Regulatory compliance information

This product complies with the essential requirements of the following applicable European Directives, and carries the CE marking accordingly:

The Low Voltage Directive 73/23/EEC, amended by 93/68/EEC The EMC Directive 89/336/EEC, amended by 93/68/EEC

To obtain Declaration of Conformity, please contact your local Agilent Technologies sales office, agent or distributor.

MANUAL CHANGES 4284AMANUAL IDENTIFICATION Model Number: 4284A Date Printed: January 2000 Part Number: 04284-90040

Precision LCR MeterOperation Manual

This supplement contains information for correcting manual errors and for adapting the manual to newer instruments that contains improvements or modifications not documented in the existing manual. To use this supplement 1. Make all ERRATA corrections 2. Make all appropriate serial-number-related changes listed below

SERIAL PREFIX OR NUMBER CHANGES ALL

MAKE MANUAL 1

SERIAL PREFIX OR NUMBER CHANGES

MAKE MANUAL

K New Item

ERRATA

CHANGES 1CHANGE 1 contains the information needed to adapt the 4284As manual.

MODEL 4284A PRECISION LCR METER OPTION 201 HANDLER INTERFACE OPERATION NOTE Page 2-6 Signal Line Used for List Sweep Comparator Function.Change the description as follows: l Comparison Output Signals: /BIN1 - /BIN9 and /AUX_BIN indicate IN/OUT judgements for each sweep point (refer to Figure 2-4). /OUT_OF_BINS indicates pass/fail judgement (one or more failed judgements of Steps 1 to 10 occurred during a single sweep)

NOTE Manual change supplement are revised as often as necessary to keep manuals as current and accurate as possible. Agilent Technologies recommends that you periodically request the latest edition of this supplement. Free copies are available from all Agilent Technologies offices. When requesting copies, quote the manual identification information from your supplement, or the model number and print date from the title page of the manual.

Date/Div: January 2000/33 Page 1 of 2PRINTED IN JAPAN

Page 2-7 Table 2-2. Contact Assignments for List Sweep Comparator FunctionChange a part of the table 2-2 as follows: Pin No. 1 2 3 4 5 6 7 8 9 11 10 Signal Name /BIN1 /BIN2 /BIN3 /BIN4 /BIN5 /BIN6 /BIN7 /BIN8 /BIN9 /AUX_BIN /OUT_OF_BINS Description Failed (out of limit) at sweep Point 1 Failed (out of limit) at sweep Point 2 Failed (out of limit) at sweep Point 3 Failed (out of limit) at sweep Point 4 Failed (out of limit) at sweep Point 5 Failed (out of limit) at sweep Point 6 Failed (out of limit) at sweep Point 7 Failed (out of limit) at sweep Point 8 Failed (out of limit) at sweep Point 9 Failed (out of limit) at sweep Point 10 /OUT_OF_BINS is asserted when one or more fail judgements of Step 1 to 10 occur in a single sweep.

Page 2-8 Figure 2-4. Signal Area Example. (For The List Sweep Comparator Function)Change the figure as follows:

MANUAL CHANGES 4284AMANUAL IDENTIFICATION Model Number: 4284A Date Printed: January 2000 Part Number: 04284-90040

Precision LCR MeterOperation Manual

This supplement contains information for correcting manual errors and for adapting the manual to newer instruments that contains improvements or modifications not documented in the existing manual. To use this supplement 1. Make all ERRATA corrections 2. Make all appropriate serial-number-related changes listed below

SERIAL PREFIX OR NUMBER CHANGES All

MAKE MANUAL 1

SERIAL PREFIX OR NUMBER CHANGES

MAKE MANUAL

K New Item

ERRATA

CHANGES 1CHANGE 1 contains the information needed to adapt the 4284As manual. Changed the company name from YOKOGAWA-HEWLETT-PACKARD, LTD., or its abbreviation YHP to Agilent Technologies Japan, Ltd.

NOTE Manual change supplement are revised as often as necessary to keep manuals as current and accurate as possible. Agilent Technologies recommends that you periodically request the latest edition of this supplement. Free copies are available from all Agilent Technologies offices. When requesting copies, quote the manual identification information from your supplement, or the model number and print date from the title page of the manual.

Date/Div: January 2000/33 Page 1 of 3PRINTED IN JAPAN

The pink sheet titled CAUTION ON OPERATIONChange the page title as follows. CAUTION ON OPERATION

3.HANDLER INTERFACE BOARD (OPTION 201)Add the following information. Fuse: Non Time Delay 0.5A 125V If you need this fuse,contact your nearest Agilent Technologies Sales and Service Office. Dangerous voltage may be present in the 4284A even through the power switch is off. Be sure to wait 1 minutes for the internal capacitors to discharge.

MODEL 4284A PRECISION LCR METER OPTION 201 HANDLER INTERFACE OPERATION NOTE Page 2-16 Procedure 1 and WarningChange the procedure 1 and warning as follows: 1. Disconnect the power cable from the 4284A and allow 1 minute for the internal capacitors to discharge. Dangerous energy/voltage exists when the 4284A is in operation,and for a time after it is powered down. Allow 1 minute for the internal capacitors to discharge.

Page 2-17Add the following CAUTION after the procedure 7. The interface board contains electronic components that can be damaged by static electricity through electrostatic discharge(ESD).To prevent ESD damage,maintain frequent contact with any bare sheet metal surface on the chassis. A grounding wrist strap (or similar device) is useful for this purpose. Handle the board carefully at all times. Avoid touching electronic components or circuit paths.

MODEL 4284A PRECISION LCR METER OPTION 202 HANDLER INTERFACE OPERATION NOTE Page 3-4 Procedure 1 and WarningChange the procedure 1 and warning as follows: 1. Disconnect the 4284As power cord and allow 1 minute for the internal supply filter capacitors to discharge. Dangerous energy/voltage exists when the 4284A is in operation,and for a time after it is powered down. Allow 1 minute for the internal capacitors to discharge.

Page 3-4Add the following CAUTION after the procedure 6. The interface board contains electronic components that can be damaged by static electricity through electrostatic discharge(ESD).To prevent ESD damage,maintain frequent contact with any bare sheet metal surface on the chassis. A grounding wrist strap (or similar device) is useful for this purpose. Handle the board carefully at all times. Avoid touching electronic components or circuit paths.

MODEL 4284A PRECISION LCR METER OPTION 301 SCANNER INTERFACE OPERATION NOTE Page 2-11PROCEDURE 1 and Warning Change the procedure 1 and warning as follows: 1. Disconnect the power cable from the 4284A and allow 1 minute for the internal capacitors to discharge. Dangerous energy/voltage exists when the 4284A is in operation,and for a time after it is powered down. Allow 1 minute for the internal capacitors to discharge.

Page 2-12Add the following CAUTION after the procedure 7. The interface board contains electronic components that can be damaged by static electricity through electrostatic discharge(ESD).To prevent ESD damage,maintain frequent contact with any bare sheet metal surface on the chassis. A grounding wrist strap (or similar device) is useful for this purpose. Handle the board carefully at all times. Avoid touching electronic components or circuit paths.

Agilent 4284A Precision LCR Meter Operation Manual

Manual ChangeAgilent Part No. N/A July 2007

Change 1Change the figure 10-5 of DC Bias Level Accuracy Test Setup without the Interface Box (page 10-9) as follows.

C Copyright 2007 Agilent Technologies

1 DC 10-5 10-9

C Copyright 2007 Agilent Technologies

CAUTIONS ON OPERATION1. UNKNOWN (MEASUREMENT) TERMINALS Do NOT apply DC voltage or current to the UNKNOWN terminals. Doing so will damage the 4284A. Before you measure a capacitor, be sure the capacitor is fully discharged. 2. MEMORY CARD Use Agilent Technologies-specified memory cards containing 4284A-specific data only. If other memory cards are used, the 4284A may be damaged. Non 4284A-specific data contained on a memory card is not guaranteed, and data may be lost. To insert a memory card into the MEMORY card slot, hold the memory card with the label facing upward and with the contacts at the slot opening Insert the card into the slot until it "clicks" in place. To remove a memory card from the 4284A, press the UNLOCK button and remove the card. Do NOT remove a memory card while LOADing or STORing data. Doing so may damage the memory card and any data stored in the memory card may be lost. Store memory cards in their furnished card cases when not in use. The card case protects memory cards from contamination and electrostatic discharge. Also, store memory cards under the following environmental conditions. Storage Temperature Range: -30 oC to +70 oC Storage Humidity Range: 30% to 85% (@ +50 oC) Do NOT shock or stress memory cards. When storing or moving your 4284A, be sure the memory card slot is empty (no memory card inserted). Do NOT touch the connector contact surface of a memory card and do NOT use chemical liquids to clean the contacts.

3. HANDLER INTERFACE BOARD (0PTION 201)If the +5V internal voltage (pin 16, 17 or 18 of the handler interface connector) is not output, output a fuse on the handler interface board (A32F1) has blown and must be replaced. Two replacement fuses are furnished with the 4284A option 201. Additional fuses are available from Agilent Technologies. Order PN 2110-0046. Fuse: Mpm Time Delay 0.5A 12.5V

If you need this fuse, contact your nearest Agilent Technologies Sales and Service Office. To replace A32F1 , perform the following procedure. 1. 2. 3. To remove the handler interface board (A32), perform procedure I through 7 on page 10-26. 10-26 Remove A32F1