Why & How ? Near Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging Massachusetts General Hospital March 29 th , 2012

Transcript of Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical...

Page 1: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Outline Why & How ?

Near Infrared Spectroscopy and Diffuse Optical Imaging

Juliette Selb

Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Massachusetts General Hospital

March 29th, 2012

Page 2: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

The Optics Division

Director: David A. Boas http://www.nmr.mgh.harvard.edu/PMI/

The Optics Division at the Athinoula A. Martinos Center for Biomedical Imaging

Diffuse Optical Imaging of the Human Brain

Optical Breast Imaging for Cancer Applications

Microscopy for Cerebro-Vascular Physiology

Page 3: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Why Optical Imaging of the Brain?

Spa

tial S

ensi

tivity

(log

mm

)

Temporal Sensitivity (log sec)

But also: Novel contrast that spans all other methods!

PET/SPECT

2

1.5

1

0.5

0 -3 -2 -1 0 1 2 3 4

millisecond second hour day

mm

cm

m

EEG/MEG

NIRS fMRI

Suitable for sensitive studies (children, neurocritical care, movement) Non-invasive, low-cost, transportable

MRI

Page 4: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Optical Contrasts: Intrinsic & Extrinsic

Invasive: high spatial resolution, more contrast parameters

Non-invasive: routine continuous and real-time use in animals & humans

Hemodynamic •  Oxy- & deoxy- hemoglobin •  Blood volume •  Blood flow

Metabolic •  Cytochrome oxidase •  NADH-FADH •  CMRGlucose •  CMRO2

Neuronal - Electrical •  Voltage sensitive dyes •  Calcium green •  Scattering

Page 5: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Near Infrared Spectroscopy (NIRS) Pulse oximeter

2 wavelengths

Detector

Global brain oxygenation monitoring

Source

Detector

Cope et al., Med.Biol.Eng.Comput. 26 (1998)

Page 6: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Light Propagation in Biological Tissues

Page 7: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Light – Tissue Interaction

Absorption + Scattering

What happens when light travels through a biological tissue?

Page 8: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Absorption

HbO, HbR

Therapeutic window

Ultraviolet Near infrared

410 nm 600 nm 770 nm

Wavelength (nm) 300 400 500 600 700 800 900 1000 1100 1200 1300

0.01

0.1

1

10

100

Abs

orpt

ion

coef

ficie

nt (c

m-1

)

Cheong et al., IEEE J. Quantum Electron 26, 1990

Hale and Querry, Appl Opt 12, 1973

HbO, HbR

Water

Main chromophore (absorber) = Hemoglobin

Two or more wavelengths enable spectroscopy of oxy- / deoxy- hemoglobin

Page 9: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Scattering

ls*

ls

µs = 1 / ls ls = Average distance between two scattering events (mostly forward scattering)

l* = Distance after which diffusion is isotropic (same in all directions)

µ’s = 1 / l*

In tissues

~ 100 µm

~ 1 mm

Scattering coefficient

Reduced scattering coefficient

After ~1mm depth, photons have lost all information about their initial direction

Microscopy techniques also use optical contrast but work at shallow depths where photons have not or little scattered ! Can only achieve a few 100 µm depth sensitivity.

DOI work in the diffusive regime at several cm depth using only diffuse photons.

Scatterers (Cell nuclei, mitochondria, whole cells, …)

Note:

Page 10: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Tissue Penetration Absorption Scattering

la ~ 10 cm ls ~ 1 mm Average propagation distance

after which a photon is absorbed Average propagation distance after which a photon loses memory of its original direction

>>

Page 11: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Light Propagation in the Head

Source

Detector

Source

Detector

•  Diffuse light reaches the brain

•  Source + detector probing a volume of a few cm3 (Superficial + cerebral)

•  Sensitive to superficial cortex only (5-8 mm)

Page 12: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Different NIRS modalities

Whole Temporal Point Spread Function

or moments of TPSF

or gated TPSF

TPSF

Time domain

Source Detector

Attenuation

Continuous wave

Source Detector

Attenuation (AC, DC)

Phase Shift

Frequency domain

Source Detector

Page 13: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

NIRS modalities: Continuous Wave

Source

Detector

!  Least expensive !  High temporal resolution (25Hz) !  Not possible to distinguish between absorption and scattering changes !  Sensitive only to changes in hemoglobin

! Functional studies (fNIRS) ! Physiological oscillations

Continuous light intensity

Attenuation

NIRS signal

Page 14: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Modified Beer-Lambert Law

Cuvette of hemoglobin (Absorbing)

OD

L

I(") = I0(") exp[- µa(").L]

µa = 1/L ln(I/I0)

I0(") Beer-Lambert Law

OD = log(I/I0)

Absorption

Optical Density

Scattering + Absorbing Medium

OD

I0(")

DPF = Differential Pathlength Factor

Modified Beer-Lambert Law

G = Losses due to Geometry

Page 15: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Modified Beer-Lambert Law

Scattering + Absorbing Medium

DPF = Differential Pathlength Factor

I0(")

I(") I(t) = I(t=0) exp[- #µa(").DPF(!).L + G]

Δµa(t) = 1/(DPF.L) ln[ I(t) / I(t=0) ]

Cuvette of hemoglobin (Absorbing)

OD

L

I(") = I0(") exp[- µa(").L]

µa = 1/L ln(I/I0)

I0(") Beer-Lambert Law

OD = log(I/I0)

Absorption

Optical Density

Modified Beer-Lambert Law

Page 16: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Near Infrared Spectroscopy

Changes in concentration

Chromophore extinction coefficient at each wavelength

Multi-wavelength measurements of

optical absorption

Changes in oxy-hemoglobin

and deoxy-hemoglobin

Changes in oxygenation and total hemoglobin

#OD("1) = ("#1HbR #[HbR] + "#1

HbO #[HbO]) x Leff

#OD("2) = ("#2HbR #[HbR] + "#2

HbO #[HbO]) x Leff

Source S

Detector D

SD

Leff = SD x DPF Change in intensity

Leff

Page 17: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Changes in Oxy-hemoglobin [HbO]

Deoxy-hemoglobin [HbR]

Differential pathlength

Changes in absorption at multiple wavelengths

To Summarize: How does NIRS work? Source

Detector Measurements of changes in detected

light intensity

Changes in Total hemoglobin

[HbT] = [HbO] + [HbR] (blood volume)

& Oxygen saturation SO2 = HbO / HbT

Sel

b, P

our l

a S

cien

ce (2

005)

, in

Fren

ch

Arrays of sources and detectors

Hemoglobin maps

Page 18: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Different NIRS geometries Local

Topography (2D)

Tomography (3D)

Global

Gibson Phys Med Biol (2005)

Franceschini Psychophysiol (2003)

Selb, Pour la Science (2005), in French

Page 19: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Instrumentation

Page 20: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Instrument evolution - CW

TechEn Inc., Milford, MA, www.nirsoptix.com

I0 It

Page 21: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

32 channel CW system: CW6

TechEn Inc., Milford, MA, www.nirsoptix.com

I0 It

!  32 laser diode sources (690 & 830nm), frequency encoded in 200 Hz steps between 4.0 kHz and 7.4 kHz

!  32 parallel APD detectors

!  Real-time display of time courses of intensity changes and #OD

!  Acquisition time per image (32x32 channels) can be as short as 10ms!!!

Page 22: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Probe development

Page 23: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Probe Considerations

Combination with other modalities

Number of optodes? (= “optical electrode”) Local Whole head

Obviously over the studied cortical region(s)… Need for control optodes elsewhere?

Location?

Adapt probe design to experiment!!!

MRI compatible (non-magnetic material) MEG compatible + space constraint Neuro-ICU: other headgear?

Co-registration!! (10-20 system, fiducial markers, 3D-digitizer)

1 cm

Optode for MEG study

EEG-NIRS cap

Page 24: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Probe Considerations

Tolerance for movement? Infants, Children Clinical Postural changes, gait experiment Speech

Probe secured: Chin strap (problem if speech) Fiber weight (support fibers to avoid weight on the head)

Trade-off subject comfort / signal quality / time to setup

Adapt probe design to experiment!!!

Page 25: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

fNIRS (functional NIRS)

Page 26: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

What does fNIRS measure?

Neuronal Activity

Hemodynamic Changes: [HbR] [HbO]

CBF CBV

Increased Oxygen

Consumption

Decrease in oxygenation

Increases in blood flow, volume and oxygenation

Arteriole Dilation

NIRS signal

Increase in blood flow exceeds increase in O2 consumption

Page 27: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

fNIRS - Sensorimotor

-1.0 -0.5 0.0 0.5

6 7

8

1 2

3

4 AB

5

Moving average of 5 points acquisition time 800 ms

Δ[Hb] (µM)

Franceschini et al, Optics Express 6, 2000

Page 28: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

fNIRS - Sensorimotor

-1.0 -0.5 0.0 0.5

6 7

8

1 2

3

4 AB

5

Moving average of 5 points acquisition time 800 ms

Franceschini et al, Optics Express 6, 2000

Δ[Hb] (µM)

Page 29: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Finger opposition

Finger tactile

Electrical median nerve

Oxy

-hem

oglo

bin

Deo

xy-h

emog

lobi

n

fNIRS - Sensorimotor

Franceschini et al. Psychophysiology (2003)

3cm 1.9cm

Detectors Sources (690 & 830nm)

Hemoglobin maps

Page 30: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Functional NIRS (fNIRS)

Developmental / cognitive studies in infants/children:

!  Language development

!  Object processing

!  Number processing

!  …

!  Infants, Children, Adults, Healthy aging

!  Normal / pathological: schizophrenia, depression, Alzheimer, Parkinson’s…

!  Sensorimotor

!  Cognitive !  Auditory !  Visual

!  Cortical regions

fNIRS has been applied to a very broad range of studies

!  Population

!  Methodology !  Stand-alone

!  Multimodality (+ fMRI, EEG, MEG, TCD, …)

Page 31: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

fNIRS - Language Processing in Infants

Infants 6-9 months

Audiovisual (animation + speech)

Visual (animation only)

Bortfeld et al., NeuroImage, 34, 407-415 (2007)

Single trial

Heather Bortfelt (Univ Connecticut, Texas A&M at the time)

Page 32: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

fNIRS - Language Processing in Infants

Audiovisual

Visual

Bortfeld et al., Developmental Neuropsychology, 34:1, 52-65 (2009)

Bortfeld et al., NeuroImage, 34, 407-415 (2007)

Subject average Grand average N = 35 Left-Right Comparison N = 21

Lateralization of language processing in

left temporal cortex Robust patterns of activation in left temporal

and primary visual regions of neocortex

Page 33: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

NIRS + fMRI

Hoge et al. NeuroImage 2005 Huppert et al, NeuroImage 2006 Huppert et al, J Biomed Opt 2006

Multimodality fNIRS Cross-validation

Cerebral metabolic rate of oxygen

Ou et al, NeuroImage 2009

NIRS - 90%, 75%, 50% of max Δ[HbR] at t= 3-5s

MEG - Neural activation at t=35 ms

NIRS + MEG Neuro-vascular coupling

NIRS + EEG

Rob Cooper Meryem Ayse Yucel David Boas

Epilepsy

Page 34: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Limitations

Page 35: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Spatial Resolution and Localization !  Poor lateral resolution ~ source-detector separation

Overlapping measurements

White et al, J Biomed Opt 15 (2010) Joseph et al, Appl Opt 45 (2006)

Very dense probe

1st-nearest 2nd-nearest DOT

Page 36: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Spatial resolution and localization !  Localization (laterally and in depth)

Overlapping measurements

True subject anatomy

Custo et al NeuroImage 49(1): 561-567 (2010)

Atlas standard head

MRI structural constraint

Page 37: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Motion Artifacts Removal •  Visual identification (not objective!!)

Motion Artifact

Inte

nsity

(AU

)

Principal or independent component analysis filters can be used to separate and remove motion artifacts from physiology.

•  Threshold on magnitude of intensity change

•  Principal component analysis

Works well in infants but tends to remove too much of the activation signal in adults implemented in HomER

Page 38: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Systemic Physiology Contamination

Boas et al. NeuroImage (2004)

Blood pressure Respiration

NIRS signal

NIRS signal very sensitive to spontaneous physiological fluctuations arising from cardiac pulse, respiration, heart rate variability, blood

pressure spontaneous oscillations, both of systemic and cerebral origins

Real biological signals, but hinder our capacity to measure cerebral activation (“physiological noise”)

Various approaches to remove this physiological contamination from activation response

Page 39: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Other Diffuse Optical Modalities

Page 40: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

NIRS modalities: Frequency Domain

RF modulated light

DC AC Phase

!  Absolute baseline properties (absorption AND scattering)

!  Some depth resolution

" Baseline oxygenation and CBV

Source

Detector

ln(r

, Idc

)

r

intercept (µs’)

slope (µa, µs’)

phas

e

r intercept = 0

slope (µa, µs’)

O’Leary et al., Phys Rev Lett, 1992

(Multi-distance FD method)

Page 41: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

NIRS modalities: Time Domain

Light pulse

Times of flight of photons

!  Most information content vs. most expensive !  Absolute optical properties

!  Depth sensitivity

!  3D tomography

! Baseline oxygenation and CBV

Source

Detector

! fNIRS with depth resolution

Page 42: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Diffuse Correlation Spectroscopy

!  Blood Flow Index (relative value) !  Sensitive to microvasculature !  Alone or combined with NIRS modalities

Coherent light

Moving scatterers (Red Blood Cells)

Boas et al. Phys. Rev. Lett. 75(9):1855–8 (1995)

Durduran et al. Opt Lett 29(15): 1766-1768 (2004)

Static solution

Increased velocity

Intensity autocorrelation

Correlation time (s) In

tens

ity a

utoc

orre

latio

n

Page 43: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Summary Diffuse Optical Modalities Modality Detection + _ Applications

CW Fast Inexpensive

Relative changes in Hb concentrations

Functional NIRS Cortical connectivity Cerebral autoregulation

FD Absolute hemoglobin quantification

Slower than CW Longitudinal studies: baseline SO2, CBV

TD Absolute hemoglobin quantification Depth sensitivity

Most complex and expensive instrumentation Lower temporal resolution

Longitudinal studies: baseline SO2, CBV Surgery monitoring fNIRS with depth Volumetric images

DCS Blood flow index Relative measure

Baseline CBF Flow variations during postural changes

Attenuation

DC AC Phase

Times of flight of photons

Autocorrelation

Correlation time

Time

Time

Time

Slope of decorrelation

Page 44: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Baseline Measurements: Brain Metabolism in Infants

Page 45: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Cerebral Development in Healthy Infants

47 healthy infants

Age: 0-50 wks Gestational age: 27.0-41.5wks

right temporal

right parietal left

temporal

occipital left parietal

Simple Probe / Measurement

•  Frequency-Domain (Imagent ISS) •  7 wavelengths (670 to 830 nm) •  6-11 locations •  ~10 s / location

Franceschini et al. Pediatr Res 61(5) (2007)

Hemoglobin fit

Measurement points

Relative cerebral metabolic rate of oxygen (rCMRO2)

7 wavelengths

Page 46: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Cerebral Development in Healthy Infants

80

StO

2 (%

)

50

60

70

0 10 20 30 40 50 Age (wks)

1

3

5

7

$=3.5

rCM

RO

2 C

BV

(ml/1

00g)

2

3

4

5 Cerebral Blood Volume

Cerebral Oxygen Saturation

Cerebral Metabolic Rate of Oxygen CMRO2 increases over the first year of life

Consistent with increase in glucose metabolism (PET)

StO2 ~ constant with age

!

CBV increases over the first year of life ! Developing vasculature

Supply matches metabolic demand

!

Page 47: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Brain Development in Infants

FD-NIRS: CBV and StO2

DCS: CBF index

Grubb flow-volume relationship from adults not valid in newborns ! Need to measure blood flow

11 premature neonates (28-34.5 wGA)

Roche-Labarbe et al Human Brain Mapping 31(3) (2010)

with CBF deduced from CBV with CBF measured

Evolution rCMRO2 with age with CBF deduced from CBV

Combination of two DOI modalities enables much more accurate estimation of oxygen consumption!

Page 48: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Brain Injury in Infants

Not statistically significant elevated StO2 in brain injured with respect to healthy group

CBV significantly higher in brain injured than in

all other groups

Cerebral perfusion may adapt to normalize StO2

after injury

Elevation in CBV is consistent with a

hyperperfusion state

rCMRO2 significantly higher in brain injured

than in all other groups

CLINICAL GROUP

0.002

0

1

2

3

0.001 0.002

stable brain unstable healthy

$=2.6

CLINICAL GROUP stable brain unstable healthy 50

54

58 62

66 70

74

78 0.012

StO2 (%)

CLINICAL GROUP brain unstable stable healthy 1.2

1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2

0.0010 0.0008 0.0005

CBV (ml/100mg) rCMRO2

Increased CMRO2 suggests post-injury

excitotoxic mechanisms

Page 49: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Brain Injury in Infants sensitivity & specificity

CBV vs. rCMRO2 by clinical group

1.2 1.6 2.0 2.4 2.8 3.2 3.6 4.0 0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

2.2

2.4

2.6

rCM

RO

2 (w

ith re

spec

t to

norm

als)

CBV (ml/100mg)

sensitivity: 78.6% specificity: 96.6%

HII

Mit-HII

choroid plexus bleeds WM injury

MCA

HII

HII

Hypoglycemia

Mit

HII

Cerebral edema HII +10d hc

HII

WM injury

stable brain injured unstable healthy

Page 50: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Cerebral Health and Development in Infants StO2 has been the most commonly used

parameter to evaluate hemodynamics in infants

but

StO2 relatively insensitive to brain maturation age or brain condition

It is the wrong parameter to look at. Probable reason why NIRS failed to be

implemented in the clinic in the ’90s

Fortunately, NIRS provide additional measures:

- Cerebral Blood Volume - Cerebral Blood Flow - Cerebral Metabolic Rate of Oxygen

Page 51: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Summary

Page 52: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

NIRS principle Measurements of

changes in detected light intensity

Changes in Oxy-hemoglobin [HbO]

Deoxy-hemoglobin [HbR]

Changes in Total hemoglobin [HbT]

(blood volume) & Oxygen saturation

Changes in absorption at multiple wavelengths

Source

Detector

Hemodynamic Metabolic

Measurement or estimation of

Cerebral Blood Flow (CBF)

CMRO2

Novel contrast: hemodynamics + metabolic

Suitable for sensitive studies (children, neurocritical care, movement)

Non-invasive, low-cost, transportable

High temporal resolution (~ 10ms)

Relatively low spatial resolution (~ 1cm)

Page 53: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Summary Diffuse Optical Imaging

•  NIRS: Intrinsic absorption of oxy- and deoxy-hemoglobin

•  DCS: Sensitive to motion of red blood cells

•  Sensitive to microvasculature (as opposed to large vessels)

•  Continuous Wave: Relative changes in Hb

! Functional imaging

•  Frequency-Domain and Time-Domain: Absolute Hb content

! Longitudinal studies, Measurements over minutes to hours

Advantages

•  Low cost

•  Non-ionizing

•  Transportable

•  Functional contrast: HbO & HbR

•  High temporal resolution

Limitations & difficulties

•  Modest spatial resolution

•  Limited penetration depth

•  Superficial contamination

•  Probe design

•  Motion artifacts

Page 54: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Summary

Applications:

•  Functional brain imaging:

•  Infants, children, adults, healthy aging

•  Sensorimotor (gait, pain studies); Developmental studies in infants (language processing, number processing, object processing (shape/color), etc…); Psychiatric studies (cognitive side effects in depression)

•  Combination with other modalities: cross-validation, neurovascular coupling, CMRO2

•  Baseline measurements: brain monitoring during surgery and anesthesia, healthy and pathological infant cerebral development

•  Applications to other parts of the body:

•  Muscle oxygenation

•  Breast imaging (tumor characterization, treatment monitoring)

Page 55: Why & How ? Near Infrared Spectroscopy and Diffuse … Infrared Spectroscopy and Diffuse Optical Imaging Juliette Selb Optics Division Athinoula A. Martinos Center for Biomedical Imaging

Thank You! Diffuse Optical Imaging of the Adult Brain

David Boas

Rob Cooper

Jay Dubb

Louis Gagnon

Katherine Perdue

Juliette Selb

Meryem Ayse Yucel

Neonate Studies

Stefan Carp

Mathieu Dehaes

Angela Fenoglio

Maria Angela Franceschini

Pei-Yi (Ivy) Lin

Genevieve Nave

Breast Imaging

Stefan Carp

Qianqian Fang

Mark Martino

Bernhard Zimmermann