEVE SIZE- 450 MM DIA COMPLETE WITH LID
Quantity = 1 set
Sl. No Description UnitDimension
RemarksL B H
1 Sieve Size 40 mm MM 40 40
2 Sieve Size 25 mm MM 25 25
3 Sieve Size 20 mm MM 20 20
4 Sieve Size 16 mm MM 16 16
5 Sieve Size 12.5 mm MM 12.5 12.5
6 Sieve Size 10 mm MM 10 10
7 Sieve Size 4.75 mm MM 4.75 4.75
8 Sieve Size 2.36 mm MM 2.36 2.36
B SIEVE SIZE- 250 MM DIA COMPLETE WITH LID
Quantity=1 set
Sl. No Description UnitDimension
RemarksL B H
1 Sieve Size 4.75 mm MM 4.75 4.75
2 Sieve Size 2.36 mm MM 2.36 2.36
3 Sieve Size 1.18 mm MM 1.18 1.18
4 Sieve Size 1 mm MM 1 1
5 Sieve Size 600 mic. MIC. 600 600
6 Sieve Size 300 mic MIC. 300 300
7 Sieve Size 150 mic MIC. 150 150
Contractor's Representative Consultant's Representative
SPOT CALIBRATION OF TRAY (ENAMAL/ALUMINIUM)
Lab. Calibration No : Date: 16/04/2008Testing Location : Ch: 1030+700A . Quantity = 2 set
Sl. No Description UnitDimension
RemarksL B H
ENAMAL TRAY
1 a) 600x450x50 MM 600 450 50
2 b) 450x300x40 MM 450 300 40
3 c) 300x250x40 MM 300 250 40
ALLUMINIUM TRAY
1 a) 600x600x50 MM 600 600 50
2 b) 900x600x50 MM 900 600 50
3 c) 300x250x40 MM 300 250 40
Contractor's Representative Consultant's Representative
SPOT INSPECTION REPORT
Lab. Calibration No:Location: 1030+700Quantity: 1DATE: 16-04-08
Name of equipment: ELECTRICALLY OPERATED OVEN
BRAND MODEL SL. NO REMARKS
HOT AIR OVEN AMBRO
CALIBRATION OF ELECTRICALLY OPERATED OVEN
REMARKS:
MACHINE WORKING IN GOOD CONDITION
100ºC 100ºC
110ºC 110ºC
150ºC 150ºC
200ºC 200ºC
CONTRACTOR'S REPRESENTATIVE CONSULTANT'S REPRESENTATIVE
TEMPERATURE OF OVEN
TEMPERATURE OF THERMOMETER
50ºC 50ºC
SPOT INSPECTION REPORT
Lab. Calibration No:Location: 1030+700Quantity: 1Date: 16-04-08
Name of equipment: VICATS APPRATUS
SL. NO DESCRIPTION OF PARTS DIMENSION OF PARTS REMARKS:
1 MOULD 80 MM DIA 40 MM HEIGHT
DIMENSIONS ARE AS PER IS2 PLUNGER 10 MM DIA 50 MM HEIGHT
3 NEEDLE
CONTRACTOR'S REPRESENTATIVE CONSULTANT'S REPRESENTATIVE
1 MM²
SPOT INSPECTION REPORT
Lab. Calibration No:Location: 1030+700Quantity: 1Date: 16-04-08
Name of equipment: MORTAR CUBE VIBRATOR FOR CEMENT TESTING
BRAND MODEL SL. NO REMARKS
PRITEC AC INDUCTION MOTOR S-2007
CONTRACTOR'S REPRESENTATIVE CONSULTANT'S REPRESENTATIVE
MACHINE WORKING IN GOOD CONDITION
SPOT INSPECTION REPORT
Lab. Calibration No:Location: 1030+700Quantity: 1DATE: 16-04-08
Name of equipment: SLUMP CONE
SL. NO DESCRIPTION OF PARTS DIMENSION OF PARTS (MM) REMARKS
1 TOP DIA 100
2 BOTTOM DIA 200
3 HEIGHT 300
4 SLUMP ROD 610 16 MM DIA
CONTRACTOR'S REPRESENTATIVE CONSULTANT'S REPRESENTATIVE
DIMENSIONS ARE AS PER IS
SPOT INSPECTION REPORT
Lab. Calibration No:Location: 1030+700Quantity: 1DATE: 16-04-08
Name of equipment: AGGREGRATE IMPACT VALUE TESTING MACHINE
SL. NO DESCRIPTION OF PARTS A/U DIMENSION OF PARTS REMARKS
1 TOTAL WEIGHT KG 60
2 WEIGHT OF HAMMER KG 14
3 HEIGHT OF FALL CM 38
CONTRACTOR'S REPRESENTATIVE CONSULTANT'S REPRESENTATIVE
DIMENSIONS ARE AS PER IS
SPOT INSPECTION REPORT
Lab. Calibration No:Location: 1030+700Quantity: 1Date: 16-04-08
Name of equipment: LE-CHATELIER APPRATUS
SL. NO DESCRIPTION OF PARTS A/U DIMENSION OF PARTS REMARKS
1 HEIGHT OF MOULD MM 30
DIMENSIONS ARE AS PER IS
2 DIA OF MOULD MM 30
3 THICKMESS OF MOULD MM 0.5
4 MM 165
CONTRACTOR'S REPRESENTATIVE CONSULTANT'S REPRESENTATIVE
CENTRE OF MOULD TO END OF POINTOR