Introduction to
Electron Microscopy
Andres Kaech
Instrumentation and Image Formation
Center for Microscopy and Image Analysis
The types of electron microscopes
Scanning electron microscope (SEM)Transmission electron microscope (TEM)
21.02.2012
2
Scanning electron microscope (SEM)Transmission electron microscope (TEM)
The types of electron microscopes
Electron beam
Specimen ~100 nm
Electron beam
Specimen
Projection Surface
Hela Cells
Elektronenmikroskopie ETH Zurich
1 µm
Specimens courtesy of Bärbel Stecher, Institute of Microbiology, ETH Zurich
Examples TEM
Mouse intestine
Microvilli
GlycocalixJunction
Actin filaments
21.02.2012
3
Elektronenmikroskopie ETH Zürich
500 nm
Mouse intestine
Membrane(lipid bilayer)
Actin filaments
Examples TEM
H/K -ATPase in cimetidine-treated resting gastric parietalcells (rabbit).
Immunolabelling: Localization of proteins
Sawaguchi et al. 2004, Journal of Histochemistry & Cytochemistry
100 nm
Examples TEM
21.02.2012
4
Examples SEM
Mouse kidney
500 µm
Mouse kidney (glomerulus)
10 µm
Examples SEM
21.02.2012
5
Pseudomonas aeruginosa
500 nm
Examples SEM
Wave-particle duality
Resolution depends on aperture and wavelength (Diffraction limited resolution)
Optical properties(Diffraction, chromatic abberation, spherical abberation, astigmatism etc.)
Abbe’s equation d = 0.61 λ/NA sin nNA
e-
Properties of electrons
Very similar to photons:
21.02.2012
6
Resolution of biological objects limited by specimen preparation: Practical resolution: > 1 nm
TEM: 40 – 300 kV
Effective instrument resolution TEM: 0.5 nm (120 kV)
Effective instrument resolution SEM: 1 nm
Resolution of electron microscopes
The higher the energy of the electrons, the lower the wavelength, the higher the resolution
SEM: 0.5 – 30 kV
Widefield light microscopeTransmission electron microscope
Condenser lens
Objective lens
Projector lens
Specimen
Illumination
Final image
Transmission electron microscope vs. Widefield light microscope
21.02.2012
7
Confocal laser scanning microscopeScanning electron microscope
Beam scanner
Detector
Lens system
Lens system
Specimen
Illumination
Scanning electron microscope vs. Confocal laser scanning microscope
Example: Transmission electron microscope
Cathode
Specimen holder
Viewing screen
TMP
RP
IGP
IGP
Ion getter pump
Turbo molecular pump
Oil diffusion pump
Rotary pump
Atmosphere: 1000 mbar
10-5 - 10-7 mbar
10-0 - 10-2 mbar
10-7 - 10-10 mbar
Electron microscopes are high vacuum systems
21.02.2012
8
High voltage(eg. 120 kV)
Electron source
(e.g. tungsten)
Electromagnetic lenses
Electromagnetic lens of a transmission electron microscope
The focal length can be changed by changing the current:
No movement or exchange of the lens is required for focusing or changing magnification!
21.02.2012
9
Chromatic aberration
Electromagnetic lenses
Spherical aberrationsDue to energy difference of electrons (wavelength)
e- (98 kV)
e- (100 kV)
e- (102 kV)
Curvature and distortion of field
Axial astigmatism - confusion of the image
Electromagnetic lenses
Under focused imageelliptic deformation
Over focused imageelliptic deformation
Focuscircle of least confusion
Cellulose filter paper imaged in SEM
Focus, corrected astigmatismcircle of confusion minimized
With astigmatism
Withouth astigmatism
21.02.2012
10
Electromagnetic lenses
Reasons:
• Inhomegenities of the lens
• Contamination of lenses and apertures
• Charging of specimen
Object: Source of e-
(off-axis)
Objective lens
Optical axis
“True” image of object
z
x
y
Focus
Corrector coils
Axial astigmatism - confusion of the image
Specimen holder
Specimen on a TEM grid
Specimen holders and stages
3 mm3 mm
Transmission electron microscope
21.02.2012
11
Specimen holders and stages
Transmission electron microscope
Goniometer: x, y, z, r
Specimen size:
• 3 mm in diameter
• 100 nm in thickness
Specimen holders and stages
Scanning electron microscope
Specimen stage (x, y, z, r, tilt)
Objective lens
Stage
Specimen stub
Stub holder
Specimen size:
• 100 mm in diameter
• 2 cm in z-direction (not electron transparent)
21.02.2012
12
Electron – specimen interactions
Inelastic(low angle, E=E0-∆E)
Unscattered(E=E0)
Primary electrons (E0)Backscattered electrons (E=E0)
Elastic(higher angle, E=E0)
Electron – specimen interactions
Emission of electrons and radiation
Inelastic scattering:
Energy is transferred from the primary electron to the specimen
K
L
M
N
1
2
K
L
M
N
21.02.2012
13
Electron – specimen interactions
Primary electrons
Unscattered electrons
Inelastically scattered electrons
Elastically scattered electrons
Secondary electronsBackscattered electrons
Auger electronsHeat
Cathode luminescenseX-rays
Specimen Interaction volume
SEM analysis
TEM analysis
Contrast formation in TEM
NOTE: Mechanisms occur at the same time (superposition)
Imaging in the transmission electron microscope
Absorption contrast
Scattering/phase contrast
21.02.2012
14
Contrast formation in TEM
Biological specimen consist of light elements:
Absorption contrast weak
Scattering/phase contrast weak
Contrast enhancement required:
Treatment with heavy metals (Ur, Pb, Os)!
“LOW CONTRAST”
Heavy metals attach differently to different components
Imaging in the transmission electron microscope
Main contrast formation in plastic embedded specimens
Scattering of electrons through heavy metals
Specimen
Image plane
phospholipids ribosome
…Heavy metal ions
Imaging in the transmission electron microscope
Objective aperture(back focal plane)
Primary electron beam
Brightness
21.02.2012
15
Thin section of alga stained with heavy metals (Ur, Pb)
Imaging in the transmission electron microscope
Thin section of alga without heavy metal staining
1 µm
Imaging in the transmission electron microscope
21.02.2012
16
Contrast enhancement by underfocusing
Thin section of a frozen-hydrated apple leaf (“unstained”)
1 µm
Phase contrast only between H2O and biological material
Electron microscopy ETH Zurich
Imaging in the transmission electron microscope
The CCD camera for electron microscopy
Outside the microscope
Inside the microscope (vacuum)
• Electrons need to be converted to photons (scintillator)
• CCD has to be protected from electron bombardment
Imaging in the transmission electron microscope
• Nowadays direct electron CCD available, no scintillator required (very expensive)
21.02.2012
17
Primary electrons
Unscattered electrons
Inelastically scattered electrons
Elastically scattered electrons
Secondary electronsBackscattered electrons
Auger electronsHeat
Cathode luminescenseX-rays
Specimen Interaction volume
SEM analysis
TEM analysis
Imaging in the scanning electron microscope
Scanning and signal detection
…Primary electron beam
secondary electrons
Imaging in the scanning electron microscope
21.02.2012
18
Signal and detection
Different properties of the different signals
►Specific detectors
►Different/specific information
Imaging in the scanning electron microscope
Secondary electron detector
+7-12kV HVPhotomultiplier
+200-500V – Collector voltage
Photons ElectronsElectrons
Primary electrons
SE
Imaging in the scanning electron microscope
21.02.2012
19
Contrast formation in SEM using SE
Different number of electrons from different spots of the specimen
topography of the specimen
location of the detector
acceleration voltage of primary electrons
composition of the specimen
Dependent on
Imaging in the scanning electron microscope
PE Primary electronsSE Secondary electronsR Excited volume
Contrast based on SE - topography
SE
PE
R
RSE
Specimen
Imaging in the scanning electron microscope
21.02.2012
20
Wing of butterfly
Contrast based on SE
Electron microscopy ETH Zurich
Imaging in the scanning electron microscope
Mouse kidney (glomerulus)
10 µm
Contrast based on SE – detector position Virtual light source
Imaging in the scanning electron microscope
21.02.2012
21
Biological material (light elements):
Only few electrons escape from specimen
Almost no contrast, similar contrast everywhere on specimen
Contrast enhancement
Localization of the signal to the surface
Coating of biological specimen with thin heavy metal layer (a few nm)
Reducing acceleration voltage
Unsharp image
Contrast formation in SEM
Imaging in the scanning electron microscope
Primary electron beam
Contrast formation
Platinum
Primary electron beam
Uncoated Coated with 4 nm platinum
Imaging in the scanning electron microscope
21.02.2012
22
Freeze-fractured yeast
500 nm
Contrast based on SE: Non-coating vs. coating with heavy metals
Uncoated Coated with 4 nm platinum
Electron microscopy ETH Zurich
Imaging in the scanning electron microscope
Focusing and magnification in electron microscopy
Focusing and magnification in TEM and SEM
Focusing and magnification in light microscopy
TEM:
Focusing:
Change current in magnetic lenses for focusing (objective lens)
Move holder in z
Magnification:
Change current in magnetic lenses (projective lenses) and combine several projective lenses.
SEM:
Focusing:
Change current in magnetic lenses for focusing (objective lens)
Move stage in z
Magnification:
Change scanning field (scan a smaller or larger area with the same number of pixels), pixel size changes.
Widefield light microscopy:
Focusing:
Moving objective or
stage in z
Magnification:
Changing the whole objective.
Confocal scanning laser microscopy:
Focusing:
Moving objective or
stage in z
Magnification:
Change scanning field (scan a smaller or larger area with the same number of pixels), pixel size changes.
Changing the whole objective