User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
-
Upload
cruzrenato -
Category
Documents
-
view
221 -
download
0
Transcript of User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
1/52
Model: 3029 and 3039
Countertop Incubator
10 cu. ft capacity
Operating and Maintenance Manual
Manual No: 7013029
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
2/52
Models 3029 and 3039_____________________________________________________________________________________
Read This Instruction Manual.
Failure to read, understand and follow the instructions inthis manual may result in damage to the unit, injury to operat-
ing personnel, and poor equipment performance.
CAUTION! All internal adjustments and maintenance must
be performed by qualified service personnel.
Refer to the serial tag on the back of this manual.
The material in this manual is for information purposes
only. The contents and the product it describes are subject to
change without notice. Forma Scientific, Inc. makes no repre-
sentations or warranties with respect to this manual. In no event
shall Forma Scientific, Inc. be held liable for any damages, direct
or incidental, arising out of or related to the use of this manual.
MANUAL NUMBER 7013029
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
3/52
Models 3029 and 3039_____________________________________________________________________________________
Table of Contents
Section 1 - Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 1
1.1 Preliminary Inspection . . . . . . . . . . . . . . . . . . . . . . . .1 - 1
1.2 Shipping Claims . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 1
1.3 Responsibility for Shipping Damage . . . . . . . . . . . . .1 - 1
Section 2 - Basic Set-Up Procedure . . . . . . . . . . . . . . . .2 - 1
2.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
2.2 Leveling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
2.3 Drain Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
2.4 Water Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
2.5 CO2 Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
2.6 Door Heater . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
2.7 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
2.8 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
2.9 Stabilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
2.10 Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
2.11 Stabilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
2.12 CO2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 12.13 Door Heater . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
Section 3 - Specifications . . . . . . . . . . . . . . . . . . . . . . . .3 - 1
Section 4 - General Description . . . . . . . . . . . . . . . . . . .4 - 1
4.1 The Model 3029/3039 CO2 Incubator . . . . . . . . . . . .4 - 1
4.2 The CH/P Control System . . . . . . . . . . . . . . . . . . . . .4 - 1
4.3 Standard Features of the 3029/3039 . . . . . . . . . . . . . .4 - 1
4.4 Auxiliary Equipment . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 1
Section 5 – Physical Installation . . . . . . . . . . . . . . . . . . .5 - 1
5.1 Leveling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 1
5.2 Attaching the Drain Lines . . . . . . . . . . . . . . . . . . . . . .5 - 1
5.3 Connect to Power . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 1
5.4 Shelf Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 1
Section 6 - Power Module . . . . . . . . . . . . . . . . . . . . . . . .6 - 1
6.1 Main Power Switch and Indicating Light . . . . . . . . . .6 - 16.2 Temperature Limit Safety Thermostat . . . . . . . . . . . . .6 - 1
6.3 Door Ajar Indicating Light . . . . . . . . . . . . . . . . . . . . .6 - 1
6.4 Door Heaters and Control Potentiometer . . . . . . . . . .6 - 1
Section 7 - Temperature Module . . . . . . . . . . . . . . . . . .7 - 1
7.1 Temperature Power Switch and Indicating Light . . . .7 - 1
7.6 Temperature Calibration Procedure . . . . . . . . . . . . . .7 - 1
Section 8 - Humidity Module . . . . . . . . . . . . . . . . . . . . .8 - 1
8.1 Humidity Power Switch and Indicating Light . . . . . .8 - 1
8.2 CH/P Humidity Controller and Digital Display . . . . .8 - 1
8.3 Humidity Indicating Light . . . . . . . . . . . . . . . . . . . . .8 - 1
8.4 Setting the Humidity . . . . . . . . . . . . . . . . . . . . . . . . . .8 - 1
8.5 Setting Up the Humidification System . . . . . . . . . . . .8 - 1
8.6 Humidity Reservoir and Probe Check . . . . . . . . . . . .8 - 1
8.7 Humidity Probe Placement . . . . . . . . . . . . . . . . . . . . .8 - 18.8 Humidity Calibration Check and Procedure . . . . . . . .8 - 2
Section 9 - Carbon Dioxide . . . . . . . . . . . . . . . . . . . . . . .9 - 1
9.1 CO2 Power Switch and Indicating Light . . . . . . . . . .9 - 1
9.2 CH/P CO2 Controller and Digital Display . . . . . . . . .9 - 1
9.3 Audible CO2 Alarm and Alarm Indicating Light . . . .9 - 1
9.4 Alarm Silence Push Button . . . . . . . . . . . . . . . . . . . . .9 - 1
9.5 Alarm Disable Switch . . . . . . . . . . . . . . . . . . . . . . . . .9 - 1
9.6 CO2 Inject Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 19.7 Sample Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 1
9.8 Sealing the Cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 1
9.9 CO2 Standardization Adjustment . . . . . . . . . . . . . . . .9 - 1
9.10 Connecting the CO2 Source . . . . . . . . . . . . . . . . . . . .9 - 1
911 Setting the CO2 Content . . . . . . . . . . . . . . . . . . . . . . .9 - 2
9.12 CO2 Standardization Check . . . . . . . . . . . . . . . . . . .9 - 2
9.13A CO2 Standardization Procedure . . . . . . . . . . . . . . .9 - 2
9.13B Recommended CO2 Standardization Procedure . . .9 - 2
9.14 CO2 Calibration Check and Procedure . . . . . . . . . . .9 - 2
Section 10 - Preventive Maintenance . . . . . . . . . . . . . . .10-1
10.1 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-1
10.2 Sterilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-1
Section 11 - Troubleshooting . . . . . . . . . . . . . . . . . . . . . .11-1
11.1 3029 Troubleshooting Guide . . . . . . . . . . . . . . . . . . .11-1
11.2A Temperature Troubleshooting Overview . . . . . . . .11-1
11.2A Temperature System Troubleshooting Overview . .11-111.3A Humidity System Troubleshooting Overview . . . .11-1
11.3B Humidity System Troubleshooting Guide . . . . . . .11-2
11.4A CO2 System Troubleshooting Overview . . . . . . . . .11-1
11.4B CO2 System Troubleshooting Guide . . . . . . . . . . . .11-2
Section 12 - Servicing . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-1
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
4/52
Section 1 - Inspection
1.1 Preliminary Inspection
Immediately upon receipt of the equipment, carefully
examine the shipment for loss or damage. Make note of any
visible damage. Unpack the shipment and inspect both interior
and exterior for any in-transit damage.
1.2 Shipping Claims
If any loss or damage is discovered, note any discrepancies
on the delivery receipt and call the delivering carrier and
request that their representative perform an inspection. Do not
discard any of the packing material and do not move the ship-
ment from the receiving area.
1.3 Responsibility for Shipping Damage
For products shipped F.O.B. Marietta, Ohio, the responsi-
bility of Forma Scientific, Inc. ends when the merchandise is
loaded onto the carrier’s vehicle.
On F.O.B. Destination shipments, Forma Scientific and the
carrier’s responsibility ends when your Receiving Department
personnel sign a free and clear delivery receipt.
Whenever possible, Forma Scientific will assist in settling
claims for loss or in-transit damage.
Models 3029 and 3039 ________________________________________________________________________ Inspection
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
5/52
2.8 Temperature (Reference Section 7)
Push in temperature set button and adjust setscrew untilthe desired temperature setpoint appears on the display.
Remove screwdriver (knob at left of sample port). Turn
temp power switch on.
2.9 Stabilization
Shut both doors and allow the incubator to run for 1
hour.
2.10 Humidity (Reference Section 8)
Turn RH power switch on. Push in the humidity knob
and use the coarse and fine adjustments to obtain the
desired setpoint.
2.11 Stabilization
Shut both doors and allow the incubator to run for 6 to 8
hours or overnight.
2.12 CO2 (Reference Section 9)
Turn the CO2 power switch on. Set the display to ‘00.0’
by using the standardization adjustment, and use thecoarse and fine adjustments to obtain the desired set-
point. Release the knob. Turn on the CO2 at the source.
Allow the incubator to stabilize for 30 minutes. Check
the CO2 reading with a Fyrite. Adjust the display to
match the Fyrite reading by using the standardization
adjustment.
2.13 Door Heater (Reference Section 6.4)
If there is condensation of the inner glass door, adjust
the door heater to 10.
Models 3029 and 3039_________________________________________________________________Basic Set-Up Procedure
Section 2 - Basic Set-Up Procedure
2.1 Unpacking (Reference Section 1)
Unpack the incubator and inspect for damage.
Remove and discard the aluminum shipping plate from
the front of the incubator.
2.2 Leveling (Reference Section 5.1)
Locate the cabinet and level, allowing 1/2” clearance
beneath the cabinet.
2.3 Drain Lines (Reference Section 5.2)
Remove the valve shipped inside the cabinet and install
in the drain line in place of the pipe plug. Attach tubing
to thevalve
.The overflow drain should be trapped and piped to an
open drain and the trap filled with water.
2.4 Water Supply (Reference Section 8.5)
Remove the floor section from the incubator. Fill the
floor pan with 1-1/2 gallons of distilled water.
Fill the water bottle and allow the floor pan to fill. Thewater level should be 1/8” to 3/8” from the top of the
overflow.
Position the humidity probe so the tip is submerged 1/8”
to 1/16” below the surface of the water.
2.5 CO2 Supply (Reference Section 9.10)
Connect a gaseous CO2 supply to the compression fittingon the back of the unit. Set the regulator to 4 to 5 PSIG.
DO NOT turn on the CO2 supply at this point.
2.6 Door Heater (Reference Section 6.4)
T h d h k b ( b h b l ) h
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
6/52
Section 3 - Specifications
a. Capacity and Weight
Chamber Capacity: 10 cu. ft. (285 liters)
Shelf Capacity: 24 with 97 sq. ft. available space
Shipping Weight: 275 lbs (117kg)
b. Dimensions
Exterior: 40.0”W x 38.5”H x 26.0”F-B
(102cm x 98cm x 66cm)
Interior: 27”W x 29”H x 22”F-B
(69cm x 74cm x 56cm)
c. Construction
Exterior: 20 Gauge Stainless Steel 2-B
Interior: 18 Gauge Stainless Steel 2-B
Insulation: 1” (2.5cm) Urethane (Cabinet)
2” (5.1cm) Fiberglass (Door)
Shelves: (6) Stainless Steel
d. Heaters
Water: 300W, 120V, Stainless Steel
Air: 300W, 120V, Wirewound Stainless Steel
Door: 100W, 120V, Silicone Strip
e. Temperature
Control Sensitivity: ±0.01°CRange: +4°C above ambient to +60°C
Repeatability: ±0.06°C
Uniformity: ±0.25°C
f. CO2
Control Sensitivity: ±0.1%
Range: 0% to 20%
g. Humidity (at +21°C/50% RH ambient)
Control: ±5%
Performance: 50% to 96% at +60°C chamber temp
70% to 96% at +37°C chamber temp
Control Sensitivity: ±0.1%
h. Electrical
3029: 120VAC, 50/60Hz, 1 Phase, 6 FLA
3039: 220VAC, 50/60 Hz, 1 Phase, 3.5 FLA
i. Utility Connections
CO2: 1/4” Compression Fitting
Drains: 1/4” MPT Chamber Drain
1/4” MPT Overflow Drain
Water: Self-Contained 1 gallon (3.8 liters) Bottle -supplies chamber reservoir,
external supply optional
Models 3029 and 3039_________________________________________________________________________Specifications
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
7/52
Section 4 - General Description
4.1 The Model 3029/3039 CO2 Incubator
The Model 3029/3039 incubator is designed to provide
continuous temperature, CO2, and humidity control in a
10 cu. ft., (285 liters), forced draft environment.
Separate solid state CH/P controllers provide direct digi-
tal setpoint control and readout of temperature in 0.1°C
increments, CO2 (within 0.1%), and relative humidity
(within 1.0%). Setpoint and readout of all three parame-
ters are presented on individual L.C.D. displays.
The temperature range of the chamber is approximately
4°C above ambient to +60°C. Temperature control is
+0.1°C, and temperature uniformity is better than
+0.25°C.
The CO2 control system is also fully automatic, allowing
the operator a CO2 content range of 0% to 20% with a
control sensitivity of +0.1%.
Rated in a +21°C/50 RH (relative humidity) ambient, a
humidity range of 50% to 96% is achieved at +60°C
chamber temperature, and a range of 70% to 96% is
achieved at +37°C chamber temperature.
4.2 The CH/P Control System
The Forma CH/P control system for the 3029/3039 incu- bator provides independent control of temperature, CO2
percentage, and humidity percentage. The system has
been designed for easy troubleshooting and repair.
The controls are easy to operate. Setpoints can be set by
simply pushing in the set knob for the desired parameter,
and turning to the coarse and fine adjustment to obtain
the chosen setpoint. Individual digital displays continu-
ally monitor all three parameters.
An independent alarm system monitors deviations from
CO2 setpoint and gives audible and visual warnings if
CO2 content deviates from setpoint by more than 1%
(nominal). An alarm bypass switch deactivates the CO2
alarm system and a spring-loaded reset temporarily
silences the alarm warning signal.
4.3 Standard Features of the 3029/3039
Self-Diagnostic Control Design: The control systemshave been designed for easy troubleshooting by the cus-
tomer. Not only does self-diagnosis decrease downtime,
but it also saves on service calls.
Positive Forced Air Convection: The air flow system
circulates air horizontally across the shelving in a left-
to-right direction. This design assures uniform distribu-
tion of temperature and humidity, eliminates CO2 strati-
fication, and assures rapid CO2 and RH recovery in the
event of loss. Conditioned air is evenly distributed at allshelf levels under varying load conditions.
Cabinet Construction: The interior is polished stainless
steel and the exterior is constructed with heavy gauge
stainless steel and finished with polyurethane enamel.
Double wall construction is employed to maximize envi-
ronmental stability and provide more economical opera-
tion.
Simplified Cleaning and Sterilization: Interior shelf sup- ports and side panels are easily removable for cleaning
and sterilization. Side panels are bracket mounted and
the floor panel is connected with two screws to make the
humidity reservoir easily accessible.
Door Construction: A full view safety glass inner door
seals onto a peripheral silicone gasket. The outer door is
fitted with a peripheral magnetic gasket and seals
against the cabinet exterior. The door heater minimizes
the condensation on the inner door.
Shelving Design: Solid stainless steel shelves are
adjustable on 1" vertical center channel brackets. Six
shelves are standard equipment, and the incubator has a
24 shelf capacity to provide as much as 97 sq. ft. (9 sq.
meters) of available space.
4.4 Auxiliary Equipment
Stacking Adapter: Required for vertical stacking of two
cabinets, the adapter includes a frame support and an air
blower system. It adds 5" (12.7 cm) to the total height
of two stacked units. (Forma Catalog Number 513028)
Additional Shelves: Additional stainless steel, non-per-
forated shelves may be purchased Incubator capacity is
Models 3029 and 3039____________________________________________________________________General Description
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
8/52
Models 3029 and 3039___________________________________________________________________ Physical Installation
Section 5 – Physical Installation
5.1 Leveling
Locate the incubator on a firm, level surface in an area of
minimum ambient temperature fluctuation. Level the cabinet
by using the four adjustable feet on the bottom of the unit. The
incubator MUST be level to ensure proper operation. Be sure
adequate clearance exists on the left side of the unit for access
to the water supply and electrical components.
Allow ½” minimum clearance at the bottom of the cabinet.
If this area is blocked, the rated low limit for humidity will be
affected.
5.2 Attaching the Drain Lines
There are two drain lines on the incubator; the chamber
drain and the overflow drain.To attach the chamber drain line:
1) Remove the valve that was shipped inside the unit.
2) Install the valve in place of the pipe plug in the drain on
the back of the unit.
3) Attach tubing to the valve and direct toward a drain.
To attach the overflow drain line:
1) The overflow drain should be rigid piped and trapped.2) Fill the trap with water.
5.3 Connecting to Power
Connect the incubator to an adequate power source.
3029: 120VAC, 50/60Hz, 1 Phase, 6 FLA
3039: 220VAC, 50/60 Hz, 1 Phase, 3.5 FLA
5.4 Shelf Mounting
The shelves in the incubator have been designed to make
cleaning easier.
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
9/52
Section 6 - Power Module
6.1 Main Power Switch and Indicating Light
The Main Power switch controls all electrical power to the
incubator. The switch must be ON for any of the incubator sys-
tems to be functional.
The main power indicating light is activated whenever the
main power switch is in the ON position and the incubator is
connected to power.
6.2 Temperature Limit Safety Thermostat
The temp limit safety thermostat is designed to be set for a
temperature slightly higher than the operating setpoint tempera-
ture of the incubator. In the event of an overtemp condition, the
temp limit thermostat performs two functions:
• Activates the audible alarm and overtemp indicator light.
• Assumes control of power to immersion and air heaters.
The temp limit control is not directly calibrated. The num-
bers (0 to 10) on the dial are for reference only and do not cor-
respond to any specific temperature.
In the event of an overtemp condition, the alarm will
sound, the overtemp indicating light will activate and the incu-
bator heaters will cycle under the control of the temp limit ther-
mostat. The temperature will be maintained at the temp limit
setting until the source of the overtemp condition has been
determined and corrected.Note: The temp safety control should be checked quarterly to
ensure proper operation. To check the temp safety, turn the
temp safety knob down toward the setpoint until the alarm and
overtemp light are activated. Reset the temp safety after the
test.
a. Setting the Temp Limit Safety Thermostat
Once chamber temperature is stable, the temp limit safety
thermostat should be adjusted as follows:
1. Turn the temp limit control knob slowly counterclock-
wise until the alarm sounds and the overtemp light is
activated.
2 Turn the control knob clockwise no more than two units
6.3 Door Ajar Indicating Light
A Door Open indicating light is located on the power mod-
ule. The red light activates when the outer door is open or hasnot been securely closed.
When the Door Ajar light is on, power to the heaters, blow-
ers and the CO2 injection solenoid is interrupted. This feature
reduces gas usage and conserves energy.
6.4 Door Heaters and Controller
Two heaters, installed in the outer door, can be used whenthere is condensation present on the inner glass door. The heater
control has to be adjusted to operate in conjunction with the
regular temperature control system. The door heater control
should never be set higher than necessary, or above 50. If set
higher, it could have an adverse effect on the temperature con-
trol of the incubator.
The control potentiometer for the door heaters is located
above the water bottle in the left side compartment.
Use the door heaters only if there is excess condensationon the inner door.
a. To Set the Door Heaters
1. Allow the incubator to stabilize at the desired operating
temperature.
2. Locate the heater adjustment.
3. Adjust the heater knob to 10. Allow the cabinet to stabi-lize overnight.
4. If condensate is still present, adjust the door heater to
15. Continue to make adjustments in increments of 5,
allowing the cabinet to stabilize overnight with each
adjustment, until there is no condensate present.
It may be necessary to adjust the door heater after each
change in the operating temperature setting.
Note: Use of the door heater can slightly alter the temperaturein the chamber.
Models 3029 and 3039_______________________________________________________________________ Power Module
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
10/52
Section 7 - Temperature Module
7.1 Temperature Power Switch and Indicating Light
The temperature power switch controls electrical power to
the temperature module. The power indicating light on the
module is activated when both the main and temperature power
switches are in the ON position.
7.2 CH/P Temp Controller and LCD Temp Display
The LCD digital display on the temperature module contin-
ually shows actual chamber temperature in °C, and the setpoint
temperature when the temp set button is pushed.
The temperature setpoint is changed by pushing the temp
set knob in and rotating the knob to the desired temperature.
Fine adjustments can be made to the setpoint immediately after
the knob is released.
When the setpoint temperature is changed, the digital dis-
play will show the new setpoint for 5-15 seconds before revert-ing to continuous readout of actual chamber temperature.
Since the temperature sensor is located in the conditioning
plenum, small changes in setpoint may result in the desired
temperature being reached in less then 20 seconds. In such
cases, the digital display will not indicate the progressive tem-
perature change as actual temperature moves toward setpoint.
Rather, only the new setpoint will be displayed.
7.3 Heat Indicator Light
The Heat indicator light activates when the temperature
controller calls for more heat in the chamber.
7.4 Setting the Temperature Control
To set the temperature control, push in on the temp setknob and dial to the desired setpoint. Fine adjustments can be
made immediately after releasing the knob.
If the last digit on the display flickers after the chamber
temperature has stabilized, a slight readjustment of the setpoint,
using the fine adjustment, will correct the problem.
Models 3029 and 3039___________________________________________________________________ Temperature Module
7.5 Temperature Calibration Check
Note: To obtain the best results, all calibrations should be per-
formed in the following order; temperature, humidity, CO2.The CH/P temperature controller has been calibrated prior
to shipment of the unit and should not require recalibration for
most laboratory applications. A periodic check of temperature
control calibration, however, is recommended as a matter of
routine maintenance.
A check of temperature calibration is especially advisable
when the ambient temperature is less than 19°C or more than
29°C.
The calibration check should be performed with a calibrat-
ing thermometer of known accuracy, using the following
method.
1. Place the calibrating thermometer inside the chamber,
away from the door and side panels.
2. Allow adequate time for the chamber temperature to sta-
bilize. If the setpoint temperature has not been readjust-
ed before the check, stabilization should be achieved
within 15 minutes. If the temperature setpoint has been
changed by 15°C or more before the calibration check,
stabilization may require as long as 4 hours.
3. Check to see if the temperature indicated by the digital
display agrees with the temperature shown by the ther-
mometer in the chamber.
If the temperature controller requires recalibration, use the
procedure in Section 7.6.
7.6 Temperature Calibration Procedure
Temperature calibration should be performed following the
replacement of either the temperature module or the tempera-
ture probe.
A high quality, calibrating thermometer of known accuracy
is required for this procedure. Place the thermometer near the
center of the incubator chamber and allow the chamber temper-ature to stabilize. Refer to Section 7.5.
1. Turn off power to the temperature module. Pull the mod-
ule out and snap it onto the extender module. Re-install
the module and turn the power on.
2. Adjust switch S2 to position 7. See diagram in Section
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
11/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
12/52
Section 8 - Humidity Module
8.1 Humidity Power Switch and Indicating Light
The humidity power switch controls electrical power to the
humidity system. The power indicating light on the module is
activated when both the main and humidity power switches are
in the ON position.
8.2 CH/P Humidity Controller and Digital Display
The LCD digital display on the humidity module continual-
ly shows the percent of humidity (RH) and the setpoint when
the RH set button is pushed.
The RH setpoint is changed by holding the RH set knob in
and rotating the knob to the desired percentage. Fine adjust-
ments can be made to the setpoint immediately after the knob is
released.
When the setpoint is changed, the digital display will showthe new setpoint for 5-15 seconds before reverting to continu-
ous readout of the actual chamber RH.
8.3 Humidity Indicating Light
The humidity indicating light activates when the humidity
controller calls for greater humidity percentage in the chamber.
8.4 Setting the Humidity
Put the humidity power switch in the ON position. Push in
on the RH% knob and rotate the knob to the desired humidity
setpoint percentage as indicated on the display.
When running low humidity in the incubator, the excess
water in the cabinet must be eliminated. The incubator should
be free of all condensate in the chamber and under the floor
panel. The humidity reservoir should be at a normal operatinglevel.
Excess water may be eliminated by leaving the chamber
door ajar for approximately one hour. Due to the door safety
switch, the fan and heater will not operate when the door is
open.
Remove the floor section from the incubator. Fill the floor
pan with 1-1/2 gallons of water. This amount of water should
almost fill the floor pan, but still be below the point where thefloat level is activated.
Open the water bottle compartment on the left side of the
cabinet and fill the bottle with distilled or demineralized water.
Allow the chamber to fill with water until the float level is
activated and the water ceases to flow. Let the water level sta-
bilize overnight. Check the water level. It should be 1/8 to 3/8
inches from the top of the overflow.
If the water level in the floor pan is too high or too low, thefloat level control needs to be adjusted. Do this by adjusting
the thumbscrew at the base of the float shaft and, if necessary,
tilting the float mounting bracket.
8.6 Humidity Reservoir and Probe Check
The humidity reservoir water level should be checked peri-
odically to ensure proper humidification levels in the chamber.
The humidity reservoir is reached by removing the floor panel.
Be sure to disconnect power before removing the floor panel.
Make sure that the humidity probe tip is positioned so that
the tip extends 1/8” to 1/16” below the float.
The chamber reservoir water level should be maintained at
1/8” to 3/8” below the overflow level as determined by the
position of the 90° elbow located in the rear left corner of the
reservoir.
8.7 Humidity Probe Placement
The humidity probe must be properly positioned for the incu-
bator to perform optimally in registering humidity changes.
To adjust the humidity probe, it is necessary to remove the
chamber floor panel. The floor panel is connected by screws in
the front corners of the chamber.
Be sure to disconnect power before removing the floor panel
Models 3029 and 3039_____________________________________________________________________ Humidity Module
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
13/52
8.8 Humidity Calibration Check and Procedure
Humidity calibration should be performed following thereplacement of the humidity module or the humidity probe.
A hygrometer is required for this procedure. It should be
placed inside the chamber, and the chamber should be permitted
to stabilize for approximately 4 hours. During this time, the
inner door should not be opened. If the hygrometer reading dif-
fers from the digital display, recalibration is necessary.
1. Turn off power to the humidity module. Pull the module
out and snap it into the extender card. Turn the power
back on.
3. Adjust switch S2 to position 1. See diagram in Section
11.3 for the switch location.
4. Turn the zero adjustment on the side of the module until
the reading on the digital display agrees with the
hygrometer reading.
5. Turn switch S2 back to 0.
6. Allow time for the chamber to stabilize. Compare the
hygrometer reading with the reading on the digital dis-
play. When the two readings agree, the system is cali-
brated.
Models 3029 and 3039____________________________________________________________________ Humidity Module
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
14/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
15/52
Section 9 - Carbon Dioxide
9.1 CO2 Power Switch and Indicating Light
The CO2 power switch controls electrical power to the CO2
system. The power indicating light on the module is activated
when both the main and CO2 power switches are in the ON
position.
9.2 CH/P CO2 Controller and Digital Display
The LCD digital display on the CO2 module continually
shows the percent of CO2 and the setpoint when the CO2 set
knob is pushed.
The CO2 setpoint is changed by pushing the CO2 set knob
in and rotating the knob until the desired percentage is indicat-
ed on the display. Fine adjustments can be made to the setpoint
immediately after the knob is released.
When the setpoint is changed, the digital display will show
the new setpoint for 5-15 seconds before reverting to continu-
ous readout of actual percent CO2.
9.3 Audible CO2 Alarm and Alarm Indicating Light
The audible CO2 alarm and alarm indicating light activate
when the percentage of CO2 deviates from setpoint by ±1%
(nominal), or the CO2 set knob is pushed in..
9.4 Alarm Silence Push Button
The alarm silence button is located on the front of the CO2
module. When pushed, it will silence the CO2 alarm, and de-
energize the alarm light. The alarm will remain deactivated
until another alarm condition causes it to sound.
9.5 Alarm Disable Switch
The alarm disable switch is located on the side of the CO2
module. It is necessary to pull the CO2 module out slightly to
allow access to the switch.
9.6 CO2 Inject Light
The CO2 Inject light is activated whenever there is ademand for CO2 to meet the setpoint percentage requirements.
Since the CO2 inject light is independent of the CO2 alarm, it
will continue to signal a need for CO2 when the CO2 alarm is
set to either the Defeat or Silence position.
9.7 Sample Port
A sample port is located on the control panel for checking
the CO2 percentage by independent means, such as with a
Fyrite or similar instrument.
To prevent CO2 loss, the sample port should be capped when
not in use.
9.8 Sealing the Cabinet
When using CO2, the cabinet must be sealed before the unit
is turned on. Check to ensure that the sample port is capped.
Disconnect power to the unit and open the electrical access
panel on the left side of the incubator. Check the Permagum
seals on the lower right. Also check the orange plug in the
upper left. If these openings are not sealed, CO2 will be drained
from the cabinet and temperature and humidity will not stabi-
lize properly.
9.9 CO2 Standardization Adjustment
The CO2 calibration adjustment is located to the right of
the CO2 set knob. This adjustment is used for standardizing the
CO2 controller for specific ambient conditions.
See Sections 9.12 and 9.13 for further information about
standardizing the controller.
9.10 Connecting the CO2 Source
A main supply of liquid CO2, supplied from tanks without
siphon tubes, will provide for the most economical use of CO2.
It is also recommended that a t o stage press re reg lator ith
Models 3029 and 3039______________________________________________________________________ Carbon Dioxide
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
16/52
To connect the CO2 supply, insert the 1/4” O.D. metal tub-
ing (or the 1/4” O.D. copper tubing piece supplied with the unit
if plastic tubing is used) from the source into the CO2 compres-sion fitting, and tighten the nut one-and-a-quarter turn. As the
nut is tightened, the ferrule (conical shaped ring) inside the nut
exerts pressure on the tubing to hold it firmly in place. (If plas-
tic tubing is used, the tubing should be firmly clamped to the
copper tubing piece to ensure a leakproof seal.)
The CO2 source must be regulated at a pressure level of 4 to 5
PSIG. Higher pressure levels may damage CO2 filter system.
9.11 Setting the CO2 Content
The following conditions must be satisfied before the CO2
percentage can be set:
1. The temperature and humidity in the chamber must be
allowed to stabilize for 6-8 hours or overnight, with the
outer door closed. Since the CO2 sensor is affected bychanges in density and water content of the air, the incu-
bator must be adjusted for site conditions of temperature
and humidity.
2. Check the CO2 control standardization (Section 9.12).
Although all units are standardized at the factory before
shipment, the CO2 control standardization may not be
correct for the ambient conditions in which the incubator
is to be used. A check is therefore mandatory to ensure
proper operation of the CO2 system. Read Section 9.12
and 9.13 before using the incubator with CO2.
To set the CO2 percentage, turn the CO2 power switch ON.
Depress the CO2 set knob and rotate the knob until the desired
percentage is indicated on the digital display.
9.12 CO2 Standardization Check
Check standardization with every adjustment of the tempera-
ture setpoint and/or change of the humidity setting.
The check procedure requires the use of a Fyrite CO2 tester
or similar measuring device However it is possible to stan
To check the standardization, allow chamber temperature
and humidity to stabilize, then sample the CO2 content using
the measuring device and compare the test instrument readingwith the incubator display. Two samples should be taken con-
secutively to allow for any variation in temperature and humidi-
ty. If the readout does not match the test instrument, perform
one of the standardization procedures in Section 9.13.
9.13 Standardization Procedure
a. Procedure A (CO2 Tester Not Required)
1. Disconnect CO2 at source (if connected) and fan the
door about 10 times to eliminate CO2 from chamber.
2. Allow the chamber temperature and humidity to stabi-
lize for at least 4 hours. Do not open door during stabi-
lization.
3. Locate the CO2 standardization adjustment located on
the control panel.
4. Using a jeweler’s screwdriver, turn the standardizationadjustment slowly counterclockwise until the digital dis-
play reads 00.0%. Allow one minute for stabilization.
The unit is now standardized and can be adjusted to the
desired CO2 operating setpoint.
b. Procedure B (CO2 Tester Required)
1. Disconnect CO2 at source (if connected) and fan thedoor about 10 times to eliminate CO2 from chamber.
2. Allow the chamber temperature and humidity to stabi-
lize for at least 4 hours. Do not open door during stabi-
lization.
3. Locate the CO2 standardization adjustment located on
the control panel.
4. Using a jeweler’s screwdriver, turn the standardization
adjustment slowly counterclockwise until the digital dis- play reads 00.0%. Allow one minute for stabilization.
5. Connect the CO2 source and set the CO2 controller for
the desired setpoint and allow approximately 30 minutes
for stabilization.
6. Using the CO2 tester, read the actual CO2 percentage in
Models 3029 and 3039_______________________________________________________________________Carbon Dioxide
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
17/52
9.14 CO2 Calibration Check and Procedure
CO2 calibration should be performed following the replace-ment of the CO2 module or the CO2 sensor.
A Fyrite or similar CO2 test device is required for this pro-
cedure. The procedure must be followed carefully to ensure
proper calibration of the CO2 controller. Test the CO2 content
with the test instrument. If the reading differs from the reading
on the digital display, recalibration is necessary.
1. Disconnect the CO2 source and eliminate the CO2 from
chamber by fanning the door 10 times.
2. Close the doors and allow sufficient time for the temper-
ature and humidity in the chamber to stabilize.
Stabilization may take up to 4 hours. Do not open the
door during the stabilization period.
3. Turn off power to the CO2 module. Pull the module out
and snap it into the extender card. Re-install the module
and turn the power back on.
4. Adjust both the coarse and fine adjustments on the CO2
setpot fully counterclockwise.
5. Wait 60 seconds and change the switch S2 to position 1.
Turn the current adjustment on the side of the module
until the display reads 07.0. Return switch S2 to 0.
6. Turn the calibration adjustment on the front panel until
the display reads 00.00.
6. Adjust the CO2 setpoint to 5%.
7. Reconnect or turn on the CO2 supply to the chamber andallow time for the system to stabilize. In most cases, 30
minutes should be sufficient.
8. Using the test instrument, sample the actual CO2 level in
the chamber. Note the reading on the digital display.
9. To prevent additional CO2 from entering the chamber,
turn the CO2 setpoint knob fully counterclockwise.
10. Adjust the span pot on the CO2 module until the reading
on the digital display agrees with the test instrumentreading.
Note: This calibration should be suitable for most laboratory
applications. If more precise calibration is necessary, this proce-
dure should be repeated.
Models 3029 and 3039_______________________________________________________________________Carbon Dioxide
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
18/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
19/52
Section 10 - Preventive Maintenance
10.1 Cleaning
The chamber interior should be cleaned with alcohol and/or
soap and water. The interior panels should be disinfected with a
general-use laboratory disinfectant. Information about disinfec-
tants recommended by Forma Scientific is included as a supple-
ment at the end of the manual.
The cabinet exterior may be cleaned with soap and water,
or any non-abrasive commercial spray cleaner.
See Section 10.2 for a sterilization procedure.
10.2 Sterilization
The incubator can be easily sterilized in approximately 30
minutes. Full strength Amphl Spray or O-Syl disinfectant in 1%
or 3% solutions (dilute with sterile distilled water) is recom-mended.
Remove the shelves, interior duct sheets and bottom from
the incubator and set them aside. Spray all interior surfaces.
Sponge down all surfaces to remove any foreign matter, and
rinse the surfaces at least twice with distilled water.
Be sure that the door gasket is also thoroughly cleaned. It
may be necessary to lift up the edge of the gasket to clean it
thoroughly. Clean the interior of the glass door with the solu-tion, and rinse it with distilled water. The drip trough should be
wiped out and carefully rinsed.
Run some of the solution through the drain lines.
If desired, all surfaces can then be rubbed or sprayed with
70% alcohol. Be sure to drain and sponge all solution from the
incubator. All shelves, duct sheets, and removable components
can be washed down with the same solution before they are
returned to the incubator.
Do not direct a hard spray of the solution at the temperature
sensor, or rub it directly. To do so could bend the sensor wire.
Do not use strong alkaline or caustic agents. Stainless steel is
corrosion-resistant, not corrosion-proof.
The use of volatiles or aromatics is not recommended for
cleaning cabinet interiors. Residues could cause contamina-
tion of the cabinet environment.
General Instructions
Use the mildest cleaning procedure that will do the job
effectively. To ensure maximum effectiveness and to avoid mar-
ring the surface, always rub in the direction of the finish polish
lines. Be sure to rinse the surface thoroughly after each clean-
ing operation. To avoid water marks, wipe the surface dry.
a. Passivation Procedure for Surface
Contamination Removal
Note: Safety requires use of rubber gloves and eye protection.
Equipment required:
1. Nitric acid 70 to 80 percent
2. Non-metallic mixing container at least 2 qt. capacity
(2) required3. Acid brush
4. Rubber gloves
5. Goggles
6. Bicarbonate of soda
1) Procedure A - Vertical surfaces
1. Dilute acid 4 to 1 with water (add acid to water slowly).
2. With acid brush, apply nitric acid half strength to areasshowing any signs of oxidation.
3. Leave acid on surface for 30 to 60 minutes depending
upon severity of blemish.
4. Prepare a solution of sodium bicarbonate and water
(approximately 4 oz. per quart).
5. With cloth dipped in sodium bicarbonate solution, rinse
acid-treated surfaces to neutralize acid.
6. Final rinse with plain water.
2) Procedure B - Horizontal surfaces (tank bottoms, etc.)
1. Dilute acid 4 to 1 with water (add acid to water slowly).
2. Pour acid until surface is just covered.
Models 3029 and 3039_________________________________________________________________Preventive Maintenance
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
20/52
4. Remove acid.
5. Prepare a solution of sodium bicarbonate and water
(approximately 4 oz. per quart). Pour solution until sur-
face is just covered and let stand for 15 minutes.6. Remove this solution and finally rinse with plain water.
Note: Rusting of 300 Series stainless steel is purely a surface
effect and can be caused by iron carbide precipitation at joints
heated by welding or contact with carbon steel chips, tools, etc.
Surface treatment will remove this contamination.
Models 3029 and 3039________________________________________________________________Preventive Maintenance
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
21/52
1 0 - 3
M o d e l s 3 0 2 9 a n d 3 0 3 9 _ _ _ _ _ _ _ _ _ _ _ _ _
_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
_ _ _ _ _ _ _ _P r e v e n t i v e M a i n
t e n a n c e
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
22/52
1 0 -4
M o d e l s 3 0 2 9 a n d 3 0 3 9 _ _ _ _ _ _ _ _ _ _ _ _ _
_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
_ _ _ _ _ _ _ _P r e v e n t i v e M a i n
t e n a n c e
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
23/52
1 0 - 5
M o d e l s 3 0 2 9 a n d 3 0 3 9 _ _ _ _ _ _ _ _ _ _ _ _ _
_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
_ _ _ _ _ _ _ _P r e v e n t i v e M a i n t e n a n c e
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
24/52
Models 3029 and 3039_____________________________________________________________________Troubleshooting
Section 11 - Troubleshooting
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
25/52
Models 3029 and 3039______________________________________________________________________Troubleshooting
d l 3029 d 3039 bl h i
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
26/52
Models 3029 and 3039_____________________________________________________________________Troubleshooting
11.2 Temperature Troubleshooting Overview
a. General Instructions
Follow the instructions on the flowchart to diagnose prob-
lems with the temperature system. For this diagnostic aid to be
effective, all checks must be made in the sequence indicated by
the flowchart and all data carefully recorded in the table provid-
ed.
b. Temperature System Troubleshooting
Turn the temp module off and remove the temperature and
logo modules. Transfer the extender card from the logo module
to the module slot. Check the alignment. Add the temp module
to the extender and check the alignment.
Do not plug the module in upside-down. To do so will damage
the unit.
General Notes
• Before removing any control module from the unit, turn
the power to that module off.
• Unless otherwise noted in the procedure, the module
referred to is the temp module. The switches referred to are
located on the temp module.
• Switch locations for the temp module are labeled on the
following module diagram.• When rotating set knob fully in one direction, it will be
necessary to push the knob in at intervals to get the knob to
rotate completely.
• When taking readings with the set knob both in and out,
wait 20 seconds after taking the reading with the knob in
before taking the reading with the knob out.
• See Drawing 3029-51-0-D for location of the plug.
11.3 Humidity Troubleshooting Overview
a. General Instructions
Follow the instructions on the flowchart to diagnose prob-
lems in the humidity system. For this diagnostic aid to be effec-
tive, all checks must be made in the sequence indicated by the
flowchart and all data must be carefully recorded.
b. Humidity System Troubleshooting Procedure
Turn the RH module off and remove the RH and logo mod-ules. Transfer the extender card from the logo module to the
RH module slot. Snap the extender card into the connector.
Check the alignment. Add the RH module to the extender and
check the alignment.
Do not plug the module in upside-down. To do so will damage
the unit.
General Notes
• Before removing the control module from the unit, turn the
power off.
• Unless otherwise noted in the procedure, the module
referred to is the RH module. The switches referred to are
located on the RH module.
• Switch locations for the RH module are labeled on the
module diagram.• When rotating set knob fully in one direction, it will be
necessary to push the knob in at intervals to get the knob to
rotate completely.
• When taking readings with the set knob both in and out,
wait 20 seconds after taking the reading with the knob in
before taking the reading with the knob out.
• Recalibration of any of the controls may be necessary fol-
lowing component replacement.• Place the thermometer through the hole on the front of the
incubator floor. The thermometer should be lowered until it
touches the bottom of the humidity pan.
M d l 3029 d 3039 T bl h ti
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
27/52
Models 3029 and 3039_______________________________________________________________________Troubleshooting
Log RH readings on this table.
Display Data LimitsS2 SettingSilence/Set Button Out Silence/Set Button In
Your Results
1 80 through 90%, ±5% N/A
2 64.0 through 96.0 N/A
3 See temp setpoint N/A
4 40 0 h h 60 0 N/A
Log temp readings on this table.
Display Data LimitsS2 SettingSilence/Set Button Out Silence/Set Button In
Your Results
1 120.0 through 180.0 N/A
*2 64.0 through 96.0 N/A
3 -120.0 through –180.0 N/A
*4 -40.0 through –60.0 N/A
5 20.0 through 30.0 N/A
6 20.0 through 60.0
7 See thermometer N/A
8 -40.0 through –60.0 50.0 through 96.0
*Limits are for temp or RH.
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
28/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
29/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
30/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
31/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
32/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
33/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
34/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
35/52
Models 3029 and 3039_____________________________________________________________________Troubleshooting
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
36/52
11.4 CO2 System Troubleshooting Overview
a. General Instructions
Follow the instructions on the flow chart to diagnose prob-
lems with the CO2 controller. For this diagnostic aid to be effec-
tive, all checks must be made in the sequence indicated by the
flow chart and all data must be carefully recorded.
b. Stabilizing the Cabinet
Prior to beginning the procedure, shut the cabinet door andset the temperature and humidity at normal operating levels.
The temperature and humidity should be stabilized by Step #12
of the procedure.
c. CO2 System Troubleshooting Procedure
Turn the CO2 module off and remove the CO2 and logo
modules. Transfer the extender card from the logo module to
the CO2 module slot. Snap the extender card into the connector.Check the alignment. Add the CO2 module to the extender and
check the alignment.
Do not plug the module in upside-down. To do so will damage
the unit.
General Notes
• The first eleven steps of the procedure can be completed
with the temperature and humidity in an unstable condi-
tion.
• The first fifteen steps can be performed with the CO2
turned off.
• Before removing the control module from the unit, turn the
power to the module off.
• At various times in this procedure, the CO2 alarm may
sound. Unless otherwise noted, the alarms may be silenced
by pressing the alarm silence button.
• Unless otherwise noted in the procedure, the module
referred to is the CO2 module and the switches referred to
are located on the CO2 module.
Models 3029 and 3039_______________________________________________________________________Troubleshooting
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
37/52
Log CO2 readings on this table.
Display Data LimitsS2 SettingSilence/Set Button Out Silence/Set Button In
Your Results
1 15.0 through 25.0 N/A
2 approximately ½ reading 3 N/A
3 35.0 through 50.0 N/A
4 See Fyrite™ N/A
5 00.2 through 20.0 N/A
6 N/A N/A
7 96.0 through 140.0 N/A
8 -40.0 through –60.0 N/A
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
38/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
39/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
40/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
41/52
Models 3029 and 3039_______________________________________________________________________________Service
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
42/52
Section 12 - Service
12.1 Electrical Servicing
Access to all electrical components is provided by opening
the component compartment door on the left side of the cabinet.
The door is secured by two 1/4 turn fasteners.
The CO2, temperature, and humidity modules can be easily
removed. Turn the power to the particular module OFF. Pull the
black button out with a suitable tool to release the module, and
carefully ease the module out of the panel.The Model 3029 incubator has been designed so that no
internal electrical servicing should be required under conditions
of normal use.
If electrical servicing should become necessary, it should be
performed by qualified service personnel only!
For electrical information, consult the electrical schematic(Drawing #3029-50-0-D) and/or the wiring diagrams (Drawing
#3029-51-0-D, #3029-53-0-D) included at the end of manual.
For maximum safety, always disconnect the power before
servicing.
12.2 Temperature Sensor Replacement
The temperature sensor is located behind the left side duct
sheet, inside the incubator. To remove and replace the sensor:
1. Turn the power switch to OFF and disconnect the unit
from the power source.
2. Remove the shelves from the incubator. Loosen the two
screws that secure the floor of the incubator. Lift out the
floor and set it aside.
3. Carefully lift the left side duct sheet from its mounting
brackets and set it aside.
4. Locate the temp sensor (#245054 on drawing #3029-53-
0-D, Detail 1). It is mounted on the top rear of the left
interior wall.
5. Unscrew the two screws that attach the sensor to the
7. Set the temp sensor on the floor of the unit.
8. Open the side component compartment and locate the
motherboard (#285055 on drawing #3029-51-0-D).
9. Locate the temp sensor (plug #1 on the mother board).
10. Unplug the temp sensor, taking care not to pinch the
plug or removal will be difficult.
11. Carefully feed the temp plug and lead wire through the
access port and into the chamber. It may be necessary to
clear the Permagum from the port. Remove any tie
wraps to permit the plug to move freely into the cham-
ber. The Permagum should be replaced and the port
sealed after the sensor is replaced.
12. Remove the sensor from the chamber.
13. Replace the sensor with part number #245054 by revers-
ing the above procedure. Be sure to seal all screws with
silastic.
14. The temperature controller may need to be recalibrated
following this procedure. Perform a temperature calibra-
tion check (Section 7.5) to determine if recalibration is
necessary.
12.3 Replacing the Chamber Heater
The chamber heater is located under a protective plate
beneath the floor of the incubator. To remove and replace the
chamber heater:
1. Turn the power switch to OFF and disconnect the unit
from the power source.2. Remove the shelves from the incubator. Loosen the two
screws that secure the floor of the incubator and set the
floor aside.
3. Locate the heater (#600210 on drawing #3029-53-0-D).
It is mounted to the protective plate.
4. While holding the nut beneath the each screw, remove
the four screws from the top of the mounting plate.
Note: The heater will drop down from the mounting plate whenthe screws are removed.
5. Carefully pivot the heater out from under the bracket.
6. Remove the nuts from the heater connections and
remove the wires.
Models 3029 and 3039_______________________________________________________________________________Service
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
43/52
12.4 Replacing the Humidity Probe
The humidity probe is located inside the incubator beneath
the floor. To remove and replace the humidity probe:1. Turn the power switch to OFF and disconnect the unit
from the power source.
2. Remove the shelves from the incubator. Loosen the two
screws that secure the floor of the incubator and set the
floor aside.
3. Locate the humidity probe (#285053 on drawing #3029-
53-0-D, Detail 2). It is held to the float at the front of the
water pan.4. Remove the probe by pulling it upward and out of the
holder.
5. Unscrew the silver ground wire from the float mounting
bracket.
6. Feed the humidity probe through the bushing in the float
mounting bracket.
7. Open the side component compartment and locate the
motherboard (#285055 on Drawing 3029-51-0-D.)
8. Locate the humidity probe plug (plug #2 on the mother-
board).
9. Unplug the humidity sensor, taking care not to pinch the
plug as removal will become more difficult.
10. Carefully feed the humidity plug and lead wire through
the access port and into the incubator chamber.
Note: It will be necessary to clear the Permagum from the port
and clip any tie wraps to permit the plug to move freely into the
chamber. The Permagum should be replaced and the port sealed
after the probe is replaced.
11. Remove the probe from the incubator chamber.
12. Replace the probe with part #245053 by reversing the
above procedure.
13. When replacing the probe, be sure that the tip extends
only about 1/16” below the surface of the water.
14. The humidity controller may need to be recalibrated fol-
lowing this procedure. Make a humidity calibration
check. (See Section 8.8.)
12.5 Replacing the Immersion Heater
3. Locate the immersion heater (#610005 on drawing
#3029-53-0-D, Detail 2). It is clipped to the floor of the
humidity reservoir.
4. Remove the nuts from the clips securing the heater to
the floor.
5. Note the color of the heater lead wires.
6. Open the component compartment and locate the heater
wires (labeled Wet Heater on Drawing #3029-51-0-D).
Disconnect the heaters by removing the wire nuts. Be
sure to note the wiring configuration.
7. Carefully feed the heater wire leads through the access port and into the incubator chamber.
Note: It may be necessary to clear Permagum from the port and
remove any tie wraps that restrict passage of the heater wires.
The Permagum should be replaced and the port sealed after the
heater is replaced.
8. Remove the heater from the chamber.
9. Replace the heater with part number #610005 by revers-
ing the above procedure.
12.6 Replacing the CO2 Sensor
The CO2 sensor is located inside the incubator behind the
left duct sheet. To remove and replace the CO2 sensor:
1. Turn the power switch to OFF and disconnect the unit
from the power source.
2. Remove the shelves from the incubator. Loosen the two
screws that secure the floor of the incubator and set the
floor aside.
3. Carefully lift out the left duct sheet from its mounting
bracket and set it aside.
4. Locate the CO2 sensor (#245057 on Drawing #3029-53-
0-D, Detail 1). It is mounted near the center of the left
interior wall.
5. Remove the two screws that attach the sensor to the
wall. Note how the washers and tie wraps are posi-
tioned. Retain the screws and washers to mount the new
sensor. Remove the tie wrap holding the temp sensor
lead wire to the CO2 sensor block.
Note: It is very important that the tie wrap is replaced when the
Models 3029 and 3039_______________________________________________________________________________Service
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
44/52
10. Carefully feed the CO2 sensor plug and lead wire
through the access port and into the incubator chamber.
Note: It may be necessary to clear Permagum from the port and
remove any tie wraps that restrict movement of the lead wire.
The Permagum should be replaced and the port sealed after the
sensor is replaced.
11. Remove the sensor from the chamber.
12. Replace the sensor with part number #245057 by revers-
ing the above procedure. Be sure to seal all screws with
silastic.
13. The CO2 controller may need to be recalibrated follow-
ing this procedure. See Section 9.14 for the CO2 calibra-
tion check procedure.
12.7 Replacing the CO2 Solenoid Valve
The CO2 solenoid valve is located in the component com-
partment. To remove and replace the solenoid valve:
1. Turn the power switch to OFF and disconnect the unit
from the power source.
2. Open the component compartment and locate the sole-
noid valve (#250003 on Drawing #3029-51-0-D).
3. Follow the wire from the solenoid to the motherboard
and disconnect plug #3 from the board.
4. Remove the tygon tubing from both sides of the sole-
noid.
5. Remove the two screws securing the solenoid to the wall
of the incubator.
6. Remove the screws from the bottom of the solenoid
mounting and remove the solenoid. Note the position of
the energy-absorbing pads.
7. Replace the solenoid valve with part number #250003
by reversing the above procedure. Place the energy-
absorbing pads in their original positions.
12.8 Replacing the Blower
The blower is located inside the component compartment,
with a portion of it extending into the incubator chamber. Toremove and replace the blower:
1. Turn the power switch to OFF and disconnect the unit
from the power source.
2. Remove the shelves from the incubator. Loosen the two
screws that secure the floor of the incubator and set the
floor aside.
3. Carefully lift out the left duct sheet from its mounting
bracket and set it aside.4. Locate the blower wheel (#131006 on Drawing #3029-
53-0-D, Detail 1).
5. Loosen the blower wheel with an Allen wrench, and lift
the wheel from the shaft.
6. Open the component compartment and locate the blower
motor (#810700 on Drawing #3029-51-0-D).
7. Remove the four screws securing the blower and pull
the blower out of the incubator wall.
Note: The area around the blower shaft has been sealed with
silastic. As a result, the blower may be difficult to remove. It
may be necessary to use a razor knife to cut the silastic.
Remember to seal around the new motor with silastic to prevent
leakage from the incubator.
Do not cut, or otherwise damage, the blower mount gasket.
It will be necessary to reuse this gasket with the new blower motor.
8. Remove the wiring connections to the blower motor.
Note the positions of the wires.
9. Replace the blower motor with part number #810700 by
reversing the above procedure.
Models 3029 and 3039_________________________________________________________________________Parts List
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
45/52
Section 13 - Parts List
Model 3029
Qty Stock # Description
1 285812 Plug, 2 Pole, 3 Wire, 15 Amp, Nylon
1 950020 Cord Set, 16/3 SJT, 8 ft.
1 260039 Switch, Toggle, DPST
1 260042 Switch, Sealed Snap Action
1 270010 Thermostat, Low-temp, B-10
1 285055 Board, Circuit, Mother
1 285660 Knob, 1-1/4”, Blue w/ Clear Skirt
1 290113 Light, Pilot, Red, 5/16 Neon, 115V
8 124001 Guides, Card
1 285057 Extender, Module Board
1 510410 Grommet, Nylatch (for modules)
1 510411 Plunger, Nylatch (for modules)
1 231002 Module, CO2 Control
1 231003 Module, Temp Control
1 231004 Module, Humidity Control
1 131006 Wheel, Blower
2 136000 Heater, Exterior Door
1 205025 Filter, RFI/EMI
2 250013 Relay, DPDT, Enclosed
1 270080 Switch, Infinite Heat
1 285639 Switch, Dial Heat
1 285660 1-1/4” Knob
1 285798 Snap-In Outlet
1 285805 Receptacle, Door
9.5 400296 Gasket, Feather, Inner Door
Model 3029 (continued)
Qty Stock # Description
2.8 431125 Magnet, 1”
2.9 431124 Magnet, 1/2”
1 431144 Gasket, Outer Door
1 463028 Door, Glass, Inner
1 600210 Heater, 300W, Air
1 610005 Heater, Immersion, 300W
1 810700 Motor, Blower
1 950015 Cord, Electrical, Door, w/ Plug
1 890139 Gallon Bottle
1 60000 Clamp, Water Bottle
1 980055 Valve, Float
1 285020 Alarm, Sonalert, 110V
1 245057 Sensor Assembly, CO2, w/ Plug
1 250003 Valve Assembly, Solenoid, CO2
1 245054 Sensor Assembly, Temperature
1 245053 Sensor Assembly, Humidity
1 400008 Module, Power Supply
2 250021 Relay, Solid State, 10A, 240VAC
4 515081 Foot, Leveling, Adjustable
1 195105 Filter, CO2 Inlet
1 960001 Orifice, CO2 Injection
1 136005 Heater, Silicone Rubber, 6W, 120V
1 475576 Door Hinge, Interior
2 505546 Door Hinge, Exterior
6 503028 Shelf
Models 3029 and 3039_____________________________________________________________________________Parts List
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
46/52
Model 3039
1 950045 Cord, 16/3 SJT, 8 ft.
1 260039 Switch, Toggle, DPST1 260042 Switch, Sealed Snap Action
1 270010 Thermostat, Low-temp, B-10
1 285055 Board, Circuit, Mother
1 285660 Knob, 1-1/4”, Blue w/ Clear Skirt
3 290113 Light, Pilot, Red, 5/16 Neon, 115V
1 285057 Extender, Module Board
1 510410 Grommet, Nylatch (for modules)
1 510411 Plunger, Nylatch (for modules)1 231002 Module, CO2 Control
1 231003 Module, Temp Control
1 231004 Module, Humidity Control
1 131006 Wheel, Blower
2 136001 Heater, Exterior Door
1 205025 Filter, RFI/EMI
2 300065 Relay, DPDT, Enclosed
1 270090 Switch, Infinite Heat1 285639 Switch, Dial Heat
1 285660 1-1/4” Knob
1 285821 Receptacle, 125V, 15A
1 285805 Receptacle, Door
9.5 400296 Gasket, Feather, Inner Door
2.8 431125 Magnet, 1”
2.9 431124 Magnet, 1/2”
1 431144 Gasket, Outer Door 1 463028 Door, Glass, Inner
1 600211 Heater, 300W, Air
1 610006 Heater, Immersion, 300W
Model 3039 (continued)
Qty Stock # Description
1 810701 Motor, Blower 1 950015 Cord, Electrical, Door, w/ Plug
1 60000 Clamp, Water Bottle
1 980055 Valve, Float
1 285020 Alarm, Sonalert, 110V
1 245057 Sensor Assembly, CO2, w/ Plug
1 250003 Valve Assembly, Solenoid, CO2
1 245054 Sensor Assembly, Temperature
1 245053 Sensor Assembly, Humidity1 400008 Module, Power Supply
2 250021 Relay, Solid State, 10A, 240VAC
4 515081 Foot, Leveling, Adjustable
1 195105 Filter, CO2 Inlet
1 960001 Orifice, CO2 Injection
1 136006 Heater, Silicone Rubber, 6W, 120V
3 1210104 Resistor, 100K Ohm 1/4W 10%
1 1211822 Resistor, 8.2K Ohm 1/2W 10%1 505522 Latch Tab
1 505523 Latch Stud
1 890139 Gallon Bottle
1 60000 Clamp
1 34014 Plug 1/2”
1 34028 Cap Vinyl
1 34053 Plug 1-1/4”
1 515072 Latch, Pushbutton6 503028 Shelf
1 195125 Bacterial Air Vent Filter
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
47/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
48/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
49/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
50/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
51/52
-
8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy
52/52