WP4- NA4: Trace gases networking: Volatile organic carbon and … standards... · 2013-07-26 ·...
Transcript of WP4- NA4: Trace gases networking: Volatile organic carbon and … standards... · 2013-07-26 ·...
Measurement Guidelines VOC Version: Draft: 2012/07/18
WP4- NA4: Trace gases networking: Volatile organic carbon and nitrogen oxides
Deliverable D4.1: Draft for standardized operating procedures (SOPs) for VOC
measurements Summary:
This SOP provides a guideline for good measurement practice for the analysis of VOCs under the EU FP7
infrastructure project ACTRIS. Only active sampling is part of this SOP. For passive sampling respective
guidelines from the EU should be used.
The SOP contains the following topics:
1. General introduction....................................................................................................................................................2 2. Sampling .......................................................................................................................................................................3
2.1. off-line sampling....................................................................................................................................................3 2.1.1. VOCs...............................................................................................................................................................3 2.1.1.1 adsorption tubes..........................................................................................................................................3 2.1.1.2 stainless steel canisters................................................................................................................................3 2.1.2 OVOCs (DNPH) ................................................................................................................................................6
2.2. on-line sampling inlet line (VOCs and OVOCs)......................................................................................................6 3. Preconcentration..........................................................................................................................................................6
3.1. removal of water/O3/CO2......................................................................................................................................6 3.1.1 Water removal ................................................................................................................................................6 3.1.2 Ozone removal................................................................................................................................................7 3.1.3 CO2 removal ....................................................................................................................................................7
3.2. Trapping ................................................................................................................................................................7 3.3. capillary columns for GC analysis of VOCs and OVOC ..........................................................................................8
4. Analysis.........................................................................................................................................................................9 4.1. GC-FID....................................................................................................................................................................9 4.2 GC-MS...................................................................................................................................................................10 4.3 PTR-MS (provided by R. Holzinger Uni Utrecht and T. Petäjä) ............................................................................10
5. Quality Assurance and Quality Control......................................................................................................................11 5.1 Standards and Scale .............................................................................................................................................12 5.2 Zero Gas................................................................................................................................................................13 5.3 Data Quality Objectives........................................................................................................................................13 5.4 Method for Measurement of Standards .............................................................................................................14 5.5. Method for Measurement of zero gas (blanks)..................................................................................................16 5.6. Method for Detecting Effects of Ozone on Reactive Compounds .....................................................................17 5.7 Logs at each station..............................................................................................................................................18
6. Post-analysis ...............................................................................................................................................................19 6.1 Data checks of final mixing ratio data..................................................................................................................19 6.2 Uncertainty evaluation.........................................................................................................................................24
6.2.1. Calculation of mixing ratios for linear detection systems ...........................................................................24 6.2.2 Determination of Precision...........................................................................................................................25 6.2.3 Determination of Uncertainty ......................................................................................................................25 6.2.4 Determination of detection limit..................................................................................................................29
6.3 Data submission ...................................................................................................................................................33 7.References...................................................................................................................................................................33 8. Appendices .................................................................................................................................................................36
Measurement Guidelines VOC Version: Draft: 2012/07/18
1. General introduction
The scientific background for the need of VOC monitoring in global and regional networks has been
extensively presented for example in the GAW Reports 111 and 171, Helmig et al., EOS, 2009). In
brief summary, VOCs are, besides CO, the major group of reduced gaseous carbon-containing
compounds emitted into the atmosphere by natural and anthropogenic sources. Typically, more than
100 compounds can be measured with mixing ratios in the low ppt range up to a few ppb. They have
multiple roles in atmospheric chemistry, and of main interest is their contribution in photochemical
processes like generation of photo-oxidants, e.g. ozone, their impact on the oxidizing capacity of the
atmosphere and their contribution to secondary organic aerosol (SOA) production. Thus, they couple
into photochemical ozone production, aerosol formation and cloud processes and thus impact air
quality and climate. Therefore, measurements of VOCs are essential and are thus among the long-
term monitoring parameters in GAW (GAW Report 172) and regional programs like EMEP.
There are three groups of VOCs that are distinguished in current literature, GAW Reports and in this
measurement guideline: the non-methane hydrocarbons (NMHCs), the oxygenated organic
compounds (OVOCs) and biogenic VOCs (mainly isoprene and the group of monoterpenes, often
summarized as BVOCs). Key substances of these groups have been identified in GAW Report 171 and
detailed guidelines for their measurements are provided in this paper, following the general QA
recommendations and the strategic plan by GAW (GAW Report 172). As analytical systems for the
measurement of the different groups of VOCs (see above) are generally capable of analyzing not only
the key species identified in GAW Report 171 but also a list of chemically similar compounds, this
guideline covers a broader range of compounds than specified in GAW Report 171. This is in line with
the EMEP objectives and addressed in the ACTRIS Description of Work.
Measurement of VOCs is mostly done by gas chromatographic methods, but there also exist well
established methods for carbonyl OVOCs by DNPH cartridge sampling and HPLC and the relatively
new on-line method PTR-MS. As gas chromatography is most widely used, this method will be in the
focus of this measurement guideline but the other techniques are also covered.
The measurement of VOCs by gas chromatography is generally performed in a series of steps with (1)
intake manifold and sampling line, (2) traps to remove water and ozone, (3) pre-concentration, (4)
gas chromatographic separation, (5) analysis in detector, and (6) data processing and data delivery. A
sample of atmospheric VOCs can be introduced to the analytical system directly from ambient air, a
canister or a preconcentration tube. The sample is normally passed through a moisture
removal/ozone removal system and then concentrated using an adsorbent medium that is
cryogenically cooled using liquid nitrogen, liquid carbon dioxide, or thermoelectric closed-cycle
coolers. The sample optionally can be refocussed cryogenically by a cooled secondary trap to narrow
the band width for injection onto the capillary GC analytical column. The concentrated sample is then
thermally desorbed into the analytical gaschromatographic column and finally analysed by FID or MS
(or any other suitable detector).
For quantification a standard should be used which either contains VOCs in ambient air, or contains
artificially mixed VOCs in N2. Concentrations should be in the region of the expected ambient
concentrations. If needed the standard has to be diluted with zero air or N2 into a concentration
which is in the range of concentrations measured at the specific station.
Measurement Guidelines VOC Version: Draft: 2012/07/18
2. Sampling
The air from which VOCs are analysed can be sampled directly at the measurement site or off-line,
using either adsorption tubes or stainless steel canisters. Samples are subsequently transported to
the lab where they are analyised. The specific requirements of the different methods are described
below.
2.1. off-line sampling
2.1.1. VOCs
2.1.1.1 adsorption tubes
Off-line sampling of NMHCs by adsorption tubes is a well-established method. Different providers have
commercially available products. For this draft it is advised to follow the specific prodecures of the
individual products.
2.1.1.2 stainless steel canisters
For analyzing NMHCs in ambient air using canisters the procedures described in a new Standard
Operation Procedure (SOP) valid in the future for the WMO GAW network may be applied.
Publication by WMO is expected to take place in the second half of 2012. For immediate further
information on the SOP contact the WCC-VOC (imk-ifu.kit.edu/wcc-voc). In the following only the
sampling procedure is described as listed in this SOP. It has to be noted, that a sound analyses of
NMHCs in air samples requires carefully preconditioning steps for the canisters. These procedures
are described in detail in the above mentioned SOP. This SOP is based largely on the
recommendations from the “Accurate Measurements of Hydrocarbons in the Atmosphere” project
AMOHA (Plass-Dülmer et al., 2006) and recommendations from US-EPA (1998, 1999) on
determination of volatile organic compounds (VOCs) in ambient air. Further, beside this SOP, many
other methods exist for whole air sampling for VOC analysis (e.g. glass flask air sampling in the US
National Oceanic and Atmospheric Administration Cooperative Global Air Sampling Network
(Pollmann et al., 2008)).
General considerations
Generally, the use of materials other than stainless steel, glass, silica coated stainless steel, PFA and
PTFE should be avoided for the measurement of NMHCs in air samples. Especially, plastics other than
PFA and PTFE shall not be used to prevent memory effects. The recommendations given below
constrain on electro-polished stainless steel (ss) canisters. A variety of ss canisters with one or two
valves may be purchased from several suppliers, e.g. Restek or TO-Can® Air Monitoring Canisters. The
use of two valve canisters allows i.e. more flexibility in air sampling and is recommended. The inner
surface of canisters is passivated, e.g. electro-polished. Stainless steel valves (e.g. Swagelok) shall be
used to seal the canisters inlets and outlets, respectively. The sampling procedures described here
are valid for two valve canisters.
Air samples collected for measurements of NMHCs in small canisters shall be pressurized. Pumps
used for filling the canisters should reach a final pressure of > 3 bar ensuring a minimum gas flow
rate at ambient pressure of > 10 l/min, e.g. MB-158 (metal bellows) or N 022 STE (KNF; PTFE). All
parts of the pump with air contact must be of stainless steel or PTFE. The pressure should be checked
by using an oil free pressure gauge.
Measurement Guidelines VOC Version: Draft: 2012/07/18
A filter in the sampling system is recommended to protect the pressurizing pump and the canisters
from contamination by micro-organisms and particles. A PTFE filter (pore size 0.45 µm, diameter 25
to 50 mm, stainless steel filter holder; e.g. Pall or Millipore) shall be placed at the inlet of the
pressurizing pump. Prior to use, a new filter shall be purged with pure nitrogen (15 min with 1 L/min)
or alternatively with sample air after assembling it in the sampling line because it may release traces
of low volatility hydrocarbons (C7 to C12 and oxygenated VOC). The filter membrane shall be
exchanged depending on exposure (guideline: every 25 samples for 1 L canisters).
Canisters must be checked for leaks and blank values before air sampling. The entire sampling system
and the main inlet tubing shall be clean and tight. The material and the sampling pump used must
comply with specifications above.
An air monitoring station on different platforms (e.g. aircraft, shipboard, tower) typically use a main
inlet tubing line and a downstream pump with high flow rate (several cubic meters per hour) to pull
ambient air to a number of sampling ports. The gas samples are then usually drawn from a manifold
near to the sampling device with a smaller flow rate (0.1 l/min to 0.5 l/min).
When sampling air with canisters an ozone scrubber during sampling sometimes is used (e.g. Plass-
Duelmer et al., 2006). Hence only reactive alkene components such as 1,3-butadiene tend to oxidize
during sampling and storage. For less reactive compounds e.g. alkanes no loss in stainless steel
canisters has been observed. Surprisingly, this has also been reported for isoprene when ambient air
with approximately 100 μg/m³ ozone was sampled into 1 L electro-polished stainless steel canisters
(Leibrock, 1996; Palluau et al., 2007). This result may be explained by a rapid destruction of ozone on
the stainless steel surface of sampling line and the canister. For the GAW NMHC target compounds
the effect of ozone on sample integrity is assumed to be negligible and an ozone scrubber during air
sampling is not required.
Canister preparation prior to sampling
The set up of continues flow air sampling system for two valve canisters is schematically shown in
Figure 1. Cleaned canisters shall be prepared at the sampling location prior to air sampling. For this
purpose all canisters are first flushed with sample ambient air prior to sampling (see Section below).
Sampling procedure
Procedure for continues flow air sampling with 2 valve canisters (Figure 3):
1) Connect sampling system to a sampling port.
2) Open the canister valve (2).
3) Open canister valve (1), switch pump on and flush canister with sample air for a period
sufficient to exchange the volume of the canister ten times at the given sampling flow rate.
4) Close valve (2) and pressurize the canister with sample air up to 2.0 bar (29 psi).
Watch pressure gauge (3).
5) Open valve (2) to release the pressurized sample air.
6) Repeat Step 3 and 4 three additional times.
7) Finally, pressurize the canister (valve (2) closed) up to 2.0 bar (29 psi) and close valve (1) to
retain the air sample.
8) Pack the canister in an appropriate shipping case (see below) and ship the canister
with a copy of the sample data sheet (see Section below) to the analysing laboratory.
Measurement Guidelines VOC Version: Draft: 2012/07/18
Note 1: Eight pressurization/release cycles are required when dry air at temperatures below 0 ºC is
sampled to stabilize higher boiling compounds.
Note 2: The technically specified maximum pump pressure shall be at least 20% above the sample
pressure of 2 bar (29 psi) for avoiding an overheating of the air during sampling. Sampling/pressuri-
zing times shall be checked every 10 samples for consistency. Increasing sampling times indicate the
need for pump service.
Note 3: When sampling under high humidity air conditions water may condense during the
pressurization step which is not a problem when analysing NMHC.
Fig. 1: Flushing and repeated pressurization/release for 2-valve-canisters (valves closed/open
according to the process step).
Sampling protocol
The protocol shall contain at least the following information:
(1) Complete identification of the canister (each canister must have an engraved number).
(2) The sampling location and sampling time period.
(3) Final sampling pressure.
(4) Any unusual features noted during the sampling (e.g.: canister was not pre-pressurised).
Canister leak test
Canisters must be leak tested. First, canisters are pressurized to approximately 3 bar (44 psi) with
clean nitrogen. The canister valves are kept closed and the initial pressure is measured. After 24 h
the final pressure is checked. It is recommended that the leak rate for 1 L canisters should not exceed
0.01 mbar/h. Depressurize and condition the canister as described in Sections 5.2 or 5.5. It is
recommended to perform a leak test after every 10th sample cycle.
Transport
To prevent contamination of the inlet and outlet tubing sealing with appropriate caps is obligatory. It
is also recommended to use a special transport case to avoid damage to cylinders and shut-off
valves.
Measurement Guidelines VOC Version: Draft: 2012/07/18
Equipment and canister storage
The sampling equipment shall be stored after sampling with capped in- and outlets in normal
conditions (extreme temperatures and humidity should be avoided) away from organic solvents. The
canisters should be analyzed as soon as possible and storage time shall not exceed 30 days (US-EPA,
1999). Storage time shall be documented on the sampling protocol for each canister.
Check of storage stability of NMHCs
With increasing number of carbon atoms, the NMHCs tend to be lost on the canister surface during
transport and storage. To check for the stability of the NMHCs, the canisters are filled with an
ambient air sample. The aliquots of the sample are then analyzed immediately after sampling and
repeatedly during anticipated storage time from sampling to measurement.
Note: Ample evidence suggests that, in addition to the canister type, the NMHC stability in canisters
is strongly dependent on the humidity of the sample. A minimum relative humidity of about 12% at
room temperature is required to achieve stable NMHC concentrations (McClenny et al., 1991).
2.1.2 OVOCs (DNPH)
Off-line sampling of OVOCs by DNPH-coated samplers with subsequent liquid chromatography is a well-
established method within EMEP, therefore the chapter 3.8 (Determination of aldehydes and ketones in
ambient air) of the EMEP manual for sampling and chemical analysis (Revision nov 2001) should be used.
2.2. on-line sampling inlet line (VOCs and OVOCs)
For VOCs the inlet line should be either silco-treated steel or stainless steel. In the case of steinless
steel, the line has to be heated up to 70 °C to prevent condensation of VOCs on internal surfaces
(HOPKINS, 2011). Transfer lines for the analysis of OVOC should be either silco-steel or PFA
(Perfluoralkoxy), but not stainless steel. Silco-treated steel should be humidified before first usage
(e.g. by passing ambient air).
The line has to be as short as possible and the diameter should not be larger than 1/8 inch in order to
minimize the dead volume of the sampling line unless it is permanently flushed. The residence time
in the inlet should not exceed a few seconds. It is recommended to have a mesh with a maximum of
5 µm mesh size in the line in order to hold back particles. This has to be exchanged regularly.
Experience from an urban site (Zurich) is that the grid has to be exchanged every 4 months.
3. Preconcentration
3.1. removal of water/O 3/CO2
3.1.1 Water removal
Different systems for reducing the water content of the sample can be used. The dew point should
be measured and it should be at least 10°C lower than the trapping temperature in case of
adsorptive sampling, as otherwise adsorbed water on the trap will lead to a smaller breakthrough
volume or even a blockage of the trap. For cryogenic trapping the dew point should be below -30°C.
Systems in use are (1) a Nafion ® Dryer with a volumetric counterflow of dry air or N2, which is
around 3 times higher than the flow of humid ambient air and (2) a water trap at subambient
temperature, where water is adsorbed but not the analytes.
Examples: Rigi, Jungfraujoch (Nafion), water trap at -40°C in 1/8“ silcosteel line (Hohenpeissenberg)
Measurement Guidelines VOC Version: Draft: 2012/07/18
3.1.2 Ozone removal
For the measurement of compounds reactive to ozone, such as alkenes including isoprene and the
terpenes, the O3 in the sampled ambient air should be destroyed before analysis. Several methods
are used.
- Heated stainlesssteel line or a metal grid in the ambient air flow (>70°C)
- Addition of a flow of NO (O3 + NO�NO2) into the ambient air flow
- Small cartiridges filled with or filters impregnated with special salts like sodium thiosulfate or
potassium-iodide
A compilation of available method and their evaluation can be found in Appendix 1 (Ozone removal
techniques for GC analysis of OVOC (oxygenated volatile organic compounds) in ambient air samples
(by J. Englert)).
3.1.3 CO2 removal
If the trapping temperature is lower than -78°C it is advisable to remove CO2 either from the ambient
air flow (by using an adsorbent in-line (e.g. Ascarite) or between trapping and analysis (e.g. by slowly
heating the trap to a temperature high enough for the CO2 to be released but not for the analytes). If
CO2 is trapped and released into the analytical system this could lead to distortion of
chromatography and/or diminished sensitivity of the detector.
3.2. Trapping
As VOCs are only occurring in the atmosphere in the range of ppt (parts per trillion, 1 x 10-12) up to
some ppb (parts per billion, 1 x 10-9) VOCs from ambient air have to be preconcentrated before the
analysis, using gas chromatograph-flame ionization detection (GC-FID) or gas chromatograph-mass
spectrometry (GC-MS). On-line sampling of VOCs is performed on a trap which contains enough of a
suitable material or a combination of different adsorbents for fully retaining VOCs at a given
temperature. Either a combination of week adsorbents with low sub-ambient temperature can be
chosen or stronger adsorbent with ambient temperature, often also multi-bed adsorbents with
increasing adsorbent strength in sampling flow direction are used. For each system break-through
volume has to be tested, using either increasing amounts of humidified synthetic standards or of
ambient air spiked with standards.
A compilation of different trapping adsorbents and their usage is done in Appendix 2 (Adsorbents for
sorbent-based enrichment of VOCs and OVOCs (oxygenated volatile organic compounds) in ambient
air samples (by Jenny Englert))
For the trapping procedure a pump should be used after the trap connected to a critical orifice or a
mass flow controller (or any other suitable instrument) to regulate the flow over the trap. It is
essential to determine the sampling volume with low uncertainty either by regularly calibrated
massflowcontrollers or by pressure rise measurement in a defined reference volume. If the pump is
used before the trap it has to be ensured that no additional contamination is produced by the pump.
Examples: Perkin Elmer TD system (Empa, OHP), ADS (Empa, selfmade, Rigi),
Measurement Guidelines VOC Version: Draft: 2012/07/18
After trapping, the trap should be flushed in forward mode at the same temperature for an adequate
amount of time to allow the purge out of remaining water and potentially adsorbed gases (e.g. CO2,
noble gases) from the trap.
Release of the analytes from the trap is normally done by heating the trap (either by ohmic
resistance or by other means of heating) in counterflow. The final temperature should be reached as
fast as possible and should be high enough to release all analytes. Analytes are transferred to the gas
chromatography system by carrier gas flow. After transfer of the analytes the trap should be
reconditioned (e.g. by flushing it further with carrier gas and heating it to a higher temperature than
needed to release the analytes. The flow of cleaning gas is vented to the environment. In case that
analyte injection is not rapid enough to obtain sharp peaks which may be due to too large traps or to
slow heating rates, a second focusing trap should be installed between preconcentration and
column. This again may be adsorptive or cryogenic but needs to have a substantially smaller internal
volume than the preconcentration trap.
3.3. capillary columns for GC analysis of VOCs and OVOC
Several possible analytical columns are discussed in the Appendix 3 (chromatographic separation (by
J. Englert).
Capillary columns exhibit compared to packed species better separation efficiencies and higher
inertness. Despite their lower capacity they are suitable for most applications in environmental
research. There are basically two types of capillary columns that are currently used for the analysis of
OVOC: PLOT (Porous Layer Open Tubular) and WCOT (Wall Coated Open Tubular) columns.
PLOT columns feature a solid stationary phase consisting of a thin layer of small and porous particles
(adsorbent) adhered to the surface of the tubing. Chromatographic results are achieved by
adsorption of the analytes on the surface of the stationary phase by either surface charge
interactions or shape selectivity and size exclusion interactions. PLOT columns in contrast to weaker
retaining dimethylpolysiloxane columns are able to separate VOC at ambient and above ambient
oven temperatures which reduces liquid nitrogen consumption that is necessary in case of WCOT
columns. Special highly polar OVOCs PLOT columns do not retain saturated aliphatic NMHCs as they
have little or limited interactions with the surface of the stationary phase. By this way OVOC are
isolated and generally no co-elutions with NMHC will appear. Thus, in principle a non-specific
detector (flame ionisation detector FID) can be used as single detector.
The disadvantage of PLOT columns is the need for humidity management. Otherwise there would be
a sharp water peak that co-elutes with OVOCs, e.g. propanal and acrolein on GS-OxyPLOT (Agilent)
and for most PLOT columns a shift in retention times depending on the moisture content of the
ambient air sample. Furthermore, some PLOT columns may occasionally lose particles of the
stationary phase (problem especially for MS), but this effect was decreased by better bonding the
porous polymer layer.
WCOT columns have a liquid stationary phase. They separate the solutes with different polarities and
solubility depending on the physical properties of the stationary phase, e.g. in non-polar films the
rate of diffusion into the stationary phase and solubility according to the boiling points. The
polar/non-polar interactions are much weaker than the adsorptive interactions in PLOT columns.
There a two types of films: non polar dimethylpolysiloxane or polar polyethylene glycol.
Dimethylpolysiloxane columns are versatile, very stable and can be operated at very low
Measurement Guidelines VOC Version: Draft: 2012/07/18
temperatures. But there are co-elution problems of OVOC with NMHC and so there is the need for a
specific detector (MS). Another disadvantage is the low retention of alcohols on
dimethylpolysiloxane columns.
On the contrary on polyethylene glycol columns alcohols have high retention. Concurrently NMHC
have lower retention so that there are less co-elutions with OVOC. But a drawback is the fact that
aldehydes have also low retention. Furthermore polyethylene glycol columns have short lifetimes,
susceptible to damage upon overheating or exposure to oxygen and they cannot be operated at sub-
ambient oven temperatures.
4. Analysis
4.1. GC-FID
Currently most GC systems are operated with FID detection since FID’s have a number of advantages:
- FID’s are very sensitive, robust, simple in design and easy to use
- they perform very stable with typically less than 2% sensitivity drift over one month
- they have a response proportional to the mass or carbon number and allow easy quantification
- with the effective carbon number (ECN) concept of the response, they allow for effective QA
- they are not sensitive to traces of water, N2 and O2, and noble gases from the sample gas
FIDs need air, make-up (typically N2) and H2 gas supplies and flow rates should be well controlled to
achieve stable operation of the detector. Detector air can be generated by catalytically cleaning (Pd or Pt
catalyst at 350-450°C) ambient air, or synthetic air can be used. Essential is to have low VOC levels or at
least low fluctuation in VOC levels. Hydrogen can either be supplied by gas cylinder or by H2-generator.
The most commonly used make up gas is nitrogen from cylinders, but also other gas may be used. It is
important to check the total flows of the individual gases, including the carrier gas, and stay within the
specified margins by the FID manufacturer. GC-FID can be operated with most carrier gases, e.g. He, H2
or N2, however for reasons of good separation, hydrogen or helium should be used. Operating very
active PLOT columns with hydrogen carrier gas may cause problems in analysis of alkene or alkyne
compounds due to catalytic reactions at active surface sites. The sensitivity of an FID is generally
sufficient to do analysis in background atmosphere at ppt levels, e.g. detection limits of GC-FID systems
for analyzing 1 Liter of air are typically better than 3 pptv (e.g. Plass-Dülmer et al., 2002).
The effective carbon number concept (ECN) (Sternberg et al., 1962, Dietz et al., 1967) states that the
response (peak area) of the FID is proportional to the number of molecules times the effective number
of carbon atoms per analyte molecule, e.g. 2 ppb of ethane have the same integrated response as 1 ppb
of butane (when comparing identical sample volumes). If other than hydrogen or carbon bonds occur,
the response of the respective carbon atom is adjusted to yield an effective carbon number. For
example, an alcohol-group (-O-H) at a terminal C, like in ethanol, results in an effective response of 0.5
for this carbon or a total of 1.5. ECN are listed in original literature (see above), e.g. 1.0 for carbon in
aliphatic and aromatic bonds, 0.95 per C in olefinic bonds, 1.3 in acetylenic, 0 for carbonyl, and 0.5 for
primary, 0.25 for secondary and 0.75 for tertiary alcohol (Sternberg et al., 1962). Using the ECN-concept,
reliable calibration factors for compounds not present in the calibration gas mixture can be estimated.
A GC-FID system can be easily characterized for losses or artifacts by making use of the known carbon
response (=peak area/(Volume sample*mixing ratio*carbon number)). When the carbon responses for
the various organic compounds are calculated, they should agree within a few %. Deviations are often
due to bad peak separation, adsorptive losses in the system, or artificial changes at active sites. Efforts
Measurement Guidelines VOC Version: Draft: 2012/07/18
should be taken to optimize the system. Especially standard addition measurements (few percent of
high concentrated standard in ambient air sample) are of high value as they characterize interferences
with other constituents of ambient air like water vapor and ozone.
Disadvantage of GC-FID systems, however, is that the FID is not substance-specific. Thus it requires a
very good chromatographic separation in order to minimize problems due to co-eluting peaks.
Accordingly, GC FID is a suitable system whenever unambiguous identification can be achieved simply
based on the retention times, otherwise combination with MS is recommended.
4.2 GC-MS
The analysis of VOCs with GCMS is possible, although not ideal for isomers of smaller compounds
such as pentenes or isohexanes and isoheptanes with similar boiling point/retention times on
analytical columns. As the sensitivity of the mass spectrometer decreases relatively rapidly,
calibrations using a working standard have to be performed nearly as often as ambient air
measurements. At Jungfraujoch for example the following system is in use: std-air-air-std-air-air-std-
air-air-std-
This ensures that the decline in sensitivity is accurately tracked. Furthermore, target tank
measurements should be performed bi-weekly and blank measurements should be performed
monthly. After a considerable change of the system (e.g. source replacement, changing of analytical
column or trap) blank measurements have to be performed as soon as the system runs in a steady
mode again.
The source has to be tuned after the loss of sensitivity has reached a certain degree, but at least
every second month. If boundary conditions of the source (repeller voltage, lenses) do not allow a
proper tuning of the source anymore it has to be cleaned using the procedure specified by the
manufacturer.
4.3 PTR-MS (provided by R. Holzinger Uni Utrecht an d T. Petäjä)
Inlet (recommendation):
~4mm ID PFA tubing, max 15m, flow 2 L/min, protected by PFA filter holder with 47mm PTFE filter 5
μm pore size
PTR-MS samples from this inlet line with a short low volume (1/8“or 1/16“) line. Recommended
materials: PFA, PEEK
Background determination:
The 2L main sample flow is passed through a Pt catalyst at a temperature of 350 C.
Critical instrument settings (recommendation):
Drift tube pressure: 2.2 mbar
Drift tube voltage: 600V
Drift tube temperature: 50C
Voltage between last drift ring and exit lense: 30V (instrument dependent, maybe define a procedure
to optimize this)
Measurement Guidelines VOC Version: Draft: 2012/07/18
Ratio O2+/H3O+ below 0.03
PTR-TOF-MS:
�SV valve setting (describe optimazation procedure)
� Peak shape standards
� MCP voltage
Q-PTR-MS
�Tuning of mass scale and resolution
� SEM voltage
Working standard (recommendation)
Methanol, m33; Acetonitrile, m42; Acetaldehyde, m45; Acetone, m59; MBO, m87, m69, m41; MVK,
m71; MEK, m73; Benzene, m79; Toluene, m93; Xylene, m107, TMB, m121; a-pinene, m137, m81,
Trifluorobenzene, m133; Trichlorobenzene, m181, m183, m185
Intercalibration standard (recommendation): same compounds
Prepared and distributed by EMPA twice per year. (I think it is not necessary to ship pressurized
cylinders ~1-2 liters of gas standard should be sufficient)
In field operation:
When and how often:
� background measurements
� working standard
� Full mass scan
Calculation of VMR
� recommended rate constants
� tranmission
� recommended procedure
5. Quality Assurance and Quality Control
Quality assurance (QA) follows the principles of the GAW QA system (see Figure 2 and
http://www.wmo.int/pages/prog/arep/gaw/qassurance.html):
1. Network-wide use of only one reference standard or scale (primary standard). In
consequence, there is only one institution that is responsible for this standard (CCL).
2. Full traceability to the primary standard of all measurements made by Global, Regional and
Contributing GAW stations.
3. The definition of data quality objectives (DQOs).
4. Establishment of guidelines on how to meet these quality targets, i.e., harmonized
measurement techniques based on Measurement Guidelines (MGs) and Standard Operating
Procedures (SOPs).
5. Establishment of MGs or SOPs for these measurements.
6. Use of detailed log books for each parameter containing comprehensive meta information
related to the measurements, maintenance, and 'internal' calibrations.
Measurement Guidelines VOC Version: Draft: 2012/07/18
7. Regular independent assessments (system and performance audits, Performance audit:
check measurements versus DQOs and traceability System audit: overall conformity of a
station with the principles of GAW).
8. Timely submission of data and associated metadata to the responsible World Data Centre as
a means of permitting independent review of data by a wider community.
Fig. 2: Principles of the GAW QA system.
5.1 Standards and Scale
The nomenclature of standards should be used as follows: The “primary standard”, “secondary
standard” etc. should only be used at the Central Calibration Laboratory (CCL), they comprise a system of
standards defining the “scale” for VOC in GAW.
Generally, “tertiary standards” are provided by the CCL to the stations and laboratories and they are
used there as “laboratory standards” or “working standards”. The highest level standard at a given site
is called the “laboratory standard” (the laboratory standard is not necessarily a tertiary standard from
the CCL, though it is recommended). The laboratory standard is supplemented by a system of working
standards at the stations that may consist of synthetic or whole air mixtures, certified or custom-made.
“Traveling standards” are generally working standards with reference mole fractions determined by the
CCL, WCC or a certified laboratory used in comparisons and round robins.
The term “target gas” is used for a working standard which is treated as a sample of unknown
compositions, e.g. the data evaluation procedures of a station are used to determine the VOC mixing
ratios and the deviation from the assigned values is a test for the quality of measurements. Accordingly,
a target gas should be close to ambient gas composition which means a whole air standard with similar
concentration levels.
The scale is kept by the CCL by means of a system of standards (see above). The scale is transferred to
the stations and labs by the tertiary standards. In case a station does not hold a tertiary standard by the
Measurement Guidelines VOC Version: Draft: 2012/07/18
CCL, it has to demonstrate that the laboratory standard is linked to the scale by regular and direct
comparison.
It is recommended that each station or laboratory holds the following Calibration Gases:
1. A laboratory standard which should be a multi-component laboratory standard (synthetic mixture)
that covers the main components and should presumably be produced by the CCL (NPL) or another
(NMI) linked to the CCL.
2. A certified multi-component working standard (synthetic mixture with certified mixing ratios)
with similar components as the laboratory standard.
3. Multi-component working standards that cover all components measured and which are
calibrated versus laboratory standard, travelling standards, or other methods (Carbon response FID,
permeation/diffusion source, mixtures). One of these working standards should be of high mole
fractions (upper nmol/mol range) for standard addition measurements.
4. A Target Gas which is a whole-air working standard calibrated versus laboratory standard, other
standards or by other means
Though not recommended, there are minimum requirements for a station that need to be fulfilled:
1. A laboratory standard to define the scale for each component measured at the station
2. One working standard (which may be custom-made) to check for drifts in the scale. In case of
a GC-FID where calibration can in many cases be reasonably transferred from the laboratory
standard to other compounds not present in the laboratory standard by means of the carbon
response concept, a well documented procedure to assign calibration factors and
uncertainties to these compounds are needed.
3. A target gas which is presumably whole air but could also be a synthetic mixture
5.2 Zero Gas
“Zero gas” is a hydrocarbon free gas from either catalytically cleaned ambient air (Pt or Pd catalyst at
400°C) or alternatively, not as good but easier to handle, synthetic gas of at least 5.0 quality.
Catalytically cleaned air is preferred as this is identical to the sample gas matrix. Often, trace
amounts of hydrocarbons in the pmol/mol range are present as impurities in the zero gas. Stations
should go for optimum zero gases by comparing the blank values obtained in measurements of
different hydrocarbon free gases aiming at the lowest levels.
The routine measurement of zero gas is part of the QA program to be followed at all stations. It
yields information about artifacts due to release of adsorbed hydrocarbons or leaks in the sample
path. Blank values should be as low as possible. However, there is an inherent problem when using
zero gas with trace level impurities of hydrocarbons: these can not easily be separated from blank
artifacts as mentioned before. Accordingly, care has to taken to identify the origin of blanks found in
zero gas measurements.
Zero gas should be applied at the inlet of the sampling system (on-line syst.) or to flasks and canisters
via any used sampling line under field use conditions. Often a high flow inlet manifold is used which
cannot be easily flushed by zero gas, then zero gas can be introduced after the sampling split to the
instrument, but checks with independent sampling not over the routinely used inlet should be
performed.
5.3 Data Quality Objectives
Measurement Guidelines VOC Version: Draft: 2012/07/18
The following data quality objectives have been approved by the GAW-VOC expert group:
The enhanced DQOs are the objective for good performing stations. Within ACTRIS, generally the
GAW DQO shall be reached. These stated DQOs are valid for individual ambient air measurements.
This is different from the interlaboratory comparability objectives used in the greenhouse gas
community where the objectives refer to uncertainties in measurements of standards comprising
multiple measurements.
5.4 Method for Measurement of Standards
Calibration using a working standard is required every 2 months using a GC-FID. If possible and for
securing highest possible data coverage it is suggested to perform the calibration every 2 weeks. As
mass spectrometers drift considerably, it is recommended to perform at least one working standard
measurements per day. For achieving optimal performance a more frequent application of the
working standard is recommended. Furthermore, a target tank has to be run once a week. This tank
can either be a certified standard mixture in air or N2 or a compressed ambient air within a cylinder,
with known concentrations.
Generally it is recommended to leave pressure regulators and transfer lines attached to the working
standard/target gas cylinders in order to minimize the risk of contamination and reduce equilibration
times. Laboratory gloves (i.e. powder-free latex) should be worn whenever working with parts in
contact with test gases in order to avoid contamination. Furthermore, it is strongly recommended to
use no liquid leak tester solutions as they might contaminate the system.
If a working standard/target gas cyclinder is newly connected, at least 24 hours before the
measurement the pressure regulator and the transfer line with capped fitting on the GC connection
side should be mounted. The transfer line should be made of Silcosteel or Sulfinert or other stainless
steel tubing with a passivated internal surface. The use of Vespel/Graphite (VG) ferrules is
recommended as these provide a tight sealing while not damaging the tubing. They can be used
several times and should only be replaced in case that sealing or contamination problems are present
GAWUncert.
%
GAWRepeat.
%
Enhanced*Uncert.
%
Enhanced*Repeat.
%
alkanes 10 5 5 2
Alkenes incl. IsopreneMonoterpenes 20 15
510
25
Alkynes 15 5 5 2
Aromatics 15 10 5 2
OVOC 20 15 10 5
Acetonitrile and DMS 20 15 10 5
Mole fraction < 0.1ppb +/-20 ppt +/- 15 ppt +/- 10 ppt +/- 5 ppt
GAWUncert.
%
GAWRepeat.
%
Enhanced*Uncert.
%
Enhanced*Repeat.
%
alkanes 10 5 5 2
Alkenes incl. IsopreneMonoterpenes 20 15
510
25
Alkynes 15 5 5 2
Aromatics 15 10 5 2
OVOC 20 15 10 5
Acetonitrile and DMS 20 15 10 5
Mole fraction < 0.1ppb +/-20 ppt +/- 15 ppt +/- 10 ppt +/- 5 ppt
Measurement Guidelines VOC Version: Draft: 2012/07/18
(follow the mounting instructions of the manufacturer). After installation, the regulator needs to be
flushed. For this, the transfer line is uncapped, the regulator’s low pressure port is gently opened and
the cylinder valve is opened very short such that the primary pressure jumps up and goes down again
to ambient. This procedure should be repeated at least 2 times. After flushing, plug the transfer line
and pressurize the pressure regulator another time by shortly opening the cylinder valve. Check the
pressure for a few minutes; if not constant, tighten the 30 mm nut and repeat until pressure is
constant. Then, apply a low pressure (0.2-0.5 bar) to the transfer line, as well. As an initial leak check,
close the regulator. After it is closed, watch the low pressure port for some 10 min. If not constant,
check plug and connection to the transfer line, tighten gently, and repeat the check. Finally, the
pressure regulator and the transfer line (plugged at the end) should be pressurized with the standard
gas for at least 24 hours. During this equilibration time, the cylinder valve is closed to avoid back
diffusion of potential contaminants into the cylinder and to avoid loosing sample through possible
leakages. This setup also serves as a static leak test as the upstream regulator pressure should not
change during the 24 h equilibration period. Connect the test gas cylinder to an appropriate
instrument inlet port. Then flush the whole inlet line for at least another 3 times and leave the gas
cylinder connected to your instrument. It is recommended to open the standard cylinder valve only
during the sampling periods unless you use an automated measurements sequence in unattended
operation.
In many set-ups the standard cylinder is permanently connected to the GC system. The measurement
of the standard gas should be performed after an initial flushing period through the GC valve system
which is sufficiently long to achieve equilibration in the lines (typically 10 min with 30 ml/min). To
stay within the DQO, the sensitivity of a GC system should not drift by more than 3% between
calibrations. Similarly, blank values (see below) and reproducibility should not change such that they
introduce more than 3 % effects on the measured data. As both, calibration and target gas
measurements enable to detect drifts in the system it is up to the operators to decide the share of
these measurements. Another issue is the reproducibility of such standard measurements. Often,
first measurements are off in a set of measurements due to insufficient equilibration of internal
surfaces of sampling lines. For such conditions, series of standard measurements are to be
performed containing at least one appropriate measurement.
In case stations use working standards/target gases not comprising all components measured, it is
justified to determine the sensitivity drift of the instrument by this reduced compound mix if it
comprises major constituents of the various groups of VOC and it covers the range of volatility and
polarity encountered in the samples. Calibration factors of compounds not present in the working
standard may then be scaled by calibration factors of physically similar behaving compounds present
in the standard.
If a drift in the laboratory standard is observed as determined by a drift / inconsistency with a
working standard or a discrepancy with a new laboratory standard beyond the combined
uncertainties, the discrepancy has to be resolved as soon as possible. Options in such a situation are
to send the laboratory standard for recalibration to the CCL or WCC, ask other stations for a high
level standard for an independent check, or check available results from past intercomparisons.
Anyways, station operators should try to identify where the drift occurred and apply a correction for
those periods in which the drift can be well described. If this is not possible, the uncertainty during
this period needs to cover the range of unexplained drift.
Measurement Guidelines VOC Version: Draft: 2012/07/18
In Figure 3 an example of a series of working standards measured at Hohenpeissenberg (DWD,
Germany) is shown. Here, the determined mixing ratios for all analysed compounds are plotted over
time in a log scale. Relative changes should become detectable by non-constant values. The first plot
shows compounds with mixing ratios of more than 50 ppt, the next C3-C6 hydrocarbons with mixing
ratios below 50 ppt, and finally the plot of C6-C9 hydrocarbons with mixing ratios below 50 ppt.
Except for some severe i-butene problems (associated with blank values), it should be pointed out
that repeatability gets worse for compounds with higher molecular weight and towards lower mixing
ratios, but still repeatability is mostly within some 2 ppt or a few percent. This plot demonstrates
repeatability within each series of 5 replicates (monthly), and reproducibility throughout the year for
ambient mixing ratio levels and ambient air matrix.
10
100
1000
10000
01/09 02/09 03/09 04/09 05/09 06/09 07/09 08/09 09/09 10/09 11/09 12/09
Ethan
Ethen
Propan
Propen
i-Butan
Acetylen
n-Butan
n-Pentan
3-M -Pentan
i-Buten
i-Pentan
2-M -Pentan
M-C-pentan+2,2-DiM -B.
Toluol
Benzol
m-Xylol
Referenzluft-04, ppt
1
10
100
01/09 02/09 03/09 04/09 05/09 06/09 07/09 08/09 09/09 10/09 11/09 12/09
t-2-Buten
c-2-Buten
Cyclopentan
Propin
1,3-Butadien
1-Penten
Cyclohexan+?
2,3-DiM -Butan
1-Buten
Isopren
Referenzluft-04, ppt
1
10
100
01/09 02/09 03/09 04/09 05/09 06/09 07/09 08/09 09/09 10/09 11/09 12/09
n-Hexan
Isopren
1-Hexen
M -C-Hexan+2,2,3-TriM -B.2,3-DiM -Pentan
2+3-M -Hexan
n-Heptan
3-M -Heptan
n-Oktan
n-Nonan
Ethylbenzol
o-Xylo l
Fig. 3: Measurements of compressed whole air from cylinder (“Referenzluft-04” or “Ref-04”) through
2009, generally 5 replicates are measured once per month.
5.5. Method for Measurement of zero gas (blanks)
Measurement Guidelines VOC Version: Draft: 2012/07/18
For blank measurements, a zero gas is sampled via the usual air sample path. Thus, it passes the
ozone and particle filter, the water trap and sampling unit just like ambient air samples. At
Hohenpeissenberg, Helium is applied at an open “T” into the 1/8” sampling line at a flow rate of
about 100 ml/min yielding a sample flow of about 80 ml/min towards the GC and 20 ml/min towards
the ambient air sample manifold. A similar amount of zero gas is analysed as in the ambient air
samples.
For convenience, we use purified He 5.0 (charcoal and mole sieve at -40°C) as zero gas as we use the
same gas as carrier gas and purge gas. Occasionally, we also test the same line using catalyst-purified
ambient air (Pd at 400°C) and essentially observe the same results as with He.
The Figure 4 gives an example of the behaviour of blanks over time. Peaks in benzene are
occasionally observed and are generally associated to some kind of overheating of traps but the
nature of this contamination is not really understood. Additionally, it can not be excluded that small
amounts of ambient air diffuse into the He against the He flow (at the open “T”) and some of the
ethane, ethylene, propane and benzene is attributed to ambient air mixing into the He.
0
5
10
15
20
25
30
01 02 03 04 05 06 07 08 09 10 11 12
year 2009
He-
blan
k, p
pt
Ethane
Ethene
Propane
Propene
Acetylene
1-Butene
Benzene
i-Butene
Fig. 4: Results of blank peak areas from He blanks transferred to mixing ratios by applying the same
calibration factors as for other samples.
5.6. Method for Detecting Effects of Ozone on React ive Compounds
In order to check for interferences with ozone and other reactive constituent of the ambient air
sample gas, it is suggested to perform standard addition measurements 4 times a year. In these, 1-5
ml/min of NPL standard are added via a fused silica capillary into the ambient air sample flow of 50
ml/min. Thus, roughly 2-10% of the whole gas sample is NPL standard, however, in the ambient air
matrix including ozone. It is not so important to know the flows exactly, as the mixing ratios of
Measurement Guidelines VOC Version: Draft: 2012/07/18
hydrocarbons with low concentrations in ambient air and no reactivity towards ozone (e.g. propyne,
cyclohexane and n-heptane) can be used from the NPL standard to scale the mixing. Then the ratios
of the mixing ratios in standard addition measurement compared to a pure NPL standard are plotted
after normalizing them to the corresponding ratios of the above mentioned compounds (Fig. 4).
Some of them have fairly high contributions from ambient air (ratios up to about 3), but the
important alkenes are clearly dominated by the added standard. If ozone interferences (losses) exist,
these reactive alkenes should show lower ratios than 1. None of the alkenes shows in the log (upper)
and linear plot (Fig. 5) any significant deviations from 1 and thus no indication of reactive losses.
NPL-Standard-Addition
0.1
1.0
10.0
10/08 10/08 12/08 12/08 01/09 03/09 04/09 05/09 06/09 07/09 08/09 09/09 10/09 11/09 12/09
Ethan Ethen Propan Propen i-Butan Acetylen
n-Butan t-2-Buten 1-Buten i-Buten c-2-Buten Propin
1,3-Butadien t-2-Penten c-2-Penten Cyclohexan+? n-Hexan Isopren
n-Heptan Benzol Toluol Ethylbenzol p,m-Xylol o-Xylol
NPL-Standard-Addition
0.5
1.0
1.5
2.0
10/08 10/08 12/08 12/08 01/09 03/09 04/09 05/09 06/09 07/09 08/09 09/09 10/09 11/09 12/09
Ethan Ethen Propan Propen i-Butan Acetylen
n-Butan t-2-Buten 1-Buten i-Buten c-2-Buten Propin
1,3-Butadien t-2-Penten c-2-Penten Cyclohexan+? n-Hexan Isopren
n-Heptan Benzol Toluol Ethylbenzol p,m-Xylol o-Xylol
Fig. 5: Results of NPL standard addition measurements from 2009 performed once per month.
Typically, 2-10% of sample volume is added to an ambient air sample by continuous addition of a
small flow of NPL standard via a quartz capillary to the ambient air.
5.7 Logs at each station
It is required that each station has the following logs either in electronic or paper-based form:
Measurement Guidelines VOC Version: Draft: 2012/07/18
1. Instrument Log with all operation parameters, significant changes, characterizations, tests
results, etc.
2. Measurement Log with all measurements including the type of measurement, the time of
measurement in UTC (start sampling, end sampling, start GC run), sampled volume (dry
volume), and comments (anything unusual)
3. A Log of the used calibration factors and blank value determinations from zero gas
measurements
4. A Log of all working standard and target gas measurements
5. An Error Log with ascribed uncertainty contributions to compound measurements due to
peak-overlap, scatter of blank values, unusual low reproducibility, instable sensitivity a.s.o. as
well as all other unexplained deviations from normal instrument performance.
6. Meteorological data Log (temp, hum., wind velocity and dir.)
6. Post-analysis
6.1 Data checks of final mixing ratio data
VOCs should be grouped in a convenient number (typically 3 or 4) of functionally similar compounds,
e.g. alkanes or alkenes, in a log plot over a time interval of half a year or a year. The procedure is
illustrated in the Figures 6-, where the quality checks were performed for the Hohenpeissenberg data
in year 2009 and Rigi 2011 data. When looking at the plots please consider that at Hohenpeissenberg
usually measure at 1:00 and 13:00 CET which explains part of the variability in the data as diurnal
variations. Furthermore, Hohenpeissenberg is on top of a hill, thus we are frequently at night
decoupled from the boundary layer and do not observe accumulations of trace gases as observed at
flat terrain sites.
Generally, it is expected that the variability of the data should increase with higher reactivity
(variability-lifetime-relation) and changes should be more pronounced for shorter lived compounds
(lower background). Spikes in positive direction may be attributed to plumes with local/regional
pollution and should be checked for consistency with other compounds from similar sources, if not
consistent, the raw data should be rechecked. Spikes in negative direction stand for clean air and
should again be checked for consistency with other compounds from similar sources. If they are
found to be not consistent with other compounds, we check the peak integration, break through in
trap or other potential loss problems (e.g. the negative spike in ethane at the beginning of august
(Fig. 6) was due to break through).
Measurement Guidelines VOC Version: Draft: 2012/07/18
10
100
1000
10000
01.01. 31.01. 03.03. 02.04. 03.05. 02.06. 03.07. 02.08. 02.09. 02.10. 02.11. 02.12.
Ethan, ppt Propan, ppt i-Butan, ppt n-Butan, ppt
Fig. 6: Time series (annual cycle) of C2-C4 alkanes measured at Hohenpeissenberg, generally,
measurements from 1:00 and 13:00 CET are shown.
For compounds with similar relative annual cycles but on different mixing ratio levels the ratio plots
as shown below can be used (Fig. 7).
In such plots we either compare structural similar compounds (as shown above for pairs of alkanes,
alkynes, and aromatics) or compounds originating from similar sources or compounds having similar
lifetimes. We check all the spikes with the log book to identify irregular operation conditions, recheck
peak integration, check with other compounds deviating in these individual measurements and try to
identify the reason for the spike. For example, in mid April the propyne/acetylene spike can be
attributed to small wood fires in the vicinity of the station,), the n-pentane/i-pentane spikes are due
to occasional n-pentane plumes (we do not know where they originate from but we observe them in
two independent GC-systems), the broad minimum in propyne/acetylene ratios in spring is due to
the longer lifetime of acetylene and the correspondingly relaxed reaction to the OH annual cycle, and
of the aromatics only benzene has a well developed summer minimum, all others are flattened in
summer but consistently as seen in the xylene/toluene ratios.
Measurement Guidelines VOC Version: Draft: 2012/07/18
0.01
0.1
1
10
01.01. 31.01. 03.03. 02.04. 03.05. 02.06. 03.07. 02.08. 02.09. 02.10. 02.11. 02.12.
n-Pentan/i-Pentan Propyn/Acetylen Toluol/Benzol p,m-Xylol/Toluol
Fig. 7: Time series of ratios between pairs of hydrocarbons with similar structure
xy-plots for the data evalualtion (used at Rigi, Switzerland by Empa)
A similar approach as used at Hohenpeissenberg is applied at the Rigi site by Empa. In addition many
xy-plots are produced to check for consistency with former years and within the year. In Figures 8
and 9 examples are shown of correlation plots using toluene vs. benzene and benzene vs. ethyne.
xy-plots used at the Rigi (Switzerland) site by Empa:
propane/ethane butenes/ethene soheptane/isohexane
n-butane/ethane pentenes/ethene 1,3-butadiene/isoprene
propane/n-butane entenes/butenes toluene/benzene
2-methylpropane/n-butane methylpropane/2-methylbutane m/p-xylene/benzene
2-methylbutane/n-pentane 1,3-butadiene/ethene Ethylbenzene/mp-xylene
propene/ethene n-hexane/n-pentane o-xylene/mp-xylene
acetylene/ethene isohexane/n-hexane o-xylene/ethylbenzene
acetylene/benzene
Measurement Guidelines VOC Version: Draft: 2012/07/18
Fig: 8: toluene vs benzene at Rigi (Switzerland). Blue 2011 data, brown 2008-2011 data
Fig: 9: acetylene vs benzene at Rigi (Switzerland). Blue 2011 data, brown 2008-2011 data
repeatability and reproducibility:
One more check which is aplied at Hohenpeissenberg for the plausibility of the measurements is
related to the repeatability and reproducibility of the measurements. We use the reference gas
measurements of a complete year and determine the standard deviations for all compounds of the
whole air reference (“Ref-04”), a synthetic mixture prepared from liquid PIANO standard material
mixtures from Supelco mixed with N2 in a high pressure cylinder (“SUPELCO”), and the Apel-Riemer
standard (“74-NMHC”).which we use to check our laboratory standard from NPL. The relative (%) and
absolute (ppt) standard deviations are plotted on a log scale versus the mixing ratio for all identified
Measurement Guidelines VOC Version: Draft: 2012/07/18
compounds. In the lower figure there are added two curves that represent a variability by 1ppt + 1%
(red line) and 3ppt + 3 % (broken red line). For a number of compounds the first line represents a
good fit. However, some of the compounds exhibit higher scatter but are generally within the 3ppt +
3% line which still is a quite good level for the reproducibility of our measurements of cylinder gases.
Compounds with higher deviations should be checked for the reason of the deviation, often peak-
overlap or –integration problems are associated with worse reproducibility. Also, heavier compounds
tend to be less reproducible due to adsorption/desorption problems.
Such a presentation helps to identify problems with individual compounds. At Hohenpeissenberg, we
calculate the precision of the measurements by error propagation based on blank value scatter, peak
integration problems and other characterised problems. This precision estimate should then be
comparable or larger (conservative estimate) than the determined reproducibility (Fig. 10) to be on
the safe side with the precision estimate.
reference gases in 2007rel. standard dev. versus mixing ratio
0%
1%
10%
100%
1000%
0.1 1 10 100 1000 10000 100000
Ref-04
SUPELCO
74-NMHC
abs. standard dev. versus mixing ratio
0
1
10
100
1000
10000
0.1 1 10 100 1000 10000 100000
Ref-04
1ppt+1%
SUPELCO
74-NMHC
3ppt+3%
Fig. 10: Standard deviations obtained from all measurements of standard (not NPL as this is the
laboratory standard used for quantification of the other standards and reference gases) and
reference gases in 2007 versus compounds’ mixing ratios in ppt in respective standard; the upper
panel shows the relative standard deviations, and the lower panel the ppt standard deviations versus
mixing ratios (ppt); two lines in the lower part show parameterizations of the standard deviations for
the “good” results (1 ppt + 1%) and as an upper limit for most of the “not so good” compounds 3 ppt
+ 3%.
Measurement Guidelines VOC Version: Draft: 2012/07/18
6.2 Uncertainty evaluation
This section describes the routine assessment of measurement precision and uncertainty. While the
precision reflects random errors in the measurement process, the uncertainty includes also possible
systematic errors in the measurement. In the following it is illustrated which factors influence precision
and uncertainty and how they are derived following the concept of the “Guide to the Expression of
Uncertainty in Measurement” (GUM, 2008).
Derived uncertainty values are 1 errors, for the expanded uncertainty the values have to be multiplied
by the coverage factor k=2 (representing the 2σ error). The error calculation is based on the calculation
of mixing ratios for linear detection systems as presented in section 6.2.1.
For data submission the precision (1σ) will be reported as well as the total expanded uncertainty (2σ)
6.2.1. Calculation of mixing ratios for linear detection systems
For substances quantifiable via a standard gas mixture, the mixing ratio χsample,i of a compound “i” in a
sample is calculated via:
icalsample
iblankisampleisample f
V
AA,
,,, *
−=χ (F1)
With the calibration factor
irespinumiblankical
icalcalical CCAA
Vf
,,,
,, *
1*=
−=
χ (F2)
Asample,i= peak area of sample measurement of compound “i”
Acal,i = peak area of calibration gas measurement of compound “i”
Ablank,i=possible blank value of compound “i” determined in zero gas measurements
cal,i = certified mixing ratio of calibration gas standard
Vcal = sample volume of calibration gas
Vsample = sample volume of sample
Cinum = Number of C atoms in the molecule “i” (e.g. for i = n-Pentane, Cnum = 5)
Cresp,i = mean C-reponse factor of compound “i”
In case of substances not quantifiable by the standard gas mixture, the mixing ratio is calculated via the
mean C-Response factor respC , which is determined from selected compounds in the standard gas
measurements averaging the respC factors for those substances.
The mixing ratio of a substance is then
respinumisample
isampleisample
CCV
A
**,
,, =χ (F3)
Measurement Guidelines VOC Version: Draft: 2012/07/18
6.2.2 Determination of Precision
The precision can either be derived from the target gas or working standard (whole air) measurements
or series of air samples taken in similar, stable air mass conditions (meteorological and chemical) in
series of 5 or more measurements.
It covers the random errors of peak integration, volume determination and blank variation. In case of
canister or adsorption tube sampling it includes the reproducibility of the sampling system as well.
The Precision precδχ is determined as the standard deviation of a series of measurements of a sample
sampleσχ :
sampleprec σχχ =∆
The above given value represents the instruments precision only at the concentration level and
complexity of the sample gas. When working with highly variable concentrations, hence variable peak
areas, a more general description of the precision has to be applied for routine measurements:
samplerel
prec DL χσχχ *31 +=∆
Where
DL = detection limit of the system (as described below) and
C = mixing ratio of considered peak
samplerel χσ = relative standard deviation of the sample (e.g. working standard)
Thus, the DL is the dominant factor for rather small peaks, while the reproducibility of the instrument
becomes the important term for larger peaks.
6.2.3 Determination of Uncertainty
The total uncertainty uncχ∆ of a measurement does not only include the random errors described by
the precision but also the systematic errors systematicχ∆ of the measurement.
222systematicprecunc χχχ ∆+∆=∆
Possible systematic errors are:
• the uncertainty in the standard gas mixing ratio calδχ
• systematic integration errors (due to peak overlay or bad peak separation) intAδ
Measurement Guidelines VOC Version: Draft: 2012/07/18
• systematic errors in sample volume determination Vδ
• systematic errors in blank determination blankδχ
• Further instrumental problems (e.g. sampling line artefacts, possible non-linearity of the
Detector (MS), changes of split flow rates) instrumentδχ
Following Gaussian error propagation, the overall systematic error is then described as 2222
int22
instrumentblankvolcalsystematic χχχχχχ ∆+∆+∆+∆+∆=∆
Referring to equations F1 to F3, the single error contributions are determined for each analysed
compound as described below (assuming blanksamplesample AAA −≅ ):
Systematic error due to the uncertainty of the calibration gas:
calcalsample
calsamplecal AV
VAδχχ *
*
*=∆ ,
where calδχ includes the certified relative uncertainty of the standard gas (or the working standard) and
possible drifts of the standard.
Systemativ integration error:
2
int,2
2
int,2int *
*
***
+
=∆ cal
calsample
calcalsamplesample
sample
cal AAV
VAA
V
f δχ
δχ ,
where calAint,δ represents the relative error in peak area due to integration of the calibration
measurement and sampleAint,δ the integration error of the sample measurement, respectively.
If the sample gas and the calibration gas chromatogram resemble each other (similar concentration
level, same peak shapes, etc.), the systematic integration error cancels.
How to determine the systematic integration errors:
i) Conservative estimate
A statistic integration error (random changes in baseline position) is covered by the precision of the
measurement.
However, for peaks strongly deviating from the theoretically expected Gaussian peak form (e.g. with a
strong tailing) or overlapping peaks a systematic integration error has to be determined.
For peaks with a pronounced tailing, some systematic error has to be expected as the slowly declining
slope generates some potential to erroneous peak integration:
Measurement Guidelines VOC Version: Draft: 2012/07/18
(a) the tailing – thus the complete amount of the eluting substance – might not be completely captured
by the routine integration parameters or (b) some of the observed tailing is some baseline artefact
independent from the observed substance. (c) else.
Of course for overlapping (non-Gaussian) peaks the not visible peak end or start generates some
potential problem.
Very conservative estimates of systematic integration errors can be derived by integrating peaks such
that an extreme minimum and maximum peak area is determined as depicted in Fig. 1 and 2. After
determination of a maximum and a minimum peak area, the standard deviation intσ , of the derived
peak areas (min,max, norm) is calculated. 3intσ
conservatively describes the possible systematic error of
peak integration.
Tailing peaks:
For tailing peaks an estimate the minimum and maximum peak area can be derived by applying the
baselines as shown in Fig 1. While the blue baseline cuts off the tailing, the red baseline captures the
visible tailing.
MIN
MAX
Figure 11 Scheme for tailing peaks. The black baseline represents the best fit. Red and blue lines illustrate the baselines to derive the maximum and minimum peak area, repectively.
Overlaying Peaks:
Measurement Guidelines VOC Version: Draft: 2012/07/18
Usually, for overlapping peaks the drop baseline integration (black straight and dotted baselines) is the
method of choice. For the theoretical Gaussian peak this procedure will exactly describe the expected
peak area. For non-Gaussian peaks it is a reasonable best estimate. A minimum for peak A can (very,
very) conservatively be derived by applying the green dashed baseline; for peak B the red dashed
baseline, respectively. Maximum areas are derived by integration of the total area of peaks A and B (via
black straight baseline) and subtraction of the tangent peaks; e.g. for peak A the maximum area is the
total area minus the minimum peak area of peak B and vice versa for peak B.
A B
Figure 12 Scheme for not separated peaks. The baseline in black (straight line) yields the area of peak A and B. Applying a baseline drop in the peak valley (black dotted line) will separate the peak areas and yield the best estimate for the overlayed peaks. Dotted baselines in in green and red can be used to derive the maximum and minimum expected peak areas.
ii) “educated estimate”; will be specified later.
For groups already applying a more educated and specific procedure to determine the systematic
integration error, it is advisable to continue to use your own procedure (and provide a description).
Systematic volume errors:
volχ∆ , the systematic error of the sample volume, can be neglected, since calsample VV δδ = , and thus the
error cancels in equation F1. The statistic volume error is covered by the measurement precision.
Measurement Guidelines VOC Version: Draft: 2012/07/18
Systematic error of blank value:
If a blank correction has to be applied, the error of this correction can be described as the deviation from
the mean blank value of the component:
1
1*
−=∆
nblankblank σχ
With the standard deviation σblank calculated from n zero gas measurements.
Further systematic errors:
instrumentχ∆ , the error in mixing ratio due to specific instrumental problems has to be evaluated for each
site individually. These errors can be derived by tests or intercomparison measurements. One possible
way to describe instrumental artefacts is illustrated in the Example below.
6.2.4 Determination of detection limit
Due to impurities or analytical problems or limits the baseline of your gas chromatographic system is
usually to a certain degree noisy. Thus, the lowest quantifiable quantity of a substance - the
detection limit of the measurement system – is different from zero.
Figure 13 Integration of noise.
Measurement Guidelines VOC Version: Draft: 2012/07/18
A simple way to calculate the detection limit is, to integrate a baseline signal over a time interval
similar to the average peak width. This integration is performed for a statistically significant number
of times (min 10 times). The derived standard deviation of the integrated area multiplied by a factor
of 3 represents the detection limit.
EXAMPLE FOR ROUND ROBIN MEASUREMENTS:
Benzene measured in Sample: VOC_in_N2 and NPL (Volsample = VolNPL = 401ml, certified mixing ratio 2ppb
± 2%; blank corrected Areas: Acal = 8263.6667; Asample = 4413.6) with the FID.
Figure 14 Benzene peak in the NPL standard (certified mixing ratio 2 ppb) in red and the VOC_in_N2 sample gas in black; both at a sample volume of 401ml.
Precision
Systematic
Error
Uncertainty
MR [ppb] 1 [ppb] [%] 1σ[ppb] [%] 2σ[ppb] [%]
Benzene 1.067 0.005 0.45 0.0123 1.2 0.026 2.5
The precision was calculated from the series of sample measurements:
Area/Vol [units] 1σ 1σ
sample 1 sample2 sample3 sample4 sample5 Average Sdev
rel.
Sdev
10.932836 11.025449 11.062344 10.982789 11.026474 11.005978 0.05 0.45%
Measurement Guidelines VOC Version: Draft: 2012/07/18
Thus, the precision is ppbsampleprec 005.0067.1*0045.0 ===∆ σχχ
(The precision is automatically calculated in the ACTRIS data template for the round robin
measurements).
The uncertainty was calculated considering (i) the error of the NPL standard, (ii) the systematic
integrations errors for the NPL and the sample chromatogram, (iii) a blank value and (iv) the error due to
potential artefacts of the measurement system:
(i) Error of the NPL as given by the certificate is: 2% (this is the 2σ error). Hence, the 1σ uncertainty for
Benzene is 1%; ppbcal 02.0=δχ .
calcalsample
calsamplecal AV
VAδχχ *
*
*=∆
,
ppbppbmlunits
cal 01067.0)02.0(*ts8263.67uni* 401ml
401*6.4413 ==∆χ
(ii)
The min and max area for benzene in the NPL calibration gas were determined as described above (Fig.
11) for one chromatogram in the measurement series:
Area [units] (blank not substracted)
norm min max Avg
sdev/3
[units]
rel. sdev/3
[%]
Benzene 8472 8325 8557 8451 39 0.5%
The relative systematic integration error for the NPL is thus 0.5%.
Since the VOC_in_N2 gas has an identical matrix, includes the same components only at a lower mixing
ratio and chromatograms resemble each other (see Fig.4), the same relative integration error of 0.5% is
assumed for the sample, thus the error cancels for this specific measurement 1.
1 Assuming the gas mixtures, hence the chromatograms, did not resemble each other (e.g. as for the
analysis of an air relative to a VOC in N2 calibration gas mixture), and the errors did not cancel out, the
integration error would be derived following (assuming further an integration error of 1% for the
sample).
Measurement Guidelines VOC Version: Draft: 2012/07/18
(iii)
The blank value was determined using 2 zero gas measurements (n=2); the standard deviation of the
blank value is σblank=0.012ppb.
012.012
1*012.0
1
1* =
−=
−=∆ ppb
nblankblank σχ
(iv)
The deviation from the mean C-response is used to describe the error due to instrumental artefacts:
E.g. a significantly lower than expected C-response for a substance can be observed due to several
instrumental artefacts, such as too weak or too strong adsorbents in the pre-concentration unit, losses
due to decomposition within the instrument or the calibration gas itself, active surfaces, etc.
The error due to instrumental artefacts influencing the C-response factor can be described as
calsample
calsamplecalinstrument AV
VACresp
*
*** χδχ =∆
We have calculated the mean C-response from C5 – C8 Alkanes and Benzene. For Benzene the deviation
from the mean is %91.0=Crespδ and thus the error in mixing ratio due to this deviation is:
ppbml
mlppb
AV
VACresp
calsample
calsamplecalinstrument 0097.0
401*6667.8263
401*6.4413*2*0091.0
*
*** ===∆ χδχ
The total expanded uncertainty (precision + systematic errors, factor 2) for benzene in the VOC_in_N2
sample is then:
ppbsystematicprecunc 033.00157.0005.0*2*2 2222 =+=∆+∆=∆ χχχ
2
int,2
2
int,2int *
*
***
*
*
+
=∆ cal
calsample
calcalsamplesample
calsample
calcal AAV
VAA
AV
V δχ
δχχ
2
2
22int )6667.8263*005.0(*
6667.8263
2*6.4413)6.4413*01.0(*
6667.8263
2
+
=∆χ 1
Measurement Guidelines VOC Version: Draft: 2012/07/18
6.3 Data submission
Data submission has to be performed using the template provided by EBAS using the website:
http://ebas-submit.nilu.no/. The deadline for submission for data from one year, is 31 July of the
following year.
Only the names of the peak list below should be used for the submission of the data under ACTRIS-
EBAS. If single compounds are not resolved for butenes, pentenes, isohexanes and isoheptanes, the
sum of these compounds can be reported, if species are measured as a group.
Peak list:
ethyne cyclo-pentane styrene
ethene 2-methylbutane ethylbenzene
ethane n-pentane m-p-xylene
propyne 2-2-dimethylpropane o-xylene
propene benzene 2-2-4-trimethylpentane
propane cyclo-hexane n-octane
1-3-butadiene methyl-cyclopentane 1-2-3-trimethylbenzene
1-butene 2-methylpentane 1-2-4-trimethylbenzene
butenes 3-methylpentane 1-3-5-trimethylbenzene
cis-2-butene n-hexane p-cymene
2-methylpropene 1-hexene tricyclene
trans-2-butene 2-2-dimethylbutane alpha-pinene
n-butane 2-3-dimethylbutane camphene
2-methylpropane isohexanes sabinene
1-butyne toluene myrcene
isoprene methyl-cyclohexane alpha-pinene
1-pentene 2-methylhexane alpha-phellandrene
2-methyl-1-butene 2-2-dimethylpentane 3-carene
2-methyl-2-butene 2-4-dimethylpentane alpha-terpinene
3-methyl-1-butene 2-2-3-trimethylbutane ocimene
cyclopentene 2-3-dimethylpentane limonene
cis-2-pentene 3-3-dimethylpentane beta-phellandrene
pentenes isoheptanes gamma-terpinene
trans-2-pentene n-heptane terpinolene
7.References
Apel, E.C. et al. (2003): A fast-GC/MS system to measure C2 to C4 carbonyls and methanol aboard
aircraft. Journal of Geophysical Research 108 (D20): 15-1 – 15-12.
Barkley, C.S. et al. (2005): Development of a Cryogen-Free Concentration System for Measurements of
Volatile Organic Compounds. Analytical Chemistry 77 (21): 6989-6998.
Folkers, A. (2002): Oxygenated volatile organic compounds in the troposphere: Development and
employment of a gas chromatographic detection method. Report of the Research Centre Jülich 3998.
Dissertation University of Köln, Jülich.
Measurement Guidelines VOC Version: Draft: 2012/07/18
Greenberg, J.P., Zimmermann, P.R., Pollock, W.F., Lueb, R.A., Heidt, L.E. (1992): Diurnal variability of
atmospheric methane, nonmethane hydrocarbons, and carbon monoxide at Mauna Loa. Journal of
Geophysical Research 97: 10,395-10,413.
Helmig, D. (1997): Ozone removal techniques in the sampling of atmospheric volatile organic trace
gases. Atmospheric Environment 31 (21): 3635-3651.
Helmig, D., Greenberg, J.P. (1994): Automated in situ gas chromatographic-mass spectrometric analysis
of ppt level volatile organic trace gases using multistage solid-adsorbent trapping. Journal of
Chromatography A 677: 123-132.
Hopkins, J.R., Jones, C.E., Lewis, A.C. (2011): A dual channel gas chromatograph for atmospheric analysis
of volatile organic compounds including oxygenated and monoterpene compounds. Journal of
Environmental Monitoring DOI: 10.1039/c1em10050e.
Koppmann, R., Johnen, F.J., Khedim, A., Rudolph, J., Wedel, A., Wiards, B. (1995): The influence of ozone
on light nonmethane hydrocarbons during cryogenic preconcentration. Journal of Geophysical Research
100: 11,383-11,391.
Kuster, W.C, Goldan, P.D., Albritton, D.L. (1986): Ozone interferences with ambient dimethyl sulfide
measurements: The problem and a solution. Eos 67: 887.
Lee, J.H., Batterman, S.A., Jia, C., Chernyak, S. (2006): Ozone artifacts and carbonyl measurements using
Tenax GR, Tenax TA, Carbopack B, and Carbopack X adsorbents. Journal of the Air & Waste Management
Association 56: 1503-1517.
Leibrock E. (1996): Entwicklung eines gaschromatographischen Verfahrens zur Spurenanalytik von
oxidierten Kohlenwasserstoffen in Luft, Wissenschafts-Verlag Maraun, Frankfurt a.M., McClenny, W.A.,
Pleil, J.D., Evans, G.F., Oliver, K.D., Holdren, M.W., Winberrry, W.T. (1991): Canister-based method for
monitoring toxic V0Cs in ambient air, J. Air Waste Manage. Assoc. 41, 1308-1318
McClenny W.A., Pleil J.D., Evans G.F., Oliver K.D., Holdren M.W., Winberry W.T. (1991): Canister-based
method for monitoring toxic VOCs in ambient air J. Air Waste Man. Ass., 41, 1308–1318.
Palluau F., Ph Mirabel , M Millet (2007): Influence of ozone on the sampling and storage of volatile
organic compounds in canisters, Environ Chem. Lett. 5, 51–55.
Plass-Dülmer, C.; Michl, K.; Ruf, R.; Berresheim, H. (2002), C2 - C8 hydrocarbon measurement and quality
control procedures at the Global Atmosphere Watch Observatory Hohenpeissenberg. J. Chromatogr.
953, 175-197.
Plass-Dülmer C., N. Schmidbauer J. Slemr, F. Slemr, H. D'Souza (2006): European hydrocarbon
intercomparison experiment AMOHA part 4: Canister sampling of ambient air, J. Geophys. Res., 111,
D04306, doi:10.1029/2005JD006351.
Pollmann J., D. Helmig, J. Hueber, Ch. Plass-Dülmer, P. Tans (2008): Sampling, storage, and analysis of
C2–C7 non-methane hydrocarbons from the US National Oceanic and Atmospheric Administration
Cooperative Air Sampling Network glass flasks, J. Chromatogr., A 1188, 75-87.
Measurement Guidelines VOC Version: Draft: 2012/07/18
Sternberg J.C., W.S. Gallaway and D.T.L. Jones, in N. Brenner (1962), The mechanism of response of
Flame Ionization Detectors, in J. Callen and M.D. Weiss (Editors) Gas-Chromatography, Academic Press,
New York, p. 231-267.
Dietz W.A. (1967), Response factors for gas chromatographic analyses, J. of Gas Chromatography 5, 68.
US-EPA (1998): Technical assistance document for analysis of ozone precursors, US-Environmental
Protection Agency, EPA/600-R-98/161.
US-EPA TO-14A (1999): Compendium of methods for the determination of toxic organic compounds in
ambient air: determination of volatile organic compounds (VOCs) in ambient air using specially prepared
canisters with subsequent analysis by gas chromatography, US-Environmental Protection Agency,
Method TO-14A, 2nd ed., EPA/625/R-96/010b.
Measurement Guidelines VOC Version: Draft: 2012/07/18
8. Appendices
Appendix1: Ozone removal techniques for GC analysis of OVOC
Appendix 2: Adsorbents for sorbent-based enrichment of VOCs and OVOCs (oxygenated volatile
organic compounds) in ambient air samples
Appendix 3: Chromatographic separation (by J. Englert)
Measurement Guidelines VOC Version: Draft: 2012/07/18
APPENDIX 1: OZONE REMOVAL TECHNIQUES FOR GC ANALYSIS OF OVOC (OXYGENATED VOLATILE
ORGANIC COMPOUNDS) IN AMBIENT AIR SAMPLES (BY J. ENGLERT)
Reactions of concentrated VOC with ozone during sampling process may alter the quantities of the
target analytes and also contribute to the formation of artefacts which may mistakenly be
interpreted as atmospheric constituents.
Ozone reactions during cryogenic enrichment of VOC:
Ozone melting and boiling points (at atmospheric pressure) are at -192.1°C and 111.9°C. During
cryogenic freezeout of VOC from ambient air samples ozone is concentrated together with the target
analytes, whereas the main constituents of air nitrogen and oxygen do not condense under these
conditions (boiling point of liquid nitrogen -196°C). Reactions of VOC with ozone occur when heating
the cryogenic trap to transfer the analytes to the GC system. Alkenes, such as isoprene and
monoterpenes can be depleted in this reactions leading to artefacts like methacrolein and
methylvinylketone. By collecting ambient air into stainless steel canisters prior to the analysis with
cryogenic freezeout techniques this effect is reduced because of the short lifetime of ozone in these
canisters (HELMIG, 1997; GREENBERG et al., 1992). But this method is not suitable for oxygenated
VOC because of the high reactivity of these compounds on unheated stainless steel surfaces.
Ozone reactions during solid adsorbent sampling of VOC:
Ozone artefacts are formed on and with some sorbents (e.g. graphitised carbon sorbents and Tenax®
TA) leading to both VOC losses and formation (LEE et al., 2006; McCLENNY et al., 2001). Adsorbed
unsaturated hydrocarbons might for example undergo reaction with ozone during ambient sampling
leading to diminished alkene concentrations and the formation of oxygenated reaction products e.g.
acetaldehyde and formaldehyde. Products from ozone - Tenax® reactions include benzaldehyde,
phenol, acetophenone and n-aldehydes (HELMIG, 1997).
Reactions with ozone can be reduced by selectively removing the oxidant in the sample flow prior to
the concentrating of the analytes of interest. The ozone removing system should be easy to use,
inexpensive, efficient in the ozone removal rate and have a high scrubbing capacity, long lifetime and
eliminate the effects of ozone without interfering with the analytes of the target compounds and
without introducing contaminants. Furthermore it should be universally applicable to allow the
analysis of a wide range of compounds. Commonly reported techniques for ozone scrubbers include
impregnated filters, impregnated glass wool, coated tubes and coated annular denuders.
Catalytic destruction of ozone on metal surfaces:
Aluminium, copper, lead and tin have low ozone depletion efficiency whereas silver, iron, zinc, gold,
nickel, mercury and platinum have high ozone destruction capacities. The ozone removal
acquirement of some metals is used e.g. by nickel tubing, which reduces ozone levels to less than 20
% of ambient air level (HELMIG, 1997). Koppmann et al. (1995) found up to 50% destruction of
ambient ozone by pulling the sample air through stainless steel inlet lines kept at 67°C.
Measurement Guidelines VOC Version: Draft: 2012/07/18
HOPKINS et al. (2011): All gas transfer lines within the system are made from stainless steel and
heated to 70°C to reduce ozone mixing ratios.
Disadvantage: Loss of OVOC on the surface of stainless steel even at high temperatures (150°C).
Ozone deletion by nitric oxide (NO) titration:
Titration of the ambient air sample with a few ppm of NO prior to the concentration step is a very
efficient method to remove ozone. Ozone deletion performance depends on sufficient reaction time
and NO concentration in the mixing chamber. An example is the titration of the ambient air sample
for 20 seconds in a 1 litre glass reaction vessel with a small flow of 200 ppm NO in nitrogen resulting
in a NO concentration of 2 ppm. NO reacts with ozone to nitrogen dioxide (NO2) and oxygen (O2)
(HELMIG, 1997). The reaction is: O3 + NO → O2 + NO2 .
Disdavantage: slow reaction, alcohol losses (but constant)
Ozone deletion by potassium iodide (KI):
In many cases KI is used for ozone removal. This technique is very effective at ambient humidity
levels while capacity is reduced in dry air respected in following equation (HELMIG, 1997): O3 + 2KI +
H2O → O2 + I2 + 2KOH. KI reacts with ozone to potassium oxide (K2O) and elemental iodine.
Example: PTFE-lined stainless steel or Silcosteel capillary, OD 1/4″, 5 cm filled with KI-coated glass
wool.
Disadvantage: formaldehyde and acetaldehyde blank values, alcohol losses (HELMIG and
GREENBERG, 1994; LEIBROCK, 1996)
Sodium sulphite (Na2SO3):
most efficient in the presence of atmospheric water vapour and hence has to be positioned
upstream of a water trap – was found to remove 99% of the ozone in a humid ambient air stream but
inconsistent removal efficiencies from different suppliers and from different batches – testing of
individual ozone traps is required (HELMIG, 1997)
Example: ¼“ glass tube filled with 1 g of Na2SO3 anhydrous crystals held in place by glass wool plugs
and maintained at 100°C to prevent clumping of the Na2SO3
Disadvantage: removal of methylvinylketone and methacrolein.
Sodium thiosulphate (Na2S2O3):
The reaction between thiosulfate and ozone produces tetrathionate oxygen and water: 2S2O32- + O3
+ 2H+ → S4O62- + O2 + H2O
Measurement Guidelines VOC Version: Draft: 2012/07/18
pH dependency of this reaction
Example: ozone filters were prepared by flowing a 10% solution of aqueous Na2S2O3 through
commercial glass fiber filters followed by dry purge with nitrogen and had capacities in excess of 1
m3 air at ambient ozone levels
(HELMIG, 1997)
Advantage: this glass fiber filters also reduce sampling artifacts from reactions with halogens
Other ozone removal agents are copper oxide (CuO), magnesium sulphate (MgSO4), manganese
dioxide (MnO2), potassium carbonate (K2CO3) and TPDDC (see table 1).
In-line ozone scrubbers like granular KCl and crystalline Na2SO4 are prone to artefacts and require
regular maintenance so that they are not suited to long-term instrument deployments (HOPKINS et
al., 2011).
Table 1: Ozone removal techniques for VOC monitoring and their characteristics.
Technique Agent Characteristics
Coated annular denuder Potassium iodide (KI) Very efficient
Cellulose filter KI Improved formaldehyde and acetaldehyde recovery
Packed Teflon tubing Crystalline KI
Quantitative transmission of formaldehyde and acetaldehyde, partial loss of methacrolein and methylvinylketone
Impinger KI 2% Aqueous, buffered KI solution
Impregnated glass wool KI Quantitative ozone removal, iodated artefacts
Coated tubing KI in copper tubing Commercial scrubber KI in polyethylene cartridge Low capacity at 5% RH
Impregnated glass fiber filter Sodium thiosulphate (Na2S2O3) High capacity, also reduces sampling artifacts from reactions with halogens
Coated copper screen Manganese dioxide (MnO2) High capacity, possible losses of terpenes (e.g. camphor, linalool), loss of formaldehyde
Packed copper tubes Anhydrous 20-60 mesh potassium carbonate (K2CO3), crystalline
100% transmittance of light hydrocarbons
Packed Teflon tubing K2CO3 ozone and water removal, 100% transmission of light hydrocarbons
Packed glass tube Crystalline sodium sulphite (Na2SO3)
Loss of unsaturated compounds prevented, most efficient in the
Measurement Guidelines VOC Version: Draft: 2012/07/18
presence of atmospheric water vapour
Cartridge Copper oxide (CuO), crystalline No losses of carbonyl compounds
Trap Crystalline magnesium sulphate (MgSO4)
Removal of at least 100 ppb, loss of ozone removal efficiency with sampling length
Gas-phase ozone titration Nitric oxide (NO)
Very efficient, quantitative recovery of formaldehyde, formation of artefacts on Tenax exposured to elevated NOx levels, possible chromatographic interferences of NO and NO2 with NMHC (KUSTER et al., 1986), losses of alcohols, slow reaction
Metal tubing Nickel (Ni) Ozone reduced to less than 20% of ambient level
Spiked cartridge TPDDC (Tetramethyl-1,4-phenylenediamine dihydrochloride)
Sampling of carbonyl compounds on microcartridges containing porous glass particles impregnated with dansylhydrazine (DNSH), agent added to the reagent solution at the time of cartridge preparation to serve as an ozone scavenger
Spiked cartridge 5% Na2S2O3 aqueous solution on Tenax
Direct pretreatment of the adsorbent, improved monoterpene recovery
Spiked cartridge Na2S2O3 Interferences eliminated
Measurement Guidelines VOC Version: Draft: 2012/07/18
APPENDIX 2 (ADSORBENTS FOR SORBENT-BASED ENRICHMENT OF VOCS AND OVOCS
(OXYGENATED VOLATILE ORGANIC COMPOUNDS) IN AMBIENT AIR SAMPLES (BY JENNY ENGLERT))
Sampling of ambient air with sorbent tubes or traps and subsequent thermal desorption to transfer
the sampled compounds to a GC system is widely-used for trace gas analysis of VOC because of the
high sensitivity of this method.
There are two different sorbent-based sampling strategies: (1) on-line sampling of ambient air
directly into (cooled) sorbent focusing traps or transfer of air samples from containers (stainless steel
canisters or PTFE bags) into these (cooled) traps; and (2) off-line pumped (active) or diffusive
(passive) sampling onto adsorbent tubes or cartridges held at ambient temperature. In the case of
off-line sampling VOC are transferred in a second step into a cooled focusing device (e.g. sorbent
trap). For oxygenated VOC a method with short transfer from sampling device to the analysis system
is important because of the high losses of these analytes on surfaces, especially on unheated and not
inert ones like untreated surfaces of stainless steel.
When selecting a suitable sorbent or sorbent combination for VOCs and OVOCs several factors have
to be considered including sorbent strength, artefacts, hydrophobicity, inertness, thermal stability
and friability. It has to be verified that there is no breakthrough (most critical compounds methanol
and acetaldehyde), getting stuck or back-diffusion of target compounds. Some special, low volatile
analytes may also be lost through aerosol formation.
The sorbents must be strong enough to retain target analytes from a specific sample volume but
must also be weak enough to release them during thermal desorption. Sorbent strength is measured
in terms of breakthrough volumes that are defined as the litres gas per gram adsorbent required to
elute a VOC off 1.0 gram adsorbent at an indicated temperature. This capacity of solid sorbents
depends on temperature and is typically specified at 20°C. It approximately halves for every 10°C
rise. Therefore, cooling the traps during sampling increases adsorbent performance. The lowest
possible temperature is limited by the dewpoint of the sampled air (BROWN and SHIREY, 2001;
HELMIG and GREENBERG, 1994; WOOLFENDEN, 2010b).
When using hydrophilic sorbents (molecular sieves) or temperatures below the dew point for
ambient air samples some kind of water trap has to be installed in the sampling line. Otherwise there
would be a reduction of sorbent performance that might reach a factor of 10 at high humidity
conditions (90% RH) and after desorption of the trapped water moisture might interfere with the
following chromatographic analysis. Weak and medium strength sorbents (porous polymers and
graphitised carbon blacks) are hydrophobic and so they prevent trapping of excess water.
Some sorbents especially carbon blacks contain chemically active materials (trace metals) and are
unsuitable for labile (reactive) species. Most porous polymers except from Tenax® TA have high
inherent artefacts with blank peaks at 5-10 ng levels (WOOLFENDEN, 2010b).
Ozone artefacts are formed on and with some sorbents (e.g. graphitised carbon sorbents and Tenax®
TA) leading to both OVOC losses and buildings (LEE et al., 2006; McCLENNY et al., 2001). So the
aspect of ozone removal has to be considered in sorbent-based ambient air sampling.
Measurement Guidelines VOC Version: Draft: 2012/07/18
Quartz wool or silica beads are not able to retain most of the compounds. They are usually used in
multi-bed traps to prevent very high boilers to come in contact with a stronger adsorbent.
Porous polymers are weak or medium strength sorbents. None of them could retain the very volatile
analytes. In multi-bed traps they are often the first sorbent in sampling direction for the mid and
higher boiling point analytes beginning from benzene. Porous polymers are hydrophobic and so are
adequate for humid ambient air samples.
CarbopackTM, CarbotrapTM and CarbographTM are graphitised carbon blacks. The three different
types differ in mesh sizes. They are suitable for most of the VOC depending on their different sorbent
strength. The strongest CarbopackTM X should have a weaker adsorbent in front of it when sampling
very high boiling point analytes. All graphitised carbon blacks are hydrophobic like porous polymers
and so are adequate for humid ambient air samples (BROWN and SHIREY, 2001).
CarboxenTM and CarbosieveTM adsorbents are very strong and not appropriate for analytes with
boiling points higher than benzene because they have very small pores. They should always be used
with a weaker adsorbent (porous polymer or graphitised carbon black) placed in front. Pore shape of
the CarbosievesTM is different from the CarboxensTM. Pores of CarbosievesTM may be blocked by
analytes with high boiling points. Both CarboxensTM and CarbosievesTM are not hydrophobic and so
do need water removal for sampling humid ambient air samples.
Charcoals are not suitable for thermal desorption because they are too strong to release most of the
analytes with only heat. However, they are sometimes used in multi-adsorbent traps for very volatile
analytes e.g. Halocarbon 12 and Chloromethane. Charcoals are hydrophilic (BROWN and SHIREY,
2001).
Multi-adsorbent traps with up to four different sorbents allow a wide range of volatile compounds to
be enriched simultaneously. Sorbents are arranged in order of increasing sorbent strength from the
sampling end. Thermal desorption is in reverse direction to sampling flow so that low-volatile
compounds do not come in contact with the stronger adsorbent for highly volatile analytes. Care
should be taken when choosing sorbents for multi-adsorbent traps or tubes. The temperature
required for conditioning the most thermally-stable sorbent must not exceed the maximum
temperature of any other. Migration of loosely bound analytes from weak to strong adsorbent (e.g.
from Tenax® TA to a carbon molecular sieve) has to be inhibited by extending the bed length of the
weaker sorbent or inserting a medium strength sorbent between (WOOLFENDEN, 2010b). Multi-
adsorbent traps applied for oxygenated VOC are for example CarbopackTM B : CarboxenTM 1000, 90
mg in total (HOPKINS et al., 2003), 75 mg CarbopackTM B : 5 mg CarbopackTM X (ROUKOS et al.,
2009) or 70 mg Tenax® TA : 110 mg CarbotrapTM : 250 mg CarbosieveTM SIII (Folkers, 2002).
There are different adsorbent bed sizes and densities depending on application and analytes. To
allow high sampling flow rates coarse sorbent grain sizes (20/40 mesh) have to be used (HELMIG and
GREENBERG, 1994).
Measurement Guidelines VOC Version: Draft: 2012/07/18
Important characteristics of the most common sorbents and their adequacy for OVOC analysis are
summarized in table 1.
Sorbent Class Strength Max. Temp. [°C]
Relative analyte size to n-alkanes
Adequacy for OVOC (e.g. methanol, ethanol, ketones, aldehydes)
Characteristics
Quartz wool/silica beads
Fused silica
Very weak
>450 C30-C40
No, too weak – but suitable for cryogen enrichment
Very inert, non-water retentive, hydrophobic, minimal inherent artefacts, friable, 40/60 mesh recommended to minimise back pressure
CarbographTM 2TD CarbopackTM C CarbotrapTM C
Graphitised carbon black
Weak >450 C8-C20 No, too weak
Very inert, hydrophobic, minimal inherent artefacts, friable, 40/60 mesh recommended to minimise back pressure, ozone artefacts
Tenax® TA Porous polymer
Weak 350 C6-C30
No, too weak (LEIBROCK, 1996; WOOLF-ENDEN, 2010b)
Too weak for acetone and n-pentane, high benzene blank value, inert, hydrophobic, low inherent artefacts (e.g. aldehydes - HELMIG and GREENBERG, 1994), NMHC, aldehyde and ketone artefacts in combination with ozone (LEE et al., 2006), prone to chemical degradation and aging effects (HELMIG and GREENBERG, 1994)
CarbographTM 1TD CarbographTM B CarbopackTM B CarbotrapTM
Graphitised carbon black
Weak/medium
>450 C5/6-C14
Yes (HOPKINS et al., 2003; ROUKOS et al., 2009)
Hydrophobic, minimal inherent artefacts, friable, formation of fines, 40/60 mesh recommended to minimise back pressure, aldehyde and ketone artefacts in combination with ozone (LEE et al., 2006)
Chromosorb® 102
Porous polymer
Medium 225 C5-C12 Yes Inert, hydrophobic, high inherent artefact
Measurement Guidelines VOC Version: Draft: 2012/07/18
levels
PoraPakTM Q Porous polymer
Medium 250 C5-C12 Yes Inert, hydrophobic, high inherent artefact levels
Chromosorb® 106
Porous polymer
Medium 225 C5-C12 Yes Inert, hydrophobic, high inherent artefact levels
PoraPakTM N Porous polymer
Medium 180 C5-C8 Yes Inert, hydrophobic, high inherent artefact levels
HayeSepTM D Porous polymer
Medium 290 Yes (LEGREID, 2006)
Inert, hydrophobic, high inherent artefact levels
CarbographTM 5TD
Graphitised carbon black
Medium/strong
>450 C3/4-C8
Hydrophobic, minimal inherent artefacts, friable, formation of fines, 40/60 mesh recommended to minimise back pressure, retention of very volatile compounds e.g. 1,3-butadiene
CarbopackTM X
Graphitised carbon black
Medium/strong
>450 C3-C9
Yes (ROUKOS et al., 2009)
Hydrophobic, minimal inherent artefacts, friable, formation of fines, 40/60 mesh recommended to minimise back pressure, retention of very volatile compounds e.g. 1,3-butadiene, no ozone artefacts (LEE et al., 2006)
CarboxenTM 569
Carbonised molecular sieve
Strong >450 C2-C5 Yes
Inert, less hydrophilic than most carbonised molecular sieves, minimal inherent artefacts
UnicarbTM
Carbonised molecular sieve
Strong >450 C3-C8 Yes
Inert, hydrophilic, performance weakened in humid conditions, individual inherent artefacts, must be conditioned slowly, requires extensive purge to remove permanent gases
CarboxenTM 1003
Carbonised molecular sieve
Very strong
>450 C2-C5 Yes
Inert, hydrophilic, performance weakened in humid conditions, individual inherent
Measurement Guidelines VOC Version: Draft: 2012/07/18
artefacts, must be conditioned slowly, requires extensive purge to remove permanent gases
CarbosieveTM SIII
Carbonised molecular sieve
Very strong
>450 C2-C5 Yes
Inert, minimal inherent artefacts, significantly water and CO2 retentive, performance weakened in humid conditions, cold trap not lower than 0°C, easily and irreversibly contaminated by higher boiling components – protect with front bed of weaker sorbent
Molecular sieve 5Å
Molecular sieve
Very strong
>400 C2-C5 No (not inert)
High inherent artefacts, significantly hydrophilic, not suitable in humid conditions, easily and irreversibly contaminated by higher boiling components
Molecular sieve 13x
Molecular sieve
Very strong
>400 C2-C5 No (not intert)
High inherent artefacts, significantly hydrophilic, not suitable in humid conditions, easily and irreversibly contaminated by higher boiling components
Charcoal Activated carbon
Very strong
>400 C2-C4
No (WOOLF-ENDEN, 2010a)
Limited to solvent extraction (too strong and reactive for thermal desorption – metal content), hydrophilic, poor sensitivity – only for ppm level concentrations, analytical interference when using MS detection
Trademarks: Tenax® TA - Buchem bv, Netherlands; Chromosorb® - Celite Corporation, USA; PoraPakTM
– Waters Corporation, USA; CarbographTM – LARA s.r.l., Italy; UniCarbTM – Markes International Ltd.,
UK, USA; HayeSepTM – Hayes Separations Inc., USA; CarbotrapTM, CarbopackTM, CarboxenTM and
CarbosieveTM – Sigma-Aldrich, USA
Measurement Guidelines VOC Version: Draft: 2012/07/18
APPENDIX 3: CHROMATOGRAPHIC SEPARATION (BY J. ENGLERT)
1. PLOT columns
Table 1: PLOT columns
PLOT column equivalents GS-OxyPLOT (Agilent),
CP-LowOx (Varian)
CP-PoraBOND U
(Agilent resp. Varian)
AlO3 PLOT
(Agilent resp. Varian)
Polarity High polar Midpolar High polar
Composition Proprietary, salt
deactivated
Styrene-glycol
methacrylate
copolymer
Proprietary, salt
deactivated
Operable temperature
range 0°C to 350°C -100°C to 300°C -100°C to 200°C
Analysis of alcohols + + -
Analysis of aldehydes + + -
Analysis of ketones + + -
Analysis of ethers + + -
Analysis of esters + + -
Analysis of aromatics + + +
Analysis of alkanes - + +
Analysis of terpenes +/- +
Analysis of nitriles + +
Expected co-elution
problems
Ethyl
acetate+MVK+MEK (2-
butanone), water
peak+propanal and
acrolein
Methanol+n-butane,
butanal+benzene+
ethylacetate+MVK,
2-butanol+MEK,
butylacetate+
ethylbenzene+m+p-
xylene+n-hexanal,
pentanal+toluene
n-butane and ethyne
isohexanes
isoheptanes
m/p-xlene
Measurement Guidelines VOC Version: Draft: 2012/07/18
Advantage
Strong selectivity to
OVOC, high retention
of OVOC even at
above ambient oven
temperatures, no
retention of saturated
aliphatic NMHC and so
no co-elutions with
OVOC, long lifetime
Water resistance,
retention times not
influenced by water,
long lifetime
Strong selectivity on
light hydrocarbons
Disadvantage
Need for humidity
management,
retention of water,
tailing of unsaturated
OVOC due to reactions
with the polar column,
unsaturated NMHC
and aromatics both
with carbon atom
numbers higher than
eleven stick in the
column
Co-elutions of OVOC
with aliphatic NMHC,
retention of water
Not usefull for
OVOCs
Measurement Guidelines VOC Version: Draft: 2012/07/18
Examples of ambient air chromatograms
1A) Al2O3 (KCl) (from Rigi, Switzerland, Empa)
Measurement Guidelines VOC Version: Draft: 2012/07/18
Measurement Guidelines VOC Version: Draft: 2012/07/18
Fig.1: Al2O3 (KCl): a typical chromatogramme at Rigi (Switzerland).
Measurement Guidelines VOC Version: Draft: 2012/07/18
1B) LowOx
Fig.2: CP-LowOx (Varian), 10 m x 0.53 mm x 10.0 µm (HOPKINS et al., 2003).
Fig.3: CP-LowOx (Varian), 30 m x 0.53 mm x 10.0 µm (ROUKOS et al., 2009).
Measurement Guidelines VOC Version: Draft: 2012/07/18
Fig. 4: CP-LowOx (Varian), 30 m x 0.53 mm x 10.0 µm (measurements École des Mines de Douai,
Environmental & Chemistry Department, site: Paris suburban, 2010).
1C) PoraBOND U
Fig. 5: CP-PoraBOND U (Varian), 25 m x 0.32 mm x 7.0 µm (measurements Empa 2012; system
description: LEGREID, 2006):
4.0 min Methylether, 5.0 min Methanol, 5.1 min n-Butane, 5.5 min. 1,3-Butadiene, 5.9 min
Acetaldehyde, 7.9 min Ethanol, 9.2 min Isoprene, 9.9 min Acrolein, 10.0 min Propanal, 10.6 min
Methylacetate, 10.8 min Isopropanol, 11.1 min Acetone, 13.0 min MTBE, 13.3 Methacrolein, 12.6 n-
Propanol, 14.8 Ethylacetate, 14.9 Butanal + Benzene, 15.1 MVK, 15.5 2-Butanol, 15.6 MEK, 17.1 2-
Methyl-3-butene-2-ol, 17.8 n-Butanol, 19.8 Pentanal + Toluene, 24.1 Butylacetate + Ethylbenzene +
m+p-Xylene + n-Hexanal, 24.8 o-Xylene, 29.0 Benzaldehyde.
Measurement Guidelines VOC Version: Draft: 2012/07/18
2. Dimethylpolysiloxane columns
Table 2: Dimethylpolysiloxane columns
WCOT column equivalents
DB-1 (Agilent), CP-Sil 5 CB (Varian), Rtx-1 (Restek), BP-1 (SGE), SPB-1 (Supelco)
HP-5ms resp. DB-5 (Agilent), CP-Sil 8 CB (Varian), Rtx-5ms (Restek), BPX-5 (SGE), SPB-5 (Supelco)
DB-624 (Agilent resp. Varian), Rtx-624 (Restek)
Polarity Non-polar Non-polar Midpolar
Composition 100% Dimethylpolysiloxane
5%-Phenyl-95%-methylpolysiloxane
6% Cyanopropylphenyl-94%-dimethylpolysiloxane
Operable temperature range
-60°C to 350°C -60°C to 350°C -20°C to 260°C
Analysis of alcohols Tailing Tailing +
Analysis of aldehydes + + +
Analysis of ketones + + +
Analysis of ethers - - -
Analysis of esters + + +
Analysis of aromatics + + +
Analysis of alkanes + + +
Analysis of terpenes + + +
Analysis of nitriles - + +
Expected co-elution problems
Propanal+acetone, ethanol+acetone, n-pentane+acetone, n-butane+ acetaldehyde, OVOC+ NMHC
n-butane+acet-aldehyde+ methanol, isobutene+ methanol, ethanol+isopentane, acetone+propanal+ isopropanol, butanal+MEK, OVOC+NMHC
Propanal+acetone, OVOC+NMHC
Advantage High thermal stability More selective than DB-1, high thermal stability
Good retention of alcohols, good selectivity, good thermal stability
Disadvantage
Low selectivity, tailing of alcohols and ketones, co-elutions of OVOC with NMHC
Tailing of alcohols and ketones, co-elutions of OVOC with NMHC
Co-elutions of OVOC with NMHC
Measurement Guidelines VOC Version: Draft: 2012/07/18
2A) DB-1
Fig. 6: DB-1 (Agilent J&W), 100 m x 0.25 mm x 0.5 µm (RIEMER et al., 1998).
Measurement Guidelines VOC Version: Draft: 2012/07/18
2B) Rtx-1
Fig. 7: Rtx-1:
2C) BPX-5
Fig. 8: BPX-5 (SGE), 50 m x 0.22 mm x 1.0 µm (own measurements at Hohenpeissenberg
Meteorological Observatory, 2011):
19.49 min isobutene + methanol, 19.55 min acetaldehyde, 19.60 min n-butane, 21.29 min ethanol,
21.49 min isopentane, 22.01 min CCl3F, 22.55 min n-pentane, 22.67 min acrolein, 22.82 min acetone.
toluène
tétradécane
octanal
α-pinène
Measurement Guidelines VOC Version: Draft: 2012/07/18
2D) DB-624
Fig. 9: DB-624 (Agilent J&W), 10 m x 0.18 mm x 1.4 µm (APEL et al., 2003).
Measurement Guidelines VOC Version: Draft: 2012/07/18
3. Polyethylene glycol column
Table 3: Polyethylene glycol column
WCOT column equivalents
DB-WAX (Agilent), CP-WAX 52 CB (Varian), Rtx-WAX (Restek), BP-20 (SGE), SUPELCOWAX 10 (Supelco)
Polarity High polar
Composition Polyethylene glycol
Operable temperature range 20°C to 260°C
Analysis of alcohols +
Analysis of aldehydes +/-
Analysis of ketones +
Analysis of ethers +
Analysis of esters +
Analysis of aromatics +
Analysis of alkanes +/-
Analysis of terpenes +
Analysis of nitriles -
Expected co-elution problems
Butanal+acetone, methanol+MEK+3-methylfuran, ethanol+benzene+MVK, methylbutenol+toluol, 2-pentanone+pentanal
Advantage
High retention of alcohols, low retention of alkanes (less co-elution problems)
Disadvantage
Low retention of aldehydes, short lifetime of the column, cannot be operated at sub-ambient temperatures
Measurement Guidelines VOC Version: Draft: 2012/07/18
3A) CP-WAX 52 CB
Fig. 10: CP-WAX 52 CB (Varian), 60 m x 0.25 mm x 0.5 µm (FOLKERS, 2002).
3B) Rtx-WAX
Fig. 11: Rtx-WAX (Restek), 60 m x 0.53 mm x 0.5 µm (GOLDSTEIN and SCHADE, 2000).
Measurement Guidelines VOC Version: Draft: 2012/07/18
3C) DB-WAX
Fig. 12: DB-WAX (Agilent J&W), 60 m x 0.32 mm x 0.5 µm (LAMANNA and GOLDSTEIN, 1999).
Measurement Guidelines VOC Version: Draft: 2012/07/18
References:
Apel, E.C. et al. (2008): Intercomparison of oxygenated volatile organic compound measurements at
the SAPHIR atmosphere simulation chamber. Journal of Geophysical Research 113 (D20307).
Apel, E.C. et al. (2003): A fast-GC/MS system to measure C2 to C4 carbonyls and methanol aboard
aircraft. Journal of Geophysical Research 108 (D20): 15-1 – 15-12.
Apel, E.C. et al. (1998): Measurements comparison of oxygenated volatile organic compounds at a
rural site during the 1995 SOS Nashville Intensive. Journal of Geophysical Research 103 (D17): 22,295
– 22,316.
Folkers, A. (2002): Oxygenated volatile organic compounds in the troposphere: Development and
employment of a gas chromatographic detection method. Report of the Research Centre Jülich 3998.
Dissertation University of Köln, Jülich.
Goldan, P.D., Kuster, W.C. (2004): Nonmethane hydrocarbon and oxy hydrocarbon measurements
during the 2002 New England Air Quality Study. Journal of Geophysical Research 109 (D21309).
Goldstein, A.H., Schade, G.W. (2000): Quantifying biogenic and anthropogenic contributions to
acetone mixing ratios in a rural environment. Atmospheric Environment 34: 4997-5006.
Helmig, D. (1999): Air analysis by gas chromatography. Journal of Chromatography A 843: 129-146.
Hopkins, J.R., Lewis, A.C., Read, K.A. (2003): A two-column method for long-term monitoring of non-
methane hydrocarbons (NMHCs) and oxygenated volatile organic compounds (o-VOCs). Journal of
Environmental Monitoring 5: 8-13.
Lamanna, M.S., Goldstein, A.H. (1999): In situ measurements of C2-C10 volatile organic compounds
above a Sierra Nevada ponderosa pine plantation. Journal of Geophysical Research 104 (D17):
21,247-21,262.
Legreid, G. (2006): Oxygenated volatile organic compounds (OVOCs) in Switzerland: From the
boundary layer to the unpolluted troposphere. Dissertation ETH No. 16982, Zürich, Dübendorf.
Leibrock, E., Slemr, J. (1997): Method for measurement of volatile oxygenated hydrocarbons in
ambient air. Atmospheric Environment 31 (20): 3329-3339.
Leibrock, E. (1996): Development of a gas chromatography system for trace analysis of oxygenated
volatile organic compounds in air. Scientific journal 40, Fraunhofer-Institute for Atmospheric
Environmental Research, Garmisch-Partenkirchen. Dissertation Hamburg University of Technology,
Hamburg-Harburg.
Lewis, A.C. et al. (1995): Programmed temperature vaporization injection (PTV) for in situ field
measurements of isoprene, and selected oxidation products in a eucalyptus forest. Atmospheric
Environment 29 (15): 1871-1875.
Riemer, D. et al. (1998): Observations of nonmethane hydrocarbons and oxygenated volatile organic
compounds at a rural site in the southeastern United States. Journal of Geophysical Research 103
(21): 28,111 – 28,128.
Measurement Guidelines VOC Version: Draft: 2012/07/18
Roukos, J. et al. (2009): Development and validation of an automated monitoring system for
oxygenated volatile organic compounds and nitrile compounds in ambient air. Journal of
Chromatography A 1216: 8642-8651.
Vickers, A. (2007a): A "solid" alternative for analyzing oxygenated hydrocarbons – Agilent´s new
capillary GC PLOT column. Agilent Technologies publication 20 (2).
Access: http://www.chem.agilent.com.
Vickers, A. (2007b): GS-OxyPLOT: A PLOT column for the GC analysis of oxygenated hydrocarbons.
Agilent technical overview. Access: http://www.chem.agilent.com.
Zou, Y., Cai, M. (2008): Investigation of the unique selectivity and stability of Agilent GS-OxyPLOT
columns. Agilent Technologies Application. Shanghai, China. Access: http://www.chem