X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES...
Transcript of X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES...
![Page 1: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/1.jpg)
X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING PARTIAL LEAST SQUARES REGRESSION
V.V.Panchuk 1,2, D.O.Kirsanov1,3, A.A.Goydenko1, M.M.Khaydukova 1,3, S.M.Irkaev2, А.V.Legin 1,3, V.G.Semenov1,2
1 Institute of Chemistry, St. Petersburg State University, St. Petersburg, Russia 2 Institute for Analytical Instrumentation RAS, St. Petersburg, Russia
3 Laboratory of artificial sensory systems, ITMO University, St. Petersburg, Russia
![Page 2: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/2.jpg)
Chemometrics in XRF studies
![Page 3: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/3.jpg)
Regression models in XRF
3
1. Ordinary least squares for low concentration:
C=a0+a*Ianalyte
2. Polynomial Regression for high concentration:
C=a0+a1*Ianalyte+a2*I2analyte
3. Multiple regression for complex matrix effects:
m
analytejjjj
m
analytjjjjanalyteanalyte IeIbIIaIaaC
analyte,1
2
,1
2210
![Page 4: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/4.jpg)
Primary X‐rayradiation
Some basic principals of X‐Ray fluorescence spectroscopy
X‐ray fluorescenceradiationh = EL‐EK
4
![Page 5: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/5.jpg)
5
X‐Ray fluorescence lines
XRF techniques for liquid sample analysis:
Energy‐dispersive X‐Ray fluorescence (EDX)
Total External Reflection X‐Ray fluorescence technique (TXRF)
![Page 6: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/6.jpg)
Energy dispersive X‐Ray fluorescence technique (EDX)
Sample cup
Sample
Fluorescence radiation
X‐Ray tube
Primaryradiation
Backscattered radiation
Detector
6
Inte
nsity
Energy
![Page 7: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/7.jpg)
https://www.bruker.com/7
Total External Reflection X‐Ray fluorescencetechnique (TXRF)
(Reflector)
![Page 8: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/8.jpg)
8
TXRF Sample preparation
Sample Addition of internal standard
Taking off some l
Pipettingon reflector
Drying by evaporation
Measuring the spectra
![Page 9: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/9.jpg)
9
EDX and TXRF techniquesAdvantages of EDX
‐ Low cost
‐ No sample preparation
‐ High reproducibility spectral line intensity
Advantages of TXRF
‐ Detection limit – up to 1 part per billion
‐Many types and manufacturers of spectrometers
![Page 10: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/10.jpg)
10
The purpose of the study
Is it possible to obtain high precision of TXRF with “cheap and dirty” EDX instrumentation?
TXRF = EDX + chemometrics?
Case study – quantification of several similar elements (lanthanides) in complex mixtures at low concentration – a tricky task for EDX
![Page 11: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/11.jpg)
11
Samples40 mixtures of six lanthanides : Ce Pr Nd Sm Eu Gd
Lanthanide concentrations in mixtures were varied in a range from 10−6 to 10−3 mol/L.
4f elementsSimilar electronic structureVery similar properties
Solutions of lanthanide nitrates in 0.01 M nitric acidConcentrations in the range 10−6 to 10−3 mol/L. Uniform distribution of samples in a concentration hyperspaceTo obtain denser distribution of experimental points in the low concentration range we used logarithmic concentrations in design.
30 samples for calibration 10 samples in an independent test set
![Page 12: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/12.jpg)
12
EDX measurements
EDX spectra of lanthanides' L‐series were obtained using energy dispersive X‐ray fluorescence spectrometer Shimadzu EDX‐800HS.
5ml of sample solution was placed in a sample cup and covered by polypropylene film of 20‐micrometer thickness.
Spectra were obtained with Rh X‐Ray tube at 15 kV voltage and 310 μA current. These conditions allowed for the maximum signal/noise ratio of spectral lines.
Each spectrum was acquired for 600 s.
Spectra were smoothed by Savitzky‐Golay filter, 2 polynomial order, 5 points window.
![Page 13: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/13.jpg)
0 2 4 6 8 10 12 14
CePr
Nd
Sm
Eu
Gd
Inte
nsity
, r.u
.
Energy, keV
0,00
0,02
0,04
0,06
0,08
4,0 4,5 5,0 5,5 6,0 6,5 7,0 7,5 8,0
CePr
Nd
Sm
Eu
Gd
Inte
nsity
, r.u
.
Energy, keV
0,02
0,04
13
Problems:1. low signal/noise in spectra due to small concentrations of lanthanides2. overlapping peaks because of the similar nature of the analyzed elements
EDX spectra of lanthanides
![Page 14: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/14.jpg)
14
TXRF measurementsRigaku Nanohunter spectrometer Internal standard – copper. 20 μl of sample solution was dropped onto quartz glass reflector, dried on a heating plate at 50 °C Incidence angle of excitation radiation 0.1°.Mo X‐ray tube at 50 kV voltage and 0.6 mA current.Accumulation time 300 s for each sample.Spectra was normalization on K line copper
4,0 4,5 5,0 5,5 6,0 6,5 7,0 7,5 8,0
0,0
0,5
1,0
1,5
2,0
2,5
Inte
nsity
, a.u
.
Energy, keV
Problems:1. low signal/noise in spectra due to small
concentrations of lanthanides2. overlapping peaks from similar nature of the analyzed elements
![Page 15: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/15.jpg)
15
PLSThe spectral range 4–8 keV was employed for calculations 201 variables for EDX and 545 variables for TXRF all (L‐series lines of the analyzed lanthanides)
All models were based on 3LVThis was enough to attain 99–100% of explained variance in Y (lanthanide concentration)
OLSIn case of EDX – the area under the characteristic XRF Lα line
The area values were determined with proprietary Shimadzu software PCEDX‐E version 1.02.
In case of TXRF – Lα line intensity was employed for OLS.
Data processing
![Page 16: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/16.jpg)
16
EDX Data processing
0,0000 0,0005 0,0010
0,0000
0,0005
0,0010
Measured, mol/L
Ce
Pred
icte
d, m
ol/L
Measured, mol/L
Calibration set Test set
Ce
0,0000 0,0005 0,0010
0,0000
0,0005
0,0010
Nd
EuEu
Nd
0,0000 0,0005 0,0010
0,0000
0,0005
0,0010
0,0000 0,0005 0,0010
0,0000
0,0005
0,0010
0,0000 0,0002 0,0004 0,0006 0,0008
0,0000
0,0005
0,0010
0,0000 0,0002 0,0004 0,0006 0,0008 0,0010
0,0000
0,0005
0,0010
OLS PLS
![Page 17: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/17.jpg)
17
4 , 0 4 , 5 5 , 0 5 , 5 6 , 0 6 , 5 7 , 0 7 , 5 8 , 0
G d
E u
S m
N d
P r
C e
PLS regression coefficients for EDX spectral variables
![Page 18: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/18.jpg)
RMSEP values (mol/L) in determination of lanthanides in mixturesbased on OLS and PLS regression
18
Ce Pr Nd Sm Eu Gd
EDX OLS 1.1 × 10−4 1.6 × 10−4 1.7 × 10−4 1.5 × 10−4 2.1 × 10−4 2.5 × 10−4
EDX PLS 5.3 × 10−5 4.2 × 10−5 5.3 × 10−5 6.3 × 10−5 5.1 × 10−5 2.6 × 10−5
TXRF OLS 2.9 × 10−5 2.3 × 10−5 9.1 × 10−5 1.1 × 10−4 1.1 × 10−4 7.6 × 10−5
TXRF PLS 2.9 × 10−5 2.9 × 10−5 5.4 × 10−5 7.7 × 10−5 2.2 × 10−5 1.6 × 10−5
![Page 19: X RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES …wsc.chemometrics.ru/media/files/conferences/wsc10/presentations/... · X‐RAY FLUORESCENCE ANALYSIS OF LANTHANIDE MIXTURES USING](https://reader031.fdocuments.net/reader031/viewer/2022020104/5bba7f4209d3f2d4678dd0bd/html5/thumbnails/19.jpg)
Concluding remarks
• Due to low signal/noise ratio and overlapping signals in EDX measurements OLS fails to provide for reasonable precision in lanthanide determination especially in low concentrations.
• PLS processing of these data allowed for significant improvement of accuracy. Moreover, simple XRF method such as EDX can provide for the same precision as a more sophisticated TXRF, when the measurement's results are fitted by PLS regression.
19