Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of...
Transcript of Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of...
![Page 1: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/1.jpg)
Werner Strunz, Zahner-elektrik
www.zahner.de
![Page 2: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/2.jpg)
Outline
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 3: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/3.jpg)
1. Resistor [ R ]
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
0 1 2 3 4 5 6
2
Spannung
Strom
![Page 4: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/4.jpg)
2. Inductance [ L ]
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
0 1 2 3 4 5 6
2
Spannung
Strom
![Page 5: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/5.jpg)
3. Capacitor [ C ]
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
0 1 2 3 4 5 6
2
Spannung
Strom
![Page 6: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/6.jpg)
4. Warburg Impedance [ W ]
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
0 1 2 3 4 5 6
2
Spannung
Strom
![Page 7: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/7.jpg)
4. Warburg Impedance [ C ] - Example
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
?
![Page 8: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/8.jpg)
5. Nernst-Diffusion[ N ]
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 9: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/9.jpg)
5. Nernst-Diffusion[ N ] - Example (FC)
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 10: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/10.jpg)
6. Finite Diffusion [ FD ]
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 11: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/11.jpg)
6. Finite Diffusion [ FD ] - Example
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 12: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/12.jpg)
7. Constant Phase Element [ CPE ]
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
jYZCPE
11
0
1 100 10K
10
100
1m
10m
100m
|Z| /
0
15
30
45
60
75
90
|phase| /
o
frequency / Hz
-0.05 0 0.05 0.1 0.15
0
-0.2
-0.25
-0.1
-0.15
-0.05
Z' /
Z''
/
![Page 13: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/13.jpg)
CPE – Flexible Element
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 14: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/14.jpg)
Impedance of an Electrolyte Capacitor
HF : Z ~ (like L)
MF : Z ~ const (like R)
LF : Z ~ 1 / (like C) j > -90 ° => CPE
100m 1 3 10 30 100 1K 3K 10K 100K
100m
1
10
100
1K
|Z| /
0
30
60
90
|phase| / o
frequency / Hz
Constant Phase Element [ CPE ]
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 15: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/15.jpg)
Constant Phase Element [ CPE ]
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
a = - 1
Z w( ) = 1
C⋅
1
j ⋅w
a = - 0,5
Z w( ) = "W"⋅1
j ⋅w( )
a = 0
Z w( ) = R
a = + 1
Z w( ) = L⋅ j ⋅w
Capacitor Diffusion (Warburg) Ohmic Inductance
0 > a ≥ - 1 : Z w( ) = 1
Yo
⋅1
j⋅w( )a
[YO] = f ()
![Page 16: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/16.jpg)
CPE – a frequency dependent capacity
1
0 logloglog YC0YC
norm
norm
Without precaution
f norm = (2)-1
(natural = mathematical
normalization)
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 17: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/17.jpg)
Normalization of CPE (R || CPE !)
K.S. Cole, R.H. Cole; J. Chem. Phys. 9 (1941) 341–352
K.S. Cole, R.H. Cole; J. Chem. Phys. 10 (1942) 98-105
G. J. Brug, A. L. G. van den Eeden, M. Sluyters-Rehbach, J. H. Sluyters; Journal of Electroanalytic Chemistry 176 (1984) 275-295
C.H. Hsu, F. Mansfeld; Corrosion 57/ No. 9 (2001) 747-748
M.R. Shoar Abouzari, F. Berkemeier, G. Schmitz, D. Wilmer; Solid State Ionics 180 (2009) 922–927
B. Hirschorn, M. Orazem, B. Tribollet, V. Vivier, I. Frateur, M. Musiani; El. Acta 55 (2010) 6218–6227
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 18: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/18.jpg)
Constant Phase Element - Normalization
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
CPEC ZYC
Z
1111
0
0YC
0YCnorm
norm
OR
![Page 19: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/19.jpg)
Normalization of CPE (R || CPE) ! - simplified derivation
Normalized capacity is independent of the exponent
11
0||
||
1
1
YRC
YRwithj
RZ
jCR
RZ
CPER
CR
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 20: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/20.jpg)
Constant Phase Element [ R||CPE ]
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
Z w( ) = R⋅1
1 + RY0 ⋅ jw( ) a = R⋅ ∫
- ∞
∞1
1 + RC⋅ jw⋅G t( )dt
with ∫- ∞
∞
G t( )dt = 1
![Page 21: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/21.jpg)
Constant Phase Element [ CPE ] Distribution Function (G())
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
G t( ) =1
2p⋅
sin p⋅ 1 - a( )( )
cosh a⋅x( ) - cos p⋅ 1 - a( )( )
with x = lnt
t0
![Page 22: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/22.jpg)
Constant Phase Element [ CPE ] Distribution Function (G())
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
G t( ) =1
2p⋅
sin p⋅ 1 - a( )( )
cosh a⋅x( ) - cos p⋅ 1 - a( )( )
with x = lnt
t0
![Page 23: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/23.jpg)
Constant Phase Element [ R||CPE ]
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 24: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/24.jpg)
Constant Phase Element [ R||CPE ]
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
Rn: equidistand spacing in log(/0)
![Page 25: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/25.jpg)
Validation of Spectra
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 26: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/26.jpg)
Problems of Daily Life
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 27: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/27.jpg)
EIS-Principle at a Single Frequency
How to validate EIS-spectra ?
What‘s the specific property ?
0 500µ 1m 1.5m 2m 2.5m 3m-500
-400
-300
-200
-100
0
100
200
300
400
500Excitation at constant frequency
time /s
fun
ctio
ns
E(t
) &
I(t)
Z
j
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 28: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/28.jpg)
Motivation
Fuel cells
Batteries
Rechargeable batteries
Solar cells
Coatings
Development and/or improvement of important technical products
Under load
Under illumination
Water uptake
NON-STATIONARY
CONDITIONS
(may) result in
NON-STATIONARY
SPECTRA
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 29: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/29.jpg)
10m 100m 1 3 10 100 1K 10K 100K
10
20
15
30
25
|Z| / m
-30
0
30
60
phase / o
frequency / Hz
artifact
Reliable detection of artifacts
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
Motivation : what we need
![Page 30: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/30.jpg)
The Kramers-Kronig Relations
0
2
0
20
)(Im
2 - (0)Re=)(Re
d
HPVHH
0
2
0
200
)(Re2 =)(Im
d
HPVH
BUT WHERE ARE THE PROBLEMS ?
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 31: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/31.jpg)
The Limited Bandwidth Problem
• Simulation of a coating during water up-take
• Measured frequency range 100 KHz – 50 mHz
• 0 : ?
• : ?
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 32: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/32.jpg)
The measurement model
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 33: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/33.jpg)
-2 -1 0 1 2 3 4
-2.6
-2.4
-2.2
-2
-1.8
-1.6
-1.4
-1.2
0
15
30
45
60
75
90
Simulation
log frequency / Hz
log
im
pe
da
nc
e Z
/ O
hm
|Ph
as
e a
ng
le| / °
Measurement Model
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 34: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/34.jpg)
Measurement Model
- Drawback: “RC“ is not linear! Z RC ( ) =
R
1+RC⋅
Z RC ( ) =R
1+RC⋅= Z real ( ) + Z imag ( ) =
R
1+ RC⋅( )2− j⋅
R2C⋅
1+ RC⋅( )2
Z RC ( ) =R
1+⋅= Z real ( ) + Z imag ( ) =
R
1+ ⋅( )2− j⋅
R⋅⋅
1+ ⋅( )2
Complex number ⇒ Z = a + j⋅b
Complex Matrix Entry ⇒Zreal
Z imag
=a −b
b a
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 35: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/35.jpg)
Measurement Model
- Solution: “RC“ Replacement :“RC=“
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
Z RC ( ) =R
1+RC⋅
Z RC ( ) =R
1+RC⋅= Z real ( ) + Z imag ( ) =
R
1+ RC⋅( )2− j⋅
R2C⋅
1+ RC⋅( )2
Z RC ( ) =R
1+⋅= Z real ( ) + Z imag ( ) =
R
1+ ⋅( )2− j⋅
R⋅⋅
1+ ⋅( )2
Complex number ⇒ Z = a + j⋅b
Complex Matrix Entry ⇒Zreal
Z imag
=a −b
b a
![Page 36: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/36.jpg)
Linear-KK Check
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
Z RC ( ) =R
1+RC⋅
Z RC ( ) =R
1+RC⋅= Z real ( ) + Z imag ( ) =
R
1+ RC⋅( )2− j⋅
R2C⋅
1+ RC⋅( )2
Z RC ( ) =R
1+⋅= Z real ( ) + Z imag ( ) =
R
1+ ⋅( )2− j⋅
R⋅⋅
1+ ⋅( )2
Complex number ⇒ Z = a + j⋅b
Complex Matrix Entry ⇒Zreal
Z imag
=a −b
b a
![Page 37: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/37.jpg)
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
Linear-KK Tool from KIT Karlsruhe
![Page 38: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/38.jpg)
Linear-KK – Battery (I)
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 39: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/39.jpg)
Linear-KK – Supercap (Sub m-Range)
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 40: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/40.jpg)
Linear-KK – Coating (Huge-Z-Range)
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 41: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/41.jpg)
Linear-KK – Battery (II)
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 42: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/42.jpg)
Drift in Batteries
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
2,0 2,2 2,4 2,6 2,8 3,0 3,2 3,4 3,6 3,8 4,0 4,23,20
3,21
3,22
3,23
3,24
3,25
3,26 Discharge 4A (0.1 C)
Relaxation 0A
Battery
voltage / V
time of experiment / h
Measurement time ~ 6 h
![Page 43: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/43.jpg)
The Z-HIT Approximation (evaluation of impedance modulus from the phase angle
jj
ln
)(ln)(
2 + onst. )(ln
0
0d
ddcH
O
S
Detection of artifacts
Detection of instationarities (drift)
History (time) preserving
Reconstruction of causal spectra
=> Reliable interpretation of spectra
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 44: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/44.jpg)
Validation of Spectra – Z-HIT
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 45: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/45.jpg)
0 2 4 6 8 10 12 1411
12
13
14
15
16
ln Z
ln
-1,4
-1,2
-1,0
-0,8
-0,6
-0,4
-0,2
0,0(O)
E
S
Phase / rad
0 2 4 6 8 10 12 1411
12
13
14
15
16
Integral
(shifted)
ln Z
0 2 4 6 8 10 12 14-0,7
-0,6
-0,5
-0,4
-0,3
-0,2
-0,1
0,0
0,1
0,2
0,3
0,4
0,5
[C] ln Z - Integral (j)
dj / dln
0 2 4 6 8 10 12 14-0,7
-0,6
-0,5
-0,4
-0,3
-0,2
-0,1
0,0
0,1
0,2
0,3
0,4
0,5
[D]
[B][A]
ln Z - Integral (j)
/6 * dj / d ln
Deduction of the Z-HIT
Randle
circuit
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 46: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/46.jpg)
The Sensitivity of Objects (Z & j) - Excellent Examples: Sensors !
- Temperature Dependent Resistor (NTC, PTC)
Pt 100, Pt 1000, KTY 81, …
- Light Dependent Resistor (LDR)
- Magnetic Dependent Resistor (MDR)
- Humidity Dependent Capacity
- …..
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 47: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/47.jpg)
The Course of Phase and Impedance when Heating NTC/PTC
Z & j : Phase j is more stable than impedance Z
1 2 3 4 53,4
3,5
3,6
3,7
3,8
3,9
4,0
0
15
30
45
60
75
90
KTY 10k (NTC)
LO
G Im
ped
ance
/
LOG frequency / Hz
|Ph
ase| / °
0 1 2 3 4 5
3,05
3,10
3,15
0
15
30
45
60
75
90
Pt 1000 (PTC)
LO
G Im
ped
ance
/
LOG frequency / Hz
|Ph
ase| / °
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 48: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/48.jpg)
onst.ln
)(ln)(
2 )(ln
0
0 cd
ddH
O
S
jj
Implementation of the Z-HIT
1) The experimental data are filtered by a smoothing algorithm. The result is a set of continuous samples equidistant in log f.
2) The integral term is calculated by numerical integration.
3) The first derivate is taken from the smoothing function.
4) The integration constant is determined by a least squares fit.
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 49: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/49.jpg)
Battery under Load - Mutual Inductance & Drift
High-frequency Data (inductance) With kind permission of R. Gross, bno-consult, Dettelbach
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 50: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/50.jpg)
Z-HIT Examples
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
Battery (I)
![Page 51: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/51.jpg)
Z-HIT Examples
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
Supercap
![Page 52: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/52.jpg)
Z-HIT Examples
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
Battery (II)
![Page 53: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/53.jpg)
Restriction (2-Gate) Z-HIT
What is History (Time) Preserving?
Considering Kramers Kronig relations
0
2
0
200
)(Re2 =)(Im
d
HPVH
jj
ln
)(ln)(
2 + onst. )(ln
0
0d
ddcH
O
S
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
Integral-Term preserved
integration along the frequency axis leads
to “weighting“ (measuring time)
![Page 54: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/54.jpg)
History (Time) Preserving
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
Randle circuit with NTC as Charge Transfer Resistance
![Page 55: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/55.jpg)
History (Time) Preserving
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
Randle circuit with NTC as Charge Transfer Resistance
Only Smoothing Z-HIT refinement
![Page 56: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/56.jpg)
History (Time) Preserving
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
Randle circuit with NTC as Charge Transfer Resistance
Only Smoothing Z-HIT refinement
Dangerous: expanding the model
without physical justification
![Page 57: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/57.jpg)
Series measurement
Water Uptake - Waterborne Coating
With kind permission of U. Christ, Fraunhofer-Institut für Produktionstechnik und Automatisierung IPA, Stuttgart
1st..4th spectrum
5th spectrum 1 3 10 30 100 1K 3K 10K 100K
1K
10K
100K
1M
10M
100M
1G
|Z| /
0
30
60
90
|phase| / o
frequency / Hz
a a a aaaaaaaaaaaaaaa
aaaaaa
a
a
a
a
a
a
a
a
a
a
a
a
aaaaaaaaaaaaa a
a a
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 58: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/58.jpg)
Water Uptake - Waterborne Coating
With kind permission of U. Christ, Fraunhofer-Institut für Produktionstechnik und Automatisierung IPA, Stuttgart
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 59: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/59.jpg)
Z-HIT: Estimate of Accuracy (I)
= 1 = 0.5
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
![Page 60: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/60.jpg)
Z-HIT: Estimate of Accuracy (II)
100.
SIM
HITZSIM
Z
ZZErrorrel
W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling …
= 1 = 0.5
![Page 61: Werner Strunz, Zahner-elektrik 7. Constant Phase Element [ CPE ] W. Strunz, IWIS 2015: Check of Causality of Measured EIS and Modeling … Z D Y j Z CPE 1 1 0 1 100 10K 10 100 1m 10m](https://reader033.fdocuments.net/reader033/viewer/2022053117/609b0b9de35b9242b77b4d5f/html5/thumbnails/61.jpg)