User Guide KI 734x Series Two way Loss & ORL Tester . www.kingfisher.com.au. Two-way Loss Tester. T he fastest and easiest two-way loss tester Real time display of bi-directional average multi-  link loss on both instruments 3 year warranty & calibration cycle No warm up required - PowerPoint PPT Presentation

Transcript of User Guide KI 734x Series Two way Loss & ORL Tester

Page 1: User Guide  KI 734x Series  Two way Loss & ORL Tester

User Guide KI 734x Series

Two way Loss & ORL Tester

www.kingfisher.com.au

Page 2: User Guide  KI 734x Series  Two way Loss & ORL Tester

Two-way Loss Tester

The fastest and easiest two-way loss tester

Real time display of bi-directional average multi- link loss on both instruments 3 year warranty & calibration cycle No warm up required 360 hours battery life & interchangeable connectors 1200 storage space for dual test results

Page 3: User Guide  KI 734x Series  Two way Loss & ORL Tester

Fast & Easy For Bi-directional Testing

Step 1: Set referencesStep 2: Press the “Auto test” button

Real time bi-directional average multi- link loss display on both instruments

Page 4: User Guide  KI 734x Series  Two way Loss & ORL Tester

Step 1 - Set Reference At Both Ends

Two-way port

Meter port

Test lead

End A

End B

An optic fiber link

Page 5: User Guide  KI 734x Series  Two way Loss & ORL Tester

Source off

Source Return Loss Power Meter Auto test

- 52.73 dBm

1310nm

Source on

Auto test Menu Abs / Rel Hold

- 4.01 dBm

1310nm - +

-4.00 dBm

Source on

Auto test Menu Set Ref Abs / Rel Hold

- 0.00 dB R

1310nm - +

-4.00 dBm

Source off

Source Return Loss Power Meter Auto test

- 52.73 dBm

1310nm

Pushing “AUTOTEST”

Pushing “Abs / Rel”

Alternates between 1310 and 1550 nm if 1310 & 1550 nm lasers are installed

Holding “Set Ref” for 3 seconds to set reference for multi-

Pushing “Menu” to be back to the turn-on display

Set Reference Key Press Sequence

Page 6: User Guide  KI 734x Series  Two way Loss & ORL Tester

An optic fiber link

Step 2 - Measure Insertion And Return Loss

End A

End B

Source off

Source Return Loss Power Meter Auto test

- 52.73 dBm

1310nm

Pushing “Auto test” at one End.

Page 7: User Guide  KI 734x Series  Two way Loss & ORL Tester

Source on

Auto test Menu Hold

- 34.011310nm - +

-4.00 dBm

Alternates between 1310 and 1550 if 1310 & 1550 nm lasers are installed

- + Source on

Auto test Menu Hold

- 54.17 - +

-4.00 dBm

1310nm

Testing is completed for the link.

A ◄► BdB

A ◄ BdB RL

STORE

Pushing “STORE” to store average Loss, A-to-B Loss, B-to-A

Loss, ORL at end A, ORL at end B of multi-

Displays the average bi-directional link loss of multi-Displays Loss/Optical Power from A-to-B (‘A ►B’)Displays Loss/Optical Power from B-to-A (‘A◄ B’)

Displays ORL at A end (‘A ►’)Displays ORL at B end (‘◄ B’)

Back to the average bi-directional link loss display

Press ‘Select’

Press ‘M

enu’Measure Insertion And Return Loss -

Continue

Page 8: User Guide  KI 734x Series  Two way Loss & ORL Tester

End A

End B

An optic fiber link

Communication During Testing

Source on

Auto test Menu Hold

- 34.011310nm - +

-4.00 dBm

A ◄► BdB

After completing one link test,

pushing “Hold” will freeze thedisplay on both units and the

unitat End B beeps twice. This

tells End B that a test has been completed.

Source on

Auto test Menu Hold

- 34.01 - +

-4.00 dBm

A ◄► BdB

Then, End B connects to the next link and pushes “Hold”. This

releasesthe display on both units and the

unit at End A beeps once. This tells

End A that he is ready for the next test.

Increases the productivity in high fiber-count systems.

Page 9: User Guide  KI 734x Series  Two way Loss & ORL Tester

Set Ref In Autotest

Source on

Auto test Menu Set Ref Hold

- 34.011310nm - +

A ◄►BdB

Displays average link loss for multi-.Hold “Set Ref” for 3 seconds will set references on both units for multi-.

Page 10: User Guide  KI 734x Series  Two way Loss & ORL Tester

View Ref Value In Autotest

Source on

Auto test Menu Abs / Rel Hold

- 34.011310nm - +

A ►BdB R

Displays link loss and reference value for multi- for A-to-B.

Source on

Auto test Menu Abs / Rel Hold

- 32.241310nm - +

A ◄ BdB R

Displays link loss and reference value for multi- for B-to-A.

-8.20 dBm

-7.80 dBm

Displays the average bi-directional link loss of multi-Displays Loss/Optical Power from A-to-B (‘A ►B’)Displays Loss/Optical Power from B-to-A (‘A◄ B’)

Displays ORL at A end (‘A ►’)Displays ORL at B end (‘◄ B’)

Back to the average bi-directional link loss display

Press ‘Select’

Press ‘M

enu’

- +

Unit A reference value

Unit B reference value

Page 11: User Guide  KI 734x Series  Two way Loss & ORL Tester

Pass/Fail Assessment

In Two-way testing mode, press ‘Level’ to enter Pass/Fail display mode.To exit press ‘Level’ again, or press ‘Select’.

Source on

<<< Auto test Menu Hold

PASS1310nm - +

A

Level

Select

Page 12: User Guide  KI 734x Series  Two way Loss & ORL Tester

Insertion Loss Pass/Fail Threshold Set UpMulti- of Insertion Loss Pass/Fail criteria can

be set.

Shift Default Level

Cancel Set

Shift Source off

SEt Menu

15.731310nm - +

dB

Select another two-way operational wavelength to set up Average Insertion Loss Pass/Fail threshold.

Enter Power Meter mode.Select two-way operational wavelength.

Use ‘Shift’ + ‘Level’ to set Average Insertion Loss Pass/Fail threshold, push ‘Set’ to store.

- +

Page 13: User Guide  KI 734x Series  Two way Loss & ORL Tester

Return Loss Pass/Fail Threshold Set UpMulti- of Return Loss Pass/Fail criteria can be

set.

Shift Default Level

Cancel Set

Select another wavelength to set up Return Loss Pass/Fail threshold.

Use ‘Shift’ + ‘Level’ to set Return Loss Pass/Fail threshold, push ‘Set’ to store.

- +

Shift Source off

SEt Menu

-10.571310nm - +

dB RL

Enter Return Loss mode.Select wavelength.

Page 14: User Guide  KI 734x Series  Two way Loss & ORL Tester

The Contents of User Memory

Average loss for multi-

Reference of both units for multi-

ORL for multi-

Nominal output power for multi-

Location info by flashing symbol of either A or B

Page 15: User Guide  KI 734x Series  Two way Loss & ORL Tester

KITS™ testing software