Applied Technology Development Division UV/VIS SPECTROPHOTOMETER MECASYS, Co., Ltd.

Transcript of UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance...

Page 1: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

UV/VIS SPECTROPHOTOMETER

MECASYS, Co., Ltd.

Page 2: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

What is UV/VIS?What is UV/VIS?

� Electromagnetic Spectrum

X-Rays UV VIS NIR IR Micro wave

200nm 400nm 700nm 2500nm 500 µm

E0

E1

∆E= E1 -E0

Molecular rotationMicro Wave

Molecular vibrationIR

Outer ElectronUV/VIS

Inner ElectronX-Rays

Type of TransitionType of Radiation

Page 3: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Principle of UV/VIS SPECTROPHOTOMETERPrinciple of UV/VIS SPECTROPHOTOMETER

� Principle

• Light Intensity Change : By Absorbance or Transmittance

•Quantity : Using Absorbance

λ

I

λ

I

Sample

Page 4: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

LambertLambert--BeerBeer’’s Laws Law

� Transmittance, t

• Ι t / Ι 0 = t

� Absorbance, A

• log 1 / t = A

� Lambert – Beer’s Law

• A = abC, A = ξbC

� b : Sample Path length, C : Sample Concentration,

� a : Absorbance Constant

� ξ : Molecular Absorbance Constant

I0 It

Sample (Conc.C)

Path length b

Page 5: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Standard Curve, Calibration curveStandard Curve, Calibration curve

� Lambert-Beer’s Law

• A=abC

• Concentration determination of Sample

Conc.Conc.Conc.Conc.

AbsorbanceAbsorbanceAbsorbanceAbsorbance

Standard Curve

� Standard Curve

• Standard Samples

• Proportional constant

• Absorbance measurement of

Samples

Concentration

determination of Sample

Page 6: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

� Deviations in absorptivity coefficients

• at high concentrations (>0.01M) due to electrostatic interactions

between molecules in close proximity

� Scattering of light

• due to particulates in the sample

� Fluorescence or Phosphorescence of the sample

� Changes in refractive index at high analysis concentration

� Shifts in chemical equilibria as a function of concentration

� Stray light

Limitation of LambertLimitation of Lambert--BearBear’’s Laws Law

Page 7: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Application of Application of UV/VIS SpectrophotometerUV/VIS Spectrophotometer

Quantity analysisQuantity analysisQuantity analysisQuantity analysis

(Using a Spectrum)

Quality analysisQuality analysisQuality analysisQuality analysis

(Standard Curve)

Chemical analysis

Food safety analysis

Blood analysis

DNA/RNA Conc. analysis

Residual Pesticide analysis

Residual chlorine analysis

Page 8: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Functions of SpectrophotometerFunctions of Spectrophotometer

� ABS/%T/CONC

• Measuring Absorbance/Transmittance/Concentration

� Survey Scan

• Measuring Absorbance(Transmittance) Spectrum

� Standard Curve (Calibration Curve)

• Determining Proportional Constant between ABS and Conc

� Simple Kinetics

• Measuring Time Dependent ABS Variation

� ABS Ratio

• Measuring ABS Ratio (DNA/Protein)

Page 9: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Structure of UV/Visible SpectrophotometerStructure of UV/Visible Spectrophotometer

Light

source

Lens

or

Mirror

MonochromatorDetector

Sample

Page 10: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

ExampleExample

Spectrophotometer

Light

SourceMonochromator

Sample

Compartment

Sensing

Part

Page 11: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Light SourceLight Source

� Deuterium Arc Lamp

• UV Region

•Wavelength Range :

190~420nm

� Tungsten LampTungsten LampTungsten LampTungsten Lamp

• Wavelength Range : Part of the UV and the whole of the Visible Wavelength Range : Part of the UV and the whole of the Visible Wavelength Range : Part of the UV and the whole of the Visible Wavelength Range : Part of the UV and the whole of the Visible

range range range range ((((약약약약350 ~ 2,500350 ~ 2,500350 ~ 2,500350 ~ 2,500nm)nm)nm)nm)

� Xenon Lamp

• Wavelength Range : 190~800nm

Tungsten LampTungsten LampTungsten LampTungsten Lamp

Page 12: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

MonochromatorMonochromator

� Accepts polychromatic input light from a lamp and outputs

monochromatic light

� Components : Entrance slit, Dispersion device, Exit slit

Entrance slit

Exit slit

Dispersion device

White Light

Monochromatic

Light

Page 13: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Dispersion Device (I)Dispersion Device (I)

� Filter

• Absorptive Filter

• Interference Filter : FWHM 10

– 50 nm

Page 14: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Dispersion Device (II)Dispersion Device (II)

� Prism

Page 15: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Dispersion Device (III)Dispersion Device (III)

� Grating

• Plane Grating, Concave Grating

Page 16: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Monochromator structureMonochromator structure

Littrow mounting

Slit

Grating

Mirror

Czerny-turner mounting

Mirror Mirror

Grating Entrance

slit

Exit slit

Page 17: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Monochromator Design Monochromator Design

� Modified Czerny-Turner Type

- Focal Length : 200mm

- Sine Drive Mechanism

- 1nm Bandwidth

� Czerny-Turner Type

- Focal Length : 100mm

- Direct Drive Mechanism

- 4nm Bandwidth

Page 18: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Absorption CellAbsorption Cell

� Receptacle for Sample.

� Material

• Quartz, Fused Silica(Using over

190nm)

• Glass (Using Over Visible Region)

Page 19: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

DetectorDetector

� Photodiode

� Photodiode & CCD

� PMT(Photo Multiplier tube)

Page 20: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Type of Type of UV/VIS SpectrophotometerUV/VIS Spectrophotometer

• Single Beam Type • Double Beam Type

Page 21: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Type of Type of UV/VIS SpectrophotometerUV/VIS Spectrophotometer

Grating

Light Sample

Detector

� Scanning type : � PDA type :

Grating

Light

Multi channel

DetectorSample

Page 22: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Specifications ExampleSpecifications Example

� Spectral Bandwidth : 1 nm

� Wavelength

• Range : 190 – 1100 nm

• Accuracy : ± 0.5 nm

• Reproducibility : ± 0.1 nm

� Photometric

• Range : -0.3 – 3.0 Abs (0 – 125 %T)

• Accuracy : < 1% at 1 Abs

• Reproducibility : ± 0.001Abs

� Stray Light : <0.05% at 220 nm & 340 nm

Page 23: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Accessories (I)Accessories (I)

� Multi-Cell Holder � Sipper System

Page 24: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division

Accessories (II)Accessories (II)

� Micro Volume Cell � Round Cell � Long Path Cell

� Temperature Cell Holder, Film/Thin-Layer Holder

Page 25: UV VIS Spectrophotometer ̷ .ppt [ б ]) · Czerny-turner mounting Mirror Mirror Grating Entrance slit Exit slit. Applied Technology Development Division Monochromator Design Modified

Applied Technology Development Division