Advantages of Using Triple Quadrupole over Single ......Single quadrupole systems contain only one...

6
p 1 Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber SCIEX Concord, Ontario (Canada) Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Higher selectivity resulting in less interference of co-eluting compounds and matrix, thus less HPLC separation is required Better Signal-to-Noise (S/N) allowing quantitation with lower limits of quantitation More reliable identification of detected analytes using Multiple Reaction Monitoring (MRM) in comparison to Selected Ion Monitoring (SIM) Wider linear range of quantitation Better accuracy and reproducibility especially at low concentrations Overview This paper describes the use of single quadrupole and triple quadrupole Mass Spectrometry coupled to Liquid Chromatography for the analysis of 17 pesticides in drinking water and soil samples. Both detection technologies are compared with respect to selectivity, sensitivity, identification, linear range, accuracy, and reproducibility for quantitative analysis. Introduction The coupling of Liquid Chromatography (LC) and tandem Mass Spectrometry (MS/MS) is a widely used analytical technique for quantitative and qualitative analysis. Electrospray Ionization (ESI), Atmospheric Pressure Chemical Ionization (APCI), or Photo Ionization (APPI) allow the ionization of various semi- volatile, thermally labile, and polar to nonpolar compounds, such as pharmaceuticals, pesticides, personal care products, steroids, explosives, drugs of abuse etc., in trace levels. Generated ions will be transferred after ionization through a vacuum interface into the mass analyzer. Quadrupoles are mass analyzers which consist of four rods with DC and RF voltages applied. An ion of a specific mass-to-charge ratio (m/z) will be stable and can pass through the quadrupole only when a specific DC/RF voltage combination is applied. Quadrupoles are therefore called mass filters. Figure 1. Schematic of Triple Quadrupole Configuration

Transcript of Advantages of Using Triple Quadrupole over Single ......Single quadrupole systems contain only one...

  • p 1

    Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber SCIEX Concord, Ontario (Canada)

    Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: • Higher selectivity resulting in less interference of co-eluting

    compounds and matrix, thus less HPLC separation is required

    • Better Signal-to-Noise (S/N) allowing quantitation with lower limits of quantitation

    • More reliable identification of detected analytes using MultipleReaction Monitoring (MRM) in comparison to Selected IonMonitoring (SIM)

    • Wider linear range of quantitation

    • Better accuracy and reproducibility especially at lowconcentrations

    Overview This paper describes the use of single quadrupole and triple quadrupole Mass Spectrometry coupled to Liquid Chromatography for the analysis of 17 pesticides in drinking water and soil samples. Both detection technologies are compared with respect to selectivity, sensitivity, identification, linear range, accuracy, and reproducibility for quantitative analysis.

    Introduction The coupling of Liquid Chromatography (LC) and tandem Mass Spectrometry (MS/MS) is a widely used analytical technique for quantitative and qualitative analysis. Electrospray Ionization (ESI), Atmospheric Pressure Chemical Ionization (APCI), or Photo Ionization (APPI) allow the ionization of various semi-volatile, thermally labile, and polar to nonpolar compounds, such as pharmaceuticals, pesticides, personal care products, steroids, explosives, drugs of abuse etc., in trace levels. Generated ions will be transferred after ionization through a vacuum interface into the mass analyzer.

    Quadrupoles are mass analyzers which consist of four rods with DC and RF voltages applied. An ion of a specific mass-to-charge ratio (m/z) will be stable and can pass through the quadrupole only when a specific DC/RF voltage combination is applied. Quadrupoles are therefore called mass filters.

    Figure 1. Schematic of Triple Quadrupole Configuration

  • p 2

    Single quadrupole systems contain only one mass filtering quadrupole while triple quadrupole systems consist of three quadrupoles. Q1 and Q3 are working as mass filters while Q2 is acting as collision cell.

    Quadrupoles can be used in scanning or filtering mode. During a mass scan, DC and RF voltages are ramped resulting in the acquisition of full scan mass spectra. Such spectra are typically used for qualitative data analysis. However, scanning a quadrupole suffers from low sensitivity and slow scan speed. Thus, quantitative studies are performed with quadrupoles working in filtering mode.

    The most selective mode to use a single quadrupole MS is called Selected Ion Monitoring (SIM). Hereby, a fixed set of DC and RF voltages is applied to the quadrupole and thus only a single m/z can pass. Ions with different m/z are filtered out.

    Multiple Reaction Monitoring (MRM) is the most common mode of using a triple quadrupole MS/MS for quantitative analysis, allowing enhanced sensitivity and selectivity. The first quadrupole filters a specific precursor ion of interest. Ions generated in the ion source having a different m/z can not pass Q1. The collision cell is optimized to produce a characteristic product ion by collision of the precursor ion with a neutral collision gas, such as nitrogen. This process is called Collision Induced Dissociation (CID). Generated product ions are transferred into the third quadrupole where only a specific m/z is allowed to pass. All other product ions are filtered out in Q3.

    Thus MRM mode works like a double mass filter which drastically reduces noise and increases selectivity. The principle and resulting chromatograms of various scan modes analyzing a mix of pesticides spiked into soil is given in Figure 2. It illustrates how increasing selectivity of the mass spectrometric experiment reduces the noise of eluting background and matrix components.

    Single quadrupole and triple quadrupole systems allow the detection of many SIM and MRM transitions, respectively. This enables quantitation of many targeted analytes in a single experiment. Typically, additional SIM and MRM transitions have to be detected to perform identification of quantified compounds. Hereby, the most intense ion is called the ‘quantifier’ and all additional ions are called ‘qualifiers’. The EU Commission Decision 2002/657/EC defined performance criteria for confirmatory methods, such as MS and MS/MS, by introducing the concept of identification points. The required number of 4 identification points can be achieved by detecting 4 SIM on a single quadrupole MS or 2 MRM transitions on a triple quadrupole MS/MS.

    Figure 2. Pesticide in a soil extract detected in different single and triple quadrupole scan modes (left to right): Full scan with MS spectrum (50 μg/kg), Selected Ion Monitoring (5 μg/kg), and Multiple Reaction Monitoring (5 μg/kg)

    Time

    m/z

    m/z 1

    m/z 3m/z 2

    last m/z last m/z

    m/z 1

    m/z 3m/z 2

    Time

    m/z

    SIM 1

    SIM 2

    last SIM

    SIM 1

    SIM 2

    last SIM

    Time

    Q1/

    Q3

    MRM 1

    MRM 2

    last MRM

    MRM 1

    MRM 2

    last MRM

    80 90 100 110 120 130 140 150 160 170 180 190 200 210 220 230 240 250 260m/z, amu

    0.00

    2.00e4

    4.00e4

    6.00e4

    8.00e4

    1.00e51.15e5 253

    171 20210288

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15Time, min

    0.0

    5.0e7

    1.0e8

    1.5e810.2

    11.20.6 12.4

    13.3 14.09.98.2 9.49.27.66.86.5

    1 2 3 4 5 6 7 8 9 10 11 12 13 14Time, min

    0.0

    1.0e4

    2.0e4

    3.0e4

    4.0e4

    5.0e4

    6.0e4

    7.0e4

    8.0e4

    9.0e4

    1.0e5

    1.1e5

    1.2e5

    1.3e5

    1.4e5

    1.5e5

    1.6e5

    11.1

    7.1

    9.60.6

    1 2 3 4 5 6 7 8 9 10 11 12 13 14Time, min

    0

    500

    1000

    1500

    2000

    2500

    3000

    3500

    4000

    4500

    5000

    5500

    6000

    65006.9

  • p 3

    Experimental HPLC

    An Agilent 1200 standard HPLC system with binary pump, well plate autosampler, and column oven was used. Chromatographic separation was performed on a Phenomenex Synergi 4u Fusion RP-80 (50x2 mm) column. A gradient of eluent A (water with 5 mM ammonium formate) and eluent B (methanol with 5 mM ammonium formate) from 80/20 to 10/90 (A/B) over 8min was used. The column temperature was set to 25°C. A sample volume of 10 μL was injected.

    MS/MS

    MS and MS/MS detection was performed using an API 3200™ LC/MS/MS system with Turbo V™ source and an Electrospray ionization probe. All pesticides were detected in positive polarity using an ionization voltage of 5000V. The ion source temperature was set to 500°C. The detected quantifier and qualifier ions are listed in Table 1. A dwell time of 25 ms was used to detect 34 MRM transitions and 15ms to detect 68 SIM.

    Results and Discussion Selectivity

    Detection on a triple quadrupole MS/MS results in higher selectivity due to the double mass filtering (Figure 3). In MRM typically a single signal per analyte is detected while SIM can result in multiple signals per compound. The reason for this is the generation of identical in-source fragment ions of pesticides of the same compound class. Thus more time consuming HPLC is required to separate such compounds to allow quantitation and identification. Needed HPLC development and analysis times are then comparable to traditional UV detection.

    Sensitivity

    Detection on a MS/MS system also results in better sensitivity. Detected Signal-to-Noise (S/N) is higher resulting in lower Limits of Quantitation (LOQ). The typical difference in sensitivity observed is at least one order of magnitude as illustrated in Figure 4.

    Table 1. Pesticides detected in Multiple Reaction Monitoring (MRM) and Selected Ion Monitoring (SIM)

    Pesticide CAS MRM 1 MRM 2 SIM 1 SIM 2 SIM 3 SIM 4 tR (min)

    Atrazine 1912-24-9 216/174 216/104 216 174 146 104 7.8

    Chlortoluron 15545-48-9 213/72 213/46 213 72 140 168 7.6

    Cyanazine 21725-46-2 241/214 241/104 241 214 104 132 6.7

    Desethylatrazine 6190-65-4 188/146 188/104 188 146 104 110 5.5

    Diuron 330-54-1 233/72 233/46 233 72 235 160 8.0

    Hexazinone 51235-04-2 253/171 253/71 253 171 71 85 6.9

    Isoproturon 34123-59-6 207/72 207/46 207 72 165* 134* 7.9

    Linuron 330-55-2 249/160 249/182 249 160 251 182 8.5

    Metazachlor 67129-08-2 278/134 278/210 278 134* 210 105 7.7

    Methabenzthiazuron 18691-97-9 222/165 222150 222 165* 150 124 7.8

    Metobromuron 3060-89-7 259/170 259/148 259 261 148 170 7.7

    Metolachlor 51218-45-2 284/252 284/176 284 252 176 134 9.2

    Metoxuron 19937-59-8 229/72 229/46 229 72 156 106 6.2

    Monolinuron 1746-81-2 215/126 215/99 215 126 99 148 7.4

    Sebuthylazine 7286-69-3 230/174 230/104 230 174 104 146 8.4

    Simazine 122-34-9 202/132 202/124 202 132 104 174 6.9

    Terbuthylazine 5915-41-3 230/174 230/104 230 174 104 146 8.6

    * Not enough HPLC separation to differentiate these in-source fragment ions

  • p 4

    If identification with 4 identification points is required and more than a single ion has to be detected the difference in LOQ based on quantifier MRM and SIM is significantly higher (Table 2).

    Linearity and Reproducibility

    Generally a wider linear range can be used for quantitation if a triple quadrupole MS/MS is used because of increased selectivity allowing detection at lower LOQ (Figure 5). The upper limit of quantitation of SIM and MRM is typically comparable due to saturation of the ion source or detector. The typical linear range in MRM using Electrospray Ionization is 3 to 4 orders of magnitude while SIM provides only 2 to 3 orders of magnitude.

    Additionally, MS/MS allows more accurate and reproducible quantitation. Table 3 compares Coefficients of Variation (%CV) of detected pesticides at different concentrations. Triple quadrupole MS/MS provides %CV below 10 at LOQ and below 5 at higher concentrations. The reproducibility on single quadrupole MS is significantly lower.

    Figure 4. Comparison of sensitivity detecting 1 μg/L Hexazinone (left) and 10 μg/L Chlortoluron (right) in MRM and SIM mode in drinking water

    Figure 3. Comparison of selectivity detecting Atrazine (left) and Diuron (right) in MRM and SIM mode at 100 μg/L in drinking water

    XIC of +MRM (34 pairs): 253.2/171.2 amu from Sample 14 (QqQ 1) of Data quant Q vs QqQ.wiff (Turb... Max. 5882.5 cps.

    1 2 3 4 5 6 7 8 9 10 11 12 13 14Time, min

    0

    1000

    2000

    3000

    4000

    5000

    5883 6.9

    XIC of +Q1 MI (68 ions): 253.2 amu from Sample 13 (Q 1) of Data quant Q vs QqQ.wiff (Turbo Spray), ... Max. 5.6e4 cps.

    1 2 3 4 5 6 7 8 9 10 11 12 13 14Time, min

    0.0

    1.0e4

    2.0e4

    3.0e4

    4.0e4

    5.0e4

    5.6e4 0.3

    6.9 253 S/N=18171 S/N=685 S/N

  • p 5

    Table 2. Limits of Quantitation (LOQ) and Limit of Identification (LOI) of detected pesticides*

    Pesticide LOQ (MRM 1) LOI (MRM 2) LOQ (SIM 1) LOI (SIM 4) Enhanced Quantifier Enhanced Qualifier

    Atrazine 0.02 0.05 0.2 1 10x 20x

    Chlortoluron 0.1 0.5 1 10 10x 20x

    Cyanazine 0.1 0.2 0.5 2 5x 10x

    Desethylatrazine 0.1 0.1 0.5 2 5x 20x

    Diuron 0.2 0.5 1 10 5x 20x

    Hexazinone 0.02 0.05 0.2 1 10x 20x

    Isoproturon 0.05 0.1 0.5 N/A 10x N/A

    Linuron 0.1 0.2 2 10 20x 50x

    Metazachlor 0.02 0.02 0.5 5 25x 250x

    Methabenzthiazuron 0.02 0.05 0.5 2 25x 40x

    Metobromuron 0.2 0.5 2 5 10x 10x

    Metolachlor 0.02 0.1 0.5 50 25x 500x

    Metoxuron 0.1 0.5 0.5 20 5x 40x

    Monolinuron 0.1 0.2 2 5 20x 25x

    Sebuthylazine 0.1 0.1 0.5 2 5x 20x

    Simazine 0.1 0.2 0.5 1 5x 5x

    Terbuthylazine 0.02 0.1 0.5 2 25x 20x

    *using Multiple Reaction Monitoring (MRM) and Selected Ion Monitoring (SIM) in drinking water samples based on S/N=6

    Table 3. Reproducibility (%CV) of detected pesticides in MRM and SIM at different concentrations

    Pesticide %CV at 1μg/L

    (MRM) %CV at 10μg/L

    (MRM) %CV at 100μg/L

    (MRM) %CV at 1μg/L

    (SIM) %CV at 10μg/L

    (SIM) %CV at 100μg/L

    (SIM)

    Atrazine 4.4 3.7 0.9 16.2 1.8 0.6

    Chlortoluron 4.2 3.4 1.1 20.3 2.9 1.7

    Cyanazine 9.0 2.1 1.8 16.6 4.6 0.9

    Desethyl-atrazine 3.8 2.0 1.5 5.8 0.8 1.4

    Diuron 4.4 2.5 1.2 23.7 2.9 1.6

    Hexazinone 1.7 1.3 1.2 3.0 1.5 1.5

    Isoproturon 8.7 0.9 1.3 17.2 3.0 1.3

    Linuron 9.6 7.0 0.9

  • AB Sciex is doing business as SCIEX.

    © 2017 AB Sciex. For Research Use Only. Not for use in diagnostic procedures. The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective owners.

    AB SCIEX™ is being used under license.

    Publication number:RUO-MKT-18-5619-A / 0701310-02

    Summary The detection in Multiple Reaction Monitoring (MRM) using a triple quadrupole MS/MS has a number of advantages in comparison to Selected Ion Monitoring (SIM) of a single quadrupole MS.

    Due to double mass filtering MS/MS detection provides much higher selectivity with less interference of co-eluting compounds and matrix components, resulting in less time consuming method development and faster analysis times. Better Signal-to-Noise allows quantitation with lower Limits of Quantitation (LOQ).

    Fewer ions have to be detected per compound on MS/MS in comparison to MS for confirmatory analysis. The concept of identification points introduced by the EU Commission Decision requires the detection of 2 MRM transitions and 4 SIM, respectively. If identification is required the detected qualifier signals have to be compared to specify LOQ. The difference between both technologies can be then at least two orders of magnitude. Finally, a wider linear range, higher accuracy, and reproducibility can be obtained on triple quadrupole MS/MS.

    Figure 5. Comparison of linearity detecting Linuron (left) and Terbuthylazine (right) in MRM (top) and SIM (bottom) mode in spiked drinking water samples

    quantitation QqQ spiked water.rdb (Linuron 1): "Linear" Regression ("1 / x" weighting): y = 1.7e+003 x + -50.9 (r = 0.9993)

    0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200Concentration, ng/mL

    0.0

    5.0e4

    1.0e5

    1.5e5

    2.0e5

    2.5e5

    3.0e5

    3.4e5

    quantitation Q spiked water.rdb (Linuron 1): "Linear" Regression ("1 / x" weighting): y = 1.86e+004 x + 1.33e+004 (r = 0.9987)

    0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200Concentration, ng/mL

    0.0

    5.0e5

    1.0e6

    1.5e6

    2.0e6

    2.5e6

    3.0e6

    3.5e6

    249/160 r=0.9993(0.1 - 200μg/L)

    249 r=0.9987(2 - 200μg/L)

    quantitation QqQ spiked water.rdb (Terbuthylazine 1): "Linear" Regression ("1 / x" weighting): y = 1.22e+004 x + 750 (r = 0.9989)

    0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200Concentration, ng/mL

    0.0

    5.0e5

    1.0e6

    1.5e6

    2.0e6

    2.4e6

    quantitation Q spiked water.rdb (Terbuthylazine 1): "Linear" Regression ("1 / x" weighting): y = 1.08e+005 x + 4.59e+004 (r = 0.9972)

    0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200Concentration, ng/mL

    0.0

    2.0e6

    4.0e6

    6.0e6

    8.0e6

    1.0e7

    1.2e7

    1.4e7

    1.6e7

    1.8e7

    2.0e7

    230/174 r=0.9989(0.1 - 200μg/L)

    230 r=0.9972(0.5 - 200μg/L)