ZERO50 - MENDELSON › files › ZEUS Specifications .pdf · ZEUS-400/300/200 Specification...
Transcript of ZERO50 - MENDELSON › files › ZEUS Specifications .pdf · ZEUS-400/300/200 Specification...
-
ZERO50
ZERONE CO., LTD.
-
Electrosurgical Unit
ZERO50 Specification
ZE
RO
50
Rated Voltage AC120V or AC230V
Rated Frequency 50Hz or 60Hz
Power Consumption 150VA +10%
Fuse T2.0AL when AC120V or T1.0AL when AC230V
Protection class Class 1, Type BF
Leakage current in acc. with IEC601,Part2-2
Carrier Frequency 1.6MHz
Repeat Frequency 100Hz / 120Hz
Size (W×D×H) 200mm × 220mm × 90mm
Weight 2.6Kg
Operation Cycle 10sec ON 30sec Idle
Specification
Using Environment
Operation temperature 10℃ to 40℃
Storage temperature -10℃ to 60℃
Humidity 20% to 95% RH,
Operation altitude 700mbar ~ 1060mbar
Mode Output Power Carrier Freg. Crest Factors Duty Rate
Cut 50W / 300Ω 1.6MHz 2.1 100%
Coagulation 40W / 300Ω 1.6MHz 2.3 30%
Output Power
-
Electrosurgical Unit
Standard Accessories
Silicone patient plate
(PL02-03R)
ZE
RO
50
Disposable twin button handle
(HD02-01D)
Optional AccessoriesKnife Electrode (EL01-01D)
Ball Electrode (EL02-01D)
Needle Electrode (EL04-01D)
Needle Electrode Angled (EL04-02D)
Power Cord
(CA04-01R)
-
Electrosurgical Unit
Feature
ZE
RO
50 ① This provides necessary functions like Pure Cut, Coagulation, Under operation by one equipment
② Twin Button Handles provide remote-control function.(Select Cutting or Coagulation)
③ When cut, coagulation or functioning, it can be recognized by the mark light.
Safety Function
① A fuse built in a power circuit prevent an over current from flowing through the equipment.
② It protects patient safely by grounding the whole case and letting the leakage electric current flow on the ground which can flow into patient.
④ The equipment has an Analog slide switch by which the user can set print function for cut and coagulationeasily and safely.
-
ZEUS-100/80
ZERONE CO., LTD.
-
Electrosurgical Unit
ZEUS-100/80 Specification
ZE
US
-10
0/8
0
ZEUS-100 ZEUS-80
Rated Voltage AC120V or AC230V
Rated Frequency 50Hz or 60Hz
Power Consumption 300VA +10% 270VA +10%
Fuse T4.0AL when AC120V or T2.0AL when AC230V
Protection class Class 1, Type BF
Leakage current in acc. with IEC601,Part2-2
Carrier Frequency 1.6MHz
Repeat Frequency 100Hz / 120Hz
Size (W×D×H) 270mm × 310mm × 110mm
Weight 6Kg
Using Environment
Operation temperature 10℃ to 40℃
Storage temperature -10℃ to 60℃
Humidity 20% to 95% RH,
Operation altitude 700mbar ~ 1060mbar
Operation Cycle 10sec ON 30sec Idle
Specification
-
Electrosurgical Unit
Mode Output Power
Carrier Freg. Crest Factors Duty RateZEUS-100 ZEUS-80
Pure Cut 100W / 300Ω 80W / 300Ω 1.6MHz 1.6 100%
Blend 100W / 300Ω 80W / 300Ω 1.6MHz 2.2 60%
Contact Coagulation 60W / 300Ω 50W / 300Ω 1.6MHz 2.5 30%
Bipolar Coagulation 80W / 100Ω 70W / 100Ω 1.6MHz 1.6 100%
ZE
US
-10
0/8
0
Output Power
Standard Accessories
Silicone patient plate
(PL02-02R)
Single foot switch
(FS01-01R)
Knife Electrode (EL01-01D)
Twin Button handle
(HD02-01D)
Ball Electrode (EL02-01D)
Needle Electrode (EL04-01D)
Needle Electrode Angled (EL04-02D)
Power Cord
(CA04-01R)
-
Electrosurgical Unit
Optional AccessoriesZ
EU
S-1
00
/80
Straight 9.0cm 0.5 Tip (BF02-01R)
Angled 9.0cm 0.5 Tip (BF03-01R)
Straight 12.1cm 1.0 Tip (BF02-02R)
Straight 17.8cm 1.0 Tip (BF02-03R)
Angled 17.8cm 1.0 Tip (BF02-03R)
Bayonet 15.2c m 1.0 Tip (BF01-02R)
Disposable return
patient plate
(PL03-02D)
Bipolar forceps cable
(CA01-01R)
Bayonet 17.8cm 1.0Tip
(BF01-01R)
Return patient plate
Cable (CA02-02R)
Monopolar handle
(CA02-01R)
Cart (CT01-01R)
-
Electrosurgical Unit
Feature
① A DISPLAY window allows you to easily check output values, indicated as digits.
② This provides necessary functions like Pure Cut, Blend, Contact Coagulation, Bipoar Coagulation, under operation by one equipment
③ Holding the button for each mode for more than 3 seconds, the output ratio on Display will be changed to 1W to befacilitated to shift from the low to high output.
④ Twin Button Handles provide remote-control function. (Select Cutting or Coagulation)
⑤ Foot Switch is available for Cutting and Coagulating.
⑥ Bipolar coagulation of by Foot Switch.
⑦ Microprocessor can straighten, stabilize the output.
⑧ Operations of Cutting, Coagulation and Bipolar Coagulation can be distinct from each other by sound and indication lamp.
⑨ Each application mode (Cutting, Coagulation, Bipolar Coagulation) has a different sound, which enables you to easilydistinguish an operation type.
⑩ If the area between a patient and the pad is not appropriate, REM(Return Electrode Monitoring) gives the alarm with awarning sound and stop an operation of the product to prevent a burning incident.
⑪ The selected output ratio of Cut, Coagulation, and Bipolar Coagulation will displayed by reenergizing the device after power OFF.
ZE
US
-10
0/8
0
-
Electrosurgical Unit
Safety Function
① A fuse built in a power circuit prevent an over current from flowing through the equipment.
ZE
US
-10
0/8
0
③ REM (Return Electrode Monitoring) monitors the size of contacting area between a patient and the pad. If the size isinappropriate, it automatically blocks the high- frequency current to minimize the danger of burning incidents.
② When the plate attached to the patient is separated from the equipment, the red alarm light begins flickering. Pressing the button of Twin Button Handle or pedal of Foot Switch will stop the alarm sound and the equipment.
④ To protect a patient, the case is fully grounded so that a leakage current can flow into the earth.
-
ZEUS-200S/150
ZERONE CO., LTD.
-
Electrosurgical Unit
ZEUS-200S/150 Specification
ZE
US
-20
0S
/15
0
ZEUS-200S ZEUS-150
Rated Voltage AC120V or AC230V
Rated Frequency 50Hz or 60Hz
Power Consumption 650VA +10% 550VA +10%
Fuse T6.3AL when AC120V or T4.0AL when AC230V
Protection class Class 1, Type CF
Leakage current in acc. with IEC601,Part2-2
Carrier Frequency 400㎑, 500㎑
Repeat Frequency 33㎑
Size (W×D×H) 330mm × 330mm × 135mm
Weight 8Kg
Using Environment
Operation temperature 10℃ to 40℃
Storage temperature -10℃ to 60℃
Humidity 20% to 95% RH,
Operation altitude 700mbar ~ 1060mbar
Operation Cycle 10sec ON 30sec Idle
Cooling 1 inner fan
Specification
-
Mode Output Power
Carrier
Freg.
Crest
Factors
Duty
Rate
Repeat
Freq.ZEUS-200S ZEUS-150
Pure Cut 200W / 200Ω 150W / 200Ω 400㎑ 1.7 100% Continuous
Blend1 150W / 200Ω 120W / 200Ω 400㎑ 2.0 67% 33 ㎑
Blend2 120W / 200Ω 100W / 200Ω 400㎑ 2.7 42%33 ㎑
Contact Coagulation 80W / 200Ω 80W / 200Ω 400㎑ 2.6 25%33 ㎑
Spray Coagulation 80W / 500Ω 60W / 500Ω 400㎑ 4.4 17%33 ㎑
Bipolar Coagulation 80W / 100Ω 80W / 100Ω 500㎑ 2.0 100% Continuous
Electrosurgical Unit
ZE
US
-20
0S
/15
0 Output Power
-
Electrosurgical Unit
Standard Accessories
Double foot switch
(FS02-01R)
Disposable twin button handle
(HD02-01D)
Return patient plate cable
(CA02-02R)
Monopolar handle
(HD01-01R)
Bipolar forcep cable
(CA01-01R)
Disposable return patient
Plate ( PL03-02D)
Bayonet 17.8cm 1.0Tip (BF01-01R)
ZE
US
-20
0S
/15
0
Silicone patient plate
(PL02-01R)
Knife Electrode (EL01-01D)
Ball Electrode (EL02-01D)
Needle Electrode (EL04-01D)
Needle Electrode Angled (EL04-02D)
Loop Electrode (EL03-01D)
Power Cord
(CA04-01R)
-
Electrosurgical Unit
Optional AccessoriesStraight 9.0cm 0.5 Tip (BF02-01R)
Angled 9.0cm 0.5 Tip (BF03-01R)
Straight 12.1cm 1.0 Tip (BF02-02R)
Straight 17.8cm 1.0 Tip (BF02-03R)
Angled 17.8cm 1.0 Tip (BF02-03R)
Bayonet 15.2c m 1.0 Tip (BF01-02R)
ZE
US
-20
0S
/15
0
Cart
(CT01-01R)
-
Electrosurgical Unit
Feature
① A DISPLAY window allows you to easily check output values, indicated as digits.
⑤ Automatic selection of input power frequency 50Hz / 60HZ
④ Freely switch of Auto Mode/Manual selection.
⑥ Holding the button for each mode for more than 3 seconds, the output ratio on Display will be changed to 1W to be facilitated to shift from the low to high output.
ZE
US
-20
0S
/15
0
② One equipment provides such functions that are required to perform an operation as Functional Cutting (Pure Cut,Blend1, Blend2), Coagulation (Spray Coagulation, Contact Coagulation), Bipolar Coagulation.
③ Auto Mode function- Without separate selection of Cut and Coagulation button, it is available the Cut and Coagulation only with
Twin Button Handle and Double Foot Switch.
- Switch over display between set Cut and Coagulation output value.
⑦ Equipped with large FND, good visualization from long distance. A display window Allows You to easily check outputvalues, indicated as digits.
⑧ Foot Switch is available for Cutting and Coagulating.
⑨ Bipolar coagulation of by Foot Switch.
⑩ Microprocessor can straighten, stabilize the output.
⑪ Operations of Cutting, Coagulation and Bipolar Coagulation can be distinct from each other by sound and indication lamp.
⑫ Each application mode (Cutting, Coagulation, Bipolar Coagulation) has a different sound, which enables you to easilydistinguish an operation type.
⑬ If the area between a patient and the pad is not appropriate, REM(Return Electrode Monitoring) gives the alarm with awarning sound and stop an operation of the product to prevent a burning incident.
⑭ The level of alarm sound can be adjusted during an operation of application mode.
⑮ The selected output ratio of Cut, Coagulation, and Bipolar Coagulation will displayed by reenergizing the device afterpower OFF.
-
Electrosurgical Unit
Safety Function
① A fuse built in a power circuit prevent an over current from flowing through the equipment.
③ REM (Return Electrode Monitoring) monitors the size of contacting area between a patient and the pad. If the size isinappropriate, it automatically blocks the high- frequency current to minimize the danger of burning incidents.
④ To protect a patient, the case is fully grounded so that a leakage current can flow into the earth.
ZE
US
-20
0S
/15
0
② When the plate attached to the patient is separated from the equipment, the red alarm light begins flickering.Pressing the button of Twin Button Handle or the pedal of Cutting and Coagulation of Double Foot Switch will
stop the alarm sound and the equipment.
-
ZEUS-400/300/200
ZERONE CO., LTD.
-
Electrosurgical Unit
ZEUS-400/300/200 SpecificationZ
EU
S-4
00
/30
0/2
00
ZEUS-400 ZEUS-300 ZEUS-200
Rated Voltage AC120V or AC230V
Rated Frequency 50Hz or 60Hz
Power Consumption 1200VA +10% 1100VA +10% 1000VA +10%
Fuse T10AH when AC120V or T6.3AL when AC230V
Protection class Class 1, Type CF
Leakage current in acc. with IEC601,Part2-2
Carrier Frequency 400㎑, 500㎑
Repeat Frequency 33㎑
Size (W×D×H) 330mm × 455mm × 150mm
Weight 13Kg
Using Environment
Operation temperature 10℃ to 40℃
Storage temperature -10℃ to 60℃
Humidity 20% to 95% RH,
Operation altitude 700mbar ~ 1060mbar
Operation Cycle 10sec ON 30sec Idle
Cooling 1 inner fan
Specification
-
Mode Output Power
Carrier
Freg.
Crest
Factors
Duty
Rate
Repeat
Freq.ZEUS-400 ZEUS-300 ZEUS-200
Pure Cut 400W / 300Ω 300W / 300Ω 200W / 300Ω 400㎑ 1.6 100% Continuous
Blend1 250W / 300Ω 230W / 300Ω 150W / 300Ω 400㎑ 2.0 80% 33 ㎑
Blend2 200W / 300Ω 180W / 300Ω 120W / 300Ω 400㎑ 2.2 60% 33 ㎑
Blend3 150W / 300Ω 120W / 300Ω 100W / 300Ω 400㎑ 2.4 50% 33 ㎑
Contact Coagulation 120W / 300Ω 100W / 300Ω 100W / 300Ω33~153
㎑3.3~1.5 100%
pulse
modulated
Spray Coagulation 100W / 300Ω 80W / 300Ω 80W / 300Ω 400㎑ 4.5 8.0% 33 ㎑
Bipolar Coagulation 99W / 100Ω 80W / 100Ω 80W / 100Ω 500㎑ 1.5 100% Continuous
Electrosurgical Unit
ZE
US
-40
0/3
00
/20
0
Output Power
-
Electrosurgical Unit
Standard Accessories
Ground Cable
(CA03-01R)
Single foot switch
(FS01-02R)
ZE
US
-40
0/3
00
/20
0
Double foot switch
(FS02-01R)
Disposable twin button
Handle (HD02-01D)
Return patient plate
cable (CA02-02R)
Monopolar handle
(HD01-01R)
Bipolar forcep cable
(CA01-01R)
Disposable return patient
Plate ( PL03-02D)
Bayonet 17.8cm 1.0Tip (BF01-01R)
Silicone patient plate
(PL02-01R)
Knife Electrode (EL01-01D)
Ball Electrode (EL02-01D)
Needle Electrode (EL04-01D)
Needle Electrode Angled (EL04-02D)
Loop Electrode (EL03-01D)
Power Cord
(CA04-01R)
-
Accessories
MODEL
ZEUS-400 ZEUS-300 ZEUS-200
Double foot switch (FS02-01R) O O O
Single foot switch (FS01-02R) O O O
Disposable twin button handle (HD02-01D) O O O
Monopolar handle (HD01-01R) O O O
Return patient plate cable (CA02-02R) O O O
Silicone patient plate (PL02-01R) O O O
Bipolar forcep cable (CA01-01R) O O O
Ground Cable (CA03-01R) O O X
Bayonet 17.8cm 1.0Tip (BF01-01R) O O O
Disposable return patient Plate ( PL03-02D) O O O
Knife Electrode (EL01-01D) O O O
Needle Electrode (EL04-01D) O O O
Needle Electrode Angled (EL04-02D) O O O
Ball Electrode (EL02-01D) O O O
Loop Electrode (EL03-01D) O O O
Electrosurgical Unit
ZE
US
-40
0/3
00
/20
0 Standard Accessories
-
Electrosurgical Unit
Optional Accessories
Straight 9.0cm 0.5 Tip (BF02-01R)
Angled 9.0cm 0.5 Tip (BF03-01R)
Straight 12.1cm 1.0 Tip (BF02-02R)
Straight 17.8cm 1.0 Tip (BF02-03R)
Angled 17.8cm 1.0 Tip (BF02-03R)
Bayonet 15.2c m 1.0 Tip (BF01-02R)
ZE
US
-40
0/3
00
/20
0
Cart
(CT01-01R)
-
Electrosurgical Unit
Feature
① A DISPLAY window allows you to easily check output values, indicated as digits.
⑩ Holding the button for each mode for more than 3 seconds, the output ratio on Display will be changed to 1W to be facilitated to shift from the low to high output.
② One equipment provides such functions that are required to perform an operation as Functional Cutting (Pure Cut,Blend1, Blend2, Blend3), Coagulation (Spray Coagulation, Contact Coagulation), Bipolar Coagulation (Hand Bipolar,
Foot Bipolar).
⑧ Each application mode (Cutting, Coagulation, Bipolar Coagulation) has a different sound, which enables you to easilydistinguish an operation type.
⑨ If the area between a patient and the pad is not appropriate, REM(Return Electrode Monitoring) gives the alarm with awarning sound and stop an operation of the product to prevent a burning incident.
⑪ The level of alarm sound can be adjusted during an operation of application mode.
⑫ The selected output ratio of Cut, Coagulation, and Bipolar Coagulation will displayed by reenergizing the device afterpower OFF.
③ Two Twin Button Handles provide remote-control function. (Select Cutting or Coagulation)
④ Foot Switch is available for Cutting and Coagulating.
⑤ Hand Bipolar(Auto Bipolar) can be used only by contacting Forceps.
⑥ Microprocessor can straighten, stabilize the output.
⑦ Operations of Cutting, Coagulation and Bipolar Coagulation can be distinct from each other by sound and indication lamp.
ZE
US
-40
0/3
00
/20
0
-
Electrosurgical Unit
Safety Function
① A fuse built in a power circuit prevent an over current from flowing through the equipment.
③ REM (Return Electrode Monitoring) monitors the size of contacting area between a patient and the pad. If the size isinappropriate, it automatically blocks the high- frequency current to minimize the danger of burning incidents.
④ To protect a patient, the case is fully grounded so that a leakage current can flow into the earth.
② When the plate attached to the patient is separated from the equipment, the red alarm light begins flickering.Pressing the button of Twin Button Handle or the pedal of Cutting and Coagulation of Double Foot Switch will
stop the alarm sound and the equipment.
ZE
US
-40
0/3
00
/20
0
-
ZEUS-400PLUS
ZERONE CO., LTD.
ZEUS-300PLUSZEUS-200PLUS
-
Electrosurgical Unit
ZEUS-400P/300P/200P Specification
ZE
US
-40
0P
/30
0P
/20
0P
ZEUS-400P ZEUS-300P ZEUS-200P
Rated Voltage AC120V or AC230V
Rated Frequency 50Hz or 60Hz
Power Consumption 1200VA +10% 1100VA +10% 1000VA +10%
Fuse T10AH when AC120V or T6.3AL when AC230V
Protection class Class 1, Type CF
Leakage current in acc. with IEC601,Part2-2
Carrier Frequency 400㎑, 500㎑
Repeat Frequency 33㎑
Size (W×D×H) 330mm × 455mm × 150mm
Weight 13Kg
Using Environment
Operation temperature 10℃ to 40℃
Storage temperature -10℃ to 60℃
Humidity 20% to 95% RH,
Operation altitude 700mbar ~ 1060mbar
Operation Cycle 10sec ON 30sec Idle
Cooling 1 inner fan
Specification
-
Mode Output Power
Carrier
Freg.
Crest
Factors
Duty
Rate
Repeat
Freq.ZEUS-400P ZEUS-300P ZEUS-200P
Pure Cut 400W / 300Ω 300W / 300Ω 200W / 300Ω 400㎑ 1.6 100% Continuous
Blend1 250W / 300Ω 230W / 300Ω 150W / 300Ω 400㎑ 2.0 80% 33 ㎑
Blend2 200W / 300Ω 180W / 300Ω 120W / 300Ω 400㎑ 2.2 60% 33 ㎑
Blend3 150W / 300Ω 120W / 300Ω 100W / 300Ω 400㎑ 2.4 50% 33 ㎑
Contact Coagulation 120W / 300Ω 100W / 300Ω 100W / 300Ω33~153
㎑3.3~1.5 100%
pulse
modulated
Spray Coagulation 100W / 300Ω 80W / 300Ω 80W / 300Ω 400㎑ 4.5 8.0% 33 ㎑
Bipolar Coagulation 100W / 100Ω 80W / 100Ω 80W / 100Ω 500㎑ 1.5 100% Continuous
Bipolar Cut 100W / 200Ω 80W / 200Ω 80W / 200Ω 500㎑ 1.5 100% Continuous
Electrosurgical Unit
ZE
US
-40
0P
/30
0P
/20
0P
Output Power
-
Electrosurgical Unit
Standard Accessories
Ground Cable
(CA03-01R)
Single foot switch
(FS01-02R)
ZE
US
-40
0P
/30
0P
/20
0P
Double foot switch
(FS02-01R)
Disposable twin button handle
(HD02-01D)
Return patient plate
cable (CA02-02R)
Monopolar handle
(HD01-01R)
Bipolar forcep cable
(CA01-01R)
Disposable return patient
Plate ( PL03-02D)
Bayonet 17.8cm 1.0Tip (BF01-01R)
Silicone patient plate
(PL02-01R)
Knife Electrode (EL01-01D)
Ball Electrode (EL02-01D)
Needle Electrode (EL04-01D)
Needle Electrode Angled (EL04-02D)
Loop Electrode (EL03-01D)
Power Cord
(CA04-01R)
-
Accessories
MODEL
ZEUS-400P ZEUS-300P ZEUS-200P
Double foot switch (FS02-01R) O O O
Single foot switch (FS01-02R) O O O
Disposable twin button handle (HD02-01D) O O O
Monopolar handle (HD01-01R) O O O
Return patient plate cable (CA02-02R) O O O
Silicone patient plate (PL02-01R) O O O
Bipolar forcep cable (CA01-01R) O O O
Ground Cable (CA03-01R) O O X
Bayonet 17.8cm 1.0Tip (BF01-01R) O O O
Disposable return patient Plate ( PL03-02D) O O O
Knife Electrode (EL01-01D) O O O
Needle Electrode (EL04-01D) O O O
Needle Electrode Angled (EL04-02D) O O O
Ball Electrode (EL02-01D) O O O
Loop Electrode (EL03-01D) O O O
Electrosurgical Unit
ZE
US
-40
0P
/30
0P
/20
0P Standard Accessories
-
Electrosurgical Unit
Optional Accessories
Straight 9.0cm 0.5 Tip (BF02-01R)
Angled 9.0cm 0.5 Tip (BF03-01R)
Straight 12.1cm 1.0 Tip (BF02-02R)
Straight 17.8cm 1.0 Tip (BF02-03R)
Angled 17.8cm 1.0 Tip (BF02-03R)
Bayonet 15.2c m 1.0 Tip (BF01-02R)
ZE
US
-40
0P
/30
0P
/20
0P
Cart
(CT01-01R)
-
Electrosurgical Unit
Feature
① A DISPLAY window allows you to easily check output values, indicated as digits.
⑩ Holding the button for each mode for more than 3 seconds, the output ratio on Display will be changed to 1W to be facilitated to shift from the low to high output.
② One equipment provides such functions that are required to perform an operation as Functional Cutting (Pure Cut,Blend1, Blend2, Blend3), Coagulation (Spray Coagulation, Contact Coagulation), Bipolar Coagulation (Bipolar Cut,
Hand Bipolar, Foot Bipolar).
⑧ Each application mode (Cutting, Coagulation, Bipolar Coagulation) has a different sound, which enables you to easilydistinguish an operation type.
⑨ If the area between a patient and the pad is not appropriate, REM(Return Electrode Monitoring) gives the alarm with awarning sound and stop an operation of the product to prevent a burning incident.
⑪ The level of alarm sound can be adjusted during an operation of application mode.
⑫ The selected output ratio of Cut, Coagulation, and Bipolar Coagulation will displayed by reenergizing the device afterpower OFF.
③ Two Twin Button Handles provide remote-control function. (Select Cutting or Coagulation)
④ Foot Switch is available for Cutting and Coagulating.
⑤ Hand Bipolar(Auto Bipolar) can be used only by contacting Forceps.
⑥ Microprocessor can straighten, stabilize the output.
⑦ Operations of Cutting, Coagulation and Bipolar Coagulation can be distinct from each other by sound and indication lamp.ZE
US
-40
0P
/30
0P
/20
0P
⑬ 10 storages are available by using user storage function.
⑭ The input power frequency 50Hz / 60Hz is automatically detected.
⑮ It can be used by selecting the monopolar output (M1, M2).
-
Electrosurgical Unit
Safety Function
① A fuse built in a power circuit prevent an over current from flowing through the equipment.
③ REM (Return Electrode Monitoring) monitors the size of contacting area between a patient and the pad. If the size isinappropriate, it automatically blocks the high- frequency current to minimize the danger of burning incidents.
④ To protect a patient, the case is fully grounded so that a leakage current can flow into the earth.
② When the plate attached to the patient is separated from the equipment, the red alarm light begins flickering.Pressing the button of Twin Button Handle or the pedal of Cutting and Coagulation of Double Foot Switch will
stop the alarm sound and the equipment.
ZE
US
-40
0P
/30
0P
/20
0P
-
ZEUS Vision
ZERONE CO., LTD.
-
Electrosurgical Unit
ZEUS Vision Specification
ZE
US
Vis
ion
Rated Voltage AC120V or AC230V
Rated Frequency 50Hz or 60Hz
Power Consumption 1200VA +10%
Fuse T10AH when AC120V or T6.3AL when AC230V
Protection class Class 1, Type CF
Leakage current in acc. with IEC601,Part2-2
Carrier Frequency 400㎑, 500㎑
Repeat Frequency 33㎑
Size (W×D×H) 388mm × 410mm × 160mm
Weight 12.5Kg
Using Environment
Operation temperature 10℃ to 40℃
Storage temperature -10℃ to 60℃
Humidity 20% to 95% RH,
Operation altitude 700mbar ~ 1060mbar
Operation Cycle 10sec ON 30sec Idle
Cooling 1 inner fan
Specification
-
Electrosurgical Unit
ZE
US
Vis
ion
Mode Output Power Carrier Freg.Crest
Factors
Duty
Rate
Repeat
Freg.
Pure Cut 400W / 200Ω 400kHz 1.6 100% Continuous
Blend1 250W / 200Ω 400kHz 2.4 70% 33kHz
Blend2 200W / 200Ω 400kHz 2.3 55% 33kHz
Blend3 150W / 200Ω 400kHz 2.8 45% 33kHz
Contact Coagulation 120W / 200Ω 400kHz 3.3 40% 33KHz
Spray Coagulation 100W / 300Ω 400kHz 5.9 8.0% 33kHz
Bipolar Standard 100W / 100Ω 500kHz 1.3 100% Continuous
Bipolar Auto Start 100W / 100Ω 500kHz 1.3 100% Continuous
Bipolar Forced 80W / 100Ω 500kHz 2.7 30% 33KHz
Bipolar Cut 120W / 200Ω 500kHz 1.6 100% Continuous
Bipolar Blend 100W / 200Ω 500kHz 2.2 65% 33KHz
Output Power
-
Electrosurgical Unit
ZE
US
Vis
ion
Mode Output Power Carrier Freg.Crest
Factors
Duty
Rate
Repeat
Freg.
Soft Coagulation 80W / 100Ω 400kHz 1.4 100% Continuous
Spray Coagulation 100W / 300Ω 400kHz 5.9 8.0% 33kHz
Papillotomy 150W / 100Ω 400kHz
Polypectomy 150W / 100Ω 400kHz
Output Power(Optional)
-
Electrosurgical Unit
Standard AccessoriesZ
EU
S V
isio
n
Single foot switch
(FS01-03R)
Double foot switch
(FS02-01R)
Disposable twin button handle
(HD02-01D)
Monopolar handle
(HD01-02R)
Knife Electrode (EL01-02D)
Ball Electrode (EL02-02D)
Needle Electrode (EL04-03D)
Needle Electrode Angled (EL04-04D)
Loop Electrode (EL03-02D)
Ground Cable
(CA03-01R)
Return patient plate
cable (CA02-02R)
Bipolar forcep cable
(CA01-02R)
Disposable return patient
Plate ( PL03-02D)
Bayonet 17.8cm 1.0Tip (BF01-01R)
Silicone patient plate
(PL02-01R)
Power Cord
(CA04-01R)
-
Electrosurgical Unit
Optional AccessoriesZ
EU
S V
isio
n
Straight 9.0cm 0.5 Tip (BF02-01R)
Angled 9.0cm 0.5 Tip (BF03-01R)
Straight 12.1cm 1.0 Tip (BF02-02R)
Straight 17.8cm 1.0 Tip (BF02-03R)
Angled 17.8cm 1.0 Tip (BF02-03R)
Bayonet 15.2c m 1.0 Tip (BF01-02R)
Cart
(CT01-01R)Monopolar Connecting
Cable (CA08-01R)
-
Electrosurgical Unit
Feature
● One equipment provides such functions that are required to perform an operation as Functional Cutting (Pure Cut,
Blend1, Blend2, Blend3), Coagulation (Spray Coagulation, Contact Coagulation), Bipolar Coagulation (Bipolar Standard,
Bipolar Auto Start, Bipolar Forced, Bipolar Cut, Bipolar Blend).
● Each application mode (Cutting, Coagulation, Bipolar Coagulation) has a different sound, which enables you to easily
distinguish an operation type.
● If the area between a patient and the pad is not appropriate, REM(Return Electrode Monitoring) gives the alarm with a
warning sound and stop an operation of the product to prevent a burning incident.
● The level of alarm sound can be adjusted during an operation of application mode.
● The selected output ratio of Cut, Coagulation, and Bipolar Coagulation will displayed by reenergizing the device after
power OFF.
● Two Twin Button Handles provide remote-control function. (Select Cutting or Coagulation)
● Foot Switch is available for Cutting and Coagulating.
● Microprocessor can straighten, stabilize the output.
● Operations of Cutting, Coagulation and Bipolar Coagulation can be distinct from each other by sound and indication lamp.
● 99 storages are available by using user storage function.
● The input power frequency 50Hz / 60Hz is automatically detected.
● It can be used by selecting the monopolar output (M1, M2).
ZE
US
Vis
ion
● This equipment consists of 7 inch of TFT LCD and 5 functional keys which are placed in the front part.
● With the adoption of touch screen, it is easy to select menu, operation functions and RF output.
● Menu and operation functions can be selected by Menu Button, Esc Button, Enter Button, Up Button and Down Button.
● The ON/OFF function of Touch Screen is provided.
-
Electrosurgical Unit
Safety Function
① A fuse built in a power circuit prevent an over current from flowing through the equipment.
③ REM (Return Electrode Monitoring) monitors the size of contacting area between a patient and the pad. If the size isinappropriate, it automatically blocks the high- frequency current to minimize the danger of burning incidents.
④ To protect a patient, the case is fully grounded so that a leakage current can flow into the earth.
② When the plate attached to the patient is separated from the equipment, the red alarm light begins flickering.Pressing the button of Twin Button Handle or the pedal of Cutting and Coagulation of Double Foot Switch will
stop the alarm sound and the equipment.
ZE
US
Vis
ion
-
ARGON-Z
ZERONE CO., LTD.
-
Argon Plasma Coagulation
ARGON-Z Specification
AR
GO
N-Z Rated Voltage AC100V - AC240V
Rated Frequency 50Hz or 60Hz
Power Consumption 50VA +10%
Fuse T0.5AL 250V
Protection class Class 1, Type CF
Input Pressure Min2.5bar(250kpa) ~ Max 4.5bar(450kpa)
Minimum purity of argon 99.998%
Argon gas flow, CUT 0.1 l/min ~ 9.9 l/min
Argon gas flow, COAG. 0.1 l/min ~ 9.9 l/min
Minimum cylinder pressure 200bar(200MPa)
Weight 6Kg
Size(W× D × H) 330mm × 380mm × 130mm
Using Environment
Operation temperature 10℃ to 40℃
Storage temperature 10℃ to 40℃
Humidity 30% to 75% RH,
Operation altitude 700mbar ~ 1060mbar
Operation Cycle 10sec ON 30sec Idle
-
Standard AccessoriesA
RG
ON
-Z
Double foot switch
(FS02-02R)
Foot Switch Signal Cable
(CA05-01R)
Monopolar Cable
(CA06-01R)
Ground Cable
(CA03-02R) * 2
Luer Lock Tube
(TB01-01R)
Power Cord
(CA04-01R)
Argon Cart
(CT01-02R)
Ground Cable
(CA03-01R)
Silicone patient plate
(PL02-01R)
Reusable Flexible Argon
Probe (AP01-02R)Reusable Flexible Argon
Probe Cable (CA07-02R)
Regulator
( RG01-01R)
국가별Argon Cylinder 사용규격(CGA No.)에따라출고
Argon Plasma Coagulation
-
Optional AccessoriesA
RG
ON
-Z
Argon Electrode Handle
(HD03-01R)
Argon Cutting Electrode(177.5mm) (EL01-03R)
Cylinder 5ℓ
(CL1-01R)
Argon Needle Electrode(177.5mm) (EL04-05R)
Argon Coagulation Electrode(368.5mm)(EL05-01R)
Argon Plasma Coagulation
-
Electrosurgical Unit
Feature
AR
GO
N-Z
● Convenient LCD display on the front plate.
● Resection and coagulation by Argon Twin Button Handle
● Foot Switch is available for Cutting and Coagulating.
● Automatic adjustment of output (flow) by Microprocessor.
Safety Function
Advantage
● 9 storages are available by using user storage function.
● Distinction of cut and coagulation on display
● Even Coagulation by Contactless Coagulation
● Fast hemostasis, when massive bleeding
● Small Carbonization and almost none Sticking
● Easy and clean range of vision due to small smoke
● A fuse built in a power circuit prevent an over current from flowing through the equipment.
● Where there is no Gas, press the button of Argon Twin Button Handle or pedal of Foot Switch, and it shows warning,
generates beep sound, and stops working.
● It earths the whole case and leak current which may flow through the patient, flow through the ground to protect the
patient safely.