Troubleshooting Guide - Veeder
Transcript of Troubleshooting Guide - Veeder
Manual No: 577013-819 ● Revision: E
Troubleshooting Guide
In-Station Diagnostics
Notice
Veeder-Root makes no warranty of any kind with regard to this publication, including, but not limited to, the implied warranties ofmerchantability and fitness for a particular purpose.
Veeder-Root shall not be liable for errors contained herein or for incidental or consequential damages in connection with thefurnishing, performance, or use of this publication.
Veeder-Root reserves the right to change system options or features, or the information contained in this publication.
This publication contains proprietary information which is protected by copyright. All rights reserved. No part of this publicationmay be photocopied, modified or translated to another language without the prior written consent of Veeder-Root.
Contact TLS Systems Technical Support for additional troubleshooting information at 800-323-1799.
DAMAGE CLAIMS / LOST EQUIPMENT
Thoroughly examine all components and units as soon as they are received. If any cartons are damaged or missing, write acomplete and detailed description of the damage or shortage on the face of the freight bill. The carrier's agent must verify theinspection and sign the description. Refuse only the damaged product, not the entire shipment.
Veeder-Root must be notified of any damages and/or shortages within 30 days of receipt of the shipment, as stated in our Termsand Conditions.
VEEDER-ROOT’S PREFERRED CARRIER
1. Contact Veeder-Root Customer Service at 800-873-3313 with the specific part numbers and quantities that were missingor received damaged.
2. Fax signed Bill of Lading (BOL) to Veeder-Root Customer Service at 800-234-5350.
3. Veeder-Root will file the claim with the carrier and replace the damaged/missing product at no charge to the customer.Customer Service will work with production facility to have the replacement product shipped as soon as possible.
CUSTOMER’S PREFERRED CARRIER
1. It is the customer’s responsibility to file a claim with their carrier.
2. Customer may submit a replacement purchase order. Customer is responsible for all charges and freight associated withreplacement order. Customer Service will work with production facility to have the replacement product shipped as soon aspossible.
3. If “lost” equipment is delivered at a later date and is not needed, Veeder-Root will allow a Return to Stock without a restockingfee.
4. Veeder-Root will NOT be responsible for any compensation when a customer chooses their own carrier.
RETURN SHIPPING
For the parts return procedure, please follow the appropriate instructions in the "General Returned Goods Policy” pages in the"Policies and Literature" section of the Veeder-Root North American Environmental Products price list. Veeder-Root will notaccept any return product without a Return Goods Authorization (RGA) number clearly printed on the outside of the package.
©Veeder-Root 2010. All rights reserved.
Warranty
TLS-350R, TLS-350 PLUS, TLS-350J MONITORING SYSTEMS.
We warrant that this product shall be free from defects in material and workmanship for a period of one (1) year from the date ofinstallation or twenty-four (24 months) from the date of invoice, whichever occurs first. During the warranty period, we or ourrepresentative will repair or replace the product, if determined by us to be defective, at the location where the product is in useand at no charge to the purchaser. LAMPS, FUSES, AND LITHIUM BATTERIES ARE NOT COVERED UNDER THISWARRANTY.
We shall not be responsible for any expenses incurred by the user.
This warranty applies only when the product is installed in accordance with Veeder-Root’s specifications, and a WarrantyRegistration and Checkout Form has been filed with Veeder-Root by an authorized Veeder-Root Distributor. This warranty will notapply to any product which has been subjected to misuse, negligence, accidents, systems that are misapplied or are not installedper Veeder-Root specifications, modified or repaired by unauthorized persons, or damage related to acts of God.
If “Warranty” is purchased as part of the Fuel Management Service, Veeder-Root will maintain the equipment for the life of thecontract in accordance with the written warranty provided with the equipment. A Veeder-Root Fuel Management ServicesContractor shall have free site access during Customer’s regular working hours to work on the equipment. Veeder-Root has noobligation to monitor federal, state or local laws, or modify the equipment based on developments or changes in such laws.
MODULES, KITS, OTHER COMPONENTS (PARTS PURCHASED SEPARATE OF A COMPLETE CONSOLE).
We warrant that this product, exclusive of lithium batteries, shall be free from defects in material and workmanship for a period offifteen (15) months from date of invoice. We warrant that the lithium batteries shall be free from defects in material andworkmanship for a period of 90 days from date of invoice. We will repair or replace the product if the product is returned to us;transportation prepaid, within the warranty period, and is determined by us to be defective. This warranty will not apply to anyproduct which has been subjected to misuse, negligence, accidents, systems that are misapplied or are not installed per Veeder-Root specifications, modified or repaired by unauthorized persons, or damage related to acts of God.
We shall not be responsible for any expenses incurred by the user.
Table of Contents
iv
IntroductionContractor Certification Requirements ..............................................................................1Related Manuals ...............................................................................................................1Safety Precautions ............................................................................................................2
ISD Post-Installation ChecklistISD with Healy Assist System CheckList .........................................................................4ISD with VST Balance system/VST EMC Membrane Processor CheckList .....................6ISD with VST Balance system/VR Vapor Polisher CheckList ...........................................7
ISD Alarm Message DiagnosticsISD VAPOR LEAKAGE ....................................................................................................8
ISD Monitoring Category ..........................................................................................8Diagnostic Procedure ...............................................................................................8Common Causes......................................................................................................8
ISD GROSS PRESSURE .................................................................................................9ISD Monitoring Category ..........................................................................................9Common Causes......................................................................................................9Field Notes ...............................................................................................................9
DEGRD PRESSURE ......................................................................................................10ISD Monitoring Category ........................................................................................10Diagnostic Procedure .............................................................................................10Common Causes....................................................................................................10Field Notes .............................................................................................................10
GROSS COLLECT .........................................................................................................11ISD Monitoring Category ........................................................................................11Diagnostic Procedure .............................................................................................11Common Causes....................................................................................................11
DEGRD COLLECT .........................................................................................................12ISD Monitoring Category ........................................................................................12Diagnostic Procedure .............................................................................................12
FLOW COLLECT ............................................................................................................13ISD Monitoring Category ........................................................................................13Diagnostic Procedure .............................................................................................13
ISD VP STATUS .............................................................................................................14ISD Monitoring Category ........................................................................................14Diagnostic Procedure .............................................................................................14Common Causes....................................................................................................14Field Notes .............................................................................................................14
ISD VP PRESSURE .......................................................................................................15ISD Monitoring Category ........................................................................................15Diagnostic Procedure .............................................................................................15Field Notes .............................................................................................................15
ISD SENSOR OUT .........................................................................................................16ISD Monitoring Category ........................................................................................16Diagnostic Procedure .............................................................................................16Common Causes....................................................................................................16
ISD SETUP .....................................................................................................................17ISD Monitoring Category ........................................................................................17Diagnostic Procedure .............................................................................................17Common Causes....................................................................................................17
MISSING RELAY SETUP ...............................................................................................18ISD Monitoring Category ........................................................................................18Diagnostic Procedure .............................................................................................18Common Causes....................................................................................................18
Table of Contents
v
MISSING TANK SETUP .................................................................................................19ISD Monitoring Category ........................................................................................19Diagnostic Procedure .............................................................................................19
MISSING HOSE SETUP ................................................................................................20ISD Monitoring Category ........................................................................................20Diagnostic Procedure .............................................................................................20Common Causes....................................................................................................20
MISSING VAPOR FLOW MTR .......................................................................................21ISD Monitoring Category ........................................................................................21
MISSING VAPOR PRES SEN ........................................................................................22ISD Monitoring Category ........................................................................................22
fnn: CHK VAPOR FLOW MTR ........................................................................................23ISD Monitoring Category ........................................................................................23
hnn: VPRFLOW MTR SETUP ........................................................................................24ISD Monitoring Category ........................................................................................24Diagnostic Procedure .............................................................................................24
PMC Alarm Message DiagnosticsVeeder- Root Vapor Polisher Site Requirements ..........................................................25VST ECS Membrane Processor Site Requirements .......................................................25PMC SETUP ...................................................................................................................26
ISD Monitoring Category ........................................................................................26VST ECS MEMBRANE DIAGNOSTIC PROCEDURE ...........................................26V-R VAPOR POLISHER DIAGNOSTIC PROCEDURE .........................................27
PMC Sensor Fault ..........................................................................................................28MISSING VP INPUT .......................................................................................................29
ISD Monitoring Category ........................................................................................29Diagnostic Procedure .............................................................................................29Field Notes .............................................................................................................29
VP EMISSIONS ..............................................................................................................30ISD Monitoring Category ........................................................................................30Diagnostic Procedure .............................................................................................30Common Causes....................................................................................................30Field Notes .............................................................................................................30
VP PRESSURE ..............................................................................................................31ISD Monitoring Category ........................................................................................31Diagnostic Procedure .............................................................................................31Field Notes .............................................................................................................31
VP DUTY CYCLE ...........................................................................................................32ISD Monitoring Category ........................................................................................32ISD Monitoring Category ........................................................................................32Diagnostic Procedure .............................................................................................32Field Notes .............................................................................................................32
Exhibit 11 Failures ..........................................................................................................32Pressure Integrity test.............................................................................................32Flow Test ................................................................................................................32Thermometer Test ..................................................................................................33
Exhibit 12 Failures ..........................................................................................................33
Miscellaneous MaintenanceRepairing Collection Vapor Recovery Equipment ...........................................................34
All Systems (Healy) ................................................................................................34Assist Systems (Healy)...........................................................................................34Balance Systems....................................................................................................34
Removing & Replacing Vapor Flow Meters ....................................................................34
Table of Contents
vi
ISD/PMC Diagnostic Menus ...................................................................................35
ISD ReportsDaily Detail Report ..........................................................................................................38
Daily Detail Status Codes.......................................................................................39Collection Report FP Ordering ...............................................................................41Containment Degradation Results..........................................................................42ISD Status Missing Text .........................................................................................42Missing All Reports for One Day ............................................................................42High A/L on All Nozzles ..........................................................................................44
Diagnosing CVLD Reports ..............................................................................................45
TablesTable 1. Smart Sensor Device Alarm and Fault Summary ....................................28Table 2. Clear Test Repair Menu ..........................................................................37
1
Introduction
This manual contains troubleshooting procedures for Veeder-Root In-Station Diagnostics (ISD).
Contractor Certification Requirements
Veeder-Root requires the following minimum training certifications for contractors who will install and setup the equipment discussed in this manual:
Installer (Level 1) Certification: Contractors holding valid Installer Certification are approved to perform wiring and conduit routing; equipment mounting; probe, sensor and carbon canistor vapor polisher installation; tank and line preparation; and line leak detector installation.
TLS-350 Technician (Level 2/3 or 4) Certification: Contractors holding valid TLS-350 Technician Certifications are approved to perform installation checkout, startup, programming and operations training, troubleshooting and servicing for all Veeder-Root TLS-300 or TLS-350 Series Tank Monitoring Systems, including Line Leak Detection and associated accessories.
In-Station Diagnostics (ISD-PMC) Technician Certification: ISD PMC Contractors holding a valid ISD/PMC Certification are approved to perform (ISD/PMC) installation checkout, startup, programming, and operations training. This training also includes troubleshooting and service techniques for the Veeder-Root In-Station Diagnostics system. A current Veeder-Root Technician Certification is a prerequisite for the ISD/PMC course.
Veeder-Root ISD/PMC Including Carbon Canister Vapor Polisher Contractor Certification: This Certification includes Executive Orders 203, 204 and the Veeder-Root Vapor Polisher. This certification is required for setup and service of the Veeder-Root Vapor Polisher.
Warranty Registrations may only be submitted by selected Distributors.
Related Manuals
577013-800 ISD Install, Setup and Operation Manual
577013-842 ISD Quick Reference Guide
577093-796 Vapor Flow Meter Installation Manual
577093-797 Pressure Sensor Installation Manual
576013-635 Veeder-Root Serial Interface Manual
576013-818 TLS-3XX Troubleshooting Manual (Feature / Version list Section 3, Table 3.-3)
577013-948 PMC Install, Setup & Operation Manual For Veeder-Root Vapor Polishers
577013-801 PMC Install, Setup & Operation Manual For VST ECS Membrane Processors
577013-937 ISD Install, Setup & Operation Manual for VST ECS Membrane Processors and Veeder-Root Vapor Polisher
577013-920 Carbon Canister Vapor Polisher Installation Manual.
Introduction Safety Precautions
2
Safety Precautions
FAILURE TO COMPLY WITH THE FOLLOWING WARNINGS AND SAFETY PRECAUTIONS COULD CAUSE DAMAGE TO PROPERTY, ENVIRONMENT, RESULTING IN SERIOUS INJURY OR DEATH.
The following safety symbols are used throughout this manual to alert you to important safety hazards and precautions.
EXPLOSIVEFuels and their vapors are extremely explosive if ignited.
FLAMMABLEFuels and their vapors are extremely flammable.
ELECTRICITYHigh voltage exists in, and is supplied to, the device. A potential shock hazard exists.
TURN POWER OFFLive power to a device creates a potential shock hazard. Turn Off power to the device and associ-ated accessories when servicing the unit.
WARNINGHeed the adjacent instructions to avoid equip-ment damage or personal injury.
READ ALL RELATED MANUALSKnowledge of all related procedures before you begin work is important. Read and understand all manuals thoroughly. If you do not understand a procedure, ask someone who does.
OFF
3
ISD Post-Installation Checklist
Refer to Post Installation Checklist to review installation. If you should experience Low Or No VL Ratio during the Operability test, additional information on those results is discussed after the checklist.
DATE
SERVICE COMPANY NAME TELEPHONE
SERVICE TECHNICIAN VEEDER-ROOT TECH CERTIFICATION #
TRAINING LEVEL (CIRCLE ONE): ISD TRAINED?
STATION NAME
STATION ADDRESS CITY
WARNING!
Before installing any vapor recovery equipment the installer must be familiar with all state and federal regulations regarding the safe installation and operation of each component. (NFPA 30A).
A level 2, 3 or 4 certified Veeder-Root contractor or Authorized Service Contractor (ASC) with Veeder-Root In-Station Diagnostics training must be available (on-site) to over-see this post-installation checklist.
Review and comply with all the safety warnings in the installation manuals and any other national, State or Local requirements.
For a complete list of precautions, please consult the Veeder-Root ISD manuals.
Required Reference Manuals
In-Station Diagnostic Install, Setup and Operation, Manual No. 577013-800
In-Station Diagnostic System Troubleshooting Guide, Manual No. 577013-819
Tools Needed
The following program is available and provided to level 2, 3 or 4 certified contractors who have passed the Veeder-Root In-Station Diagnostics training course:
ISD PC Setup Tool, V-R P/N 332333-001 (requires a laptop with the appropriate cabling to make an RS-232 connection to the Veeder-Root TLS Console)
2/3 OR 4
STATE ZIP
ISD Post-Installation Checklist ISD with Healy Assist System CheckList
4
ISD with Healy Assist System CheckList
ProcedureThe following recommended procedure can be followed at the completion of the ISD software setup:
STEP 1. EVR TYPE is set to VACUUM ASSIST?
STEP 2. VACUUM ASSIST TYPE is set to HEALY VAC?
STEP 3. NOZZLE A/L RANGE MAX is set to 1.15 and MIN is set to 0.95?
STEP 4. VAPOR PROCESSOR TYPE is set to NO VAPOR PROCESSOR?
STEP 5.There is a wired and ENABLED “AIRFLOW METER” (i.e. ISD Vapor Flow Meter) in each vapor recovery dispenser?
STEP 6. There is a wired and ENABLED “PRESSURE SENSOR” (i.e. ISD Vapor Pressure Sensor)?
STEP 7. The FUEL HOSE TABLE is setup and filled out for each vapor recovery hose?
STEP 8. The ISD FUEL GRADE HOSE MAP does not contain entries for non vapor recovery hoses (e.g. diesel)?
STEP 9.On the TLS press the MODE key until the DIAGNOSTIC MODE menu is displayed (this will cause a TLS console System Self Test).
STEP 10. Using the Troubleshooting Guide respond to all ISD Setup ALARMS posted on the printer tape.
STEP 11.Repeat Steps 8 & 9 until there are no ISD setup or self-test alarms. The TLS Console display reads ALL FUNCTIONS NORMAL.
STEP 12.Using the ISD PC Setup Tool and the ISD Vapor Pressure Sensor calibration valve: The ISD VaporPressure Sensor reads an ambient pressure reading with an offset no greater then +/- 0.20 IWC?
STEP 13.Returned the ISD Vapor Pressure Sensor calibration valve so that the sensor is reading UST vapor pressure?
STEP 14. Using the ISD PC Setup Tool: An ISD A/L reading is coming in for each gas hose at the location?
STEP 15. The TLS console clock is set to the correct date & time?
ISD Post-Installation Checklist ISD with Healy Assist System CheckList
5
Question
Why are the V/L extremely low or non existent when running tests with the PC Setup Tool immediately after installation?
Explanation
1. Refer to Post Installation Checklist (on page 3) to review installation.
2. If the V/L are non-existent or extremely low (50%) for all nozzles during the Operability Test - it indicates an air flow meter problem.
- If the BirProtocolDim is installed make sure there is a 'G' in the DIM string.
- Healy Vacuum pump V/L needs to be set.
- Check the ball valve between the Healy pump and air flow meter is not closed or partially closed.
- Check that the installed meter does not still have the dust caps on. This will significantly reduce airflow for both all nozzles on the dispenser.
See fnn: CHK VAPOR FLOW MTR troubleshooting procedures (on page 23).
Question
How come some of the SmartSensor airflow meters do not show up in the EVR/ISD > AIRFLOW METER SELECT setup menu so I can ENABLE them?
Explanation
• ISD can only use SmartSensors numbered 1 - 26. Ensure that all the ISD SmartSensors are numbered 1 - 26 by moving the ISD SmartSensors to be ahead of other SmartSensors installed in the system. If SmartSensors were moved to change the SmartSensor numbers of the airflow meters into the range of 1 - 26 you may need to cold boot the system before ISD will add them to the list of available airflow meters.
ISD Post-Installation Checklist ISD with VST Balance system/VST EMC Membrane Processor CheckList
6
ISD with VST Balance system/VST EMC Membrane Processor CheckList
ProcedureThe following recommended procedure can be followed at the completion of the ISD software setup for VST Balance Sys-tems with VST EMC Membrane Processor:
STEP 1. EVR TYPE is set to BALANCE?
STEP 2. The Balance Nozzle Type is VST?
STEP 3. The Vapor Processor set to VST Vapor Processor?
STEP 4. There is a wired and ENABLED Hydrocarbon Sensor?
STEP 5.There is a wired and ENABLED ‘AIRFLOW METER’ (i.e. ISD Vapor Flow Meter) in each vapor recovery dispenser?
STEP 6. There is a wired and ENABLED ‘PRESSURE SENSOR’ (i.e. ISD Vapor Pressure Sensor)?
STEP 7. The FUEL HOSE TABLE is setup and filled out for each vapor recovery hose?
STEP 8. The ISD FUEL GRADE HOSE MAP does not contain entries for non vapor recovery hoses (e.g. diesel)?
STEP 9. PMC setup Vapor Processor Max Runtime is set to 30 minutes?
STEP 10. PMC setup Turn Off Vapor Processor is set to -0.2 IWC?
STEP 11. PMC setup Turn On Vapor Processor is set to +0.2 IWC?
STEP 12.On the TLS press the MODE key until the DIAGNOSTIC MODE menu is displayed (this will cause a TLS console System Self Test).
STEP 13. Using the Troubleshooting Guide respond to all ISD Setup ALARMS posted on the printer tape.
STEP 14.Repeat Steps 12 & 13 until there are no ISD setup or self-test alarms. The TLS Console display reads ALL FUNCTIONS NORMAL.
STEP 15.Using the ISD PC Setup Tool and the ISD Vapor Pressure Sensor calibration valve: The ISD Vapor Pressure Sensor reads an ambient pressure reading with an offset no greater then ± 0.20 IWC?
STEP 16.Returned the ISD Vapor Pressure Sensor calibration valve so that the sensor is reading UST vapor pressure?
STEP 17. Using the ISD PC Setup Tool: An ISD A/L reading is coming in for each gas hose at the location?
STEP 18. The TLS console clock is set to the correct date & time?
ISD Post-Installation Checklist ISD with VST Balance system/VR Vapor Polisher CheckList
7
ISD with VST Balance system/VR Vapor Polisher CheckList
ProcedureThe following recommended procedure can be followed at the completion of the ISD software setup for VST Balance Sys-tems with Veeder-Root Vapor Polisher:
STEP 1. EVR TYPE is set to BALANCE?
STEP 2. The Balance Nozzle Type is VST?
STEP 3. The Vapor Processor set to Veeder-Root Polisher?
STEP 4.There is a wired and ENABLED ‘AIRFLOW METER’ (i.e. ISD Vapor Flow Meter) in each vapor recovery dispenser?
STEP 5. There is a wired and ENABLED ‘PRESSURE SENSOR’ (i.e. ISD Vapor Pressure Sensor)?
STEP 6. The FUEL HOSE TABLE is setup and filled out for each vapor recovery hose?
STEP 7. The ISD FUEL GRADE HOSE MAP does not contain entries for non vapor recovery hoses (e.g. diesel)?
STEP 8.On the TLS press the MODE key until the DIAGNOSTIC MODE menu is displayed (this will cause a TLS console System Self Test).
STEP 9. Using the Troubleshooting Guide respond to all ISD Setup ALARMS posted on the printer tape.
STEP 10.Repeat Steps 8 & 9 until there are no ISD setup or self-test alarms. The TLS Console display reads ALL FUNCTIONS NORMAL.
STEP 11.Using the ISD PC Setup Tool and the ISD Vapor Pressure Sensor calibration valve: The ISD Vapor Pressure Sensor reads an ambient pressure reading with an offset no greater then ±0.20 IWC?
STEP 12.Returned the ISD Vapor Pressure Sensor calibration valve so that the sensor is reading UST vapor pressure?
STEP 13. Using the ISD PC Setup Tool: An ISD A/L reading is coming in for each gas hose at the location?
STEP 14. The TLS console clock is set to the correct date & time?
8
ISD Alarm Message Diagnostics
ISD VAPOR LEAKAGE
A Vapor Leakage Detection test failure occurs when the vapor recovery containment system leaks at 2 times the allowable CARB standard defined in the TP-201.3. For a typical 12-hose site, that means it exceeds 8.5cfh (limit ranges over 8-10 cfh for <6 to >24 hoses).
ISD MONITORING CATEGORY
Vapor Containment Monitoring as implemented by over-pressurization and Vapor Leakage Detection tests, is responsible for ensuring that UST ullage pressure and system leak rate stay within regulatory boundaries. The Vapor Leakage Detection test will check all components for leaks including: P/V valves, dispenser piping, vacuum assist motors, nozzles, hoses, breakaways, vapor processor piping and elements as well as any other piping and fitting or component connected into the UST ullage space.
DIAGNOSTIC PROCEDURE
• Run a TP 201.3 test and use common field techniques for determining the source of the containment system leak.
• Fix all FLOW COLLECT alarms in a balance site first prior to diagnosing this alarm.
COMMON CAUSES
3. Phase I equipment, including but not limited to: dry breaks, spill buckets, tank inventory gauge caps, P/V valves, fill caps.
4. Phase II equipment, including but not limited to: breakaways, hoses, nozzles, whips, vacuum assist motors, solenoid valves, vapor shear valves.
5. Check drop tubes for leaks.
Refer to the Clear Test Repair Menu (on page 37) for clearing this alarm on the TLS after repairs are complete.
ISD Alarm Message Diagnostics ISD GROSS PRESSURE
9
ISD GROSS PRESSURE
A gross over-pressure failure occurs when the 95th percentile of 7-days’ ullage pressure data exceeds the gross over-pressure requirement. That means the pressure is greater than 1.3"wc for more than 1.2 hours a day on average or more than 8.4 hours for 7 days.
ISD MONITORING CATEGORY
Vapor Containment Monitoring as implemented by over-pressurization and Vapor Leakage Detection tests, is responsible for ensuring that UST ullage pressure and system leak rate stay within regulatory boundaries. The Vapor Leakage Detection test will check all components for leaks including: P/V valves, dispenser piping, vacuum assist motors, nozzles, hoses, breakaways, vapor processor piping and elements as well as any other piping and fitting or component connected into the UST ullage space.
COMMON CAUSES
1. Failure of pressure management control devices (e.g. check to make that the Healy Clean Air Separator is properly connected to UST's vapor space)
2. Failure of ORVR blocking Vapor Collection systems, including but not limited to: ORVR detecting assist nozzles
a. Check nozzle flow rates, change fuel filters on dispensers with low flow rates.
b. Check nozzle boots to make sure they are not torn
3. Faulty Phase I equipment leading to persistent over pressure conditions during bulk deliveries. (Refer maintenance personnel to EVR manufacturer troubleshooting guides for diagnostic procedures.)
4. Failure of ISD pressure sensor.
Refer to the Clear Test Repair Menu (on page 37) for clearing this alarm on the TLS after repairs are complete.
FIELD NOTES
1. Check that the EVR fittings are tight in the dispenser. - Loose fittings caused excessive leaks between the nozzle and vacuum pump which led to over collection, and the nozzle
not identifying ORVR vehicles. These conditions lead to an overpressure condition. - If the nozzle hoods are broken or worn this can lead to an over pressure condition. It can also lead to Gross Collect
alarms on nozzles that do not have a problem. - If the nozzle that triggered the alarm checks out OK, check the boots on the other nozzles at the site and repair the
damaged ones. If the rubber boot on a nozzle is broken or worn, or the fuel filter is clogged lowering the fuel flow rate, it will cause the nozzle to not correctly identify ORVR vehicles. When there are several broken at the site, nozzles with good boots looked blocked to ISD and trigger this alarm.
- This condition will also lead to IsdGrossPressure warning also. When nozzles fail to identify ORVR vehicles because of the worn boot or clogged fuel filter, too much air is ingested into the underground tank which lead to an over pressure condition.
- Clogged fuel filters will reduce fuel flow rate can contribute to overpressure conditions.
2. Some regulator tests can cause warning.
Check if a 2 inch pressure decay test on the day the IsdGrossPressure alarm posted. If that is the case - the alarm will clear on its own in 7 days unless there is another contributing problem. We do not recommend using Clear Test Repair Menu to reset test data unless a repair has been documented.
ISD Alarm Message Diagnostics DEGRD PRESSURE
10
DEGRD PRESSURE
A degradation over-pressure failure occurs when the 75th percentile of 30-days’ ullage pressure data exceeds the degradation over-pressure requirement. That means the pressure is greater than 0.3"wc for more than 6 hrs a day on average or more than 7.5 days for 30 days.
ISD MONITORING CATEGORY
Vapor Containment Monitoring as implemented by over-pressurization and Vapor Leakage Detection tests, is responsible for ensuring that UST ullage pressure and system leak rate stay within regulatory boundaries. The Vapor Leakage Detection test will check all components for leaks including: P/V valves, dispenser piping, vacuum assist motors, nozzles, hoses, breakaways, vapor processor piping and elements as well as any other piping and fitting or component connected into the UST ullage space.
DIAGNOSTIC PROCEDURE
Perform an operability test on the pressure sensor to ensure the ISD system is not causing the fault. See ISD Setup & Operation Manual in Related Manuals.
COMMON CAUSES
• See ISD Gross Pressure (page 9)
• Refer to the Clear Test Repair Menu (on page 37) for clearing this alarm on the TLS after repairs are complete.
FIELD NOTES
1. Check that the EVR fittings are tight in the dispenser (see ISD GROSS PRESSURE, Field Note 1 on page 9).
ISD Alarm Message Diagnostics GROSS COLLECT
11
GROSS COLLECT
1-Day Gross A/L Test failure occurs when the A/L ratio is at least 75% below the lower certified A/L ratio or at least 75% above the upper certified ratio.
ISD MONITORING CATEGORY
Vapor Collection Monitoring for sites equipped with assist vapor recovery systems is responsible for ensuring that proper front-end vapor capture is occurring during fueling events. Among other components, front-end equipment includes the nozzle, hose and the breakaway. Vapor collection is assisted by vacuum motor(s) located inside the dispenser vapor return piping.
DIAGNOSTIC PROCEDURE
1. Find the hose that triggered the alarm by looking at the Collection Tests in the ISD Daily Report in the normal menu.
2. A certified technician for the EVR equipment manufacturer must inspect the equipment.
Refer to the Clear Test Repair Menu (on page 37) for clearing this alarm on the TLS after repairs are complete.
COMMON CAUSES
• Failure of the EVR equipment.
• GrossCollectAllHoses: All hoses have extremely high A/L.
The most common cause of High A/L on all Nozzles at the same time is the dispenser events are artificially low. This can be caused by the DIM not having a 'G' in the DIM setup string. The only DIM that requires this is the BIR Protocol Dispenser Interface Module P/N 330280-001.
FIELD NOTES
If the nozzle that triggered the alarm checks out OK, check the boots on the other nozzles at the site and repair the damaged ones.
This alarm should not occur at a VST Balance site. If it does check EVR Type setup in ISD.
---------------------COLLECTION TEST GROSS DGRDA/L (#) A/L(#)
FP01: UNLEADED1.13 (288) W-0.10(288) : :FP: LABEL1.13 (188) -0.10 (188) ISD-EVR/fig1.eps
These lines appear ifEVR type=VACUUM ASSIST
W - Warning is on FP 1 unleaded hose.
ISD Alarm Message Diagnostics DEGRD COLLECT
12
DEGRD COLLECT
A 7-day Degradation A/L Test failure occurs when the A/L ratio is at least 25% below the lower certified A/L ratio or at least 25% above the upper certified ratio.
ISD MONITORING CATEGORY
Vapor Collection Monitoring for sites equipped with assist vapor recovery systems is responsible for ensuring that proper front-end vapor capture is occurring during fueling events. Among other components, front-end equipment includes the nozzle, hose and the breakaway. Vapor collection is assisted by vacuum motor(s) located inside the dispenser vapor return piping.
DIAGNOSTIC PROCEDURE
• Perform the TP 201.5 A/L test. To find the hose that triggered the alarm by looking at the Collection Tests in the ISD Daily Report in the normal menu.
• Refer to the Clear Test Repair Menu (on page 37) for clearing this alarm on the TLS after repairs are complete.
FIELD NOTES
This alarm should not occur at a VST Balance site. If it does check EVR Type setup in ISD.
ISD Alarm Message Diagnostics FLOW COLLECT
13
FLOW COLLECT
A 1-Day Vapor Collection Flow Performance Test failure occurs when the vapor collection flow performance is less than 50%.
ISD MONITORING CATEGORY
Vapor Collection Monitoring for sites equipped with balance vapor recovery systems is responsible for ensuring that front-end vapor collection is operating within certified range with respect to the baseline collection performance during fueling events. The baseline collection performance is established by CARB and is the V/L standard for the system. Among other components, front-end equipment includes the nozzle, hose and the breakaway. A balance vapor recovery system relies upon a tight bellows seal in order to allow displaced vapors to return to the UST during fueling events.
DIAGNOSTIC PROCEDURE
• Perform the TP201.4 test.
• Find the hose that triggered the alarm by looking at the Collection Tests in the ISD Daily Report in the normal menu.
• Refer to the Clear Test Repair Menu (on page 37) for clearing this alarm on the TLS after repairs are complete.
FIELD NOTES
This alarm should not occur at a Healy Assist site. If it does check EVR Type setup in ISD.
ISD Alarm Message Diagnostics ISD VP STATUS
14
ISD VP STATUS
1-Day Processor Test
ISD MONITORING CATEGORY
Vapor Processor Monitoring is responsible for ensuring that the vapor processor is operating within normal parameters with no risk of releasing excessive HC effluent emissions.
DIAGNOSTIC PROCEDURE
• Refer the problem to a certified technician for the processor manufacturer.
• Refer to the Clear Test Repair Menu (on page 37) for clearing this alarm on the TLS after repairs are complete.
COMMON CAUSES
• TLS Console Controlled Processor (VST)
- VP EMISSIONS FAIL
- VP DUTY CYCLE FAIL
• Non-TLS Console Controlled Processor (OPW)
- Inn: EXTERNAL INPUT ALARM
FIELD NOTES
The Healy EVR does not have a vapor processor so you should not see this type of alarm.
ISD Alarm Message Diagnostics ISD VP PRESSURE
15
ISD VP PRESSURE
1 Day Over-Pressure Test
ISD MONITORING CATEGORY
Vapor Processor Monitoring is responsible for ensuring that the vapor processor is operating within normal parameters with no risk of releasing excessive HC effluent emissions.
DIAGNOSTIC PROCEDURE
• TLS controlled processor:
- Put the processor in manual ON mode and see if the pressure drops. Perform the operability test required by the processor.
• Non-TLS controlled processor:
- Clear all collection alarms prior to fixing this problem.
- Verify processor operation.
Refer to the Clear Test Repair Menu (on page 37) for clearing this alarm on the TLS after repairs are complete.
FIELD NOTES
The Healy EVR does not have a vapor processor so you should not see this type of alarm.
ISD Alarm Message Diagnostics ISD SENSOR OUT
16
ISD SENSOR OUT
System Setup Self-Test
ISD MONITORING CATEGORY
System Self-Test Monitoring algorithms are designed to ensure the proper setup and operation of the ISD monitor. Among other items, the self-test algorithms will check for a properly configured TLS Console monitor as well as the correct number of functioning sensors and interface modules.
System Self-Test Monitoring algorithms are broken down into two types of tests:
1. ISD sensor out self-test, designed to monitor for properly functioning sensors.
2. Setup self-test, designed to verify the monitor configuration.
DIAGNOSTIC PROCEDURE
Verify that the ISD setup is complete.
COMMON CAUSES
The following conditions will generate this warning:
• PROBE OUT (Tank Inventory device)
• Mnn: DISABLED DIM ALARM (MDIM Module device)
• Enn: DISABLED DIM ALARM (EDIM Module device)
• Enn: COMMUNICATION ALARM (EDIM Module device)
• snn: COMMUNICATION ALARM (Smart Sensor device)
• mnn: COMMUNICATION ALARM (Mod bus device)
• CHK VAPOR FLOW MTR
Refer to the Clear Test Repair Menu (on page 37) for clearing this alarm on the TLS after repairs are complete.
ISD Alarm Message Diagnostics ISD SETUP
17
ISD SETUP
One or more of the ISD setup requirements are incomplete.
ISD MONITORING CATEGORY
ISD Setup Diagnostic self-tests are designed to monitor and ensure proper monitor configuration. Setup self-test will verify:
1. That the ISD system is properly setup to shutdown affected fueling point(s) as required by CP-201 regulations.
2. At least one tank contains gasoline.
3. At least one fuel position and gas hose is setup.
4. At least one Vapor Flow Meter is setup.
5. At least one Vapor Pressure Sensor is setup.
6. An external input is setup if a non-TLS Console Controlled Processor is installed.
7. A control relay is setup if a TLS Console Controlled Processor is installed.
Setup self-testing occurs following power-up as well as at daily intervals at the Daily Test Time. A failure will result in a setup self-test warning and warning event recording. Consecutive failures over a 7-day period will result in a setup self-test alarm, failure event recording, and attempted shutdown of the affected fueling point(s).
DIAGNOSTIC PROCEDURE
1. Look for one of the Common Cause alarm conditions
2. Remove the alarm conditions and the ISD SETUP condition can be removed.
3. Once all the conditions are removed, enter and exit the Setup Menu using the MODE key, then press the red ALARM button on the TLS and the condition should clear.
There are two times when the ISD system setup is checked:
1. Just prior to the programed ISD Analysis time (See Setup Manual)
2. When exiting the SETUP MENU at the TLS front panel
Refer to the Clear Test Repair Menu (on page 37) for clearing this alarm on the TLS after repairs are complete.
COMMON CAUSES
The following conditions will generate this warning:
• MISSING RELAY SETUP
• MISSING HOSE SETUP
• MISSING TANK SETUP
• MISSING VAPOR FLOW MTR
• MISSING VAPOR PRESS SEN
• MISSING VP INPUT
• AIRFLOW MTR SETUP
• PMC SETUP FAIL
ISD Alarm Message Diagnostics MISSING RELAY SETUP
18
MISSING RELAY SETUP
One or more required shutdown alarms have not been assigned to a control device.
ISD MONITORING CATEGORY
ISD Setup Diagnostic self-tests (ref. page 17) are designed to monitor and ensure proper monitor configuration. Setup self-test will verify that the ISD system is properly setup to shutdown affected fueling point(s) as required by CP-201 regulations.
DIAGNOSTIC PROCEDURE
Once all the Common Causes are removed, enter and exit the Setup Menu using the MODE key, then press the red ALARM button on the TLS and the condition should clear.
COMMON CAUSES
• Tanks are configured with a control device (Relay, PLLD, WPLLD, or VLLD). The control device does not have all the correct alarms assigned.The following ISD alarms must be assigned to the relay:
- ISD GROSS PRESSURE FAIL
- ISD DEGRD PRESSURE FAIL
- ISD VAPOR LEAKAGE FAIL
• When there is a Vapor Processor installed the following ISD alarms must be assigned:
- ISD VP PRESSURE FAIL
- ISD VP STATUS FAIL
• When ISD system is configured as an EVR Balance type the following HOSE alarms must be assigned:
- FLOW COLLECT FAIL
• When the ISD system is configured as an EVR Vacuum Assist type the following HOSE alarms must be assigned:
- GROSS COLLECT FAIL
- DEGRD COLLECT FAIL
FIELD NOTES
• When configuring Relays, do not fill in the Tank ID field for relays that are not used by ISD to shut down tank STP. For example, when PLLD is used as the primary control for shutting down the STP and a relay is used to signal a tank overfill alarm, do not fill in the Tank ID field for the relay controlling the overfill alarm signal. This will cause a MISSING RELAY SETUP warning.
ISD Alarm Message Diagnostics MISSING TANK SETUP
19
MISSING TANK SETUP
There are no vapor recovery (gasoline) tanks defined, or a gasoline pump has not been assigned to a control (shut down) device in at least one tank.
ISD MONITORING CATEGORY
ISD Setup Diagnostic self-tests (ref. page 17) are designed to monitor and ensure proper monitor configuration. Setup self-test will verify at least one tank contains gasoline.
DIAGNOSTIC PROCEDURE
1. Configure the tanks and assign one of the following control devices:
- RELAY
- PLLD
- WPLLD
- VLLD
2. Then assign all the required ISD alarms to the controlling device.
3. Once all the causes are removed, enter and exit the Setup Menu using the MODE key, then press the red ALARM button on the TLS and the condition should clear.
ISD Alarm Message Diagnostics MISSING HOSE SETUP
20
MISSING HOSE SETUP
The Fuel Grade Table does not have any hoses assigned to it.
ISD MONITORING CATEGORY
ISD Setup Diagnostic self-tests (ref. page 17) are designed to monitor and ensure proper monitor configuration. Setup self-test will verify at least one fuel position and gas hose is setup.
DIAGNOSTIC PROCEDURE
• This warning indicates the mapping of the hose devices to the dispensed products has not been started. Refer to the ISD Setup manual (P/N 577013-800) to map hoses to the fuel products.
• The absence of this alarm does not mean the map is correct. The alarm indicates that the Fuel Grade Table has no hoses assigned.
• Once all the hoses are mapped, enter and exit the Setup Menu using the MODE key, then press the red ALARM button on the TLS and the condition should clear.
COMMON CAUSES
Hose mapping has not been started.
ISD Alarm Message Diagnostics MISSING VAPOR FLOW MTR
21
MISSING VAPOR FLOW MTR
There is no Vapor Flow Meter setup or detected.
ISD MONITORING CATEGORY
ISD Setup Diagnostic self-tests (ref. page 17) are designed to monitor and ensure proper monitor configuration. Setup self-test will verify at least one Vapor Flow Meter is setup.
DIAGNOSTIC PROCEDURE
Enable the air flow meters used by ISD
1. Correct Condition2. MODE out of the Setup Menu
3. Press ALARM key4. Check if alarm cleared
Install and configure the air flow meters
according to the ISD Setup and
Installation manuals
Is there an air flow meter
installed and configured in the
TLS?
Is the airflow meter serial
number available?
Is the airflow meter
ENABLED?
Contact Veeder-
Root Service
It does take a few minutes for the TLS to collect the serial number information from the sensors. If there are no airflow meter serial numbers in the ISD setup menu to enable, this is an indication that the airflow
meters may not be communicating with the
TLS.
Go to the ISD air flow meter Select Menu
No
Yes
Yes
No
No
ISD-EVR/fig7.eps
Yes
ISD Alarm Message Diagnostics MISSING VAPOR PRES SEN
22
MISSING VAPOR PRES SEN
There is no Vapor Pressure Sensor setup or detected.
ISD MONITORING CATEGORY
ISD Setup Diagnostic self-tests (ref. page 17) are designed to monitor and ensure proper monitor configuration. Setup self-test will verify at least one Vapor Pressure Sensor is setup.
DIAGNOSTIC PROCEDURE
Enable the pressure
sensor used by ISD
1. Correct Condition
2. MODE out of the Setup Menu
3. Press ALARM key
4. Check if alarm cleared
Install and configure
the pressure sensor
according to the ISD
Setup and
Installation manuals
Is there a pressure sensor
installed and configured in the
TLS?
Is the pressure
sensor serial
number
available?
Is the pressure
sensor
ENABLED?
Contact Veeder-
Root Service
It does take a few minutes
for the TLS to collect the
serial number information
from the sensors. If there is
no pressure sensor serial
number in the ISD setup
menu to enable, this is an
indication that the pressure
sensor may not be
communicating with the
TLS.
Go to the ISD Pressure Select Menu
No
Yes
Yes
No
No
ISD-EVR/fig8.eps
Yes
ISD Alarm Message Diagnostics fnn: CHK VAPOR FLOW MTR
23
fnn: CHK VAPOR FLOW MTR
Failure of locked rotor test - possible locked vapor flow meter
ISD MONITORING CATEGORY
System Diagnostic
DIAGNOSTIC PROCEDURE
FIELD NOTES
See “Removing & Replacing Vapor Flow Meters” on page 34 when replacing a vapor flow meter.
No
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Is back pressure still
failing?
Is back pressure still
failing?
Assist Site?
Is there at least one hose
with passingback pressure?
Run TP 201.5 Air to Liquid Volume Test
Replace Vapor Flow Meter
Retest equipment to ensure proper operation
Does the test air flow
meter register A>0 on one or more hoses?
Check all hoses using the vapor flow meter under test for liquid and drain if necessary.
Run a 60 DFH TP 201.4 Dynamic Backpressure Test on all hoses using the Vapor Flow Meter.
Swap hanging hardware with a known good set. (Example: whip, breakaway, hose and nozzle).
Remove Vapor Flow Meter and replace with temporary pipe.
See Repairing Collection Vapor Recovery Equipment in the Miscellaneous
Retest known good hose and backpressure again. Add 0.08 IWC (@60 CFH) to test value and compensate for the missing Vapor Flow Meter.
ISD-EVR/fig9.eps
ISD Alarm Message Diagnostics hnn: VPRFLOW MTR SETUP
24
hnn: VPRFLOW MTR SETUP
Incoming transaction from a hose with an unavailable Vapor Flow Meter.
ISD MONITORING CATEGORY
ISD Setup Diagnostic self-tests (ref. page 17) are designed to monitor and ensure proper monitor configuration. Setup self-test will verify at least one Vapor Flow Meter is setup.
DIAGNOSTIC PROCEDURE
1. Using the ISD Hose Setup Menus print out the ISD Hose Table (see example printout below).
2. Find the AFM assigned to Hose nn.
3. Verify that the AFM is still configured in the TLS.
hosetableprtout.eps
ID = Hose IDFP = Mapped fuel position as TLS Console recognizes it (-1 = unassigned)FL = Fuel position label as written on dispenserHL = Hose labelAA = Airflow meter ID assignedRR = Relay IDUU = unassigned
ID = Airflow meter ID assignedSerial Number = Airflow meter's serial number
ISD HOSE TABLE ID FP FL HL AA RR ------------------------01 01 01 02 01 UU 02 02 02 02 01 UU 03 03 03 02 02 UU 04 04 04 02 02 UU 05 05 05 02 03 UU 06 06 06 02 03 UU 07 07 07 02 04 UU 08 08 08 02 04 UU 09 09 09 02 05 UU 10 10 10 02 05 UU 11 11 11 02 06 UU 12 12 12 02 06 UU ISD AIRFLOW METER MAP ID SERIAL NUM LABEL ------------------------ 1 03001401 AFM1 FP1 - 2 03001402 AFM2 FP3 - 3 03001403 AFM3 FP5 - 4 03001404 AFM4 FP7 - 5 03001405 AFM5 FP9 - 6 03001406 AFM6 FP11
25
PMC Alarm Message Diagnostics
This manual provides instructions to install, setup, and operate the components of Veeder-Root Pressure Management Control (PMC) equipment. The PMC feature is an option for the TLS console platform, and as such, many of the installation/setup/operation instructions for non-PMC specific tasks are covered in TLS-3XX supplied literature. Do not use this manual when PMC is installed with ISD. Use the ISD Setup & Operation Manual, 577021-800.
Veeder- Root Vapor Polisher Site Requirements
Below are the requirements for all PMC installations:
• V-R TLS-350R/EMC w/BIR, TLS-350 Plus/EMC Enhanced, TLS-350/EMC and ProMax consoles with ECPU2 - install as per TLS-3XX Site Prep manual, setup following instructions in TLS-3XX System Setup Manual.
• A flash memory board (NVMEM203) for PMC software storage - installed on the ECPU2 board in place of the console’s 1/2 Meg RAM board - install as per TLS-350 Series Board and Software Replacement Manual, no setup required.
• Vapor Pressure Sensor and Smart Sensor Module- install and connect following instructions in the Vapor Pressure Sensor Installation Guide.
• Carbon Canister Vapor Polisher - install and connect following instructions in the Carbon Canister Vapor Polisher Installation Guide.
• An RS-232 Port will be available for use by contractor or government inspectors.
VST ECS Membrane Processor Site Requirements
Below are the requirements for all PMC installations:
• V-R TLS-350R/EMC w/BIR, TLS-350 Plus/EMC Enhanced, TLS-350/EMC and ProMax consoles with ECPU2 - install as per TLS-3XX Site Prep manual, setup following instructions in TLS-3XX System Setup Manual.
• A flash memory board (NVMEM203) for PMC software storage - installed on the ECPU2 board in place of the console’s 1/2 Meg RAM board - install as per TLS-350 Series Board and Software Replacement Manual, no setup required.
• Smart Sensor Module and Vapor Pressure Sensor. Install and connect following instructions in the Vapor Pressure Sensor Installation Guide.
• Mod UBus Module connected to a hydrocarbon sensor module installed according to processor manufacturers specifications.
• A 4-Relay or I/O Combination Module to control the vapor processor motor and setup as instructed in this manual.
• An RS-232 Port will be available for use by contractor or government inspectors.
PMC Alarm Message Diagnostics PMC SETUP
26
PMC SETUP
Incomplete PMC Setup requirements
ISD MONITORING CATEGORY
ISD Setup Diagnostic self-tests (ref. page 17) are designed to monitor and ensure proper monitor configuration. Setup self-test will verify that the Pressure Management Control is configured.
VST ECS MEMBRANE DIAGNOSTIC PROCEDURE
Gasoline tanks installed and configured?
Vapor pressure sensor installed and configured?
Does PMC Setup Alarm
Clear?
Is there a relay assigned to the
vapor processor?
Is a hydrocarbon sensor installed and configured?
Is there more than one relay assigned to the
vapor processor?
Contact Veeder- Root ServiceComplete
No
No
No
Yes
Yes
Yes
819-5.eps
Yes
Yes
Yes
1. Correct Condition2. MODE out of the Setup Menu
3. Press ALARM key
No
No
No
PMC Alarm Message Diagnostics PMC SETUP
27
V-R VAPOR POLISHER DIAGNOSTIC PROCEDURE
Gasolinetanks installed and
configured?
Complete
No
No
Yes
Yes
Yes
Yes
No
No
No
Call V-R Tech Support
Vaporpressure sensor
installed andenabled?
7 Sensorinput moduleinstalled andconfigured?
Vapor valveinstalled and
enabled?
Yes
No
Correct conditionMODE out of Setup MenuPress ALARM key
Doesalarmclear?
819-10.eps
ATM sensorconfigured and
enabled?
PMC Alarm Message Diagnostics PMC Sensor Fault
28
PMC Sensor Fault
Table 1 contains a listing of the Smart Sensor Device generated alarms including their cause and suggested troubleshooting. TLS Console PMC alarms may be interspersed amongst non-PMC alarms, please see TLS Series manuals for more information.
Table 1. Smart Sensor Device Alarm and Fault Summary
Alarm Type1 Smart Sensor Device Cause Suggested Troubleshooting
Communication Alarm
Vapor Valve, Pressure Sen-sor, Tank Probe
Device not communicating with Smart Sensor Module
Check wiring and connections from the Smart Sensor board to the device in alarm.
Smart Sensor Fault
Vapor Valve Vapor Valve component inop-erative, see examplet below:
• Valve command fault
• Cap not charging
• Cap not holding
• Temperature out of range
Check installation of all Vapor Valve components including Thermal Probe and Vapor Sensor Assembly. Refer to manual 577013-920.
1All Smart Sensor Device Alarms and Faults are indicated with a Yellow Light on the TLS Console.
IB6100FEB 4, 2008 1:09 PMs 2:Vapor valve
VAPOR VALVESERIAL NUMBER 123456VALVE POSITION: OPENOPEN CAP: CHARGEDCLOSE CAP: CHARGEDAMBNT TMP: 65.08 FOUTLET TMP: 75.05 FSENSOR FAULTS:CAP NOT HOLDING
IB6100FEB 4, 2008 1:09 PMs 2:Vapor valve
VAPOR VALVESERIAL NUMBER 123456VALVE POSITION: OPENOPEN CAP: CHARGEDCLOSE CAP: CHARGEDAMBNT TMP: 65.08 FOUTLET TMP: 75.05 FSENSOR FAULTS:VALVE COMMAND FAULT
IB6100FEB 4, 2008 1:09 PMs 2:Vapor valve
VAPOR VALVESERIAL NUMBER 123456VALVE POSITION: OPENOPEN CAP: CHARGEDCLOSE CAP: CHARGEDAMBNT TMP: 65.08 FOUTLET TMP: 75.05 FSENSOR FAULTS:CAP NOT CHARGING
IB6100FEB 4, 2008 1:09 PMs 2:Vapor valve
VAPOR VALVESERIAL NUMBER 123456VALVE POSITION: OPENOPEN CAP: CHARGEDCLOSE CAP: CHARGEDAMBNT TMP: 65.08 FOUTLET TMP: 75.05 FSENSOR FAULTS:TEMPERATURE RANGE
PMC Alarm Message Diagnostics MISSING VP INPUT
29
MISSING VP INPUT
An external input for the OPW and ARID vapor processor cannot be found.
ISD MONITORING CATEGORY
System Self-Test Monitoring algorithms are designed to ensure the proper setup and operation of the ISD monitor. Among other items, the self-test algorithms will check for a properly configured TLS Console monitor as well as the correct number of functioning sensors and interface modules.
System Self-Test Monitoring algorithms are broken down into two types of tests:
1. ISD sensor out self-test, designed to monitor for properly functioning sensors.
2. Setup self-test, designed to verify the monitor configuration.
DIAGNOSTIC PROCEDURE
• This warning only applies to EVR systems with an OPW or ARID vapor processor.
• Ensure the input line is properly configured and installed for the processor.
• Once all the causes are removed, enter and exit the Setup Menu using the MODE key, then press the red ALARM button on the TLS and the condition should clear.
FIELD NOTES
This warning will not appear on a Healy EVR system.
PMC Alarm Message Diagnostics VP EMISSIONS
30
VP EMISSIONS
The effluent emissions concentration test occurs at daily intervals at the daily assessment time after at least one day's HC%, processor run time and station throughput has been collected. A failure occurs when the mass emission exceeds the defined threshold for the system. A failure of the effluent emissions concentration test will result in a warning and warning event recording. Two consecutive 1-day periods of effluent emissions concentration test failures will result in a failure alarm, failure event recording, and shutdown of the site. The processor is not allowed to operate while the emission FAILURE is active
ISD MONITORING CATEGORY
Vapor Processor Monitoring is responsible for ensuring that the vapor processor is operating within normal parameters.
DIAGNOSTIC PROCEDURE
• Refer problem to the processor manufacturer certified technician.
• Refer to the Clear Test Repair Menu (on page 37) for clearing this alarm on the TLS after repairs are complete.
COMMON CAUSES
• The processor is hydrocarbon emissions are exceeding defined limits.
FIELD NOTES
• This warning will not appear on a Healy EVR system.
PMC Alarm Message Diagnostics VP PRESSURE
31
VP PRESSURE
The processor over-pressure test occurs at daily intervals at the daily assessment time after at least 1-day's UST ullage vapor pressure data has been collected. A VST ECS Membrane Processor failure occurs when the 90th percentile of 1-day's ullage pressure data (i.e. 10% of the pressure data) is equal to or exceeds 1" wc. A Veeder-Root Polisher failure occurs when the 90th percentile of 1-day's ullage pressure data (i.e. 10% of the pressure data) is equal to or exceeds 2.3" wc. A failure of the processor over-pressure test will result in a warning and warning event recording. Two consecutive 1-day periods of processor over-pressure test failures will result in a failure alarm, failure event recording, and shutdown of the site.
ISD MONITORING CATEGORY
Vapor Processor Monitoring is responsible for ensuring that the vapor processor is operating within normal parameters with no risk of releasing excessive HC effluent emissions.
DIAGNOSTIC PROCEDURE
• For a VST ECS Membrane Processor, refer problem to the processor manufacturer certified technician. For a Veeder-Root Polisher, perform the operability tests as outlined in Exhibit 11 of VR 203/ VR 204.
• Refer to the Clear Test Repair Menu (on page 37) for clearing this alarm on the TLS after repairs are complete.
FIELD NOTES
• This warning will not appear on a Healy EVR system.
PMC Alarm Message Diagnostics VP DUTY CYCLE
32
VP DUTY CYCLE
The processor duty cycle test occurs at daily intervals at the daily assessment time after at least 1-day's on-time data has been collected. A failure occurs when the duty cycle exceeds 18 hours (75%). A failure of the duty cycle test will result in a warning and warning event recording. Two consecutive 1-day periods of processor duty cycle test failures will result in a failure alarm, failure event recording, and shutdown of the site
ISD MONITORING CATEGORY
Vapor Processor Control
ISD MONITORING CATEGORY
Vapor Processor Monitoring is responsible for ensuring that the vapor processor is operating within normal parameters with no risk of releasing excessive HC effluent emissions.
DIAGNOSTIC PROCEDURE
• Refer problem to the processor manufacturer certified technician.
• Refer to the Clear Test Repair Menu (on page 37) for clearing this alarm on the TLS after repairs are complete.
FIELD NOTES
This warning will not appear on a Healy EVR system.
Exhibit 11 Failures
PRESSURE INTEGRITY TEST
• Check that lockable three way valve is in the test position. If not in test position switch valve to test position and rerun Pressure Integrity test.
• Verify the Polisher solenoid control valve is in the closed position by accessing the TLS PMC Diagnostics. If Control valve is in open position, close Valve by accessing PMC Diagnostics and rerun Pressure Integrity test.
• Check for leaks on your test equipment. Find leaks using leak detection solution (I.E. Soap and Water) and repair. Rerun Pressure Integrity test.
• Check all fittings and connections on the VR Vapor Polisher for leaks. Find leaks using leak detection solution (I.E. Soap and Water) and repair. Rerun Pressure Integrity test.
• Check the Filter basket for leaks Tighten four filter basket screws and check filter o-ring and replace if necessary. Rerun Pressure Integrity test
• Check for leaks around the manifold If leaks found call Veeder-Root Technical support at 800-323-1799.
FLOW TEST
• Verify all test equipment has current calibrations. Have equipment calibrated per NIST specifications.
• Check that lockable three way valve is in the test position If not in test position switch valve to test position and rerun Pressure Integrity test.
PMC Alarm Message Diagnostics Exhibit 12 Failures
33
• Verify the Polisher solenoid control valve is in the open position by accessing the TLS PMC Diagnostics. If Control valve is in closed position, put valve in manual mode, and then open the valve. Rerun Pressure Integrity test.
• Verify your nitrogen flow rate is at 18.0 SFCH Adjust flow rate per test procedure and rerun Flow test. If flow test continues to fail contact Veeder-Root Technical support at 800-323-1799.
THERMOMETER TEST
• Replace Tank probe if gas tank thermometer fails all tests.
• Replace Vapor Valve thermometer if Vapor Valve thermometer fails all tests.
• Replace ambient temperature sensor if ambient temperature sensor is out of range.
Exhibit 12 Failures
• Run Vapor Polisher operability test as outlined in Executive Order 203/204 Exhibit 11 to check for leaks. Run only the pressure integrity portion of Exhibit 11 at this time.
• If the Vapor Polisher fails any part of the test run the force purge procedure as outlined in Executive Order 203/204 Exhibit 12 Appendix A.
• Perform Vapor Pressure Sensor operability test as outlined in Executive Order 203/204 Exhibit 10.
34
Miscellaneous Maintenance
Repairing Collection Vapor Recovery Equipment
ALL SYSTEMS (HEALY)
Check common vapor recovery piping both before and after the ISD vapor flow meter.
ASSIST SYSTEMS (HEALY)
Check common vacuum assist electronics
BALANCE SYSTEMS
Check hose liquid extraction
Removing & Replacing Vapor Flow Meters
When removing and replacing a vapor flow meter follow these steps:
1. Deconfigure the air flow meter in the SmartSensor Setup Menu.
2. Replace the air flow meter according the instructions in the Related Manuals.
3. Reconfigure the air flow meter in the SmartSensor Setup Menu.
No changes are required to the ISD Setup.
35
ISD/PMC Diagnostic Menus
The diagnostic menus below are viewed from the TLS Console front panel. Consult manual 577013-800 for the procedure used to selecting/changing Diagnostic Mode setup parameters via the front panel keys.
SMARTSENSOR DIAGNOSTICPRESS <STEP> TO CONTINUE
SYSTEM DIAGNOSTICPRESS <STEP> TO CONTINUE
DIAG MODEPRESS <FUNCTION> TO CONT
s 1: VP: FP1 - 2COMM DATA PRESS <PRINT>
Prints out comm diag - see example below
Prints out constants diag - see example below
Prints out channel diag - see example below
To select another sensor
F
F
S
T
P
To select another sensor
S
T
P
To select another sensor
S
T
P
s 1: VP: FP1 - 2CONSTANTS PRESS <PRINT>
s 1: VP: FP1 - 2CHANNELS PRESS <PRINT>
SS COMM DIAG-----------------------s 1: AFM1 FP1-2SAMPLES READ 58SAMPLES USED 54PARITY ERR 0PARTIAL READ 0COMM ERR 0RESTARTS 0
SS CONSTANTS DIAG-----------------------s 1: AFM1 FP1-2
VAPOR PRESSURESERIAL NUMBER 1007PROTOCOL VERSION 0
SS CHANNEL DIAG-----------------------s 1: AFM1 FP1-2YY-MM-DD HH:MM:SSC00 B50B 3D68 00E0 0000C04 0000 03EF 0000 0004C08 0A3C 3D68 5693 0081C12 80C4 80A4 0104 2579C16 0000 0000 00A3 03D6C20 0709 0032 04C9 880F
ISD-EVR/fig2.eps
E
F
P
S
T
Enter
Function
Step
Tank/Sensor
Repeat keypresses
Key Legend
ISD/PMC Diagnostic Menus Removing & Replacing Vapor Flow Meters
36
P
ISD DIAGNOSTICPRESS <STEP> TO CONTINUE
SYSTEM DIAGNOSTICPRESS <STEP> TO CONTINUE
DIAG MODEPRESS <FUNCTION> TO CONT
CLEAR TEST AFTER REPAIRPRESS <ENTER>
CONTAINMENT OVER PRESSPRESS <ENTER>
F
F
S
E
VAPOR LEAKAGE TESTPRESS <ENTER>
S
VAPOR COLLECTION TESTPRESS <ENTER>
S
SENSOR OUT TESTPRESS <ENTER>
S
SETUP TESTPRESS <ENTER>
S
PROCESSOR STATUS TESTPRESS <ENTER>
S
CLEAR TEST AND LOGARE YOU SURE?: YES
E
E
E
FP: ff h:HH nnnnnnnnnnnPRESS <ENTER>
S
E
CLEAR TEST AND LOGARE YOU SURE?: YES
E
ARE YOU SURE?: YESPRESS <STEP> TO CONTINUE
E
2
2
2
2
2
1
2
ISD-EVR/fig4.eps
Notes:1. All repair dates are saved in the Miscellaneous Event Log.2. Reference the Clear Test Repair Menu table on the next page.
ff = fuel position labelHH = hose on ffnnnnnnnnnnn hose label
Selects all hoses on this fuel position
See example printout at right.
Press step to select an individual hose on this fuel position.
FP: ffPRESS <ENTER>
TEST FAIL CLEAR DATES
CONTAINMENT OVER PRESS02-06-11 12:43
VAPOR COLLECTION TESTFP: 1 h: 1 MIDGRADE02-06-10 11:13
E
F
P
S
T
Enter
Function
Step
Tank/Sensor
Repeat keypresses
Key Legend
ISD/PMC Diagnostic Menus Removing & Replacing Vapor Flow Meters
37
Table 2.- Clear Test Repair Menu
Menu Selection Clears Alarms Reset Dates
Containment Over Press ISD GROSS PRESSURE WARNISD GROSS PRESSURE FAILISD DEGRD PRESSURE WARNISD DEGRD PRESSURE FAILISD VP PRESSURE WARNISD VP PRESSURE FAIL
Containment Test Time
Vapor Leakage Test ISD VAPOR LEAKAGE WARNISD VAPOR LEAKAGE FAIL
Vapor Leak Test Time
Vapor Collection Test GROSS COLLECT WARN GROSS COLLECT FAIL DEGRD COLLECT WARN DEGRD COLLECT FAIL FLOW COLLECT WARN FLOW COLLECT FAIL AIRFLOW MTR SETUP
Hose Test Time
Sensor Out Test ISD SENSOR OUT WARNISD SENSOR OUT FAIL
Sensor Out Test Time
Setup Test ISD SETUP WARNISD SETUP FAIL
Setup Self Test Time
Processor Status Test ISD VP STATUS WARNISD VP STATUS FAILVP EMISSIONS WARNVP EMISSIONS FAILVP DUTY CYCLE WARNVP DUTY CYCLE FAIL
Valid Vapor Processor Test Time
PMC DIAGNOSTICPRESS <STEP> TO CONTINUE
SYSTEM DIAGNOSTICPRESS <STEP> TO CONTINUE
DIAG MODEPRESS <FUNCTION> TO CONT
F
F
ISD-EVR/fig6.eps
E
F
P
S
T
Enter
Function
C Change
Step
Tank/Sensor
Repeat keypresses
Key Legend
VAPOR PRESSUREINCHES H2O: -X.XXX
S
VAPOR PROCESSOR MODEMANUAL
S
VAPOR PROCESSOR STATEVP STATE: OFF
S
HYDROCARBON SENSORHC SENSOR XX.XXX%
S
P
Current vapor pressure reading
See exampleat right
Automatic is the default
Updates to reflect current HC reading
If VP mode = Manual and relay configured, then pressing Changeselects On/Off
PMC DIAGNOSTICS
VAPOR PROCESSOR MODEMANUAL
VAPOR PROCESSOR STATEVP STATE OFF
38
ISD Reports
Daily Detail Report
A Daily Detail example report is shown below:IV0700
MMMDDYYYY HH:MM XM
S--- STATION HEADER ---
ISD DAILY REPORT DETAILS
EVR TYPE: VACUUM ASSIST
ISD TYPE: 01.03
VAPOR PROCESSOR TYPE: NO VAPOR PROCESSOR
OVERALL STATUS :WARN EVR VAPOR COLLECTION :PASS
EVR VAPOR CONTAINMENT :WARN
ISD MONITOR UP-TIME : 99% STAGE I TRANSFERS: 9 of 20 PASS
EVR/ISD PASS TIME : 90%
Status Codes: (W)Warn (F)Fail (D)Degradation Fail (G)Gross Fail
(ISD-W)ISD Self-Test Warning (ISD-F)ISD Self-Test Fail (N)No Test
ISD ISD ---CONTAINMENT TESTS--- STAGE ---COLLECTION TESTS-DAILY AVERAGE HOSE A/L RATIO-----------------------
EVR %UP GROSS DGRD MAX MIN LEAK I VAPOR FP1 FP2 FP3 FP4
DATE STATUS TIME 95% 75% "WC "WC CFH XFR PRCSR BLEND BLEND BLEND BLEND
12/16 PASS 100% 3.2N 2.8N 3.0 -5.0 0 PASS 0.88 1.04 0.98 0.91
12/17 PASS 100% 3.5N 3.2N 3.6 0.4 0 0.93 0.96 0.95 1.01
12/18 PASS 100% 3.5N 3.2N 3.8 2.4 0 0.88 0.90 0.90 1.01
12/19 PASS 100% 3.5N 3.3N 3.7 2.2 0 0.92 0.95 0.96 0.97
12/20 PASS 100% 3.5N 3.3N 5.0 -5.0 0 FAIL 0.95 0.91 0.97 1.01
12/21 W 99% 3.5W 3.2N 0.3 -2.9 0 FAIL 0.83 1.02 0.93 1.02
12/22 PASS 100% 0.3N 0.2N 0.4 -0.2 0 0.97 0.99 0.93 0.95
12/23 PASS 100% 3.4N 3.2N 3.6 0.0 0 0.93 0.93 0.89 1.05
12/24 PASS 100% 3.4N 3.1N 3.7 1.8 1 0.94 1.00 0.92 0.84
12/25 PASS 100% 3.4N 3.1N 3.4 -5.0 0 PASS 0.94 0.82 0.94 0.98
12/26 PASS 100% 3.4N 3.2N 3.9 1.9 0 0.96 0.95 0.92 1.01
12/27 PASS 100% 3.5N 3.2N 3.7 2.4 0 0.92 0.97 0.94 0.99
12/28 W 100% 3.5W 3.2N 5.0 -1.8 0 PASS 0.95 1.01 0.95 0.98
12/29 PASS 100% 0.4N 0.2N 0.5 -0.2 17N 0.95 0.98N 0.95 0.99
12/30 PASS 100% 0.4N 0.2N 0.7 -5.0 3N FAIL 0.96 1.00 0.93 0.90
12/31 PASS 100% 0.3N 0.1N 0.4 -0.7 3N 0.94 0.99 0.95 1.01
01/01 PASS 100% 0.3N 0.1N 0.3 -0.4 2N 0.96 0.98 0.93 1.00
01/02 PASS 100% 0.3N 0.1N 0.3 -0.3 2N 0.98 1.04 1.00 1.03
01/03 PASS 100% 0.3N 0.1N 0.5 -0.3 3N 0.95 1.01 0.93 1.03
01/04 PASS 100% -0.0N -0.3N 0.0 -0.8 3 0.92 0.89 0.93 0.98
01/05 PASS 100% -0.1N -0.3N 0.5 -5.0 2 PASS 0.95 0.99 0.92 1.03
01/06 PASS 100% -0.2N -0.5N -2.1 -5.0 2 0.96 0.98 0.88 1.04
01/07 PASS 100% -0.2N -0.7N -0.4 -2.1 1 0.98 1.00 0.94 1.01
01/08 PASS 100% 0.0N -0.4N 0.3 -1.2 2 0.96 0.97 0.93 1.01
01/09 PASS 100% 0.1N -0.1N 0.3 -0.3 2 0.98 1.00 0.87 1.02
01/10 PASS 100% 0.2 0.0N 0.3 -0.4 2 0.97 0.98 0.90 0.98
01/11 PASS 100% 0.2 0.0N 0.8 -4.3 2 PASS 0.98 1.00 0.93 1.02
01/12 PASS 100% 0.2 0.0N 0.3 -3.2 2 0.99 1.02 0.92 1.00
01/13 PASS 100% 0.2 0.0N 0.4 -0.4 3 0.96 1.00 0.92 0.97
01/14 PASS 100% 0.6 0.0N 3.7 -0.9 3 PASS 0.99 0.99 0.93 1.01
01/15 W 100% 3.5W 0.1N 4.1 2.0 1 0.95N 1.01 0.95 0.97N
ISD Reports Daily Detail Report
39
DAILY DETAIL STATUS CODES
Question
When and why do the status codes appear? (see example below.)
Explanation
Assist Systems - Hose W,F,G,D Results
Next to each hose collection daily average A/L result there is an assessment. The absence of any of the results listed above indicates a pass. If any of the listed conditions W,F,G or D exist, they will post before a pass. The assessment can indicate the result for one (or combination) of three tests called Statistical Test, Gross Test and Degradation Test.
• F - Statistical Failure - A BLKD is posted instead of an A/L value.
- Automobiles equipped with ORVR equipment appear as a blocked dispensing event (A/L very low). ISD separates out this type of activity from blockages that are caused by something other than ORVR vehicles. It does this statistically using multiple dispensing events. When there are more blockages than would be expected from normal ORVR traffic, a BLKD warning is issued.
• G - Failure for Gross Test and will appear with A/L value.
• D - Failure for Degradation Test and will appear with A/L value.
• W - Warning for Statistical (BLKD), Gross (A/L) and/or Degradation (A/L).
- Indicates a statistical warning if BLKD appears instead of the A/L value.
- Indicates either a Gross or Degradation warning if the A/L value appears next to it. You can distinguish which of the two warning conditions it is by comparing the A/L value with the gross and degradation ranges defined in the IV00 or IV02 reports. It is important to make this distinction as a gross test will convert to a failure and shut down on the 2nd day and the degradation test will convert to a failure and shut down on the 8th day.
ISD Reports Daily Detail Report
40
Assist Systems - Hose N No Test Result
• If the Degradation test did not run and the Gross test did not run, then an N is posted.
• If there were no events received to make any conclusion about the A/L value for the hose, the A/L value is replaced with N.
• If there are some events received, enough to calculate an average, but not enough to complete a test, the No Test result will appear with an A/L average.
Assist Systems - Hose Pass Result
• When none of the above results W,F,G,D,N appear then it is a pass.
• If the Degradation test did not run (no test), but the Gross test passed, the result is a PASS.
• If the Gross test did not run (no test), but the Degradation test passed, the result is a PASS.
Assist Systems - Code Precedent (listed high to low)
1. G, F - Gross Failure (A/L) and Gross Stat Failure (BLKD)
2. D - Degradation Failure (A/L)
3. W - Gross Warning (A/L) and Gross Stat Warning (BLKD)
4. W - Degradation Warning (A/L)
5. Pass - Pass Gross and/or Degradation test.
6. N - No Test - Neither Gross nor Degradation test completed.
Assist Systems - Self Test Codes
• ISD-W - A self test warning condition exists. See ISD SETUP
• ISD-F - A self test warning condition persisted for 7 days and resulted in a Failure.
ISD Reports Daily Detail Report
41
COLLECTION REPORT FP ORDERING
Question
Should I be concerned if the fueling position numbers are out of order in the header?
Explanation
When using a Gilbarco current loop DIM the fueling position numbers may not be in order in the header. The fueling position numbers come from the current loop and may not match the numbers on the side of the dispenser or on the POS.
See following example:
IV0700
MMMDDYYYY HH:MM XM
---STATION 1 HEADER---
ISD DAILY REPORT DETAILS
EVR TYPE: VACUUM ASSIST
ISD TYPE: 01.03
VAPOR PROCESSOR TYPE: NO VAPOR PROCESSOR
OVERALL STATUS :WARN EVR VAPOR COLLECTION :PASS
EVR VAPOR CONTAINMENT :WARN
ISD MONITOR UP-TIME : 99% STAGE I TRANSFERS: 23 of 23 PASS
EVR/ISD PASS TIME : 86%
Status Codes: (W)Warn (F)Fail (D)Degradation Fail (G)Gross Fail
(ISD-W)ISD Self-Test Warning (ISD-F)ISD Self-Test Fail (N)No Test
ISD ISD ---CONTAINMENT TESTS--- STAGE --COLLECTION TESTS-DAILY AVG.HOSE A/L RATIO---
EVR %UP GROSS DGRD MAX MIN LEAK I VAPOR FP4 FP1 FP2 FP3
DATE STATUS TIME 95% 75% "WC "WC CFH XFR PRCSR BLEND BLEND BLEND BLEND
02/01 PASS 100% 0.2N 0.0N 0.4 -0.0 0N PASS N N N N N N N N
02/02 PASS 100% 0.2N 0.1N 0.2 -0.0 10N N N N N N N N N
02/03 PASS 100% 0.2N 0.0N 0.3 -0.8 33N 0.89 0.96 0.95 0.98
ISD Reports Daily Detail Report
42
CONTAINMENT DEGRADATION RESULTS
Question
Why is the Degradation Test Result a No Test?
Explanation
Containment Degradation Results will not post until after the system has been running 7 Days.
When the containment pressure remains at or below -5"wc for most of the day (Look at MAX "WC and MIN "WC) there may not be enough samples available over the past 7 days to perform an assessment, so a N test is reported.
ISD STATUS MISSING TEXT
Question
Why are the ISD EVR status and % Up time sometimes missing for the last day in the report?
Explanation
The daily detail report sometimes does not have the ISD Average and ISD EVR STATUS and ISD % UP TIME for the last day, but it does have all the other results for the day (containment and collection). If there is a post delay configured in ISD, the daily detail report will not show ISD EVR STATUS or % UP TIME if it is pulled during the time period between the assessment and the post. The ISD EVR STATUS and % UP TIME are part of the daily post and have not occurred yet. After the post time the ISD EVR STATUS or % UP TIME will show up in the daily detail report.
The post delay is not to be used in ISD. Instead of using the post delay the assessment time must be scheduled for the time they want the alarms to post.
MISSING ALL REPORTS FOR ONE DAY
Question
Why am I missing all the results for one day?
Explanation
In the example report below there is one day where all the reports are missing.
1. One possible cause is the assessment time was programmed for a new hour. The assessment time may not be reached until the next day, so it looks like a day is missing. However, the test has been done, but more than 24 hours were included in the test following the missing one. You can tell this has happened by looking at the time warnings were posting previously and the time they are posting presently. The time will be different.
2. The TLS clock time was changed to a time beyond the assessment time, skipping the test for that day. The Miscellaneous Event log will show the time change.
3. The TLS was not powered during the programmed assessment time and the assessment was skipped for that day. The Miscellaneous Event log will show an ISD SHUTDOWN and ISD STARTUP event that will show it was down during the assessment time.
ISD Reports Daily Detail Report
43
IV0700
MMMDDYYYY HH:MM XM
--STATION HEADER--
ISD DAILY REPORT DETAILS
EVR TYPE: VACUUM ASSIST
ISD TYPE: 01.03
VAPOR PROCESSOR TYPE: NO VAPOR PROCESSOR
OVERALL STATUS :WARN EVR VAPOR COLLECTION :PASS
EVR VAPOR CONTAINMENT :WARN
ISD MONITOR UP-TIME :100% STAGE I TRANSFERS: 26 of 26 PASS
EVR/ISD PASS TIME : 96%
Status Codes: (W)Warn (F)Fail (D)Degradation Fail (G)Gross Fail
(ISD-W)ISD Self-Test Warning (ISD-F)ISD Self-Test Fail (N)No Test
ISD ISD ---CONTAINMENT TESTS--- STAGE ---COLLECTION TESTS-DAILY AVERAGE HOSE A/L RATIO-----------------------
EVR %UP GROSS DGRD MAX MIN LEAK I VAPOR FP7 FP8 FP4 FP5 FP6 FP1 FP2 FP3
DATE STATUS TIME 95% 75% "WC "WC CFH XFR PRCSR BLEND BLEND BLEND BLEND BLEND BLEND BLEND BLEND
02/17 PASS 100% -5.0 -5.0N -4.1 -5.0 1 0.99 N N 0.93 0.85 0.91 0.92 1.06 0.91
02/18 PASS 100% -5.0 -5.0N -5.0 -5.0 1N PASS 1.00 N N 1.00 0.86 0.88 0.96 1.08 0.89
02/19 PASS 100% -4.6 -5.0N 0.7 -5.0 1 PASS 0.98 N N 0.96 0.84 0.87 0.90 1.04 0.88
02/20 PASS 100% -4.7 -5.0N -5.0 -5.0 1N 0.99 0.93 0.96 0.85 0.90 0.90 1.04 0.90
02/21 PASS 100% -2.6 -5.0N -0.5 -5.0 1N PASS 0.98 0.98 0.92 0.86 0.92 0.92 1.02 0.88
02/22 PASS 100% -2.4 -5.0N 2.6 -5.0 1 PASS 0.96 0.94 0.92 0.86 0.85 0.91 1.02 0.92
02/23 PASS 100% -2.2 -5.0N 0.8 -5.0 1 PASS 0.96 0.90 0.92 0.87 0.88 0.94 1.07 0.88
02/24 PASS 100% -1.6 -5.0N -0.0 -5.0 1 PASS 0.95 0.89 0.91 0.85 0.95 0.90 0.98 0.86
02/25 PASS 100% -1.4 -5.0N 1.0 -5.0 1 PASS 0.95 0.83 0.95 0.91 0.92 0.93 1.07 0.85
02/26 PASS 100% -2.4 -5.0N -2.7 -5.0 2 PASS 0.95 0.89 0.93 0.93 0.99 0.96 1.07 0.91
02/27 PASS 100% -2.4 -5.0N -3.9 -5.0 2 0.99 0.90 0.97 0.88 0.89 0.91 1.04 0.91
02/28 PASS 100% 0.0 -5.0N 5.0 -5.0 3 PASS 0.98 0.91 0.90 0.91 0.97 0.96 1.05 0.95
03/01 PASS 100% 0.1 -5.0N 0.2 -5.0 4 0.95 0.96 0.88 0.87 0.90 0.89 1.05 0.97
03/02 PASS 100% 0.1 -5.0N 0.8 -5.0 5 PASS 0.96 0.95 0.95 0.84 0.92 0.94 1.01 1.04
03/03 W 100% 0.1 -5.0N -2.9 -5.0 13W 1.00 0.91 0.94 0.82 0.94 0.93 1.08 1.03
03/04 PASS 100% 0.1 -5.0N -4.3 -5.0 4 PASS 0.95 0.90 0.94 0.88 0.89 0.94 1.08 1.03
03/05 PASS 100% 0.1 -4.9N -3.1 -5.0 4 0.99 0.92 0.98 0.88 0.90 0.94 1.03 1.02
03/06 PASS 100% 0.1 -4.9 -5.0 -5.0 4 PASS 0.95 0.83 0.93 0.85 0.91 0.91 1.08 1.08
03/07 PASS 100% -0.0 -4.9 -4.9 -5.0 0 0.99 0.86 0.95 0.91 0.91 0.94 1.09 1.02
03/08 PASS 100% -3.5 -4.9 0.0 -5.0 2 PASS 0.99 0.94 0.97 0.89 0.93 0.93 1.13 1.07
03/09 PASS 100% -3.5 -4.8 -1.7 -5.0 2 0.99 0.92 0.96 0.89 0.94 0.92 1.05 1.06
03/10 PASS 100% -2.9 -4.6 -2.1 -5.0 2 PASS 1.01 0.93 0.96 0.89 0.91 0.89 1.08 1.02
03/11 PASS 100% -2.9 -4.8 -5.0 -5.0 2 0.99 0.88 0.98 0.91 0.89 1.02 1.04 1.05
03/12 PASS 100% -2.9 -4.9 -5.0 -5.0 2 0.99 0.91 1.00 0.90 0.91 1.00 1.07 1.07
03/13 PASS 100% -2.9 -4.9 -0.0 -5.0 2 PASS 0.99 0.90 0.97 0.93 0.90 1.00 1.01 1.06
03/14 PASS 100% -2.9 -4.9 -2.5 -5.0 2 PASS 0.91 0.86 1.00 0.93 0.89 0.95 1.14 1.03
03/15 PASS 100% -3.0 -4.9 -5.0 -5.0 2 PASS 0.97 0.90 0.96 0.94 0.84 0.98 1.12 1.02
03/16 PASS 100% -4.3 -4.9 -3.8 -5.0 2 0.91 0.91 0.99 0.91 0.94 1.00 1.10 1.02
03/17
03/18 PASS 100% -5.0N -4.9 -4.9 -5.0 0N 0.97 0.91 1.00 0.89 0.91 1.00 1.09 1.09
03/19 PASS 100% -5.0N -4.9 -4.7 -5.0 0N 0.98 0.94 0.97 0.90 0.90 1.01 1.04 1.05
ISD Reports Daily Detail Report
44
HIGH A/L ON ALL NOZZLES
Question
Why is my report is showing large A/L for all Dispensers?
Explanation
• The most common cause of High A/L on all Nozzles at the same time is the dispenser events are artificially low. This can be caused by the DIM not having a 'G' in the DIM setup string. The only DIM that requires this is the BIR Protocol Dispenser Interface Module P/N 330280-001.
• Gross Collect warnings and alarms can result from this condition. See example reports from two sites below.
-----------------------------------------------------Example Site Report with High A/L-------------------------------------------------------------IV0500
MMMDDYYYY HH:MM XM
--STATION HEADER--
ISD DAILY REPORT DETAILS
EVR TYPE: VACUUM ASSIST
ISD TYPE: 01.03
VAPOR PROCESSOR TYPE: NO VAPOR PROCESSOR
OVERALL STATUS :FAIL EVR VAPOR COLLECTION :FAIL
EVR VAPOR CONTAINMENT :WARN
ISD MONITOR UP-TIME : 99% STAGE I TRANSFERS: 30 of 30 PASS
EVR/ISD PASS TIME : 50%
Status Codes: (W)Warn (F)Fail (D)Degradation Fail (G)Gross Fail
(ISD-W)ISD Self-Test Warning (ISD-F)ISD Self-Test Fail (N)No Test
ISD ISD ---CONTAINMENT TESTS--- STAGE
EVR %UP GROSS DGRD MAX MIN LEAK I VAPOR
DATE STATUS TIME 95% 75% "WC "WC CFH XFR PRCSR
05/24 ISD-W 100% 0.0N 0.0N 0.0 0.0 0N PASS
05/25 ISD-W 100% 0.1N 0.0N 5.0 -5.0 N PASS N N N N N N
05/26 PASS 99% 0.0N 0.0N 0.4 -0.2 0N PASS N N N N N N
05/27 PASS 100% 0.1N 0.0N 0.0 -0.0 N N N N N N N
05/28 PASS 100% 0.1N 0.0N 0.0 -0.0 N N N N N N N
05/29 PASS 100% 0.0N 0.0N 0.0 -0.0 N N N N N N N
05/30 PASS 100% 0.1N 0.0N 0.7 -0.3 N 7.93N 8.00N N N
05/31 PASS 99% 0.1N 0.0N 0.4 -0.4 N 5.54N 6.39N 1.94N
06/01 W 99% 0.1 0.0N 0.3 -0.3 W 4.91N 6.22N 3.12N
06/02 W 100% 0.1 0.0N 2.7 -1.1 14N PASS 4.41W 6.35N 3.25N
06/03 F 100% 0.2 0.0N 0.9 -0.8 8N 4.40G 6.23W 2.90N
06/04 W 100% 0.1N -0.4N 0.2 -2.9 1N 4.14N 5.00N 2.71W
ISD Reports Diagnosing CVLD Reports
45
---COLLECTION TESTS-DAILY AVERAGE HOSE A/L RATIO------------------------------
05/24
05/25
05/26 N N N N N N N N N N
05/27 N N N N N N N N N N N N N N N N N N
05/28 N N N N N N N N N N N N N N N N N N
05/29 N N N N N N N N N N N N N N N N N N
05/30 3.74N 4.17N 4.02N N N 4.26N N N 4.23N 1.08N N N
05/31 3.62N 3.91N 3.90N 2.82N 4.26N 3.79N 3.28N 1.65N 4.06N
06/01 3.70N 2.92N 3.90N 1.92N 2.86N 3.02N 3.54N 2.18N 4.06N
06/02 3.12N 2.56W 3.80N 2.28N 2.25N 2.81W 3.29N 2.05W 2.50N
06/03 2.51W 2.35G 3.06N 2.17W 2.21N 2.92G 3.22W 2.32G 2.57N
06/04 3.09N 1.90N 2.94N 2.99N 2.20N 2.20N 3.81N 2.86N 2.70N
Diagnosing CVLD Reports
Question
NO TEST is reported every day. What can cause this?
Explanation
• The first 7 days after startup CVLD will always report NO TEST.
• If the pressure is below -5 " wc every day (look at max min value in Daily Detail Report under Containment) a test is not performed.
• If there is pressure above -5 " wc check and make sure the thermal coefficients on the gasoline tanks are in the correct range greater than or equal to 0.00060 and less than or equal to 0.00079. (Typically gasoline coefficients are programmed to 0.000700).
For technical support, sales orother assistance, please visit:
www.veeder.com