Www.ihe.net PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02...

11
www.ihe.ne t www.ihe.ne t PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02 Wednesday

Transcript of Www.ihe.net PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02...

Page 1: Www.ihe.net PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02 Wednesday.

www.ihe.netwww.ihe.net

PCD Infusion Pump WGContainment / Mode / Status

Discussion (rev. 3)

Updated: 2012.05.02 Wednesday

Page 2: Www.ihe.net PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02 Wednesday.

www.ihe.net2

DS-IP-xyz Containment/Mode/StatusDiscussion Scope:

Review proposed containment models for : DS – Infusion Pump – LVP DS – Infusion Pump – Syringe DS – Infusion Pump – PCA

Define role and values for pump Mode parameter Define purpose and values for pump Status parameter

Page 3: Www.ihe.net PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02 Wednesday.

www.ihe.net3

DS-IP-LVP Containment Model

LVP VMD is paired with a generic pump delivery channel

Source Channels package semantics that define / drive the behavior of a Delivery Channel

Multiple source channels can be “active” at any point in time – thus the addition of a “bolus” channel

Source & Delivery Channel definitions consistent across pump types.

Spreadsheet => RTMMS => ICSGenerator

Page 4: Www.ihe.net PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02 Wednesday.

www.ihe.net4

DS-IP-SYR Containment Model

Syringe-specific parameters are contained in the Syringe Channel

Source & Delivery Channels are the same as those specified in the General Purpose pump specification.

Spreadsheet => RTMMS => ICSGenerator

Page 5: Www.ihe.net PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02 Wednesday.

www.ihe.net5

DS-IP-PCA Model PCA specific parameters

are packaged in a PCA Channel

PCA Channel can be combined with Source Channels as needed

Syringe-based PCA would also include a Syringe Channel (next slide)

PCA Channel may operate exclusively, sequentially or concurrently with one or more Source Channels.

Spreadsheet => RTMMS => ICSGenerator

Page 6: Www.ihe.net PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02 Wednesday.

www.ihe.net6

DS-IP-PCA-SYR Model

PCA capability added to a syringe pump

Only difference from the generic PCA pump is the addition of the Syringe Channel parameter package.

Spreadsheet => RTMMS => ICSGenerator

Page 7: Www.ihe.net PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02 Wednesday.

www.ihe.net7

Pump Mode & Status ParametersDiscussion Points:

Operational Mode & Status contained in delivery channels Mode reflects …

what source channels are actively determining the operation of their associated delivery channel

What the user would “see” as the configured “mode” All source channels OR only currently active source channels?

Status reflects … primarily what is observed in the distal line …

ISSUE #1: What do we want these values to be for DS-IP-xyz? Including active / concurrent source channel indication

ISSUE #2: Harmonization with the 11073-10101:2004 Definitions

Page 8: Www.ihe.net PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02 Wednesday.

www.ihe.net8

X73 Pump “Mode” ParameterMDC_PUMP_MODE

(from 11073-10101:2004, A.7.7.12)

hRTM Group: “Operational Mode”: pump-mode-nominal pump-mode-drug-dosing pump-mode-ramp-taper pump-mode-multi-step pump-mode-multi-dosing pump-mode-bolus pump-mode-loading-dose pump-mode-multi-channel pump-mode-pca pump-mode-continuous pump-mode-pca-and-continuous pump-mode-piggyback pump-mode-concurrent

Page 9: Www.ihe.net PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02 Wednesday.

www.ihe.net9

X73 Pump “Status” Parameter

(from 11073-10101:2004, A.7.7.18)

MDC_PUMP_STAT

hRTM Group: “Operational Mode”: pump-status-ready pump-status-infusing pump-status-paused pump-status-kvo pump-status-delayed pump-status-standby pump-status-vtbi-complete pump-status-off pump-status-priming

Page 10: Www.ihe.net PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02 Wednesday.

www.ihe.net10

Proposed Resolution Define two Operational Mode parameters + 1 status parameter MDC_PUMP_MODE_SETTING (proposed)

Indicates the configured infusion “protocol” Enumerations include: pump-mode-continuous / -piggyback / -

pca / … Multiple mode “bits” can be set Example: pump_mode_continuous + pump_mode_pca +

pump_mode_bolus MDC_PUMP_MODE_CURRENT (proposed)

Indicates the actual operational – including simultaneous – mode components (tied to Source Channels)

Example: pump_mode_continuous + pump_mode_pca MDC_PUMP_STAT (… can we make this _STATUS?)

Indicates what the pumping mechanism is doing Does NOT include alarm status indication (separate parameter) Enumerations include: pump-status-ready / -infusing / -paused /

-delayed / -standby / -off / …

Page 11: Www.ihe.net PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02 Wednesday.

www.ihe.netwww.ihe.net

PCD Rules!