Wavelength dispersion seen in spectroscopy
description
Transcript of Wavelength dispersion seen in spectroscopy
Sample Placement in Spectroscopy
FTIR
Fourier transform infrared spectroscopy shines a beam containing many frequencies of light at once, and measures how much of that beam is absorbed by the sample
Solves problems of IR difficult mechanics hard to transport limited resolution slow and inaccurate analysis stray light limited sensitivity lack of reproducibility
Sample Location
Sample Location
Reasons for Sample Locations
The sample is located after the interferometer
The moving mirrors reflect an array of frequencies
As this mirror moves, each wavelength of light in the beam is periodically blocked or transmitted by the interferometer, acting similar to a monochromator, due to wave interference
Different wavelengths are modulated at different rates by the moving mirror
This set up for FTIR is used to see how the sample absorbs at an array of wavelengths in a short period of time
The Moving mirror causes the array of wavelengths within the FTIR spectrometer.
A. True B. FalseC. NeitherD. BothE. This is not the answer
Infrared Spectroscopy
Pass a beam of infrared light through the sample.
When the frequency of the IR is the same as the vibrational frequency of a bond, absorption occurs.
Examination of the transmitted light reveals how much energy was absorbed at each frequency (or wavelength)
This can lead to identification/study of different molecules
Sample location
The sample is placed in line with a reference for comparison
Sample placed before the chopper/splitter The chopper/splitter alternates which beam (sample or reference) enters
the monochromator and detector This is done to prevent fluctuations in the output of the source affecting
the data and allows the effects of the solvent of the sample to be cancelled out
The sample is placed before the monochromator so that it can adjust which wavelength range is detected by the IR detector
UV/Visible Spectroscopy
Utilizes light in the UV/Visible region to measure the absorption
Measures transition of non-binding electrons from ground state to excited state as opposed to fluorescence, which measures excited to ground state
UV/Visible Spectroscopy A beam of light from a visible
and/or UV light source is separated into its component wavelengths by a prism or diffraction grating.
Each wavelength is split into two equal intensity beams by a half-mirrored device.
One beam passes through the sample and the other through the reference or blank.
The intensities of these light beams are then measured by electronic detectors and compared.
Sample is placed behind the dispersive element
Why the sample is located behind the dispersive element
UV/visible spectroscopy is used in quantitative determinations of solutions especially transition metal ions, conjugated organic compounds, and biological macromolecules
These samples used in UV/visible spectroscopy only absorb at specific wavelengths in the UV/visible range of 200-800 nm
A single wavelength is needed to identify the compound within the sample, so the sample must be behind the monochromator
For Example:NADH absorbs at 340 nm
NADH is broken down into NAD+ to allow many reactions to occur
Utilizing Beer’s Law the rate of reaction (decrease in NADH concentration) can be determined by the decrease in absorbance as the reaction
One can determine whether an enzyme, such as MDH above, used in this kinetic assay was successful or not depending on the rate of reaction
Where is the sample in UV/Visible Spectroscopy located?
A. After the monochromator
B. Before the monochromator
C. Right in front of the deuterium lamp
D. You don’t need a sample in UV/Visible spectroscopy
E. This one is wrong
Fluorescence Spectroscopy
Emission at lower energy than absorption
Greater selectivity but fluorescent yields vary for different molecules
Detection at right angles to excitationS/N is improved so sensitivity is better
Fluorescent tags
In fluorescence spectroscopy…
A) We can select for wavelength by having the excitation source pass through a filter or monochromator
B) We want the detection at ~180° to the excitation for an improved signal/noise ratio
C) Emission occurs at an equal energy to absorption
D) All of the above
Fluorescence within Chromatography
The use of an excitation monochromator can be avoided using a laser because it emits light at a very narrow wavelength interval
Disadvantage: Cannot drastically change the wavelength
Resources
http://www2.chemistry.msu.edu/faculty/reusch/virttxtjml/spectrpy/uv-vis/spectrum.htm
http://en.wikipedia.org/wiki/Ultraviolet–visible_spectroscopy
http://en.wikipedia.org/wiki/Fourier_transform_infrared_spectroscopy
http://david-bender.co.uk/metonline/CHO/shuttles/shuttles9.htm
https://www.princeton.edu/~achaney/tmve/wiki100k/docs/Ultraviolet-visible_spectroscopy.html
http://en.wikipedia.org/wiki/Infrared_spectroscopy