Page 2: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM S P6 · Technical DataType Earhook damped ThinTube

2 ccm coupler Ear simulator 2 ccm coupler Ear simulatorOutput sound pressure

levelat 1.6 kHz – 129 dB SPL – 116 dB SPL

Peak 124 dB SPL 132 dB SPL 124 dB SPL 125 dB SPLHFA-OSPL 90 121 dB SPL – 113 dB SPL –

GainFull-on gain (FOG) at 1.6 kHz – 49 dB – 48 dB

Full-on gain (Peak) 55 dB 62 dB 45 dB 53 dBHFA-FOG 42 dB – 41 dB –

Reference test gain 42 dB 42 dB 36 dB 41 dBFrequency, noise and

directivityFrequency range 100-7100 Hz 1000-7100 Hz 100-7100 Hz 280-7100 Hz

Equivalent input noise 20 dB 23 dB 15 dB 15 dBTotal harmonic distortion at

500 / 800 / 1600 Hz 2 / 1 / 1 % 2 / 1 / 1 % 1 / 1 / 2 % 1 / 1 / 2 %

AI-DI – –Inductive coil sensitivityMASL (1 mA/m) at 1.6 kHz – – – –

HFA MASL (1 mA/m) – – – –HFA SPLITS (left/right) – – – –

RSETS (left/right) – – – –AGC-O (fully activated)

Attack / release time 3 / 90 ms – 3 / 90 ms –Battery

Battery voltage 1.3 V 1.3 VBattery current drain 0.9 mA 0.9 mA

Battery life (cell zinc air) ~ 125 h ~ 125 hIRIL IEC 118-13:2011

(bystander)800 - 960 MHz < -10 dB SPL < -10 dB SPL

1400 - 2000 MHz < -10 dB SPL < -10 dB SPLANSI C63.19 M3 M3

Page 3: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM M P6 · Technical DataType Earhook damped ThinTube

2 ccm coupler Ear simulator 2 ccm coupler Ear simulatorOutput sound pressure

levelat 1.6 kHz – 133 dB SPL – 121 dB SPL

Peak 130 dB SPL 136 dB SPL 125 dB SPL 128 dB SPLHFA-OSPL 90 126 dB SPL – 116 dB SPL –

GainFull-on gain (FOG) at 1.6 kHz – 61 dB – 54 dB

Full-on gain (Peak) 60 dB 68 dB 53 dB 62 dBHFA-FOG 53 dB – 46 dB –

Reference test gain 49 dB 54 dB 39 dB 46 dBFrequency, noise and

directivityFrequency range 100-7000 Hz 320-7200 Hz 100-7000 Hz 120-7200 Hz

Equivalent input noise 19 dB 19 dB 19 dB 19 dBTotal harmonic distortion at

500 / 800 / 1600 Hz 2 / 2 / 1 % 2 / 2 / 1 % 1 / 1 / 2 % 1 / 1 / 3 %

AI-DI – –Inductive coil sensitivityMASL (1 mA/m) at 1.6 kHz – 89 dB – 82 dB

HFA MASL (1 mA/m) 82 dB – 74 dB –HFA SPLITS (left/right) 102 / 105 dB – 93 / 95 dB –

RSETS (left/right) -7 / -4 dB – -6 / -4 dB –AGC-O (fully activated)

Attack / release time 3 / 90 ms – 3 / 90 ms –Battery

Battery voltage 1.3 V 1.3 VBattery current drain 1.0 mA 1.0 mA

Battery life (cell zinc air) ~ 220 h ~ 220 hIRIL IEC 118-13:2011

(bystander)800 - 960 MHz < -20 dB SPL < -20 dB SPL

1400 - 2000 MHz < -15 dB SPL < -15 dB SPLANSI C63.19 M4 / T2 M4 / T2

Page 4: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM P P6 · Technical DataType Earhook damped Earhook undamped ThinTube

2 ccm coupler

Ear simulator

2 ccm coupler

Ear simulator

2 ccm coupler

Ear simulator

Output sound pressure level

at 1.6 kHz – 130 dB SPL – 132 dB SPL – 119 dB SPLPeak 130 dB SPL 136 dB SPL 136 dB SPL 140 dB SPL 125 dB SPL 129 dB SPL

HFA-OSPL 90 123 dB SPL – 128 dB SPL – 114 dB SPL –Gain

Full-on gain (FOG) at 1.6 kHz – 68 dB – 70 dB – 53 dB

Full-on gain (Peak) 70 dB 75 dB 75 dB 78 dB 65 dB 66 dBHFA-FOG 62 dB – 66 dB – 49 dB –

Reference test gain 47 dB 54 dB 50 dB 57 dB 37 dB 43 dBFrequency, noise and

directivityFrequency range 100-6900 Hz 150-7000 Hz 100-5800 Hz 150-6500 Hz 100-5800 Hz 100-6000 Hz

Equivalent input noise 19 dB 19 dB 19 dB 19 dB 28 dB 28 dBTotal harmonic distortion at

500 / 800 / 1600 Hz 2 / 1 / 1 % 2 / 1 / 1 % 3 / 2 / 1 % 5 / 2 / 2 % 1 / 1 / 2 % 1 / 2 / 3 %

AI-DI – – –Inductive coil sensitivityMASL (1 mA/m) at 1.6 kHz – 99 dB – 99 dB – 83 dB

HFA MASL (1 mA/m) 92 dB – 95 dB – 80 dB –HFA SPLITS (left/right) 105 / 105 dB – 108 / 108 dB – 96 / 96 dB –

RSETS (left/right) -2 / -2 dB – -2 / -2 dB – -1 / -1 dB –AGC-O (fully activated)

Attack / release time 3 / 90 ms – 3 / 90 ms – 3 / 90 ms –Battery

Battery voltage 1.3 V 1.3 V 1.3 VBattery current drain 1.0 mA 1.2 mA 1.3 mA

Battery life (cell zinc air) ~ 220 h ~ 190 h ~ 170 hIRIL IEC 118-13:2011

(bystander)800 - 960 MHz < -23 dB SPL < -23 dB SPL < -23 dB SPL

1400 - 2000 MHz < -15 dB SPL < -15 dB SPL < -15 dB SPLANSI C63.19 M4 / T2 M4 / T2 M4 / T2

Page 5: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM XP P6 · Technical DataType Earhook undamped

2 ccm coupler Ear simulatorOutput sound pressure

levelat 1.6 kHz – 139 dB SPL

Peak 140 dB SPL 144 dB SPLHFA-OSPL 90 133 dB SPL –

GainFull-on gain (FOG) at 1.6 kHz – 76 dB

Full-on gain (Peak) 80 dB 84 dBHFA-FOG 72 dB –

Reference test gain 56 dB 64 dBFrequency, noise and

directivityFrequency range 100-5600 Hz 100-5900 Hz

Equivalent input noise 24 dB SPL 26 dB SPLTotal harmonic distortion at

500 / 800 / 1600 Hz 3 / 2 / 1 % 7 / 3 / 2 %

AI-DI 3.6 dBInductive coil sensitivityMASL (1 mA/m) at 1.6 kHz – 107 dB SPL

HFA MASL (1 mA/m) 102 dB SPL –HFA SPLITS (left/right) 115 / 112 dB SPL –

RSETS (left/right) -1 / -4 dB –AGC-O (fully activated)

Attack / release time 3 / 90 ms –Battery

Battery voltage 1.3 VBattery current drain 2.4 mA

Battery life (cell zinc air) ~ 160 hIRIL IEC 118-13:2011

(bystander)800 - 960 MHz <-34 dB SPL

1400 - 2000 MHz <-34 dB SPLANSI C63.19 M3 / T4

Page 6: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

Fitting RangeXTM S XTM M

Hea

rin

g lo

ss

60

80

0

20

40

100

1200,125 0,5 1 2 843

f

dB HL

kHz

60

80

0

20

40

100

1200,125 0,5 1 2 843

f

dB HL

kHz

Hea

rin

g lo

ss

ThinTube open tip ThinTube open tip+ ThinTube double + ThinTube double+ + Earhook + + Earhook

XTM P XTM XP

60

80

0

20

40

100

1200,125 0,5 1 2 843

f

dB HL

kHz

Hea

rin

g lo

ss

60

80

0

20

40

100

1200,125 0,5 1 2 843

f

dB HL

kHz

Hea

rin

g lo

ss

ThinTube double Earhook undamped+ Earhook damped+ + Earhook undamped

Page 7: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM S P6 (Earhook damped) · Basic Data2 ccm coupler Ear simulator

100

90

130

140

120

110

80

700,125 0,50,25 1 2 8 1043

f

kHz

LO

dB SPLOutput sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

100

90

130

140

120

110

80

700,125 0,50,25 1 2 8 1043

f

kHz

LO

dB SPLOutput sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

100

90

130

140

120

110

80

700,125 0,50,25 1 2 8 1043

f

kHz

LO

dB SPLFrequency response (LI = 60 dB)

100

90

130

140

120

110

80

700,125 0,50,25 1 2 8 1043

f

kHz

LO

dB SPLBasic acoustic response (LI = 60 dB)

Page 8: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM S P6 (ThinTube) · Basic Data2 ccm coupler Ear simulator

100

90

130

140

120

110

80

700,125 0,50,25 1 2 8 1043

f

kHz

LO

dB SPLOutput sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

100

90

130

140

120

110

80

700,125 0,50,25 1 2 8 1043

f

kHz

LO

dB SPLOutput sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

100

90

130

140

120

110

80

700,125 0,50,25 1 2 8 1043

f

kHz

LO

dB SPLFrequency response (LI = 60 dB)

100

90

130

140

120

110

80

700,125 0,50,25 1 2 8 1043

f

kHz

LO

dB SPLBasic acoustic response (LI = 60 dB)

Page 9: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM M P6 (Earhook damped) · Basic Data2 ccm coupler Ear simulator

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1070

80

90

100

110

120

130

140

f

LO

Output sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1070

80

90

100

110

120

130

140

f

LO

Output sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1050

60

70

80

90

100

110

120

f

LO

Frequency response (LI = 60 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1060

70

80

90

100

110

120

130

f

LO

Basic acoustic response (LI = 60 dB)

Inductive response

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1050

60

70

80

90

100

110

120

f

LO

Inductive response(H = 10 mA/m)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1050

60

70

80

90

100

110

120

f

LO

SPLITS curve left(H = 31.6 mA/m)

SPLITS curve right(H = 31.6 mA/m)

Page 10: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM M P6 (ThinTube) · Basic Data2 ccm coupler Ear simulator

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1060

70

80

90

100

110

120

130

f

LO

Output sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1070

80

90

100

110

120

130

140

f

LO

Output sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1040

50

60

70

80

90

100

110

f

LO

Frequency response (LI = 60 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1050

60

70

80

90

100

110

120

f

LO

Basic acoustic response (LI = 60 dB)

Inductive response

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1040

50

60

70

80

90

100

110

f

LO

Inductive response(H = 10 mA/m)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1040

50

60

70

80

90

100

110

f

LO

SPLITS curve left(H = 31.6 mA/m)

SPLITS curve right(H = 31.6 mA/m)

Page 11: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM P P6 (Earhook damped) · Basic Data2 ccm coupler Ear simulator

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1070

80

90

100

110

120

130

140

f

LO

Output sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1070

80

90

100

110

120

130

140

f

LO

Output sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1050

60

70

80

90

100

110

120

f

LO

Frequency response (LI = 60 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1060

70

80

90

100

110

120

130

f

LO

Basic acoustic response (LI = 60 dB)

Inductive response

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1060

70

80

90

100

110

120

130

f

LO

Inductive response(H = 10 mA/m)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1050

60

70

80

90

100

110

120

f

LO

SPLITS curve left(H = 31.6 mA/m)

SPLITS curve right(H = 31.6 mA/m)

Page 12: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM P P6 (Earhook undamped) · Basic Data2 ccm coupler Ear simulator

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1070

80

90

100

110

120

130

140

f

LO

Output sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1080

90

100

110

120

130

140

150

f

LO

Output sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1060

70

80

90

100

110

120

130

f

LO

f

LO

Frequency response (LI = 60 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1060

70

80

90

100

110

120

130

f

LO

Basic acoustic response (LI = 60 dB)

Inductive response

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1060

70

80

90

100

110

120

130

f

LO

Inductive response(H = 10 mA/m)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1050

60

70

80

90

100

110

120

f

LO

SPLITS curve left(H = 31.6 mA/m)

SPLITS curve right(H = 31.6 mA/m)

Page 13: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM P P6 (ThinTube) · Basic Data2 ccm coupler Ear simulator

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1060

70

80

90

100

110

120

130

f

LO

Output sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1070

80

90

100

110

120

130

140

f

LO

Output sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1050

60

70

80

90

100

110

120

f

LO

Frequency response (LI = 60 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1060

70

80

90

100

110

120

130

f

LO

Basic acoustic response (LI = 60 dB)

Inductive response

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1050

60

70

80

90

100

110

120

f

LO

Inductive response(H = 10 mA/m)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1040

50

60

70

80

90

100

110

f

LO

SPLITS curve left(H = 31.6 mA/m)

SPLITS curve right(H = 31.6 mA/m)

Page 14: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM XP P6 (Earhook undamped) · Basic Data2 ccm coupler Ear simulator

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1080

90

100

110

120

130

140

150

f

LO

Output sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1080

90

100

110

120

130

140

150

f

LO

Output sound pressure level (LI = 90 dB)

Full on gain (LI = 50 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1060

70

80

90

100

110

120

130

f

LO

Frequency response (LI = 60 dB)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1070

80

90

100

110

120

130

140

f

LO

Basic acoustic response (LI = 60 dB)

Inductive response

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1070

80

90

100

110

120

130

140

f

LO

Inductive response(H = 10 mA/m)

kHz

dB SPL

0.125 0.25 0.5 1 2 3 4 8 1060

70

80

90

100

110

120

130

f

LO

SPLITS curve left(H = 31.6 mA/m)

SPLITS curve right(H = 31.6 mA/m)

Page 15: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM S/M/P/XP P6 · Features and AccessoriesEssential Facts

Signal processing (channels) / Gain/MPO (handles) 12 / 6

Hearing programs 4Wireless Sync 1) –

Volume and control coupling 1) –Speech

Voice Ranger –Automatic Directional Microphone 2)

Fixed Directional MicrophoneBandwidth Compression –

Feedback Preventer

Automatic OptimizationSmart Automatic Acclimatization –

Automatic ClassifierData Logging

Sound Quality And Comfort

Noise Management

available optional – not available

Page 16: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

XTM S/M/P/XP P6 · Features and AccessoriesS M P XP

Style Specific FeaturesSecure Tec protection IP67 IP67 – –

Telecoil –Battery Size 312 13 13 675

Battery door on/off functionNanocoated housing

Audio streaming with Smart Connect – – – –Instrument Configurations

Rocker switch / flat cover – / – / – / – / –Push button –

Battery door - direct audio input –Battery door - child lock – – – –

Small earhook –

Programming Accessories

Programming adapter / cable size 312 size 13 Programming shoe / CS44

Programming shoe / CS44

ConnexxAir, ConnexxLink 3) – – – –

AccessoriesSmart Connect 3) – – – –

Smart Remote 3) – – – –

Smart Key

Smart Transmitter (req. Smart Connect) 3) – – – –

Transmitter (req. Smart Connect) 3) – – – –

Speech Connect (req. Smart Connect) 3) – – – –

AppSmart Connect App (req. Smart Connect) 3) – – – –

Smart Remote App

available optional – not available1) Bilateral fitting required2) Directional microphone required3) Wireless option required

Page 17: XTM S/M/P/XP P6s,m,p … · ANSI C63.19 M3 / T4 . Fitting Range XTM S XTM M Hearing los s 60 80 0 20 40 100 120 0,125 0,5 12 3 4 8 f dB HL kHz 60 80 0 20 40 100 120 0,125 0,5 12 3

The information in this document contains generaldescriptions of the technical options available,which do not always have to be present in individualcases and are subject to change withoutprior notice.The required features should therefore be specifiedin each individual case at the time of conclusionof the respective contract.

Legal ManufacturerSivantos GmbH Henri-Dunant-Strasse 100, 91058 ErlangenGermany

Subject to change without prior notice

Order No. 02916-99T5-7600© 04.2018, Sivantos GmbHAll rights reserved

Abbreviations and StandardsXTM S/M/P/XP P6AbbreviationsThe following abbreviations are used in this datasheet:

Output Sound Pressure LevelOSPLHigh Frequency AverageHFAFull-On GainFOGArticulation Index Directivity IndexAI-DIMagneto Acoustical Sensitivity LevelMASLCoupler SPL for an Inductive Telephone SimulatorSPLITSRelative Equivalent Telephone SensitivityRSETSAutomatic Gain Control - OutputAGC-OInput Related Interference LevelIRILMaximum Pressure OutputMPO

Standardsu All measurements with the 2 ccm coupler were performed according to ANSI S3.22-2009 and

IEC 60118-7:2005 if applicable.u All measurements with an ear simulator were performed according to IEC 118-0/A1 and to DIN 45605

(frequency range) if applicable.u The following ear pieces were used:

- ThinTube - Earhook

WARNINGInstrument has an output sound pressure level of 132 dB SPL or more. Risk of impairing the residual hearing of the user.uTake special care when fitting this instrument.

WARNINGInstrument has an output sound pressure level of 132 dB SPL or more. Risk of impairing the residual hearing of the user.uTake special care when fitting this instrument.

WARNINGChoking hazard posed by small parts.u This instrument is not intended for the fitting of infants,

children under 3 years and persons of mental incapacity.