VOA Analysis with a Headspace Sampler GC/MSD System · Tgt ppt R.T. Ion R2 Vinyl Chloride 1.91 62...

15
Agilent 7697A Headspace Sampler With 7890B GC and 5977A MSD Mike Szelewski Agilent Technologies Wilmington,DE February 14, 2013 January 8, 2013 Agilent Confidential 1 VOA Analysis with a Headspace Sampler GC/MSD System

Transcript of VOA Analysis with a Headspace Sampler GC/MSD System · Tgt ppt R.T. Ion R2 Vinyl Chloride 1.91 62...

Agilent 7697A Headspace Sampler

With 7890B GC and 5977A MSD

Mike Szelewski

Agilent Technologies

Wilmington,DE

February 14, 2013

January 8, 2013

Agilent Confidential

1

VOA Analysis with a

Headspace Sampler

GC/MSD System

Highlights for Headspace-GC/MS

7697A HSS - 7890B GC / 5977A MSD

1. MassHunter for Acquisition - it’s easy

2. MassHunter for Data Analysis - many features

3. Sensitivity (S/N) - better with the new extractor lens

4. LOD (IDL) as measured by replicate injections - meets your needs

5. Linearity as measured by R2 - near perfect

2

MassHunter 5977A Acquisition

3

1. Existing MSD Chemstation acquisition methods for SIM are easily converted to

MassHunter methods for 5977A

2. Simply load an existing 7890A/5975C method directly into MassHunter

3. SIM groups, SIM ions, and times are all preserved

4. Modification of SIM ions/groups is much easier than in MSD Chemstation

Note:

Acquisition parameters do not match the existing VOAs Headspace Analyzer. The

column flow chosen here matches that used by many customers.

MassHunter 5977A Acquisition, after directly loading a

7890/5975 method

4

All of the ions,

groups and

times loaded

without error

Page 5

Ramp 'C/min 'C Hold min

Initial 30 0.3

Ramp 1 10 185 0.0

Runtime 15.8 min

MSD Agilent 5977

Inlet MMI

Solvent Delay 0 min

Temp 150 'C

SIM Ions 3-15 ions/group

Mode Hot Split

Dwell 20-100 msec

Flow 1.0

# Groups 19

Quad Temp 180 'C

Source Temp 280 'C

Transfer Line 190 'C

Standards: methanol dilutions of commercial

standards from UltraScience. 2000 ppm stock

solutions containing 60 compounds and/or 6

gases diluted into methanol containing 2 ppm

of 3 ISTDs. Methanol spikes of 15 uL into 15

mL water for headspace use

Column DB-5MSUI part # (122-5533UI)

Injection volume 1 mL loop

Outlet Pressure Vacuum

30m x 0.25 mm id x 1.0 um film

Tune etune.u Gain = 5

Postrun none

Acquisition Mode SIM

Split Ratio 40:1

TID On

Method Parameters for 7697-7890/5977 VOAs

5

HS oven temp 70 'C

Equilibration 10 min

Liner 1 mm id, straight, no packing Extractor lens 3 mm

Equilibration 10 min

Sample size 15 mL in 20 mL vial

Alternate

column

DB-624 part # (122-1334)

20m x 0.18 mm id x 1.4 um film

Alternate column results were almost identical to the initial

column however peak shape was slightly worse for

the earliest eluters

SIM-TIC of VOA std, 53 compounds at 50ppt,

3 ISTDs at 2ppb and 6 gases at 100 ppt.

7697-7890/5977 Headspace-GC/MSD

Vinyl chloride

6

MassHunter 5977 Data Analysis - Existing MSD Chemstation

data analyses methods for are easily converted to MassHunter

methods. qdb.mth is the calibration database , ESTD in this case

7

Input the path to the

existing method you

want to convert

Use this default name

for the converted

quant method

Click this button

and WHAM !

you get this

MassHunter 5977A Data Analysis - converted from

MSD Chemstation, then calibrated.

8

The Scan

column had to

be changed to

SIM, the only

needed change

Mass Hunter 5977 Data Analysis - Benzene calibration from

50-20,000 ppt (0.05-20 ppb)

9

5 cal levels

Benzene

overlay with

TIC

50 ppt

Benzene

Cal curve

and ISTDS

Sensitivity Improvement up to 2.7x using the 5977A

Extractor Lens for Headspace Volatiles*

1. Signal-to-Noise varies from no improvement to 2.7x better, comparing runs with

and without using the extractor lens

2. Vinyl chloride - no S/N increase, signal and significant noise both have higher

amplitude using the extractor lens

3. Toluene - 2.7x S/N increase, very low noise background

4. S/N improvement can not be predicted by ion nor retention time

5. All ions show higher signal using the extractor lens compared to not using it

10

* 4 runs were made with each tune file. S/N was calculated for 14 compounds

in each run, then averaged

Limit Of Detection (LOD) based on 7 replicate

analysis of a 200 ppt standard in water

11

2013

2009 EC UK-EA 2009 EC

Named 5977 bid spec Reporting

Compounds LOD LOD Level

Vinyl Chloride 60 100 500

1,2-Dichloroethane 23 100 3000

Benzene 20 100 1000

Trichloroethene 21 100 10000

Tetrachloroethene 17 100 10000

LOD is based on area repeatability and not on S/N

All values are parts-per-trillion

Desired LOD is typically 5-10x less than reporting level

40:1 split delivers less water on column, good peak shape

and good reproducibility

Vinyl chloride lowest calibration level, 100 ppt

12

*1.889 min. Vinyl Chloride 0.1074 ppb

This is also the

requested LOD

level for the

UK-EA bid

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

1.2

1.8 1.9 2 2.1

Linearity reported as R2. 5 level calibration from 50

ppt to 20,000 ppt (0.05-20 ppb), ISTD

13

50-20,000

Tgt ppt

R.T. Ion R2

Vinyl Chloride 1.91 62 0.9999

1,2-Dichloroethane 4.81 62 1.0000

Benzene 4.99 78 1.0000

Trichloroethene 5.67 95 0.9999

Tetrachloroethene 7.83 166 1.0000

Vinyl chloride calibration levels are 2x higher

MassHunter 5977A Data Analysis - Vinyl chloride calibration

from 100-40,000 ppt (0.1-40 ppb)

14

5 cal levels

Vinyl

chloride

overlay with

TIC

100 ppt

Vinyl chloride

Cal curve and

ISTDS

Summary for Headspace-GC/MSD

7697A-7890B/5977A

1. MassHunter Acquisition - all MSD information on one page. If you can run MSD

Chemstation, you can easily run this. GC and Headspace screens the same as

before.

2. MassHunter for Data Analysis - very versatile for target compound quantitation.

Batch view speeds data review

3. Sensitivity (S/N) - up to 2.7x better with the new extractor lens compared to the

old drawout lens

4. LOD (IDL) as measured by replicate injections - all less than 100 ppt (0.1 ppb) for

60 common drinking water volatiles

5. Linearity as measured by R2 - 0.9996 - 1.0000 across a calibration range of 50 -

20,000 ppt

15