Download - SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Transcript
Page 1: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with

Coronary Artery Disease

Michael Maeng, Klaus Rasmussen, Per Thayssen, Henning Kelbæk, Jan Ravkilde, Ulrik Abildgaard, Lisette

O. Jensen, Evald H. Christiansen, Knud N. Hansen, Hans-Henrik Tilsted, Peter R. Hansen, Lars R. Krusell,

Thomas Engstrøm, Jens Aarøe, Jan S. Jensen, Hans E. Bøtker, Steen D. Kristensen, Steen Z. Abildstrøm, Anne

Kaltoft, Morten Madsen, Søren P. Johnsen, Henrik T. Sørensen, Leif Thuesen & Jens F. Lassen

Page 2: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Conflicts of interests for Michael Maeng, M.D.Cordis, Johnson & Johnson:

Speaker fees

Medtronic:

Consultant fee

SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with

Coronary Artery Disease

Page 3: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Background

• Three published studies (Endeavor III, ISAR-TEST-2, ZEST-AMI) have shown that the zotarolimus-eluting Endeavor stent (END) is inferior to the sirolimus-eluting Cypher stent (CYP) with regard to angiographic endpoints - None of these studies were powered to assess clinical endpoints

Page 4: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Purpose

To compare the clinical outcome in routine clinical care (“all-comer”) patients randomized to Endeavor or Cypher stent implantation

Page 5: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

PCI Organization in Denmark

Population of 5,5 million inhabitants

5 high volume PCI centers

12-month dual anti-platelet therapy is recommended

PCI center

100 US miles

Page 6: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Methods • The study was performed within the framework of the Danish Organization for Randomized Trials with Clinical Outcome (SORT OUT)• We designed the study to reflect daily clinical practice. Therefore, no control angiography or study-related patient contact were scheduled• We used patient driven clinical event detection by use of the Danish Civil Registration system, the National Patient Registry, and the Danish Heart Registries

Page 7: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Methods

Inclusion: Indication for treatment with a DES

Exclusion: Inability to provide informed consent Life expectancy < 1 year Allergy to aspirin or clopidogrelParticipation in another trial

Page 8: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

EndpointsPrimary endpoint:

Major adverse cardiac events (MACE) defined as a composite of cardiac mortality, myocardial infarction (MI), or target vessel revascularization (TVR)

Secondary endpoints: All cause mortality Cardiac mortality MIDefinite stent thrombosis (ARC definition)TVRTarget lesion revascularization (TLR)

Page 9: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Results

• 2,332 patients were enrolled• Complete 18-month follow-up in 2,200

(94%) patients

Page 10: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Selected Patient Characteristics

END CYPNo of patients 1162 1170Age (yrs) 64 64Male (%) 73 72Diabetes (%) 15 14Hypertension (%) 54 51Lipid-lowering therapy (%) 70 68Previous CABG (%) 7 7Previous PCI (%) 21 17Previous MI (%) 26 27

Page 11: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

PCI Indication

END CYPStable angina (%) 53 51NSTEMI/UAP (%) 38 38STEMI (%) 6 9Other (%) 4 3

Page 12: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Selected Lesion Characteristics

END CYPNo of lesions/patient 1.6 1.5

Stents/patient 1.8 1.7

Lesion length (mm) 13 14

Stent length (mm) 18 18

Stent diameter (mm) 3.2 3.2

Lesion type A (%) 20 18

Lesion type B (%) 45 48

Lesion type C (%) 36 35

Page 13: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

MACE

HR = 2.19 (1.58 – 3.04)

P < 0.0001END : 9.7%

CYP : 4.5%

Page 14: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

All Cause Mortality

HR = 1.61 (1.03 – 2.50)

P = 0.035

END : 4.4%

CYP : 2.7%

Page 15: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Cardiac Death

HR = 1.51 (0.73 – 3.14)

P = 0.27

END : 1.6%

CYP : 1.0%

Page 16: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Myocardial Infarction

HR = 2.22 (1.09 – 4.53)

P = 0.029

END : 2.1%

CYP : 0.9%

Page 17: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Definite Stent Thrombosis

HR = 2.19 (0.83 – 5.77)

P = 0.13

END : 1.1%

CYP : 0.5%

Page 18: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Target Vessel Revascularization

HR = 2.42 (1.67 – 3.52)

P<0.0001END : 7.9%

CYP : 3.3%

Page 19: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Target Lesion Revascularization

HR = 3.66 (2.23 – 6.01)

P < 0.0001

END : 6.1%

CYP : 1.7%

Page 20: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

Conclusion

The Endeavor stent was inferior to the Cypher stent in routine clinical care (“all-comer”) patients at 18-month follow-up

Page 21: SORT-OUT III: A Prospective Randomized Comparison of Zotarolimus-Eluting and Sirolimus-Eluting Stents in Patients with Coronary Artery Disease

For further details …..

The SORT OUT III 18-month manuscript is published on-line in The Lancet today

Thank you for your attention