Water Activity Measurements at NMISA
Regina MnguniNLA Test & Measurement
2018T206
![Page 2: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/2.jpg)
Background
• The NMISA started a cross cutter programme called Africa Food and Feed Reference Material (AFFRM).
• Water activity is very important in food substances, therefore its relevance in the programme.
• To measure this accurately, water activity meters needs to be calibrated.
• To support this programme, the humidity laboratory had to be accredited for the water activity meters.
![Page 3: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/3.jpg)
Introduction
• Humidity has different parameters which includes:
- Dewpoint Temperature
- Relative humidity
- Water Activity
- Moisture content
• Water activity (Aw) is free water/moisture available in a material (e.g food) as opposed the chemically boundmoisture.
• Directly related to the equilibrium relative humidity (%ERH)
![Page 4: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/4.jpg)
Introduction
• %ERH – relative humidity of air at which there is no net exchange of moisture with a nearby substance.
• Water activity is expressed on a scale of 0 aw to 1 awwhile %ERH is 0 to 100 %ERH.
• Aw is one the most critical factors in determining quality and safety of goods we consume every day.
• It affects shelf life, safety, texture and smell of food.
• Controlling the water activity is very important in food industry as low water activity prevents microbial growth (increases the shelf life).
![Page 5: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/5.jpg)
Water Activity Measurements
• Water activity is the ratio of the water vapour pressure of a material to the water vapour pressure of pure water under the same conditions.
• 𝐴𝑤 =𝑝
𝑝0=
%𝐸𝑅𝐻
100
• p = water vapour pressure of the material
• p0 = vapour pressure of pure water
• %ERH = Equilibrium relative humidity
![Page 6: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/6.jpg)
Water Activity Measurements
• Water activity measurements are mainly used in the food, paper and pharmaceutical industries.
• Packaging and storage conditions of the products are very critical for these industries
![Page 7: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/7.jpg)
Instruments
• Water activity instruments measure the amount of free water present in a sample.
• They are humidity instruments with a compartment/chamber for sample holding sealed from the surrounding.
• These are either chilled mirror hygrometers or relative humidity meters with capacitive or resistive sensor.
• Some instruments have detachable measuring chambers while others are embedded in the instrument.
![Page 8: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/8.jpg)
Instruments
• Different types are available commercially
![Page 9: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/9.jpg)
Instruments
![Page 10: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/10.jpg)
Calibration
• The traceability chain is the same as for relative humidity
Chilled mirror hygrometers
Contact thermometers
Unsaturated salts/ampoules
Relative humidity meters, saturated salts & water
activity meters
Humidity dewpoint
generators
![Page 11: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/11.jpg)
Calibration
• Relative humidity meters – normally calibrated in a climatic chamber or against reference salt solutions depending on the temperature and humidity range.
• With the water activity meters the procedure might be different because of the location of the sensor.
• At NMISA – unsaturated salt solutions (ampoules) as reference standard for relative humidity and water activity
![Page 12: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/12.jpg)
Calibration
• A pad is placed in a sample holder and soaked with the ampoule content of certain humidity (e.g 5 %rhor 0,05 aw).
• It is then placed in a chamber and sealed off to create %ERH
• After stabilisation, water activity and temperature readings are recorded
• The results of the meter are then compared with the ampoule’s certificate
![Page 13: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/13.jpg)
Ampoules
![Page 14: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/14.jpg)
Quality assurance of results
• The NMISA humidity lab went through international assessment to include calibration of water activity meters on the scope of accreditation.
• Any new method needs validation before accreditation
• The laboratory participated in the bilateral interlaboratory comparison
• The comparison was between NMISA and RotronicUK (reference lab)
![Page 15: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/15.jpg)
Quality assurance of results
• Rotronic UK supplied two water activity meters as the comparison artefacts
• The two were measured from 0,05 aw to 0,95 aw
using the ampoules as the reference standards.
• The corrections at each measured value were used as the comparison results
• The normalised error, En ratio method was used for the evaluation of measurement results.
![Page 16: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/16.jpg)
Quality assurance of results
• En ratio equation
• 𝐸𝑛 =𝑥𝐿𝑎𝑏−𝑥𝑅𝑒𝑓
𝑈𝐿𝑎𝑏2 +𝑈𝑅𝑒𝑓
2
• Where En = normalised error
xLab = measurement results of the participating lab (NMISA)
xRef = measurement result of reference lab (Rotonic UK)
ULab = Expanded Uncertainty of participating lab (NMISA)
URef = Expanded Uncertainty of the reference lab (Rotronic UK)
![Page 17: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/17.jpg)
Results
Table 1
Nominal Humidity
(aw)
Reference value (aw)
NMISA (aw) En ratio
0,050 0,001 -0,002 -0,38
0,100 0,001 -0,003 -0,55
0,200 -0,005 -0,002 0,33
0,350 -0,001 0,001 0,15
0,500 0,002 0,007 0,40
0,750 0,000 0,003 0,16
0,950 -0,002 0,006 0,41
Figure 1
-0,020
-0,015
-0,010
-0,005
0,000
0,005
0,010
0,015
0,020
0,000 0,200 0,400 0,600 0,800 1,000
Co
rre
ctio
n (
aw
)
Nominal Value (aw)
Aw-Therm SN 0061382591
NMISA Results Rotronic UK results
![Page 18: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/18.jpg)
Results
-0,020
-0,015
-0,010
-0,005
0,000
0,005
0,010
0,015
0,020
0,000 0,100 0,200 0,300 0,400 0,500 0,600 0,700 0,800 0,900 1,000
Co
rre
ctio
n (
aw
)
Nominal Value (aw)
Aw-Therm SN 0061382591
NMISA Results Rotronic UK results
![Page 19: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/19.jpg)
Comparison Artefacts
![Page 20: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/20.jpg)
Results
Table 2
Nominal Humidity
(aw)
Reference value (aw)
NMISA (aw)
En ratio
0,050 0,001 -0,001 -0,28
0,100 0,001 0,001 0,04
0,200 -0,004 0,003 0,87
0,350 -0,001 0,005 0,62
0,500 0,001 0,009 0,56
0,750 0,002 0,005 0,18
0,950 -0,002 0,007 0,46
Figure 2
-0,020
-0,015
-0,010
-0,005
0,000
0,005
0,010
0,015
0,020
0,000 0,200 0,400 0,600 0,800 1,000
Co
rre
ctio
n (
aw
)
Nominal Value (aw)
Hygrolab 3 SN 60308720
NMISA Results Rotronic UK results
![Page 21: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/21.jpg)
Results
-0,020
-0,015
-0,010
-0,005
0,000
0,005
0,010
0,015
0,020
0,000 0,100 0,200 0,300 0,400 0,500 0,600 0,700 0,800 0,900 1,000
Co
rre
ctio
n (
aw
)
Nominal Value (aw)
Hygrolab 3 SN 60308720
NMISA Results Rotronic UK results
![Page 22: Water Activity Measurements at NMISA€¦ · Calibration •Relative humidity meters –normally calibrated in a climatic chamber or against reference salt solutions depending on](https://reader034.fdocuments.net/reader034/viewer/2022042122/5e9cca9ffe328270123510f7/html5/thumbnails/22.jpg)
Conclusion
• Satisfactory results were produced from the bilateral interlaboratory comparison.
• The NMISA humidity laboratory is now accredited for the calibration of water activity meters.
• The accredited range is from 0,05 aw to 0,95 aw
• The calibration and measurement capability (CMC) ranges from 0,005 aw to 0,014 aw
• See SANAS website for the accreditation scope