Venflon Pro Safety P2P

2
BD Venflon TM Pro Safety Points to Practise Product Information Code No. Gauge Colour Catheter Specification Units per Length mm ODmm IDmm Flow Rate Box/Case H2O ml/min 393222 22G Blue 25 0.9 0.6 42 50/500 393224 20G Pink 32 1.1 0.8 67 50/500 393227 18G Green 45 1.3 1.0 103 50/500 393229 16G Grey 45 1.8 1.4 236 50/500 393230 14G Orange 45 2.0 1.6 270 50/500 BD Medical The Danby Building Edmund Halley Road Oxford Science Park Oxford OX4 4DQ Tel: 01865 748844 Fax: 01865 717313 www.bd.com Product Features and Benefits 1 Sharp polished needle to ensure easy skin and vein penetration 2 BD Vialon TM cannula with clinically proven benefits 3 Cannula tip has a tapered design to reduce pain during insertion 4 Injection port with snap cap 5 Grip plate allows a more ergonomic grip to aid insertion 6 Needle grip specially designed to accommodate a variety of insertion techniques 7 End cap placed conveniently for easy handling 1 3 2 4 5 6 7 8 9 8 Passive safety mechanism automatically deploys on needle removal 9 A needle protection shield which: - fully encases the needle tip - is plastic and has no sharp edges - is designed to minimise blood exposure Rubik's Cube ® used by permission of Seven Towns Ltd. www.rubiks.com

Transcript of Venflon Pro Safety P2P

BD VenflonTM Pro Safety

Points to Practise

Product Information

Code No. Gauge Colour Catheter Specification Units per Length mm ODmm IDmm Flow Rate Box/Case H2O ml/min

393222 22G Blue 25 0.9 0.6 42 50/500

393224 20G Pink 32 1.1 0.8 67 50/500

393227 18G Green 45 1.3 1.0 103 50/500

393229 16G Grey 45 1.8 1.4 236 50/500

393230 14G Orange 45 2.0 1.6 270 50/500

BD MedicalThe Danby BuildingEdmund Halley RoadOxford Science ParkOxford OX4 4DQ

Tel: 01865 748844Fax: 01865 717313

www.bd.com

Product Features and Benefits

1 Sharp polished needle to ensure easy skin and vein penetration

2 BD VialonTM

cannula with clinically proven benefits

3 Cannula tip has a tapered design to reduce pain during insertion

4 Injection port with snap cap

5 Grip plate allows a more ergonomic grip to aid insertion

6 Needle grip specially designed to accommodate a variety of insertion techniques

7 End cap placed conveniently for easy handling

1

32

4

5

6

7

8

9

8 Passive safety mechanism automatically deploys on needle removal

9 A needle protection shield which: - fully encases the needle tip - is plastic and has no sharp edges - is designed to minimise blood exposure

Rubik's Cube ® used by permission of Seven Towns Ltd. www.rubiks.com

BD Venflon ProTM SafetyPoints to practise

Preparation

Insertion

Needle removal

• Complytoaseptictechnique throughout the procedure

• Disinfecttheinjectionsiteaccording to local policy

• Removethecannulafromthe packaging and lower the wings

• Adoptyourpreferredgripandremove the needle cover

• Insertthecannulaatalowangle

• Uponprimaryflashbacklowerthe angle almost parallel to the skin

• Advancethecannulaslightly,twoto threemillimetres,toensurethecannula tip is in the vein

• Stabilisethecannula

• Easetheneedlebacktwotothree millimetres

• Secondaryflashbackbetweentheneedle and cannula will confirm correct placement in the vein

• Advancethecannulacompletelyinto the vein

• Removethetourniquet

• Stabilisethecannula

• Occludetheveinjustabovecannula tip and withdraw the needle holding the needle grip or grip plate

• DoNOT hold onto the needle protection shield

• Disposeoftheneedleinasharps container

• ClosetheIVaccesswiththeend cap,otheradd-ondevicesoradd administration set

• Apply sterile dressing

• Flushthecannula to verify correct placement in the vein

BD MedicalThe Danby BuildingEdmund Halley RoadOxford Science ParkOxford OX4 4DQ

Tel: 01865 748844Fax: 01865 717313

www.bd.com

Safety, in safe hands

2 3

4 5 6

7 9

10

8

BD,BDlogoandallothertrademarksarepropertyofBecton,DickinsonandCompany©2011BD.