TS Series Installation, Operation & Parts Manual Operation & Parts Manual ... TS30A 200 GPM 760 lpm...
Transcript of TS Series Installation, Operation & Parts Manual Operation & Parts Manual ... TS30A 200 GPM 760 lpm...
TS Series V Assy.
Anodized Aluminum, Mechanical Drive
Manual FPPMSV02, Rev. 2010-12
Installation, Operation & Parts Manual Flow meter P/No.: TS_________________ Flow meter S/No.: __________________
The flow meter shipped under this general manual, has specifically:
MODEL ACCESSORIES REGISTER*
TS10A none none
TS15A Strainer, standard Register (7887 or 7886)
TS20A Strainer, High Capacity with 10:1 pulser
TS30A Air Eliminator with 100:1 pulser
Backpressure Valve Printer (7888)
Air Check Valve Preset Counter (7889)
Preset Valve/Linkage with Micro Switch kit * Register manuals published by Veeder-Root
Internals: Rotors: PPS/carbon LV HT HV
PPS/Teflon LV HT HV
Seals : Viton Teflon
Tuthill Transfer Systems, 8825 Aviation Drive, Fort Wayne, IN 46809, USA Tel: 260-747-9060 Fax: 800-866-4861 www.tuthill.com [email protected]
INDEX Section . Page FPP Meters and Principle of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 3
Material Specifications, Capacity & derating on high viscosity liquids . . . . . . . . . . . . 1 . 4
Part No. and Assy No. break-down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 6-7
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 8
Installation, Start-Up & Operation Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 8-9
Reversing Direction of Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 10
Flow meter Calibration & Re-Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 11-12
Servicing the Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 13-15
Troubleshooting the Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 17
Flange kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 16
Parts List, models TS10A, TS15A, TS20A & TS30A . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 18
Exploded View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 19
Torque & Tools chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 20
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 20
Strainers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 . 1
Parts List, standard Strainer (2” & 3”) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 . 2
Parts List, High Capacity Strainer (2” & 3”) . . . . . . . . . . . . . . . . . . . . . . . . . 2 . 3
Air Eliminator, incl. Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 . 4-5
Backpressure Valve, incl. Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 . 6
Air Check Valve, incl. Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 . 7
Name of Distributor who supplied this flow meter: ________________________________ Phone number for Distributor Service Department: ________________________________
1 . 2
Meter S/No. is on the spec plate, which is located on the side of the Right Angle Drive register adaptor.
Register S/No. is on the register spec plate, which is on the side of the register.
PRINTER REGISTER PRESET COUNTER PRESET VALVE
AIR ELIMINATOR
STRAINER
BACKPRESSURE VALVE
(not visible)
Serial No. Terminology This flow meter is the subject of a continuing improvement program. The possible components in this flow meter are: To ensure correct supply of replacement parts, every parts inquiry must include the Serial No. from the original flow meter assembly.
FLOW METER
REGISTER AIR CHECK VALVE
AIR ELIMINATOR
STRAINER
About FPP Meters We thank you for purchasing an FPP Meter for liquid meas-urement service. FPP Meters, formerly Fluid Power Prod-ucts, is now a trade name of Tuthill Transfer Systems. FPP was established in 1980. Since its inception, the company has been dedicated to manufacturing cost-effective, high performance flow metering devices for petroleum, industrial, commercial and municipal service.
Our facilities include computerized order entry and inventory control, so that you are assured of accurate and prompt de-liveries. Furthermore, our production personnel ensure that each order, regardless of size, receives individual attention.
Constant attention to new product development and produc-tion design, our high standards of manufacture and final testing are the reasons why FPP Meters meet your most demanding requirements. With the ‘Waveform’ oval gear (2003 patent) meter accuracy is better than ever before.
Our meters are used in batching, blending, process control and to dispense fluids in liquid handling facilities throughout the World. Service includes gasohol blends, bio-diesel and special formulation racing fuels.
Principle of Operation Positive Displacement meters have a measuring chamber, where inlet & outlet are separated by rotors, a rotating ele-ment or sliding vanes. As the liquid passes through the flow meter, it causes the rotors/element/vanes to turn, which forms the basis for volumetric measurement.
The Oval Gear metering principle is based on two elliptical (oval) gears, which turn on center on two horizontal shafts inside a measuring chamber formed by two overlapping cyl-inders. The oval gears have meshing teeth along their en-tire circumference, ensuring that the gears will maintain cor-rect position in relation to each other at all times, without the use of timing gears.
The volume being transferred from the inlet to the outlet side (= volume meas-ured), forms between the oval gear and the side of the measuring cham-ber, alternately in the upper and the lower half of the measuring chamber.
In a full 360° rotation of the gears, four such known volumes are released to the downstream side of the flow meter.
With precision machining and close internal tolerances, the slippage is minimal for superior linearity (accuracy) over a broad turn-down ratio. Oval gear meters are largely unaf-fected by changes in liquid viscosity (see page 1.8).
The TS Series meter is designed to provide:
V●● Assy: Drive to a mechanical register (this manual).
E & F●● Assy: Electric pulse signal to an electronic register installed on the flow meter.
S & W●● Assy: Electric pulse signal to a remote electronic register.
In the two electronic versions, the flow meter is 100% gland-less with static O-ring seals only. A pulser (pulse pick-up) detects the movement of internal magnets, to generate the signal to the electronic register. Several types of pulse sig-nal are available to ensure signal compatibility with all elec-tronic registers.
Fluid Compatibility FPP Meter flow meters are available in anodized aluminum and stainless steel, with a variety of rotor types, and seal & bearing materials, to ensure compatibility with a broad range of liquids and operating conditions. If in doubt about com-patibility with a specific fluid or the operating conditions, please refer to FPP210 (Liquid Compatibitily list) or FPP200 (FPP Technical Manual). If those publication do not provide a clear answer, please consult with your author-ized FPP Meter distributor, or Customer Care at the factory.
While most refined petroleum products can be handled with the same flow meter, some require different rotors and/or seals. Do not change service liquid, without consulting with your authorized FPP Meter distributor. Model Nominal Capacity TS10A 40 GPM 150 lpm TS15A 60 GPM 230 lpm TS20A 150 GPM 570 lpm TS30A 200 GPM 760 lpm
1 . 3
TS Series, Principle of Operation
Only 2 moving parts. Patented ‘Waveform’ oval gears = sustained accuracy with a minimum of maintenance. No metal-to-metal contact in measuring chamber or in bearings. The lowest differential pressure values amongst rotary PD meters.
=> Lower Cost of Ownership!
Normal turn-down is 10:1. Operation in excess of nominal capacity will cause excessive wear or premature failure. Meters can be calibrated for use below 10%, if the flow rate is steady or viscosity > 50 cSt.
1 . 4
TS Series, Anodized Aluminum - Specifications for V●● assemblies
Materials Case : Anodized aluminum Rotors : PPS with carbon bearings, (oval gears) TeflonTM bearings optional. Posts : HC303SS Seals : VitonTM standard, TeflonTM optional. Connections Bolted companion flanges with NPT or BSP port; Optional: Anodized aluminum welding flanges Carbon steel welding flanges Carbon steel 150# RF ANSI adaptors Pressure Rating at 100°F (38°C) with 3:1 Safety Factor Meter/Strainer : 150 PSI/100°F = 10.3 BAR/40°C Air Elim/Valves: 150 PSI/100°F = 10.3 BAR/40°C Temperature Rating V assy : -40°F to +180°F = -40°C to +80°C Use HT or HV rotors if temp. can exceed 120°F = 50°F Meter Performance When field calibrated on actual liquid of service Linearity Over 5:1 turn-down from max. capacity : ±0.175% Over 10:1 turn-down from max. capacity : ±0.30% Over 30:1 turn-down from max. capacity : ±0.5% Repeatability 0.05% or better under constant conditions Nominal Capacity With LV rotors on 1-300 cP viscosity: With HV rotors on 5-1000 cP viscosity:
GPM LPM m3/h TS10A 40 150 9 TS15A 60 230 14 TS20A** 150 570 34 TS30A** 200 760 45
** = On low viscosity lubricating liquids, intermittent use to 125% of capacity is permitted. Viscosity With LV rotors 100% capacity to 1,500 SSU/300 cP. With HV rotors 100% capacity to 5,000 SSU/1,000 cP. With HV rotors, the flow meter may be used to
1,500,000 SSU/350,000 cP at reduced flow rates (see meter viscosity/coefficient table in next column).
Opt. Pulsers on mechanical register 10:1 pulser 100:1 pulser 1/10 gallon register 10 PPG 100 PPG Whole gallon register 1 PPG 10 PPG 1/10 liter register 10 ppl 100 ppl Whole liter register 1 ppl 10 ppl Dekaliter register 0.1 ppl 1 ppl
●
●
●
●
Rotor
B L I C M J CodeSSU cSt LV HT HV LV HT HV TYPE
32 1 1.00 1.00 1.00 1.00 1.00 1.0060 10 1.00 1.00 1.00 1.00 1.00 1.00
240 50 1.00 1.00 1.00 1.00 1.00 1.00465 100 1.00 1.00 1.00 1.00 1.00 1.00925 200 1.00 1.00 1.00 0.90 0.90 1.00
1,415 300 0.86 0.86 1.00 0.73 0.73 0.981,890 400 0.77 0.77 1.00 0.62 0.62 0.962,360 500 0.71 0.71 1.00 0.57 0.57 0.942,830 600 0.66 0.66 1.00 0.53 0.53 0.923,300 700 0.63 0.63 1.00 0.50 0.50 0.903,775 800 0.60 0.60 1.00 0.48 0.48 0.854,250 900 0.56 0.56 1.00 0.45 0.45 0.804,720 1,000 0.54 0.54 1.00 0.43 0.43 0.759,440 2,000 0.35 0.77 0.28 0.65
14,150 3,000 0.28 0.65 0.22 0.5518,875 4,000 0.24 0.58 0.19 0.4623,600 5,000 0.19 0.53 0.15 0.4228,325 6,000 0.16 0.49 0.13 0.3933,050 7,000 0.14 0.47 0.11 0.3737,750 8,000 0.12 0.44 0.09 0.3542,475 9,000 0.10 0.42 0.08 0.3447,200 10,000 0.41 0.3294,400 20,000 0.30 0.24
141,600 30,000 0.24 0.19188,800 40,000 0.20 0.16236,000 50,000 0.18 0.14283,200 60,000 0.17 0.13330,400 70,000 0.14 0.11377,600 80,000 0.13 0.10424,800 90,000 0.12 0.10472,000 100,000 0.11 0.09944,000 200,000 0.08 0.06
1,416,000 300,000 0.07 0.051,888,000 400,000 0.06 0.052,360,000 500,000 0.06 0.042,832,000 600,000 0.06 0.043,304,000 700,000 0.05 0.043,776,000 800,000 0.05 0.044,248,000 900,000 0.05 0.044,720,000 1,000,000 0.05 0.04
Intermittent service = < 6 hours/day, and deliveries not to exceed 15-20 minutes.On higher viscosity liquids, model nominal capacity is reduced as per tables below .
Liquid Viscosity
Flow Meter CoefficientCarbon brgs. Teflon brgs.
Normal operating range is w ith 10:1 turn-dow n from model nominal capacity.
Flow Meter nominal capacity:
On low viscosity liquids, best performance (linearity/service life) is betw een 50% and 85% of model nominal capacity.On diesel fuel, TS20A & TS30A models w ith mechanical register only may be operated to 125% of nominal capacity in intermittent service (this does NOT apply to models with preset and/or printer).
See cover for rotor & seal materials in this meter.
1 . 6
Flow Meter Part No.
Flow Meter Part No. Most flow meter assemblies can be fully defined with a 10 position Part Number. Only when non-standard options are se-lected (pulser, seals, non-standard rotors or strainer basket), does it become necessary to extend the Part No.:
Position 1-5 = Model No. Capacity & Materials
Position 6-8 = Assy. No. Register & Accessories, refer to table on page 1.7.
Position 9-17 = VariablesVariablesVariables See codes below.
MECHANICAL, 2010 P/No. overviewTS Series
RO
TOR
S
STR
AIN
ERM
esh
size
PULS
ER
SEA
LS
CA
LIB
R.
FLA
NG
ES
MODEL No. ASSY No.
Area Shaded YELLOW = non-std => Longer Lead Time!
Whole Imp gallon, not on TS10 or TS15
Rotor/Bearing materials & Rotor type
U 1/10 Imp gallon, not on TS30
100 mesh, optional (2" only)Std for HT service ('YELLOW' in p/l) 0PPS/carbon, HVPPS/carbon, HT
8 80 mesh, optional (3" only)2 20 mesh, std/HV rotors
1211
Std. Strainer (blank if 40 mesh)
40 mesh, std/LV & HT rotorsStrainer BASKET
See Assy Chart
4
SDekaliter, not on TS10 or TS15
AA welding flanges
H Hi Cap strainer
1/10 liter, TS10 onlyL
8
C18
Whole liter, not on TS10
Flanges
150# RF ANSI adaptors, CS
CS welding flangesH
Calibration for mechanical register
S 2 0 A V _
AB
NPT companion flangesBSP companion flanges
A Anodized Aluminum
T
RO
TOR
S
If all STANDARD selections, these identifiers are not filled in.
PPS/Teflon, HT
M14 15
PPS/Teflon, LV
1/10 gallon, not on TS30
STR
AIN
ERM
esh
size
5 6 7 17
PULS
ER
SEA
LS
CA
LIB
R.
FLA
NG
ES
MODEL No. ASSY No.
131 2 3 4 10
_169
100:1 solid stateCF
XDS
none (standard)10:1 dry reed
M
LIC
Standard = 'GREEN' price in price lists.
Teflon,
I
B PPS/carbon, LV
D
GT
Whole gallon, not on TS10 or TS15Y
Std for HV service ('BLUE' in p/l)
Strainer TYPE
J PPS/Teflon, HV
Seal Material
Pulser on Register
AB
Viton,
1 . 7
Flow Meter Assy No. = Register & Accessory combination
if LV rotors, LV valve (1-50 cSt)
if HT rotors, LV valve (1-50 cSt)
if HV rotors, LV valve (50-300 cSt)
If liquid viscosity is in 50-300 cSt range , either the distributor or the installer must remove set screw from valve piston prior to start-up.
Assy Nos. / Mechanical Meters+ B A C KPR ESSU R E
V A LV E
+ A IR C HEC K V A LV E
Issued
Without register
7887
7887, 7888, 7889 & 7856
7887, 7889 & 7856
7887 & 7888
V57
V53
VR
3
with mechanical preset valve/linkage
V51 V52
V11
VR
4
V56
7887, 7888 & 7889
V41BPV
V42 ACV
V32ACV
Same + Micro Switch kit on preset counter
With Calibrator, RAD adaptor, Ratio Gear Plate, 2-stg Preset Counter & Register
+ ST R A IN ER
Same + Zero Start Ticket Printer, V09
V36
V08
V40
+ A IR ELIM IN A T O
R
VR
3V
R4
Same + Micro Switch kit on preset counter
V15
V55
With Calibrator, RAD adaptor, Ratio Gear Plate, 2-stg Preset Counter, Register & Zero Start Ticket Printer
Same + Zero Start Ticket Printer, Preset Counter/Micro Switch assy
Same + Preset Counter/Micro Switch assy
V39
V35
WITHOUT preset valve/linkage
CA
L V22V21
M ET ER
V03With calibrator & RAD adaptor for register, but no ratio gear plate or register (distributor to add)
7887, 7888, 7889 & 7856
7887, 7889 & 7856
V12V44
7887 & 7889V13
V17V16
V54
ACV
V26
V37
V25V06
V31V38
BPV
ACV
BPV
VR
1V
R2 V07
With Calibrator, RAD Adaptor, Ratio Gear Plate & Register V04 V05
V29
BPV
BPV
ACV
ACV
BPV
ACV
BPV
August, 2009
BPV
ACV
OR
BPV
ACVV23V27
V28
V58
V34
V43V30
V45
V46
SAFETY INSTRUCTIONS Make sure that all necessary safety precautions have been taken, including proper clothing, personal safety equipment and fire safety equipment if required.
Before Start-Up of the Flow Meter, make certain that: 1. The meter is properly mounted, secured and piped. 2. All connections are tight. 3. All bleed and drain valves are closed. 4. Do NOT smoke near meter, or use meter near an open
flame, when metering flammable liquids. Fire or Explo-sion could result.
5. This meter is not intended for use on liquids, which re-quire 3A Sanitary equipment for human consumption. It may be used on edible liquids, which do not spoil easily, such as vegetable oils, liquid sweeteners & similar.
Install the Flow Meter and Accessories in compli-ance with all applicable Local, State & Federal
Construction, Electrical and Safety Codes.
Installation Positive Displacement meters are designed to operate full
of liquid. The meter should be installed in a manner, so that it remains full of liquid at all times.
The flow meter is not designed to operate on air, but the design and materials of construction allow for operation on vapor for short periods of time without damage to the oval gears or flow meter internals.
The meter should always be supported by bolting firm-ly to a solid platform or foundation. Never use the con-necting pipe as the sole means of support.
For most accurate measurement, the flow meter should operate with some backpressure (avoid delivering into an open vessel).
The flow meter can operate with liquid going Left-to-Right, Right-to-Left or Vertical Up, but it must be installed with rotor shafts in horizontal position (= with vertical end covers). Failure to observe this will impact negatively on flow meter accuracy.
1 . 8
Installation, Start-Up & Operation
OPERATING TEMPERATURE TS Series V●● assemblies are rated for operation from -40°F/+180°F (-40°C/+80°C). However:
They are not suitable for cryogenic service. When operating temperature exceeds +120°F
(+50°C), pressure rating is reduced. Please refer to FPP Technical Manual for details.
OPERATING PRESSURE Maximum non-shock Operating Pressure is: Flow Meter/Strainer : 400 PSI (28 BAR) at 100°F Air Eliminator/Valves: 150 PSI (10 BAR) (= +38°C)
The flow meter should never be operated in excess of this pressure. Care should be taken to eliminate thermal and hydraulic shock conditions, so that system pressure never exceeds Maximum Working Pressure rating.
Hydraulic shock can be harmful to flow meter and other system components. Consideration to eliminate hydraulic shock should be given in selection of pump and design of the piping system.
Protective caps installed in flow meter flanges prior to shipment should remain in place until you are ready to install in the piping system.
It is recommended that a Strainer be installed upstream of each flow meter, to prevent damage from foreign mat-ter, such as welding slag, pipe scale or parts breaking off other equipment. If using an FPP right-angle strainer:
Inlet can be either from the front or the rear of the strainer (inlet flange & strainer basket cover have same bolt pattern, and can be reversed in the field).
Allow adequate space for removal of strainer basket cover, so strainer basket can be cleaned. 10” for models TS10A, TS15A & TS20A. 11” for model TS30A
If no fluid is specified, strainer basket comes standard with 40 mesh.
We also offer:
20 mesh for high viscosity liquids, such as heavy fuel oil, automotive lube oils, liquid sugars, molas-ses, etc.
100 mesh for gasoline, alcohol & solvent service.
200 mesh for LPG (Propane) service.
Flush the system to remove all debris, scale and welding slag prior to flow meter installation. If this is not possi-ble, temporarily remove rotors (oval gears), and reinstall after the system has been flushed.
Continued
1 . 9
Installation & Start-Up/Operation
Apply pipe compound to male threads, to install the two companion flanges. Tighten to a position, that allows the meter to bolt to the companion flanges, free of pipe stress.
When installing the flow meter, consider future mainte-nance of both flow meter and accessories. The meter can be serviced in place, provided block (isolation) valves are included, and adequate space allowed.
Flow meters are designed to withstand a certain level of vibrations, such as might be experience on retail delivery tank trucks. If the flow meter will be used in a system with higher levels of vibration, it must be protected with: Vibration isolating pad under the mounting pedestal. Vibration isolators (flexible hose) on both inlet & outlet.
In critical installations a by-pass line is recommended, so flow can continue while flow meter is being serviced.
When an Air Eliminator is included in the flow meter assembly, the strainer/air eliminator must be in horizontal position, since the air eliminator operates on a gravity principle. A few drops of liquid may be expelled when the air eliminator vents, so vent ports should be piped back to storage or to a collection tank (sloping towards the tank).
If an Air Check Valve is included in the flow meter as-sembly, a modified air eliminator is used. One of the air eliminator vents must be connected to the port on the air check valve, using pipe, tubing or a flexible hose.
This connection must be made in the field, as we do not know what would be a convenient/practical path for the connection. Similarly, we do not know whether local reg-ulations or codes dictate materials to be used for this con-nection.
Thermal relief valves are recommended, and should be installed whenever it is possible to block (isolate) the flow meter between two valves. The pressure rise in a closed system, from just a few degrees increase in temperature, can be many times normal working pressure.
Connections for calibration should be provided during installation. An easy means for diverting flow into a cali-bration vessel (or master meter) should be considered.
The ideal installation incorporating all of these recommenda-tions is shown at the bottom of the page.
Start-Up/Operation Very slowly fill the system with liquid, to avoid operating the flow meter on air or vapor. This can be accomplished in the following manner:
1. Throttle the meter inlet valve, and allow the system to fill slowly by gravity.
2. Crack open the outlet valve. Start the pump, and then slowly crack open the inlet valve, filling the meter slowly before fully opening the inlet and outlet valves.
In normal operations:
Avoid sudden changes in temperature. Avoid sudden changes in flow rate. Gradually increase or decrease the flow rate.
TRV protects portion of system, which can be isolated between valves IV1, CV1 & IV2.
Calibration Connection valves
Thermal Relief valve
IV1 & IV2:CV1 & CV2:
BPV:TRV:
B Y - PA SS LIN E
IV1 IV2
TRV
TEMPORARY MASTER METER
CV2OPERATING FLOW METER
BPV
CV1
Isolation Valves
By-Pass valve
(or Prover Tank)
OPT ION A L
: IV2 & BPV closed, other valves open: IV1 & IV2 closed (by-pass open)
During Calibration
: BPV, CV1 & CV2 closedNormal OperationService/Maintenance
Now loosen the set screw (18) with a 3/32” Allen wrench, to release the calibrator drive shaft bushing (21B).
Slide the calibrator drive shaft bushing (21B) out of the bottom of the RAD (14).
Pull the calibrator drive shaft assembly out of the RAD.
Remove the face gear (11E), and turn it 180º, so that the gear teeth are facing in the opposite direction from the original installation.
Re-assemble the parts in reverse order.
Make sure that the calibrator drive shaft bushing flange (21B), is tight against the RAD (14).
Tighten the hex nut (21C)
Re-install the E-ring (21F) in the appropriate grove (in the above example, the E-ring now goes into groove above the pinion).
Inspect to make sure that there is slight backlash be-tween the pinion gear (11D) and the face gear (11E). They must engage without binding or slipping.
1 . 10
Changing Direction of Flow
TS Series flow meters can operate in either direction. Standard assembly, when facing the flow meter, has the flow going Left-to-Right, unless specified differently at time of the order. A label with the word INLET is placed on the inlet flange, as assembled at the factory.
When the meter is first installed, check that the register is turning correctly when you start the flow. If it is not (or it is attempting to run backwards), check position of the face gear (11E) on the calibrator drive shaft (21A), and if neces-sary correct the position of the face gear versus the packing gland pinion (11D) as explained below.
Left-to-Right flow = Face gear below packing gland pinion
Right-to-Left flow = Face gear above packing gland pinion
The direction of flow can be changed in the field. This re-quires the following steps:
Any accessories (strainer and/or preset valve) attached to the flow meter must be moved to the opposite side.
Strainer inlet is standard from the front. When mov-ing the strainer to Right-to-Left position, switch the inlet flange and strainer basket cover. If rear inlet is preferred, leave flange & basket cover as is, when moving the strainer.
Standard outlet on Preset Valve or Air Check Valve is towards the front. When moving the valve to Right-to-Left position:
Air Check Valve: Rotate the valve 180º, then in-stall tubing connecting air eliminator to the valve.
Preset Valve: Rotate the valve 180º. Disassem-ble the linkage, and re-assemble mirror image to original assy.
Reverse the position of the face gear (11E) on the cali-brator drive shaft (21A).
First remove screws (16 & 17) to detach the dust cover (15) from the RAD (14 = Right Angle Drive) adaptor. Now refer to photo in next column.
Notice position of the face gear (11E) on the calibrator drive shaft (21A) in relation to the packing gland pinion (11D).
Remove E-ring (21F) with a screwdriver or small pliers. As shown, the E-ring is installed below the face gear, where it is not visible in the photo.
11E
11D
21A
18
INLET
19
14
INLET
21B
16
17
15
14
21E
21B
21C
21D
21E
21F
11E
21A = Drive shaft with 2 grooves: Upper Lower
Mechanical TS Series meters have a mechanical drive train directly from the rotors (oval gears) to the register. To con-vert rotor movement to volume reading on the register, this drive train includes 3 calibration components:
Packing gland pinion (11D) can have 12 or 24 teeth, while the face gear (11E) always has 24 teeth, re-sulting in a 1:1 or 2:1 ratio. The appropriate ratio de-pends upon flow meter model size & specific unit of calibration selected for the register.
Mechanical calibrator (19). This component makes minute adjustments to final calibration to compensate for individual flow meter characteristics, viscosity of fluid being metered, and flow meter wear factor.
A ratio gear plate installed below the mechanical register. Gearing on this gear plate is specific to flow meter model & unit of calibration selected. Optimum ratio is between 0.5000 & 5.0000 for best service life.
The mechanical register is common to all flow meter mod-els, so it is possible to move registers from one flow me-ter to another. Do NOT move the ratio gear plate from one flow meter to another.
.Frequency of Re-Calibration In installations subject to local Weights&Measures regula-
tions, frequency of recalibration must conform to such regulations.
If local authorities issue regulations for non-W&M flow meters, such regulations must be observed.
If user is ISO9000 certified, user ISO standards will indi-cate frequency of re-calibration for instrumentation. Those standards should be observed.
When no regulations or standards apply, our recommenda-tions are:
A. Calibrate immediately after installation.
B. Re-calibrate after 15-30 days.
C. Re-calibrate after 180 days and again after 360 days.
After the run-in calibration (B) and follow-up calibrations (C), it is possible to evaluate degree of change under normal operating conditions. Based on values found, and total vol-ume being metered under normal operating conditions, de-cide whether a 6, 12 or possibly 24 month schedule should be adopted.
1 . 11
Calibration, Mechanical Flow Meters
W&M Certifications:
TS10A US TS15A US & Canada TS20A US, Canada, EU & Australia TS30A US, Canada, EU & Australia
Certifications are liquid specific. Please refer to national certificates for liquid categories covered. Flow Meter Calibration at the Factory All TS Series meters are accuracy tested prior to shipment. Data from accuracy testing is not supplied with flow meters, since results achieved on our test fluid do not apply to actual liquid of operation, unless the two liquids have identical vis-cosity characteristics.
When calibrated on a liquid with 1 cP viscosity, if the flow meter is not field recalibrated on actual liquid of service, we expect that the flow meter will be:
Under by 1-2% on a liquid with 0.5 cP viscosity.
Over by 0.35% to 0.6% on a liquid with 10 cP viscosity.
Over by 0.7% to 1.2% on a liquid with 100 cP viscosity.
The accuracy curve will not shift significantly at higher vis-cosities, even if the actual operating liquid has viscosities up to 500,000 cP.
Accuracy curves of individual flow meters vary slightly. Such minor variations are corrected in the mechanical cali-brator. In our calibration procedure, the meter calibrator is set for zero error, when operating at 95-100% of capacity on 1 cSt viscosity fluid.
Flow Meter Calibration in The Field Since we cannot test on actual fluid of operation, it is the responsibility of the buyer to field calibrate:
In place of service
On actual operating liquid.
This will minimize errors arising from:
A. Operation at a different flow rate. B. Operation on a fluid with different viscosity
It is recommended that written records be maintained on all flow meters. These records should include:
Supplier and Service Department phone number. Date of installation. Details of maintenance performed. Date & result of each re-calibration, with % change made
on flow meter calibrator assy.
11D
11E
19
When prover/master meter reading is less than flow meter register reading, add percentage calculated by turning the calibrator in the + volume direction.
When prover/master meter reading is more than flow meter register reading, subtract percentage calcu-lated by turning the calibrator in the - volume direction.
After correcting calibrator setting, circulate product through the flow meter for a few minutes. Then perform at least 3 more tests, to confirm flow meter accuracy & repeatability.
If the flow meter does not repeat, it will likely require a new set of rotors (oval gears).
Before ordering new gears, inspect the measuring chamber for scratches or wear. If the measuring chamber is scratched or scored beyond what can be smoothed with emery paper, flow meter case should be replaced.
Finally: Re-seal the flow meter.
Enter date and % correction on the permanent flow meter record.
As long as degree of change is moderate, the flow meter is in good condition.
If there is a sudden significant jump in correction required, the rotors are likely about worn out. Rotor replacement should be considered now, rather than letting further wear cause rotors to start rubbing on flow meter housing.
1 . 12
Flow Meter Calibration
Procedures & Methods Flow meters used in systems where the flow rate can fluctu-ate, should be tested at minimum, intermediate & maximum flow rates. In non-W&M service, a flow meter always oper-ating at a steady flow rate, can be tested at that flow only.
All tests should be repeated 3 times to confirm repeatability. All tests should be of at least 60 second duration, to mini-mize the effect of flow meter error during start-up and shut-down.
After calibrating a known volume (X) into an accurate prover, through a master meter, or against a certified scale with adequate resolution, compare with register reading (Y) and calculate correction:
When re-calibration has established that a correction is required, remove the dust cover (15) from the RAD (14 = right angle drive adaptor), by detaching screws (16 & 17).
This provides access to the calibrator assembly (19), which has a dial on the front, and is marked with direc-tions for + and - Volume. Each graduation of this dial represents approx. 0.06% change
The dial has a slot for a screw driver in the center, so the setting can be changed. As you turn the center, you will hear two ‘clicks’ for each grad-uation mark. Each ‘click’ rep-resents approximately 0.03% change in the setting.
Do NOT change setting, unless re-calibration of the flow meter has established the need for correction.
19
X - YX x 100 = % correction
17
15 16
14
19
Always coat bolt threads with an anti-seize, or an appropri-ate lubricant to prevent thread damage, and to assure prop-er torque values are applied when reassembling.
If flow meter threads are damaged, repair using inserts.
Servicing the Mechanical Drive Components Flow meter wear depends upon several variable factors (flow rate vs model capacity, lubricity of the liquid, total vol-ume being metered annually and maintenance of mechani-cal register, preset and/or printer), so it is impossible to pre-dict the expected life for each component.
There are 8 parts or components, which will require inspec-tion and/or replacement at some point. In the order they are likely to occur, these are:
Regular Maintenance (1-3 years) A. Face Gear (11E) & Packing Gland Pinion (11D) B. Packing Gland O-ring (11B) Longer Term Wearing Components (2-5 years) C. Calibrator Drive Shaft Assembly (21) D. Gear Plate, internal (4) Very Long Term Wearing Components (3-10 years) E. Packing Gland (11) F. Oval Gears (2), or inspection of measuring chamber Longest Term Wearing Components (5-20 years) G. Post Plate assembly (3) H. Calibrator Assembly (19)
Of these, only A, C & H can be performed without taking the pressure off the system and draining the flow meter. The remaining all require opening of flow meter interior.
The first step is to remove the dust cover (see pages 1.8-9) from the RAD (14 = Right Angle Drive adaptor).
A. Face Gear (11E) & Packing Gland Pinion (11D) C. Calibrator Drive Shaft Assembly (21) Always replace Face Gear & Packing Gland Pinion together. They can be ordered under a single P/No. (see page …).
Remove E-ring (21F) from the drive shaft assembly (21) with a screwdriver or small pliers.
Loosen the set screw (18) with a 3/32” Allen wrench, to release the calibrator drive shaft bushing (21B).
Slide the calibrator drive shaft bushing (21B) out of the bottom of the RAD (14), and pull the calibrator drive shaft assembly out of the RAD.
Replace Face Gear (11E) on the calibrator drive shaft assembly.
Remove E-ring (11C) from the packing gland (11) with a screwdriver or small pliers, and replace the Packing Gland Pinion (11D).
Reassemble in reverse order.
Servicing the Flow Meter
Installation, Maintenance & Service must be performed by personnel:
A. Qualified to work on this type of equipment.
B. Familiar with all applicable local codes and ordi-nances covering the type of service, where the flow meter is used (gasoline, LPG, etc.).
Avoid pipe strain and stress when making flow meter re-pairs. The weight of piping and the flow meter should be supported independently. This means that the meter can be serviced without affecting piping alignment.
Avoid prying or exerting heavy pressure on precision parts, as this could affect the performance of the flow meter.
Assure that all machined parts are free of burrs and nicks. Stone all machined surfaces if necessary to remove burrs.
14a
14b
26
19
18
11C/D
11E
21F
1 . 13
12
13
! SAFETY INSTRUCTIONS ! REMOVE ALL INTERNAL PRESSURE
BEFORE OPENING THE FLOW METER.
DRAIN & RINSE THE FLOW METER BEFORE SERVICING.
IF IT IS NECESSARY TO TRANSPORT OR SHIP THE FLOW METER TO A DIFFERENT LOCATION FOR SERVICE, IT MUST BE RINSED 3 TIMES WITH A NEUTRAL FLUID PRIOR TO TRANSPORTATION/SHIPMENT.
27/28
You now have access to the Gear Plate as-sembly (4). If rebuild-ing with repair kit, no further disassembly is required.
If replacing the oval gears, inspecting the measuring chamber or replacing the complete gear plate:
Remove 4 screws (6) from the gear plate. The gear plate can now be pulled off dowel pins (5), using a 5/16-8 x 1” jack bolt if necessary.
The gear plate assembly can now be rebuilt or replaced as required.
Prior to removing the rotors (oval gears), use a soft pencil to make an alignment mark where the gears engage. This will make re-assembly easier.
The oval gears can now be inspected or replaced as re-quired. When placing the oval gears back on the posts, make sure that they are aligned and spin freely on the posts. If they are not aligned correctly, they will disen-gage.
The measuring chamber can now be inspected. Remove any deposits using a fine emery cloth or wire brush. Be careful to remove any particulate material, which can cause the meter to jam.
A minor scratch in chamber surface can be carefully smoothed out, but be careful not to cause scoring or to change the dimensions. If the measuring chamber shows signs of scoring from rotors rubbing, it should be re-placed.
Re-assemble in reverse order.
G. Post Plate assembly (3) First relieve system pressure and drain the flow meter.
Remove the cover bolts (10) from the rear cover (9), now remove the rear cov-er from the flow me-ter and inspect the cover O-ring (7) for flaws. Teflon fitted meters: Always re-place seals when opening the flow me-ter.
Servicing the Flow Meter
B. Packing Gland O-ring (11B) E. Packing Gland Assembly (11) First relieve system pressure and drain the flow meter. Then follow instructions for above for removal of calibrator drive shaft assembly. Remove E-ring (11C) from the pack-ing gland assembly, and pull off the pinion (11D).
Remove three screws (13) from the packing gland retain-er (12). Remove the retainer
Packing Gland O-ring and/or Packing Gland assembly can now be replaced.
When replacing the Packing Gland, the drive dog must be aligned with the slot in the gear on the gear plate.
Reassemble in reverse order.
D. Gear Plate, Internal (4) F. Oval Gears (2), or inspection of measuring chamber The wearing parts in the Gear Plate assembly are the reduc-tion gears. On a longer time scale, the bushing in the gear plate will wear out. The gear plate assembly can be rebuilt or replaced in one of 3 ways, with d.3 being what is most commonly required:
d.1 Complete gear plate assembly
d.2 Gear plate/bushing assembly only, re-using gears.
d.3 Repair kit, with reduction gears & fasteners.
First relieve system pressure and drain the flow meter.
Also, detach the mechanical register stack, to take the weight off of flow meter front cover.
If the mechanical register stack includes a preset counter, detach the linkage from the trip ring (see top of next column) prior to taking the register stack off.
Remove dust cover & calibrator drive shaft as explained above.
Remove 4 screws/washers (27/28) holding the RAD to the front cover (8).
Remove the cover bolts (10) from the front cover (8), now remove the front cover from the flow meter and inspect the cover O-ring (7) for flaws. Teflon fitted meters: Al-ways replace seals when opening the flow meter.
Valve linkage connects to preset trip ring at the rear.
Slide the sleeve towards the shaft to release the mounting stud.
Register stack attaches to the RAD with 4 screws, 2 front and 2 rear.
STUD
SLEEVE
6
4
5
8
7
1 . 14
7
5 6
9
3
10
For optional flow meter accessories:
Strainer
Air Eliminator
Backpressure Valve
Air Check Valve
For optional control valves:
Mechanical Preset Valve
1-Stg solenoid operated Valve
2-Stg solenoid operated Valve
For mechanical register stack:
Register
10:1 pulser
100:1 pulser
Preset Counter
Micro Switch kit
Ticket Printer
1 . 15
Servicing the Flow Meter
Remove 4 screws (6) from the post plate (3). The post plate assembly can now be pulled off dowel pins (5), us-ing a 5/16-8 x 1” jack bolt if necessary.
The posts are expected to last through several sets of rotors. When the posts show signs of wear, or if they have been bent due to hydraulic shock, the complete postplate assembly must be replaced.
If hydraulic shock is suspected, the cause should be identified and eliminated.
If the posts show uneven wear (bearings cutting grooves into the post), the differential pressure across the flow meter is too high, due to a combination of ex-cessive flow rate & viscosity. Solutions are:
1. Reduce the flow rate.
2. Increase the temperature to reduce the viscosity.
3. Use a larger flow meter.
Re-assemble in reverse order.
H. Calibrator Assembly (19) The calibrator assembly is a mechanical device, which will wear out eventually. Calibrator wear can be accelerated by failure to clean and lubricate the components in the mechan-ical register stack (refer to separate manuals). If flow meter re-calibration gives erratic results (non-repeating), it is likely that the calibrator is worn out.
To replace the calibrator:
Remove dust cover, drive shaft assembly and register stack, as explained under A and D.
Remove 2 screws (20) holding the Calibrator Assembly to the Veeder-Root register flange (14B), using a Phillips screwdriver.
Re-assemble in reverse order.
Pls see Section 2
Pls see separate Tuthill manual
Pls see separate Veeder-Root manuals
1 . 16
Flange Kits & Parts for Flanges
Tuthill TS Series flow meters come standard with bolted companion flanges, with either NPT or BSP threaded ports. Other options are welding flanges (aluminum or steel), and 150# RF ANSI adaptors (steel).
Since meter (square) & accessory (round) flanges differ, a
new flange is required when adding to or deleting accesso-ries from an existing flow meter. When adding accessories to TS10A & TS15A models, an adaptor might be required.
Single flange kits & adaptors, and relevant seals & screws, are identified below.
Flow Meter Strainer Standard High Cap. Air Check & Electronic Inlet/Outlet or Valve Strainer Strainer Preset Vlv Preset Valve Flange Adaptor Inlet Flange Inlet Flange Outlet Flg Outlet Flange TS10A, OS NPT/Viton FK1100-1 FK1209 FK1104-1 na na FK1106-S1 1” BSP/Viton FK1101-1 same FK1105-1 na na FK1107-S1 ANSI/Viton FK5100-1 same FK5104-1 na na FK5106-S1 Viton O-ring SL1122 SL1122 SL1334 na na SL7500
NPT/Teflon FK2100-1 FK1209-T FK2104-1 na na na BSP/Teflon FK2101-1 same FK2105-1 na na na ANSI/Teflon FK5101-1 same FK5105-1 na na na Teflon O-ring SL2122 SL2122 SL2334 na na na Flange screw FS2950 FS0920 FS2960 na na 1123413
TS10A, NS NPT/Viton FK1110-1 na na na na na 1” BSP/Viton FK1111-1 na na na na na Viton O-ring SL1223
NPT/Teflon FK2110-1 na na na na na BSP/Teflon FK2111-1 na na na na na Teflon O-ring SL2223 Flange screw FS2950
TS15A NPT/Viton FK1160-1 FK1215² FK1154-1 na FK1156-1 FK1156-S1 1½” BSP/Viton FK1161-1 same FK1155-1 na FK1157-1 FK1157-S1 ANSI/Viton FK5160-1 same FK5154-1 na FK5156-1 FK5156-S1 Viton O-ring SL1233 SL1233 SL1334 na SL7500 SL7500
NPT/Teflon FK2160-1 FK1215-T² FK2154-1 na FK2156-1 na BSP/Teflon FK2161-1 same FK2155-1 na FK2157-1 na ANSI/Teflon FK5161-1 same FK5155-1 na FK5157-1 na Teflon O-ring SL2233 SL2233 SL2334 na SL2500 na Flange screw/stud FS9902 FS5201 FS2960 na FS2960 1123413 Flange washer FS5008 FS5203 na Flange nut na FS5202 na
TS20A NPT/Viton FK1200-1 na FK1204-1 FK1204HC-1 FK1106-1 FK1106-S1 2” BSP/Viton FK1201-1 na FK1205-1 FK1205HC-1 FK1107-1 FK1107-S1 ANSI/Viton FK5200-1 na FK5204-1 FK5204HC-1 FK5106-1 FK5106-S1 Viton O-ring SL1233 na SL1334 SL1350 SL7500 SL7500
NPT/Teflon FK2200-1 na FK2204-1 FK2204HC-1 FK2106-1 na BSP/Teflon FK2201-1 na FK2205-1 FK2205HC-1 FK2107-1 na ANSI/Teflon FK5201-1 na FK5205-1 FK5205HC-1 FK5107-1 na Teflon O-ring SL2233 na SL2334 SL2350 SL2500 na Flange screw FS2950 na FS2960 FS2965 FS2960 1123413
TS30A NPT/Viton FK1300-1 na FK1304-1 FK1304-1 FK1106-1 FK1106-S1 3” BSP/Viton FK1301-1 na FK1305-1 FK1305-1 FK1107-1 FK1107-S1 ANSI/Viton FK5300-1 na FK5304-1 FK5304-1 FK5106-1 FK5106-S1 Viton O-ring SL1240 na SL1350 SL1350 SL7501 SL7501
NPT/Teflon FK2300-1 na FK2304-1 FK2304-1 FK2106-1 na
BSP/Teflon FK2301-1 na FK2305-1 FK2305-1 FK2107-1 na ANSI/Teflon FK5301-1 na FK5305-1 FK5305-1 FK5107-1 na Teflon O-ring SL2240 na SL2350 SL2350 SL2503 na Flange screw FS1951 na FS2965 FS2965 FS2965 FS5210
NOTES: 1 = Used for both strainer & air check valve when assembled to TS10A model. 2 = Used for strainer only when assembled to TS15A model (valves fit directly to meter body).
1 . 17
Troubleshooting the Flow Meter
Prior to opening or disassembly of any flow meter, all internal pressure must be relieved and all liquid must be drained. This must be done in accordance with applica-ble company and local codes & ordinances.
Make sure that all necessary safety precautions have been taken, including proper clothing, personal safety equipment and fire safety equipment if required.
No Flow Blocked strainer basket. Clean the basket. Faulty or non-functioning pump. Repair pump. Valve stuck in closed position. Check and repair valves. Flow meter ‘frozen’ due to build-up of chemical salts (or
frozen water) inside the measuring chamber. Clean the flow meter (see D & F on page 1.12), and inspect for damage.
Meter jammed on a particle, that has passed through a damaged strainer basket. Remove particle (see D & F on page 1.12) and replace rotors if necessary. Replace strainer basket, see page 2.1).
Reduced Flow Rate Strainer basket partially blocked. Clean the basket (see
page 2.1). Pump not functioning correctly. Repair pump. Valve stuck in partially closed position. Check valves and
repair. Meter rotors (oval gears) partially ‘salted’ with chemical
deposits, slowing the movement. Clean the meter (see D & F on page 1.12).
Product Flows, but the register does not record Check packing gland pinion (11D) & face gear (11E). If disengaged, re-assemble the drive shaft assembly
correctly. If worn, replace pinion & face gear as a set.
If the calibrator drive shaft turns, the problem is either in the calibrator assembly (19) or in the mechanical register. Remove the calibrator assembly (see H on page 1.13).
Insert the drive shat and turn by hand. If it turns smoothly without binding, the problem is in the register.
Refer to Veeder-Root register manual & parts list. If the calibrator drive shaft does not turn, the problem is
either in the packing gland or in reduction gears on the bearing plate.
Check the connections between the pulser and the elec-tronic register.
Check pulser output. Replace pulser (or SCL) if needed. If product is flowing, and the flow meter is generating a
pulse signal, the problem is in the electronic register. Please refer to the manual for the electronic register.
Product Flows, register does not record correctly If the error factor is constant, the flow meter is fine. The likely cause is either: Incorrect K Factor in the electronic register. Re-calibrate
the meter and correct the K Factor. A constant problem with air getting into the system. Re-
view system design and control valves. If the error is random, the likely cause is either: Poor cable connections (insulation not trimmed, or stray
strands getting close to incorrect contacts). This can be either on the scaler board or where pulse signal is con-nected to the register. Inspect and correct connections as necessary.
Valve leaking, allowing a portion of the system to drain. Check & repair valves.
An intermittent problem with air in the system, combined with inadequate air elimination. Review system design and control valves.
Interference from other electrical equipment nearby., pos-sibly combined with sub-standard cables.
Broken Rotors (Oval Gears = #2) System started on air/vapors.
Replace rotors and inspect meter case (#1) for damage; repair/replace as necessary. Slowly fill the system with liquid before starting the pump. Gradually increase flow rate to full capacity.
Rotors jammed on a foreign particle (welding slag, part broken off other equipment, etc.). Replace rotors and inspect meter case (#1) for damage; repair/replace as necessary. Add a strainer in front of the flow meter. Slowly fill the system with liquid before start-ing the pump. Gradually increase flow rate to full capaci-ty.
Breaking Teeth on Rotors (Oval Gears = #2) Breaking Teeth/shaft on Reduction Gears (#4D/F/G These are signs of hydraulic shock conditions in the sys-tem. Common sources: Starting or stopping flow too rapidly. Replace damaged
components and correct operational practices. Pump by-pass not adjusted properly. Re-adjust as nec-
essary.
Open connection to another portion of the system with higher flow rates. Close connection prior to use.
Leakage from End Covers The seals (and possibly end covers) have been damaged due to excessive pressure. There are two possible sources: Starting or stopping flow too rapidly. Replace damaged
components and correct operational practices. The flow meter is in a system, where it can be isolated
between two valves. Add a Thermal Relief Valve to bleed off excess pressure when the temperature rises.
1 . 18
Parts List
TS10A, OS TS10A, NS TS15A TS20A TS30A
ITEM DESCRIPTION QTY P/No. QTY P/No. QTY P/No. QTY P/No. QTY P/No. 1 Meter Body 1 MB2560 1 MB2565 1 MB2575 1 MB2395 1 MB5201-1 Depth across body: 1.47” (37 mm) 1.97” (50 mm) 3.46” (88 mm) 3.46” (88 mm) 4.96” (126 mm)
2 Gear Set, PPS/carbon, LV 1 GS560RC 1 GS10RC 1 GS15RC 1 GSTS2HRC 1 GS5201RC ** Opt. PPS/carbon, HT GS560RC-MV GS10RC-MV GS15RC-MV GSTS2HRC-MV GS5201RC-MV PPS/carbon, HV GS560RC-HV GS10RC-HV GS15RC-HV GSTS2HRC-HV GS5201RC-HV PPS/Teflon, LV GS560RT GS10RT GS15RT GSTS2HRT GS5201RT PPS/Teflon, HT GS560RT-MV GS10RT-MV GS15RT-MV GSTS2HRT-MV GS5201RT-MV PPS/Teflon, HV GS560RT-HV GS10RT-MV GS15RT-HV GSTS2HRT-HV GS5201RT-HV
3 Post Plate assy. 1 CP25521 1 CP25801 1 CP25802 1 CP2396-2 1 CP23961 *** 4 Gear Plate assy., internal 1 CP25532 1 CP25532 1 CP25532 1 CP23162 1 CP23172 *** 4a Gear plate with bushing CP25531 CP25531 CP25531 CP23161 CP23171 4b-I Repair kit, replacement of combination gear set RKCP1 RKCP1 RKCP1 RKCP2 RKCP2 ** 5 Dowel pin 4 MS902118-8 4 MS902118-8 4 MS902118-8 4 MS902118-8 4 MS902118-8 6 Screw, for post/gear plates 8 FS9661 8 FS9661 8 FS9661 8 FS9661 8 FS9661
7 O-ring, cover, Viton™ 2 SL1157 2 SL1157 2 SL1157 2 SL1167 2 SL1167 * Opt. Teflon™ SL2157 SL2157 SL2157 SL2167 SL2167 8 Front cover 1 CP2550 1 CP2550 1 CP2550 1 CP2398 1 CP2398 9 Rear cover 1 CP2551 1 CP2551 1 CP2551 1 CP2397 1 CP2397 10 Cover screws 16 FS2900 16 FS2900 16 FS2900 24 FS2928 24 FS2928
11 Packing gland, AA/Viton 1 RK2001 1 RK2001 1 RK2001 1 RK2003 1 RK2003 ** Opt. SS/Teflon RK8002 RK8002 RK8002 RK80031 RK80031 11b O-ring, Viton SL1212 SL1212 SL1212 SL1212 SL1212 * Opt. Teflon SL2212 SL2212 SL2212 SL2212 SL2212 11cde For 1/10 gallon GS2001 GS2001 GS2002 GS2002 (std) na * + 21f Whole gallon na na na GS2001 (opt) GS2002 1/10 liter GS2002 (std) GS2002 (std) na na na Whole liter GS2001 (opt) GS2001 (opt) GS2001 GS2002 (std) GS2002 (std) Whole dekaliter na na na GS2001 (opt) GS2001 (opt) 12 Retainer, packing gland 1 MP9061 1 MP9061 1 MP9061 1 MP9061 1 MP9061 13 Screws, PG retainer 3 FS9515 3 FS9515 3 FS9515 3 FS9515 3 FS9515
14 RAD (right angle drive) 1 RD15911 1 RD15911 1 RD15911 1 RD15911 1 RD15911 15 RAD dust cover 1 MP2505 1 MP2505 1 MP2505 1 MP2505 1 MP2505 16 Screws, w seal wire hole 2 FS59121 2 FS59121 2 FS59121 2 FS59121 2 FS59121 17 Screws, plain 2 FS5912 2 FS5912 2 FS5912 2 FS5912 2 FS5912 18 Set screw for drive shaft 1 FS9652 1 FS9652 1 FS9652 1 FS9652 1 FS9652 19 Calibrator assy. 1 CA6000 1 CA6000 1 CA6000 1 CA6000 1 CA6000 *** 20 Screws, calibrator to RAD 2 FS2501 2 FS2501 2 FS2501 2 FS2501 2 FS2501 21 Drive shaft assy, (A-F) 1 RD15912 1 RD15912 1 RD15912 1 RD15912 1 RD15912 **
22 Meter pedestal 1 MP2528 na na 1 MP2526 1 MP2529 23 Split lock washers na na na na na 24 Screws, for pedestal 2 FS9923 na na 2 FS9923 2 FS9923 25 Drain plug 2 MS4019 2 MS4019 2 MS4019 2 MS4019 2 MS4019
26 Screws, RAD, w seal hole 2 FS59121 2 FS59121 2 FS59121 2 FS59121 2 FS59121 Screws, RAD, plain 2 FS5912 2 FS5912 2 FS5912 2 FS5912 2 FS5912 27 Screws, RAD to cover 4 FS2800 4 FS2800 4 FS2800 4 FS2800 2 FS2800 28 Lock washer, 1/4” 4 FS3602 4 FS3602 4 FS3602 4 FS3602 4 FS3602
29 Flange or adaptor assy. See pg. 1.14 See pg. 1.14 See pg. 1.14 See pg. 1.14 See pg. 1.14
30 Register mounting bolt 4 FS0856 4 FS0856 4 FS0856 4 FS0856 4 FS0856
Time frame for replacement of regular maintenance items and long term consumables (rotors, packing gland & calibrator) varies a great deal. Factors include user maintenance schedule, fluid lubricity, flow rate versus model capacity, intermittent versus continuous duty & total volume. In heavy duty service, rotors may last from 2-5 years, while in lighter duty service, they may last from 10-20 years or more.
* = Indicates recommended Spare Parts (regular maintenance items).
** = Indicates suggested Spare Parts (longer time frame, but should be on hand if meter is in a location where service is not readily available).
*** = Indicates very long term wearing component (should be on hand if meter is in a location where service is not readily available).
1 . 19
TS Series - Exploded View, Mechanical (V●● Assy.)
16
14a
12
1
2
3
8
15
17
18
11
20
19
13
4
5
5 6
7
9 10
7
14b
27
21
11E 24
23
25
29
22
26
28
A B
C
#4 = Gear Plate assembly (internal) Can be replaced complete, or as two sub-assemblies (the individual parts are not sold separately, as they must be replaced as a set):
Gear plate with bushing (4A)
Rebuild kit (items 4B-4I) #29 = Companion Flanges Companion flange assemblies (A+B+C) vary, depending on whether the flange attaches to:
Flow meter body
Inlet strainer
Preset valve or air check valve on flow meter outlet.
A complete list of companion flange assemblies, seals and screws is found on page 1.16
#11 = Packing Gland The packing gland ratio (D = 12 or 24 tooth pinion) de-pends upon flow meter model, and unit of calibration select-ed on the mechanical register. A list of packing gland ratios can be found in FPP Technical Manual (see list of ratio gear plates).
Pinion/Face Gear (11D/E) can be replaced separately:
GS2001 2:1, metal/PPS
GS2002 1:1, PPS/PPS (standard)
GS2003 1:1, metal/PPS (for tank truck service)
The packing gland is replaced as a kit under the P/No. shown in the parts list, which includes both GS2001 & GS2002.
#21 = Calibrator Drive Shaft assy. This assembly is replaced as a kit containing all components (A-F). Except for the E-ring (F) and face gear (11E), parts are not sold separately.
3A
10
11B
30
1 . 20
Tools & Torque Chart
Warranty Tuthill Transfer Systems (“Manufacturer”) warrants to each buyer of its FPP Meters products (the “Buyer”) for a period of 12 months from date of invoice or sales receipt, but in no event more than 18 months from date of manufacture, that goods of its manufacture (“Goods”) will be free from defects of material and workmanship. Manufacturers sole obligation under the foregoing warranties will be limited to either, at Manufacturer’s option, replacing or repairing defective Goods (subject to limitations hereinafter provided) or refunding the purchase price for such Goods theretofore paid by the Buyer, and Buyer’s exclusive remedy for breach of any such warranties will be enforcement of such obligations of Manufacturer. If Manufacturer so requests the return of the Goods, the Goods will be redelivered to Manufacturer in accordance with Manufacturer’s instructions F.O.B. Factory. The remedies contained herein shall constitute the sole recourse of the Buyer against Manufacturer for breach of warranty. IN NO EVENT SHALL MANUFACTURER’S LIABILITY ON ANY
CLAIM FOR DAMAGES ARISING OUT OF THE MANUFACTURE SALE, DELIVERY OR USE OF THE GOODS EXCEED THE PURCHASE PRICE OF THE GOODS. The foregoing warranties will not extend to Goods subjected to misuse, neglect, accident or improper installation or maintenance, or which have been altered or repaired by anyone other than Manufacturer or its authorized representative. THE FOREGOING WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES OF MERCHANTABILITY, FITNESS FOR PURPOSE OF ANY OTHER TYPE, WHETHER EXPRESS OR IMPLIED. No person may vary the foregoing warranties and remedies except in writing signed by a duly authorized officer of Manufacturer. Warranties or remedies that differ from the foregoing shall not otherwise be binding on Manufacturer. The Buyer’s acceptance of delivery of the Goods constitutes acceptance of the foregoing warranties and remedies, and all conditions and limitations thereof.
Tuthill Transfer Systems (TTS) warranty does not extend to damage due to corrosion, salt deposits or separation of chemicals - whether occurring during periods of operation or storage.
Components added after shipment from TTS, such as hose, nozzles & similar, are the responsibility of the distributor or contractor selling the installation.
D ESC R IP T ION TS10A TS15A TS20A TS30APost/Bearing plate screw Tool 5/32" Allen 5/32" Allen 5/32" Allen 5/32" Allen
10/32 19.2 in-lbs 10/32 19.2 in-lbs 10/32 19.2 in-lbs 10/32 19.2 in-lbsCover screw Tool 1/4" Allen 1/4" Allen 5/16" Allen 5/16" Allen
5/16-18 80 in-lbs 5/16-18 80 in-lbs 3/8-16 143 in-lbs 3/8-16 143 in-lbsPacking gland retainer screw Tool Impact Battery Tool Impact Battery Tool Impact Battery Tool Impact Battery Tool
8/32 10.8 in-lbs 8/32 10.8 in-lbs 8/32 10.8 in-lbs 8/32 10.8 in-lbsDust cover screws Tool Impact Battery Tool Impact Battery Tool Impact Battery Tool Impact Battery Tool
10/14 10.8 in-lbs 10/14 10.8 in-lbs 10/14 10.8 in-lbs 10/14 10.8 in-lbsCalibrator bushing set screw Tool 3/32 Allen Tool 3/32 Allen Tool 3/32 Allen Tool 3/32 Allen Tool
10/32 19.2 in-lbs 10/32 19.2 in-lbs 10/32 19.2 in-lbs 10/32 19.2 in-lbsCalibrator mounting screw Tool Impact Battery Tool Impact Battery Tool Impact Battery Tool Impact Battery Tool
Sheet Metal Screw s Sheet Metal Screw s Sheet Metal Screw s Sheet Metal Screw sPedestal screw Tool Built-in pedestal Built-in pedestal Impact Battery Tool Impact Battery Tool
N/A N/A 3/8-16 143 in-lbs 3/8-16 143 in-lbsCover drain plug Tool 1/4" Allen 1/4" Allen 1/4" Allen 1/4" Allen
Hand tight Hand tight Hand tight Hand tightRAD housing/flange screw Tool Impact Battery Tool Impact Battery Tool Impact Battery Tool Impact Battery Tool
Sheet Metal Screw s Sheet Metal Screw s Sheet Metal Screw s Sheet Metal Screw sRAD to meter cover screw Tool Impact Battery Tool Impact Battery Tool Impact Battery Tool Impact Battery Tool
1/4-20 45.6 in-lbs 1/4-20 45.6 in-lbs 1/4-20 45.6 in-lbs 1/4-20 45.6 in-lbsCompanion flange screw Tool 5/16" Allen 1/4" Allen 5/16" Allen
3/8-16 143 in-lbs 5/16-18 80 in-lbs 3/8-16 143 in-lbs
Note: All torque specification are Reference guide listed from the Fastenal Technical Guide.
Torque specification30
27Torque specification
25Torque specification
26Torque specification
24Torque specification
13Torque specification
20
18Torque specification
16 17 Torque specification
Torque specification
IT EM
6Torque specification
10Torque specification