Download - Intro to spectrophotometry and electronchemistry

Transcript
Page 1: Intro to spectrophotometry and electronchemistry

Mercury Y. Lin MD FCAP LCDR MC USN

NHCP

Page 2: Intro to spectrophotometry and electronchemistry

Is the measurement of light intensity at a selected wave length

In clinical lab, radiant energy can be measured based on

-emission

-transmission

-absorption

-light scatter

-reflection

Page 3: Intro to spectrophotometry and electronchemistry

TSample=IS/Io

TRef=IR/Io

So,

Tcompound of interest = IS/IR.

Page 4: Intro to spectrophotometry and electronchemistry

The amount of light absorbed as light passes through is:

Board Exam Q.!

Page 5: Intro to spectrophotometry and electronchemistry

Beer’s Law [concentration of a substance is directly proportional to the amount of light absorbed].

b is usually set a 1 cm, and c is in moles/liter. So [Absorbance] α [concentration]

Page 6: Intro to spectrophotometry and electronchemistry

Absorbance is only directly proportional to concentration when within linearity (range established experimentally).

Incident radiation on the substance must be monochromatic.

No chemical reaction occurs. Because of relative concentration errors at

very high and very low end of absorbance, optical system should be designed to analyze between 0.1 to 0.7 absorbance unit (20% and 80% T).

Page 7: Intro to spectrophotometry and electronchemistry
Page 8: Intro to spectrophotometry and electronchemistry

1. analyte absorbs at a unique wavelenth (uncommon)

2. analyte reacts with a reagent to produce an adduct that absorbs at unique wavelength.

3. analyte is involved in reaction that produces chromophore.

Page 9: Intro to spectrophotometry and electronchemistry

Example: Total Protein Quantitation. Assays can be designed to look for the same analyte via

different wavelengths

Lowry Assay

Bradford/Coomassie blue assay

Bicinchoninic acid Assay

Copper based, classic “Biuret” reaction

Page 10: Intro to spectrophotometry and electronchemistry

Direct Spectrophotometry: Biliruin has maximum absobance @ 455 nm. But oxyhemoglobin absorbs @ 455 nm and 575 nm. (we can mathematically subtract out oxyHb interference).

Page 11: Intro to spectrophotometry and electronchemistry

A more common way to measure bilirubin is by “Diazo” reaction

<absorbs @ 600 nm>

Board Exam Q.!

Page 12: Intro to spectrophotometry and electronchemistry

There are actually 4 types of bilirubin. α-bilirubin: unconjugated β-bilirubin: monoconjugated to glucuronic

acid γ-bilirubin: diconjugated to glucuronic acid δ-bilirubin: bound to protein.

The conjugated forms (β,γ,δ) are water

soluble. The unconjugated form (α) becomes soluble

when EtOH is present.

Page 13: Intro to spectrophotometry and electronchemistry

Now, Diazo reaction is accelerated in the presence of EtOH, So this is how we get:

#1: Diazo rxn + accelerator => total bilirubin

#2: Diazo rxn w/o accelerator => direct (conjugated ) bilirubin

Take #1-#2=indirect (unconjugated) bilirubin

Board Exam Q.!

Page 14: Intro to spectrophotometry and electronchemistry

To ensure optimal function, periodic monitoring is required in:

- Photometric Accuracy (is the set wavelength accurate?)

- Absorbance check (compare result with known standards)

- Linearity

- Stray light (e.g. light source external to the system).

Page 15: Intro to spectrophotometry and electronchemistry

Beer’s Law loses linearity at the high and low concentration ends of the relationship. For absorbances below 0.1 and above 1.0 considerable errors may be expected (but this will be dependent upon the performance characteristics of the individual spectrophotometer).

Page 16: Intro to spectrophotometry and electronchemistry

Atomic Absorption (AA) Spectrophotometry: element of interest is atomized (dissociated from its chemical bond), set to lower energy state, and its ability to absorb radiation is measured. Used to measure aluminum, copper, lithium, lead, Zn, magnesium.

Flame Emission Spectrophotometry: metal elements are subject to hot flame and its light emission is measured (e.g. lithium=red, sodium=yellow, potassium=violet). Now rarely used.

Page 17: Intro to spectrophotometry and electronchemistry

Fluorescence: when a molecule absorbs light at one wave length and reemits at a longer wavelength.

Flurometry is the measurement of emitted fluorescence light.

“Stokes Shift”: the difference between wavelength of excited light vs the wavelength of the emitted fluorescence light.

Page 18: Intro to spectrophotometry and electronchemistry

Modifying Beer’s Law, the relationship of fluorescence emission to substance concentration is

**This is equation is only valid when the solution absorbs <2% of exciting light

Page 19: Intro to spectrophotometry and electronchemistry

Fluorescence measurements are 100X – 1,000X more sensitive than absorbance measurement (due to higher “signal” and much less “noise” compared to conventional photometry).

Flow cytometry is essentially a laser-induced fluorometer + light scatter analyzer.

Board Exam Q.!

Page 20: Intro to spectrophotometry and electronchemistry

Both are methods of measuring light scatter. Mathematic equations show light scatter is

related to: -particle size -wavelength dependence -distance of observation -polarized incident light source -concentration of particle -molecular weight of particle IMHO, beyond the scope for practicing

pathologists. What we need to know are:

Page 21: Intro to spectrophotometry and electronchemistry

Are optical techniques to measure precipitate formation in Ag-Ab reactions.

Turbidimetry: is the measurement of ↓ intensity of incident light beam as it passes through a solution containing substance of interest.

Nephelometry: is the measurement of light energy reflected/scattered that is NOT in the direction of the transmitted light. (commonly 90° angle to incident light).

Generally speaking, nephelometry is more sensitive than turbidimetry with a lower limit of detection (1-10 mg/L for serum protein).

Board Exam Q.!

Page 22: Intro to spectrophotometry and electronchemistry
Page 23: Intro to spectrophotometry and electronchemistry

1. Antigen Excess: - ”too much antigens which saturate all available antibody

binding sites.” - during antigen excess, light scattering signal falls

because of immune complexes lattice are not well formed.

Board Exam Q.!

Page 24: Intro to spectrophotometry and electronchemistry

2. Matrix Effect: to some extend, all molecules scatter light. “Matrix” is all components of a material system, except the substance of interest(analyte).

Page 25: Intro to spectrophotometry and electronchemistry

Involves the measurement of current or voltage generated by the activity of specific ions.

4 analytical subtypes of electrochemistry utilized in clinical labs are:

-Potentiometry

-Voltammetry/Amperometry

-Conductometry

-Coulometry

Page 26: Intro to spectrophotometry and electronchemistry

Measures Δ potential (voltage) between 2 electrodes (half-cell)

is widely used in measurement of electrolytes and blood gas.

An electrode with a constant (“known”) voltage is termed “reference electrode.” This electrode is insensitive to the substance in the solution that is being studies. Examples of reference electrodes used in clinical labs include: calomel (HgCl)electrodes and Ag/AgCl electrodes.

The “meauring” electrode is termed “indicator electrode.”

Page 27: Intro to spectrophotometry and electronchemistry

The concentration of ion in a solution is then calculated from the Δ potential between 2 electrodes, using Nernst Equation.

E = E0 – (0.0592/n) log Cred/Cox

Cell potential measured

Standard reduction potential

Number of electrons involved in reaction

Concentration of reduced species

Concentration of oxidized species

Page 28: Intro to spectrophotometry and electronchemistry

In Ion selective electrodes(ISE), either glass or polymeric membranes are utilized to enables optimal “permselectivity” of ion of interest.

Glass electrodes are used in H+ and Na+ measurements.

Polymer membrane electrodes are available for K+,Ca2+,Li+,Mg2+,bicarb measurements.

Page 29: Intro to spectrophotometry and electronchemistry

Valinomycin: selective for K+. Board Exam Q.!

Page 30: Intro to spectrophotometry and electronchemistry
Page 31: Intro to spectrophotometry and electronchemistry

Direct ISE:

- measures analyte in undiluted whole blood or plasma sample.

- typically used in blood gas analyzers and POCT electrolyte analyzers (But also used in OCD Vitros systems).

- no centrifugation needed if using whole blood.

Page 32: Intro to spectrophotometry and electronchemistry

Indirect ISE:

- measures plasma or serum that has been preanalytically diluted with a “diluent.”

- requires that the plasma and erythrocytes are separated by centrifugation.

- typically used in high volume centralized laboratory chemistry analyzers.

- generally comparable to flame photometry (the reference method).

Page 33: Intro to spectrophotometry and electronchemistry

The principle of potentiometry is identical in both direct ISE and indirect ISE.

However, disagreement can arise in clinical situations. Example: - Normally, serum is 93% water phase and 7% nonaqueous

phase. Na is located in serum water phase only. - If the nonaqueous phase is expanded (e.g.

hyperlipidemia or hyperproteinemia), the water phase % is decreased.

- so less Na is present in a given volume of serum (even though [Na] in the water phase is unchanged] results in pseudohyponatremia. The phenomenon is seen in Indirect ISE systems.

Page 34: Intro to spectrophotometry and electronchemistry
Page 35: Intro to spectrophotometry and electronchemistry

Measures the quantity of electricity (in coulombs) required to convert an analyte to a different oxidation state.

1 coulomb = quantity of electricity that is transported in 1 sec by a 1 amp current.

Coulometry is used to measure [Cl].

Page 36: Intro to spectrophotometry and electronchemistry

To measure chloride, a constant current is applied to 2 Ag electrodes => which liberates Ag+ at a constant rate.

Cl- in solution combines with Ag+ to form insoluble AgCl. [Basically: a silver ion for every chloride ion].

Once there is excess Ag+, the reaction stops.[see Amperometry on how the systems knows when to stop]

The # of liberated Ag+ = # of Cl- in the solution. Which is calculated using Faraday’s Law.

Page 37: Intro to spectrophotometry and electronchemistry

Amperometry measures the current produced by an oxidation-reduction RXN.

This is *part 2* of [Chloride] measurement system. As Cl- are consumed and excess Ag+ are in the system, there will be a increase in current. And the endpoint is reached.

Amperometry is also used to measure pO2 in blood gas. In this setup, polypropylene membrane on the pO2 electrode allows dissolved oxygen to pass through where it reacts with platinum electrode and is reduced. The change in current is proportional to partial pressure of oxygen in the biologic specimen.

Page 38: Intro to spectrophotometry and electronchemistry

In voltammetry, potential is applied to an electrochemical cell and the resulting current is measured.

Advantage: high sensitivity, able to measure multiple analytes simultaneously in a single study.

Used in detection analytes in parts per billion range (e.g. lead and heave metal assays.)

Page 39: Intro to spectrophotometry and electronchemistry

Technique used to determine amount of a substance by measuring the conductivity of the solution/mixture.

Example(1): iSTAT measurement of hematocrit.

- Based on the principle of RBC’s lipid membranes acting as electrical insulators.

- But abnormal plasma protein concentrations will interfere with measurement. (intraop infusion of electrolyte solutions will lower baseline protein concentration and result in erroneous low Hct value by conductivity).

Page 40: Intro to spectrophotometry and electronchemistry

Example(2): “Coulter Principle” – cell counting:

- conductivity of blood cells are lower than salt (suspension) solution.

- each time a blood cell passes through a narrow orifice, there is ↓conductivity/ ↑resistance . This spike in “counted.”

Page 41: Intro to spectrophotometry and electronchemistry

Coulter method of sizing and counting particles is based on measurable changes in electrical impedance produced by nonconductive particles suspended in an electrolyte. In the sensing zone each particle displaces its own volume of electrolyte. Volume displaced is measured as a voltage pulse; the height of each pulse being proportional to the volume of the particle. The quantity of suspension drawn through the aperture is precisely controlled to allow the system to count and size particles for an exact reproducible volume.

Page 42: Intro to spectrophotometry and electronchemistry
Page 44: Intro to spectrophotometry and electronchemistry

Thank you for staying awake….