Download - Tissue Imaging by 5 kHz High-Performance MALDI-TOF Poster Number TP191 Christina Vestal 1, Kenneth Parker 1, Kevin Hayden 1, George Mills 1, Marvin Vestal.

Transcript
Page 1: Tissue Imaging by 5 kHz High-Performance MALDI-TOF Poster Number TP191 Christina Vestal 1, Kenneth Parker 1, Kevin Hayden 1, George Mills 1, Marvin Vestal.

Tissue Imaging by 5 kHz High-Performance MALDI-TOF Poster Number TP191

Christina Vestal1, Kenneth Parker1, Kevin Hayden1, George Mills1, Marvin Vestal1, Shannon Cornett2, and Richard Caprioli2. 1Virgin Instruments, Sudbury, MA, 2Vanderbilt University School of Medicine, Nashville, TN.

IntroductionTissue imaging and profiling by MALDI-TOF MS provides unique information that facilitates our understanding of normal biological and pathological processes. This technology provides a high-throughput tool to interrogate protein expression in tissue at the molecular level. It can also provide direct measurements of distribution of drugs and metabolites in tissues and correlate these distributions with changes in the proteome. While many significant results have been reported over the past ten years, more widespread applications have been inhibited by limitations on the combined speed and spatial resolution available with existing instruments. In this work two new high-performance MALD-TOF instruments are evaluated for these applications. MethodThe new analyzers employ a 5 kHz laser for higher speed and better sample utilization.

A linear MALDI-TOF mass spectrometers is used for imaging proteins. This analyzer operates at ion energy of 42 keV and includes a new hybrid detector that is immune to saturation effects without compromising time resolution.

The second instrument is a reflector MS operating at 8 kV that provides very high resolving power and mass accuracy for peptides and small molecules. This instrument was used for imaging lipids in tissue samples using DHB matrixSample PlatesThese instruments employ a sample plate 124x127 mm outside dimensions with 102x108 mm active area, allowing relative large samples such as whole body tissue sections from small mammals to be imaged with relatively high resolution. For this work adaptor plates accommodating up to 4 conductive 45x45 glass slides or 45x47 metal plates were used. An image of the plate is generated by a flat-bed scanner and that image is used to define the sample to be analyzed.

Conclusions and Future WorkThis work has demonstrated that tissue images with 50 m resolution can be generated with practical speeds for many applications using a laser operating at 5 kHz. There is no apparent loss in data quality relative to earlier work at 20-200 Hz. We have demonstrated that up to 1 million spectra can be generated, stored, and processed per day, but at present data interpretation may be significantly slower than data acquisition. The images and spectra presented here represent a very small part of the total information available in one of these measurements. Further improvements in software and matrix deposition will be required to fully utilize this

capability.References1. J. A. Hankin, R. M. Barkley, R. C. Murphy, “Sublimation as a Method of Matrix Application for Mass Spectrometric Imaging”, J. Am. Soc. Mass Spectrom. 2007, 18, 1646-1652.2. P. Chaurand, J. L. Norris, D. S. Cornett, J. A. Mobley, and R. M. Caprioli, “New Developments in Profiling and Imaging of Proteins from Tissue Sections by MALDI Mass Spectrometry”, J. Proteome Res. 2006, 5, 2889-2900.AcknowledgementsThe assistance of Dr. Pierre Chaurand in providing the tissue sections with matrix deposition and the histological images of the sections.

This work was supported in part by the National Institutes of Health under grants RR025705 and GM079833.

Area Chromatogram - All Runs - db:///Marvin/Vanderbilt/Lipids3 - 52887.026 s.d. with 96 %hits

20 30 40

35

40

X Position

Y P

osit

ion

Area Chromatogram - Unknown Traversal - db:///Marvin/shannon/lipid1La 50 Most Intense V?task=0 (SpectrumDataReducer) Reduced - 14.658 s.d. with 56 %hits

75 80 85 90

84

85

86

87

88

89

90

91

X Position

Y P

ositio

n

772.6 / (798 - 802)866.7 / (798 - 802)782.7 / (798 - 802)

Area Chromatogram - All Runs - db:///Marvin/RichardVisit/linprot2 - 33.598 s.d. with 37 %hits

70 80 90 100

25

30

35

X Position

Y P

os

itio

n

8563 / (3000 - 30000)12124 / (3000 - 30000)14044 / (3000 - 30000)10158 / (3000 - 30000)15012 / (3000 - 30000)10894 / (3000 - 30000)4964 / (3000 - 30000)12414 / (3000 - 30000)

Area Chromatogram - All Runs - db:///Marvin/RichardVisit/linprot2 - 6982.828 s.d. with 42 %hits

70 80 90 100

25

30

35

X Position

Y P

osit

ion

15050 - 30000

Area Chromatogram - All Runs - db:///Marvin/RichardVisit/linprot2 - 286.258 s.d. with 7 %hits

70 80 90 100

25

30

X Position

Y P

os

itio

n

9973 / (15050 - 30000)

Untitled

7748

8377

15461

16202

24074

31829

4735963028 78637

20000 40000 60000 80000

0.000

0.002

0.004

Daltons

Volts

Group:4 Spot:13560 Shots:3000

Untitled

6112 6593

7354 7932

8554

8978

9322

9973

11328

12123

14041

15004

15737

5968 10968 15968

0.000

0.002

0.004

0.006

0.008

Daltons

Volts

Group:4 Spot:15915 Shots:500

\\domainserver\public\Marvin\MiniTof-2009_03_18_13_51\MiniTof-0007-2009_03_18_13_53_53.330-.sqlspectrum#2951

798.55

799.56

800.56

760.58

826.57

772.54

848.63782.57

828.54806.52

830.52850.62

734.57

822.57 852.58

866.65762.60 783.56768.55

790.56756.54

834.58735.57 804.49

767.56

78.004443 80.004443 82.004443 84.004443 86.004443

0.0

0.5

1.0

1.5

Microseconds

Vo

lts

\\domainserver\public\Marvin\MiniTof-2009_03_18_13_51\MiniTof-0007-2009_03_18_13_53_53.330-.sqlspectrum#2042

868.54

866.72

864.71

870.55

866.54871.56868.72

864.54 870.77

84.8000000 85.0000000

0.0

0.2

0.4

Microseconds

Volts

Area Chromatogram - All Runs - db:///Marvin/RichardVisit/linprot2 - 3130.757 s.d. with 58 %hits

70 80 90 100

25

30

35

X Position

Y P

osi

tio

n

10000 - 15020

Area Chromatogram - All Runs - db:///Marvin/RichardVisit/linprot2 - 51.142 s.d. with 4 %hits

70 80 90 100

25

30

35

X Position

Y P

osi

tio

n

16768 / (10000 - 15020)22159 / (10000 - 15020)

Untitled Linear-Linear

4748

4937

4964

51055155

702175087537

8562

8602

10157

10204

10270

10396

10463

10863

10938

10975

11324

11367

12960

13000

13063 14000

1404214994

15120

1526416767

5000 10000 15000 20000

0.000

0.002

0.004

0.006

0.008

Daltons

Volts

Group:4 Spot:17036 Shots:3000

Untitled Linear-Linear

89789905

9972

10253

11324

11535

12362

14009

14046

1499816773

16808

22159

10000 15000 20000 25000

0.0000

0.0005

0.0010

0.0015

Daltons

Volts

Group:4 Spot:15522 Shots:22000

Area Chromatogram - All Runs - db:///Marvin/RichardVisit/linprot2 - 5915.735 s.d. with 80 %hits

95.0 95.5 96.0 96.5

27.2

27.4

27.6

27.8

X Position

Y P

osi

tio

n

3000 - 30000

Area Chromatogram - All Runs - db:///Marvin/RichardVisit/linprot2 - 11507.209 s.d. with 76 %hits

70 80 90 100

25

30

X Position

Y P

osi

tio

n

2000 - 30000

Untitled Linear-Linear

2787

2825

2854

4111

4150

4180

4219

5351

5462

5607

6211

6248

6967

72608439

11202

11255

1132011400

12384

12418

12488

12627

12689

5000 10000

0.00

0.01

0.02

Daltons

Volts

Group:4 Spot:13037 Shots:20500

TIC 500-1000

772.6 866.6

50 m resolution

1

2

3

4protein

The metal plate in the upper left position contains three mouse brain sections similar to those analyzed here, but all of the workpresented was from samples on conductive glass plates. Wholemouse pup sections labeled 1-4 were analyzed for lipids usingsublimed DHB matrix as described earlier.1 This image was generated after analysis and most of the matrix has been desorbed. All 4 of these sections were analyzed in one day and generated 1,000,000 mass spectra. Data from one of these (3) is presented here.

All of the data presented on protein imaging from whole mouse pup was generated from the sample on a glass slide at lower right. Matrix was spotted with sinipinic acid at 200 mm intervals using the automatic spotter described previously.2

Example of mouse brainimage generated withprofile spectra. 250 shotsper pixel, 1 mm/s raster at50 m spacing. 100,000spectra in 1.4 hours

20,000 RPnote doublets

High Speed Imaging of lipids in mouse brain sections. 2.5 mm/s, raster at 50 m spacing, 50 shots/pixel (spectra transfer time 0.01 s) 50 pixels/s at 50 m resolution 150,000 spectra in 50 minutes

stained section for similar (but not identical)mouse brain section

772.6

772, 800,845,881

Lipid images of whole mouse pup section at 50 m resolution, 100 shots/pixel, 150,000 spectra in 1.3 hours. Note: Our mouse is backside down.

stain

Protein images of whole mousepup at 5 kHz, 37 J/pulse.Raster at 50 m spacing, 1 mm/s 500 shots/spectrum, 50x100 m pixels, 75,000 spectra in 2.2 hours.

Expanded view of 1x1.5 mmportion of TIC for whole mouse image. Sinipinic acid matrix was spottedat 200 m intervals and sample was scannedwith 50x100 m pixels. Results indicate thatmatrix spots typically less than 150 m in diaand resolution of scanning is 50 m.

200 m

brainbladder spine spinal cord

Imaging of lipids and other small molecules using high-resolution Reflector TOF employs“on-the-fly” peak detection and calibration. Spectra are only saved if they meet predetermined criteria; typically a minimum intensity within a selectedmass range. This has no effect on maximumacquisition rate (limited only by transfer timefrom digitizer to computer--typically <10 ms) butdramatically reduces storage requirements and accelerates data reduction and interpretation.Protein imaging with the linear TOF typically employs full profile spectra with 4 ns bins