Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

26
Ultra-long MicroStrip Ga s Counter for Spallation Neutron Source Facilitie s Hiroyuki Takahashi, Prasit Siritiprussamee, Kaoru Fujita, Hisako Niko, Masaharu Nakazawa, Shunji Kishimoto*, Takashi Ino*, Michihiro Furusaka*, Hidenori Toyokawa** University of Tokyo, KEK*, JASRI**

description

Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities. Hiroyuki Takahashi, Prasit Siritiprussamee, Kaoru Fujita, Hisako Niko, Masaharu Nakazawa, Shunji Kishimoto*, Takashi Ino*, Michihiro Furusaka*, Hidenori Toyokawa** University of Tokyo, KEK*, JASRI**. Purpose. - PowerPoint PPT Presentation

Transcript of Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Page 1: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Ultra-long MicroStrip Gas Counter for Spallation Neutron Source F

acilities

Hiroyuki Takahashi, Prasit Siritiprussamee, Kaoru Fujita, Hisako Niko, Masaharu

Nakazawa, Shunji Kishimoto*, Takashi Ino*, Michihiro Furusaka*, Hidenori Toyokawa**

University of Tokyo, KEK*, JASRI**

Page 2: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Purpose

• High-counting rate and high-resolution neutron detectors are required for the J-PARC neutron scattering facility.

• Low-cost reliable modules for neutron spectrometers

Page 3: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Reconsideration of Linear PSD

• Fast He-3 tubes

• Replace anode wires with high-tech plate

• Easy to maintain

図Tube(diameter:6-10mm)

1D-MSGC

sealed gas mixture

(thickness:0.5-1mm)

600mm

MSTube

Page 4: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

• Simple replacement of anode wires with anode strips is easy, but

• Take advantage of microstrip plate– Cathode signal is usable.

RESISTIVE CHARGE DIVISION METHOD is not compatible withHigh counting rate operation

Page 5: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

The use of cathode signal to position encoding

anode

cathode1-oddgrid

grid

cathode1-even

cathode2-odd

cathode2-even

perform a rough charge division for global positioning

incident position

global positioning

Page 6: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

8cm long test plate

1.6mm

0.8mm

Page 7: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

0

0.2

0.4

0.6

0.8

1

1.2

0 0.4 0.8 1.2 1.6 2 2.4 2.8 3.2 3.6

prpbability determined to cathode#1probability determined to cathode#2

position(mm)

probability

Electrode pattern

CF4 4bar3He 0.3bar

Neutron beam test

1.8mm

Distortion freeHigh resolution

Page 8: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Trace (800um)

a coarse position by using charge division method

Cathode modulation

Page 9: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

• Resistive charge division requires high gas gain

• Single-end packaging is preferable

• Extension of cathode signal encoding method with GLG method

Page 10: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Global Local Grouping (GLG) Method

Separate total charge into two parts: Global and Local

Global information is used to obtain coarse position

Local information is used to obtain

fine position

Global Information

1 2

0 1 2 3 4 5 6 7 8 9

Local Information

Page 11: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

New Micropattern Global-Local Char

ge Division

Let us assume Total length: 640 mm S/N = 40:1

Achievable resolution = 640/40 x= 16mm

Half cathode S/N = 20:1 x= 32mm

The other half cathode S/N = 20:1 x = 1.6mm

x 10 resolution is expected

STEPWISEmodulation

ZIG-ZAGmodulation

Zig-Zag + Stepwise -> Pseudo gray code encoding

Page 12: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

S/N RATIO

RE

SO

LUT

ION

(m

m)

ACHIEVABLE RESOLUTION vs S/N RATIO

EFFECTIVE REGION

Page 13: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Test plateSensitive length ~ 648 mm.2 or 4 Anodes (3mm pitch) and 4 grids.Global pitch ~ 35mm.Global – Local pad layout

700mm

800m

Global Local

UP TO ~3m is possible

Page 14: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

First lot had high anode resistance problem

0

1020

30

40

5060

70

8090

100

0 10 20 30 40 50 60 70

Position (cm)

Gas

gai

n (%

)

17% of gain difference.

Rise time greatly varied with position (from 0.3 to 5us.).

Reduced to 1/5 in the second lot

Page 15: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Synchrotron Experiments

Tested at KEK Photon Factory, BL-14A.bBeam energy 8keV.Beam size 400um.

Page 16: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Energy resolution

8keV X-rays

Gas gain: 6000

13.3% (FWHM)

Page 17: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Global position spectrum

68mm FWHM

Page 18: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Local scan test 4mm pitch

1-4 mm (FWHM) resolution

Page 19: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Global - Local relationship

LOCAL

GLOBAL

Page 20: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Improvement of the detector

• Divide global to ½ x 4 resolution

• 2nd lot

Page 21: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

FIRST NEUTRON BEAM TESTwith second lot

• 3He 0.5bar + CF4 2.5 bar

GAS GAIN > 300

Page 22: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

CHARGE SENSITIVE AMPLIFIER OUTPUT SIGNAL

Page 23: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

32mm

GLOBAL POSITION TEST

Page 24: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

8mm

LOCAL POSITION TEST

Page 25: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities

Summary

• A new 1-D microstrip gas counter using global-local charge division method is developed.b

• 64cm long tube is fabricated and successfully tested with a neutron beam.

Page 26: Ultra-long MicroStrip Gas Counter for Spallation Neutron Source Facilities