WG-16 Update fMRI & QIBA 2010 Nov 08 Doug Tucker Medical Numerics, Inc.

Transcript of WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that...

Page 1: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

WG-16 Update

fMRI & QIBA

2010 Nov 08

Doug Tucker

Medical Numerics, Inc.

Page 2: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

Outline

� QIBA Background

� QIBA fMRI

� fMRI Workflow

Page 3: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

Q IB A B a ckg round

� Quantitative Imaging Biomarkers Alliance

� Wiki: http://qibawiki.rsna.org

� Mission: “Improve the value and practicality of quantitative

biomarkers by reducing variability across devices, patients and

time.”

� Organized into 4 working areas (Committees):

� Quantitative-MR

� DCE-MRI subcommittee

� fMRI subcommittee

� Quantitative-PET

� Quantitative-CT

� COPD-Asthma

Page 4: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

How does QIBA ‘work’

Courtesy: Andrew Buckler

Page 5: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

What is a Profile?

� A Profile is a document.

� It tells a user what can be accomplished by following the Profile. ("Profile Claims")

� E.g. you will be able to detect volume changes of greater than <20%> in Stage IV Lung Nodules which are <5mm> in diameter or greater.

� It tells a vendor what they must implement in their product to state compliance with the Profile. ("Profile Details")

� E.g. to comply, the scanner must be able to: � scan a <Mark-324> Chest Phantom, identify the smallest resolvable target, display the diameter of that target

� demonstrate resolving targets at least as small as <2mm> diameter on the Mark-324 phantom

� scan patients according to the ACRIN NLST acquisition protocol

� E.g. to comply, the quantification application must be able to: � segment a nodule (automatically or manually), derive the volume, store it in a DICOM object

� run a user through a set of test data with known volumes and at the end display an accuracy score

� It may also tell the user staff what they must do for the Profile Claims to be realized. ("Profile Details")

� E.g. to comply, the site CT techs must be able to: � scan the patient within 10 minutes of contrast injection

� E.g. to comply, the radiologist must be able to: � achieve a score of <95%> or better using their segmentation application on the <Lung Invaders> test set.

�Retrieved from "http://qibawiki.rsna.org/index.php?title=What_Are_Profiles%3F"

Page 6: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

What about the IHE Process?

� How do Profiles fit in the IHE Process?

� The Profile concept and process has been borrowed from IHE

� Users identify a high priority multi-system problem.

� Vendors & Users collaborate to document a standards-based solution (a Profile).

� Vendors confirm it works by testing their implementations with other vendors (a Connectathon) (a Connectathon)

� How does QIBA differ from IHE?

� IHE mostly profiles engineering problems. QIBA also tackles scientific problems.

� When a profile claim is unvalidated or a profile detail cannot be defined until further research is done:

� IHE declares the profile premature and sets it aside until such groundwork is completed

� QIBA locates, encourages, and if necessary coordinates, getting the research completed

�Retrieved from "http://qibawiki.rsna.org/index.php?title=What_Are_Profiles%3F"

Page 7: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

QIBA fMRI Subcommittee

� Formed at RSNA 2009

� Chairpersons:

� Cathy Elsigner – NordicNeuroLab, Inc

� Jeff Petrella – Duke University

� Joy Hirsch – Columbia University

� Organization

� Participants from both academic and commercial backgrounds:

� Avotec, Inc

� BioClinica, Inc

� Buckler Biomedical LLC

� CHOP

� Columbia University

� Duke University

� GE Healthcare

� Harvard-Brigham and Women's Hospital

� Johns Hopkins University� Organization

� Committee with two working groups:

� fMRI-DICOM (DTucker)

� Reproducibility (JVoyvodic)

� Johns Hopkins University

� Lehigh Valley Diagnostic Imaging

� MAC

� MD Anderson Cancer Center

� Medical College of Wisconsin

� NordicNeuroLab, Inc

� Prism Clinical

� Roche

� Siemens

� Temple

� University of California, San Diego

� University of Texas Health Services Center at San Antonio

Page 8: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

What is the QIBA fMRI Committee working on?

Courtesy: Andrew Buckler

Page 9: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

Scientific vs Clinical Neuroimaging workflow1

‘Although scientific and clinical neuroimaging will share the majority of resources, the annotation of concepts

and instances will be slightly different, which can be explained by the contrast of “research flow” and “clinical

flow” (see figure).’

1. From: Nakai T, Bagarinao E, Tanaka Y, Matsuo K, and Racoceanu D. Ontology for fMRI as a Biomedical Informatics Method. Magn

Reson Med Sci, 7:3, 141-155, 2008.

Page 10: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

Procedure

Ordering

Patient

Assessment

and Training

Tests (paradigms) to be performed

( Modified) Paradigm(s) to be performed.

Workflow Model (proposed by me)

Data

Acquisition

Post-

Acquisition

Data

Processing

Interpretation

and Results

Distribution

Performed Paradigm(s)

Statistical Map(s) and Log(s)Cancel

Page 11: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

Methodology

� fMRI-DICOM working group asked members of the QIBA fMRI

Committee to describe their current clinical practice

� 6 respondents

� 3 Academic

� Duke

� Johns Hopkins

� Medical College of Wisconsin

� 3 Manufacturers

� Medical Numerics, Inc

� NordicNeuroLab, Inc

� PRISM Clinical Imaging

� Evaluation of responses to determine if model works to describe

data, or if a new model is needed.

Page 12: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

Ordering Step

� Respondents referenced the ordering process:

� “The patient’s initial entry into the workflow … through conventional methods established at each institution.”

� “… mapping is ordered as part of more comprehensive suite of brain imaging data…”

� “… the ordering physician initiates the fMRI workflow by making a choice of one or more fMRI tasks …”

� One respondent provided indications for use:

� ‘indicated either to assist diagnosis or to guide treatment planning and delivery’� ‘indicated either to assist diagnosis or to guide treatment planning and delivery’

� Most respondents alluded to a set of pre-existing tests (paradigms):

� “…general purpose paradigms such as visual field mapping and functionally specific tests…”

� “… language paradigms include: - auditory presentation of stimuli … -combined visual and auditory … - a picture naming task…”

� “… provide stimulus paradigms embedded within paradigm guidance software …”

� One respondent described the content of the fMRI task (paradigm)

� “… include patient instructions, acquisition parameters, and data processing parameters”

Page 13: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

Assessment & Training

� All workflows described some patient training and assessment step.

� Input to this step includes the patient, the trainer and the list of paradigms to be performed

� Rational:

� “… assessment should be made of the patient’s alertness, cognitive ability, and behavioral capabilities in order to flag any factors that could compromise …”

� “Patient training should be conducted … to familiarize the patient with the fMRI tasks … and to confirm patient’s ability to perform the task”

� Changes to paradigms:� Changes to paradigms:

� “changes in paradigm selection and/or instructions may be required”

� “Tune or replace paradigm selection as needed; may change difficulty, length, etc”

� Caveats:

� One respondents noted that for certain types of tasks (e.g. memory) the practice stimuli should be unique in order to avoid habituation.

� Output or work products:

� Trainer assessment

� “Trained paradigms == worklist”

� “If changes are required … should be noted … and remain part of the patient record.”

Page 14: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

Data Acquisition

� Typical scan include

� Preparatory QA/QC activities

� Augmented by additional equipment used to conduct fMRI examination include stimulus delivery hardware, response devices and synchronization devices

� Need to ensure patient can see and hear stimuli

� One respondent described their imaging protocol in depth:

� Most respondents described

� short training / refresh performed prior to each task

� Real-time patient monitoring for alertness, compliance and task-related physiological changes :

� Responses to tasks (typically hand or foot movement),

� Eye tracking

� Use of real-time fMRI analysisimaging protocol in depth:

� Structural Imaging

� 3-plane localizer

� Calibration scan for parallel imaging

� Whole-brain T1 (1mm^3 voxels)

� T2 FLAIR

� Shim scan

� Functional mapping scan (continuing on previous example)

� Vary by paradigm

� Prescribed using the same slices as the T2 images

� DTI

� Axial 15 direction centered to match fMRI scan

� Use of real-time fMRI analysis

� Evaluate patient motion

� QA / QC check of data

� Evaluate fMRI activation

� Post-scan evaluation for alertness, etc.

� Part of QA/QC record

� One respondent uses a 1 to 5 rating scale (both technician and self-assessment).

� One respondent specifically indicated that this should be done during the scan

Page 15: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

Post-Acquisition Data Processing

� Steps described include:

� Multiple reads/write cycles of data

� Reconstruction of time-series

� Corrections for warping and misalignment

� Data smoothing and normalization

� Evaluation for artifacts, data rejection, overall data quality measures

Clustering of regions of interest

� Output

� Documentation of processing including technical QA and evaluation

� One respondent described a standard multipoint evaluation of quality control protocol – evaluation of outliers

� 3D images

� Overlaying structural and functional data

� Color-coded activation� Clustering of regions of interest

� Image segmentation

� Computation of metrics and statistical measures

� Computational environments

� Respondents described performing processing:

� both on the acquisition device and on dedicated processing stations

� Using a combination of software tools (some FDA cleared, other research tools.

� Multiple orientations

� Variety of image formats

� Proprietary

� JPEG

� Integration of results onto PACS is important

Page 16: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

Display / Reporting / Interpretation

� Results become part of patient record

� Need for annotation and markups

� Referring physicians may not understand what they are looking at or for

� Integration into departmental image management systems (PACS) seen as important

� Respondents indicated they use standardize or template reports that include:

� Demographic information

� Indications

� Techniques

� Structural image finding

� Functional image finding

� Impressions

� Annotations

� May include references to DTI and Perfusion results if also performed.

� Export of images to other departments

� Surgical planning systems

Page 17: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

Procedure

Ordering

Patient

Assessment

and Training

Design Values of:

•Statistical Model

•Stimulus Presentation Model

•Acquisition Parameters

Exam/Patient Specific:

•Statistical Model

•Stimulus Presentation Model

•Acquisition Parameters

Workflow Model (Augmented)

Data

Acquisition

Post-

Acquisition

Data

Processing

Interpretation

and Results

Distribution

Design Values of:

•Processing Parameters

Exam/Patient Specific

•Image Data

•Patient Responses

•Statistical Model

•Performance Logs

Exam/Patent Specific:

•Statistical Maps

•Processing Logs

•Performance / Quality Logs

Page 18: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

fMRI

Control

System

Nondeterministic

Statistical Model

Acquisition

Parameters

Training &

Assessment

System

“Worklist”

Order

“Modified

Worklist”

Paradigm

Library

Stimulus

Presentation

SystemPatient

MR Image

Acquisition

System

Processing

System

MR

excitation

acquisition

Stimulation

Response

synchronization

Statistical

Maps

Reports

Image Data

Patient

Response

Data

(Default)

Processing

Parameters

Statistical Model

Description

Stimulus

Presentation

Model

(Updated*)

Statistical

Model

*For non-deterministic paradigms

Deterministic

Statistical

ModelTechnical

QC Reports

Reports

Page 19: WG-16 Update fMRI & QIBAqibawiki.rsna.org/images/7/7a/DrTuckerSlides_2010_11_WG...T2 images that this should be done during the DTI Axial 15 direction centered to match fMRI scan Evaluate

Next Steps & Discussion

� Open invitation to all to visit QIBA posters RSNA

� Initial Profile Development

� ETA – 6-9 months (??)

� Narrowing in on a set of claims for the first profile:

� fMRI activation is a valid indicator of the functional status of the associated brain tissue

� Presence of fMRI activation indicates that the brain tissue is responsive to the stimulus/task used.

� Absence of a BOLD response should indicate damaged, non-functioning tissue.

� One might even hope that graded response amplitude could be an indicator of partial function.

� fMRI activation evoked by a specific task (sensory, motor, or cognitive) indicates that the associated brain tissue is a necessary component of the brain network responsible for the tested function

� If the task is finger flexion, then resulting fMRI activation indicate those brain regions that contribute to the ability to flex the fingers.

� Caveats:� Caveats:

� Neurovascular uncoupling

� Disconnection

� Functional specificity

� Performance variablity

� Parallel work to define parameters, representation, etc.

� What needs to be store, when it needs to be stored, where it gets stored..

� Special thanks to

� Andrew Buckler – for contributing QIBA slides

� Bob Haworth – for reviewing slides and contributing ideas