A NEW LC MS APPROACH FOR ENHANCING … in Milford, MA. Reagent grade DFA was purified via...
Transcript of A NEW LC MS APPROACH FOR ENHANCING … in Milford, MA. Reagent grade DFA was purified via...
TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS ©2018 Waters Corporation
INTRODUCTION
Protein reversed phase chromatography, while preferred
for LC-MS, is heavily dependent on the conditions under
which it is performed. Methods employing polymeric
columns and trifluoroacetic acid (TFA) have been
preferred by chromatographers but are inherently
restricted to low pressure, low throughput analyses and
compromised MS detection.
Accordingly, a novel platform for LC-MS has been
developed that includes three critical breakthroughs: a
new particle technology to afford increased throughput, a
unique high coverage phenyl surface to lessen ion
pairing dependence, and a more MS-friendly mobile
phase system based on highly purified difluoroacetic
acid (DFA). Together, these advances achieve the
simultaneous optimization of both LC and MS
capabilities, making it possible to characterize mAbs and
ADCs to an unprecedented level of detail.
A NEW LC-MS APPROACH FOR ENHANCING SUBUNIT-LEVEL PROFILING OF mAbs AND ADCs Jennifer Nguyen1, Jacquelynn Smith2, Olga V. Friese2, Jason C. Rouse3, Daniel P. Walsh1, Ximo Zhang1, Nilini Ranbaduge1, and Matthew A. Lauber1 1Waters Corporation, Milford, MA, USA, 2Biotherapeutics Pharm. Sci., Pfizer WRD, St Louis, MO, USA, 3Biotherapeutics Pharm. Sci., Pfizer WRD, Andover, MA, USA
METHODS
Reduced, IdeS digested NIST mAb was acquired in the form of the Waters mAb
Subunit Standard (p/n 186008927). Therapeutic monoclonal antibodies (mAbs) and
antibody drug conjugates (ADCs) (manufactured by Pfizer) were subjected to IdeS
digestion and reduction according to standard procedures and performed at Waters
Corporation in Milford, MA. Reagent grade DFA was purified via distillation. ICP
quantitation of metals was performed by EAG Laboratories. Various ion pairing
conditions and concentrations using DFA, TFA, and formic acid (FA) were
investigated along with separation temperature, flow rate, and alternative eluents
such as isopropanol to demonstrate the optimization of methods.
Analyses were performed using an ACQUITY UPLC H-Class Bio, ACQUITY UPLC
TUV detector, and Xevo G2-XS QToF mass
spectrometer. Waters mAb Subunit Standard
separations were performed on a 2.7 µm, 2.1 x 50 mm
BioResolve RP mAb Polyphenyl column. ADC
separations were performed at 80 °C or 70 °C on a 2.7
µm, 2.1 x 150 mm BioResolve RP mAb Polyphenyl or a
1.7 µm, 2.1 x 150 mm ACQUITY BEH C4 300 Å
column. Samples were run using 0.1% or 0.15%
DFA, TFA, or FA in water (mobile phase A) and
0.1% or 0.15% of the same modifier in
acetonitrile or 90/10 (v/v) acetonitrile/isopropanol
(mobile phase B). The gradient was run from 15
-55% in 20 min at a flow rate of 0.6 mL/min for 150 mm columns and 0.2 mL/min for
50 mm columns. Analyses were performed with UV detection at 280 nm using
MassLynx 4.1 and UNIFI 1.8. LC/MS analyses were performed in sensitivity mode
and optimized for the reduction of adducts. MaxEnt was used for deconvolution.
References
1. Smith, J., Friese, O., Rouse, J., Nguyen, J., Lauber, M. , and Jayaraman, P. Characterization of Hydrophobic Monoclonal Antibodies and Antibody Drug
Conjugates. Presented at WCBP 2018, Washington D.C., United States, January 28-31, 2018.
2. Nguyen, J. M.; Rzewuski, S.; Walsh, D.; Cook, D.; Izzo, G.; DeLoffi, M.; Lauber, M. A. Designing a New Particle Technology for Reversed Phase Separations of
Proteins. (2018). Waters Application Note (PN: 720006168EN).
3. Fekete, S.; Veuthey, J.; Guillarme, D. New trends in reversed-phase liquid chromatographic separations of therapeutic peptides and proteins: Theory and
applications. J. Pharm. Biomed. Anal. 2012, 69, 9-27. of therapeutic peptides and proteins: Theory and applications. J. Pharm. Biomed. Anal. 2012, 69, 9–27.
With LC-MS quality DFA and the novel polyphenyl solid-core stationary phase of the BioResolve RP mAb column, a new platform method has been developed for both mAb and ADC subunit separations. Figure 8 shows how this new platform method can be implemented to improve ADC subunit profiling, starting with a previous employed method and working toward a newly optimized technique performed with DFA mobile phases and a 2.7 µm, 2.1 x 150 mm BioResolve RP mAb Polyphenyl column to give better MS sensitivity, better resolution and higher recovery of the Fd’(3) sepcies at both high (80 °C) or lowered temperatures (70 °C), such that on-column degradation can also be minimized.
CONCLUSION
A new platform technology has been developed to enhance LC-MS subunit profiling of mAbs and ADCs that is based on the use of highly purified DFA as a mobile phase ion pairing agent and the novel BioResolve RP mAb Polyphenyl stationary phase.
LC-MS quality DFA has been successfully prepared as verified through ICP metal quantitation and LC-MS application testing.
DFA confers notable gains in MS sensitivity versus TFA, even at higher concentrations, while providing comparable (and sometimes better) resolution.
An optimized concentration of DFA can also increase the recovery of challenging protein analytes, as has been exemplified with the increased recovery of a three payload Fd’(3) subunit encountered in the analysis of an ADC.
LC-MS quality DFA was then used to improve the subunit separation of an ADC sample. High resolution
separations with favorable baseline properties were achieved by using 0.15% DFA, as is highlighted with the
zoomed chromatograms of Figure 5. Moreover, it was found that a 0.15% DFA modified mobile phase also
gave the highest recovery of the challenging to recover three payload Fd’(3) subunit.
The raw MS spectra in Figure 6 indicated that, as predicted, a FA mobile phase gives the highest sensitivity
among the various ion-pairing agents. Interestingly, 0.15% DFA gives the next highest signal intensity. For
this comparison, there was, in fact, only a 15% decrease in sensitivity versus using 0.1% FA. Observations on
charge state distributions are equally noteworthy. That is, there is actually only a slight shift in average charge
state in comparing the spectra obtained with FA versus DFA versus TFA.
Deconvoluted MS spectra for the unmodified light chain [LC(0)] of the ADC are shown in Figure 7. The spectra
resulting from DFA modified mobile phase showed less sodium (Na) and potassium (K) adducts than FA which,
as a lower strength ion pairing agent, may more readily allow the undesirable creation of adducts. The TFA
separation also shows low metal content but gives a distinct signal for a gas phase TFA ion pair. With DFA
separations, there is little to no sign of a DFA gas phase ion pair. Use of a 10 eV collision energy ensures that
any DFA gas phase ion pairs that are formed upon ionization are eliminated upon low energy collisional
activation.
RESULTS AND DISCUSSION
It has been observed that the use of DFA in place of TFA can actually
afford higher chromatographic resolution, as exemplified in a separation
of NIST mAb subunits (Figure 2).
Moreover, versus TFA, DFA has been confirmed to yield higher
sensitivity MS detection of proteins due to its lower ion-pairing strength
and reduced ion suppression (Figure 3).
These preliminary data clearly indicated that there could be significant
promise in the use of DFA for protein LC-MS. However, two concerns
remained that had to be addressed if new methods based on DFA were
to be developed: 1) that there is no commercially available MS-quality
DFA and 2) that it might be necessary to optimize MS settings
specifically for DFA based mobile phases.
The issue of DFA purity and MS suitability was solved by performing
multiple distillations to produce a low metal content, LC-MS grade form
of the reagent. ICP quantitation of metals (Table 1) shows the
successful reduction of metals in DFA to make it comparable to LC-MS
quality FA and TFA reagents that are currently commercially available.
Figure 4 shows the benefit to using an LC-MS quality DFA reagent in
combination with optimized MS settings. A focus is made on the
deconvoluted mass spectrum of the NIST mAb light chain (LC) to
highlight the minimization of adduct levels. These spectra show that an
optimized MS system is also extremely important to minimizing metal
adducts.
A
0.1% DFA0.1% FA 0.1% TFAPc*=105Pc*=114Pc*=71
DFA0.1%
TFA FA
~2x
0.1% 0.1%
Figure 2. UV chromatograms of NIST mAb subunits separated using a
2.1 x 50 mm BioResolve RP mAb Polyphenyl column, 0.1% modified
mobile phases, a flow rate of 0.2 mL/min and temperature of 80°C.
Figure 3. MS sensitivity compari-
son of 0.1% modified mobile
phase additives. These results
were obtained from the analysis
of IdeS digested, reduced subu-
nits of NIST mAb using a QDa
mass detector.
Table 1. ICP quantitation of metals for DFA, distilled DFA, LC-MS grade
FA, and LC-MS grade TFA shows the effectiveness of using distillation
to produce an LC-MS quality DFA reagent.
Conc. (ppm)
Commercially Available
DFA
LC-MS Quality DFA*
LC-MS Grade FA
LC-MS Grade TFA
Na 41 <1 <1 3
Mg 2 <0.1 <0.1 0.3
K 2 <1 <1 <1
Ca 11 <1 <1 1
Fe 2 <1 <1 <1
Figure 4. Deconvoluted spectra of the NIST mAb LC subunit as obtained
with a 2.1 x 50 mm BioResolve RP mAb Polyphenyl column, 0.1% DFA
modified mobile phases, and Xevo G2-XS QTof mass spectrometer.
*Patent Pending
0.1% FA
Pc*=102.8
0.1% DFA
Pc*=158.2
0.15%
DFA
Pc*=158.4
0.1% TFA
Pc*=149.9
% Peak Area
Fd’(3) = 2.6
% Peak Area
Fd’(3) = 4.0
% Peak Area
Fd’(3) = 5.2
% Peak Area
Fd’(3) = 4.2
scF
c(0
)
LC
(0)
LC
(1)
Fd’(0
) Fd’(1
a)
Fd’(3
)
Fd’(1
a)
min
us H
2O
Fd’(2
b)
Fd’(2
a)
scF
c(0
) D
/P
scF
c(0
)
LC
(0)
LC
(1)
Fd’(0
) Fd’(1
a)
Fd’(3
)
Fd’(1
a)
min
us H
2O
Fd’(2
b)
Fd’(2
a)
scF
c(0
) D
/P
scF
c(0
)
LC
(0)
LC
(1)
Fd’(0
)
Fd’(1
a)
Fd’(3)Fd’(1
a)
min
us H
2O
Fd’(2
b)
Fd’(2
a)
scF
c(0
) D
/P
scF
c(0
)
LC
(0)
LC
(1)
Fd
’(0
) Fd’(1
a)
Fd’(3
)Fd’(2
b)
Fd’(2
a)
Fd’(1a) m
inus H
2O
Figure 5 . UV chromatograms of ADC subunits with zoomed views as
obtained with various acid modified mobile phases using a BioResolve
RP mAb Polyphenyl 2.1 x 150 mm column, flow rate of 0.6 mL/min, and
temperature of 80 °C.
Pc* = 138.1
% Peak Area
Fd’(3) = 4.3
B1.7 µm, 2.1 x 150 mm ACQUITY BEH C4 300 Å
0.1% TFA at 80 °C with 10% IPA in Mobile Phase B
2.7 µm, 2.1 x 150 mm BioResolve RP mAb Polyphenyl
0.1% TFA at 80 °C with 10% IPA in Mobile Phase B
A
Pc* = 152.5
% Peak Area
Fd’(3) = 4.3
2.7 µm, 2.1 x 150 mm BioResolve RP mAb Polyphenyl
0.1% TFA at 80 °C without IPA in Mobile Phase B
Pc* = 153.8
% Peak Area
Fd’(3) = 4.2
2.7 µm, 2.1 x 150 mm BioResolve RP mAb Polyphenyl
0.15% DFA at 80 °C without IPA in Mobile Phase B
Pc* = 158.4
% Peak Area
Fd’(3) = 5.2
Pc* = 134.3
% Peak Area
Fd’(3) = 4.2
Pc* = 142.1
% Peak Area
Fd’(3) = 5.4
DC
FE
Starting Point
New Platform Method Control1.7 µm, 2.1 x 150 mm ACQUITY BEH C4 300 Å
0.15% DFA at 70 °C without IPA in Mobile Phase B
2.7 µm, 2.1 x 150 mm BioResolve RP mAb Polyphenyl
0.15% DFA at 70 °C without IPA in Mobile Phase B
0.1% FA
1.29e6
0.1% DFA
5.23e5
0.15% DFA
9.85e5
0.1% TFA
6.61e4
3.24%
+Na2.3%
+K
3.07%
Glycation
+ 162 Da
5.47%
-H2O
1.74%
+Na 0.3%
+K
0.33%
Glycation
+ 162 Da
5.98%
-H2O
4.77%
-H2O 1.83%
+TFA1.16%
+Na 1.09%
+K
0.36%
Glycation
+ 162 Da
5.92%
-H2O
0.57%
+Na 0.2%
+K
0.1% FA
2.59e5
0.1% DFA
1.46e5
0.1% TFA
6.25e4
0.15% DFA
2.22e5
Figure 6. Raw mass spectra of the of the unmodified light chain [LC(0)] of the ADC using various modified mo-
bile phases and optimized MS settings using a BioResolve RP mAb Polyphenyl 2.1 x 150 mm column and a
Xevo G2-XS QTof mass spectrometer.
Figure 7 . Deconvoluted MS spectra of the unmodified light chain [LC(0)] of the ADC as obtained with various
acid modified mobile phases and optimized MS conditions using a BioResolve RP mAb Polyphenyl 2.1 x 150
mm column and Xevo G2-XS QTof mass spectrometer.
Figure 8. UV spectra of the ADC subunit separations showing the incremental improvements in performance
upon changing from the previous platform method (0.1% TFA modified mobile phase using a BEH C4 column)
to the new platform method (based on 0.15% DFA and a BioResolve RP mAb Polyphenyl column), in alphabeti-
cal order from A to F. Panel F provides a control comparison to show the enablement of the method via the use
of a BioResolve RP mAb Polyphenyl column.
Figure 1. A) Schematic representation of the BioResolve RP mAb Poly-
phenyl 450Å 2.7 µm stationary phase. B) Chemical structure of DFA.
MS
Initial Conditions
Desolvation Gas Flow (L/h): 600
Collision Energy: 6 eV
Final Conditions
Desolvation Gas Flow (L/h): 100
Collision Energy: 10 eV
Reag
en
t Q
uali
ty D
FA
LC
-MS
Qu
ali
ty D
FA
11.2%
+Na
10.0%
+K
7.02%
-H2O
1.98e6
2.89%
Glycation
6.71%
+Na
6.82%
+K
7.19%
-H2O
3.07%
Glycation
9.60e5
10.08%
+Na
7.7%
+K
6.78%
-H2O 2.39%
Glycation
2.59e6
2.46%
+Na
1.95%
+K
7.28%
-H2O
2.96%
Glycation
1.20e6
B
Fc/2
LC
Fd’