1 Chapter 25 HPLC High-Performance Liquid Chromatography.
-
Upload
emery-newman -
Category
Documents
-
view
269 -
download
8
Transcript of 1 Chapter 25 HPLC High-Performance Liquid Chromatography.
5
Increased Efficiency
• In any method the object is to increase mobile phase / stationary phase interaction.
• Decrease particle size. – Better packing – slower flow.
8
Smaller is Better
• N ~ 3500 L (cm) / dp (m)
• Smaller particle size leads to – Higher plate number– Higher pressure– Shorter optimum run time
11
What is a bar?
• A measure of pressure.
• One bar = 1.01325 Atm
• One bar is 100 000 Newtons/m2
• Under water you gain about one atm for each ten meters in depth.
13
Stationary Phase / Support
• Support is the scaffolding that the stationary phase sits on.
• Support - Microporous Silica– Solvent can get inside– Surface area of 100’s m2 / gram– Silica degrades above pH 8 so keep pH below.– Special supports have been developed for
higher pH
17
Bonded Phases
• SiOH + ClSi(CH3)2 - R = Si-O-(CH3)2 - R • Change surface polarity
– Amino (CH2)3NH2
– Cyano (CH2)3C#N– Diol (CH2)2OCH2CH(OH)CH2OH– Octadecyl (CH2)17CH3
– Octyl (CH2) 7CH3
– Phenyl (CH2)3C6H5
20
Elution
• Competition for the surface.
• Which substance has the greater surface affinity – a solvent with a higher affinity will push the analytes along.
23
Isocratic and Gradient Elution
• Isocratic - same composition• Gradient - composition changes over time
• Method Development.• A mixture of compounds. Separated with
acetonitrile (B) and aqueous buffer (A)• 1) benzyl alcohol 2) phenol 3) 3’,4’-dimethyloxyacetophenone 4)
benzoin 5) ethyl benzoate 6) toluene 7) 2,6-dimethoxytoluene 8) o-methyoxybiphenyl
27
Too Long - Over Two Hours!
• We can do a gradient. Examine when we get the best separations and then change composition over time.
31
Practical Issues
• Solvents– Must be very pure, lack UV absorbing species– Should be filtered– Guard columns should be used to protect column from strong
absorbing compounds– Solvents should be degassed – bubbles and oxygen (sparging)– Normal phase solvents should be 50% saturated with water– Gradients in reversed phase can require 10 – 20 column volumes
to return to starting conditions• Add 3% 1-propanol to each solvent and you will cut this to 1.5 empty
volumes
32
Reducing Waste Solvent(Save some money – be a hero)
• Shorter columns with smaller particles
• Switch from 4.6 mm to 3.0/2.0 mm id columns
• In isocratic systems – use an electronic recycler.
33
Quality Assurance
• Inject a QC sample each day to insure that you have consistent peak shapes and retention times. – Keep a log of your column performance
34
Symmetric Band-shape
• Asymmetry Factor A/B should rarely be worse than 0.9 to 1.5.
• Tailing a bigger issue and fronting generally.– Amines interacting with support active sites.– Add 30 mM TEA– Acidic compounds – Add 30 mM Ammonium Acetate– Or for mixes or unknowns – 30 mM triethylammonium acetate.– Persistent problem – add dimethyloctylamine or
dimethyloctylammonium acetateThese take a long time to wash on on changing mobile phases.
35
Other issues
• Voids can develop at the column inlet.– Repack with fresh stationary phase to get rid of this
problem. (Might want to get new column)
• Columns should be washed to get rid of salts and strongly adsorbed compounds
• Frits should be cleaned. Back wash or replace• Samples should be dissolved in mobile phase or a
weaker solvent.
37
Overloading
• Care should be taken not to overload the injection on the column.– Inject 10x less and see if peaks look any better.
• Reversed Phase can deal with 1 to 10 mg sample per gram of silica. (About 10 cm on a 0.46 mm column)
• Injection volume– < 15% of the peak volume at baseline
39
Injection and Detection
• Pumps – Piston Type under program control. Up to 400
bar (40 MPa or 6000 psi). Gradients made by proportioning valves.
48
Evaporative Light Scattering
• Solutes less volatile than mobile phase• Light scattered from mass of analyte.• Poor linearity – polynomial calibration• Good with gradients. No solvent front.• Same buffers as used with mass spec.
Acetic acid, formic acid and TFA, ammonium acetate, diammonium phosphate, ammonia or TEA.
50
Electrochemical
• Analytes that can be oxidized or reduced– Phenols, aromatic amines, peroxides,
mercaptans, ketones, aldehydes etc.
• Can be very sensitive but are difficult to work with.
53
Gradient Separations
• Same purpose as temperature or pressure programming in GC. Speed up analyses.
• Increase the eluent strength as the run progresses. That is increase the amount of organic phase as the run progresses in reversed phase methods.
54
Dwell Time
• Time it takes for the composition change at the pump to reach the head of the column.
• Important on method transfer
• Can be determined by running a gradient of 0.1% acetone (detect at 260 nm) without a column. Start gradient at time 0 and see when the acetone starts to be absorbed.
56
Since gradients require a post equilibrium time then perhaps isocratic will work better.
• What is the span of all the peaks. That time span is t
– If t/tg > 0.25 then use a gradient. If less you can make an isocratic system. (See next slide)
58
Optimize Gradient
• Run a broad gradient. (5% to 100% over 40 to 60 min)
• Eliminate gradient before first peak and after last peak. Run over same time.
• Cut time if above works well to save time.