Transcutaneous Monitoring and Pulse Oximetry
RC 290
![Page 2: Transcutaneous Monitoring and Pulse Oximetry RC 290.](https://reader035.fdocuments.net/reader035/viewer/2022062221/56649cc15503460f949889c5/html5/thumbnails/2.jpg)
TCM: PO2 and PCO2 via Skin Electrodes
• Continuous and non-invasive• Digital readout plus hard copy• TCM values show values of O2 and CO2
at tissue level– These values are a affected both by lung
function and perfusion
• TCM used more in infants than adults because infants have thinner skin and less sub-Q fat
![Page 3: Transcutaneous Monitoring and Pulse Oximetry RC 290.](https://reader035.fdocuments.net/reader035/viewer/2022062221/56649cc15503460f949889c5/html5/thumbnails/3.jpg)
TCM System
![Page 4: Transcutaneous Monitoring and Pulse Oximetry RC 290.](https://reader035.fdocuments.net/reader035/viewer/2022062221/56649cc15503460f949889c5/html5/thumbnails/4.jpg)
TCM Electrodes• PO2 is a Clark
electrode• PCO2 is a
Severinghaus electrode
• Electrodes should not be placed over bone
• Electrodes may be heated to “arterialize” the site– Heated electrodes need
to be moved every 4-8 hours
• Water or a sealant solution should be placed on the skin to prevent an air bubble from forming between the skin and electrode
• Electrodes usually “sloped”, ie 2 point calibration
![Page 5: Transcutaneous Monitoring and Pulse Oximetry RC 290.](https://reader035.fdocuments.net/reader035/viewer/2022062221/56649cc15503460f949889c5/html5/thumbnails/5.jpg)
TCM Monitor Unit• Controls for electrode
temperature and high and low limits for alarms
• Digital display of TCM values
• Power or energy indicator (shows how much is being used to heat electrode– This is the
perfusion monitor!
![Page 6: Transcutaneous Monitoring and Pulse Oximetry RC 290.](https://reader035.fdocuments.net/reader035/viewer/2022062221/56649cc15503460f949889c5/html5/thumbnails/6.jpg)
Power/Energy Level Changes
• After the unit is started and stabilizes, the relative heating power indicator is zeroed
• If it starts to have a positive deflection, this means it is taking more power to keep the electrode heated to the set temperature
• This means perfusion has increased!
![Page 7: Transcutaneous Monitoring and Pulse Oximetry RC 290.](https://reader035.fdocuments.net/reader035/viewer/2022062221/56649cc15503460f949889c5/html5/thumbnails/7.jpg)
Power/Energy Level Changes (cont.)
• If relative power starts to have a negative deflection, this means it is taking less energy to keep the electrode at the set temperature
• This indicates decreased perfusion!
• TCM values and arterial values have poor correlation with decreased perfusion!
![Page 8: Transcutaneous Monitoring and Pulse Oximetry RC 290.](https://reader035.fdocuments.net/reader035/viewer/2022062221/56649cc15503460f949889c5/html5/thumbnails/8.jpg)
If relative heat energy does not have a negative deflection, correlation between TCM and arterial values is good
• Decreased TCM PO2 without negative deflection of relative heat energy usually indicates a pulmonary problem
![Page 9: Transcutaneous Monitoring and Pulse Oximetry RC 290.](https://reader035.fdocuments.net/reader035/viewer/2022062221/56649cc15503460f949889c5/html5/thumbnails/9.jpg)
A negative deflection in relative heating power with a
drop in TCM PO2 usually indicates a perfusion problem
Not a lung problem!
![Page 10: Transcutaneous Monitoring and Pulse Oximetry RC 290.](https://reader035.fdocuments.net/reader035/viewer/2022062221/56649cc15503460f949889c5/html5/thumbnails/10.jpg)
TCM Uses
• Making ventilator changes• Effects of activities on infant
– Feeding, handling, etc
• Effects of respiratory therapies– O2 changes, PD & P, suctioning , etc
• Identifying the direction of shunt in PDA
![Page 11: Transcutaneous Monitoring and Pulse Oximetry RC 290.](https://reader035.fdocuments.net/reader035/viewer/2022062221/56649cc15503460f949889c5/html5/thumbnails/11.jpg)
Shunt Direction in PDA Using TCM
• Two TCM PO2 electrodes– One is placed pre-
ductal and the other is post-ductal
• If difference is greater than 15 mmhg, PDA is R-L
![Page 12: Transcutaneous Monitoring and Pulse Oximetry RC 290.](https://reader035.fdocuments.net/reader035/viewer/2022062221/56649cc15503460f949889c5/html5/thumbnails/12.jpg)
TCM Does Not Replace ABGs
It does cut down on the number of ABGs and is a good indicator of
when they need to be done
![Page 13: Transcutaneous Monitoring and Pulse Oximetry RC 290.](https://reader035.fdocuments.net/reader035/viewer/2022062221/56649cc15503460f949889c5/html5/thumbnails/13.jpg)
Pulse Oximetry in Neonates
Advantages• Less costly than
TCM• Less maintenance
than TCM
Disadvantages• Does not give a direct
indication of perfusion• Can not assess
ventilation/PCO2• Does not give good
indication of hyperoxia– Alarms should be set
between 92% and 95%
![Page 14: Transcutaneous Monitoring and Pulse Oximetry RC 290.](https://reader035.fdocuments.net/reader035/viewer/2022062221/56649cc15503460f949889c5/html5/thumbnails/14.jpg)
Pulse Oximetry Is Used More for Surveillance than
Diagnosis