5702_Error lists Troubleshooting.pdf
-
Upload
jose-yeisin-hamburger-rivera -
Category
Documents
-
view
348 -
download
2
Transcript of 5702_Error lists Troubleshooting.pdf
3Error lists
/Troubleshooting
3-1
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
Note: R = Response of centrifuge when the error is reported.
Display: Cause and response of centrifuge.The consecutive number (cause of error) provides for more detailed localization of the error. It can be read fromservice program 8 (5702) or service program 9 (5702 R / 5702 RH).
After power-off the display shows "OFF". The time before the display is cleared depends on the operational status (centrifuge accelerating, braking or stopped).
All operational states can only be reset if the "OFF" display is cleared between power-off and power-on ! Only then does the controller perform a reset.
Note: The error display shows Er 3, 4, etc.Er 3-0, 3-1 only appears when the "speed" dial is held down for six seconds and thenturned clockwise/counterclockwise.
Error Cause Remedy
Er 2 Imbalance switch, at standstill, defective.
R: Centrifuge will not start.
Check switch, cable, plug.
– Check/adjust imbalance switch with service display SE 3.
– Load rotor symmetrically.
– Check switch, cable, connector.
– Replace Control PCB.
Er 3 No rotational speed pulses during running or when starting.Speed measurement system defective.Drive defective.The rotational speed is measured at the motor.Error remains active after power is cut.
R: The centrifuge decelerates unbraked. The centrifuge remains locked for 5 minutes.The lid cannot be opened until the error display is cleared.
Tubes collide with lid of centrifuge during start.
– Leave centrifuge switched on for > 5 minutes. Start again.
– Check motor speed sensor.
– Replace Converter PCB.
– Check tubes. Tubes too long or wrong adapter used?
– Check plug connections.
Er 3-0 When the centrifuge is powered up Err. 3 is signaled. See Er 3.
Er 3-1 No pulses after starting centrifuge.Centrifuge brakes.
See Er 3.
Error list
Reset
Error list
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-2
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
Er 3-2 Pulse failure during running.The centrifuge remains locked for 5 min.
See Er 3.
Er 5 No acknowledge message from lid latch while centrifuge is running.
R: The centrifuge decelerates unbraked.
– Check lid latch function with Service program 3.
– Check/adjust/replace microswitches.
– Check/replace reed magnet/sensor.
– Check/replace Control PCB.
– Check/replace motor lid latch.
Er 5-1 When the centrifuge is powered up the rotor rotates.The lid switch is open.
See Er 5.
Er 5-2 Rotor rotates. The lid switch is open See Er 5.
Er 5-3 No lid latch acknowledge signal. See Er 5.
Er 6 Fault in Converter PCB, thermal overload.
R: The centrifuge decelerates unbraked.
– Allow centrifuge to cool and restart.
– Determination of converter error code with customer service indicators.
– Check/replace Converter PCB.
– Check/replace braking resistor.
– Motor sluggish / check / replace.
Er 6-1 Converter PCB not ready. – See Er 6.
– Check connection between control PCB and Converter PCB.
Er 6-2 The Converter PCB has reported an error (error code set). See Er 6.
Er 6-4 Converter PCB has performed a reset (mains power interrupted).
– Check mains power connection.
– Check/replace fuses.
Er 6-5 Wrong Converter PCB ID after power-up. See Er 6.
Error Cause Remedy
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-3
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
Er 6-6 Safety circuit for release of Converter PCB defective.Motor is too hot.
See Er 6.
Er 7 Overspeed control.
R: Centrifuge decelerates unbraked.
– Check mains voltage.
– Check smooth running of motor.
– Check/replace motor.
– Replace Converter PCB.
Er 7-1 Maximum rotor speed exceeded. See Er 7.
Er 7-2 5702: Speed deviation ±1000 rpm from nominal.5702 R: Speed deviation ± 600 rpm from nominal.5702 RH: Speed deviation ± 600 rpm from nominal.
See Er 7.
Er 8 Ramp error during acceleration.
R: Centrifuge brakes.
– Check concentric running of motor / replace.
– Motor sluggish, check / replace.
– Replace Converter PCB.
Er 8-1 Time error. See Er 8.
Er 9 EEPROM write error.
R: No storage of running data.
– Initialize EEPROM in service program 2.
– Replace control PCB.
Er 9-1 Save running data. See Er 9.
Er 9-3 Initialization. See Er 9.
Er 9-4 Save operating hours. See Er 9.
Er 10 EEPROM read error.
R: Set defaults, data of last run are lost.
See Er 9.
Er 10-2 Read last nominal values. See Er 9.
Error Cause Remedy
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-4
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
Er 10-5 Read operating hours. See Er 9.
Er 11 Mains power interrupted.
Reset by internal software or hardware.
R: Centrifuge decelerates unbraked.
– Check mains power connections.
– Check ground connections.
– Check connection from Converter PCB to control PCB.
– If recurring, replace control PCB.
– Replace Converter PCB as necessary.
Er 14 EPROM checksum error after power-on. Error in program memory, EPROM defective.
R: No further operation possible.
Replace control PCB.
Er 15In b
Imbalance shutdown occured during a run.
R: Centrifuge decelerates unbraked.
See Er 2.
Er 16 Error in inquiry to Converter PCB.
R: Centrifuge shuts down and decelerates unbraked.
See Er 3.
Er 16-2 Error after telegram repetition. See Er 3.
Er 17 Error transmitting select telegram to Converter PCB.
R: Centrifuge shuts down and decelerates unbraked.
See Er 6.
Er 17-1 Converter PCB-telegram queue overflow. See Er 6.
Er 17-2 Error after three repeats of a select telegram. See Er 6.
Error Cause Remedy
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-5
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
Er 18(5702 R / 5702 RH only)
Excess temperature in rotor chamber.
R: Centrifuge shuts down and decelerates unbraked.
– Room temperature too high?
– Direct sunlight?
– Sufficient space for ventilation?
– Condenser fan clogged by dust?
– Service program SE 10,check temperature sensors at rotor chamber and condenser.
– Check all connections.
– Check/replace Converter PCB.
Er 18-1(5702 R / 5702 RH only)
After reaching the temperature, the ± 5 °C interval is left. See Er 18.
Er 18-2(5702 R / 5702 RH only)
Temperature in rotor chamber > +12 °C above nominal temperatur.
See Er 18.
Er 18-3(5702 R / 5702 RH only)
Temperature in rotor chamber > 50 °C. See Er 18.
Er 19(5702 only)
Mains interruption. See Er 25.
Er 19(5702 R / 5702 RH only)
Excess temperature at condenser.
R: Cooling unit is switched off, condenser fan keeps running.
– Allow cooling unit to cool down, check air inlet
– Sufficient space to the wall, to other devices?
– Does the condener fan also work when the cooling aggregate is running? Replace, if required.
– Check service program SE 10.
– Check all connections.
Er 19-1(5702 R / 5702 RH only)
Temperature measured > 62 °C. See Er 19.
Er 20(5702 R / 5702 RH only)
Temperature sensor at rotor chamber defective.
R: Centrifuge shuts down, braked deceleration, cooling unit switches off.
– Check with service program SE 10.
– Replace temperature sensor at rotor chamber.
Error Cause Remedy
MSTM/KW/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-6
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
Er 21(5702 R / 5702 RH only)
Temperature sensor at condenser defective.Measured value outside –8 °C.....+83 °C
R: Centrifuge shuts down, braked deceleration.
– Check service programSE 10.
– Replace temperature sensor at condenser.
Er 22(5702 R / 5702 RH only)
Mains frequency for condenser fan faulty on Converter PCB.
R: Fan is switched on or off.Centrifuge can be operated.Error is displayed in standstill only.
Replace Converter PCB.
Er 23(5702 R / 5702 RH only)
Excess temperature motor 5702.
R: Centrifuge decelarates unbraked.
Check / replace motor.
Er 24(5702 R / 5702 RH only)
Interference converter – compressor.compressor overheated.
R: Centrifuge shuts down, braked deceleration.
– Check service program SE 10.
– Replace Converter PCB.
Er 24-1(5702 R / 5702 RH only)
Compressor fault, cannot be deleted at converter. See Er 24.
Er 24-2(5702 R / 5702 RH only)
Compressor converter faulty during start. See Er 24.
Er 24-3(5702 R / 5702 RH only)
Compressor does not start (excess current). See Er 24.
Er 25Int
Mains interruption during a run > 5 seconds.
Mains voltage generally too low or too high?Variations in mains voltage (with 5702 R / 5702 RH possible during start of cooling unit).
– Check mains voltage.
– Check/replace mains switch.
– Check/replace Converter PCB.
Er 27(5702 RH only)
Heating switch relay defective. Check/replace converter PCB.
Error Cause Remedy
MSTM/KW/0706
3Error lists
/Troubleshooting
3-7
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
Warnings are not saved to the EEPROM!
Error display
Cause of error Remedy
LID Lid latch error.
R: Centrifuge does not start, lid will not open.
Note: In Service mode Er 30 is displayed.
– Check/adjust/replace switch/reed sensor/magnet.
– Check/replace lid motor.
– Check/replace Control PCB.
– See Er 5.
Er 30 Lid not locked on starting, Limit switch not reached. See Er 5.
Er 31 EEPROM error. – Carry out initialization SE 2.
– See Er 5.
OFF Display after power-off.How long the control remains active depends on the operational status of the centrifuge (accelerating, braking, stopped).
Er 40 Internal interface error. – Replace control PCB.
– Replace Converter PCB.
Er 41 Time incremental rotational transmitter defective. – Service program SE 4.
– Replace control PCB.
Er 42 Speed incremental rotational transmitter defective. – Service program SE 4.
– Replace control PCB.
Er 44 No lid acknowledge message. – See Er 5.
Er 45 No lid acknowledge message. – See Er 5.
Warnings and notices
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-8
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
Press and hold down "rpm/rcf" button for > 6 seconds (also possible during running).
Switch test displays by tapping button. To return to normal display press and hold down button for > 6 seconds.
1. Frequency converter (FC) status display
1 = Lid closed 5 = Released2 = Lid latch closed 6 = Cycle on3 = Converter stand-by message 7 = Temperature warning – converter power supply4 = Collective error message – 8 = Converter excess current
Converter has shut down with an error
2. Converter temperature or converter error code (in case of error)
Error codes: 1 = Undervoltage 10 = No release / lid switch open2 = Overvoltage FC = Reset by watchdog4 = Excess temperature FF = Serial interface – timeout7 = Excess current
3. Intermediate circuit voltage / speed control
4. Evaluating acknowledges
1.
2. x x x
x xx = Converter temperature (in °C)E xx = Converter error code, if converter shut down with an error
3. x x x
Lid closed: xxx = Intermediate circuit voltageLid opened: xxx = Actual speed for checking speed (turn rotor manually)
4. x 0 y
In standstill:
During run:
x = 1 Excess current lid latch motor during opening.y = 1 Excess temperature drive motor.z = 1 Imbalance released.
oxyz = Actual speed from converter
Customer service indications 5702
1 2
|3 4
|5 6
| |7 8
| ||
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-9
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
5. Converter version
6. Software version centrifuge
7. EEPROM checksum
8. Error number, cause of error as per list (only accessible directly after an error)
5. x x x
xxx = Converter, Version x.xx
6. y x x
y = Internal program variablexx = Software version centrifuge x.x
A B C D
ABCD = Four-digit hex number = Program memory checksum
7. u x x
u = Cause of errorxx = Error number
Customer service indications 5702
MSTM/KW/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-10
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
Press "rpm/rcf" button > 6 s (as well possible during run).
Continue to next test displays with short keystroke. Back to normal display by pressing button > 6 s.
0: Temperatures
1: Condenser temperature
2: Converter – compressor
y y minsec
y x x x
yyy = corr. nom. value = control valuexxx = Actual temperature sensor rotor chamber in °Cmin = Converter – compressor readysec = Compressor running
1. y minsec
y x x x rpmrcf
yy = Fan status(-- fan off)xx.x = Condenser temperature, cooling unit switches off at 62 °Cmin = Converter – compressor readysec = Compressor runningrpm = Mains 50Hzrcf = Mains 60Hz
2. v minsec
x x y y rpmrcf
v = Internal program parameterxx = Internal program parameter yy = Rotating field frequency compressormin = Converter – compressor readysec = Compressor runningrpm = Limit compr. frequency to max. 65 Hz f. mains volt. approx. 96 V ...110 Vrcf = Limit compr. frequency to max. 60 Hz f. mains volt. approx. < 96 V
Customer service indications 5702 R
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-11
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
3: Converter parameter drive
4: Converter parameter drive
5: Software version centrifuge
6: Lid signals / switches
3. y minsec
z x x x rpmrcf
Lid opened:During run:
y = Drive status (internal)z = Machine status (internal)xxx = Actual speed, when turning rotor manually xxx = Actual speed from converter
min = Ready converter – drivesec = Release for converter – driverpm = Cycling onrcf = Error message converter – drive
4. x min x y y y
xx = Temperature converter (on PCB Power supply – drive)min = Temperature warningyyy = Intermediate circuit voltage
5. x x y y
xx = Software version centrifuge x.xyy = Temperature (on PCB, Power supply compressor)
6. u v x y z
u = Internal variable (lid latch off)v = Motor-temp. switch (0 = Temperature too high, 1 = Switch OK)x = Imbalance switch (0 = Imbalance, 1 = Imbalance OK)y = Lid-lock switch (0 = open, 1 = closed)z = Lid switch – Reed magnet (0 = open, 1 = closed)
Customer service indications 5702 R
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-12
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
7: Last error messages
8: Voltage version / check sum EPROM
9: Current error message (selectable only directly after error)
7. x x °C y y ■
xx = Errors occured lastyy = Penultimate error message etc.°C, ■ = If a converter error code is recognized together with error 6/16/17,
it is displayed alterning with the error no. (converter error is marked with °C or ■)
8. y x x x x
y = Voltage version y=0 = 120 V y=1 = 230 Vxxxx = Check sum EPROM 0...FFFDH
9. E x x y y rpmrcf
E = Errorxx = Error numbery = No. for locating the causeyy = Converter error number,
if (with error 6) the converter switches off with errorrpm = Converter error number drivercf = Converter error number compressor
1 – Undervoltage2 – Overvoltage4 – Excess temperature converter5 – Excess temperature motor7 – Excess current converter / intermediate circuit overload
10 – no release, lid switch open (only drive)40 – Short-circuit shut offFA – Reset by watch dogFF – Serial interface time out
Customer service indications 5702 R
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-13
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
Press "rpm/rcf" button > 6 s (as well possible during run).
Continue to next test displays with short keystroke. Back to normal display by pressing button > 6 s.
H: Settings for heating unit
0: Temperatures
1: Condenser temperature
H. y minsec
z x °C x. x
y = Rotor number – with open lid(for input see operating instructions)
yz = Heating intensity with closed lidxxx = Temperature of heating cablerpm = Heating unit on /° C = Heating unit on – acknowledgement convertermin = Converter-compressor is readysec = Compressor running
y y minsec
y x x x
yyy = corr. nom. value = control valuexxx = Actual temperature sensor rotor chamber in °Cmin = Converter – compressor readysec = Compressor running
1. y minsec
y x x x rpmrcf
yy = Fan status(-- fan off)xx.x = Condenser temperature, cooling unit switches off at 62 °Cmin = Converter – compressor readysec = Compressor runningrpm = Mains 50Hzrcf = Mains 60Hz
Customer service indications 5702 RH
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-14
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
2: Converter – compressor
3: Converter parameter drive
4: Converter parameter drive
5: Software version centrifuge / Converter parameters of compressor
2. v minsec
x x y y rpmrcf
v = Internal program parameterxx = Internal program parameter yy = Rotating field frequency compressormin = Converter – compressor readysec = Compressor runningrpm = Limit compr. frequency to max. 65 Hz f. mains volt. approx. 96 V ...110 Vrcf = Limit compr. frequency to max. 60 Hz f. mains volt. approx. < 96 V
3. y minsec
z x x x rpmrcf
Lid opened:During run:
y = Drive status (internal)z = Machine status (internal)xxx = Actual speed, when turning rotor manually xxx = Actual speed from converter
min = Ready converter – drivesec = Release for converter – driverpm = Cycling onrcf = Error message converter – drive
4. x min x y y y
xx = Temperature converter (on PCB Power supply – drive)min = Temperature warningyyy = Intermediate circuit voltage
5. u x x y y
xx = Software version centrifuge x.xyy = Temperature (on PCB, Power supply compressor)u = Limit of fan step ø to 45 °C, later display in hexadecimal
Customer service indications 5702 RH
=̂
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-15
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
6: Lid signals / switches
7: Last error messages
8: Voltage version / check sum EPROM
6. u v x y z
u = Internal variable (lid latch off)v = Motor-temp. switch (0 = Temperature too high, 1 = Switch OK)x = Imbalance switch (0 = Imbalance, 1 = Imbalance OK)y = Lid-lock switch (0 = open, 1 = closed)z = Lid switch – Reed magnet (0 = open, 1 = closed)
7. x x °C y y ■
xx = Errors occured lastyy = Penultimate error message etc.°C, ■ = If a converter error code is recognized together with error 6/16/17,
it is displayed alterning with the error no. (converter error is marked with °C or ■)
8. y x x x x
y = Voltage version y=0 = 120 V y=1 = 230 Vxxxx = Check sum EPROM 0...FFFDH
Customer service indications 5702 RH
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-16
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
9: Current error message (selectable only directly after error)
9. E x x y y rpmrcf
E = Errorxx = Error numbery = No. for locating the causeyy = Converter error number,
if (with error 6) the converter switches off with errorrpm = Converter error number drivercf = Converter error number compressor
1 – Undervoltage2 – Overvoltage4 – Excess temperature converter5 – Excess temperature motor7 – Excess current converter / intermediate circuit overload
10 – no release, lid switch open (only drive)40 – Short-circuit shut offFA – Reset by watch dogFF – Serial interface time out
Customer service indications 5702 RH
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-17
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
The service functions are accessed by pressing the "rpm/rcf" button and the power-on switch simultaneously (not the "Standby" button).
Service function selection SE by rotating the "time" dial.
Start a service function by "start/stop" button.
Quit service program by switching off mains power.
Program overview
SE 1 Measurement, acceleration and braking times
SE 2 EEPROM initialization
SE 3 Test lid latch/imbalance switch
SE 4 Display and button test
SE 5 Cycle duration test
SE 6 Display error table
SE 7 Display last 4 errors
SE 8 Display operating hours and number of centrifuge runs
SE 9 Test overspeed cut-off
Service functions 5702
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-18
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
SE 1 Measurement, acceleration and braking times
Call SE 1.
Press "start/stop" to start the program.A short display test follows, all characters appear for approx. 3 seconds and a long beep sounds (the standby button lights up red).
Display:
Start the measurement run by pressing the "start/stop" button.Machine accelerates up to final speed.
Display:
xx = Acceleration time 0..4268 rpm (97 %) in s
After 3 seconds, display alternates every second between:
zz = Remaining timeyy = Rounded actual rotational speed
Stop the run by pressing the "start/stop" button. The centrifuge brakes down to a standstill, the lid opens, 5 beeps sound.
xx.x = Braking time in s.
Quit the program with the "short" button.
12 min 4.4 rpm
HL xx
1. HL xx Acceleration time
2. zz yy
br xx
Service functions 5702
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-19
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
SE 2 EEPROM initialisation
Call SE 2.
Start the program by pressing the "start/stop" button.
Display:
Confirm execution with the "rpm/rcf" button.
* The default display values are stored.
* The error table, operating hours, number of centrifuge runs and the table of the most recent errors are cleared.
Display:
After initialization the program is automatically terminated.
Cancel without initialization by "start/stop" button.
SE 3 Test lid latch and imbalance switch
Call SE 3.
Start the program by pressing the "start/stop" button.
Limit switch display:x=0 = Lid open (R) v=0 Imbalance switch not actuatedx=1 = Lid closed (R) v=1 Imbalance switch actuatedy=0 = Unlocked (M) o= For factory check onlyy=1 = Locked (M) (R = Reed magnet; M = Microswitch)
Pressing the "open" button actuates the lid motor (also possible with lid opened).
Quit the program with the "short" button.
ln lt
5.0 min 4.4 rpm
xy ov
Service functions 5702
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-20
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
SE 4 Display and key test
Call SE 4.
Start the program by pressing the "start/stop" button.A short display test follows, all characters appear for approx. 3 s, a long beep sounds and the standby button lights up red briefly.
Press all the buttons in succession, in the sequence: "rpm/rcf" button, "open" and "short".
Each time a button is pressed a digit on the display is activated.
When the "Standby" (lights up red) and "start/stop" buttons are pressed the display shows:
After "start/stop" the program is automatically terminated.
SE5 Cycle duration test
Call SE 5.
Start the program by pressing the "start/stop" button.
Display:
The run-time and the rotational speed can be changed.The cycle duration test is started by pressing the "start/stop" button again.
Sequence: 2 programs at 4400 rpm are scheduled, with 15 runs and 36 runs respectively.The switchover is made with the Standby button.
r = 15 runs, 10 min run-time, 3 min break after each run, after 15 runs 5 min breakr = 36 runs, 15 min run-time, 1 min break after each run, after 36 runs 5 min break
1. Start run. The other starts are automatic; the lid remains closed.In a break between runs the program can be quit with the "short" button.
ln
ln I88
88 88
10 min 4.4 rpm
Service functions 5702
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-21
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
SE6 Display error table
Call SE 6.
Start the program by pressing the "start/stop" button.
Display:
xx = Error numberyyy = Number of errors
The display changes once per second. It can be cancelled by pressing the "start/stop" button.The table can be cleared with service program SE 2 (Initialize).
When the errors have been run through the program is automatically terminated.
SE7 Display last 4 errors
Call SE 7.
Start the program by pressing the "start/stop" button.
Display:
x=1 = Last errorx=2 = Last-but-one error, etc.yy = Error number
When the last 4 errors have been run through the program is automatically terminated.
Note:Error 15 is not displayed here, as it is usually caused by customer loading errors.
SE8 Display operating hours and number of centrifuge runs
Call SE 8.
Start the program by pressing the "start/stop" button.
Display:
xxxx = Operating hours (min + sec)
Display:
yyyy = Number of centrifugation runs : 10
Note:The display switches automatically after 3 seconds.
The "Number of centrifugation runs" display also appears for 3 seconds,then the program is automatically terminated.
xx yyy
x yy
xx minsec xx
yy yy
Service functions 5702
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-22
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
SE9 Test overspeed cut-off
The test is performed without the rotor!
Call SE 9.
Start the program by pressing the "start/stop" button.
Display:
Start test run by pressing "start/stop" button.
Centrifuge accelerates up to cut-off speed and must shut down with Error 7. It decelerates unbraked.
After standstill, 5 signal tones are emitted.
Quit the program with the "short" button when the rotor has come to a standstill.
10 min 4.8 rpm
Service functions 5702
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-23
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
The service functions are accessed by pressing the "rpm/rcf" button and the power-on switch simultaneously (not the "Standby" button).
Service function selection SE by rotating the "time" dial.
Start a service function by "start/stop" button.
Quit service program by switching off mains power.
Program overview
SE 1 Measurement, acceleration and braking times
SE 2 EEPROM initialization
SE 3 Test lid latch/imbalance switch
SE 4 Display and button test
SE 5 Cycle duration test
SE 6 Display error table
SE 7 Display last 4 errors
SE 8 Display operating hours and number of centrifuge runs
SE 9 Test overspeed cut-off
SE 10 Fan and cooling unit / display temperature sensors
SE 11 Filling cooling unit (no service function)
SE 12 Imbalance settings
SE 13 Limitation of the fan speed of the condenser
Service functions 5702 R
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-24
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
1. Measurement, acceleration and braking times
Call SE 1.
Press "start/stop" to start the program.
Display:
Start the measurement run by pressing the "start/stop" button. Machine accelerates up to the final speed.The cooling unit is switched on only during braking ( xx=-- oder Kompressorfrequenz).
xx.x = Acceleration time 0..4268 rpm (97 %) in sec
Press "start/stop" to stop the run. Machine brakes until standstill.
xx.x = Braking time
Press "short" to exit the program.
2. EEPROM initialization
Call SE 2.
Press "start/stop" to start the program.
Display:
Press "rpm/rcf" to carry out initialization.
* The error table, operating hours, number of centrifuge runs and the table of the most recent errors are cleared.
* Fixed programs 10min / 4°C / 4400 rpm are set.
Display:
1 2 x x 4. 4
H L x x. x
b r x x. x
I n l t
5. 0 4 4. 4
Service functions 5702 R
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-25
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
3. Test lid latch
Call SE 3.
Press "start/stop" to start the program.
v = Motor temp. switch (0 = Temperature too high,1 = Switch OK)x = Imbalance switch (0 = Imbalance, 1 = Imbalance OK)y = Lid lock switch (0 = opened, 1 = closed)z = Lid switch – Reed magnet (0 = opened, 1 = closed)
Switching off the lid motoru = 1 – Switching off the lid by intermediate step during opening and closing.u = 2 – Lid closedu = 3 – Excess current cut-off, lid open
The lid motor is controlled during closing by pressing "rcf" (leave a pause of 1 s between each pulse).
Press "short" to exit the program.
SE 4 Display and button test
Call SE 4.
Press "start/stop" to start the program.A short display test follows, all characters appear for approx. 3 s, a long beep sounds and the program buttons light up blue.
Press all the buttons in succession, in the sequence: "rpm/rcf" button, "open", "fast temp" and "short".
Each time a button is pressed a digit on the display is activated.
When the "Standby" and "start/stop" buttons are pressed the display shows:
After "start/stop" the program is automatically terminated.
L u v x y z
I n
I 8 8 8 8
8 8 8 8 8 8
Service functions 5702 R
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-26
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
SE 5 Cycle duration test
Call SE 5.
Start the program by pressing the "start/stop" button.
Display:
The run-time and the rotational speed can be changed.The cycle duration test is started by pressing the "start/stop" button again.
Sequence:2 programs at 4400 rpm are scheduled, with 15 min. runs and with 10 min. runs respectively.The switchover is made with the Standby button.
r = 36 runs, 10 min run-time, 3 min break after each run, after 36 runs 5 min breakr = 36 runs, 15 min run-time, 1 min break after each run, after 36 runs 5 min break
1. Start run. The other starts are automatic; the lid remains closed.In a break between runs the program can be quit with the "short" button.
SE 6 Display error table
Call SE 6.
Start the program by pressing the "start/stop" button.
Display:
xx = Error numberyyy = Number of errors
The display changes once per second. It can be cancelled by pressing the "start/stop" button.The table can be cleared with service program SE 2 (Initialize).
When the errors have been run through the program is automatically terminated.
SE7 Display last 4 errors
Call SE 7.
Start the program by pressing the "start/stop" button.
Display:
x=1 = Last errorx=2 = Last-but-one error, etc.yy = Error numberzz = Converter error code (see table "Customer service indications 5702 R", No. 9 "Current error message")
Note:Error 15 is not displayed here, as it is usually caused by customer loading errors.
1 5 4 4. 4
E r x x y y y
x E y y z z
Service functions 5702 R
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-27
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
SE8 Display operating hours and number of centrifuge runs
Call SE 8.
Start the program by pressing the "start/stop" button.
Display:
xxxxx = Operating hours
Display:
yyyyyy = Number of centrifuge runs
The display switches over after 3 s!
The number of centrifugations is also displayed for 3 s, after this the program terminates automatically!
SE 9 Test overspeed cut-off
The test is performed without rotor!
Call SE 9.
Start the program by pressing the "start/stop" button.
Display:
Start test run by pressing "start/stop" button.
Centrifuge accelerates up to cut-off speed and must shut down with Error 7. It decelerates unbraked.
Quit the program with the "short" button when the motor has come to a standstill.
x x x x x h
y y y y y y
1 5 4 4. 7
Service functions 5702 R
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-28
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
SE 10 Fan and cooling unit / display temperature sensors
Call SE 10.
Start the program by pressing the "start/stop" button.
Display unit:
u=b = Converter – compressor problem-freev=b = Converter – drive problem-free
Buttons:fast temp = Compressor, x=1 (on) x=0 (off)open = Condenser fan, z=0 (on) x=– (off)temp = Setting
condenser fan status, zz = settable from 0 up to 32 (switch on condenser fan with "open" before)prog 1 = Fan for Converter PCB, y=0 (on) y=1 (off)
"short" button: Selecting Display cooling unit <=> Temperatures
Display temperatures:
xx.x = Condenser temperature with factor "x10"yy.y = Chamber temperature
"start/stop" terminates the program
SE 11 Filling cooling unit
No service function.
"short" key for leaving the program.
SE 12 Setting imbalance
Call up SE 12.
– Set rotational speed
– Insert and load rotorTable for imbalance adjustment/testing, see Chapter 5
– Close lid
Start program with "start/stop" key.
Display when emergency switch activated:
Service program is exited. It is possible to open the lid while the rotor is at a stillstand.
Note:The imbalance setting can be checked with SE 12 without accessing the cooling aggregate.
u x v y z z
x x y y. y
I n b
Service functions 5702 R
x .
MSTM/KW/0706
3Error lists
/Troubleshooting
3-29
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
SE 13 Limitation of the fan speed of the condenser
Call up SE 13.
Start program with "start/stop" key.
Display when emergency switch is activated:
Change setting: "time" dialSave setting and end program: "start/stop" key.
Note:The setting of the condenser fan is possible within a range from "20 – 32". A change makes sense when the room tempe-rature is very high or when the voltage is lowThe standard lowest setting for the fan speed is "20". "25" must be set for 100 volt devices following the replacement of the PCB Control 5702 R.The setting is maintained after the device is switched off or following initialization.
2 0
Service functions 5702 R
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-30
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
The service functions are accessed by pressing the "rpm/rcf" button and the power-on switch simultaneously (not the "Standby" button).
Service function selection SE by rotating the "time" dial.
Start a service function by "start/stop" button.
Quit service program by switching off mains power.
Program overview
SE 1 Measurement, acceleration and braking times
SE 2 EEPROM initialization
SE 3 Test lid latch/imbalance switch
SE 4 Display and button test
SE 5 Cycle duration test
SE 6 Display error table
SE 7 Display last 4 errors
SE 8 Display operating hours and number of centrifuge runs
SE 9 Test overspeed cut-off
SE 10 Fan, cooling and heating unit / display temperature sensors
SE 11 Filling cooling unit (no service function)
SE 12 Imbalance setting
SE 13 Limiting the fan speed of the condenser
Service functions 5702 RH
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-31
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
1. Measurement, acceleration and braking times
Call SE 1.
Press "start/stop" to start the program.
Display:
Start the measurement run by pressing the "start/stop" button. Machine accelerates up to the final speed.The cooling unit is switched on only during braking ( xx=-- oder Kompressorfrequenz).
xx.x = Acceleration time 0..4268 rpm (97 %) in sec
Press "start/stop" to stop the run. Machine brakes until standstill.
xx.x = Braking time
Press "short" to exit the program.
2. EEPROM initialization
Call SE 2.
Press "start/stop" to start the program.
Display:
Press "rpm/rcf" to carry out initialization.
* The error table, operating hours, number of centrifuge runs and the table of the most recent errors are cleared.
* Fixed programs 10min / 4°C / 4400 rpm are set.
Display:
1 2 x x 4. 4
H L x x. x
b r x x. x
I n l t
5. 0 4 4. 4
Service functions 5702 RH
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-32
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
3. Test lid latch
Call SE 3.
Press "start/stop" to start the program.
v = Motor temp. switch (0 = Temperature too high,1 = Switch OK)x = Imbalance switch (0 = Imbalance, 1 = Imbalance OK)y = Lid lock switch (0 = opened, 1 = closed)z = Lid switch – Reed magnet (0 = opened, 1 = closed)
Switching off the lid motoru = 1 – Switching off the lid by intermediate step during opening and closing.u = 2 – Lid closedu = 3 – Excess current cut-off, lid open
The lid motor is controlled during closing by pressing "rcf" (leave a pause of 1 s between each pulse).
Press "short" to exit the program.
SE 4 Display and button test
Call SE 4.
Press "start/stop" to start the program.A short display test follows, all characters appear for approx. 3 s, a long beep sounds and the program buttons light up blue.
Press all the buttons in succession, in the sequence: "rpm/rcf" button, "open", "fast temp" and "short".
Each time a button is pressed a digit on the display is activated.
When the "Standby" and "start/stop" buttons are pressed the display shows:
After "start/stop" the program is automatically terminated.
L u v x y z
I n
I 8 8 8 8
8 8 8 8 8 8
Service functions 5702 RH
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-33
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
SE 5 Cycle duration test
Call SE 5.
Start the program by pressing the "start/stop" button.
Display:
The run-time and the rotational speed can be changed.The cycle duration test is started by pressing the "start/stop" button again.
Sequence:2 programs at 4400 rpm are scheduled, with 15 min. runs and with 10 min. runs respectively.The switchover is made with the Standby button.
r = 36 runs, 10 min run-time, 3 min break after each run, after 36 runs 5 min breakr = 36 runs, 15 min run-time, 1 min break after each run, after 36 runs 5 min break
1. Start run. The other starts are automatic; the lid remains closed.In a break between runs the program can be quit with the "short" button.
SE 6 Display error table
Call SE 6.
Start the program by pressing the "start/stop" button.
Display:
xx = Error numberyyy = Number of errors
The display changes once per second. It can be cancelled by pressing the "start/stop" button.The table can be cleared with service program SE 2 (Initialize).
When the errors have been run through the program is automatically terminated.
SE7 Display last 4 errors
Call SE 7.
Start the program by pressing the "start/stop" button.
Display:
x=1 = Last errorx=2 = Last-but-one error, etc.yy = Error numberzz = Converter error code (see table "Customer service indications 5702 R", No. 9 "Current error message")
Note:Error 15 is not displayed here, as it is usually caused by customer loading errors.
1 5 4 4. 4
E r x x y y y
x E y y z z
Service functions 5702 RH
MSTM/MMWT/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-34
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
SE 8 Display operating hours and number of centrifuge runs
Call SE 8.
Start the program by pressing the "start/stop" button.
Display:
xxxxx = Operating hours
Display:
yyyyyy = Number of centrifuge runs
The display switches over after 3 s!
The number of centrifugations is also displayed for 3 s, after this the program terminates automatically!
SE 9 Test overspeed cut-off
The test is performed without rotor!
Call SE 9.
Start the program by pressing the "start/stop" button.
Display:
Start test run by pressing "start/stop" button.
Centrifuge accelerates up to cut-off speed and must shut down with Error 7. It decelerates unbraked.
Quit the program with the "short" button when the motor has come to a standstill.
x x x x x h
y y y y y y
1 5 4 4. 7
Service functions 5702 RH
MSTM/MMWT/0706
3Error lists
/Troubleshooting
3-35
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
SE 10 Fan, Cooling aggregate and heating unit/ display temperature sensors
Call SE 10.
Start the program by pressing the "start/stop" button.
Display aggregate:
u=b = Converter – compressor problem-freev=b = Converter – drive problem-free
Buttons:fast temp = Compressor, x=1 (on) x=0 (off)open = Condenser fan, z=0 (on) x=– (off)temp = Setting
condenser fan status, zz = settable from 0 up to 32 (switch on condenser fan with "open" before)prog 1 = Fan for Converter PCB y =0 (on) y=1 (off)prog 2 = Heating unit on/off w=1 (on) w=0 (off)
°C = Heating unit of PCB converter on
"short" key switchover to display of aggregate <=> Temperatures
Display temperatures:
xx.x = Condenser temperatureyy.y = Chamber temperature
"short" key switchover to display of heating unit temperature:
yy = Heating unit temperature
"start/stop" terminates the program
SE 11 Filling cooling unit
No service function.
"short" key to leave the program.
u x v y °C w z z
x x y °C y. y
y y
Service functions 5702 RH
x .
MSTM/KW/0706
3Er
ror
lists
/Tr
oubl
esho
otin
g
3-36
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting
SE 12 Setting imbalance
Call up SE 12.
– Set rotational speed
– Insert and load rotorLoad rotor on one side with G1 = No errorTable for imbalance adjustment/testing, see Chapter 5
– Close lid
Start program with "start/stop" key.
Display with activated emergen1cy switch:
Service program is left. It is possible to open the lid while the rotor is at a standstill.
Note:The imbalance setting can be checked with SE 12 without accessing the cooling aggregate.
SE 13 Limiting the fan speed of the condenser
Call up SE 13.
Start program with the "start/stop" key.
Display with activated emergency switch:
Change setting: "time" dialSave setting and end program: "start/stop" key.
Note:The setting of the condenser fan is possible within a range of "20 – 32". A change makes sense when the room tempera-ture is very high or when the voltage is low.The lowest fan speed of "20" is set as a standard. "25" must be set for 100 volt devices following replacement of the PCB Control 5702 R.The setting is maintained after the device is switched off or following initialization.
I m b a l
2 0
Service functions 5702 RH
MSTM/KW/0706