VOA Analysis with a Headspace Sampler GC/MSD System · Tgt ppt R.T. Ion R2 Vinyl Chloride 1.91 62...
Transcript of VOA Analysis with a Headspace Sampler GC/MSD System · Tgt ppt R.T. Ion R2 Vinyl Chloride 1.91 62...
Agilent 7697A Headspace Sampler
With 7890B GC and 5977A MSD
Mike Szelewski
Agilent Technologies
Wilmington,DE
February 14, 2013
January 8, 2013
Agilent Confidential
1
VOA Analysis with a
Headspace Sampler
GC/MSD System
Highlights for Headspace-GC/MS
7697A HSS - 7890B GC / 5977A MSD
1. MassHunter for Acquisition - it’s easy
2. MassHunter for Data Analysis - many features
3. Sensitivity (S/N) - better with the new extractor lens
4. LOD (IDL) as measured by replicate injections - meets your needs
5. Linearity as measured by R2 - near perfect
2
MassHunter 5977A Acquisition
3
1. Existing MSD Chemstation acquisition methods for SIM are easily converted to
MassHunter methods for 5977A
2. Simply load an existing 7890A/5975C method directly into MassHunter
3. SIM groups, SIM ions, and times are all preserved
4. Modification of SIM ions/groups is much easier than in MSD Chemstation
Note:
Acquisition parameters do not match the existing VOAs Headspace Analyzer. The
column flow chosen here matches that used by many customers.
MassHunter 5977A Acquisition, after directly loading a
7890/5975 method
4
All of the ions,
groups and
times loaded
without error
Page 5
Ramp 'C/min 'C Hold min
Initial 30 0.3
Ramp 1 10 185 0.0
Runtime 15.8 min
MSD Agilent 5977
Inlet MMI
Solvent Delay 0 min
Temp 150 'C
SIM Ions 3-15 ions/group
Mode Hot Split
Dwell 20-100 msec
Flow 1.0
# Groups 19
Quad Temp 180 'C
Source Temp 280 'C
Transfer Line 190 'C
Standards: methanol dilutions of commercial
standards from UltraScience. 2000 ppm stock
solutions containing 60 compounds and/or 6
gases diluted into methanol containing 2 ppm
of 3 ISTDs. Methanol spikes of 15 uL into 15
mL water for headspace use
Column DB-5MSUI part # (122-5533UI)
Injection volume 1 mL loop
Outlet Pressure Vacuum
30m x 0.25 mm id x 1.0 um film
Tune etune.u Gain = 5
Postrun none
Acquisition Mode SIM
Split Ratio 40:1
TID On
Method Parameters for 7697-7890/5977 VOAs
5
HS oven temp 70 'C
Equilibration 10 min
Liner 1 mm id, straight, no packing Extractor lens 3 mm
Equilibration 10 min
Sample size 15 mL in 20 mL vial
Alternate
column
DB-624 part # (122-1334)
20m x 0.18 mm id x 1.4 um film
Alternate column results were almost identical to the initial
column however peak shape was slightly worse for
the earliest eluters
SIM-TIC of VOA std, 53 compounds at 50ppt,
3 ISTDs at 2ppb and 6 gases at 100 ppt.
7697-7890/5977 Headspace-GC/MSD
Vinyl chloride
6
MassHunter 5977 Data Analysis - Existing MSD Chemstation
data analyses methods for are easily converted to MassHunter
methods. qdb.mth is the calibration database , ESTD in this case
7
Input the path to the
existing method you
want to convert
Use this default name
for the converted
quant method
Click this button
and WHAM !
you get this
MassHunter 5977A Data Analysis - converted from
MSD Chemstation, then calibrated.
8
The Scan
column had to
be changed to
SIM, the only
needed change
Mass Hunter 5977 Data Analysis - Benzene calibration from
50-20,000 ppt (0.05-20 ppb)
9
5 cal levels
Benzene
overlay with
TIC
50 ppt
Benzene
Cal curve
and ISTDS
Sensitivity Improvement up to 2.7x using the 5977A
Extractor Lens for Headspace Volatiles*
1. Signal-to-Noise varies from no improvement to 2.7x better, comparing runs with
and without using the extractor lens
2. Vinyl chloride - no S/N increase, signal and significant noise both have higher
amplitude using the extractor lens
3. Toluene - 2.7x S/N increase, very low noise background
4. S/N improvement can not be predicted by ion nor retention time
5. All ions show higher signal using the extractor lens compared to not using it
10
* 4 runs were made with each tune file. S/N was calculated for 14 compounds
in each run, then averaged
Limit Of Detection (LOD) based on 7 replicate
analysis of a 200 ppt standard in water
11
2013
2009 EC UK-EA 2009 EC
Named 5977 bid spec Reporting
Compounds LOD LOD Level
Vinyl Chloride 60 100 500
1,2-Dichloroethane 23 100 3000
Benzene 20 100 1000
Trichloroethene 21 100 10000
Tetrachloroethene 17 100 10000
LOD is based on area repeatability and not on S/N
All values are parts-per-trillion
Desired LOD is typically 5-10x less than reporting level
40:1 split delivers less water on column, good peak shape
and good reproducibility
Vinyl chloride lowest calibration level, 100 ppt
12
*1.889 min. Vinyl Chloride 0.1074 ppb
This is also the
requested LOD
level for the
UK-EA bid
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
1.1
1.2
1.8 1.9 2 2.1
Linearity reported as R2. 5 level calibration from 50
ppt to 20,000 ppt (0.05-20 ppb), ISTD
13
50-20,000
Tgt ppt
R.T. Ion R2
Vinyl Chloride 1.91 62 0.9999
1,2-Dichloroethane 4.81 62 1.0000
Benzene 4.99 78 1.0000
Trichloroethene 5.67 95 0.9999
Tetrachloroethene 7.83 166 1.0000
Vinyl chloride calibration levels are 2x higher
MassHunter 5977A Data Analysis - Vinyl chloride calibration
from 100-40,000 ppt (0.1-40 ppb)
14
5 cal levels
Vinyl
chloride
overlay with
TIC
100 ppt
Vinyl chloride
Cal curve and
ISTDS
Summary for Headspace-GC/MSD
7697A-7890B/5977A
1. MassHunter Acquisition - all MSD information on one page. If you can run MSD
Chemstation, you can easily run this. GC and Headspace screens the same as
before.
2. MassHunter for Data Analysis - very versatile for target compound quantitation.
Batch view speeds data review
3. Sensitivity (S/N) - up to 2.7x better with the new extractor lens compared to the
old drawout lens
4. LOD (IDL) as measured by replicate injections - all less than 100 ppt (0.1 ppb) for
60 common drinking water volatiles
5. Linearity as measured by R2 - 0.9996 - 1.0000 across a calibration range of 50 -
20,000 ppt
15