Overview of New YSI Laboratory Products
Trulab, MultiLab, Ion Selective Electrodes (ISEs)
August 12, 2014
Chris Cushman, Assistant Product Manager [email protected]
Laura St. Pierre, Senior Product Manager [email protected]
Outline
•TruLab
-Analog pH/mV(ORP)/ISE benchtop instruments
•MultiLab
-Digital multiparameter benchtop instruments
•New TruLine ISEs
TruLab: Analog pH/mV(ORP)/ISE Benchtop Instruments with new suite of pH electrodes
YSI TruLab Product Line
• Analog line of pH, mV and ISE instruments – released March 2014
• Specialty line of pH and ISE probes for various applications
• Positioned to compete against Thermo and Hanna
• Complimentary to MultiLab products
YSI TruLab Instruments
TruLab pH 1110 • Single channel analog instrument • pH, mV(ORP), Temp • No memory • Segmented display • 3 stored buffer sets
TruLab pH 1310 • Single channel analog instrument • pH, mV(ORP), Temp • Memory (500 manual, 5000 auto) • USB output; data importer software • 22 stored pH buffer sets • Backlit graphic display • Printer option (1310P)
YSI TruLab Instruments
TruLab pH/ISE 1320 • Dual channel analog instrument • pH, mV(ORP), Temp, ISEs • Connect 2 ISEs simultaneously • Direct ion concentration readout for:
Ag+, Br-, Ca2+, Cd2+, Cl-, Cu2+, CN-, F-, I-, K+, Na+, NO3-, Pb2+, S2-, NH3, NH4
+, CO2 • Memory (500 manual, 5000 auto) • USB output; data importer software • 22 stored pH buffer sets • Backlit graphic display • Printer option (1320P)
YSI TruLab Instruments
Features of all models:
• 3 year warranty
• BNC connectors
• Calibration timer
• AutoRead for repeatability
• Electrode stand included
• Universal power supply
• USB cable and PC data importer included when applicable (1310/20)
• Wide range of pH electrode options for specialized applications
• Stored buffer sets (3 for 1110, 22 for 1310/20)
• Battery or AC operation; USB power for 1310 and 1320
TruLine, IoLine, and ScienceLine Electrodes
TruLine: 6 single junction pH electrodes and 2 ORP electrodes
• Glass or epoxy body, refillable, and temperature/non-temp
• 1 spear tip pH electrode for penetrating solids; 1 flat tip for surfaces
• Cover about 80% of pH lab applications
• Can be used with variety of meters (BNC)
• Fast response
• Refillable electrodes offer long, usable life
TruLine, IoLine, and ScienceLine Electrodes
IoLine: 2 triple junction glass electrodes with iodine/iodide reference • No contamination of sample with metal ions (Tris buffers) • 3-chamber reference system allows for stability in all media and at
changing temps; used in variety of applications • 1 micro version for small sample size
ScienceLine: 3 double junction electrodes with high electrolyte outflow • Unique applications requiring particular junction • Ground joint: highest outflow, best for solutions with a lot of solids,
useful for solutions with low ions • Platinum: high outflow to prevent clogging, but not as high as ground
joint; micro size available • Fast response; double junction reduces the chances that the
reference will touch the sample, avoiding possible contamination.
TruLine, IoLine, and ScienceLine Electrodes
Product line features:
• German engineered/manufactured
• Variety of electrode designs to meet a large range of applications
• 1 meter cable
• Longevity and fast response
• Competitive pricing
• Temperature included with all IoLine and ScienceLine, optional for TruLine
Measurement reminder: • Refilling opening must be OPEN during calibration and measurement
MultiLab: Digital Multiparameter Benchtop Instruments
• 4 instruments: - 4010-1(P), single channel - 4010-2, dual channel - 4010-3, triple channel
• All instruments include: - 3-year warranty! - MultiLab importer PC software - universal power supply - USB cable - electrode stand
• All instruments work with BOD Analyst Pro
YSI MultiLab Product Line
YSI MultiLab Instruments
MultiLab 4010-1
• Single channel digital instrument
• pH, mV(ORP), Temp, DO/BOD, cond. • Data storage -500 data sets in manual mode
-5,000 data sets in automatic logging mode
-Send data to PC (ASCII and .CSV); import directly into Excel®
• Large, easy to read backlit graphic display
YSI MultiLab Instruments
MultiLab 4010-2/3
• Two or three channel digital instrument
• pH, mV(ORP), Temp, DO/BOD, cond.
• ISE option with analog adapter -Covers one digital input
• Data storage -500 data sets in manual mode
-10,000 data sets in automatic logging mode
-Send data to PC, USB stick, or USB printer
• Large, easy to read color display
• Antibacterial keypad
•Digital sensor recognition, processing and data transfer
•Plug-and-play connectivity; probe auto-recognized
•Sensors store unique ID with serial number and calibration status
•Monitors sensor performance and calibration to notify when it is time to change sensor cap or refill/replace pH sensor
•Includes pH, ORP, conductivity and optical based self-stirring BOD
YSI MultiLab Intelligent Digital Sensors (IDS)
Ion Selective Electrodes: YSI TruLine ISEs
Measurement Basics: Nernst Equation
• ISEs measure ion activity (i.e. “effective concentration”) -Ion activity < actual ion concentration due to interaction of ions in solution, reduces ions available to react
• ISEs are designed to measure specific ions, but other ions can interfere
-Adding ionic strength adjustor acts to minimize/mask effect of interfering ions
• pH electrode is a Hydrogen ISE
• ISE output is in mV; follows Nernst Equation
• Slope of electrode: 2.303 ∗ 𝑅𝑅𝑅𝑅𝑛𝑛𝑛𝑛
-Divalent (e.g. Ca2+) ~ 29 mV/decade (typical 26-29)
-Must have decade (i.e. order of magnitude) difference in calibration standards; can “fill-in” between standards if many used
𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 Ca2+ = 2.303 ∗ 8.314 𝐽𝐽𝐾𝐾 𝑚𝑚𝑆𝑆𝑆𝑆 ∗ 298.15 𝐾𝐾 ∗
12 ∗
𝑚𝑚𝑆𝑆𝑆𝑆9.649 ∗ 104 𝐶𝐶 = 2.958 ∗ 10−2 𝐽𝐽 𝐶𝐶−1 = 29.58 𝑚𝑚𝑚𝑚
ISE Electrode Types
Half cell ISEs:
• Contain only the ion sensor (half cell); separate reference must be used
-Both have ion exchange with the sample
-mV output determined by difference in potential b/w the sensing half cell and the reference
Combination ISEs:
• Contain the sensing electrode and reference electrode in the same body
• Much easier to use; preferred by most customers
• YSI TruLine electrodes
Measurement Methods
Direct measurement:
• Simple and most commonly used ISE method
• Measure mV output in samples and determine concentration based on calibration (i.e. just as is done with pH)
Incremental methods:
• Standard addition, standard subtraction, sample addition, sample subtraction
• Methods of spiking the sample
• Sample ion concentration calculated from change in potential
• YSI TruLab 1320 and MultiLab 4010-2/3 have all incremental methods preprogrammed
Currently Available ISEs: YSI and Competition
YSI Brand A Brand B Brand C
NH3
NH4+
Br-
Cd2+
Ca2+
Cl-
Cu2+
CN-
F-
I-
Pb2+
NO3-
K+
Ag+/S2-
Na+
• 16 parameters
• All electrodes are refillable and include double-junction reference
• 4 sensor technologies
-Gas sensing, polymer/PVC membrane, solid state, glass sensor
• Included with electrode purchase: -Unfilled electrode
-1 replacement sensor module for polymer/PVC type electrodes due to shorter shelf life
-Reference fill solution, ionic strength adjustor (ISA), and small bottle of standard
• 9-12 months warranty depending on ISE
• Made in the U.S.A.
YSI TruLine ISE Electrodes
• Technology: Gas-sensing • Features: Replaceable Membrane Modules, pH sensor inside
• Range: 0.02-17,000 mg/L • Slope: 54-59 mv/decade • pH range: Must be 11 or higher to make NH3 gaseous. ISA has blue color at this pH
• Interferences: few, volatile amines • Applications/Notes: Popular for wastewater, aquaculture, agriculture, industrial. pH shift caused by ammonia gas is correlated to concentration
Ammonia (NH3)
• Technology: Polymer/PVC Membrane • Features: Refillable, Replaceable Sensor Modules
• Range: 0.02-1,800 mg/L • Slope: 54-59 mv/decade • pH range: 4-10 • Interferences: few, Na+, K+
• Applications/Notes: Popular for wastewater, environmental, industrial, food/beverage
Ammonium (NH4+)
• Technology: Solid State • Features: Refillable, non-replaceable sensor • Range: 0.4-79,900 mg/L • Slope: 54-59 mv/decade • pH range: <12 • Interferences: I-, Cl-, CN-, NH3
-
• Applications/Notes: Pool & spa, industrial
Bromide (Br-)
• Technology: Solid State • Features: Refillable, non-replaceable sensor • Range: 0.1-11,200 mg/L • Slope: 26-29 mv/decade • pH range: 2-8 • Interferences: Hg2+, Ag+, Cu2+, Pb2+
• Applications/Notes: Industrial, environmental
Cadmium (Cd2+)
• Technology: Polymer/PVC Membrane • Features: Refillable, Replaceable Sensor Modules
• Range: 0.02-40,000 mg/L • Slope: 25-29 mv/decade • pH range: 2.5-11 • Interferences: Pb2+, Hg2+, Si2+, Mg2+, Cu2+, K+, Tris+, Na+, others
• Applications/Notes: Popular for Aquarium, industrial, food/beverage
Calcium (Ca2+)
• Technology: Solid State • Features: Refillable, non-replaceable sensor • Range: 1.8-35,500 mg/L • Slope: 54-59 mv/decade • pH range: 2-12 • Interferences: CN-, Br-, I-, OH-, S2-, NH3
• Applications/Notes: Very popular for industrial, wastewater, drinking water, food/beverage, environmental. This sensor is slightly photosensitive – store in dark place when not in use.
Chloride (Cl-)
• Technology: Solid State • Features: Refillable, non-replaceable sensor • Range: 0.6-6,400 mg/L • Slope: 26-29 mv/decade • pH range: 2-6 • Interferences: Hg2+, Ag+, Fe2+, Br-, Cl- • Applications/Notes: Industrial, environmental. Also known as Cupric ISE.
Copper (Cu2+)
• Technology: Solid State • Features: Refillable, non-replaceable sensor • Range: 0.2-260 mg/L • Slope: 54-59 mv/decade • pH range: 10-14 • Interferences: I-, Br-, Cl-, S2- • Applications/Notes: Industrial, environmental. Standard solutions not included in electrode package (safety issue)
Cyanide (CN-)
• Technology: Solid State Crystal • Features: Refillable, non-replaceable sensor • Range: 0.02 mg/L - saturation • Slope: 54-59 mv/decade • pH range: 5-7 at low range, 11 at higher range • Interferences: OH- • Applications/Notes: Very popular for drinking water, industrial. Lanthanum fluoride crystal is the sensor.
Fluoride (F-)
• Technology: Solid State • Features: Refillable, non-replaceable sensor • Range: 0.006-127,000 mg/L • Slope: 54-59 mv/decade • pH range: <14 • Interferences: CN-, Cl-, S2O3
2-, S2-, NH3 • Applications/Notes: Drinking water, industrial, food/beverage
Iodide (I-)
• Technology: Solid State • Features: Refillable, non-replaceable sensor • Range: 0.02-20,700 mg/L • Slope: 26-30 mv/decade • pH range: 4-7 • Interferences: Hg2+, Ag+, Cu2+, Fe2+, Cd2+ • Applications/Notes: Industrial, environmental
Lead (Pb2+)
• Technology: Polymer/PVC Membrane • Features: Refillable, Replaceable Sensor Modules
• Range: 0.4-62,000 mg/L • Slope: 54-59 mv/decade • pH range: 2.5-11 • Interferences: few, ClO4
-, ClO3-, I-, F-
• Applications/Notes: Very popular for environmental, industrial, agriculture, food/beverage
Nitrate (NO3-)
• Technology: Polymer/PVC Membrane • Features: Refillable, Replaceable Sensor Modules
• Range: 0.04-39,000 mg/L • Slope: 54-59 mv/decade • pH range: 2-12 • Interferences: Cs+, NH4
+, Tl+, H+, Ag+,Tris+, Na+, Li+
• Applications/Notes: Popular for agriculture, food/beverage, environmental, industrial
Potassium (K+)
• Technology: Solid State • Features: Refillable, non-replaceable sensor • Range: 0.01-107,900 mg/L (Silver) 0.003-32,100 mg/L (Sulfide)
• Slope: 54-59 mv/decade (Silver) 26-29 mv/decade (Sulfide)
• pH range: 2-12 • Interferences: Hg2+ • Applications/Notes: Industrial, environmental, food/beverage. Sulfide does not come with standard solution (stability)
Silver/Sulfide (Ag+/S2-)
• Technology: Glass Sensor • Features: Refillable, non-replaceable sensor • Range: 0.1-23,000 mg/L • Slope: 54-59 mv/decade • pH range: >9 • Interferences: H+, K+ • Applications/Notes: Industrial, agriculture, power, food/beverage. Looks like a pH electrode!
Sodium (Na+)
TruLine ISE Tips
• Takes ~1-2 mL to fill electrode
• One order of magnitude (i.e. decade) difference between two standards; can include standards between
• Add ISA to both samples and standards to ensure similar ionic strength
• Short term (1-3 day) storage: place in low standard and fill the reference
• Long term: store dry (i.e. drain reference) and avoid sunlight
• Fill hole should be open during calibration and measurement, closed during storage