CADD SOLIS FREQUENTLY ASKED QUESTIONS ENEXIA SPECIALTY PHARMACY 10/2019

Transcript of Cadd solis frequently asked questions - Enexia Specialty solis...Cadd solis frequently asked...

  • CADD SOLIS FREQUENTLY ASKED QUESTIONSENEXIA SPECIALTY PHARMACY 10/2019

  • TABLE OF CONTENTS

    CODES – SLIDE 3

    PRIMING TUBING – SLIDE 4

    CHANGING CASSETTES – SLIDE 11

    CHANGING DOSE – SLIDE 18

    REPROGRAMMING PUMP – SLIDE 25

  • CODES

    MOST PUMPS CAN BE UNLOCKED WITH THE

    FOLLOWING CODES: 513 OR 921.

    IF THE ABOVE TWO CODES DO NOT WORK,

    THEN CALL ENEXIA SPECIALTY PHARMACY

    TO ENTER A CODE, EACH OF THE 3 DIGITS

    MUST BE CHOSEN INDIVIDUALLY USING THE

    UP AND DOWN ARROWSTHEN PRESSING

    THE SELECT BUTTON WHICH MOVES THE

    PUMP TO THE NEXT DIGIT.

  • PRIMING TUBING

    IF PUMP IS NOT ALREADY STOPPED, HIT THE

    RED/GREEN “STOP/START” BUTTON

    “YES” TO STOP PUMP

  • PRIMING TUBING

    PRESS THE WHITE BUTTON BELOW “TASKS” ON

    THE BOTTOM RIGHT OF THE SCREEN

  • PRIMING TUBING

    USING THE UP AND DOWN ARROWS, SCROLL

    DOWN TO “PRIME TUBING” THEN PRESS THE

    “SELECT” BUTTON

  • PRIMING TUBING

    DISCONNECT TUBING FROM PATIENT. PRESS

    “PRIME”

    ASPTICALLY HOLD TUBING OVER SECURE

    IMPERMEABLE CONTAINER TO CATCH PRIMED

    LIQUID

  • PRIMING TUBING

    THE PUMP WILL SHOW VOLUME PRIMED IN

    REAL TIME

    ONCE ALL VISIBLE BUBBLES HAVE EXITED THE

    TUBING PRESS “STOP PRIMING”

  • PRIMING TUBING

    PRESS “NO” TO STOP PRIMING

  • PRIMING TUBING

    RECONNECT PUMP TUBING TO PATIENT’S LINE

    HIT THE RED/GREEN “STOP/START” BUTTON

    RESTART THE PUMP BY PRESSING “YES”

  • CHANGING A CASSETTE

    STOP THE PUMP BY PRESSING THE RED/GREEN

    STOP/START BUTTON ON THE FRONT OF THE

    PUMP. SEE “HOW TO” ON SLIDE 4, IF

    NECESSARY.

    DISCONNECT THE OLD CASSETTE TUBING

    FROM THE PATIENT AND HOOK UP TUBING

    FROM NEW CASSETTE

    UNLOCK THE CASSETTE BY TURNING THE

    ATTACHED KEY COUNTERCLOCKWISE AS

    INDICATED ON THE RIGHT SIDE OF THE PUMP

  • CHANGING A CASSETTE

    PULL THE METAL LATCH DOWN TO RELEASE

    THE CASSETE

  • CHANGING A CASSETTE

    ATTACH NEW CASSETTE TO PUMP BY

    ALIGNING THE HOOKS ON THE TOP LEFT OF

    CASSETTE TO LATCHES ON THE PUMP’S

    BOTTOM LEFT.

  • CHANGING A CASSETTE

    PUSH THE RIGHT SIDE OF CASSETTE UP AND

    SECURELY CLOSE THE LATCH

    LOCK THE CASSETTE WITH THE KEY BY

    TURNING IT CLOCKWISE. REMOVE THE KEY

    FROM SLOT

  • CHANGING A CASSETTE

    THE PUMP WILL AUTOMATICALLY ASK TO RESET

    RESERVOIR VOLUME TO 250ML, WHICH IS THE

    AMOUNT IN EACH “NEW” BLUE CASSETTE

    PRESS “YES” (RESEVOIR VOLUME ON TOP LEFT

    CORNER WILL RESET ACCORDINGLY)

  • CHANGING A CASSETTE

    ALL NEW CASSETTES ARE PRIMED DURING

    COMPOUNDING, SO YOU DO NOT HAVE TO

    PRIME NEW CASSETTES

    PRESS “NO” TO NOT PRIME

  • CHANGING A CASSETTE

    START THE PUMP AND YOU ARE DONE.

  • CHANGE BASAL RATE, BOLUS DOSE AND/OR LOCKOUT TIME

    STOP THE PUMP BY PRESSING THE RED/GREEN

    STOP/START BUTTON ON THE FRONT OF THE

    PUMP. SEE “HOW TO” ON SLIDE 4, IF NECESSARY.

    PRESS THE WHITE BUTTON BELOW “TASKS” ON

    THE BOTTOM RIGHT OF THE SCREEN. SEE “HOW

    TO” ON SLIDE 5, IF NECESSARY.

    USING THE UP AND DOWN ARROWS, SCROLL

    DOWN TO “VIEW DELIVERY SETTINGS” THEN

    PRESS THE “SELECT” BUTTON

  • CHANGE BASAL RATE, BOLUS DOSE AND/OR LOCKOUT TIME

    THE FOLLOWING SCREEN HAS CURRENT PUMP

    SETTINGS:

    BASAL = CONTINUOUS RATE

    BOLUS = DEMAND/PCA (Patient Controlled

    Administration) DOSE

    LOCKOUT TIME = FREQUENCY OF BOLUS

    TO CHANGE ANY OF THE SETTINGS, SCROLL

    USING ARROW KEY TO THE DESIRED VALUE

    THAT YOU WANT TO CHANGE AND PRESS

    SELECT

  • CHANGE BASAL RATE, BOLUS DOSE AND/OR LOCKOUT TIME

    ENTER THE CODE OF 513 OR 921 (SEE SLIDE 3)

    CHANGE THE DOSE TO THE DESIRED VALUE

    USING THE UP AND DOWN ARROWSTHEN

    PRESSING THE CONFIRM BUTTON

  • CHANGE BASAL RATE, BOLUS DOSE AND/OR LOCKOUT TIME

    PRESS “YES” TO BYPASS THE SOFT LIMIT SCREEN.

  • CHANGE BASAL RATE, BOLUS DOSE AND/OR LOCKOUT TIME

    REPEAT STEPS IN SLIDES 20 AND 21 FOR EACH

    DESIRED SETTING THAT YOU WANT TO

    CHANGE.

    DO NOT CHANGE THE KVO RATE = “KEEP

    VEIN OPEN”, WHICH IS AUTOMATICALLY

    CALCULATED BY THE PUMP.

    ONCE COMPLETED, HIT THE RED/GREEN

    STOP/START BUTTON.

    THE PUMP WILL ASK YOU TO DOUBLE CHECK THE

    SETTINGS, EVEN THE ONES NOT MODIFIED

    PRESS “REVIEW”

  • CHANGE BASAL RATE, BOLUS DOSE AND/OR LOCKOUT TIME

    PRESS “ACCEPT VALUE” ON ALL SETTINGS

    INCLUDING KVO RATE TO PROCEED.

    ONCE ALL SETTINGS ARE ACCEPTED, PRESS

    “NEXT”

  • CHANGE BASAL RATE, BOLUS DOSE AND/OR LOCKOUT TIME

    STOP THE PUMP BY PRESSING THE RED/GREEN

    STOP/START BUTTON ON THE FRONT OF THE

    PUMP. SEE “HOW TO” ON SLIDE 4, IF

    NECESSARY.

    PRESS “YES” TO START PUMP. TASK COMPLETED.

  • REPROGRAM PUMP INCLUDING MED CONCENTRATION CHANGE

    STOP THE PUMP BY PRESSING THE RED/GREEN

    STOP/START BUTTON ON THE FRONT OF

    THE PUMP. SEE “HOW TO” ON SLIDE 4, IF

    NECESSARY.

    PRESS THE WHITE BUTTON BELOW “TASKS”

    ON THE BOTTOM RIGHT OF THE SCREEN. SEE

    “HOW TO” ON SLIDE 5, IF NECESSARY.

    USING THE UP AND DOWN ARROWS, SCROLL

    DOWN TO “VIEW ADVANCED TASKS” THEN

    PRESS THE “SELECT” BUTTON

  • REPROGRAM PUMP

    SCROLL DOWN TO “START NEW PATIENT” AND

    PLESS THE “SELECT” BUTTON

  • REPROGRAM PUMP

    PRESS “CONTINUE”

    THE PUMP REMINDS THE USER THAT ALL

    PREVIOUS SETTINGS WILL BE ERASED

  • REPROGRAM PUMP

    ENTER THE CODE (SEE SLIDE 3)

    SELECT “PCA” THEN PRESS THE “SELECT”

    BUTTON

  • REPROGRAM PUMP

    THE ONLY OPTION HERE IS TO “PROGRAM

    MANUALLY”. PRESS “SELECT BUTTON”

  • REPROGRAM PUMP

    MORPHINE, HYDROMORPHONE AND

    METHADONE MEDICATIONS ARE ALWAYS

    DISPENSED IN MILLIGRAMS. SELECT “MG” AND

    PRESS “SELECT BUTTON”

    FENTANYL MEDICATIONS ARE ALWAYS

    DISPENSED IN MICROGRAMS. SELECT “MCG”

    AND PRESS “SELECT BUTTON”

  • REPROGRAM PUMP

    SET CONCENTRATION AS THE PATIENT’S PRESCRIPTION

    STATES USING THE “UP” AND “DOWN” ARROWS THEN

    PRESS “CONFIRM”

    CONCENTRATIONS THAT WE DISPENSE - PLEASE CHECK THE

    COLORED LABEL ON THE CASSETTE FOR THE CORRECT

    CONCENTRATION:

    MORPHINE 1MG/ML MORPHINE 10MG/ML

    HYDROMORPHONE 1MG/ML HYDROMORPHONE 10MG/ML

    METHADONE 10MG/ML

    FENTANYL 50MCG/ML FENTANYL 200MCG/ML

  • REPROGRAM PUMP

    VERIFY THE CONCENTRATION THEN PRESS

    “YES”

  • REPROGRAM PUMP

    PRESS “REVIEW” TO CONTINUE PROGRAMMING

    EACH PUMP SETTING INDIVIDUALLY.

  • REPROGRAM PUMP

    THE FOLLOWING SCREEN HAS CURRENT PUMP

    SETTINGS:

    BASAL = CONTINUOUS RATE

    BOLUS = DEMAND/PCA (Patient Controlled

    Administration) DOSE

    LOCKOUT TIME = FREQUENCY OF BOLUS

    TO CHANGE ANY OF THE SETTINGS, SCROLL

    USING ARROW KEY TO THE DESIRED VALUE

    THAT YOU WANT TO CHANGE AND PRESS

    SELECT

  • REPROGRAM PUMP

    CHANGE THE DOSE TO THE DESIRED VALUE

    USING THE UP AND DOWN ARROWSTHEN

    PRESS THE CONFIRM BUTTON

  • REPROGRAM PUMP

    PRESS “YES” TO BYPASS THE SOFT LIMIT SCREEN.

  • REPROGRAM PUMP

    THE CASSETTE RESERVOIR VOLUME SHOULD BE

    RESET TO 250ML

  • REPROGRAM PUMP

    REPEAT STEPS IN SLIDES 35 AND 36 FOR EACH

    DESIRED SETTING.

    AFTER ALL VALUES ARE PROGRAMMED, YOU ARE

    PROMPTED TO PRESS “ACCEPT VALUE” ON ALL

    SETTINGS IN ORDER TO PROCEED.

    DO NOT CHANGE THE KVO RATE = “KEEP VEIN

    OPEN”, WHICH IS AUTOMATICALLY CALCULATED

    BY THE PUMP.

    ONCE ALL SETTINGS ARE ACCEPTED, PRESS

    “NEXT”

  • REPROGRAM PUMP

    ALL NEW CASSETTES ARE PRIMED DURING

    COMPOUNDING SO YOU DO NOT HAVE TO

    PRIME NEW CASSETTES

    PRESS “NO” TO NOT PRIME

  • REPROGRAM PUMP

    STOP THE PUMP BY PRESSING THE RED/GREEN

    STOP/START BUTTON ON THE FRONT OF THE

    PUMP. SEE “HOW TO” ON SLIDE 4, IF

    NECESSARY.

    START THE PUMP AND YOU ARE DONE.