User Guide to Operating the Zeiss Sigma 500VP SEM - User...User Guide to Operating the Zeiss Sigma...
Transcript of User Guide to Operating the Zeiss Sigma 500VP SEM - User...User Guide to Operating the Zeiss Sigma...
User Guide to Operating the Zeiss Sigma 500VP SEM
Nanoscale Fabrication and Characterization Facility, Petersen Institute of NaoScience and
Engineering, University of Pittsburgh, 348 Benedum Hall, 3700 O’Hara Street, Pittsburgh, PA 15261
Sample Preparation
Put on a pair of gloves. Samples should be clean, as small as possible, stable in vacuum. If needed,
clean samples, stubs and the specimen holder with isopropanol and kimwipe papers.
Mount fragment or powder samples firmly on Al-stubs using C-tape. Fix stubs onto the ZEISS
specimen holder. 9 stubs may be loaded. Mount thin sections directly on the specimen holder.
Make maps if necessary to be used to locate each sample in the SEM.
Warning Salt and oils from fingerprints will contaminate the FESEM vacuum and your samples if you
do not wear gloves. Fingerprints can be removed with isopropanol and kimwipe.
Starting ZEISS SmartSEM Program
First, the tool needs to be activated by logging onto the Zeiss SEM in FOM system.
The EM Server, implementing the internal communication between software and hardware, is always
running too. It is sometimes minimized to a small element (icon) on the right side of the Windows task
bar.
Note: If the EM Sever was closed by the last User, Starting the ZeissSmartSEM software will first
reload the EM server and recover software/hardware communication.
Double click on ZeissSmartSEM icon.
Alternatively, select Start/Programs/SmartSEM/SmartSEM User Interface. The EM Server Lon
On dialogue appears.
By logging, the SmartSEM user interface opens and is ready to operate the tool. By default, a TV view
inside the specimen chamber is shown.
The data zone is a special and useful group of annotation objects which are used to display useful
parameters. If it is not open, select View/Data Zone/Show Data Zone from the menu. Alternatively,
type <Ctrl+D> to toggle the data zone.
Loading the Specimen Holder
Make sure the specimen holder with samples fixed is ready.
(Set up SEM Controls in case it is not setup. Select Tools> Goto Control Panel from the menu to
display the SEM Control panel.)
Annotation bar
Tool bar
Status bar
Data zone
Use mouse to click VAC, and hit VENT, then select YES to the question window. N2 gas is being
introduced into the chamber. OR select Vacuum tab on the SEM Control panel. Click Vent. Click
Yes to confirm the Vent command.
Put on gloves. It takes about 2-3 minutes to vent. Wait and check
until the door can be opened.
Pull the chamber door open carefully and slide the holder onto
the stage.
When chamber open check that o-ring on the door is in place, if not place it in the grove
(must have gloves on while doing this).
Close the door immediately and hold it, then click VAC and hit PUMP. OR Click Pump in the
SEM Control panel.
Wait until vacuum gets to 5∙10-5
Torr before turning EHT on.
Switch ON the EHT
‘EHT’ means the extra-high tension acceleration voltage. This voltage must be applied to the gun to
make the gun emit electrons.
Confirm that EHT Vac ready = Yes is displayed on the Vacuum tab of the SEM Control panel.
Select the Gun tab on the SEM Control panel.
Set the acceleration voltage:
Double-click in the EHT Target = 10 kV.
Enter the required acceleration voltage in the
EHT Target field, for example 10kV,
and click OK.
Switch ON the EHT:
Click EHT in the status bar.
Select EHT on from the pop-up menu.
The EHT switches on and increases to 10 kV.
The accelerating voltage ranges from 200V to 30 kV. Use 10 kV for imaging, 15 or 20 kV for EDS
at the beginning, then tune EHT when needed for different applications.
Position one of your samples below the objective lens using the joy stick with caution. Take great
care not to hit the lens or the detectors. If a detector is damaged by a user, he/she will be billed for
replacement parts.
Once focused, the WD (working distance between the sample surface and the low portion of the
lens) should be greater than 4 mm.
Generating an image
Select the Detector tab on the SEM Control panel.
Select InLensDuo or SE2 detector to display live image.
Find your sample.
(Zeiss recommends to select the SE2 detector to obtain the first image. This is because the
SE2 detector provides a good signal-to-noise ratio even at long working distances.
A few parameters (Magnification / Focus, Brightness / Contrast, Stigmation, Aperture
Alignment) need to be adjusted in order for the SEM to give sharp and high contrast images at a
given EHT and WD. Always do corrects and adjustments at a higher magnification, then acquire an
image at a lower magnification.
Stage settings and motions
The Stage Navigation window can be used to initiate XY movement of the stage by double-clicking
on the sample holder top view. Be sure to select the correct holder from the list and to adjust the
zoom view sliding bar to your needs. All five axes coordinates are reported in this window in order
to save specific positioning on the sample if needed.
The STOP button on the Stage Navigation window or the Break button on the dual joystick can be
clicked at any time to quickly stop any engaged motion of the stage.
Finding your way on the sample is mainly done by the use of the dual joystick panel, where all five
axes can be actuated.
Saving the image
Select the Scanning tab on the SEM Control panel.
Select Freeze on = End Frame in the drop-down list.
Click Freeze.
(A red dot near the lower right corner of the image area
indicates that the image is frozen.)
Select File > Save Image from the SmartSEM menu.
Enter a path and a file name.
Click Save…tif to save the image file.
To continue imaging, select Image > Unfreeze from the SmartSEM menu, or click Unfreeze on
the Scanning tab.
Taking sample OUT
Go to the ChamberScope view.
If Variable Pressure (VP) mode is used, go back to High Vacuum (HV) mode from VP mode.
Click EHT and hit EHT off to turn off the accelerating voltage.
Or go to Gun Beam State = Beam Off EHT off
Move the stage down first. Set the stage back to X=50, Y=50, Z=1, T=0, R=0 using Stage at the
SEM Control window. (Stage automatically will go down to a safe Z position)
Click VAC and hit Vent to vent the SEM specimen chamber.
Put on a pair of gloves.
Once the chamber vented, open the door and remove the specimen holder off the stage.
Close the chamber door immediately. Then click VAC and hit Pump. The chamber must be
pumped down at the end of a session.
Wait for a green tick for VAC.
Go to File and click Logoff. Hit Yes to the two questions.
CHOOSING AN IMAGING DETECTOR, ADJUSTING B&C
Under the SEM Control Panel, select the “Detectors” tab
Choose the desired detector from the drop-‐down menus. Alternatively you could use the data zone to
change signal A.
Detectors
InLensDuo detector is a high resolution secondary electron detector. Below about 1000x
magnification, there will be significant detector artifacts. The detector is top down looking so you lose
shadow contrast making the sample look flatter than it is. Operation is restricted to high vacuum mode
and accelerations ≤ 20kV. Typical parameters are 10kV with a 30µm aperture and a WD of 3 – 5mm.
Acceleration of .2 – 20kV is possible. Working distance is limited by acceleration but 3 – 10mm can
be useful.
SE2 is useful for topographic imaging in high vacuum mode at working distances >4mm. Detector is
for general propose imaging (low-magnification, high-magnification, SE and BSE imaging). Typical
starting acceleration is 5kV, with a working range of 1 – 30kV.
QBSD back-scatter electron detector is good for compositional contrast imaging. WD 6-15mm, EHT:
5kV-30kV.
VPSE G4 detector is for variable pressure imaging of non-conductive samples. Image information is
similar to SE, but areas of the sample with higher charge levels will be brighter. Typically start with a
detector bias of 85%, 40pa N2, Acceleration ≥ 15kV, 60µm aperture in high current mode.
Acceleration from 8 – 30kV can be useful. Working distance is typically 5 – 15mm. Pressure in the
range of 10 – 80pa is typically useful. In high vacuum mode some samples with have a CL signal that
can be used to understand variations in elemental distribution that cannot be seen in BSD or EDS.
CCD is used to observe the chamber when moving the stage or closing the door.
Adjusting brightness and contrast (BC) can be accomplished in many ways on our Sigma. The
preferred method of most users is the Brightness and Contrast knobs on the control panel. However,
under the scanning tab it is possible to set Auto-‐brightness and contrast. This feature tends to be
annoying since it is continuously trying to optimize BC, but it is useful to get a starting point. Set
“Auto BC = Off” once you have a starting point. The line scan feature found on the scanning tab can
be a useful feature to aid optimization of BC.
NOISE & CHARGE REDUCTION
Basic noise reduction modes:
• Scan speed – balance S/N with charging and acquisition time
o Higher number is slower (note cycle time)
• Line average
o Repeats the line N times at the given scan speed
• Frame average (good for charge reduction at low mag)
o Repeats the whole image N times at the given scan speed
• Note: “Avg” mode cycles continuously; “Int” mode completes just one frame and blanks
beam.
• You usually want to set freeze on end of frame so you do not end up with scan artifacts in
your saved image.
• Generally Scan 4 – 8 with line average N=4 is a good starting place.
• If you have charging, try a faster scan speed with lots of frame averaging or frame integration.
COMMONLY USED HOT-KEY
Generally, it is more productive to use hot keys so you can keep your hands on the control
panel. Many users reject use of hot keys, and prefer to use menus. The table below is a list of
commonly used hot keys.
Hot Key
Action
Tab
Switch between course/fine adjustments
Ctrl + Tab
Center image (to cancel, simply right-‐click mouse)
Ctrl + Shift + Tab
Select region of interest (ROI) to image (centering/magnification to ROI)
Ctrl + e:
Save image (opens save dialog to select save folder, enter sample
ID, and set indices) F4
Cycles forward through magnification table
Shift + F4
Cycles backward through magnification table
Shift + F2
Lens hysteresis removal
Ctrl + F
Auto focus (not recommended, but it works on some samples)
Ctrl + S
Auto Stigmator (not recommended, but it works on some samples)
Zeiss-specific recipes – starting conditions for conductive specimens
(high vacuum mode)
Information required Suggested working conditions
(HC = high current mode)
Detector EHT Apt /HC WD
Basic topographic imaging SE2 10-‐20kV 30µm 10mm
No HC
High resolution imaging InLensDuo 20kV 30µm 5 mm
No HC
Fine surface details InLensDuo 2kV 30µm 5 mm
SE2 No HC
beam sensitive specimen SE2 2kV 20µm 5mm
No HC
Maximum depth of field SE2 10-‐20kV for
high-‐Z spec
2-‐5kV for low-‐Z
10µm
HC on
10mmforhi-‐kV
6mm for low kV
Compositional contrast imaging,
conductive
CZ BSD, 4 10-‐20 kV 60µm ~8.5mm
spec segments active HC on
Surface topography with
composition
CZ BSD, 5 10-‐20 kV 60µm 10mm
information segments active HC on
Trace element variations CL detector 10-‐20 kV 60µm 10mm
(VPSE) HC on
nts
Zeiss-specific recipes – starting conditions for non-conductive specimens
Information required Suggested working conditions
(HC = high current mode, HV = high vacuum
mode, VP = variable pressure mode)
Detector Vacuum EHT Apt
/HC
WD mode
Basic topographic imaging VPSE or
CZ BSD, 5
segme
VP
P = 40 Pa
20kV 60µm
HC
on
15mm
High(er) resolution, VP mode VPSE VP 20kV 30µm 8 mm
P = 20 HC on
Fine surface details SE2 HV 2kV 30µm 5 mm
No HC
Beam sensitive specimen SE2 HV 2kV 20µm 5mm
No HC
Maximum depth of field -‐ VP mode VPSE VP 20kV 30µm 15mm
P= 40 HC on
Compositional contrast imaging CZ BSD, 4 VP 20 kV 60µm ~8.5mm
segments P = 40 HC on
active
Surface topography with composition CZ BSD, 5 VP 20 kV 60µm 10 -‐ 15mm information segments P=40 HC on
active