Introduction

The V600® Disinfection Controller is the latest dedicated controller designed by Siemens Water Technologies, built on the depth of knowledge the company has acquired over many years yet incorporating advanced electronics and process control application features. The V600® Disinfection Controller is designed to optimise control for chlorination/dechlorination and chloramination control.

It is suitable for use with gaseous chlorinators/ sulphonators/ammoniators or liquids using dosing pumps for commercially produced sodium hypochlorite or that derived from Siemens Water Technologies OSEC® hypochlorite generation system, Sodium Bisuphite or Ammonium Sulphate solutions.

The V600® Disinfection Controller provides a comprehensive and unrivalled list of features for manual or automatic control of simple borehole applications through to complex treatment work applications.

CONTROL SYSTEM & MOUNTING OPTIONSThe V600® Disinfection Controller system has been designed with consideration to the practicalities of site installation and site operation. The system consists of a V600® Disinfection Controller, which may be installed at the most suitable site location and a Dosing Interface Unit, which ideally should be located adjacent to the dosing apparatus. (The maximum distance between the two devices is 1000m).

The V600® Disinfection Controller provides the output to the controlling devices, information gathering and dissemination point, and the location for RS232 and RS 485 communications.

The Dosing Interface Unit (DIU) provides the link and connection point for the dosing devices and system/operational features which interface with your bespoke dosing and operational control system.

MOUNTING OPTIONSBoth the V600® Disinfection Controller and Dosing Interface Unit (DIU) can be supplied in the following options; Wall mounted plastic enclosures to IP 67 Panel Mounting for use in customers panel. Enclosure to IP 65 Single or Double Wall mounted steel enclosures to IP 65

BenefitsBuilt in flexibilityChoice of 6 control optionsComprehensive list of featuresSimple to operateVariety of installation and mounting optionsCombined with the Wallace & Tiernan® ChemWeb-Server and OPC software, the system offers a wide range of remote communications options

V600® Disinfection Controller

s

Steel enclosure

Page 3: V600 Disinfection Controller System

Control optionsA choice of 6 control options to meet the most simple to the most demanding of applications:-

1. Manual controlEnables the operator to take full control and set the desired dose rate to the required value.

2. Control in response to flowA 0/4-20 mA signal or 0-10V signal from a flow transmitter is fed into the controller and the dosing equipment injects chemical proportional to the treated water flow.

An option to accept 2 flow signals is also available. The sum of the two signals is used for the calculation of the dose rate. If this option is selected the V600® Disinfection Controller will automatically detect the two flow signals and configure the flow control accordingly.

FLOW 1

COMMON FLOW

FLOW 2

To Flow Input 1on Controller

To Flow Input 2on Controller

To Flow Input 1on Controller

Common Flow Meter(Standard)

2 Flow Meters(Option)

1. Manual control screen

1. Manual control screen

FLOW

Flow Meter

Controller

DosingDevice(s)

Flow InputControl Output

Control in response to flow

Control in response to flow

Page 4: V600 Disinfection Controller System

3. Dual feed forward control (ratio control)Provides the opportunity to feed an input signal from an ammonia monitor or chlorine analyser to provide a ratio adjustment to increase or decrease the chlorine dose rate. The advantage of this input signal is that it initiates an immediate response to the dosage calculation without waiting for process times. The additional use of flow meter signal input is selectable.

This mode of control is used in the treatment of river water applications where a signal from an ammonia analyser will enable the controller to respond to varying ammonia levels that may necessitate significant adjustments to the chlorine dose rate. It may also be used in rechlorination schemes, enabling the controller to take into account existing varying chlorine levels arriving at the rechlorination point.

FLOW

Flow Meter

Controller

DosingDevice(s)

Flow Input

Control Output

Analyser Input 1

Analyser(Cl2 / NH3)

3. Dual feed forward control (ratio control)

4. Residual ControlIn this control mode, sample water is fed continuously to the analyser and transmitter which provides a 0/4-20 mA signal to the controller. The measured signal is compared with a pre-set residual and any deviation from the control set point initiates a proportional/integral correction to the chemical dosing device.

Residual control provides stable residual measurement where the quality of the water varies but the flow variations of the treated water are relatively stable.

The V600® Disinfection Controller will accept a 0/4-20mA signal from a Chlorine/Chlorine Dioxide/Ozone or DEOX/2000® analyser. (The DEOX/2000® analyser is a centre zero analyser for the continuous on - line measurement of dechlorination or deozonation processes. See publication SB.50.000.GE)

FLOW

Controller

DosingDevice(s)

Control Output

Analyser (Cl2 / ClO2 / Deox / O3)

Analyser 1

4. Residual control

Page 5: V600 Disinfection Controller System

5. Control with flow and residual control. (Compound loop)Provides stable residual control where both the water quality and the treated water flows are subject to variations. The inputs are the same for those specified for No. 4. Residual Control and flow meter input.

Therefore in this control mode, sample water is fed continuously to the analyser and transmitter which provides a 0/4-20 mA signal to the controller. The measured signal is then compared with a pre-set residual and any deviation from the control set point initiates a proportional/integral correction to the chemical dosing device.

FLOW

Flow Meter

Controller

DosingDevice(s)

Flow Input

Control Output

Analyser(Cl2 / ClO2 / Deox / O3)

Analyser 1

FLOW

Flow Meter

Controller

DosingDevice(s)

Flow Input

Control Output

Analyser (Cl2)

Analyser 1

Analyser 2

Analyser (Cl2)

Contact Tank

5. Control with flow and residual control (compound loop)

6. Control with flow and residual control with set point trimProvides primary control as per flow and residual control outlined previously but with the added advantage of secondary control from a second analyser to compensate for any additional chlorine demand that may occur after a period of time.

Typically used in applications where responsive control is required prior to the inlet to a chlorine contact tank but there maybe continuing demand within the contact tank. Considerations should be given to the duration of both process loops and Siemens Water Technologies would be pleased to offer advise in this application.

6. Control with flow and residual control with set point trim

Page 6: V600 Disinfection Controller System

V600® Disinfection Controller connection diagramsThe input/output connection to the V600® Disinfection Controller are shown below.

Feedback1K / 5K Pot.

Enable RelaySPCO

Inc / Dec Relay SPCO

mA Output0/4..20mA

On/Off RelaySPCO

Vent/ Booster Pump 1Relay SPCO

Fault Digital InputVolt Free Contact

Feedback1K / 5K Pot.

Enable RelaySPCO

Inc / Dec Relay SPCO

mA Output0/4..20mA

On/Off RelaySPCO

Vent / Booster Pump 2Relay SPCO

Fault Digital InputVolt Free Contact

Dosing Machine 1

Dosing Machine 2

Power 115V / 230V ACOR 24V DC

Power Supply

Dosing Interface Unit

Booster Pump FailVolt Free Contact

Page 7: V600 Disinfection Controller System

Power 115V / 230V ACOR 24V DC

GP Digital Input 1Volt Free Contact

GP Digital Input 2Volt Free Contact

mA Output 10/4..20mA

mA Output 20/4..20mA

Alarm Relay 1SPCO

Alarm Relay 2SPCO

Alarm Relay 3SPCO

Alarm Relay 4SPCO

RS485

RS232

Alarm Relays

mA Outputs

GP Digital Inputs

Communications

Power Supply

mA Output 30/4..20mA

Alarm Relay 5SPCO

Alarm Relay 6SPCO

Vent / Booster Pump 1 Relay SPCO

Booster Pump FailVolt Free Contact

Booster Pump /Vent Control

Vent / Booster Pump 2Relay SPCO

Controller

Flow 2 (Option)0/4..20mA or 0..10V

Analyser 10/4..20mA

Analyser 20/4..20mA

External Setpoint0/4..20mA

Analogue Inputs

Flow 0/4..20mA or 0..10V

V600® Disinfection Controller connection diagrams continued

Page 8: V600 Disinfection Controller System

System optionsCHOICE OF OPERATIONAL FEATURESA number of customer selectable options are built into the V600® Disinfection Controller system to enhance the operational efficiency of your dosing system.

Injector Vacuum Line Relief (gas feed devices)The use of remote injectors is desirable when there are long distances between the gas feed device and the injector point. It reduces the system process time, enabling quick responses to water quality or flow variations.

The use of the injector vacuum line relief option is desirable to prevent excess vacuum being formed that may occur in long injector vacuum lines.

Dosing Pump Vent OptionCertain liquids such as Sodium Hypochlorite, can gas lock dosing pumps at low flows. Whilst many dosing pumps can be fitted with gas venting devices, the V600® Disinfection Controller provides that added protection against this problem by providing the option to boost the pump speed at an adjustable frequency and duration or to open a solenoid valve and vent the gas back to the chemical storage tank.

Control of injector water supplyThe V600® Disinfection Controller will control the operational of water solenoid valve or injector booster pump.

V600® Disinfection Controller system - operationCHOICE OF SYSTEM OPTIONSThe V600® Disinfection Controller is designed to provide continuous operation and provide selectable features for that added degree of security. It enables the selection and full operational use of standby plant in one of the following selectable options.

Duty/Standby modeAuto duty cycling enables automatic changeover of the duty/standby dosing machines for periods varying from 1-90 days. Automatic changeover of the standby machine can be initiated by machine fault (loss of vacuum on gas machines) or low and high residuals. Automatic Return to the set duty machine once the fault is cleared.

Duty/AssistAuto duty cycling enables automatic changeover of the duty/assist dosing machines for periods varying from 1-90 days. In this mode the duty machine brings in the assist machine when the high control alarm point on the duty machine is reached.

At this point both machines run at 50% of the required dose until the combined output reaches 90% of the duty machine when it reverts solely to the duty machine.

Duty/DutyProvides live standby operation as each machines run at 50% of the required dose providing even wear of both machines. In the event of one machine failing, the other increases to provide the full dose. (Check that each machine is capable of providing the full dosage range required)

System FaultThe V600® Disinfection Controller system is designed to detect a fault in the hardware of V600® Disinfection Controller and the Dosing Interface Unit. It will also alarm if there is a fault with the connection between the two units.

Automatic Control Inhibit and Auto shutdownThe V600® controller has 2 Digital Inputs for used only with normally closed contacts. These can be assigned to:-

Digital Input 1Control inhibit, Alarm Inhibit, Control inhibit and output to 0%, Control inhibit and output to 50%, Control inhibit and output to 100%. This feature is used when residual inputs or flow measurement are unreliable for a short period of time such as during filter backwashing procedures.

Digital Input 2Shutdown, alarm inhibit.

LED Display

Main Screen

Page 9: V600 Disinfection Controller System

V600® Disinfection Controller - system featuresThe following two diagrams show the range of features built into the V600® Disinfection Controller system and indicate system potentials

Ch

oic

e o

f vo

ltag

es

23

0v

AC

, 1

15

v A

C,

24

v D

C

V6

00

®

Co

ntr

oll

er

Secu

rity

. 2

pas

swo

rds

&

3 le

vels

of

acce

ss

Stru

ctu

re m

enu

sy

stem

. On

ly it

ems

sele

cted

are

sh

own

w

hen

app

ropr

iate

. Ev

eryt

hin

g el

se is

h

idde

n

Men

u 1

fo

r co

mm

issi

on

ing.

Men

u 2

fo

r co

nfi

gura

tio

n. C

on

tro

l M

ode

en

able

s qu

ick

chan

ges

to o

pera

tio

nal

pa

ram

eter

s

Opt

ion

of

Term

inal

Box

to

acc

ept

stee

l wir

ed a

rmo

ure

d ca

blin

g.

Encl

osu

re t

o IP

55

Plas

tic

wal

l m

ou

nte

d En

clo

sure

to

IP 6

7

Mo

un

tin

g o

ptio

ns

Ch

oic

e o

f co

ntr

ol

opt

ion

s

Co

ntr

ol i

n r

espo

nse

to

flo

w0

/4-2

0m

A o

r 0

-10

v fr

om

fl

ow t

ran

smit

ter

2 F

low

met

er in

puts

an

d th

e su

m u

sed

for

calc

ula

tio

ns

(Au

to D

etec

tio

n -p

lug

&

play

pri

nci

ple)

Flo

w s

ign

als

disp

laye

d as

I/s

orM

LD,

m3

/h

Opt

ion

of

Do

ubl

e W

all M

ou

nte

d St

eel

encl

osu

re t

o a

ccep

t bo

th t

he

V6

00

& D

IU.

Suit

able

fo

r st

eel

wir

e ar

mo

ure

d ca

blin

g. E

ncl

osu

re

IP 6

5

Opt

ion

of

sin

gle

wal

l-mo

un

ted

stee

l en

clo

sure

to

acc

ept

eith

er V

60

0 o

r D

IU

and

suit

able

fo

r st

eel w

ire

arm

ou

red

cabl

ing.

En

clo

sure

to

IP 6

5

Pan

el m

ou

nte

d o

ptio

n f

or

mo

un

tin

g in

cu

sto

mer

’s o

wn

pa

nel

Encl

osu

re t

o IP

65

Pan

el m

ou

nte

dD

ual

fee

d fo

rwar

d (r

atio

co

ntr

ol)

0/4

-20

mA

inpu

t fr

om

A

mm

on

ia o

r C

hlo

rin

e m

on

ito

r an

d 0

/4-2

0m

A o

r 0

-10

V f

rom

flo

w

tran

smit

ter

Flo

w s

ign

als

disp

laye

d as

I/s

orM

LD, o

rM3

/h

Stat

us

disp

lay

opt

ion

s in

dica

tin

g va

lues

, bar

gra

ph,

grap

hs,

tex

t, c

on

tro

l alg

ori

thm

s an

d do

sin

g m

ach

ine

info

rmat

ion

Show

n o

n a

24

0

x 6

4 b

ack-

lit L

CD

di

spla

y

Info

rmat

ion

Sin

gle

feed

back

co

ntr

ol

0/4

-20

mA

inpu

t fr

om

ch

lori

ne

/ ch

lori

ne

diox

ide,

Deo

x o

r O

zon

e m

on

ito

r

Opt

ion

of

mov

ing

alar

ms

low

& h

igh

al

arm

s 1

in r

elat

ion

to

va

ryin

g se

t po

ints

2 lo

w r

esid

ual

ala

rms

and

2 h

igh

re

sidu

al a

larm

s

Indi

cati

on

of

25

al

arm

co

ndi

tio

ns.

(T

he

last

24

ala

rms

are

show

n in

th

e al

arm

his

tory

log)

Co

ntr

ol w

ith

flo

w &

re

sidu

al c

on

tro

l

0/4

-20

mA

or

0-1

0V

fro

m

flow

tra

nsm

itte

r +

0/4

-20

m

A r

esid

ual

inpu

t si

gnal

Pro

cess

tim

e pr

opo

rtio

nal

to

flo

w(n

ew f

orm

ula

)

Flo

w s

ign

als

disp

laye

d as

I/s

orM

LD,

orM

3/h

9 L

ED D

ispl

ay, P

ow

er

on

/off

, Sys

tem

fau

lt,

Alr

am-W

hen

an

y al

arm

act

ivat

ed.

Rela

y 1

, 2, 3

, 4, 5

, 6

acti

vate

d

Rela

y 1

, can

be

use

d fo

r w

atch

dog

ope

rati

on

No

rmal

ly o

pen

or

clo

sed

ope

rati

on

. Lat

ched

wit

h

ackn

owle

dge.

Un

latc

hed

or

latc

hed

wit

h r

eset

ope

rati

on

.

6 a

larm

rel

ays.

Ea

ch w

ith

fro

nt

pan

el

stat

us

LED

. A

ll al

arm

s ar

e m

appa

ble

Resi

dual

set

-po

int

trim

Seco

nda

ry c

on

tro

l fro

m

0/4

-20

mA

add

itio

nal

cel

l in

put

sign

al

Pro

cess

tim

e pr

opo

rtio

nal

to

flo

w(n

ew f

orm

ula

)

Flo

w s

ign

als

disp

laye

d as

I/s

orM

LD, o

r

M3

/h

Ass

ign

able

0-2

0m

A

or

4-2

0m

A s

ign

als

Flow

, flo

w 2

, flo

w (

tota

l),

anal

yser

1, a

nal

yser

2, s

et

poin

t, c

on

tro

l ou

tpu

t, c

hlo

rin

e de

man

d. (

Ava

ilabi

lity

depe

nds

o

n c

on

tro

l sys

tem

sel

ecte

d)

3 a

nal

ogu

e o

utp

uts

Man

ual

co

ntr

ol

Dow

nlo

ad d

ata

to

lapt

op

Upl

oad

pro

gram

into

fla

sh

mem

ory

Dat

a lo

ggin

g

Syst

em

Co

ntr

ol

Oti

on

s

2 D

igit

al In

puts

Inpu

t 1

map

ped

to C

on

tro

l In

hib

it, A

larm

Inh

ibit

an

d o

utp

ut

to 0

%, o

r 5

0%

o

r 1

00

%

Inpu

t 2

map

ped

to

shu

tdow

n o

r al

arm

inh

ibit

New

ala

rms

inh

ibit

edH

ardw

are

fau

lt C

AN

bus

erro

r,

Do

sin

g in

terf

ace

Fau

ltSy

stem

Fau

ltA

uto

sh

ut

dow

nlo

w f

low

ala

rm, o

r st

andb

y m

ach

ine

fau

lure

, or

anal

ogu

e si

gnal

fai

lure

Opt

ion

s to

sel

ect

shu

t do

wn

or

man

ual

se

ttin

g

Wh

en t

he

4-2

0m

A s

ign

al f

alls

be

low

3.1

mA

Si

gnal

fai

lure

al

arm

sSi

gnal

Fai

lure

Ala

rms

Upo

n f

ailu

re o

f th

e fl

ow &

A

nal

yser

sig

nal

s, c

on

tro

l m

aybe

set

to

sh

utd

own

or

to a

def

ault

val

ue

Failu

re o

f A

nal

yser

2 in

Set

Po

int

Trim

Mo

de c

an b

e se

t to

def

ual

t to

pri

mar

y lo

op

con

tro

l on

ly

Dis

play

s al

l an

alo

gue

inpu

t st

atu

s, C

on

tro

l I/O

sta

t

Co

mpr

ehen

sive

ra

nge

of

diag

no

stic

s

Co

mm

sD

own

load

dat

a to

la

pto

pU

plo

ad p

rogr

am in

to f

lash

m

emo

ryRS

23

2

Ch

emw

eb S

erve

rO

PC C

om

plia

nt

SCA

DA

Supp

ort

ing

Prot

oco

lRS

48

5

Page 10: V600 Disinfection Controller System

V600® Disinfection Controller system features - dosing interface unit

This

info

rmat

ion

is a

lso

disp

laye

d on

the

V60

0 co

ntro

ller

Indi

cate

s po

wer

O

n/O

ff, S

yste

m F

ault

, Co

mm

unic

atio

ns in

dica

tion

, D

uty

mac

hine

1 s

elec

ted,

op

erat

iona

l, m

achi

ne fa

iled.

D

uty

mac

hine

2 s

elec

ted,

op

erat

iona

l, m

achi

ne fa

iled

9 LE

D’s

di

spla

ys

& a

larm

m

essa

ges

Info

rmat

ion

Do

sin

g In

terf

ace

Un

itCh

oice

of

cont

rolli

ng

devi

ces

Dos

ing

Pum

ps

Incr

ease

/ D

ecre

ase

(pos

itio

ner)

1k

or 5

k fe

edba

ck

pote

ntio

met

ers

whi

ch a

uto

dete

ctio

n

Rela

y co

ntac

t ra

nge

of

30-1

22 P

PM

0/4-

20m

A

Incr

ease

/ D

ecre

ase

+ m

A o

utpu

t co

ntro

l (e

ven

cont

rol

on e

ach)

This

info

rmat

ion

is a

lso

disp

laye

d on

the

V60

0 co

ntro

ller

Dos

ing

Stat

us

is d

ispl

ayed

on

LCD

Ope

rati

onal

Fe

atur

es

Gas

Fee

d -

Chlo

rina

tors

,Su

lpho

nato

rs,

Am

mon

iato

rs

Incr

ease

/ D

ecre

ase

(pos

itio

ner)

1k

or 5

k fe

edba

ck

pote

ntio

met

ers

whi

ch a

uto

dete

ctio

n

0/4-

20m

A

Incr

ease

/ D

ecre

ase

+ m

A

outp

ut c

ontr

ol

(eve

n co

ntro

l on

eac

h)

If e

ithe

r m

achi

ne

fails

the

oth

er

incr

ease

s to

pr

ovid

e th

e fu

ll do

se.

Both

mac

hine

s op

erat

e at

50

% e

ach

prov

idin

g liv

e st

andb

y op

erat

ion.

(Ch

eck

rang

e of

eac

h m

achi

ne)

Dut

y / D

uty

Syst

em

Opt

ions

Gas

Fee

d Sy

stem

sIn

ject

vac

uum

line

re

lief

Mac

hine

s ru

n at

50

% e

ach

Dut

y ru

ns u

ntil

high

con

trol

al

arm

is re

ache

d

Aut

o du

ty c

yclin

g fo

r 1

- 90

days

. Ind

epen

dant

ly

set

to p

rovi

de e

ven

or u

neve

n w

ear.

Set

chan

geov

er t

ime

Dut

y A

ssis

tLi

quid

Fee

d Sy

stem

sD

osin

g pu

mp

vent

op

tion

Ope

n so

leno

id v

alve

and

/ or

Bo

ost

pum

p to

max

spe

ed

Adj

usta

ble

freq

uenc

y an

d du

rati

on

sett

ings

Whe

n ou

tput

re

ache

s 90

%

of h

igh

cont

rol

poin

t, s

wit

ches

ba

ck to

dut

y m

achi

ne

Bum

perl

ess

tran

sfer

on

pos

itio

ner

cont

rol e

limin

ates

do

si n

g va

riat

ions

du

e to

tim

e de

lays

in

pos

itio

ner

oper

atio

n

Chan

geov

er v

ia

mac

hine

faul

t (l

oss

of v

acuu

m),

lo

w o

r hi

gh

resi

dual

Aut

o D

uty

Cycl

ing

for

1 - 9

0 da

ys. I

ndep

enda

ntly

se

t to

pro

vide

eve

n or

une

ven

wea

r. Se

t ch

ange

over

tim

e

Dut

y /

Stan

dby

Boos

ter

pum

p ch

ange

over

(v

ia V

600

cont

rolle

r or

Dos

ing

Inte

rfac

e U

nit)

Will

cha

ngeo

ver

2 bo

oste

r pu

mps

. Si

ngle

rela

y co

ntro

l fo

r ea

ch b

oost

er

pum

p, fr

om

com

mon

dig

ital

in

put

failu

re

Aut

o Cy

clin

g w

ith

dosi

ng

mac

hine

s

Faul

t cl

eare

d re

turn

to d

uty

mac

hine

New

Ala

rms

Inhi

bite

dH

ardw

are

faul

t, C

AD

bus

erro

r, D

osin

g In

terf

ace

Faul

tSy

stem

Fau

lt

Cont

rol o

f in

ject

or

oper

atin

g w

ater

Opt

iona

l wat

er

sole

noid

val

ve o

r bo

oste

r pu

mp

Mac

hine

1 F

ail,

Mac

hine

2

Fail,

Boo

ster

Pum

p Fa

il3

Dig

ital

In

puts

V6

00

®

Co

ntr

oll

er

Page 11: V600 Disinfection Controller System

V600® Disinfection Controller system - communications and security

USER OPERATING AND SYSTEM INFORMATIONThe V600® Disinfection Controller is designed to be easy to operate and a 240 x 64 LCD backlight display provides the user with access to system and process information.

User Friendly OperationA user friendly menu structure is provided to enable easy and quick access to the information required. Access is via 3 ‘soft keys. Up, down, left, right arrows and enter and escape buttons provide easy navigation to the information required.

COMMUNICATIONS & SYSTEM SECURITYRS232 CommunicationsThe RS232 interface enables a standard laptop computer to be connected to the V600® Disinfection Controller. This interface may then be used to upload the main program into flash memory, or download the data logging data from the controller where it can be imported into standard software programs such as Microsoft Excel where the information can be displayed in any format.

RS485 CommunicationsAn RS 485 connection will allow remote access and transmission of data. When used in conjunction with the Wallace & Tiernan® ChemWeb-Server it offers an flexible choice of remote communications via modem/TCP IP and to OPC complaint SCADA systems using the Wallace & Tiernan® OPC software.

All Operation and Commissioning parameters will be accessible using the RS 485 interface and some additional variables to give status/alarm information.

SecurityInbuilt security system to the V600® Disinfection Controller is provided by 2 passwords and 3 levels of access.1. System Password – allows access to the commissioning and operational menus2. Menu 2 password – allows access to the operational parameters only3. Manual/Auto parameters – Open access. The 3 levels of access correspond to the 3 menu systems using the soft keys

V600® Disinfection Controller key pad

The Status Display options include the display of residual values, bar graphs, trending graphs, text, and dosing machine information.

Built in Data Logging can provide the user with one of the following selectable values to record.

Analyser 1Analyser 2Flow (total)Control Output (total)

The data logging is fixed to 1 month of data at 2 minute sampling times. The data is displayed in 8 hr periods providing the user with clearly visible information.

The data may be viewed as a graph on the controller display or downloaded to a laptop via the RS 232 interface.

Alarm information provides indication of 25 alarm conditions. The last 24 alarms are shown in an alarm history log. There are 2 low residual alarms and 2 high residual alarms.

The V600® Disinfection Controller also provides a selectable option to move low and high alarm 1 in relation to varying set points.

RS232 connection to V600® Disinfection Controller

Page 12: V600 Disinfection Controller System

Siemens Water Technologies

Germany+49 8221 [email protected]

United Kingdom+44 1732 [email protected]

USA+1 856 507 [email protected]

© 2008 Siemens Water Technologies Corp.

Literature Number: WT.040.700.000.IE.PS.0309

Subject to change without prior notice.

V600, DEOX/2000, OSEC and Wallace & Tiernan are trademarks of

Siemens, its subsidiaries or affiliates.

The information provided in this brochure contains merely general

descriptions of characteristics of performance which in case of actual

use do not always apply as described or which may change as a result

of further development of the products. An obligation to provide the

respective characteristics shall only exist if expressly agreed in the terms

of contract.

www.siemens.co.uk/water

Technical dataThe following specifications apply to both the Controller and the Dosing Interface Unit (DIU).

ELECTRICALMains Supply (Voltage specified with order): 115V AC ±10%, 50-60Hz, 30VA230V AC ±10%, 50-60Hz, 30VA24V DC, 30W

Fuses F1 & F2: 115V / 230V AC 1A(T), 250V, TR524V DC 2.5A(T), 250V, TR5

Fuses FS1 & FS2 (Steel enclosure only): All voltages -3.15A(T), 250V, 5 x 20mm

Safety: BS EN 61010Installation category II

EMC:BS EN 61326

PHYSICAL (H x W x D)Plastic Enclosure Size: 273 x 316 x 167 Weight: 5 Kg

Single Steel Enclosure Size: 600 x 435 x 279Weight: 19 Kg

Double Steel Enclosure Size: 600 x 656 x 279Weight: 29 Kg

Panel Mounting - Bezel Size: 210 x 280 x 25 Weight: 1 Kg

- Base Unit Size: 273 x 316 x 115Weight: 4 Kg

Terminal Box Size: 200 x 300 x 120 Weight: 4 Kg

ENVIRONMENTALTemperature Range: Operation 0…50°C (max 90% RH, non condensing)Storage –20…70°C

IP / NEMA Rating:Plastic enclosure: IP 67 / NEMA 4XSteel enclosure: IP 65 / NEMA 13Panel Mount Bezel: IP 65 / NEMA 13Terminal box: IP 55 / NEMA 12

INPUTS / OUTPUTS:Digital Inputs:For use with volt free contacts onlyIsolated voltage supplied by controller (15V DC nominal)

Relay Outputs:Resistive rating:5A, 250V AC, 1250VA max5A, 220V DC, 150W max

UL/CSA Rating: 5A 1/6 HP 125,250 V AC5A 30V DC 30W max1A 30V DC to 0.24A 125V DCSuppression with Schottky diodes

Analogue Inputs: 0…20 / 4…20mA and 0…5VInput impedance 47Ω for mA signalsAccuracy 0.5% FSGalvanically isolated from earth to 50V

Analogue Outputs: 0…20 / 4..20mAAccuracy 0.5% Full ScaleMaximum load 400Ω or 1KΩ (switch selectable)Galvanically isolated from earth to 50V

Feedback Signals:1KΩ or 5KΩ potentiometer (automatic detection)Not isolated

RS232 Interface: Supports Siemens Water Technologies protocolNot isolated

RS485 Interface: Supports Siemens Water Technologies protocolGalvanically isolated from earth to 50V