International Atomic Energy Agency IAEA X ray production and X ray production and Angiography Equipment Angiography Equipment L 4

description

X ray production and Angiography Equipment. L 4. Answer True or False. Radiation collimation is not necessary in the new flat panel detectors. The extra filtration in some of the new X ray systems is used to improve image quality. - PowerPoint PPT Presentation

Transcript of X ray production and Angiography Equipment

Page 1: X ray production and Angiography  Equipment

International Atomic Energy AgencyIAEA

X ray production andX ray production andAngiography EquipmentAngiography Equipment

L 4

Page 2: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 2Radiation Protection in Cardiology IAEA

Answer True or FalseAnswer True or False

1. Radiation collimation is not necessary in the new flat panel detectors.

2. The extra filtration in some of the new X ray systems is used to improve image quality.

3. Radiation dose to the patients can only be measured by a specialized person standing in the catheterization laboratory during the procedure.

Page 3: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 3Radiation Protection in Cardiology IAEA

Answer True or FalseAnswer True or False

4. When changing the field of view from 23 cm to 18 cm, it is necessary to collimate the radiation field because it is not done automatically.

5. Most of the modern X ray systems for cardiology include a transmission ionization chamber to measure staff dose.

6. Most of the modern X ray systems have the capability to produce patient dose report at the end of the procedure.

Page 4: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 4Radiation Protection in Cardiology IAEA

Educational objectivesEducational objectives

• X ray generation and imaging (fundamentals)

• What are equipment standards for catheterization equipment (FDA, IEC), particular needs for pediatric patients equipment?

Page 5: X ray production and Angiography  Equipment

International Atomic Energy AgencyIAEA

X ray Generation and image X ray Generation and image formationformation

Page 6: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 6Radiation Protection in Cardiology IAEA

Filmless X ray Cinefluorographic Unit

Page 7: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 7Radiation Protection in Cardiology IAEA

Operator / Automatic control of• tube voltage (kVp) • tube current (mA) • exposure duration

Provide• environment• source of electrons (cathode)• source of X rays (anode)• induction motor to rotate anode• heat dissipation• electrical insulation• x-ray shielding

Page 8: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 8Radiation Protection in Cardiology IAEA

A typical X-Ray Tube

Current to cathode (mA) no. of electrons liberated no. of X ray photons

Voltage across(kilovolts-peak [kVp])

energy of electrons energy of X ray photons

Page 9: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 9Radiation Protection in Cardiology IAEA

Typical Photon Energy Spectrum from a Machine Operating at KV = 80

(from The Physical Principles of Medical Imagings, 2Ed, Perry Sprawls)

Page 10: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 10Radiation Protection in Cardiology IAEA

Comparison of Photon Energy Spectra Produced at Different kVp Values

(from The Physical Principles of Medical Imagings, 2Ed, Perry Sprawls)

Page 11: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 11Radiation Protection in Cardiology IAEA

Filter

Page 12: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 12Radiation Protection in Cardiology IAEATransmission chamber and collimators

Other important elements of the X

ray systems

Page 13: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 13Radiation Protection in Cardiology IAEA

To obtain the images …To obtain the images …

• Two technologies are used:• Image intensifier• Flat panel detector

Page 14: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 14Radiation Protection in Cardiology IAEA

Motorized Iris

Video Camera

Image Intensifier

DE

TE

CT

OR

DE

TE

CT

OR

Photons

Cesium Iodide (CsI)

Light

Amorphous Silicon Panel(Photodiode/Transistor Array)

Digital DataDigital Data

Electrons

Read Out Electronics

Photons

Cesium Iodide (CsI)

Light

Photo-cathode

Video SignalVideo Signal

Electrons

Output screen

Light

CCD or PUT

Electrons

Readout Electronics

1

3,000

400

400,000

2,400

Particles #

ImageIntensifier

Flat-panel

Page 15: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 15Radiation Protection in Cardiology IAEA

Fields of view … magnificationFields of view … magnification

• Imaging detectors allow different fields of view (magnification) to improve the spatial resolution.

• This magnification usually increases the skin dose to the patient.

• Because only a part of the detector is used during magnification, the radiation field is automatically collimated to the visualized area.

Page 16: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 16Radiation Protection in Cardiology IAEA

X ray room dosimetric

information

Page 17: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 17Radiation Protection in Cardiology IAEA

High filtration High filtration

• The introduction of additional filtration in the X ray beam (commonly copper filters) reduces the number of low energy photons and, as a consequence, saves skin dose for the patients.

Page 18: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 18Radiation Protection in Cardiology IAEA

• Additional Cu filters can reduce the skin dose by more than 70%.

• Some systems offer variable extra filtration (0.2 mm - 0.9 mm) that is automatically set according to patient weight and angulation of the C-arm.

• Automatic filter insertion try to keep the dose as low as possible without degrading image quality.

Reduction of Radiation Exposure with Reduction of Radiation Exposure with extra filtrationextra filtration

Page 19: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 19Radiation Protection in Cardiology IAEA

Pulsed fluoroscopyPulsed fluoroscopy

• Pulsed fluoroscopy can be used as a method of reducing radiation dose, particularly when the pulse rate is reduced.

• But … pulsed fluoroscopy does not mean that dose rate is lower in comparison with continuous fluoroscopy!!.

• Dose rate depends on the dose per pulse and the number of pulses per second.

Page 20: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 20Radiation Protection in Cardiology IAEA

Collimation

Page 21: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 21Radiation Protection in Cardiology IAEA

Wedge filterWedge filter

Wedge filter. GE Advantx X ray system

Page 22: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 22Radiation Protection in Cardiology IAEA

The wedge filter has not been used to obtain this cine series. Note the important

difference in contrast.

The wedge filter has been used to obtain

this cine series.

Importance of wedge filtersImportance of wedge filters

Page 23: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 23Radiation Protection in Cardiology IAEA

Reduction of Radiation Exposure with Reduction of Radiation Exposure with virtual collimationvirtual collimation

• Radiation-free Collimation

• Manipulation of diaphragms in Last Image Hold

• No fluoroscopy required

Page 24: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 24Radiation Protection in Cardiology IAEA

Antiscatter gridAntiscatter grid

• To avoid that scatter radiation from the patient reaches the detector producing a degradation in image quality

• Shall be easily removable in new X ray systems (according with IEC standards)

Page 25: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 25Radiation Protection in Cardiology IAEA

Antiscatter grid Flat panel systemSiemens Axiom

Page 26: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 26Radiation Protection in Cardiology IAEA

Page 27: X ray production and Angiography  Equipment

International Atomic Energy AgencyIAEA

Equipment standardsfor Cath Lab

Page 28: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 28Radiation Protection in Cardiology IAEA

X ray Equipment Standards X ray Equipment Standards and Regulationsand Regulations

• Standards are consensus guides from the manufacturing community, not regulatory

• Several groups set standards regarding equipment, e.g., International Electrotechnical Equipment (IEC)

• Apply to electrical, mechanical, and radiation safety

• Apply to equipment at time of manufacture and installation

Page 29: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 29Radiation Protection in Cardiology IAEA

What to look for while establishing a cath. What to look for while establishing a cath. lab.lab.

• If the relevant Standards are fulfilled

• If a medical physicist is available

• If radiation protection tools are available

• If patient dose measuring and recording system is available

• If acceptance tests, commissioning and quality assurance programme have been foreseen

Page 30: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 30Radiation Protection in Cardiology IAEA

What to look for while establishing a cath. What to look for while establishing a cath. lab.lab.

• If the X rays system selected is appropriate for the procedures to be carried out in the catheterization laboratory

• If some other relevant information described in ACC/AHA Guidelines and AAPM-70 (described in this lecture) have been taken into account

Page 31: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 31Radiation Protection in Cardiology IAEA

Limitation in entrance exposure rateLimitation in entrance exposure rate

Federal Register: May 19, 1994. 21 CFR Part 1020.Federal Performance Standard for Diagnostic X-Ray Systems and Their Major Components; Final Rule.DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug Administration

Page 32: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 32Radiation Protection in Cardiology IAEA

Limitation in entrance exposure rateLimitation in entrance exposure rate

• The Standard for Diagnostic X Ray Systems (May 19, 1994), limits the entrance exposure rate of fluoroscopic X ray systems during normal fluoroscopy to 10 R/min unless an optional high-level control (HLC) is activated.

• If HLC is activated, the entrance exposure rate must be limited to 20 R/min.

• The entrance exposure rate limits do not apply during the recording of images.

Page 33: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 33Radiation Protection in Cardiology IAEA

Proposed RuleDecember 10, 2002

Page 34: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 34Radiation Protection in Cardiology IAEA

mGy (total)mGy/min

(at 15 cm from the isocenter towards the x-ray source)

Fluoroscopy time

Page 35: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 35Radiation Protection in Cardiology IAEA

Fluoroscopic equipment manufactured on or after May 19, 1995:

Shall not be operable if AKR is higher than 88 mGy/min (10 R/min).

Exceptions: •When a mode a high-level control is activated: 180 mGy/min (20 R/min). A continuous signal audible to the fluoroscopist shall indicate that the high-level control is being employed.• During the recording of images (archiving of fluoroscopic or radiographic images in analog format with a video-tape or video-disc recorder does not qualify as an exception).

Limits:88 mGy/min

180 mGy/min

Page 36: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 36Radiation Protection in Cardiology IAEA

IEC Standard 2000

Page 37: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 37Radiation Protection in Cardiology IAEA

IEC standard on Interventional IEC standard on Interventional RadiologyRadiology

• Radioscopically guided invasive (and interventional) procedures

• Interventional reference point• Isokerma maps shall be provided• The anti-scatter grid should be removable without the use of

tools

• Dosimetric indications: reference air kerma rate, cumulative reference air kerma. cumulative area kerma product, (shall be accurate to within 50 %)

• Supplementary indications: cumulative time of radioscopy, cumulative number of radiographic irradiations, integrated reference air kerma

Page 38: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 38Radiation Protection in Cardiology IAEA

Page 39: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 39Radiation Protection in Cardiology IAEA

• Collimation: Dual-shape collimators incorporating both circular and elliptical shutters may be used to modify the field for cardiac contour collimation. Partially absorbent contoured filters are also available to control the bright spots produced by the lung tissue bordering the heart.

Page 40: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 40Radiation Protection in Cardiology IAEA

Philips systems

Page 41: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 41Radiation Protection in Cardiology IAEA

Example of the influence of wedge filter in the skin dose

Page 42: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 42Radiation Protection in Cardiology IAEA

• Image intensifiers. Because of the necessity of imaging large fields (e.g., for ventriculography, aortography) as well as small fields (coronary arteries), multimode (double or triple) cesium iodide image intensifiers are recommended. Formats available vary with the manufacturer but are typically 9 in/ 6 in/4.5 in (9/6/4.5), 9/6, 10/4, and 9/5.

Page 43: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 43Radiation Protection in Cardiology IAEA

• A freely movable lead glass or acrylic shield suspended from the ceiling should be used. Its sterility may be maintained by using disposable plastic covers.

• Each procedure room should have a detailed determination of exposure levels performed by a qualified radiation physicist.

• There is a tendency in the busy laboratory to assign a low priority to preventive maintenance and quality assurance inspections.

Page 44: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 44Radiation Protection in Cardiology IAEA

Coronary interventionalists must also have a thorough knowledge of specialized equipment, techniques, and devices used to perform PCI

competently

Page 45: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 45Radiation Protection in Cardiology IAEA

AAPM-70 (2001)AAPM-70 (2001)

• The generator should be capable of generating 80 to 100 kilowatts (kW) of power.

• The generator design should result in “square wave” kVp pulses to achieve optimum patient dose savings.

Page 46: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 46Radiation Protection in Cardiology IAEA

AAPM-70 (2001)AAPM-70 (2001)

• For adult studies, a 9 to 11 inch (23 to 27 cm) size is used.

• Pediatric cardiac studies use smaller FoVs due to the small size of the pediatric heart.

• The 4.5 inch (11 cm) FoV would be commonly employed for most pediatric imaging studies.

Page 47: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 47Radiation Protection in Cardiology IAEA

AAPM-70 (2001). Pediatrics.AAPM-70 (2001). Pediatrics.

• Cine frame rate capability should extend up to at least 60 fps for small children.

• The generator should support an X ray tube with a minimum of three focal spots. Patients up to 3 to 4 years old can be imaged with an 0.3 mm focal spot size, and patients up to

• 8 to 9 years old can be imaged with cine using an 0.6 mm focal spot.

Page 48: X ray production and Angiography  Equipment

International Atomic Energy AgencyIAEA

Some self evaluation questions Some self evaluation questions ……

Page 49: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 49Radiation Protection in Cardiology IAEA

Answer True or FalseAnswer True or False

1. Collimation of the radiation field is always automatically made by the X ray system.

2. Some new systems include the feature of “virtual collimation” meaning that unnecessary radiation is removed numerically by the software.

3. Filtration in the X ray tube should be as low as possible.

Page 50: X ray production and Angiography  Equipment

International Atomic Energy AgencyIAEA

Additional informationAdditional information

Page 51: X ray production and Angiography  Equipment

International Atomic Energy AgencyIAEA

IAEA survey 2001-2003IAEA survey 2001-2003

X ray systems evaluated:

9-15 from 5 countries

Page 52: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 52Radiation Protection in Cardiology IAEA

Conclusions from the IAEA surveyConclusions from the IAEA survey

• Patient dose and image quality depend largely on the settings made at the commissioning of the radiological equipment.

• For different systems and different operation modes, entrance air kerma can increase by a factor of 20 (including electronic magnification) for the same patient thickness.

Page 53: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 53Radiation Protection in Cardiology IAEA

Conclusions from the IAEA surveyConclusions from the IAEA survey

• Increasing phantom thickness increases dose by an additional factor of up to 12.

• Differences in radiation doses from the evaluated systems show a potential for dose reduction whilst maintaining image quality.

Page 54: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 54Radiation Protection in Cardiology IAEA

Importance of testing X ray equipmentImportance of testing X ray equipment

• Characterization of the X ray system, that should be part of the acceptance and status tests, should inform cardiologists about the dose rates and dose/frame for the different operation modes and for the different patient thicknesses. Image quality shall also be evaluated.

• Regular constancy checks should verify if important changes may have occurred.

Page 55: X ray production and Angiography  Equipment

International Atomic Energy AgencyIAEA

Examples of patient dose Examples of patient dose reportsreports

Page 56: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 56Radiation Protection in Cardiology IAEA

1 CARD FIXED Coro ND 1k 7s 15F/s 15-Jan-03 09:16:21A 81kV 744mA 6.0ms 200CL small 0.3Cu 17cm 211.4µGym² 36.2mGy 0LAO 0CRA 105F

2 CARD FIXED Coro ND 1k 6s 15F/s 15-Jan-03 09:17:01A 86kV 734mA 6.0ms 600CL small 0.2Cu 17cm 376.9µGym² 63.8mGy 29RAO 0CRA 94F

3 CARD FIXED Coro ND 1k 5s 15F/s 15-Jan-03 09:17:43A 124kV 553mA 8.0ms ****** small 0.2Cu 17cm 490.3µGym² 94.1mGy 48RAO 22CRA 75F

4 CARD FIXED Coro ND 1k 6s 15F/s 15-Jan-03 09:18:16A 115kV 591mA 8.0ms ****** small 0.2Cu 17cm 460.4µGym² 97.8mGy 48RAO 22CRA 84F

5 CARD FIXED Coro ND 1k ***** 15F/s 15-Jan-03 09:19:05A 96kV 714mA 8.0ms ****** small 0.2Cu 17cm 9.3µGym² 1.9mGy 15RAO 30CRA 2F

6 CARD FIXED Coro ND 1k ***** 15F/s 15-Jan-03 09:19:07A 102kV 666mA 8.0ms ****** small 0.2Cu 17cm 17.2µGym² 3.5mGy 15RAO 30CRA 3F

Example of the data included in the study report (Siemens)

Page 57: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 57Radiation Protection in Cardiology IAEA

Example of the data included in a dosimetric report:

Philips Integris 5000.

Coronary angiography

65% cine; 35% fluoroscopy

13 series, 728 frames

1,54 Gy.cm2/min

0,368 Gy.cm2/10 fr

1 min fluoroscopy =

39 fr = 3 s cine

Page 58: X ray production and Angiography  Equipment

International Atomic Energy AgencyIAEA

Examples of information Examples of information contained in the DICOM headercontained in the DICOM header

Page 59: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 59Radiation Protection in Cardiology IAEA

Philips Integris 5000 cardio(0008,0032) : Acquisition Time : 12:36:27(0018,0060) : KVP : 83(0018,1030) : Protocol Name : 12.5 IPS Coronaria(0018,1110) : Distance Source to Detector : 940(0018,1150) : Exposure Time : 8(0018,1151) : X-ray Tube Current : 873(0018,1162) : Intensifier Size : 169.99998(0018,1510) : Positioner Primary Angle : -32.099998(0018,1511) : Positioner Secondary Angle : 0.69999999(0020,0013) : Image Number : 8(0028,0008) : Number of Frames : 73

Page 60: X ray production and Angiography  Equipment

Lecture 4: X ray production and Angiography Equipment 60Radiation Protection in Cardiology IAEA

Siemens Axiom Artis cardio (0008,0033) : Image Time : 15:38:32(0008,103E) : Series Description : Coro LD 1k T(0018,0060) : KVP : 68(0018,1110) : Distance Source to Detector : 920(0018,1111) : Distance Source to Patient : 759.99998(0018,1151) : X-ray Tube Current : 737(0018,1154) : Average Pulse Width : 6(0018,115E) : Image Area Dose Product : 1577(0018,1162) : Intensifier Size : 230(0018,1510) : Positioner Primary Angle : -0.2(0018,1511) : Positioner Secondary Angle : -0.3(0018,1702) : Collimator Left Vertical Edge : 0(0018,1704) : Collimator Right Vertical Edge: 1023(0018,1706) : Collimator Upper Horizontal Ed: 0(0018,1708) : Collimator Lower Horizontal Ed: 1023(0020,0012) : Acquisition Number : 13(0028,0008) : Number of Frames : 69