Weighing Terminals - Atlantic Scale · Weighing Terminals IND560x Terminal The intrinsically safe...

3
Weighing Terminals IND560x Terminal The intrinsically safe IND560x weighing terminal brings uncompromising power and versatility into Division 1, Zone 1 and Zone 21 classified areas. Choose from analog or digital weighing technologies and communicate via serial, Ethernet and a variety of PLC interfaces. Easily program the feature-rich IND560x or create custom software applications. Whatever the need, the IND560x is the perfect solu- tion for hazardous area weighing. Weigh intuitively SmartTrac™ graphical display, operator prompting and one- touch access to terminal features through softkeys, work to improve productivity and minimize errors. Increase speed The IND560x features an ultra-fast A/D conversion rate. With TraxDSP™ tunable digital filtering to suppress environmental effects on weighing accuracy, enjoy precise, repeatable measurements from milligrams to tons. Connect easily Fast PLC integration and network access into the hazardous area available via the ACM500 communication module. Uncompromised weighing in hazardous areas Control process Internal discrete I/O options establish direct control of automatic and semi-automatic processes fully within hazardous areas. Zone 1/21, Division 1 IECEx

Transcript of Weighing Terminals - Atlantic Scale · Weighing Terminals IND560x Terminal The intrinsically safe...

Page 1: Weighing Terminals - Atlantic Scale · Weighing Terminals IND560x Terminal The intrinsically safe IND560x weighing terminal brings uncompromising power and versatility into Division

Wei

ghin

g Te

rmin

als

IND560x TerminalThe intrinsically safe IND560x weighing terminal brings uncompromising power and versatility into Division 1, Zone 1 and Zone 21 classified areas. Choose from analog or digital weighing technologies and communicate via serial, Ethernet and a variety of PLC interfaces. Easily program the feature-rich IND560x or create custom software applications. Whatever the need, the IND560x is the perfect solu-tion for hazardous area weighing.

Weigh intuitivelySmartTrac™ graphical display, operator prompting and one-touch access to terminal features through softkeys, work to improve productivity and minimize errors.

Increase speedThe IND560x features an ultra-fast A/D conversion rate. With TraxDSP™ tunable digital filtering to suppress environmental effects on weighing accuracy, enjoy precise, repeatable measurements from milligrams to tons.

Connect easilyFast PLC integration and network access into the hazardous area available via the ACM500 communication module.

Uncompromised weighing in hazardous areas

Control processInternal discrete I/O options establish direct control of automatic and semi-automatic processes fully within hazardous areas.

Zone 1/21, Division 1

IECEx

Page 2: Weighing Terminals - Atlantic Scale · Weighing Terminals IND560x Terminal The intrinsically safe IND560x weighing terminal brings uncompromising power and versatility into Division

Features and benefits•Intrinsicallysafedesignmeansnobarriersorenclosuresrequiredwhenplacingterminalinhazardousareasclassi-

fied Zone 1, Zone 21 or Division 1•Mobileweighingavailablewhenharsh/analogenclosuremodelisusedwiththeIND560x-PAB(PowerAdapterBoard)andavailableexternalNiMHbatterypack

•TraxEMT™(EmbeddedMaintenanceTechnician)automaticallylogscriticalsysteminformationtoaidwithstringentprocessvalidationrequirementsandcompliancetoregulatorycontrols.Builtindiagnostictestsimprove serviceability

•ACM500safeareacommunicationmodulesupportsavarietyofPLCinterfacesalongwithEthernetandserialports,allowing for easy integration of the IND560x into facility networks and broader control systems

•ReduceinstallationcostsbyutilizingtheACM200communicationmodulewhenonlyserialcommunicationtothesafeareaisrequired

•Communicationhardwareoptionsincludebothcurrentloopandfiberoptics.InterfacewithlegacyMETTLERTOLEDOdataconvertersisensured,providingacleanmigrationpathfromlegacyequipment

•15programmablesoftkeyshelpeliminateoperationerrorsandincreaseproductivity•Easilyaccessedmemorytablesstoretargetandtarerecords,eliminatingtheneedforexternaldatastorageand

improving process efficiency•Programmableon-screenpromptsguideoperatorsconsistentlythroughcomplexoperations,supportingfacilitySOPsandimprovingoverallprocessquality

•OperationandfunctionsimilartoIND560,reducingneedfortrainingwhenoperatingbothsafeandhazardousareaapplications

•Over/UndermodeofoperationisenhancedwithgraphicaldisplayandprogrammableI/Otoclearlyindicatetransac-tion status

•MaterialTransfermodesupportslatchedtargetcontrolI/O,makingtheIND560xanexcellenthazardousareafillingordosing controller

•Optionalapplicationsoftwaremodulesaddenhancedfeatureswhenrequired:

IND560x Terminal

−Fill-560:Advancedfillinganddosing −Drive-560:Simple(In/Out)vehicleweighing−COM-560:Supportforlegacycommunicationprotocols −TaskExpert™:Tocreatecustomapplications

Page 3: Weighing Terminals - Atlantic Scale · Weighing Terminals IND560x Terminal The intrinsically safe IND560x weighing terminal brings uncompromising power and versatility into Division

Subject to technical changes©08/2011 Mettler-Toledo, Inc.INDB0042.6E(LTR)

www.mt.com/IND560xfor more information

METTLERTOLEDOisuniquelyqualifiedtoprovidetheinstallation,calibrationand maintenance services essential to ensuring accuracy, optimizing uptime andprolongingequipmentlife.Ourservicerepresentativeswillbethereatthe right time, with the right parts, the right tools and the right skills to meet your needs.

Essential Services for Dependable Performance

METTLER TOLEDOService

Enclosure dimensions

(H x W x D)

Panelmount:160x265x92mm(6.3x10.4x3.6in.)

Harshenvironment/desk/wall:170x289x183mm(6.7x11.4x7.2in.)

Enclosure construction/

environmental protection

Panelmount:Stainlesssteelfrontpanel,aluminumchassis.CertifiedIP65.Lowprofilefrontpanelreducescontamination

Harshenvironment:Fullstainlesssteelhousing.CertifiedIP69k,IP68,IP65

Operating environment -10°Cto40°C(14°Fto104°F),10%to95%relativehumidity,non-condensing

Display 128x64dotmatrixLiquidCrystalDisplaywithbacklightstandard.21mmhighweightdisplay

Power WorkswithAPS768xintrinsicallysafepowersupply.120VACand240VACmodelsavailable.

Harshenvironment/analogmodelcompatiblewithexternalNiMHbatterypack,whenoptionalIND560x-PowerAdapterBoard(PAB)isinstalled

Keypad Power on/off. Clear, Tare, Print, Zero. Numeric keys. Navigation keyset. 5 softkeys

Scale types Supportssinglescaleinterface:Analogscale,5VDCexcitation,upto4x350Wloadcells(87Ωminimumresistance),2or3mV/V,100,000d

maximumresolution;IDNetscale,currenthigh-precisionTBrick-Exbases

Calibration Separatesingle-stepZeroandSpan,available5-pointlinearityadjustment,stepcalibrationorCalFREE™(notestweightsneeded).

Programandsavea25-stepcalibrationtestsequenceforguaranteedrepeatability.Configurablecalibrationexpirationandemailalertsavailable

Update Rates A/D conversion ≤366Hz.Targetupto50Hz.PLCupto20Hz.DiscreteI/Oupto25Hz

Serial connectivity Standard:OnestandardintrinsicallysafeRS-232port(COM1).

OptionalCOM4/COM5:Bi-directional,dual-channelfiberopticorintrinsicallysafecurrentloopinterfaceforcommunicationtosafearea

Peripheral

communication devices

Fiber optic or current loop connection to ACM500 in safe area; Current loop connection to ACM200 in safe area; Fiber optic connection to legacy

METTLERTOLEDOfiberopticconverter

Can operate as a remote display for other terminals in Ex or safe area

Supported

communication*

Serial(RS-232/422/485),Ethernet10Base-T,PLC(A-BRIO,AnalogOutput,DeviceNet,PROFIBUSDP,EtherNet/IP,ModbusTCP)

Inputs:ASCIII,CPTZ,SICSLevel0/1,Remotedisplay.Outputs:METTLERTOLEDOContinuous,Demand,Continuoustemplate,5configurabletemplates

* Some communication options only available with ACM500

I/O InternalsolidstatediscreteI/Ooptionsofferatotalof4inputsand6outputs.Choosefromoneofthreeoptions:ActiveInputs/ActiveOutputs,ActiveInputs/

Passive Outputs, Passive Inputs/Passive Outputs

- Active Inputs are powered internally by the IND560x and are designed for use with simple hazardous area switches

- Active OutputsarepoweredinternallybytheIND560xandprovide12VDCswitchingat50mAtotal.Theseareintendedforusewithlow-power,

intrinsically safe solenoids

- Passive Inputs and Passive Outputsrequirepowertobeprovidedexternallythroughanappropriatebarrier.Entityparametersmustbeconfirmed

TotalI/Ocontrolexpandableto12inputsand18outputsusingARM100modulesandRS-485fromtheACM500

Ex approvals US: IS,ClassI,II,III;Div1;GroupsA-GT4Ta=-10°Cto+40°C;

IP65;Type4

I/1/AExibIICT4

Controldrawing72191600

Canada: IS,ClassI,II,III;Div1;GroupsA-GT4Ta=-10°Cto+40°C;

Controldrawing72191600

ATEX:CertificateBVS09ATEX

E 010 and first supplement

II2GExib[opisGa]IICT4Gb

II2DExib[opisDa]IICT60°C Db

IP65

IECExCertificateIECExBVS10.0096X

Exib[opisGa]IICT4Gb

Exib[opisDa]IIICT60°CDb

IP65

Weights and measures

approvals

USA: ClassII,100,000d;ClassIII/IIIL,10,000d;CoC#05-0572A4

Canada: ClassII,100,000d;ClassIII,10,000d;ClassIIIHD,20,000d;AM-5593Rev3

Europe: ClassII,resolutiondeterminedbyplatformapproval;ClassIII,10,000e;TC6812Rev9

MID:Catchweighing-T10230;GravimetricFilling-T10231

Technical Specifications

Mettler-Toledo, Inc.1900PolarisParkwayColumbus,OH43240Tel.1-800-METTLERFax.1-614-438-4900