    NIST DTSA-II ( Son of DTSA ):

    Step-by-Step

    Dale E. Newbury (grateful user)National Institute of Standards and Technology

    Gaithersburg, MD 20899-8370

  • 7/18/2019 Nist Dtsa-II Guide

    2/118

    NIST-NIH Desktop Spectrum Analyzer (DTSA)

  • 7/18/2019 Nist Dtsa-II Guide

    3/118

    Photon Energy (keV)

    Log

    Intensity

    FeK

    FeK

    CaK

    CaK

    SiK AlK

    FeL

    OK

    CaL

    NIST Glass K411

    E0= 20keV

    Inside specimen

    After specimen absorption

    After EDS window absorption

    After EDS broadening (145 eV)

    Spectral Simulation with DTSA

    For 16 years, I have heard: When will you have DTSA for the pc?

  • 7/18/2019 Nist Dtsa-II Guide

    4/118

    NIST DTSA-II

    Created by Nicholas Ritchie of NIST ([email protected]),

    inspired by NIST-NIH Desktop Spectrum Analyzer (DTSA) invented

    1990-92 by Chuck Fiori (NIH and NIST) and Carol Swyt-Thomas (NIH),

    and then further developed by Carol and Bob Myklebust (NIST).

  • 7/18/2019 Nist Dtsa-II Guide

    5/118

    NIST DTSA-II

    Created by Nicholas Ritchie of NIST ([email protected]),

    inspired by NIST-NIH Desktop Spectrum Analyzer (DTSA) invented

    1990-92 by Chuck Fiori (NIH and NIST) and Carol Swyt-Thomas (NIH),

    and then further developed by Carol and Bob Myklebust (NIST).

    DTSA ran only on Macintosh, and then only up to system 10. (New

    Macs wont run DTSA) A painful question heard many, many times:

    When wil l you have DTSA for the pc? DTSA-II is the long awaited

    answer!!

  • 7/18/2019 Nist Dtsa-II Guide

    6/118

    NIST DTSA-II

    Created by Nicholas Ritchie of NIST ([email protected]),

    inspired by NIST-NIH Desktop Spectrum Analyzer (DTSA) invented

    1990-92 by Chuck Fiori (NIH and NIST) and Carol Swyt-Thomas (NIH),

    and then further developed by Carol and Bob Myklebust (NIST).

    DTSA ran only on Macintosh, and then only up to system 9. (New

    Macs wont run DTSA) A painful question heard many, many times:

    When wil l you have DTSA for the pc? DTSA-II is the long awaited

    answer!!

    DTSA-II is written in Java and operates on Mac, pc, UNIX, Linux.

  • 7/18/2019 Nist Dtsa-II Guide

    7/118

    NIST DTSA-II

    Created by Nicholas Ritchie of NIST ([email protected]),

    inspired by NIST-NIH Desktop Spectrum Analyzer (DTSA) invented

    1990-92 by Chuck Fiori (NIH and NIST) and Carol Swyt-Thomas (NIH),

    and then further developed by Carol and Bob Myklebust (NIST).

    DTSA ran only on Macintosh, and then only up to system 10. (New

    Macs wont run DTSA) A painful question heard many, many times:

    When wil l you have DTSA for the pc? DTSA-II is the long awaited

    answer!!

    DTSA-II is written in Java and operates on Mac, pc, UNIX, Linux.

    DTSA-II is NOTDTSA! Nicholas started from scratch and used DTSA

    as a guide to develop DTSA-II.

  • 7/18/2019 Nist Dtsa-II Guide

    8/118

    NIST DTSA-II

    Created by Nicholas Ritchie of NIST ([email protected]),

    inspired by NIST-NIH Desktop Spectrum Analyzer (DTSA) invented

    1990-92 by Chuck Fiori (NIH and NIST) and Carol Swyt-Thomas (NIH),

    and then further developed by Carol and Bob Myklebust (NIST).

    DTSA ran only on Macintosh, and then only up to system 10. (New

    Macs wont run DTSA) A painful question heard many, many times:

    When wil l you have DTSA for the pc? DTSA-II is the long awaited

    answer!!

    DTSA-II is written in Java and operates on Mac, pc, UNIX, Linux.

    DTSA-II is NOTDTSA! Nicholas started from scratch and used DTSA

    as a guide to develop DTSA-II.

    DTSA-II is being continually improved and the latest version can be

    downloaded for free athttp://www.cstl.nist.gov/div837/837.02/epq/dtsa2/index.html

  • 7/18/2019 Nist Dtsa-II Guide

    9/118

  • 7/18/2019 Nist Dtsa-II Guide

    10/118

    NIST DTSA-II

    Created by Nicholas Ritchie of NIST ([email protected]),

    inspired by NIST-NIH Desktop Spectrum Analyzer (DTSA) invented

    1990-92 by Chuck Fiori (NIH and NIST) and Carol Swyt-Thomas (NIH),

    and then further developed by Carol and Bob Myklebust (NIST).

    DTSA ran only on Macintosh, and then only up to system 10. (New

    Macs wont run DTSA) A painful question heard many, many times:

    When wil l you have DTSA for the pc? DTSA-II is the long awaited

    answer.

    DTSA-II is written in Java and operates on Mac, pc, UNIX, Linux.

    DTSA-II is NOTDTSA! Nicholas started from scratch and used DTSA

    as a guide to develop DTSA-II.

    DTSA-II is being continually improved and the latest version can be

    downloaded for free athttp://www.cstl.nist.gov/div837/837.02/epq/dtsa2/index.html

    Tools currently embedded in DTSA-II:

  • 7/18/2019 Nist Dtsa-II Guide

    11/118

  • 7/18/2019 Nist Dtsa-II Guide

    12/118

    Basic operations

    Opening and manipulating spectral files

  • 7/18/2019 Nist Dtsa-II Guide

    13/118

    Display and overlay spectra with various scaling options on linear/log/sqrt axes

    Basic DTSA-II display window

  • 7/18/2019 Nist Dtsa-II Guide

    14/118

    Spectrum f ile selection window

  • 7/18/2019 Nist Dtsa-II Guide

    15/118

    A single spectrum is selected.

    Spectrum preview opens in this window.

  • 7/18/2019 Nist Dtsa-II Guide

    16/118

    Display and overlay spectra with various scaling options on linear/log/sqrt axes

    OK or double click loads spectrum

    Basic display of single spectrum

    Intensity (x-ray counts)

    Energy scale in eV

    Opened

    spectra are

    listed hereSpectrum information

    from header

  • 7/18/2019 Nist Dtsa-II Guide

    17/118

    But wait!

    A brilliant feature, the Report is going to

    record your actions. A daily diary of actions

    (file named by date) is automatically saved.

  • 7/18/2019 Nist Dtsa-II Guide

    18/118

    But wait! A brilliant feature, the Report is going to record your actions. A

    daily diary of actions (file named by date) is automatically saved.

    The Report is a great

    feature for those of us

    approaching geezerhood.

    Now, what was I saying?

  • 7/18/2019 Nist Dtsa-II Guide

    19/118

    Open several spectra at a time:

    Holding down SHIFT key gives a continuous run

    Preview of spectra in this window

  • 7/18/2019 Nist Dtsa-II Guide

    20/118

    To open several spectra at a time:Holding down CTRL key allows multiple separate

    selections

  • 7/18/2019 Nist Dtsa-II Guide

    21/118

    3 spectrum files have opened

    Note data from header

    List of loadedspectra.

    Highlighted

    are displayed

    When spectra have

    differences in the

    header data, only

    the terms in common

    will be displayed

    when more than one

    spectrum is selected

  • 7/18/2019 Nist Dtsa-II Guide

    22/118

    Basic operations

    Opening and manipulating spectral files

    Display of spectra

  • 7/18/2019 Nist Dtsa-II Guide

    23/118

    Changing the vertical axis scale:

    Method 1: Use arrows: x2 or x5 /2 or /5

    This arrow button restores full spectrum display

    Method 2: Mouse in y-axis number field; click and drag up or down

    Basic display of single spectrum

  • 7/18/2019 Nist Dtsa-II Guide

    24/118

    Basic display of single spectrum

    Changing the vertical axis scale:

    Method 3

    Right click in spectrum display brings up this

    Window; choose zoom in or zoom out; this

    operation can be repeated.

    Note: Zoom to all restores full spectrum

  • 7/18/2019 Nist Dtsa-II Guide

    25/118

    Basic display of single spectrum

    Changing the vertical axis

    scale from linear to log

    1. Right click in spectrum

    window.

    2. Select Ordinate scale

    3. Select Log

  • 7/18/2019 Nist Dtsa-II Guide

    26/118

    Basic display of single spectrum

    Changing the vertical axis

    scale from linear to log

    1. Right click in spectrum window.

    2. Select Ordinate scale

    3. Select Log

  • 7/18/2019 Nist Dtsa-II Guide

    27/118

    Basic display of single spectrum

    Changing the vertical axis

    scale from linear to square root

    1. Right click in spectrum window.

    2. Select Ordinate scale

    3. Select Square root

  • 7/18/2019 Nist Dtsa-II Guide

    28/118

    Basic display of single spectrum

    Changing the horizontal axis scale: Method 1

    1. Click a preset to show 0-5, 010, 0-15, or 0-20 keV

  • 7/18/2019 Nist Dtsa-II Guide

    29/118

    Basic display of single spectrum

    Changing the horizontal axis scale: Method 1

    Click a preset to show 0-5, 010, 0-15, or 0-20 keV

    0 5 keV is shown

  • 7/18/2019 Nist Dtsa-II Guide

    30/118

    Basic display of single spectrum

    Changing the horizontal axis scale: method 2

    Right click in spectrum field

    Choose create an ROI

  • 7/18/2019 Nist Dtsa-II Guide

    31/118

    Basic display of single spectrum

    Changing the horizontal axis scale: Method 2

    1. Right click in spectrum field

    2. Choose create an ROI

    3. Define start and end energies

  • 7/18/2019 Nist Dtsa-II Guide

    32/118

    Basic display of single spectrum

    Changing the horizontal axis scale: Method 2

    1. Right click in spectrum field

    2. Choose create an ROI

    3. Define start and end energies

    4. OK This action creates yellow ROI

    5. Click this icon to fully expand this ROIor

    6. Click within ROI and select

    Zoom to region

    B i di l f i l t

  • 7/18/2019 Nist Dtsa-II Guide

    33/118

    Basic display of single spectrum

    Changing the horizontal axis scale: Method 2

    6. Final display of 3.0 5.0 keV range

    Basic display of single spectrum

  • 7/18/2019 Nist Dtsa-II Guide

    34/118

    Basic display of single spectrum

    Changing the horizontal axis scale: Method 3

    1. Double click and sweep an ROI

    (low to high energy) creating yellow ROI

    2. Immediate display of ROI parameters

    Basic display of single spectrum

  • 7/18/2019 Nist Dtsa-II Guide

    35/118

    Basic display of single spectrum

    Changing the horizontal axis scale: Method 3

    1. Double click and sweep an ROI

    (low to high energy) creating yellow ROI

    2. Immediate display of ROI parameters

    3. Click icon to expand

    OR

    4. Right click within yellow ROI

    5. Select zoom to region

    Basic display of single spectrum

  • 7/18/2019 Nist Dtsa-II Guide

    36/118

    Basic display of single spectrum

    Changing the horizontal axis scale: Method 3

    1. Double click and sweep an ROI

    (low to high energy) creating yellow ROI

    2. Immediate display of ROI parameters

    3. Click icon to expand

    OR

    4. Right click within yellow ROI

    5. Select zoom to region

    6. Final display of expanded ROI

  • 7/18/2019 Nist Dtsa-II Guide

    37/118

    Basic operations

    Opening and manipulating spectral files

    Display of spectra

    Peak labeling (manual only)

    Peak Labeling

  • 7/18/2019 Nist Dtsa-II Guide

    38/118

    KLM selection

    One spectrum

    selected from list.

    Peak Labeling

    Peak Labeling

  • 7/18/2019 Nist Dtsa-II Guide

    39/118

    To make labels stick

    click these check boxes

    Peak Labeling

    Peak Labeling

  • 7/18/2019 Nist Dtsa-II Guide

    40/118

    Labeled Spectrum

    Peak Labeling

    Peak Labeling

  • 7/18/2019 Nist Dtsa-II Guide

    41/118

    Choosing the

    peak label style

    Peak Labeling

    Peak Labeling

  • 7/18/2019 Nist Dtsa-II Guide

    42/118

    Choosing the

    peak label style

    IUPAC Long labels

    Peak Labeling

    Peak Labeling

  • 7/18/2019 Nist Dtsa-II Guide

    43/118

    Choosing the

    peak label style

    Siegbahn labels

    Peak Labeling

  • 7/18/2019 Nist Dtsa-II Guide

    44/118

    Basic operations

    Opening and manipulating spectral files

    Peak labeling (manual only)

    Exporting spectra for publication (Gnuplot)

    Export ing spectrum for publication as gnuplot

  • 7/18/2019 Nist Dtsa-II Guide

    45/118

    Export ing spectrum for publication as gnuplot

  • 7/18/2019 Nist Dtsa-II Guide

    46/118

    Publication-quality graphics from Gnuplot

    Comparing spectra from different EDS spectrometers,

  • 7/18/2019 Nist Dtsa-II Guide

    47/118

    p g p p

    or from different dates from the same EDS:

    The issue of EDS calibration A spectrum is recorded with calibration data: eV/channel, zero offset,

    number of channels (depending on manufacturer, this data may or

    may not be embedded in the .msa header).

    When a spectrum is read into DTSA-II, the calibration information ischecked against the current calibration. If there is a mismatch, a

    message prompts the analyst.

    Spectrum Calibration

  • 7/18/2019 Nist Dtsa-II Guide

    48/118

    When attempting to open

    a spectrum file with

    a different calibration,this message appears:

    Th i f EDS lib ti

  • 7/18/2019 Nist Dtsa-II Guide

    49/118

    The issue of EDS calibration A spectrum is recorded with calibration data: eV/channel, zero offset,

    number of channels (depending on manufacturer, this data may or

    may not be embedded in the .msa header).

    When a spectrum is read into DTSA-II, the calibration information is

    checked against the current calibration. If there is a mismatch, a

    message prompts the analyst.

    The analyst then has two choices:

    1. Change the detector selection to match the incoming spectrum

    2. Accept the incoming spectrum but display it according to the current

    calibration information. (Note: the incoming spectrum will retain its

    calibration data so that when it is the only spectrum being displayed, its

    own calibration will be applied.)

  • 7/18/2019 Nist Dtsa-II Guide

    50/118

    Comparing multiple spectra Matching to a particular ROI

    Comparing multiple spectra

  • 7/18/2019 Nist Dtsa-II Guide

    51/118

    We wish to match the spectra for the

    integral of SiK,

    Step 1: Click and swipe across SiK,

    peak to define ROI for matching.

    Comparing multiple spectra

  • 7/18/2019 Nist Dtsa-II Guide

    52/118

    We wish to match the spectra

    for the integral of SiK,

    Step 2: Right click and select

    Spectrum Comparison,

    then Region integral

    Comparing multiple spectra

  • 7/18/2019 Nist Dtsa-II Guide

    53/118

    Result shows Si peaks matched

    Spectrum sub sampling tool

  • 7/18/2019 Nist Dtsa-II Guide

    54/118

    Spectrum sub-sampling tool

    Take an experimentally measured spectrum and create

    one or more sub-samples, that is, equivalent spectra

    that would have been collected at lower dose.

    Sub-sampled spectra are useful for statistical studies.

    e.g., how does detection limit vary with dose.

    Spectrum sub-sampling : creating equivalent spectra for other doses

  • 7/18/2019 Nist Dtsa-II Guide

    55/118

    Starting

    measured spectrum

    Spectrum sub-sampling : creating equivalent spectra for other doses

  • 7/18/2019 Nist Dtsa-II Guide

    56/118

    Equivalent dose factor;

    set to 5% for this calculation

    Number of synthesized

    spectra, each with

    different statistics

    Random

    Number

    seed

    Spectrum sub-sampling : creating equivalent spectra for other doses

  • 7/18/2019 Nist Dtsa-II Guide

    57/118

    3 sub-sampled

    spectra added to l ist

    Note: for this example,

    5% dose lowered the

    intensity scale by a

    factor of 20, as expected

    Spectrum sub-sampling : creating equivalent spectra for other doses

  • 7/18/2019 Nist Dtsa-II Guide

    58/118

    Each sub-sampled

    spectrum is prepared

    with different random

    statistics.

  • 7/18/2019 Nist Dtsa-II Guide

    59/118

    Background fitting tool

    Background Fitting

  • 7/18/2019 Nist Dtsa-II Guide

    60/118

    Background Fitting

  • 7/18/2019 Nist Dtsa-II Guide

    61/118

    Enter material

    composition

    Background Fitting

  • 7/18/2019 Nist Dtsa-II Guide

    62/118

    Enter material

    composition

    Background Fitting

  • 7/18/2019 Nist Dtsa-II Guide

    63/118

    Fit with automatic

    placement of

    background ROIs

    Background Fitting

  • 7/18/2019 Nist Dtsa-II Guide

    64/118

    Fit with maual

    placement of

    background ROIs

    Background Fitting

  • 7/18/2019 Nist Dtsa-II Guide

    65/118

    Comparison of

    manualand automatic

    placement of

    background ROIs

    DTSA-II Simulation Mode

  • 7/18/2019 Nist Dtsa-II Guide

    66/118

    EDS spectra calculated from

    1. First principles, using best available cross sections

    and physical data (flat, bulk target only)

    Simulation Alien

  • 7/18/2019 Nist Dtsa-II Guide

    67/118

    Simulation Alien: selecting Analytical Simulation

  • 7/18/2019 Nist Dtsa-II Guide

    68/118

    Simulation Alien: specifying composit ion

  • 7/18/2019 Nist Dtsa-II Guide

    69/118

    Simulation Alien: target composition

  • 7/18/2019 Nist Dtsa-II Guide

    70/118

    Simulation Alien: instrument configuration

  • 7/18/2019 Nist Dtsa-II Guide

    71/118

    Simulation Alien: Other opt ions not invoked

  • 7/18/2019 Nist Dtsa-II Guide

    72/118

    Simulation Alien

  • 7/18/2019 Nist Dtsa-II Guide

    73/118

    Hit Finish when Progress

    bar is fi lled

    Simulation Alien

  • 7/18/2019 Nist Dtsa-II Guide

    74/118

    Physics result

    Simulation Al ien: Other options invoked

  • 7/18/2019 Nist Dtsa-II Guide

    75/118

    Counting statistics appropriate to

    the dose and spectrometer effic iency

    are applied for the specified number of

    replicates.

    Simulation Alien

  • 7/18/2019 Nist Dtsa-II Guide

    76/118

    Counting statistics appropriate to

    the dose and spectrometer efficiency

    are applied to the physics result for the

    specified number of replicates.

    Simulation Alien

  • 7/18/2019 Nist Dtsa-II Guide

    77/118

    Counting statistics appropriate to

    the dose and spectrometer efficiency

    are applied to the physics result for the

    specified number of replicates.

    Comparison of

    Physics Resultand

    after applying

    counting statistics

    DTSA-II Simulation Mode

  • 7/18/2019 Nist Dtsa-II Guide

    78/118

    EDS spectra calculated from

    1. First principles, using best available cross sections

    and physical data (flat, bulk target only)

    2. Monte Carlo electron trajectory simulation forvarious specimen configurations:

    1. Flat, bulk

    2. Layer on bulk 3. Inclusion (hemisphere) embedded in bulk

    4. Spherical particle on substrate

    5. Cubic particle on substrate

    Simulation Alien: Monte Carlo simulation of a 1 m K411 glass sphere on a C substrate

  • 7/18/2019 Nist Dtsa-II Guide

    79/118

    Simulation Alien: Monte Carlo simulation of a 1 m K411 glass sphere on a C substrate

  • 7/18/2019 Nist Dtsa-II Guide

    80/118

    Simulation Alien: Monte Carlo simulation of a 1 m K411 glass sphere on a C substrate

  • 7/18/2019 Nist Dtsa-II Guide

    81/118

    Simulation Alien: Monte Carlo simulation of a 1 m K411 glass sphere on a C substrate

  • 7/18/2019 Nist Dtsa-II Guide

    82/118

    Invoke x-ray

    generation

    images

    Simulation Alien: Monte Carlo simulation of a 1 m K411 glass sphere on a C substrate

  • 7/18/2019 Nist Dtsa-II Guide

    83/118

    Simulation Alien: Monte Carlo simulation of a 1 m K411 glass sphere on a C substrate

  • 7/18/2019 Nist Dtsa-II Guide

    84/118

    Report tab:

    Generated and

    emitted intensities

    Simulation Alien: Monte Carlo simulation of a 1 m K411 glass sphere on a C substrate

  • 7/18/2019 Nist Dtsa-II Guide

    85/118

    X-ray

    production

    images

    under

    Reporttab.

    Simulation Alien: Monte Carlo simulation of a 1 m K411 glass sphere on a C substrate

  • 7/18/2019 Nist Dtsa-II Guide

    86/118

    X-ray

    production

    images

    under

    Reporttab.Trajectory View

    Simulation Alien: Monte Carlo simulation trajectories can be viewed with Cosmo Player

  • 7/18/2019 Nist Dtsa-II Guide

    87/118

    Simulation Alien: Monte Carlo simulation of a 1 m K411 glass sphere on a C substrate

  • 7/18/2019 Nist Dtsa-II Guide

    88/118

    View along X-axis (rotated view from Y-axis)

    Simulation Alien: Monte Carlo simulation of a 1 m K411 glass sphere on a C substrate

  • 7/18/2019 Nist Dtsa-II Guide

    89/118

    View along Y-axis

    View along the beam

  • 7/18/2019 Nist Dtsa-II Guide

    90/118

    1 m

    View along the beam

  • 7/18/2019 Nist Dtsa-II Guide

    91/118

    View from bottom of particle

  • 7/18/2019 Nist Dtsa-II Guide

    92/118

    DTSA-II Simulation Mode

    EDS spectra calculated from

  • 7/18/2019 Nist Dtsa-II Guide

    93/118

    EDS spectra calculated from

    1. First principles, using best available cross sections

    and physical data (flat, bulk target only)

    2. Monte Carlo electron trajectory simulation forvarious specimen configurations:

    1. Flat, bulk

    2. Layer on bulk 3. Inclusion (hemisphere) embedded in bulk

    4. Spherical particle on substrate

    5. Cubic particle on substrate

    Simulation Alien: Monte Carlo simulation of a 1 m SiO2 hemisphere in FeS2

  • 7/18/2019 Nist Dtsa-II Guide

    94/118

    E0= 20 keV

    1-m inclusion of SiO2in FeS2

    Simulation Alien: Monte Carlo simulation of a 1 m SiO2 hemisphere in FeS2

  • 7/18/2019 Nist Dtsa-II Guide

    95/118

    E0= 20 keV

    1-m inclusion of SiO2in FeS2

    Simulation Alien: Monte Carlo simulation of a 1 m SiO2 hemisphere in FeS2

  • 7/18/2019 Nist Dtsa-II Guide

    96/118

    Simulation Alien: Monte Carlo simulation of a 1 m SiO2 hemisphere in FeS2

  • 7/18/2019 Nist Dtsa-II Guide

    97/118

    DTSA-II: Quantitative Analysis

  • 7/18/2019 Nist Dtsa-II Guide

    98/118

    ZAF analysis against standards

    Standards are used to extract neededpeak references for MLLS fit.

    Report contains pertinent data (ZAF

    factors, weight%, atom%, normalizedweight%; 1statistics)

  • 7/18/2019 Nist Dtsa-II Guide

    99/118

  • 7/18/2019 Nist Dtsa-II Guide

    100/118

  • 7/18/2019 Nist Dtsa-II Guide

    101/118

  • 7/18/2019 Nist Dtsa-II Guide

    102/118

    Select File

  • 7/18/2019 Nist Dtsa-II Guide

    103/118

    Choose a spectrum fi le

  • 7/18/2019 Nist Dtsa-II Guide

    104/118

  • 7/18/2019 Nist Dtsa-II Guide

    105/118

    Continue

    specifying

    needed

    spectra

  • 7/18/2019 Nist Dtsa-II Guide

    106/118

  • 7/18/2019 Nist Dtsa-II Guide

    107/118

    Access

    Finder

    to select

    spectra

  • 7/18/2019 Nist Dtsa-II Guide

    108/118

  • 7/18/2019 Nist Dtsa-II Guide

    109/118

  • 7/18/2019 Nist Dtsa-II Guide

    110/118

    Report contains

    ZAF details

  • 7/18/2019 Nist Dtsa-II Guide

    111/118

    Residuals after fitt ing peaks

    Note: symmetric structure of residuals

    at peak positions implies that thepeak references used are

    appropriate for this unknown.

    Comparison of

    M d d

  • 7/18/2019 Nist Dtsa-II Guide

    112/118

    Measuredand

    Simulated FeS spectra

    Note good fit for

    FeK and S K, but poorfit for FeL

    Comparison of

    Measured and

  • 7/18/2019 Nist Dtsa-II Guide

    113/118

    Measuredand

    Simulated FeS spectra

    Comparison ofMeasuredand

  • 7/18/2019 Nist Dtsa-II Guide

    114/118

    Simulated FeS spectra

    after region integral

    scaling

    Adding a new detector to the list under Preferences

  • 7/18/2019 Nist Dtsa-II Guide

    115/118

    Adding a new detector to the list under Preferences

  • 7/18/2019 Nist Dtsa-II Guide

    116/118

    Adding a new detector to the list under Preferences

  • 7/18/2019 Nist Dtsa-II Guide

    117/118

    Adding a new detector to the list under Preferences

  • 7/18/2019 Nist Dtsa-II Guide

    118/118