1 12/2015 Rijuta Ravichandran, Devan Boyce
Allwin21 Corp-AccuThermo AW610
The following document is intended to describe the safe operating procedures of the
AccuThermo AW610 for CNM2 members. This manual is meant to used as a reference
and does not replace formal training. In order to gain access to the tool, formal
qualification by staff is required.
System Overview
The AccuThermo AW610 is a rapid thermal processing (RTP) system. It uses high
intensity visible radiation to heat single wafers for short process periods of time at
precisely controlled temperatures.
The system itself consists of an oven unit and a computer running the AW-610 controller
software. The oven unit consists of a quartz tray that slides into an isolation tube
between two sets of lamps that provide the source of heat. The lamp intensity and heat
is adjusted by recipes created on the controller software.
2 12/2015 R. Ravichandran, D. Boyce
System Configuration
Water-cooled, aluminum chamber with gold plating for temperature uniformity
Isolated quartz tube
Quartz tray for 4 to 6 in wafer (contact staff if using pieces)
Thermocouple (100-550C)
Pyrometer for non-contact temperature sensing (400-1250C)
0-300 seconds for “steady” processing time (Fig 1.)
4’’ Thermocouple wafer for pyrometer calibration
Nitrogen and Argon gas flow
Advantages (over conventional furnace)
Fast heating and cooling rates in short amounts of time
Superior heating uniformity to minimize thermal stress
Allows for multiple processes without cross contamination
Disadvantages
Single wafer only
Limited in duration of heating and processing
Not easily integrated with other heating processing steps (i.e. oxidation)
Figure 1. Time vs. Temp chart for selecting appropriate anneal times (min) for “steady” step
processing time
3 12/2015 R. Ravichandran, D. Boyce
Principle of Operation
The AccuThermo Allwin610 system is a rapid thermal processing (RTP) unit, which uses high intensity
visible radiation to heat single wafers for short process periods of time at precisely controlled temperatures.
Figure 2 describes the general temperature profile for any given recipe. First the system is purged of ambient air
in the delay stage, and then wafer is pre-heated in the intensity stage, if the
pyrometer sensor is selected. Next, the system ramps to the desired set point, there can be subsequent ramps in a
recipe. The transition stage is when the temperature control changes from ramp to steady- state, to stabilize the
temperature in the system. During steady state control, the process is set to keep the temperature constant for a
given time programmed (0-300s) in the recipe. After this step, the lamps turn off and the wafer goes through a
cool-down stage. This process provides a short thermal annealing process for either Silicon or III-V materials.
The key component throughout this process is accurately measuring and modulating the temperature of
the system. There are two sensors available to measure temperature in the system: thermocouple and
pyrometer. The thermocouple is generally used for processes below 550C and will degrade above those
temperatures. The pyrometer is used for processes between 400 and 1250C, and
has a much faster response time than the thermocouple. Once the correct temperature sensor of the
system has been identified, it is important to apply the appropriate compensation and correction factors for the
sensor.
The correct material type, emissivity values, and other temperature correction variables must be
selected in order to ensure accurate temperature readings. If you are working with III-V compounds or other IR-
transparent materials you must use a wafer susceptor. Because the thermal capacitance of the susceptor is very
different from the standard silicon wafer, it is important to indicate when you are using a Si wafer or a susceptor. If
you are using non-standard silicon wafers it is important to find the correct emissivity value for your material. The
emissivity value is used as a correction factor for the pyrometer, and acts an offset for the temperature readings.
Lastly, there are four variables: Gain, Sensitivity, Delay, and Steady Intn factor that are all used to fine tune
temperature performance. It is possible to permanently damage the tool without properly checking these settings.
Therefore, it is critical to choose the correct sensor, select the correct emissivity values, and fine-tune the
temperature profile for your process. More information on selecting the correct settings and variables for your
process will be discussed in the attached supporting documentation.
4 12/2015 R. Ravichandran, D. Boyce
Figure 2. Temperature profile of RTP process. Not shown: The DELAY stage is used to purge the
chamber of ambient air. The INTENSITY stage preheats the wafer to a temperature that is
sufficient for the pyrometer to be usabl
5 12/2015 R. Ravichandran, D. Boyce
OPERATING INSTRUCTIONS Open the required facilities gas lines
Turn all required gases to the “valve open” position shown here. Without opening these
valves there will be no gas flow into the chamber. The Clean Dry Air (CDA) line must always
be open when running the RTP.
Valve open
Valve closed
CDA: Clean dry air
Valve MUST be open
(shown: open)
Turn on lamp power
Temperature display ( kC)
Currently displayed: 69C
Lamp power switch (on)
Lamp power indicator
6 12/2015 R. Ravichandran, D. Boyce
Loading samples into chamber
Shown: Chamber door in LOCK position.
7 12/2015 R. Ravichandran, D. Boyce
Push handle downwards to move the chamber door to UNLOCK position
CAREFULLY pull chamber door out to place your wafer on the quartz tray. Make sure the wafer
is in contact with the thermocouple or the sample will overheat and damage the chamber.
Shown: 6’’ wafer that melted inside the chamber because the sample was overheated.
CAREFULLY slide the chamber door closed. Pull chamber handle upwards to lock the door.
8 12/2015 R. Ravichandran, D. Boyce
Running a recipe
4. Click to select a Recipe from the list of Recipes in the column “Recipe File”. You can
scroll through the list of filenames by using the slider to the right of the list, or using the up and
down arrow keys on the keyboard. A selected filename is highlighted in red. You can also select
a filename by manually typing in the filename into the field above the list.
From the main menu, select "Process for Production
9 12/2015 R. Ravichandran, D. Boyce
WARNING: Selecting a filename and pressing DEL on the keyboard will
permanently delete the file 5. Select the folder with your username (created during the training session) under “DirID”.
Select or create a “Lot ID” from the list of Lot IDs. This specifies where the process data is stored
upon completion
6. To run the selected Recipe, click on the Start Process button. This will display the
Process Monitor screen while the process is running. The Process Monitor screen shows a
completed process curve. In figure 3, the x-axis indicates the process time and the y-axis
indicates the measured process temperature.
The curves display aspects of the monitored process:
10 12/2015 R. Ravichandran, D. Boyce
Figure 3: Process window as the RTP runs your recipe. X-axis displays time and then y-axis
displays process temperature. The process temperature is also determined based on the
temperature sensor used and the other settings selected are displayed on the right hand side of
the screen.
Green: recipe temperature as defined by the Recipe.
Black: Model (ideal) temperature curve.
Blue: real, measured temperature during the wafer process.
Red: The lamp intensity (power percentage) during the wafer process. The range is
from 0 to 100.
Light Blue: temperature feedback from the thermocouple.
Pink: temperature feedback f r o m t h e p y r o m e t e r . The t e m p e r a t u r e from
this pyrometer is not reliable below 450 °C.
Broad Lines: Various gas flows
11 12/2015 R. Ravichandran, D. Boyce
7. A process may be interrupted and stopped at any time by pressing the ESC key on the
keyboard. This will turn off the lamps. If the “Turn Off Gases After Process” in the recipe is set
to NO, then the gases that are on will stay on and flow at the current flow rate. If it is set to YES,
then the gases will be shut off.
At the end of the process, “Process Over” will be displayed on the Process Monitor screen and
be blinking in different colors. Press any key to save the process data. This will also exit back to
the Process for Production screen.
To view the last run process data graphically, click on the Display Last Data button. This will
display the process data on a screen very similar to the Process Monitor screen. To view
process data from previous runs, select the process data to be viewed from the “Data ID”
column, and then click on the Display Process Data button.
12 12/2015 R. Ravichandran, D. Boyce
Creating and editing a recipe
Use the recipe
menu to select
file to edit
Open the recipe menu from the main menu to see a list of pre-programmed recipes that are
available to be viewed or to be edited for your own process.
Use the arrow keys to scroll through the recipe files. Select a recipe file and click on “Recipe
edit” to view the set points of the process.
WARNING: Selecting a file and then pressing "delete" will
permanently delete the file.
13 12/2015 R. Ravichandran, D. Boyce
(1a) Recipe Name: Limited to 8 characters
(2e) Steady Intensity (Intn) Factor. Used to fine tune the temperature profile.
(2d) Wafer Type: Wafer or Suceptor
(2c) Gain, Sensitivity, and Delay values for fine-tuning the temperature profile
(2b) Emissivity Values for Si substrate
(2a) Sensor Type: Pyrometer of thermocouple
(1e) Recipe Validate before exiting and saving
(1d) Gas Type (N2 or Ar) and flow rate (SLPM)
(1c) Processing time for each step
(1b) Processing Steps: Delay, Intensity, Ramp, Steady,(...) Delay
14 12/2015 R. Ravichandran, D. Boyce
1a. After you have selected an existing recipe and change its name before modifying the
parameters and saving it. Recipes must comply with the following format or they may
be deleted by staff at any time:
Leave the EXT field as RCP.
Pyrometer recipe names must begin with “p_” and TC recipe names must begin
with “t_”.
Recipes must also include the temperature, gas, and time of the process at the
final temp (ex. P600AR30, T200N20, etc).
All recipes have an 8 char max, please try to be as descriptive as possible given the
limitations.
15 12/2015 R. Ravichandran, D. Boyce
1b. In the lower half of the screen, the recipe contains steps that describe the process cycle.
Step Temp Func is the Process Function, which describes the type of process function for that
step. It can be RAMP, INTN, STEADY, DELAY, or FINISH.
Ramp: The Ramp step instructs the controller to increase the temperature at a
constant rate until the specified temperature has been reached. The rate is
calculated by dividing the difference between the temperature specified in the step
and the temperature specified in the previous step by the time specified in the step.
The process controller cannot do two consecutive RAMP steps.
Intn: The controller keeps the lamp power at a constant intensity during an Intensity
step. The Intn step is used to heat the wafer to a temperature where the wafer is
seen by the pyrometer. The minimum reliable temperature for use of the
pyrometer is 500ºC.
Steady: During the Steady step, the lamp intensity is controlled to maintain the
specified anneal temperature. It then maintains that value until the specified time
spent in the step has elapsed.
Delay: The Delay step instructs the controller to turn the lamps off while setting and
maintaining the set point of the other controlled parameters (i.e. Gases), until the
specified time spent in the step has elapsed.
Finish: This ends the recipe.
16 12/2015 R. Ravichandran, D. Boyce
1c. Time: This is the amount of time, in seconds, to act on the current step. Please refer to the
time-temperature guide below for the maximum steady-state time for each temperature:
(REFERENCE: Stanford SNF)
Temp/Intn: This is the target temperature or intensity system strives for.
Step Function Field-entry description
Steady Set-point temperature to be maintained
Ramp Ending temperature of the ramp
INTN The percentage of the lamp power intensity
Delay No effect. Lamps are turned off during this step
1d. Gas Flow rates: The various gas columns specify the flow rate of each gas for each step. The
system is configured for max 10SLPM flow rate of both N2 and Ar. If the “turn off gas after
process” button is selected to say “yes” it will turn off the gases after the process. Click on this
button to toggle between yes/no.
17 12/2015 R. Ravichandran, D. Boyce
1e. When entering data values into the data entry area, the recipe editor checks for out-of-
range entries. If a value is out-of-range, the editor will alert you and will advise you of the
proper range. The recipe can be validated by clicking on the Recipe Validate (F10) button. All
errors need to be corrected before the recipe can be used for processing. Click Save and Exit
the Recipe Editor.
WARNING: Recipe Validate does not
consider the time constraints for
various temperatures. Please refer to
the time vs. temperature chart to
choose the appropriate time settings
for the “steady” step.
18 12/2015 R. Ravichandran, D. Boyce
Shown: Recipe after successful validation. Press save and exit this menu.
2a. Sensor type must be set to either thermocouple or pyrometer depending on the
temperatures used in your process. Use the thermocouple for processes up to 550°C. The
pyrometer is unreliable for temperatures under 500°C, but can read temperatures up 1100°C.
2b. Emissivity is a correction factor for the pyrometer. A standard silicon wafer has an
emissivity of 77.04. The software treats emissivity as a factor that offsets the pyrometer
temperature reading.
2c. The variables - Gain, Sensitivity, Delay, Steady Intn Factor - are for fine tuning the
temperature control performance. It is best to start with the default values in the recipe and
fine tune them as needed. More information about adjusting these parameters are included in
the supporting documentation.
A. Gain: The value for the gain of the temperature PID control for the process steps
RAMP and STEADY.
B. Sensitivity: The value for the coefficient of the gain in the STEADY period of the
recipe. This makes the gain for a STEADY step different from a RAMP step. The
gain value for STEADY is GAIN x SENSITIVITY. If the SENSITIVITY value is 1.0, there
is no difference between the gain for a RAMP and a STEADY.
C. Delay: This is the value for the transient period from RAMP to STEADY. Normally
the default value is 1.0.
D. Steady Intn Factor: This is a coefficient used only during RAMP and STEADY
steps. It is used at the beginning of the steps to correct the initial lamp power
intensity. This feature is used to trouble-shoot temperature profiles and
explained in more detail in supporting documentation.
19 12/2015 R. Ravichandran, D. Boyce
Overview-Hardware
Supporting Documentation
LED Dis play
(wafer temp)
Chamber door handle
(shown: lock position)
Emergency Off
EMO
Lamp power
Indicat or
Lamp power
switch (on/off)
Oven overheat
indicator
Chamber door
20 12/2015 R. Ravichandran, D. Boyce
Recipe Menu for
editing processes
To run process,
and view data
STAFF ONLY
STAFF ONLY
Exit System
Login here
Eqpt manual
Overview-Software
21 12/2015 R. Ravichandran, D. Boyce
(1a) Recipe Name: Limited to 8 characters
(2d) Wafer Type: Wafer or Suceptor (2a) Sensor Type: Pyrometer or
thermocouple (2b) Emissivity Values for Si substrate
(2c) Gain, Sensitivity, and Delay values for fine-tuning the temperature profile
(1e) Recipe Validate before exiting and saving
(1d) Gas Type (N2 or Ar) and flow rate (SLPM)
(2e) Steady Intensity (Intn) Factor. Used to fine tune the temperature profile
(1c) Processing time for each step
1b) Processing Steps: Delay, Intensity, Ramp, Steady,(...) Delay
22 12/2015 R. Ravichandran, D. Boyce
TEMPERATURE MONITORING
The AccuThermo RTP system can be equipped with two types of temperature monitoring devices: a thermocouple and an optical pyrometer. The thermocouple is used for controlling processes at temperatures below 550°C and starts to degrade and fuse to samples around 800°C. The optical pyrometer is the extended range pyrometer (ERP). Unlike earlier pyrometers used that were restricted to use at process temperatures above 800°C, the ERP can be used throughout the recommended process temperature range of 400°C to 1200°C.
Thermocouple
The AccuThermo RTP system only uses the K-type (NiCr/NiAl) thermocouple. When in
use, the thermocouple wires should not be touching each other, and the thermocouple bead is
the only point in which the thermocouple should be in contact with the wafer (Fig 4). If the two
different metal alloy wires are touching each other at any other point other than the wafer, it
will read an inaccurate temperature. This could lead to the temperature overshooting, wafers
melting, and damaged quartzware. Thermocouple voltages are fed to a temperature
compensation circuit. The thermocouple compensation circuit contains an analog linearizing
circuit with an output calibrated to 1 mv/°C. The signal is digitized and sent to the computer for
use by the temperature control system.
Figure 4. Top and side views of thermocouple and wafer placement in quartzware inside chamber
23 12/2015 R. Ravichandran, D. Boyce
CAUTION
Thermocouple may fuse into th e wafer at temp er atur es above 8 0 0 °C. As a result, the thermocouple should be removed and the pyrometer should be used for temperature measurement when the oven is to be operated at temperatures higher than 700°C.
Pyrometer
The Extended Range Pyrometer (ERP) is a dual infrared detector that provides accurate and consistent temperature measurement during rapid thermal processing of semiconductor wafers at higher temperatures (Fig 5). The ERP measures wafer temperature by detecting infrared light radiation emitted from the wafer within a specified wavelength range. The resulting signal is adjusted to compensate for the effects of non-wafer radiation sources.
Optical pyrometer readings are emissivity-dependent (see “Compensating for Emissivity”). This means that these readings are affected by changes in wafer backside surface color and roughness, deposited backside layers, or variations in substrate materials.
Chamber
Figure 5. Side view of RTP to show placement of pyrometers and wafers within the chamber. The extended range pyrometer can measure temperatures between 400°C to 1200°C using infrared light radiation.
24 12/2015 R. Ravichandran, D. Boyce
Compensating for Emissivity
When adjusting the emissivity values, make sure that the “Use Emissivity Local” button is active. If the “Use Emissivity System” is selected, it will use the default factory value. This means you will end up with the same pyrometer offset readings regardless of how you change the emissivity value on your recipe page.
When the pyrometer is used as the temperature sensor for temperature control, the
EMISSIVITY compensation must be set to suit the emissivity of the type of wafer being processed. This value compensates for the reflective properties that are different from the “standard” wafer. The standard wafer is usually pure silicon and it has an emissivity of 77.
To determine the correct EMISSIVITY setting, you will need the TC wafer thermocouple
assembly. The thermocouple wafer is used to indicate the true temperature in the oven while the oven is run under pyrometer control. The wafer is heated to the processing temperature that will be used. The output of the thermocouple is read by the control software.
When the steady-state temperature is reached, read the TC temperature, as this
indicates the “real temperature”. Compare this reading with the pyrometer-indicated temperature. If the readings differ, the EMISSIVITY must be adjusted, such that the pyrometer temperature matches the “real temperature” as read by the TC wafer. The heating cycle and adjustment are repeated until the pyrometer shows the same temperature as indicated by the TC wafer.
If the Pyrometer is less than the Real Temperature, then decrease the emissivity If the Pyrometer is greater than the Real Temperature, then increase the emissivity
25 12/2015 R. Ravichandran, D. Boyce
TROUBLESHOOTING THE TEMPERATURE CONTROL
This section will describe the common problems with temperature control and how to correct them. These problems are temperature oscillation, overshoot and undershoot. When diagnosing temperature control problems, refer to the process data graph and use the pink vertical line to help understand the situation. The pink vertical line can be moved with the left and right keyboard arrow keys and moving the mouse pointer to a position on the graph and clicking the left mouse button.
Oscillation Oscillation occurs when the wafer temperature does not stabilize within the steady state or ramp, as shown in figure 6. This problem is cause by the temperature control over compensating the intensity to try to get the wafer temperature to the set point.
During the Ramp step,
Figure 6. Oscillations
o short time oscillations are eliminated by decreasing GAIN. o long time oscillations are eliminated by increasing GAIN.
During the Steady step,
o short time oscillations are eliminated by decreasing SENSITIVITY. o long time oscillations are eliminated by increasing SENSITIVITY.
Remember, SENSITIVITY is dependent on the GAIN. Therefore, if the GAIN has been changed, SENSITIVITY may also have to be changed.
Be careful. The oscillation of the intensity needs to be controlled. If it spikes too much, the gain needs to be reduced.
26 12/2015 R. Ravichandran, D. Boyce
Overshoot An overshoot is when the wafer temperature exceeds the steady state temperature during a transition and then drops back down to the steady state set point, as shown in figure 7.
Figure 7. Overshoot
The following variables can be changed to compensate for an overshoot. Change only one variable at a time and run a few cycles. If the overshoot still occurs, change another variable.
If the temperature is oscillating during the Ramp Up, there would be a good chance an overshoot would occur. Decrease the GAIN until the oscillation is eliminated (refer to the oscillation section above).
If the overshoot is caused by the intensity being too high during the transition, decrease Steady-Intn-Factor for the RAMP step. The temperature at the end of the transition should be exactly the same as the setpoint for the Steady step.
If the overshoot is caused by the intensity being too high at the end of the transition, decrease Steady-Intn-Factor for the STEADY step.
If the overshoot cannot be controlled during the transition by decreasing the Steady- Intn-Factor for the RAMP step, the transition may be too short. Increase DELAY.
27 12/2015 R. Ravichandran, D. Boyce
Undershoot An undershoot occurs when the transition from Ramp to Steady-State is too gradual and the temperature takes too long to get to the Steady-State set point, as shown in figure 8.
Figure 8. Undershoot
The following variables can be changed to compensate for an undershoot. Change only one variable at a time and run a few cycles. If the undershoot still occurs, change another variable.
If the real-temperature is oscillating during the Ramp Up, there would be a good chance
an undershoot would occur. Decrease the GAIN until the oscillation is eliminated (refer to the oscillation section above).
If the real-temperature is sluggish during the Ramp Up and not reaching the model- temperature, increase the GAIN until the real-temperature follows the model- temperature.
If the undershoot is caused by the intensity being too low during the transition, increase Steady-Intn-Factor for the RAMP step. The temperature at the end of the transition should be exactly the same as the set point for the Steady step.
If the end of the transition reaches the set point temperature, but there is a dip at the beginning of the Steady step, the intensity is too low at the end of the transition. Increase Steady-Intn-Factor for the STEADY step.
If the Steady-Intn-Factor for the Ramp and Steady steps are large, then the DELAY may be too large. Decrease DELAY.
28 12/2015 R. Ravichandran, D. Boyce
THERMOCOUPLE REPLACEMENT
At temperatures above 800C the thermocouple will melt and can fuse to silicon wafers. Additionally, the thermocouple life can be shortened when used with oxygen gas flows. Therefore, it may be necessary to remove the thermocouple and only use the pyrometer at times, and replace the thermocouple back into the chamber once you have finished your process. Additionally, if there is overheating or the thermocouple becomes damaged, we have extra pieces on-hand. Please follow the procedure described to remove or install a thermocouple wire into the system.
Removal of thermocouple
1. Remove all samples from the process chamber. Shown: Empty quartzware with thermocouple and bead placed in the chamber.
Thermocouple bead that comes into
contact with sample in chamber
Quartz holder for
thermocouple wires
Thermocouple sensor screws
29 12/2015 R. Ravichandran, D. Boyce
2. Use the screwdriver to loosen screws highlighted in red below (~3 turns), until the wires can be removed from under the screw heads.
3. Remove Thermocouple from the processing chamber and set aside in a safe location for installation later. If the thermocouple is damaged please discard.
4. Tighten down screws highlighted in red. Make sure not to overtighten the sensor screws or you may crack the quartzware.
Installation of thermocouple
1. Open process chamber and remove all samples from quartzware. 2. Using magnetic tool distinguish the two wires.
a. Place the magnetic lead (NiAl) on the left screw head and the non- magnetic wire (NiCr) on the right screw head.
3. Loosen the screws just enough that the wires can be slipped between the screws and the leads for the machine. Bend the thermocouple wire around the screw. You don’t need to loop the wire around the screw in order to get accurate readings.
Magnetic (NiAl) wire
on left sensor lead
Non-magnetic (NiCr)
wire on the right lead
30 12/2015 R. Ravichandran, D. Boyce
4. Tighten screws and trim the excess wiring so as to not produce false reading 5. Ensure that you have installed the thermocouple correctly, run a simple recipe with a
dummy wafer in the chamber. a. If the temperature goes up then you may resume processing as normal. b. If the temperature goes down and even into the negative, stop the
process immediately and reverse the wiring of the thermocouple as soon as the chamber returns to room temperature.
Correct Wiring Signal
Incorrect Wiring Signal
31 12/2015 R. Ravichandran, D. Boyce
DATA EXTRACTION 1. Insert your flash drive into the computer. Note: The software will not recognize a flash drive that is larger than 2GB.
2. Reboot the computer using ctrl + alt + del This will allow the computer to read the flash drive and take you to the main menu
3. Press the green exit icon to leave the main menu
32 12/2015 R. Ravichandran, D. Boyce
4. Locate the letter assignment of your flash drive by inputting: dir D: a. If you can’t find your flash drive in dir D:, then use dir E: b. When you find the correct letter for your drive, press enter. Then reactivate the
system by typing rtapro into the command line after the “>”.
33 12/2015 R. Ravichandran, D. Boyce
5. Log back onto the software using the log-on key
6. Open process for production and select the data file or recipe that you wish to extract.
34 12/2015 R. Ravichandran, D. Boyce
7. Select the Transfer to Text File button
8. Appropriately name the new text file so that you may find it later
35 12/2015 R. Ravichandran, D. Boyce
9. Exit the program through the main screen. 10. Back up the drive onto the flash drive. Input the following commands as described with the
exception of name which will be determined by the user. a. Input the code: d: to select the flash drive b. Input: mkdir name with “name” being your user ID to create a new folder on that flash
drive i. If this is not your first time extracting
data you will need to delete the files in that directory using the command: del name. You will not lose any of your previous data as this will still be copied over during the back up.
c. Input: xcopy \rtapro\*.* d: \name\*.* “name” is the title of the new folder that you created in the drive
d. Wait for the program to completely finish copying the files over. You will receive a message similar to the screen below. This should not take more than 5 mins to complete
36 12/2015 R. Ravichandran, D. Boyce
11. Remove flash drive and restart the computer using ctrl+alt+del 12. Check for file on flash drive which can then be converted into an excel spreadsheet. 13. Converting .txt to .xlsx
a) Open excel and start the open dialog box b) From the drop down menu of file types select text file (.txt) c) You should see a conversion dialog box
d) Click delimited and then click next e) On the next screen choose that the data is comma separated
37 12/2015 R. Ravichandran, D. Boyce
f) Finally click finish when the data is sorted into the desired column