Validation Cell Analyzers
-
Upload
candiddreams -
Category
Documents
-
view
41 -
download
7
description
Transcript of Validation Cell Analyzers
![Page 1: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/1.jpg)
Validation procedures for cell analyzers
Dr Archana VazifdarDept. of Hemato-Pathology,
SRL Religare Ltd.
![Page 2: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/2.jpg)
Validation
A documented act of demonstrating that a procedure, process, & activity will consistently lead to expected results
FDA 1987
![Page 3: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/3.jpg)
Principles of automation
• Impedance – count and size cells by change in resistance produced as they are suspended in an electrically conductive medium
• Optical scatter- measures scatter properties of cells by laser light
![Page 4: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/4.jpg)
• RBC & Platelets measured in one channel
– RBC volume > 30-36 fl
– Platelet volume 2-20 fl
• Hb & WBC measured in second channel
• DLC in third channel
![Page 5: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/5.jpg)
Validation of automated CBC analyzers:
– interpretation of numeric, graphic & flag data
– Delta checks
– Review of peripheral smear
![Page 6: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/6.jpg)
Interpretation of data
![Page 7: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/7.jpg)
Normocytic Normochromic
RBC count
Spurious increase:•Giant platelets•High WBC counts (>50)
Spurious decrease:•Cold /warm agglutinins•Very small RBC•Cryoglobulins
![Page 8: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/8.jpg)
ADVIA 120
CELL-DYN
COULTER
Platelet count
Spurious increase:•RBC/ WBC fragments•Cryoglobulins•Lipids
Spurious decrease:•Platelet clumps•Giant platelets
![Page 9: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/9.jpg)
neutrolympho
Baso,mono, eos, blasts
WBC (FCM)
Normal WBC scatterplot
Normal WBC histogram
Impedance- VCS
![Page 10: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/10.jpg)
Optical scatter: ADVIA120DLC by Peroxidase method
Spurious increase
•PLT clumps & large platelets •Nucleated red cells•Resistant RBC’s
Spurious decrease:
•Clotted sample•Fragile cells- CLL•Lymphoid aggregates- UTI, B- cell NHL, CMML•Storage associated degeneration
![Page 11: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/11.jpg)
Flags
• A signal to the operator that the analyzed sample may have a significant abnormality/ does not meet acceptance criteria/ cannot be displayed
• Cause of errors:– Analyzer– Sample– Random run error
![Page 12: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/12.jpg)
RBC flagsSuspect flags• N’rbc, R’rbc, Micro RBC, RBC fragments,
– interfere with WBC & platelet counts• H & h errors• short sample, aged sample
Definitive flags• Anemia, anisocytosis, microcytosis,
macrocytosis, poikilocytosis• Erythrocytosis
![Page 13: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/13.jpg)
FLAG:Anemia, Microcytosis, anisocytosis
Hb 8.5RBC 3.2
Left shift of curve:
MicrocytosisIron Deficiency Anemia
β thalassemia trait Anemia of chronic diseases
![Page 14: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/14.jpg)
Conclusion:
s/o Iron Deficiency AnemiaAdvise Iron studies
ACTION:
RBC indicesMentzer’s index (MCV/RBC)=
18.3MI ≤ 13- BTT, ≥ 13- IDA
![Page 15: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/15.jpg)
Flags:•N’rbc, Micro RBC/ RBC fragments•Giant platelets•ThrombocytopeniaLt of curve not touching baseline:NoiseSchistocytes &/ extremely small rbcGiant platelets
PLT 140MPV 7.9PCT .148PDW 15
Hb 6.4
![Page 16: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/16.jpg)
Conclusion:
RBC count falsely ↓Platelets falsely ↑ (mask t’penia)
Hemolytic anemia
Action:
•RBC Indices- MCV, RDW•PLT Histogram- MPV & PDW •Review PS- RBC morphology
-PLT count (100)
![Page 17: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/17.jpg)
Bimodal peak: Dimorphic RBC population
Transfused cellsCombined deficiencyTherapeutic response in IDA
Hb- 8.6, MCH- 26.5, MCHC- 32.2
Flags:Dimorphic RBC population, anisocytosis
Action:
Review PS to identify cause
![Page 18: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/18.jpg)
50/ F, Hb-8.9, MCV-73, MCH- 25.6, RDW-26.8
Blood transfusion
![Page 19: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/19.jpg)
Dual/Combined deficiency
45/F, Severe pallorHb-5.1, MCV-96.7, MCH- 29.6, MCHC-31.4, RDW-24.5 TLC/Plt-Normal
S. Fe- 25TIBC- 144
S. Fe saturtn- 20.8S. B12- 158
![Page 20: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/20.jpg)
Right portion of curve extended:RBC agglutinationN’rbcsLeukocytosis
Flags:H&H error, N’rbc, dimorphic redsAnemia, macrocytosis, anisocytosis
H&H
• Sample related problems- turbidity-↑ Hb– Lipemia/ TPN– Cryoglobulins
• Autoagglutination• Hemolysis (in-vitro/vivo)• Spurious ↓ Hct• Clotted sample
Spurious ↑MCHC:
![Page 21: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/21.jpg)
corrected
Conclusion:False ↓ RBC, Hct, False ↑ MCV, MCH & MCHC
Cold agglutinin disease
After warming in H2O bath @ 37ºC for 15 mins
Action:Review PS: L/F agglutination vs n’rbc’s
![Page 22: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/22.jpg)
Short sample (microtainer)Repeat collection
Causes of H&H mismatch:
•partial sample aspiration/ improper mixing•Hb/ MCV measurement error/ very low•High WBC counts (interfere with Hb measurment)
•Cold agglutinins
![Page 23: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/23.jpg)
PlateletsSmallest guys largest culprits!!
• As platelet counts fall, reliability of analyzer decreases.
• Conventional methods are unable to provide consistently accurate results in lower range
• Clinicians using thresholds of 5-10 X 109/l must be aware of the limitations in precision and accuracy of cell counters
Linearity : 10–1,000 X 109/l
![Page 24: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/24.jpg)
Common platelets flags
• PLT Clumps – ↓Plt counts– Interferences with WBC Results (↑WBC
counts)• Giant platelets• Small platelets• PIC/POC delta- difference > 20,000• Thrombocytopenia- true/false
![Page 25: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/25.jpg)
Increased small sized particles:
Noise, debris, lipids, bacteria, fungi ? Wiskott Aldrich syndrome
Conclusion:
Falsely elevated platelet counts
Flags:Small platelets
Debris/ noise
![Page 26: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/26.jpg)
Action:
Review PS for platelet count
Conclusion:
Falsely ↑RBC countFalsely ↑WBC count
Falsely ↓ Plt count, ↑MPV
Giant platelets
Flags:Giant platelets, platelet clumpsCellular interference
Non fitted curve with increase in large cells:
Large platelets, clumps
![Page 27: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/27.jpg)
PIC/POC delta
•Excessive noise included in impedance count•Debris, bacteria, fungi•Plt clumps•Giant plt
45/M
![Page 28: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/28.jpg)
IG, Band, BlastsAty ly, Variant lyMPO, non viable WBCN’RBC, rst RBCPlt clumpOutside Reportable RangeLeukocytosis, monocytosis, basophilia, eosinophiliaUnable to Find Clear Separation between WBC subpopulations
WBC Flags
![Page 29: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/29.jpg)
Shoulder on the left of curve:
N’rbcLyse resistant RBCPlatelet clumps/ Giant plateletsFibrinImpedance noise
![Page 30: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/30.jpg)
Flags: IG, Blasts, eosinophilia,monocytosis, lymphopenia
CML
LeukocytosisThrombocytosisAnemia
![Page 31: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/31.jpg)
Flags:Aty lymphocyte, Variant lymphocyteNon-viable wbcLeukocytosisT’penia
Acute Leukemia
![Page 32: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/32.jpg)
38/F, k/c/o DM
Flag: leukocytosis, n’rbc, dimorphic reds
Conclusion:
21 nrbc’s/100 wbc- corr WBC= 17.35
DM in sepsis with liver abscess
Plt 100
![Page 33: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/33.jpg)
VCS:•Quantitative •Operator independent•Routinely available•Inexpensive
INCREASE MEAN NEUTROPHIL VOLUME (MNV)DECREASE MEAN NEUTROPHIL SCATTER (MNS) –left shift
–Lacking leukocytosis or neutrophilia
Newer Aspects: VCS-Neutrophil population data
Suggestive of acute bacterial sepsis
![Page 34: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/34.jpg)
Automated malaria detection
• “Gold standard” - thick & thin smear • Need for rapid, sensitive & cost-effective
screening technique
• Hemazoin pigment• Activation of neutrophils & monocytes• Increase volume heterogeneity (anisocytosis) of
monocytes & lymphocytes, detected by VCS
• ‘Positional parameters’, used as objective criteria for detecting presence of plasmodium
Clin. Lab. Haem., 26, 367–372 Automated detection of malaria
![Page 35: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/35.jpg)
Normal P lasmodium falciparum
Monocytes
Reactive LY
Parasitized RBC
Vol SD lymphocyte X SD Monocyte / 100 > 3.7
Am J Clin Pathol 2006;126:691-698Briggs et al / MALARIA DETECTION USING VCS TECHNOLOGY
shoulder
![Page 36: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/36.jpg)
• Specificity is 94% and sensitivity 98%
• PPV is 70% and NPV 99.7%.
• A flag indicating potential presence of malaria is a valuable diagnostic method for detection of malaria and may become a routine parameter in it’s diagnosis
![Page 37: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/37.jpg)
Case 1 38/M, No history available
![Page 38: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/38.jpg)
Result after treatment in H20 bath @ 37 ̊C
Cold agglutinin disease
![Page 39: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/39.jpg)
27/M, Hb 7, MCV 94, MCH 32, MCHC 35.7, RDW 14.6, Plt 158
Flags: Blasts, IG, n’rbc, rbc fragments, giant platelets
Case 2
![Page 40: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/40.jpg)
Conclusion:
Severe hemolysis following Primaquine ingestion in G6PD deficiency
50 nrbc’s/100 WBCSpherocytes +Giant platelets
![Page 41: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/41.jpg)
Case 3 : 33/M, Thrombocytopenia X 6 mnths, no bleeding. All other parameters WNL, ? ITP
Flags: n’rbc, micro rbc/ rbc fragments
![Page 42: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/42.jpg)
Action:
Change anticoagulant to Sodium Citrate
Platelet count- 243
Conclusion
EDTA dependant pseudothrombocytopenia(EDP)
![Page 43: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/43.jpg)
Case 4: 15/M, Fever
![Page 44: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/44.jpg)
Conclusion:
Plasmodium falciparum , PI 15%Thrombocytopenia
Malaria discriminant factor= 6.3
![Page 45: Validation Cell Analyzers](https://reader038.fdocuments.net/reader038/viewer/2022102808/55cf968d550346d0338c3cae/html5/thumbnails/45.jpg)
THANK YOU
Archana Vazifdar, M.D.SRL Religare Ltd.