and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod...

31
Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution

Transcript of and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod...

Page 1: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Quadrupole and Ion Trap Mass Analysersand an introduction to Resolution

Page 2: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

A simple definition of a Mass Spectrometer

◗ A Mass Spectrometer is an analytical instrument that can separate charged molecules according to their mass–to–charge ratio.

The mass spectrometer can answer the questions “what is in the sample” (qualitative structural information) and “how much is present” (quantitative determination) for a very wide range of samples at high sensitivity

Page 3: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Vacuum SystemVacuum System

Sample Sample InletInlet DetectorDetector Data Data

SystemSystemMass Mass

AnalyserAnalyserIonisation Ionisation MethodMethod

Components of a Mass Spectrometer

Atmosphere

Page 4: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

How are mass spectra produced ?

◗ Ions are produced in the source and are transferred into the mass analyser

◗ They are separated according to their mass/charge ratio in the mass analyser (e.g. Quadrupole, Ion Trap)

◗ Ions of the various m/z values exit the analyser and are ‘counted’ by the detector

Page 5: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

What is a Mass Spectrum ?

◗ A mass spectrum is the relative abundance of ions of different m/z produced in an ion source

-a “chemical fingerprint”

◗ It contains– Molecular weight information (generally)– Structural Information (mostly)– Quantitative information

Page 6: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

What does a mass spectrum look like ?

◗ Electrospray mass spectrum of salbutamol

Base peak Base peak at m/z 240at m/z 240

MHMH++

HO

HO

NH tBu

OH

6060 8080 100100 120120 140140 160160 180180 200200 220220 240240 260260 280280 30030000

100100

%%

240240

241241

Page 7: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

What information do you need from the analysis ?

◗ Low or High Mass range

◗ Average or Monoisotopic mass (empirical)

◗ Accurate Mass

◗ Quantitation - precision, accuracy, selectivity

◗ Identification

◗ Structural Information

◗ Isotope Ratios

Page 8: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Mass Analyzers

Types of Mass Spectrometer

–Magnetic Sector

–Quadrupole

–Ion Trap

–Time-of-flight

–Hybrid- Sector/trap, Quad/TOF etc

Mass Spectrometers separate ions according to their mass-to-charge (m/z) ratios

Page 9: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

ESI Probe

Q0 Q1 Q2 Q3

Hyperbolic, high precision

quadrupoles

Square Rod Ion Transmission to Analytical Quads

Electron Multiplier, Detection

System

Q2 is Non-Linear Collision Cell

A Triple Stage Quadrupole Mass Analyzer

Finnigan TSQ

Page 10: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Hyperbolic Quadrupoles

Page 11: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

How does the quadrupole work ?

-ve

+ve

The quadrupole consists of four parallel rods. The opposing rodshave the same polarity whilst adjacent rods have opposite polarity.

Each rod is applied with a DC and an RF voltage.Ions are scanned by varying the DC/Rf quadrupole voltages.

Only ions with the selected mass tocharge ratio will have the correct oscillatory pathway in the Rf field.

Page 12: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Quadrupole Mass Analyser

The ion is transmitted along the quadrupole in a stable trajectory Rf field. The ion does not have a stable trajectory and is ejected from the quadrupole.

Page 13: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Non-linear collision cell in the TSQA non-linear Q2 ensures that neutral species are ejected from the mass analyser pathway

Page 14: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Non-linear collision cell removes the line of sight to the detector for neutral species - reducing the noise at the detector.

Non-linear collision cell in TSQ

Page 15: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Q1

Q3

Q2

Q0

Dynode

A 90 Degree Square Q2 in the TSQ Quantum

API

Page 16: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Ion Trap

◗ Consists of ring electrode and two end caps

◗ Principle very similar to quadrupole

◗ Ions stored by RF & DC fields

◗ Scanning field can eject ions of specific m/z

◗ Advantages

- MS/MS/MS…..

- High sensitivity full scan MS/MS

Page 17: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

The ability of a quadrupole to resolve masses is proportional to frequency of AC and length of quadrupole

The ability of a quadrupole to resolve masses is proportional to frequency of AC and length of quadrupole

Resolving Power of a quadrupole

Page 18: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Round rods can only approximate the field generated by hyperbolic rods - most ideal when Rrod/R0 = 1.16

RRrodrod

RR00

RRrodrod = Radius of rod= Radius of rod

RR00 = Position of rod= Position of rod relative relative

to axis of instrumentto axis of instrument

Resolving power of round rods

Page 19: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Manufacturing rods such that Rrod/R0 = 1.16 over the full length of quadrupole is non-trivial. Variability would assure that one M+ is selected at one position and another where the ratio is different - loss of specificity and sensitivity.

Thermal expansion coefficients must be very low – expansion /contraction within the length of the rod changes Rrod/R0

Hyperbolic rods give excellent transmission across the full range of resolution - providing optimal analytical specificity. Compare sensitivity at unit resolution versus open resolution. At tighter resolution round rods systems tend to lose a greater proportion of ions through variation in Rrod/R0.

Resolving power of round rods

Page 20: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Round rods manufacture typically requires several steps - first, machine rods. Second, machine ‘holder’ by drilling four holes into a cylinder. Third, machine ‘dropout’ to create ion path. Insert rods. Error in Rrod/R0 grows by number of processes.

RRrodrod

RR00

Manufacture of round rods

Page 21: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Magnetic Sector Instruments

Constant Resolution with mass

10% Valley Definition

Magnetic Sector Instruments

Constant Resolution with mass

10% Valley Definition

Resolution Basics

Resolution = m / ∆ m

Page 22: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Quadrupoles

Ion traps

and TOF’s

Constant peak widths

Variable Resolution

FWHM Definition

Quadrupoles

Ion traps

and TOF’s

Constant peak widths

Variable Resolution

FWHM Definition

Resolution Basics

Resolution = m / ∆ mm/z 500/1, R = 500m/z 500/0.1, R = 5000

Page 23: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

The advantage of high resolution measurement is to eliminate chemical background of the same nominal mass but different accurate mass and, therefore to increase the signal to noise ratio and the sensitivity respectively.

Better selectivity is obtained by applying high resolution in case of isobaric compounds, i.e. two compounds of same nominal mass but different accurate mass. With low resolution only a combined Spectral result is obtained under Product ion conditions . With high resolution separate detection and therefore separate isolation and MS/MS spectra are obtained.

Example: R= 1000 R= 5000

Compound mass1: 372.351 Da OVERLAPPING SEPARATIONCompound mass2: 372.421 Da

What is the advantage of using high resolution mass spectrometry?

Page 24: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

1.0 FWHM 0.7 FWHM 0.5 FWHM

Effect of changing resolution on peak shape

0.3 FWHM 0.2 FWHM 0.1 FWHM

Page 25: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Effect of Peak Width On Transmission

0

20

40

60

80

100

%

Transmission

2 1.5 0.7 0.5 0.2 0.1

Peak Width FWHM

QUADRUPOLE ROUND RODS

5000 *2500 *

1000 *Unit

* Resolution at m/z 500

Page 26: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Resolving Target Compounds in the Presence of Interferences

◗ Examples: Mometasone and Alprazolam were spiked into PPG

◗ >5000 (FWHM) parent ion resolution allows mass selection of these compounds differing by less than 0.2 Da from PPG

◗ Increasing the resolution to exclude the PPG interference increases the selectivity and specificity for the quantitative assay

Page 27: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Mometasone resolution in the presence of PPGs(Mometasone [M+H]+ = 521.2, PPG (n=8) [M+Na]+ = 521.3)

518.0 518.5 519.0 519.5 520.0 520.5 521.0 521.5 522.0 522.5 523.0 523.5 524.0 524.5 525.0

m/z

Rel

ativ

e A

bund

ance

521.2

523.2

522.2521.3

522.9 524.1523.4522.3

521.2

521.3

521.2

523.2

522.9 523.4

522.8

523.2

resolved Q1 mass spectrum

transmitted parent ionsQ1 at unit resolution

Q1 selecting onlymometasone 35Cl

Q1 selecting onlymometasone 37Cl

transmitted parent ionsQ1 at unit resolution

PPG

PPG PPG

mometasone 37Cl

mometasone 35Cl

mometasone 37Cl

mometasone 37Cl

mometasone 35Cl

Page 28: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Mometasone resolution in the presence of PPGsproduct ion spectra of 37Cl mometasone at various precursor ion resolutions

The increased resolution on product ion scan of 37Cl mometasone can selectively exclude PPG interferents with an actual mass difference of less than 0.2 Da.

Q1 resolution unit

Q1 resolution 0.08 FWHMpassing only mometasone 37Cl

* = PPG - related peaks

Page 29: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Alprazolam resolution in the presence of PPGs(alprazolam [M+H]+ = 309.0907, PPG (n=5) [M+H]+ = 309.2277)

250 pg/uL alprazolam in 20 uM PPGs

306.0 306.5 307.0 307.5 308.0 308.5 309.0 309.5 310.0 310.5 311.0 311.5 312.0

m/z

Rel

ativ

e A

bund

ance

309.3

309.1

309.3

309.1

309.3

Transmitted precursor ions

309→309

PPG

alprazolam

PPG

0.08 Da FWHM

Unit resolution on Q1

0.08 FWHM resolution on Q1Set to pass PPG ions

0.08 FWHM resolution on Q1Set to pass alprazolam ions

alprazolam

Page 30: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Alprazolam resolution in the presence of PPGsProduct ion spectra at various precursor ion resolutions

The increased resolution on product ion scan of alprazolam can selectively excludePPG interferents with an actual mass difference of less than 0.2 Da.

Q1 set at unit resolution, spectrum is a composite of alprazolam and interfering PPG ions

Q1 set to pass only PPG ions

Q1 set to pass only alprazolam ions

Page 31: and an introduction to Resolution Quadrupole and Ion Trap ......Manufacturing rods such that R rod /R 0 = 1.16 over the full length of quadrupole is non-trivial. Variability would

Ion Traps Quadrupoles

Mass Separation in time Mass Separation in space

High sensitivity Full Scan Lower sensitivity Full Scan

Lower sensitivity SRM High sensitivity SIM and SRM

Offer mutiple stages of MSn No mutiple stages of MS

Offers higher resolution Lower resolution

Parent and neutral loss scans

Comparison of Quads and Traps