are in search mode. Auto Connect Before using Auto Connect, be sure no other nearby Medtronic devices are in search mode. Cancel Continue Use the arrow buttons to scroll to Continue. Press the Select button . Menu Insulin Settings Sensor Settings Event Markers Reminders Utilities Utilities Self Test Carb Unit Manage Settings Sensor Demo Device Options Device Options Manage Devices Connect Device Active Insulin 1.5 U Bolus Basal mmol/L BG Search New Device See User Guide to connect new device. Select Search below when ready. Confirm Device SN BGJ133333F found: SN: BGJ133333F Cancel Confirm Connect Device Manual Connect Auto Connect 5.2 mmol/L Active Insulin 1.5 U Bolus Basal BG 12:00 PM 12:00 PM CONNECT TO PUMP CONNECT TO PUMP ? OK OK Skip Skip ? OK OK Go Back Go Back Connect to a MiniMed Pump? Select “Connect Device” on your Pump. CONNECT TO PUMP CONNECT TO PUMP Cancel Cancel Searching... Turn the page over to complete setup. ® QUICK REFERENCE GUIDE 5.2 mmol L Before Meal Before Meal 12:00PM l 10/13 12:00PM l 10/13 Bolus Bolus Reminder Reminder Notes Notes Uses only CONTOUR ® NEXT Blood Glucose Test Strips CONNECT TO PUMP CONNECT TO PUMP 5.2 mmol L Before Meal Before Meal 12:00PM l 10/13 12:00PM l 10/13 Bolus Bolus Reminder Reminder Notes Notes Initial Setup of Meter and Connection to MiniMed ® 630G Pump 1 Remove the USB cap to reveal the USB connector. Charging Meter 2 To begin setup, hold down the top Menu button for several seconds until the meter turns on. Unplug Meter From Charging Source Before Beginning Setup Connect to Pump 3 Scroll down to select your language. Press OK. 5 Connect to a MiniMed Pump? Press OK. 6 Press OK. 7 Put the meter down and pick up your pump. Press any button to turn the screen on, then press the Menu button . Use the down arrow to scroll to Utilities. Press the Select button . 8 In the Utilities menu, use the down arrow to scroll to Device Options. Press the Select button . 9 In the Device Options menu, scroll to Connect Device. Press the Select button . 10 Place the meter and pump next to each other. Select Auto Connect on your pump. 11 Read the pump screen, using the down arrow to scroll to the next page. 12 Place the meter and pump next to each other. Select Search on the pump. Search Search Cancel Cancel CONNECT TO PUMP CONNECT TO PUMP Select “Search” on your Pump and meter. The search may take up to 2 minutes. 13 Check to see if Device SN (serial number) on the pump matches the Device SN on the meter. If they match, scroll to Confirm on the pump. Press the Select button . CONNECT TO PUMP CONNECT TO PUMP Use Pump to confirm Device SN: BGJ133333F 14 Turn the pump over. Make sure that the SN on the back of the pump matches the pump serial number now on the meter screen. Press Next. ? Next Next Delete Delete MiniMed Pump connected: AA12345678 15 Select Always and press OK on next screen. Always Always Ask Me Ask Me Never Never Automatically send results to the Pump? SEND OPTIONS SEND OPTIONS 16 Select the Date Format. Month/Day/Year Month/Day/Year Day.Month.Year Day.Month.Year SELECT SELECT DATE FORMAT DATE FORMAT ? 17 Press OK. Note: If time and date are incorrect, you must change them on the pump. OK OK Date / Time is set by Pump. DATE / TIME DATE / TIME ? 18 Press Accept to select AutoLog is Off. AutoLog allows you to mark a test result as Fasting, Before Meal, or After Meal. See your user guide to learn more. Accept Accept Turn On Turn On AutoLog is Off. AUTOLOG AUTOLOG ? Wireless Blood Glucose Monitoring System ? Manual Connect Manual Connect Auto Connect Auto Connect CONNECT TO PUMP CONNECT TO PUMP 4 Press OK if the correct language is displayed. Press NO if the incorrect language is displayed. * Wall charger may not be included in all meter kits. Contact Customer Service for information on obtaining a wall charger. If pump screen goes blank, press any button to turn it back on. When the pump links to the meter, the SN on the meter screen will change to a different serial number. This number is the pump SN. PUMP CONNECTED PUMP CONNECTED ® Select Auto Connect on your meter. Select Search on the meter. 10:30PM | 10/13/11 English Deutsch Español LANGUAGE LANGUAGE OK OK 81552932 Rev. 09/16 Display Screen The CONTOUR ® NEXT LINK 2.4 meter has a permanent rechargeable battery. Charge your meter with the wall charger * or by plugging the USB connector into a computer. Be sure the computer is turned on. Your meter will briefly display “Do Not Test–Charging” and the test strip port light will flash. You cannot do a blood glucose test while the battery is charging. When charging is finished, the test strip port light will turn off. Unplug your CONTOUR NEXT LINK 2.4 meter to test. USB Connector Selection/ Scrolling Buttons Test Strip Port Display Screen Test Strip Port Light Top Menu Button Protective USB Cap Meter Pump Menu Select Right Up Back Notification Light Left Down NOTE: Check the units of measure of your meter. See your meter user guide for complete instructions. p, e ton the n. From Charging Beginning Se Setu tup p ng is finished, the test strip port light will ug your CONTOUR NEXT LINK 2.4

Transcript of 81552932 CntrNextLINK2 4 QRG EN FpBp v9 · 2017-05-24 · Connect Device Active Insulin 1.5 U Bolus...

Page 1: 81552932 CntrNextLINK2 4 QRG EN FpBp v9 · 2017-05-24 · Connect Device Active Insulin 1.5 U Bolus Basal mmol/L BG Search New Device See User Guide to connect new device. Select

DATE:CLIENT:

DESCRIPTION:

SCHAWK JOB#:SKU#:

DIMENSIONS:PRINTER SPEC:

RELEASE: V-9 LANGUAGE(S): ENGLISHINTERNAL: R-X

18211 NE 68th Street, E120 Redmond, WA 98052 T: 425-881-5454

BAN#:81552932 Rev. 09/16Contour Next LINK 2.4 QRG Canada

September 23, 2016Ascensia Diabetes CareContour Next LINK 2.4 QRG Canada English9203589755 Parent BAN# 8155891416.5” (H) x 8” (W)10000752618 REV -

FONTS: Arial, Helvetica, Helvetica Neue, Zapf Dingbats, Skinny Mini, Mini Micra, Myriad ProCOLORS: Cyan Magenta Yellow Black 281 Process Build: 100C 79M 10Y 42K

NOTE: THIS COLOR PROOF INDICATES COLOR BREAK ONLY AND MAY NOT ACCURATELY REFLECT ACTUAL PRODUCTION COLOR.Regional Master Template No: M-QRG-0347-R0002Global / EU Master Template No: M-QRG-0304-R0003

Alt#1 - 04/15/14 Copy ChangesAlt#2 - 12/01/15 Copy ChangesAlt#3 - 12/10/15 Copy ChangesAlt#4 - 12/18/15 Copy ChangesAlt#5 - 07/08/16 Copy ChangesAlt#6 - 08/17/16 Copy ChangesAlt#7 - 09/08/16 Copy Changes

Alt#8 - 09/15/16 Copy ChangesAlt#9 - 09/23/16 Copy Changes

are in search mode.

Auto ConnectBefore using AutoConnect, be sure no other nearby Medtronic devices are in search mode.

Cancel Continue

Use the arrow buttons to scroll to Continue.

Press the Select button .

MenuInsulin SettingsSensor SettingsEvent MarkersRemindersUtilities

UtilitiesSelf TestCarb UnitManage SettingsSensor DemoDevice Options

Device OptionsManage DevicesConnect Device

Active Insulin 1.5 UBolus Basal

mmol/L

BG

Search

New DeviceSee User Guide to connect new device. Select Search below when ready.

Confi rm Device SNBGJ133333F found: SN:BGJ133333F

Cancel Confi rm

Connect DeviceManual ConnectAuto Connect

5.2 mmol/LActive Insulin 1.5 UBolus Basal

BG

12:00PM

12:00PM

CONNECT TO PUMPCONNECT TO PUMP

?OKOK

SkipSkip

?OKOK

Go BackGo Back

Connect to a MiniMed Pump?

Select “ConnectDevice” on your Pump.

CONNECT TO PUMP CONNECT TO PUMP

CancelCancel

Searching...

Turn the page over to complete setup. �

®

QUICK REFERENCE GUIDE

5.2 mmolL Before Meal

Before Meal

12:00PM l 10/13

12:00PM l 10/13

BolusBolusReminderReminderNotesNotes

Uses only CONTOUR®NEXT Blood Glucose Test Strips

CONNECT TO PUMPCONNECT TO PUMP

5.2

mm

olL

Bef

ore

Mea

l B

efor

e M

eal

12:0

0P

M l

10

/13

12:0

0P

M l

10

/13

Bolu

sBo

lus

Rem

inde

rRe

min

der

Note

sNo

tes

Initial Setup of Meter and Connection to MiniMed® 630G Pump

1 Remove the USB cap to reveal the USB connector.

Charging Meter

2 To begin setup, hold down the top Menu button for several seconds until the meter turns on.

Unplug Meter From Charging Source Before Beginning Setup

Connect to Pump

3 Scroll down to select your language.

Press OK.

5 Connect to a MiniMed Pump? Press OK.

6 Press OK.

7 Put the meter down and pick up your pump.

Press any button to turn the screen on, then press the Menu button .

Use the down arrow to scroll to Utilities.

Press the Select button .

8 In the Utilities menu, use the down arrow to scroll to Device Options.

Press the Select button .

9 In the Device Options menu, scroll to Connect Device.

Press the Select button .

10 Place the meter and pump next to each other.

Select Auto Connect on your pump.

11 Read the pump screen, using the down arrow to scroll to the next page.

12 Place the meter and pump next to each other.

Select Search on the pump.

SearchSearch

CancelCancel

CONNECT TO PUMPCONNECT TO PUMP

Select “Search” on your Pump and meter.

The search may take up to 2 minutes.

13 Check to see if Device SN (serial number) on the pump matches the Device SN on the meter.

If they match, scroll to Confirm on the pump.

Press the Select button .

CONNECT TO PUMP CONNECT TO PUMP Use Pump to confirm Device SN:BGJ133333F

14 Turn the pump over.

Make sure that the SN on the back of the pump matches the pump serial number now on the meter screen.

Press Next.?

NextNext

DeleteDelete

MiniMed Pump connected:AA12345678

15 Select Always and press OK on next screen.

AlwaysAlways

Ask MeAsk Me

NeverNever

Automatically send results to the Pump?

SEND OPTIONSSEND OPTIONS

16 Select the Date Format.Month/Day/YearMonth/Day/Year

Day.Month.YearDay.Month.Year

SELECTSELECT

DATE FORMATDATE FORMAT ?

17 Press OK.

Note: If time and date are incorrect, you must change them on the pump.

OKOKDate / Time is set by Pump.

DATE / TIMEDATE / TIME?

18 Press Accept to select AutoLog is Off.

AutoLog allows you to mark a test result as Fasting, Before Meal, or After Meal. See your user guide to learn more.

AcceptAccept

Turn OnTurn On

AutoLog is Off.AUTOLOGAUTOLOG

?

Wireless Blood Glucose Monitoring System

?

Manual ConnectManual Connect

Auto ConnectAuto Connect

CONNECT TO PUMPCONNECT TO PUMP

4 Press OK if the correct language is displayed. Press NO if the incorrect language is displayed.

*Wall charger may not be included in all meter kits. Contact Customer Service for information on obtaining a wall charger.

If pump screen goes blank, press any button to turn it back on.

When the pump links to the meter, the SN on the meter screen will change to a different serial number. This number is the pump SN.

PUMP CONNECTEDPUMP CONNECTED

®®

Select Auto Connect on your meter.

Select Search on the meter.

10:30PM | 10/13/11

EnglishDeutschEspañol

LANGUAGELANGUAGE

OKOK

81552932

R

ev.

09/1

6Display Screen

The CONTOUR®NEXT LINK 2.4 meter has a permanent rechargeable battery.

Charge your meter with the wall charger * or by plugging the USB connector into a computer. Be sure the computer is turned on.

Your meter will briefly display “Do Not Test–Charging” and the test strip port light will flash. You cannot do a blood glucose test while the battery is charging.

When charging is finished, the test strip port light will turn off. Unplug your CONTOUR NEXT LINK 2.4 meter to test.

USB Connector

Selection/Scrolling Buttons

Test Strip Port

Display Screen

Test Strip Port

Light

Top Menu

Button

Protective USB Cap

Meter Pump

Menu

Select

Right

Up

Back

Notification Light

Left

Down

NOTE: Check the units of measure of your meter. See your meter user guide for complete instructions.

p, e ton

the n.

From Charging Beginning SeSetutupp

ng is finished, the test strip port light willug your CONTOUR NEXT LINK 2.4

81552932_CntrNextLINK2_4_QRG_EN_FpBp_v9 1 09/23/16 9:27 AM

ENJRK
New Stamp
Page 2: 81552932 CntrNextLINK2 4 QRG EN FpBp v9 · 2017-05-24 · Connect Device Active Insulin 1.5 U Bolus Basal mmol/L BG Search New Device See User Guide to connect new device. Select

WARNING:• The lancing device is intended for self-testing by a single patient. It must not be used on more than one person

due to the risk of infection.

• Use a new lancet each time you test because it is no longer sterile after use.

• All parts of the kit are considered biohazardous and can potentially transmit infectious diseases, even after you

have performed cleaning and disinfection.

• Always wash your hands well with soap and water before and after testing, handling the meter, lancing device,

or test strips.

For complete instructions on cleaning and disinfecting your meter, please see your meter user guide.

Press Send or Don't Send.

BG

9.3 mmol/L

Calibrate SensorBolus

Done

The lancing device graphics are for illustration purposes only. Your lancing device may look different. Refer to your lancing device insert for detailed instructions on preparing the lancing device.

9.3 mmolL

SentSent

10:30PM l 11/1310:30PM l 11/13

9.3 mmolL

YOUR RESULTSYOUR RESULTS

BolusBolus

Reminder Reminder

NotesNotes

9.3 mmol/L

Calibrate SensorBolus

Done

cing device.

2 Remove the lancing device endcap.

Insert the lancet firmly into the lancing device until it comes to a full stop.

1 Insert the test strip and wait for the APPLY BLOOD message.

3 Replace the endcap. Rotate the endcap dial to adjust the puncture depth.

4 Obtain your blood drop.

5 Touch the tip of the test strip to the drop of blood. If the first blood drop is not enough, the meter will beep twice and display STRIP UNDERFILLED. You have about 30 seconds to apply more blood to the same strip.

7 Select Bolus to send a bolus to your pump. Select Reminder to set a reminder for your next test. Select Notes to add more information to the result.

9 Remove the test strip and the meter will turn off in 3 minutes. Or hold down the Menu button to turn off the meter.

8 Pick up your pump.

Press and hold the Back

button to get to the home screen. Your blood sugar result should display on the pump screen.

Obtain Blood Drop

Testing Blood Drop

Bolus, Reminder and Notes

Turn Meter Off

Confirm That Result Was Sent to Pump

See your CONTOUR®NEXT LINK 2.4 user guide, lancing device insert and MiniMed pump user guide for full instructions.

6 If Always was selected in Send Options during Setup, results will be sent directly to your pump. If Ask Me was selected in Send Options during Setup, you can choose to send your result to your pump after each blood sugar test.

Testing Your Blood

Sending Results to Your MiniMed Pump

20 Accept or Change your Target range.

The Target range is your goal for your blood sugar test results.

?AcceptAccept

ChangeChange

TARGETTARGET

3.9 - 10.03.9 - 10.021 Your system setup is now

complete and ready to run a blood sugar test.

Go to Step 1 below.

Setup is complete.Insert strip to start a test.

?

SendSend

Don't SendDon't Send

19 Accept or Change High and Low Alerts.

The meter alerts you with a sound when your blood sugar level is above or below your set levels.

AcceptAccept

ChangeChange

?HIGH AND LOW ALERTSHIGH AND LOW ALERTS

High: 13.9 High: 13.9 mmol/Lmmol/L

Low: 3.9 Low: 3.9 mmol/Lmmol/L

NOTE: If AutoLog is ON, you are asked to mark your result as Fasting, Before Meal, or After Meal before your result displays. If your result is above your High Alert setting or below your Low Alert setting, you will see your result without having to make an AutoLog selection.

NOTE: Alerts can only be set outside your selected Target Range(s). Targets can only be set inside your selected Alert level(s). For additional information, you may contact Customer Service. See contact information for Customer Service below.

CAUTION: Always check the display on your pump to ensure that the glucose result shown agrees with the glucose result shown on the meter.

For more information, go to www.diabetes.ascensia.com or contact Customer Service at: 1-800-268-7200

For In Vitro Diagnostic Use

Ascensia, the Ascensia Diabetes Care logo, Clinilog, Contour, Glucofacts, and the No Coding logo are trademarks and/or registered trademarks of Ascensia Diabetes Care Holdings AG.

TM see www.ascensiadiabetes.ca/en/tm-mcwww.diabetes.ascensia.com

Bolus Wizard, CareLink, Guardian, MiniMed, and Paradigm are trademarks and/or registered trademarks of Medtronic MiniMed, Inc.

© 2016 Ascensia Diabetes Care Holdings AG. All rights reserved.

For information on patents, and related licenses, see:

www.patents.ascensia.comAscensia Diabetes Care Holdings AGPeter Merian-Strasse 904052 Basel, Switzerland

Manufactured for:

Lancing Device CONTOUR®NEXT

Test Strip

Lancet

CONTOUR®NEXT LINK 2.4 Blood Glucose Meter