Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 1
Morehouse Calibration Software Instruction Manual For use with DSC-USB, 4215, 4215-HS, 4215 Plus, and HADI indicators
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 2
Table of Contents Setup ............................................................................................................................................................. 3
Install Morehouse Software Calibration ............................................................................................... 3
Windows Defender SmartScreen .......................................................................................................... 4
SSCE Runtime ........................................................................................................................................ 5
Indicator Drivers .................................................................................................................................... 5
Operating Morehouse Calibration Software ................................................................................................ 6
Launch Morehouse Calibration Software ............................................................................................. 6
Managing Load Cells ............................................................................................................................. 7
Finding Load Cell Coefficients ............................................................................................................... 9
Managing Indicators ........................................................................................................................... 11
Calibrating A Load Cell ................................................................................................................................ 12
Calibration Controls ............................................................................................................................ 13
Calibrate Load Cell .............................................................................................................................. 14
Data Collection .................................................................................................................................... 15
FAQ .............................................................................................................................................................. 16
SQL Server Compact Exception ........................................................................................................... 16
Removing Force Units ......................................................................................................................... 17
Force Summation ................................................................................................................................ 18
Missing Indicator ................................................................................................................................. 19
Missing HADI Indicator ........................................................................................................................ 20
Resetting COM Ports ........................................................................................................................... 21
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 3
Setup The latest version of the Morehouse Calibration Software can be found at https://mhforce.com/wp-
content/uploads/2021/04/Morehouse_Calibration_Software.zip.
If downloading from our site, it will need to be unzipped prior to continuing with the installation.
Install Morehouse Software Calibration Navigate to the USB drive, CD, or the unzipped files and double click on the Setup file.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 4
Windows Defender SmartScreen The installer may be blocked because it is unrecognized. Simply click More info and then Run anyway.
Follow all on screen instructions. Once successfully installed, the following screen should appear.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 5
SSCE Runtime The Morehouse Calibration Software requires the SQL Server Compact Edition runtime to be installed.
The runtime is named SSCERuntime_x86.exe or SSCERuntime_x64.exe. Select the one that corresponds
to your CPU’s architecture (typically x64).
Follow the instructions to complete installation. A similar screen should appear after a successful
installation.
Indicator Drivers Install the necessary drivers for the indicators to be used (DSC, HADI, and/or 4215). The driver installers
are in the Drivers/<Indicator Name> directory.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 6
Operating Morehouse Calibration Software Launch Morehouse Calibration Software By default, Morehouse Calibration Software is installed at C:\Program Files (x86)\Morehouse
Calibration Software\Morehouse Calibration Software. Double click the Morehouse Calibration
Software file to run the program.
Alternatively, a shortcut to the program should have been installed on the desktop.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 7
Managing Load Cells Click on Enter Calibration Coefficients to add, edit, and/or remove load cells.
Click Add to add a new load cell.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 8
Select the Coefficient Units from the drop down. And then press OK.
A new load cell is created with an empty name.
1. Enter a name where <Name Here> is.
2. Select Compression (or Tension) Units
3. Enter at least 2 coefficients
4. Optionally import from Cert via clipboard
5. Repeat for the other load type if necessary
Select any existing load cell to modify the coefficients as necessary or click the Delete button to remove
the load cell from the list.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 9
Finding Load Cell Coefficients Coefficients can be found on the load cell’s certificate. Verify the load direction at the top of the page.
The coefficients needed are listed as Bn and should be entered in Bn. Refer to images for an example.
Select Proceed to finish managing the load cells.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 10
Optional: Import from Certificate
To help reduce typos, the From Cert button was added to allow effective copy/pasting of the
coefficients. If the PC has a copy of the digital certificate on it, you can select the text as shown below
and copy (CTRL + V or Right Click->Copy) the text to your clipboard. From there, just press the From Cert
button to fill all the coefficients in.
Note: The only requirements are that the copied text includes the “B₀” text and all coefficients are
selected. There are no issues with additional text being accidentally selected.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 11
Managing Indicators Click on Search Indicators to detect and modify connected indicators.
1. Select an indicator from the list
2. Click the drop-down arrow to select the load cell
3. Select the currently connected load cell.
Click Proceed to save any changes.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 12
Calibrating A Load Cell Select the indicator connected to a load cell to calibrate. The load cell’s name is listed below the
indicator.
The indicator and load cell combo will be added to the calibration display. Please verify it is correct
before proceeding.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 13
Calibration Controls 1. Force/Mass Display – Direct read of force or mass
2. Unit of Force/Mass selection – Select the unit of measurement
3. Display Precision – Adjusts the number of digits shown on the Display
4. Reading Interval – How frequent the display is updated
5. Indicator Communication Toggle – Turn on to communicate with indicator
6. Tare All Button – Tare all active indicators simultaneously
7. Force/Mass Toggle – Turn on to calculate mass using local gravity value
8. Local Gravity – Used for mass calculations
9. Capture Data – Logs the current reading in the selected cell
10. Indicator Force / mV/V Readout – Display the indicator’s current reading
11. Indicator Force / mV/V Toggle – Toggle to view a raw or force reading of the indicator
12. Compression/Tension Toggle – Toggle based on load direction
13. Add to Force Sum – Add this indicator’s readings to the total force measurement
14. Current Indicator – Visual cue of selected indicator
15. Current Load Cell – Coefficients being used
16. Tare – Zero the readings
17. Remove Indicator – The indicator and load cell return to the indicator list
18. Indicator List – List of indicators that are currently connected with the name of the assigned load cell
19. Data Logging – Enter forces to be applied; select cell to capture data to
20. Export to Excel – Create a spreadsheet of the data logged
21. Clear Data – Clear all data from Data Logging
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 14
Calibrate Load Cell 1. Enter the various forces to be applied. A new row will be added once upon selecting the empty
cell
2. Select the corresponding cell to record data in.
3. Activate this toggle to enable data to be read from the indicator
4. Press Capture Data to log the indicator’s reading. The next cell down (2) will be highlighted
automatically. The program will record both the force reading and the mV/V reading of the
indicator. Select the drop down to choose between the two.
5. Adjust the force and repeat step 4, as necessary. Once all forces have been recorded for a run,
rotate the load cell, as necessary. Repeat step 2 for all 3 runs.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 15
Data Collection Once data desired has been collected, press Export to Excel to generate a spreadsheet with all the
logged data. Exporting to Excel requires Excel to be installed on the same computer.
The following is an example of the outputted data in force units.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 16
FAQ SQL Server Compact Exception Question: I am receiving an unhandled exception mentioning SQL Server Compact. What causes it and
how do I fix it?
Answer: The exception is due to the program being unable to load the required database components because they are not installed, or the incorrect version is installed. The fix is to run the installer that is included with the program. Refer to the SSCE Runtime section of this document for instructions.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 17
Removing Force Units Question: Our company does more than just Force and Mass calibration/measurements. Is there a way
to remove the unit label to help remove confusion for our technicians?
Answer: Yes. Enabling the “Ignore Calibration Units” checkbox on the Start/Setup tab, all unit labels are
removed from the program. Due to a technical limitation though, the Managing Load Cells window still
requires a unit to be selected.
Please note that when this feature is enabled, the unit of Force/Mass dropdown is disabled and simply
displays “As Calibrated.” Additionally, the Mass toggle and Gravity text box are also disabled. Finally,
any points captured while this feature is enabled will not record any unit of measurement data.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 18
Force Summation Question: Can I use multiple indicators and load cells to calculate force?
Answer: Yes. Force summation is supported and up to 4 indicators can be used at a time.
1. Ensure all load cell coefficients and indicators are set up. See Managing Load Cells and Managing
Indicators for instructions. 2. Select the indicators to be used in the calculation. See Calibrating a Load Cell for instructions.
a. Each selected indicator/load cell will have its own region for control and display
purposes 3. An indicator can be ignored for temporarily by unchecking the Add to Force Sum checkbox for
the indicator. 4. Turn on the Force/Mass toggle. 5. Capture data and export as before. 6. The exported data will contain both devices and how each of them contributed to the force
measurement.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 19
Missing Indicator Question: Why isn’t my indicator showing up?
Answer:
1. If using a Morehouse 4215 indicator, ensure that the “Add Line Feeds” option is enabled via the
front panel controls.
2. The indicator may not be properly connected. Please ensure the connection to the computer is
good.
3. The indicator’s driver may not have been installed. Please refer to Indicator Drivers in this document
for further instructions.
4. The indicator may be in use by another program. Please stop any other programs that may be using
the indicator or restart the PC.
5. As a last attempt, please see Resetting COM Ports.
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 20
Missing HADI Indicator Question: Why isn’t my HADI indicator being detected?
Answer:
1. Complete the steps in the Missing Indicator first
2. Open the computer’s Device Manager via Control Panel or right-clicking on the start menu and
selecting Device Manager
3. Find the Universal Serial Bus controllers device category and expand to view all devices
4. Locate the USB Serial Converter device and right click on it and select Properties.
5. Navigate to the Advanced tab and enable Load VCP
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 21
Resetting COM Ports Question: I am having unexplained indicator issues, how do I reset my COM ports?
Answer:
WARNING: These steps will also remove any other COM devices. Use with caution!
1. Open the computer’s Device Manager via Control Panel or right-clicking on the start menu and
selecting Device Manager
2. Find the Ports (COM & LPT) device category and expand to view all devices
a. USB Serial Port (HADI), USB COM Port (4215), or USB DSC Port (DSC) may be listed
Morehouse Calibration Software
Instruction Manual PM-4114 4/2022 Page 22
3. Select each listed device from the Ports (COM & LPT) category and right click and select Uninstall
Device
4. Disconnect the indicator’s USB cable from the PC and then reconnect it. If it is not automatically
detected, click on the Scan for hardware changes button at the top of the device manager
Top Related