VetSpin ® Duo Centrifuge Vetlab Supplies Ltd Unit 13 Broomers Hill Park Broomers Hill Lane Pulborough RH20 2RY Telephone: 01798 874567 Fax: 01798 874787 Email: [email protected] www.vetlabsupplies.co.uk User Manual August 2019

Transcript of VetSpin Duo Centrifuge - Vetlab Supplies Ltd · 2019. 8. 1. · INTRODUCTION The VetSpin® Duo...

  • VetSpin® Duo

    Centrifuge

    Vetlab Supplies Ltd Unit 13 Broomers Hill Park Broomers Hill Lane Pulborough RH20 2RY

    Telephone: 01798 874567 Fax: 01798 874787 Email: [email protected] www.vetlabsupplies.co.uk

    User Manual

    August 20

    19

    http://www.vetlabsupplies.co.uk/https://vetlabsupplies.co.uk/products/veterinary-centrifuges/combispin/https://vetlabsupplies.co.uk/products/veterinary-centrifuges/combispin/https://vetlabsupplies.co.uk/about/contact-us/http://www.vetlabsupplies.co.uk

  • INTRODUCTION

    The VetSpin® Duo Centrifuge is ergonomically designed for centrifuging relatively small volumes of

    biological fluids at speeds of up to 10,000 rpm. It is particularly suited to Veterinary Practice labor-

    atories.

    The VetSpin® Duo has been designed for fast, simple, intuitive and safe operation. The single rotor

    can hold both microhaematocrit capillaries and larger plastic tubes up to 2ml in volume therefore

    avoiding the need to change rotors. Three pre-set optimised speed/time programs are available for

    serum/plasma, microhaematocrits, and urine sediments and accessed by one click of a key.

    However, both speed and time can each be simply changed incrementally in case different settings

    are required for 'special' samples.

    Plastic tubes of up to 2ml in volume are placed in angled tube holders in the rotor and

    microhaematocrit glass capillaries are fitted horizontally in grooves on top of the rotor.

    A screw on rotor lid is provided for use when running microhaematocrits (it is not needed when

    running plastic tubes on their own). The VetSpin® has a welded steel outer casing and an inner

    casing to protect the motor and enclose the rotor and an automatic lid lock ensures that the lid

    cannot be opened during running or braking.

    FEATURES

    Combination rotor holds both 75mm glass microhaematocrit capillaries and plastic tubes up to

    2ml volume.

    Plastic tubes are held at an angle in the rotor which makes them easier to load and unload,

    and helps to protect the integrity of the precipitated sample if there is a delay in unloading.

    This also means that the rotor lid is not required if these tubes are run on their own.

    3 fixed speed/time programs: serum/plasma, microhaematocrit and urine sediment can be

    quickly accessed at the press of key.

    Time is variable in increments of 1 min. up to max. 99 mins.

    Speed is variable in increments of 100 rpm up to max. 9999 rpm.

    Brushless motor for quiet running, fast speeds, fast braking and low maintenance.

    Automatic braking reduces cycle time.

    End of run audible alarm sounds when rotor has stopped confirming that it is safe to open lid.

    Automatic lid lock ensures lid cannot be opened until rotor has stopped.

    Four large durable suction pads / feet are fitted to ensure a tight grip and stability on bench

    especially when running.

    Casing is constructed of strong welded steel with internal casing to protect motor and rotor.

    Conforms to all CE directives.

  • Function/Parameter

    Technical Data

    Environment in Use

    For indoor use

    Clean, dry and ventilated

    Stable flat work surface

    Relative Humidity : ≤80%

    Temperature Range in Use +5°C to + 40°C

    Power AC220-230V 50/60HZ

    Unit Fuse 3A 230V

    Plug Fuse 5A Timer 1-99 minutes

    Max. Speed 9999 RPM

    Max. RCF 9500xg

    Rotor Volume 10 x 2ml plastic tubes and 10 glass 75mm capillary tubes

    3 Fixed programs 1: Serum/Plasma 9999 rpm 2 mins

    2: PCV (MHct) 9999 rpm 5 mins

    3: Urine Sediment 1500 rpm 2 mins

    Noise (at Max RPM) ≤65dB(A)

    Dimensions 300mm (L) *270mm (W) *190mm (H)

    Net Weight 9 kg

    VetSpin® TECHNICAL SPECIFICATIONS

    1.0 - GETTING STARTED - Unpacking your VetSpin® Duo Centrifuge

    1.1. Please unpack the centrifuge carefully and place onto a flat level surface near a power point with at least 12” of space surrounding it.

    Please make sure you keep all the packaging and accessories.

    1.2. Pull the Manual Lid Release Cord Vertically downwards to open the Lid.

    When the lid has released you can remove the packaging from around the Rotor. It is essential that the unit is not turned on while the packaging is still in place! So use the manual lid release cord for your set up procedure.

    Please Make sure the On/Off switch is in the off position!

    1.3. Close the lid of the VetSpin and plug the power lead firmly into the back of the unit and connect to power supply.

    1.4. Please Check Accessories Supplied:

    One Power Lead—5 Amp Fuse - Only use the lead supplied

    One Rotor Cover (for microhaematocrit /PCV procedures)

    One spare 3 Amp centrifuge unit fuse

    One Spanner for adjusting lid lock pin nut (Cat No: 4502)

    One Allen Key to allow removal of rotor for cleaning (Cat No: 4503)

    One Rotor Removal Tool Set ( Cat No: 4510)

  • 2.0 - OPERATING INSTRUCTIONS

    2.1. Turn the power on at the mains and then switch on the power at the back of the VetSpin. The front panel should light up with the red 'STBY' light on.

    2.2. To open the lid press the red 'STOP/LID' key and wait at least one second for the lock to click open and the lid lift slightly. Now lift the lid open carefully.

    Please do not touch the top of the lid with fingers/hands while

    pressing the STOP/LID key. When the lid is open, unscrew and remove the PCV Rotor Cover.

    Samples can now be loaded.

    2.3.1. Plastic tubes up to 2ml volume should be loaded in the tube holders ensuring they are BALANCED correctly. The rotor cover is not required unless microhaematocrit capillaries are run at the same time.

    To BALANCE: Use tubes of the same type, with equal

    volume of sample positioned directly opposite each other.

    Never run the centrifuge with missing TUBE HOLDERS!

    2.3.2. Microhaematocrit capillaries are loaded horizontally in the grooves on top of the rotor and do not need balancing but THE ROTOR COVER MUST BE FITTED to avoid sample splatter.

    NB: Remember to place the sealant end of the capillary next to the rim gasket on the outside edge of the rotor in a horizontal position. Also remember to record the sample ID against the rotor number next to the capillary. Please note that the microhaematocrit reading grid on the rotor lid is approximate and more accurate readings should be obtained with a separate reader. 2.4. Close the centrifuge lid slowly and carefully until it clicks into position.

    2.5. Check the desired program is highlighted by the yellow light. This can be quickly changed by pressing the yellow 'PROGRAM' key.

    NB: If a non-programmed speed/time setting is required simply select a program nearest to your desired settings - see 'Features’ for fixed program details - and adjust the settings accordingly with the +/- tabs underneath. Press key twice quickly in succession to effect an incremental change or, for larger changes, press key twice and hold key down until the desired setting is achieved.

    WARNING!

    THE ROTOR NUT IS AN ESSENTIAL PART

    OF THE UNIT AS IT SECURES THE ROTOR

    TO THE SPINDLE . WITHOUT THIS NUT IN

    PLACE THE ROTOR WILL LIFT OFF AT

    HIGH SPEED CAUSING SERIOUS DAMAGE

    TO THE CENTRIFUGE. TIGHTEN WEEKLY!

  • 2.6. Press the green 'START' key. 'RUN' light comes on.

    Stay near the centrifuge until it has reached full speed to ensure no vibration or unusual noise occurs. If these do occur, immediately press red key (the brake will apply automatically) and check balancing of tubes.

    2.7. At the end of the cycle wait until the unit BEEPS, Three Triple Beeps and the speed display has

    returned to the start speed. Press red key to open lid.

    ALWAYS LEAVE THE CENTRIFUGE LID OPEN SO IT CAN AIR AND REMAIN DRY

    TROUBLESHOOTING GUIDE

    Symptom Correction

    Loud motor noise and/or vibration when running.

    Stop and check TUBE BALANCE. If the noise is still present check the tightness of rotor with Allen key supplied and run without TUBES. If the problem persists please contact Technical Support.

    Lid Lock not releasing when red key is pressed.

    Lid can be unlocked manually by locating the white plastic plug or magnetic ring positioned underneath the front of the unit.

    The cord attached is pulled to release the lid. Must be pulled down Vertically POWER CUTS: In the event of a power cut, turn the unit off at the back as well as the mains and use the release cord to retrieve your samples. The manual lid release cord is not for regular use, your lid should open automatically at all times. —————————————————————————- The lid lock pin may need adjusting and/or the lid cushion may need replacing. This is a relatively simple process but please contact Technical Support first before attempting this.

    TECHNICAL SUPPORT - TEL: 01798 874871

    MAINTENANCE

    Weekly - Tighten Rotor Nut using the Allen Key provided

    Very little other maintenance is required but it is advisable to clean the bowl inside when appropriate. This can be done by removing the rotor. SEE OVER FOR INSTRUCTIONS. Also, occasionally, it may be necessary to replace consumables like the lid cushion and the rotor rim gasket which is necessary for supporting the sealant end of microhaematocrit tubes. These are available from Vetlab Supplies Ltd. See back page for details

    Fuse Replacement

    Unplug the unit from power source. The fuse is located underneath

    the power cord at the back of the centrifuge.

    Remove the fuse holder by using the end of a screwdriver. There is

    a spare 3AMP fuse located in the holder for your convenience.

    Replace the 3AMP Fuse and push the holder back in carefully.

    Replace the power cord and plug into power source.

    Manual Lid

    Release Cord

    Always Pull

    Down Vertically!

    3AMP Spare Fuse

    LID LOCK PIN

  • Important Instructions

    HOW TO...

    REMOVE THE ROTOR FOR CLEANING

    Please Follow The Instructions Carefully

    Please contact Technical Services: 01798 874871 If you have any Questions

    STEP 1

    PLEASE TURN OFF THE UNIT!

    USE THE ALLEN KEY PROVIDED (IN YOUR ACCESSORIES PACK)

    SUPPORT THE ROTOR AND REMOVE THE ROTOR NUT - RIGHT

    TIGHT, LEFT LOOSE!

    KEEP THE ROTOR NUT IN A SAFE PLACE! THIS MUST BE FITTED

    AT ALL TIMES WHEN USING THE CENTRIFUGE.

    STEP 2

    REMOVE 2 TUBE HOLDERS FROM OPPOSITE SIDES OF THE ROTOR.

    *PLEASE NOTE HOW MUCH OF THE MOTOR SHAFT SHOWS FOR

    REFERENCE WHEN REPLACING THE ROTOR AFTER CLEANING THE BOWL.

    STEP 3

    USING THE ROTOR REMOVAL TOOL SET, INSERT BOTH TOOLS

    OUTWARDS INTO THE CAVITIES AND THEN LIFT THE ROTOR OFF/UP

    VERTICALLY!

    BE CONFIDENT, THE ROTOR SHOULD BE TIGHT!

    STEP 4

    REPLACING THE ROTOR IS VISUAL - IT MUST SIT ONTO THE SHAFT PINS. LINE UP VISUALLY AND DROP THE ROTOR CAREFULLY AND VERTICALLY DOWN THE MOTOR SHAFT. THIS MAY TAKE A FEW ATTEMPTS. YOU WILL SEE THE SHAFT IN FULL WHEN CORRECTLY FITTED (*SEE STEP 2). SCREW THE ROTOR NUT BACK ON USING THE ALLEN KEY PROVIDED NICE AND TIGHT!

    Rotor Removal Tool Set

    WARNING - IT IS ESSENTIAL THAT THE ROTOR NUT IS REPLACED AS

    THIS NUT SECURES THE ROTOR ONTO THE UNIT. SERIOUS DAMAGE

    WILL BE CAUSED TO THE UNIT IF THIS IS NOT REPLACED AND THE

    CENTRIFUGE IS USED WITHOUT THE ROTOR SECURED TIGHTLY. IF

    THE ROTOR NUT IS MISPLACED DURING CLEANING...DO NOT USE THE

    CENTRIFUGE AND CONTACT US IMMEDIATELY FOR A REPLACEMENT

    ROTOR NUT.

  • Rotor Cover

    Lid Cushion

    Lid Lock Pin

    Rotor Rim Gasket

    Manual Lid Lock Release

    On/Off Power Switch

    Please always use the VetSpin® Power Lead provided.

    This Unit Requires a Power Lead with a 5 Amp Fuse.

    Rotor Nut

    Tube Holders

    Lid Interlock Safety Switch

    5 AM

    P FUSE

    Unit Fuse and Power Lead

  • WARRANTY INFORMATION The manufacturer warrants this instrument to be free from defects in material and workmanship under normal use for one year from the date of purchase (including electrical components).

    It does not cover damage resulting from abuse or misuse, repairs or alterations performed by other than authorised repair technicians, or damage occurring in transit.

    Please read the Terms & Conditions of your warranty carefully

    If you have any problems with your VetSpin® Duo Centrifuge please contact:

    Technical Support on 01798 874871

    Vetlab Supplies Ltd Unit 13 Broomers Hill Park Broomers Hill Lane Pulborough RH20 2RY

    Telephone: 01798 874567 Fax: 01798 874787 Email: [email protected] www.vetlabsupplies.co.uk

    www.vetlabsupplies.co.uk

    VetSpin® Duo Accessories

    Cat No: 4502 10mm Spanner x 1

    Cat No: 4503 5mm Allen Key x 1

    Cat No: 4504 Rotor Gasket x 1

    Cat No: 4505 Lid Cushion x 1

    Cat No: 4507 3 Amp Fuse x 1

    Cat No: 4508 PCV Rotor Cover x 1

    Cat No: 4510 Rotor Removal Tool Set x 1

    Please telephone us for up-to-date prices