Venflon Pro Safety P2P
Transcript of Venflon Pro Safety P2P
BD VenflonTM Pro Safety
Points to Practise
Product Information
Code No. Gauge Colour Catheter Specification Units per Length mm ODmm IDmm Flow Rate Box/Case H2O ml/min
393222 22G Blue 25 0.9 0.6 42 50/500
393224 20G Pink 32 1.1 0.8 67 50/500
393227 18G Green 45 1.3 1.0 103 50/500
393229 16G Grey 45 1.8 1.4 236 50/500
393230 14G Orange 45 2.0 1.6 270 50/500
BD MedicalThe Danby BuildingEdmund Halley RoadOxford Science ParkOxford OX4 4DQ
Tel: 01865 748844Fax: 01865 717313
www.bd.com
Product Features and Benefits
1 Sharp polished needle to ensure easy skin and vein penetration
2 BD VialonTM
cannula with clinically proven benefits
3 Cannula tip has a tapered design to reduce pain during insertion
4 Injection port with snap cap
5 Grip plate allows a more ergonomic grip to aid insertion
6 Needle grip specially designed to accommodate a variety of insertion techniques
7 End cap placed conveniently for easy handling
1
32
4
5
6
7
8
9
8 Passive safety mechanism automatically deploys on needle removal
9 A needle protection shield which: - fully encases the needle tip - is plastic and has no sharp edges - is designed to minimise blood exposure
Rubik's Cube ® used by permission of Seven Towns Ltd. www.rubiks.com
BD Venflon ProTM SafetyPoints to practise
Preparation
Insertion
Needle removal
• Complytoaseptictechnique throughout the procedure
• Disinfecttheinjectionsiteaccording to local policy
• Removethecannulafromthe packaging and lower the wings
• Adoptyourpreferredgripandremove the needle cover
• Insertthecannulaatalowangle
• Uponprimaryflashbacklowerthe angle almost parallel to the skin
• Advancethecannulaslightly,twoto threemillimetres,toensurethecannula tip is in the vein
• Stabilisethecannula
• Easetheneedlebacktwotothree millimetres
• Secondaryflashbackbetweentheneedle and cannula will confirm correct placement in the vein
• Advancethecannulacompletelyinto the vein
• Removethetourniquet
• Stabilisethecannula
• Occludetheveinjustabovecannula tip and withdraw the needle holding the needle grip or grip plate
• DoNOT hold onto the needle protection shield
• Disposeoftheneedleinasharps container
• ClosetheIVaccesswiththeend cap,otheradd-ondevicesoradd administration set
• Apply sterile dressing
• Flushthecannula to verify correct placement in the vein
BD MedicalThe Danby BuildingEdmund Halley RoadOxford Science ParkOxford OX4 4DQ
Tel: 01865 748844Fax: 01865 717313
www.bd.com
Safety, in safe hands
2 3
4 5 6
7 9
10
8
BD,BDlogoandallothertrademarksarepropertyofBecton,DickinsonandCompany©2011BD.