Contents
Chapter 1 1 Intended Use, Contraindications and General Precautions
Chapter 2 3 Description of the Inogen One® G4 Oxygen Concentrator 3 ImportantPartsoftheInogenOne®G4OxygenConcentrator 4 UserInterfaces 5 Input/OutputConnections 6 PowerSupplyOptions 8 InogenOne®G4Accessories
Chapter 3 11 Operating Instructions 11 GeneralInstructions 15 AdditionalOperatingInstructions 19 BatteryOperatingInstructions 20 BatteryCareandMaintenanceChapter 4 21 Inogen One® G4 Oxygen Concentrator Audible and Visible Signals (including Alerts)
Chapter 5 29 Troubleshooting Chapter 6 31 Cleaning, Care and Maintenance 31 CannulaReplacement 31 CaseCleaning 31 FilterCleaningandReplacement 33 InogenOne®G4ColumnChangeProcedure 36 OtherServiceandMaintenance 36 DisposalofEquipmentandAccessories
Chapter 7 37 Symbols Used on Concentrator and Accessories
Chapter 8 39 Inogen One® G4 System Specifications
English
1
1 Intended Use, Contraindications and General Precautions
IntendedUse,ContraindicationsandGeneralPrecautionsChapter1
Intended Use
TheInogenOne®G4OxygenConcentratorisusedonaprescriptivebasisbypatientsrequiringsupplementaloxygen.Itsuppliesahighconcentrationofoxygenandisusedwithanasalcannulatochanneloxygenfromtheconcentratortothepatient.TheInogenOne®G4maybeusedinhome,institution,vehicleandvariousmobileenvironments.
CAUTION USAFederallawrestrictsthisdevicetosalebyorontheorderofaphysician.Mayalsobeapplicableinothercountries.
CAUTION Nasalcannulashouldberatedfor3litersperminute(e.g.SalterLabs16SOFT)toensureproperpatientusageandoxygendelivery.
WARNING Availabilityofanalternatesourceofoxygenisrecommendedincaseofpoweroutageormechanicalfailure.Consultyourequipmentproviderfortypeofback-upsystemrecommended.
CAUTION Itistheresponsibilityofthepatienttomakeback-uparrangementsforalternativeoxygensupplywhentraveling;Inogenassumesnoliabilityforpersonschoosingnottoadheretomanufacturerrecommendations.
TheexpectedlifefortheInogenOne®G4OxygenSystemis5years,withtheexceptionofthesievebeds(metalcolumns)whichhaveanexpectedlifeof1yearandthebatteries,whichhaveanexpectedlifeof500fullcharge/dischargecycles.
! C
English
2
Contraindications
WARNING ThisdeviceisNOTINTENDEDtobelifesustainingorlifesupporting. CAUTION Undercertaincircumstances,theuseofnon-prescribedoxygentherapy
canbehazardous.Thisdeviceshouldbeusedonlywhenprescribedbyaphysician.
CAUTION Additionalmonitoringorattentionmayberequiredforpatientsusingthisdevicewhoareunabletohearorseealertsorcommunicatediscomfort.Ifthepatientshowsanysignsofdiscomfort,aphysicianshouldbeconsultedimmediately.
CAUTION TheInogenOne®G4isnotdesignedorspecifiedtobeusedinconjunctionwithahumidifier,nebulizerorconnectedwithanyotherequipment.Useofthisdevicewithahumidifier,nebulizerorconnectedwithanyotherequipmentmayimpairperformanceand/ordamagetheequipment.DonotmodifytheInogenOne®G4Concentrator.Anymodificationsperformedontheequipmentmayimpairperformanceordamageequipmentandwillvoidyourwarranty.
! C
! C
! C
General Precautions
WARNING Thedeviceproducesenrichedoxygengas,whichacceleratescombustion.DoNOTALLOWSMOKINGOROPENFLAMESwithin10feetofthisdevicewhileinuse.
WARNING DonotsubmersetheInogenOne®G4oranyoftheaccessoriesinliquid.Donotexposetowaterorprecipitation.Donotoperateinexposedrain.Thiscouldleadtoelectricalshockand/ordamage.
CAUTION Donotuseoil,grease,orpetroleum-basedproductsonorneartheInogen
One®G4.
CAUTION NeverleavetheInogenOne®G4inanenvironmentwhichcanreachhightemperatures,suchasanunoccupiedcarinhightemperatureenvironments.Thiscoulddamagethedevice.
3
Important Parts of the Inogen One® G4 Oxygen Concentrator
Description of the Inogen One® G4 Oxygen Concentrator2
On/Off Button
Breath Detection Light
*Display
Audible Alert Button Alert/Alert Light
Flow Control
DescriptionoftheInogenOne®G4OxygenConcentratorChapter2
English
DescriptionoftheInogenOne®G4OxygenConcentratorChapter2
*Actualdisplayappearancemayvary.
4
User Controls
ON / OFF ButtonPressoncetoturn“ON”;Pressandholdforonesecondtoturn“OFF”.
Audible Alert Button PressingthisbuttonwilltoggletheInogenOne®G4’sbreathdetectionaudiblealertonandoff.
BreathDetectionAlertMode.TheInogenOne®G4willalertwithaudibleandvisualsignalsfor“nobreathdetected”whenthismodeisenabledandnobreathhasbeendetectedfor60seconds.
At60seconds,thedevicewillenterintoautopulsemodeandonceanotherbreathisdetected,thedevicewillexitautopulsemodeanddelivernormallyoninspiration.Thedisplay’smodeindicationareawillshowabellicon,flashingyellowlightanddisplaymessagewhenthealertisenabled.
Ifpowerislost,thebreathdetectionaudiblealertremainssetintheuserpreferredmode.Flow Setting Control Buttons Usethe–or+flowsettingcontrolbuttonstoselectthesettingasshownonthedisplay.Therearethreesettings,from1to3.
User Interfaces
DisplayThisscreendisplaysinformationregardingflowsetting,powerstatus,batterylifeanderrors.IfyouwouldliketochangethelanguageontheInogenLCDScreencontactyourEquipmentProviderforinstructions.Actualdisplayappearancemayvary.
5DescriptionoftheInogenOne®G4OxygenConcentratorChapter2
User Interfaces (continued)
Indicator LightsAgreenlightindicatesbreathdetection.Ayellowlightindicateseitherachangeinoperatingstatusoraconditionthatmayneedresponse(alert).Aflashinglightishigherprioritythannon-flashing.
Audible SignalsAnaudiblesignal(beep)indicateseitherachangeinoperatingstatusoraconditionthatmayneedresponse(alert).Morefrequentbeepsindicatehigherpriorityconditions.
BacklightAbacklightwillilluminatethescreenfor15secondswhentheon/offbuttonisbrieflypressed.
Input / Output Connections
Particle FilterThefiltermustbeinplaceattheintakeendoftheconcentratorduringoperationtokeepinputairclean.
Cannula Nozzle FittingThenasalcannulaconnectstothisnozzleforInogenOne®G4outputofoxygenatedair.
DC Power InConnectionforexternalpowerfromtheACpowersupplyorDCpowercable.
USB PortForserviceuseonly.
English
6
Power Supply Options
Single and Double Rechargeable Lithium Ion BatteriesThebatterywillpowertheInogenOne®G4withoutconnectiontoanexternalpowersource.Whenfullycharged,asinglebatterywillprovideupto2.7hoursofoperation;adoublebatterywillprovideupto5hoursofoperation.ThebatteryrechargeswhenproperlyinstalledintheInogenOne®G4andtheconcentratorisconnectedtoACorDCpower.Rechargingtimeisupto3hoursforasinglebatteryand5hoursforadoublebattery.Seeinformationinthe“BatteryCareandMaintenance”section.
Power SupplyOverviewTheInogenOne®G4ACpowersupply(BA-401)isusedtopowertheInogenOne®G4concentratorfromanACpowersource.
DescriptionTheInogenOne®G4ACpowersupplyisspecificallydesignedforusewiththeInogenOne®G4OxygenConcentrator(IO-400).TheACpowersupplyprovidestheprecisecurrentandvoltagerequiredtosafelypowertheInogenOne®G4andisdesignedtooperatefromspecifiedACpowersources.WhenusedwithACpowersources,thepowersupplyautomaticallyadaptstoinputvoltagesfrom100Vto240V(50-60HZ)permittingusewithmostpowersourcesthroughouttheworld.
TheACpowersupplywillchargetheInogenOne®G4BatterieswhenusedwithACinputpower.Duetoaircraftpowerlimitations,theACPowerSupplycannotbeusedtochargetheInogenOne®G4Batterywhenusedonanaircraft.TheACpowersupplyisusedwiththefollowingcomponents:
•PowersupplywithattachedpoweroutputcabletoconnecttotheInogenOne®G4•ACpowerinputcable
TheDCpowercable(BA-306)isspecificallydesignedforusewiththeInogenOne®G4OxygenConcentrator(IO-400).TheDCpowerinputcableconnectsdirectlytotheautomobilecigarettelighterorauxiliaryDCpowersupply.RefertoChapter3foradditionaloperatinginstructionsforusewithDCpowersource.
7DescriptionoftheInogenOne®G4OxygenConcentratorChapter2
English
*Actualproductappearancemayvary.
AC Power Plug*
Power Output Plug (Plug into Concentrator)
LED Indicator Light*AC Input
InogenOneACPowerSupplyModel#BA-401
InogenOneG4DCPowerCableModel#BA-306
DC cigarette lighter Power Plug for use in Automobile/RV/Boat/Aircraft
Power Output Plug (Plug into Concentrator)
WARNING Donotusepowersuppliesorpowercablesotherthanthosespecifiedinthisusermanual.Theuseofnon-specifiedpowersuppliesorpowercablesmaycreateasafetyhazardand/orimpairequipmentperformance.Donotwrapcordsaroundpowersupplyforstorage.Donotdrive,dragorplaceobjectsovercord.Doingsomayleadtodamagedcordsandafailuretoprovidepowertotheconcentrator.Toavoiddangerofchokingorstrangulationhazard,keepcordsawayfromchildrenandpets.
! C
8
Inogen One® G4 Accessories
WARNING Donotusepowersupplies/adaptersoraccessoriesotherthanthosespecified inthisusermanual.Theuseofnon-specifiedaccessoriesmaycreateasafety hazardand/orimpairequipmentperformance.
Nasal CannulaAnasalcannulamustbeusedwiththeInogenOne®G4toprovideoxygenfromtheconcentrator.Asinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.
WARNING Toavoiddangerofchokingandstrangulationhazard,keeptubingawayfromchildrenandpets.
NOTE Increasingthecannulalengthmayreducetheperceivednoiseduringoxygenbolusdelivery.
NOTE Whenusingacannula25ft.inlengthwiththeInogenOne®G4,anincreaseinflowsettingmayberequired.
Inogen One G4 Carry Strap (CA-401)Thecarrystrapisdesignedtoattachdirectlytotheconcentrator.Thereisametalslideforeaseoflengthadjustments,ashoulderpadforcomfortandasewnonvelcrotabtowrapupthatextracannulatubing.
! C
! C
9
Carry Bag (CA-400)TheCarryBagprovidesaprotectivecoverwithahandleandshoulderstraptoenableyoutocarrytheInogenOne®G4.TheInogenOne®G4canbeoperatedusingbatterypowerduringtransportwiththeCarryBag.
Inogen One® G4 Optional Accessories
External Battery Charger (BA-403)
TheInogenOne®G4externalbatterychargerwillchargetheInogenOne®G4singleanddoublebatteries.
1.PlugtheExternalBatteryChargerACpowersupplycordintoanelectricaloutlet.
2.PlugtheExternalBatteryChargerACpowersupplyintothebatterycharger.
3.SlideyourchargerontotheInogenOneG4Battery byclickingandlockingintothecharger.4.Whenthebatteryisinthecorrectposition,asolid redlightwillindicatethatthebatteryischarging.5.Whenthegreenlightilluminates,thebatteryis fullycharged.
CAUTION AvoidtouchingtherecessedelectricalcontactsoftheExternalBatteryCharger;damagetocontactsmayaffectchargeroperation.
NOTE Thesecontactsarenotpoweredunlessabatteryisinplaceandcharging.
NOTE TocompletelyremovepowerfromtheExternalBatteryCharger, removetheplug.
DescriptionoftheInogenOne®G4OxygenConcentratorChapter2
English
11OperatingInstructionsChapter3
English
Operating Instructions3
! C
General Instructions
1. Place the Inogen One® G4 in a well ventilated location.Airintakeandexhaustmusthaveclearaccess.LocatetheInogenOne®G4insuchawaythatanyauditoryalertsmaybeheard.AlwaysoperatetheInogenOne® G4inanuprightposition(seeimageforproperorientation).
WARNING AvoiduseoftheInogenOne®G4inpresenceofpollutants,smokeorfumes.DonotusetheInogenOne®G4inpresenceofflammableanesthetics,cleaningagentsorotherchemicalvapors.
CAUTION Donotobstructairintakeorexhaustwhenoperatingthedevice.
Blockageofaircirculationorproximitytoaheatsourcemayleadtointernalheatbuildupandshutdownordamagetotheconcentrator.
CAUTION TheInogenOne®G4Concentratorisdesignedforcontinuoususe.Foroptimalsievebed(columns)life,theproductshouldbeusedfrequently.
2. Ensure particle filter is in place.
CAUTION DonotoperatetheInogenOne®G4withouttheparticlefilterinplace.Particlesdrawnintothesystemmaydamagetheequipment.
Intake
ExhaustExhaust
ThissideUP
12
3. Install the battery. InserttheInogenOne®G4batterybyslidingbatteryintoplaceuntilthelatchreturnstotheupperposition.
CAUTION TheInogenOne®G4batteryactsasasecondarypowersupplyintheevent ofaplannedorunexpectedlossoftheACorDCexternalpowersupply.When
operatingtheInogenOne®G4fromanACorDCexternalpowersupply,aproperlyinsertedInogenOne®G4batteryshouldbemaintainedintheunit.Thisprocedurewillensureuninterruptedoperationandwilloperateallalertsandalertsintheeventofalossoftheexternalpowersupply.
4. Connect the Power Supply. ConnecttheACinputplugtothepowersupply.ConnecttheACpowerplugtothepowersourceandconnectthepoweroutputplugtotheInogenOne®G4.ThegreenLEDonthepowersupplywillbeilluminatedandabeepwillsoundfromtheconcentrator.
CAUTION Ensurethepowersupplyisinawellventilatedlocationasitrelieson aircirculationforheatdissipation.Thepowersupplymaybecomehotduring operation.Makesurethepowersupplycoolsdownbeforehandling.
CAUTION Thepowersupplyisnotwaterresistant.
CAUTION Donotdisassemblethepowersupply.Thismayleadtocomponentfailure and/orsafetyrisk.
CAUTION Donotplaceanythinginthepowersupplyportotherthanthesuppliedwallcord.AvoidtheuseofelectricalextensioncordswiththeInogenOne®G4.
Ifanextensioncordmustbeused,useanextensioncordthathasanUnderwritersLaboratory(UL)Markandaminimumwirethicknessof
18gauge.Donotconnectanyotherdevicestothesameextensioncord.
13OperatingInstructionsChapter3
NOTE Undercertainconditions(seeTechnicalSpecifications)thepower supplymayshutdown.ThegreenLEDwillblinkorwillnolongerbe
illuminated.Ifthisoccurs,disconnectthepowersupplyforatleast10secondsandreconnect.
NOTE WhenthepowersupplyisdisconnectedfromtheACoutlet,alsodisconnectitfromtheconcentratortoavoidunnecessarybatterydischarge.
5. Connect the nasal cannula tubing to the nozzle fitting.NozzlefittingislocatedonthetopoftheInogenOne®G4.Useofasinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.Additionaltitrationmaybeneededtoensureproperoxygendeliverywhenusingaparticularcannula.
CAUTION Toensureoxygenflow,ensurethatthenasalcannulaisproperlyconnectedtothenozzlefittingandthatthetubingisnotkinkedorpinchedinanyway.
CAUTION Replacethenasalcannulaonaregularbasis.Checkwithyourequipmentproviderorphysiciantodeterminehowoftenthecannulashouldbereplaced.
6. Turn on your Inogen One® G4 by pressing the ON/OFF Button.AsingleshortbeepwillsoundaftertheInogenlogoisdisplayed.“PleaseWait”willappearwhiletheconcentratorstartsup.Thedisplaywillindicatetheselectedflowsettingandpowercondition.Followingabriefstart-upsequence,awarmupperiodupto2minuteswillinitiate.Duringthistimeperiodtheoxygenconcentrationisbuildingtobutmaynothavereachedspecification.AdditionalwarmuptimemaybeneededifyourInogenOne®G4hasbeenstoredinextremelycoldtemperatures.
7. Set the Inogen One® G4 Concentrator to the flow rate prescribed by your physician or clinician. Usethe+or–settingbuttonstoadjusttheInogenOne®G4tothedesiredsetting.Thecurrentsettingcanbeviewedonthedisplay.
English
14
! C
8. Position the nasal cannula on your face and breathe through your nose. TheInogenOne®G4willsensetheonsetofinhalationanddeliveraburstofoxygenataprecisetimewhenyouinhale.TheInogenOne®G4willsenseeachbreathandcontinuetodeliveroxygeninthismanner.Asyourbreathingratechanges,theInogenOne®G4willsensethesechangesanddeliveroxygenonlyasyouneedit.Attimes,ifyouinhaleveryquicklybetweenbreathstheInogenOne®G4mayignoreoneofthebreaths,givingtheappearanceofamissedbreath.ThismaybenormalastheInogenOne®G4sensesandmonitorsthechangesinyourbreathingpattern.TheInogenOne®G4willnormallysensethenextbreathanddeliveroxygenaccordingly.
Agreenlightwillflasheachtimeabreathisdetected.Makecertainthatthenasalcannulaisproperlyalignedonyourfaceandyouarebreathingthroughyournose.
WARNING Ifyoubegintofeelillorareexperiencingdiscomfortwhileusingthisdevice,consultyourphysicianimmediately.
CAUTION TheInogenOne®G4isdesignedtoprovideaflowofhighpurityoxygen.An
advisoryalert,“OxygenLow”,willinformyouifoxygenconcentrationdrops.Ifalertpersists,contactyourequipmentprovider.
GeneralToremovepower,unplugtheinputcordfromitssource(i.e.,ACwalloutlet,DCcarcigarettelighteradapter)anddisconnectitfromtheInogenOne®G4.
CAUTION Ensurethepowersupplyispoweredfromonlyonepowersource(ACorDC)atanygiventime.
15OperatingInstructionsChapter3
Additional Operating Instructions
For Use In Home with AC PowerForoperationofthepowersupplyusinganACpowersource,followtheseinstructions:
1. ConnecttheACinputtothepowersupply.2. ConnecttheACpowerplugtothepowersourceandconnectthepoweroutputplug
totheInogenOne®G4.ThegreenLEDwillbeilluminated,indicatingthe powersupplyhasinputpower.
Traveling with your Inogen One® G4 System
TheInogenOne®G4Systemmakestravelbyboat,carortrainmoreconvenientforoxygenusersthaneverbefore.Nowyougetthesamequalityperformanceandconveniencewhileonthegothatyou’reusedtoreceivingfromyourInogenOne®G4athome.HerearesomeusefulandimportantinstructionsformaximizingperformanceandconveniencewhenusingyourInogenOne®G4whiletraveling.
Youshouldbeginplanningforatripwithachecklistofitemstoremember.Thischecklistshouldinclude:
✓ ACpowersupplyandDCpowercable✓ ExtraBattery(s)ifrequired✓ Importantphonenumbers,suchasthoseofyourphysicianandhomehealthcare provider,orprovidersintheareayouwillbetraveling✓ Planforbackupoxygenintheeventofaprolongedpoweroutageor mechanicalfailure
For Use In Automobile/RV/Boat/AircraftForoperationusingaDCpowersource,followtheseinstructions:
1.ConnecttheDCpoweroutputplugtotheInogenOne®G4.2.ConnecttheDCpowerplug(CigaretteLighterAdapter)to
thepowersource.3. Theplugshouldinsertintothesocketwithoutexcessive
forceandsecurelyremaininplace.
English
16
4. Checkthedevicedisplayscreentoconfirmthatanexternalpowersupplyisconnected. AniconofabatterywithalightningboltoranACpowerplugwillbedisplayedonthe
screen,indicatingthatexternalpowerisconnectedandfunctioningproperly.
WARNING EnsurethattheautomobilepowersocketisadequatelyfusedfortheInogenOne®G4powerrequirements(minimum15Amp).Ifthepowersocketcannotsupporta15Ampload,thefusemaybloworthesocketmaybedamaged.
WARNING ThetipoftheCigaretteAdapterPlugbecomesHOTwheninuse.Donottouchthetipimmediatelyafterremovalfromanautocigarettelightersocket.
CAUTION Ensuretheautomobilepowersocketiscleanofcigaretteashandtheadapterplugfitsproperly,otherwiseoverheatingmayoccur.
CAUTION Donotusethepowersupplywithacigaretteplugsplitterorwithan extensioncable.ThismaycauseoverheatingoftheDCpowerinputcable.
CAUTION DonotjumpstarttheautomobilewiththeDCpowercableconnected.Thismayleadtovoltagespikeswhichcouldshutdownand/ordamagetheDCpowerinputcable.
CAUTION WhenpoweringtheInogenOne®G4inanautomobileensurethevehicle’sengineisrunningfirst,beforeconnectingDCcableintocigarettelighteradapter.Operatingthedevicewithouttheenginerunningmaydrainthevehicle’sbattery.
CAUTION Achangeinaltitude(forexample,fromsealeveltomountains)mayaffecttotaloxygenavailabletothepatient.Consultyourphysicianbeforetravelingtohigherorloweraltitudestodetermineifyourflowsettingsshouldbechanged.
Traveling By AirTheInogenOne®G4conformstoallapplicableFAArequirementsforPOCuseonboardanaircraft.
Planning Your FlightWhenflyingwiththeInogenOne®G4,youmustinformtheairlinethatyouwillbeusingyourInogenOne®G4onboardtheaircraft.YoushouldcontacttheSpecialAssistancedeskoftheairlineonwhichyouwillbetravelingtolearnofanypaperwork,etc.thatmightbeneeded.
! C
! C
17OperatingInstructionsChapter3
English
•Someairlinesmayequiptheiraircraftwithonboardelectricalpower.YoumayhaveanopportunitytorequestaseatwithapowerportwhichcanbeusedtopoweryourInogenOne®G4.However,availabilityvariesbyairline,typeofaircraftandclassofservice.Youshouldcheckwithyourairlinesforavailabilityandalwaysplanonhavingsufficientbatterypowerfornolessthan150%oftheexpecteddurationoftheflight.Airlinesmayhavespecificrequirementsforbatterylifeduration,socheckwiththeairlinebeforetraveling.
•YourDCPowerCableisequippedwithacommonlyusedcigarettelighteradapter.However,aircraftsusedifferentpowerportconfigurationsanditisdifficulttodeterminewhichtypeofpowerportcompatibilityyouraircraftmaysupply.It’sagoodideatopurchaseanadapter,foundatelectronicandtravelstores.
Before Your FlightHerearesomethingstokeepinmindthedayyourflightdeparts:
•EnsureyourInogenOne®G4isclean,ingoodconditionandfreefromdamageorothersignsofexcessivewearorabuse.
•BringenoughchargedbatterieswithyoutopoweryourInogenOne®G4fornolessthan150%oftheexpecteddurationoftheflight,plusaconservativeestimateofunanticipateddelays.
•Regional/Commuterairlinesdonotofferonboardelectricalpower.Ifyourtravelplanscallforflightsonregionalairlines,bringenoughchargedbatterieswithyoutopoweryourInogenOne®G4fornolessthan150%oftheexpecteddurationoftheflight,plusaconservativeestimateofunanticipateddelays.
•Arriveattheairportearly.AirportsecurityscreeningpersonnelmayrequireextratimetoinspectyourInogenOne®G4.
•Whilewaitingtoboardyourflight,youmaybeabletoconservebatterypowerbyusingtheACPowerSupplytopoweryourInogenOne®G4fromanelectricaloutletintheairportterminalifavailable.
18
During Your Flight1. Ifusingairlinepowerport,removetheBatteryfromtheInogenOne®G4Oxygen
Concentrator.Duetoaircraftpowerlimitations,theACPowerSupplycannotbeusedtochargetheInogenOne®G4Batterywhenonboardaircraft.
2. ConnecttheDCPowerPlugforavailableairlinepower.Checkwithairlinepersonneltoensurecompatibility.
•Duringtaxi,takeoffandlanding,stowyourInogenOne®G4undertheseatinfrontofyou.YourInogenOne®G4willfituprightundermostairlineseats.
•ItisnotnecessarytoturnoffyourInogenOne®G4duringtaxi,takeoffandlanding.
CAUTION Achangeinaltitude(forexample,fromsealeveltomountains)mayaffecttotaloxygenavailabletothepatient.TheInogenOne®G4hasbeenverifiedtoprovideoxygentospecificationupto10,000ft.(3048m).Consultyourphysicianbeforetravelingtohigherorloweraltitudestodetermineifyourflowsettingsshouldbechanged.
After Your Flight•Remembertorechargeadditionalbatteriesyoumayhaveusedpriortoyournextflight.
Traveling By Bus, Train or BoatContactyourcarriertofindoutaboutpowerportavailability.
19
! C
English
OperatingInstructionsChapter3
Battery Operating InstructionsEnsurethatthebatteryisinplaceandcharged.DisconnecttheInogenOne®G4fromitspowersource.WhiletheInogenOne®G4isoperatingonbatterypower,thebatterywilldischarge.Thedisplaywillindicatetheestimatedremainingpercentage(%)orminutesofuse.
Whentheconcentratordetectsthatthebatterylifeislow,withlessthan10minutesremaining,alowpriorityalertwillsound.Whenthebatteryisempty,thealertwillchangetoahighpriority.
Whenbatterylifeislow,dooneofthefollowing:• PlugtheInogenOne®G4intoanACorDCpowersourceusingtheACpowersupplyor DCcable.•ReplacethebatterywithachargedbatteryafterturningofftheInogenOne®G4 (bypressingtheON/OFFbutton).Toremovebatterypressandholdthebatterylatch buttonandslidebatteryofftheconcentrator.• Ifthebatteryisdrained,chargethebatteryorremoveitfromtheconcentrator.
IftheInogenOne®G4isbeingpoweredbytheACpowersupplyorDCpower,batterieswillchargeduringoperation.LeavingyourInogenOne®G4pluggedinpastthefullchargetimewillnotharmtheconcentratororthebattery. WARNING Itistheresponsibilityofthepatienttoperiodicallycheckthebatteryand replaceasnecessary.Inogenassumesnoliabilityforpersonschoosingnotto adheretomanufacturersrecommendations.
Normal Battery ChargingToensurethatyourbatteryisproperlycharging,inspectthatthecorrectACandDCpoweroutputplugadapterisbeingusedandthattheadapterisproperlyinsertedintothepoweroutlet.Observethedisplayorlightsthatindicatechargingstatus.
NOTE Whenstartingtochargeafullydischargedbattery,thechargingprocessmay startandstopduringthefirstfewminutes.
20
Battery Care and MaintenanceYourInogenOne®G4LithiumIonBatteryrequiresspecialcaretoensureproperperformanceandlonglife.UseonlyInogenOne®G4BatterieswithyourInogenOne®G4Concentrator.
Keep DryAlwayskeepliquidsawayfrombatteries.Ifbatteriesbecomewet,discontinueuseimmediatelyanddisposeofbatteryproperly.
Effect of Temperature on Battery PerformanceTheInogenOne®G4singlebatterypowerstheInogenOne®G4Concentratorupto2.7hoursundermostenvironmentalconditions.Toextendtherun-timeofyourbattery,avoidrunningintemperatureslessthan41˚F(5˚C)orhigherthan95˚F(35˚C)forextendedperiodsoftime.
Battery Time Remaining ClockTheInogenOne®G4continuouslydisplaysbatterytimeremaining.Thisdisplayedtimeisonlyanestimateandtheactualtimeremainingmayvaryfromthisvalue.
Please Follow These Important Guidelines to Maximize Battery Performance and Life:• Storebatteryinacool,dryplace.Storewithachargeof40-50%.• Ifusingmultiplebatteries,makesurethateachbatteryislabeled(1,2,3orA,B,C,etc.) androtateonaregularbasis.Batteriesshouldnotbeleftdormantformorethan 90daysatatime.
21
Display
TheInogenOne®G4displaycontainspowerstatusicons,modeicons,textwithinformationalmessagesanderrornotifications.
Power Status Icons
Theseiconsareexamplesofthoseshowninthedisplay’swindowwhentheInogenOne®G4isoperatingonbatterypower.
Inogen One® G4 Oxygen Concentrator Audible and Visible Signals
AudibleandVisibleSignalsChapter4
! C
! C
! C
! C
Meaning
Batteryisempty.
Batteryhaslessthan10%chargeremaining.Thisiconflashes.
Batteryhasapproximately40%to50%chargeremaining.
Batteryisfull.
Icon
4English
22
Power Status Icons (continued)ThemodeiconsbelowareexamplesofthoseshownwhentheInogenOne®G4isoperatingfromanexternalpowersupplyandchargingthebattery.Thelightningboltindicatesthatanexternalpowersupplyisconnected.
! C
! C
! CMeaning
Batteryischargingwithchargelevelbetween60%and70%.
Thebatteryisfullychargedandischargingasnecessarytomaintainitscharge.
Batteryischargingwithchargelevellessthan10%.
TheInogenOne®G4isoperatingfromanexternalpowersourcewithnobatterypresent.
Icon
Mode IconsThesearethemodeiconsshowninthedisplay’swindow.
Display Text
NOTE Whentwoconditionsoccursimultaneously,theconditionwiththehigherprioritywillbedisplayed.
Meaning
Thebreathdetectionaudiblealerthasbeenenabled.
Thebreathdetectionaudiblealertisdisabled.Thisisthedefaultcondition.
Icon
! C
23AudibleandVisibleSignalsChapter4
English
Informational MessagesThefollowinginformationdisplaysarenotaccompaniedbyanyaudiblefeedbackoranyvisualchangeintheindicatorlights.
Condition/Action/Explanation
TheInogenlogoisdisplayedatstartup.
Displayedduringwarmup.“X”representstheselectedflowsetting(eg.,Setting2).
Defaultdisplaywhenoperatingonbatterypower.“X”representstheselectedflowsetting(e.g.,Setting2).“HH:MM”representstheapproximatetimeremainingonthebatterycharge(e.g.,1:45).
Defaultdisplaywhenoperatingonanexternalpowersupplyandthebatteryischarging.“xx%”representsthepercentbatterycharge(e.g.,86%).
Defaultdisplaywhenthebatteryisnotchargingorwhenthetimeremainingisnotavailablefromthebattery.
Displayedwhentheconcentratorispluggedinandbeingusedtochargeabattery(notbeingusedforoxygenproduction).Itisnormaltoseeafullychargedbatteryreadbetween95%and100%whenexternalpowerisremoved.Thisfeaturemaximizestheusefullifeofthebattery.
Message Display & Text
SettingXPleaseWait
SettingXBatteryHH:MM
SettingXChargingxx(or)BatteryFull
SettingXBatteryxx%
Chargingxx%(0r) BatteryFull
Notifications
! C
WARNING Audiblenotifications,rangingfrom55dbato65dbadependingontheusersposition,aretowarntheuserofproblems.Toinsurethataudiblenotificationsmaybeheard,themaximumdistancethattheusercanmoveawayfromitmustbedeterminedtosuitthesurroundingnoiselevel.
24
Low Priority AlertsThefollowinglowpriorityalertmessagesareaccompaniedbyadouble beepandasolid yellow light.
Condition/Action/Explanation
Batterypowerislow,withlessthan10minutesremaining.Attachexternalpowersupplyorpowerdownandinsertafullychargedbattery.
Columnmaintenanceisrequiredwithin30days.Contactyourequipmentprovidertoarrangeforservice.
Message Display & Text
BatteryLow AttachPlug
Replace Columns
Notifications (continued)TheInogenOne®G4monitorsvariousparametersduringoperationandutilizesanintelligentalertsystemtoindicateamalfunctionoftheconcentrator.Mathematicalalgorithmsandtimedelaysareusedtoreducetheprobabilityoffalsealertswhilestillensuringpropernotificationofanalertcondition.Ifmultiplealertconditionsaredetected,thehighestpriorityalertwillbedisplayed.Notethatfailuretorespondtothecauseofanalertconditionforlow,mediumandhighpriorityalertspotentiallywillresultindiscomfortorreversibleminorinjuryonlyanddevelopwithinaperiodoftimesufficienttoswitchtoabackupsourceofoxygen.Thefollowingnotificationmessagesareaccompaniedbyasingle, short beep.
Condition/Action/Explanation
On/Offbuttonhasbeenpressedfortwoseconds.Concentratorisperformingsystemshutdown.
AudibleAlertbuttonhasbeenpressedforfiveseconds.
Message Display & Text
PleaseWait ShuttingDown
HH:MM Vx.x:SerialNumber
25AudibleandVisibleSignalsChapter4
English
Low Priority Alerts (continued)Condition/Action/Explanation
Batteryerrorhasoccurred.Checktheconnectionofyourbatteryandensurethatitisproperlyattachedandlatchedonconcentrator.Ifbatteryerrorrecurswithsamebattery,stopusingthebatteryandswitchtoanewbatteryorremovebatteryandoperateconcentratorusingexternalpowersupply.
Concentratorisproducingoxygenataslightlylowlevel(<82%)foraperiodof10minutes.Ifconditionpersists,contactyourequipmentprovider.
Batteryhasexceededitschargingtemperatureandcharginghasstopped.Thebatterywillnotchargewhilethisalertispresentbutwillbegintochargewhenthebatterytemperaturereturnstothenormaloperatingrange.Ifbatterychargingisdesiredsooner,removethebatteryfromtheconcentratorandallowittocoolinanopenareaforapproximately10-15minutes.Then,re-insertthebatteryintotheInogenOne®G4.Iftheproblemstillpersists,contactyourequipmentprovider.
Theconcentratorrequiresservicingattheearliestconvenience.Theconcentratorisoperatingtospecificationandmaycontinuetobeused.Contactyourequipmentprovidertoarrangeforservice.
Theconcentrator’soxygensensorhasmalfunctioned.Youmaycontinuetousetheconcentrator.Iftheconditionpersists,contactyourequipmentprovider.
Message Display & Text
CheckBattery
OxygenLow
RemoveBatterytoCool
ServiceSoon
SensorFail
26
Condition/Action/Explanation
Concentratorhasnotdetectedabreathfor60seconds.Checkthatcannulaisconnectedtoconcentrator,therearenokinksintubingandcannulaispositionedproperlyinyournose.
Oxygenoutputconcentrationhasbeenbelow50%for10minutes.Ifconditionpersists,switchtoyourbackupoxygensourceandcontactyourequipmentprovidertoarrangeforservice.
Abreathhasbeenrecognized,butproperoxygendeliveryhasnotbeendetected.
Batteryhasexceededtemperaturelimitwhileconcentratorisrunningonbatterypower.Ifpossible,moveconcentratortoacoolerlocationorpowerunitwithanexternalpowersupplyandremovebattery.Ifconditionpersists,contactyourequipmentprovider.
Message Display & Text
NoBreathDetect CheckCannula
OxygenError
O2DeliveryError
BatteryHOT Warning
Medium Priority AlertsThefollowingmediumpriorityalertmessagesareaccompaniedbyatriple beep,repeatedevery25seconds,andaflashing yellow light.
27
Condition/Action/Explanation
Concentratorhasinsufficientbatterypowertoproduceoxygen.Attachexternalpowersupplyorexchangebattery,thenrestartunitifnecessarybypressingOn/Offbutton.
Batteryhasexceededtemperaturelimitwhileconcentratorisrunningonbatterypower.Concentratorhasstoppedproducingoxygen.Ifpossible,moveconcentratortoacoolerlocation,thenturnpoweroffandbackon.Ensureairintakeandoutletventshaveclearaccessandparticlefiltersareclean.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.
Concentratortemperatureistoohighandoxygenproductionisshuttingdown.Ensureairintakeandoutletventshaveclearaccessandparticlefiltersareclean.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.
Message Display & Text
BatteryEmpty AttachPlug
BatteryHOT
SystemHOT
High Priority Alerts
CAUTION IfyouarenotneartheInogenOne®G4youmaynotbeabletohearorseethehighpriorityalerts.MakesuretheInogenOne®G4isinalocationwherethealertsandalertswillberecognizediftheyoccur.
Thefollowinghighpriorityalertmessagesareaccompaniedbyafive beep pattern,repeatedevery10secondsandaflashing yellow light.
AudibleandVisibleSignalsChapter4
English
28
Condition/Action/Explanation
Thismayresultfromtheconcentratorbeingstoredinacoldenvironment(below0ºC(32ºF)).Movetoawarmerenvironmenttoallowtheunittowarmupbeforestartingit.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.
Concentratorhasstoppedproducingoxygenandisshuttingdown.Youshould:
1. Switchtobackupoxygensource2.Contactyourequipmentprovider
Message Display & Text
SystemCOLD
SystemError
High Priority Alerts (continued)
29
Troubleshooting
TroubleshootingChapter5
Useexternalpowersupplyorreplacebatterywithonethatisfullycharged
Checkpowersupplyconnectionandverifygreenlightissolid
CheckDCCableconnectionattheConcentratorandatcigarettelighterorauxiliaryDCpowersource
Contactyourequipmentprovider
Batteryisdischargedornobatteryispresent
ACPowersupplyisnotconnectedproperly
DCCableisnotconnectedproperly
Malfunction
ConcentratordoesnotpoweronwhenOn/Offbuttonispressed
Recommended Solution
RefertoChapter4
Possible Cause
RefertoChapter4
Problem
Anyproblemaccompaniedbyinformationonconcentratordisplay,indicatorlightsand/oraudiblesignals
Solutionstosomepossibleissuesyoumayencounteraredescribedinthissection.
Inogen One® G4 Oxygen Concentrator
5English
30
PressOn/Offbuttontopowerconcentrator
Checkcannulaanditsconnectiontoconcentratornozzle
Concentratorisnotpoweredon
Cannulaisnotconnectedproperlyoriskinkedorobstructed
Nooxygen
Recommended SolutionPossible CauseProblem
Troubleshooting (continued)
31
Cannula Replacement
Yournasalcannulashouldbereplacedonaregularbasis.Consultwithyourphysicianand/orequipmentproviderand/orcannulamanufacturer’sinstructionsforreplacementinformation.Asinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.
Case CleaningYoumaycleantheoutsidecaseusingaclothdampenedwithamildliquiddetergent(suchasDawnTM)andwater.
WARNING DonotsubmersetheInogenOne®G4oritsaccessoriesinwaterorallowwatertoenterintothecase;thismayleadtoelectricalshockand/ordamage.
WARNING DonotusecleaningagentsotherthanthosespecifiedinthisUserManual.Donotusealcohol,isopropylalcohol,ethylenechlorideorpetroleumbasedcleanersonthecasesorontheparticlefilter.
Filter Cleaning and ReplacementTheparticlefiltermustbecleanedweeklytoensuretheeaseofairflow.Removefiltersfromthefrontofthedevice.Cleantheparticlefilterswithamildliquiddetergent(suchasDawnTM)andwater;rinseinwateranddrybeforereuse.
Cleaning, Care and Maintenance
Cleaning,CareandMaintenanceChapter6
6
! C
! C
English
32
NOTE Itmaybenecessarytocleantheparticlefiltermoreoftenindustyenvironments.
TopurchaseadditionalparticlefilterscontactyourequipmentproviderorInogen.
Output FilterTheoutputfilterisintendedtoprotecttheuserfrominhalationofsmallparticlesintheproductgasflow.TheInogenOne®G4includesanoutputfilterconvenientlylocatedbehindtheremovablecannulanozzlefitting.Inogenrequiresthatthisfilterbereplacedbetweenpatients.
TheoutputfiltermaybereplacedbytheequipmentproviderorbytheownerusingtheOutputFilterReplacementKit(RP-404).
TheInogenOneG4Concentratormustbecleanedanddisinfectedaspertheaboveinstructionsforeachnewpatient.Nospecialmaintenanceneedstobecarriedoutbythepatient.YourequipmentproviderperformsmaintenanceoperationstoassurecontinuedreliableservicefromyourInogenOneG4.Themanufacturer’sinstructionsforthepreventativemaintenanceofthedevicesaredefinedintheservicemanual.Allworkmustbeperformedbytrainedtechnicianscertifiedbythemanufacturer.
DC Input Cable Fuse ReplacementTheCigaretteLighterDCpowerplugcontainsafuse.IftheDCinputcableisbeingusedwithaknowngoodpowersourceandtheunitisnotreceivingpower,thefusemayneedtobereplaced.
Toreplacethefuse,followtheseinstructionsandrefertothephotographbelow.1. Removethetipbyunscrewingtheretainer.Useatoolifnecessary.2. Removetheretainer,tipandfuse.3. ThespringshouldremaininsidetheCigaretteLighterAdapterhousing.Ifthespring
isremoved,makesuretoreplacethespringfirstbeforeinsertingthereplacementfuse.
33Cleaning,CareandMaintenanceChapter6
English
4. Installareplacementfuse,InogenRP#125(BUSSMDA-12)andreassemblethetip.Ensuretheretainerringisproperlyseatedandtightened.
Inogen One® G4 Column Change Procedure
NOTE Columnchangeinstructionsareonlytobeusedwhenmaintenanceis requiredandarenotintendedforpracticepurpose.
1. TurnofftheInogenOne®G4concentratorbypressingthepowerbuttontoshutdownthedevice.
2.RemovetheInogenOne®G4concentratorfromthecarrycase,ifapplicable.
3.RemovethebatteryfromtheInogenOne®G4concentrator.
4.PlacetheInogenOne®G4concentratoronitssidesothattheundersideisvisible.Themetalcolumnas-semblycanbeseenononesideofthedevice.
FuseCigarette Adapter Plug Retainer Tip Retainer
CAUTION Forcontinuedprotectionagainstriskoffire,onlyusefusespecified.
34
Inogen One® G4 Column Change Procedure (continued) 5.Unlockthecolumnassemblybypushingthelatch
buttonawayfromthecolumns.
6.Whileholdingthebuttonopen,slidethecolumnassemblyoutofthedevicebypullingonthecolumnpullhandle.
7. RemovethecolumnscompletelyfromtheInogenOne®G4.Bothcolumnsareremovedasonepiece.
Column(metaltube)Installation8.Removedustcapsofnewcolumnassembly.Make
surethereisnodustordebriswherethedustcapswerelocated.
Openandunlocked
35Cleaning,CareandMaintenanceChapter6
English
Inogen One® G4 Column Change Procedure (continued)
9.InsertcolumnassemblyintotheInogenOne®G4concentrator.Donotleavethecolumnendsexposed;columnassemblyshouldbeinsertedintotheIno-genOne®G4assoonasthedustcapshavebeenremoved.
10.Pushthecolumnassemblyintothedevicesuchthat thecolumnsarefullyseatedintotheInogenOne®
G4concentrator.Thespringloadedlatchbuttonshouldfullyreturntotheclosedposition.
11. ConnecttheACpowersupplycordtotheInogen One®G4andplugthepowersupplyACcordintoan electricaloutlet.DoNotPowerontheInogenOne® G4concentrator.
12.Pressandholdtheplus(+)and(-)minusbuttonfor5
seconds.Thescreenwilldisplay“sievereset” message.Releasebuttononcemessageisdisplayed
onscreen.
13.Pressthealert buttononceandscreenwill display“sieveresetsuccess”.
14.PressthepowerbuttontoturnontheInogenOne®G4,andusenormally.
WARNING Donotuseanycolumnsotherthanthosespecifiedinthisusermanual.Theuseofnon-specifiedcolumnsmaycreateasafetyhazardand/orimpairequipmentperformanceandwillvoidyourwarranty.! C
Closedandlocked
36
Disposal of Equipment and AccessoriesFollowyourlocalgoverningordinancesfordisposalandrecyclingoftheInogenOne®G4andaccessories.IfWEEEregulationsapply,donotdisposeofinunsortedmunicipalwaste.WithinEurope,contacttheEUAuthorizedRepresentativefordisposalinstructions.Thebatterycontainslithiumioncellsandshouldberecycled.Thebatterymustnotbeincinerated.
Maintenance Items List• InogenOne®G4singlebattery(model#BA-400)• InogenOne®G4doublebattery(model#BA-408)• Replacementintakeparticlefilters(model#RP-405)• OutputFilterReplacementKit(model#RP-404)• InogenOne®G4columns(model#RP-406)
Forassistance,ifneeded,insettingup,using,maintaining,ortoreportunexpectedoperationorevents,contactyourequipmentprovider,ormanufacturer.
Other Service and Maintenance WARNING DonotdisassembletheInogenOne®G4oranyoftheaccessoriesor
attemptanymaintenanceotherthantasksdescribedinthisusermanual;disassemblycreatesahazardofelectricalshockandwillvoidyourwarranty.Donotremovethetamperevidentlabel.Foreventsotherthanthosedescribedinthismanual,contactyourequipmentproviderforservicingbyauthorizedpersonnel.
CAUTION DonotuselubricantsontheInogenOne®G4oritsaccessories.
! C
37
Symbols Used On Concentrator and Accessories
SymbolsUsedonConcentratorandAccessoriesChapter7
7English
Meaning
Awarningindicatesthatthepersonalsafetyofthepatientmaybeinvolved.Disregardingawarningcouldresultinsignificantinjury.
Acautionindicatesthataprecautionorserviceproceduremustbefollowed.Disregardingacautioncouldleadtoaminorinjuryordamagetoequipment.
SeeUserManualforInstructions.
U.S.FederalRegulationRestrictsthisDevicetoSalebyOrderofPhysician.MayalsobeapplicableinotherCountries.
ACPower
DCPower
NoSmokingwhiledeviceisinuse.
NoOpenFlames(Concentrator);Donotincinerate(Battery).Refertoinstructionmanual/booklet.
Manufacturer
Symbol
WARNING
CAUTION
ONLY
ONLY
ONLY
ONLY
ONLY
! C
EC REP AuthorizedRepresentativeintheEuropeanCommunity
38
Meaning
KeepDry
IndoororDryLocationUseOnly,DoNotGetWet
UseNoOilorGrease
DoNotDisassemble(contactyourequipmentproviderforservicingbyauthorizedpersonnel)
DoNotDisposeofInUnsortedMunicipalWaste
TypeBFAppliedPart,NotIntendedforCardiacApplication
ClassIIDevice
ElectricalSafetyAgencyCertificationLogo
CompliesWithApplicableEUDirectivesIncludingMedicalDeviceDirective
Symbol
ONLYONLY
ONLY
ONLYONLY
User Interface Label
Meaning
ON/OFFButton
IncreaseFlowSetting
DecreaseFlowSetting
AudibleAlertButton
Symbol
0123
93-01213-00-00 rev1
0123
COLOR SUMMARY: 1. Black
39InogenOne®G4SystemSpecificationsChapter8
English
Inogen One® G4 ConcentratorDimensions:Withsinglebattery
L/W/H:5.91in.(15.01cm.)/2.68in.(6.8cm.)/6.5in.(16.3cm.)L/W/H:5.91in.(15.01cm.)/2.68in.(6.8cm.)/7.2in.(18.2cm.)
Weight: 2.8pounds(1.27Kg.)(includessinglebattery)Noise: 40dBA(onsetting2)Warm-UpTime: 2minutesOxygenConcentration: 90%-3%/+6%atallsettingsFlowControlSettings: 3settings:1to3Power:ACPowerSupply:
DCPowerCable:RechargeableBattery:
ACInput: 100to240VAC50to60Hz
Auto-Sensing: 1.0A
DCInput: 13.5-15.5VDC,10AMax.Voltage: 12.0to16.8VDC
BatteryDuration: Upto2.7hourswithsinglebatteryUpto5hourswithdoublebattery
BatteryChargingTime: Upto3hoursforasinglebatteryUpto5hoursforadoublebattery
EnvironmentalRangesIntendedforUse:
Temperature:41to104˚F(5to40˚C)Humidity:0%to95%,non-condensingAltitude:0to10,000ft(0to3048meters)
EnvironmentalRangesIntendedforShippingandStorage:
Temperature:-13to158˚F(-25to70˚C)Humidity:0%to95%,non-condensingStoreinadryenvironmentAltitude:0to10,000ft(0to3048meters)
Transportation: KeepDry,HandleWithCare
Inogen One® G4 System Specifications8
40
Inogen One® G4 Concentrator (continued)
Classifications
ModeofOperation: ContinuousDutyTypeofProtectionAgainstElectricalShock:
ClassII
DegreeofProtectiontoConcentratorComponentsAgainstElectricalShock:
TypeBFNotintendedforcardiacapplication
DegreeofProtectiontoConcentratorComponentsAgainstIngressofWater:
IP22-Verticallydrippingwatershallhavenoharmfuleffect&protectagainstingressofsolidobjects>12.5mm.diameterwhentheenclosureistiltedatanangleupto15˚fromitsnormalposition*
DegreeofSafetyforApplicationinPresenceofAnestheticGases:
Notsuitableforsuchapplication
TestedbyIndependentLaboratory:
Safety:IEC60601-1CAN/CSAC22.2No.60601-1ElectromagneticCompatibility:IEC60601-1-2RTCADO160
*NormalpositionoftheInogenOne®G4isuprightwiththeuserinterfacedisplayfacingupward.
41
PortableandmobileRFcommunicationsequipmentshouldbeusednoclosertoanypartofthedevice,includingcables,thantherecommendedseparationdistancecalculatedfromtheequationapplicabletothefrequencyofthetransmitter.
Recommendedseparationdistance:d=1.2√P150kHzto80MHzd=1.2√P80MHzto800MHzd=2.3√P800MHzto2.5GHz
WherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufactureranddistherecommendedseparationdistanceinmeters(m).
FieldstrengthsfromfixedRFtransmitters,asdeterminedbyanelectromagneticsitesurveya,shouldbelessthanthecompliancelevelineachfrequencyrangeb.
Interferencemayoccurinthevicinityofequipmentmarkedwiththefollowingsymbol:
3Vrms150kHz to80MHz
3V/m80MHz to2.5GHz
3Vrms
3V/m
ConductedRFIEC61000-4-6
RadiatedRFIEC61000-4-3
Compliance Level
IEC 60601 Test Level
Immunity Test Electromagnetic Environment - Guidance
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity: TheConcentratorisintendedforuseintheelectromagneticenvironmentspecifiedbelow.TheuseroftheConcentratorshouldmakesureitisusedinsuchanenvironment.
Floorsshouldbewood,concreteorceramictile.Iffloorsarecoveredwithsyntheticmaterial,therelativehumidityshouldbeatleast30%.
Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.
Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.
Electrostaticdischarge(ESD)
IEC61000-4-2
Electricalfasttransient/burst
EC61000-4-4
Surge
IEC61000-4-5
±6kVcontact
±8kVair
±6kVcontact
±8kVair
±2kVforpowersupplylines
±1kVforinput/outputlines
±2kVforpowersupplylines
±1kVforinput/outputlines
±1kVline(s)toline(s)
±2kVline(s)toearth
±1kVline(s)toline(s)
±2kVline(s)toearth
Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.Iftheuserofthe[MEEQUIPMENTorMESYSTEM]requirescontinuedoperationduringpowermainsinterruptions,itisrecommendedthatthe[MEEQUIPMENTorMESYSTEM]bepoweredfromanuninterruptedpowersupplyorabattery.
Powerfrequencymagneticfieldsshouldbeatlevelscharacteristicofatypicallocationinatypicalhospitalorhomeenvironment.
Voltagedips,shortinterruptionsandvoltagevariationsonpowersupplyinputlines
IEC61000-4-11
Powerfrequency(50/60Hz)magneticfield
IEC61000-4-8
<5%UT(>95%dipinUT)for0.5cycle
40%UT(60%dipinUT)for5cycles
70%UT(30%dipinUT)for25cycles
<5%UT(>95%dipinUT)for5sec
<5%UT(>95%dipinUT)for0.5cycle
40%UT(60%dipinUT)for5cycles
70%UT(30%dipinUT)for25cycles
<5%UT(>95%dipinUT)for5sec
3A/m 3A/m
English
42
NOTE At80MHzand800MHz,thehigherfrequencyrangeapplies.
NOTE Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructures,objects,andpeople.
NOTE UTisthea.c.mainvoltagepriortoapplicationofthetestlevel.
a:Fieldstrengthfromfixedtransmitters,suchasbasestationsforradio(cellular/cordless)telephonesandlandmobileradios,amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywithaccuracy.ToassesstheelectromagneticenvironmentduetofixedRFtransmitters,anelectromagneticsitesurveyshouldbeconsidered.IfthemeasuredfieldstrengthinthelocationinwhichtheconcentratorisusedexceedstheapplicableRFcompliancelevelabove,theconcentratorshouldbeobservedtoverifynormaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary,suchasre-orientingorrelocatingthedevice.
b:Overthefrequencyrange150kHzto80MHz,thefieldstrengthsshouldbelessthan3V/m.Recommended Separation Distances between Portable and Mobile RF Communications Equipment and This Device:
ThisconcentratorisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.TheuseroftheconcentratorcanhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetweenportableandmobileRFcommuni-cationsequipment(transmitters)andthisconcentratorasrecommendedbelow,accordingtothemaximumoutputpowerofthecommuni-cationsequipment.
150 kHz to 80 MHzd=1.2√P
0.120.381.23.812
0.120.381.23.812
0.230.732.37.323
0.010.11
10100
80 MHz to 800 MHzd=1.2√P
800 MHz to 2.5 GHzd=2.3√P
Separation Distance According to Frequency of Transmitter (M)Rated Maximum Power Output of Transmitter (W)
Fortransmittersratedatamaximumoutputpowernotlistedabove,therecommendedseparationdistancedinmeters(m)canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter,wherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufacturer.
Guidance and Manufacturer’s Declaration – Electromagnetic EmissionsTheconcentratorisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Theuseroftheconcentratorshouldassurethatitisusedinsuchanenvironment.
NOTE At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.
NOTE Theguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructures,objects,andpeople.
Emissions Test Compliance Electromagnetic Environment - Guidance
RFemissionsCISPR11
Group1 TheconcentratorusesRFenergyonlyforitsinternalfunction.ThereforeitsRFemissionsareverylowandnotlikelytocauseanyinterferenceinnearbyequipment.
RFemissionsCISPR11
ClassB Theconcentratorissuitableforuseinallestablishments,includingdomesticestablishmentsandthosedirectlyconnectedtothepubliclow-voltagepowersupplynetworkthatsuppliesbuildingsusedfordomesticpurposes.HarmonicEmissions
IEC61000-3-2ClassA
Voltagefluctuations/flickeremissionsIEC61000-3-3
Complies
PN 96-06728-00-02 A
Inogen, Inc.326 Bollay DriveGoleta, CA 93117Toll Free: 877-466-4362 +1-805-562-0515 (Outside the USA) E-mail: [email protected] www.inogen.com
®