TP for EBT
-
Upload
mukhlizar-ismail -
Category
Documents
-
view
218 -
download
0
description
Transcript of TP for EBT
-
Treatment Planning for EBT
Rena Widita
The aims of treatment planning To localize the tumour volume in patient To define the target volume for treatment To measure the outline of the patient and to place
within it the target volume and other anatomical structures which may affect the dose distribution
To determine the optimum treatment configuration required to irradiate the target volume
To calculate the resultant dose distribution in the patient
To prepare an unambiguous set of treatment instructions for the radiographers.
Imaging/diagnostic
In treatment planning consideration must not only be given to the accuracy and precision of dose to a specified point/s, but also to the geometric tolerance of dose delivery
It has been estimated that there is a 4 mm positional error in treatment and this should be taken into account when the target volume is defined
Optimisation of treatment may be constrained by several factors
Physical factors: Fundamental nature of the radiation used The limitation of treatment equipment
Clinical factors, which may make it inappropriate to give the most sophisticated treatment when the patient will not gain any significant benefit.
SINGLE FIELDS
In photon therapy, treatments with a single field are generally used only for palliative purposes
It is only with low energy x-rays that single-photon beams are used curatively
-
Calculation Two data charts are required for the
calculation of the monitor units or a linac or the time to be set on a cobalt unit to deliver the prescribed dose to the target An output chart A depth dose table
The data are normally tabulated for square fields
Output Chart for 6 MV X-rays beam
Giving the required number of monitor units to deliver a dose of100cGy to the position of the dose maxon the central axis at 100 cm SSD
Field Size (cm2) Output (mu/100cGy)6 x 6 104.38 x 8 102
10 x 10 10012 x 12 98.515 x 15 96.720 x 20 94.8
Depth dose chart for 6 MV (SSD = 100 cm)
Depth (cm) Field Size (cm)6 x 6 8 x 8 10 x 10
1.5 100 100 1002 98.1 98.2 98.34 89.3 89.8 90.26 80.6 81.8 82.68 71.6 72.9 74
10 63.7 65.1 66.2
If the field is rectangular, an equivalent square field size which would have the same output and depth dose characteristics is used
s = the side of the equivalent square a,b = the sides of the rectangular field
Example Prescribed dose : 1500 cGy No. Of fractions : 5 Target depth : 6 cm Field size : 12 cm x 7 cm SSD : 100 cm Equivalent square : 8.8 cm PDD : 82.1% Output Factor : 101.2 mu/100cGy
Daily monitor units = (1500/5)*(101.2/100)*(100/82.1)= 370 mu
Output and depth dose data have been interpolated
Non-standard treatment distance
When the field size needed is larger than can be obtained at the standard SSD, and so an extended SSD is required
Shortened SSD can be used to increase the dose rate and therefore to decrease treatment time
Altering SSD causes a change in output factor and to a lesser, in the depth dose
The variation in output factor with SSD can be assumed to depend on the inverse square law (ISL) provided that the deviation from the standard SSD is not large (less than about 10 cm)
-
The correction for the output factor
f = the treatment distance fo = the nominal SSD dmax = the depth of dose max
The variation of dose with depth depends on both the attenuating properties of tissue and on the ISL
For a point at a depth d in a beam with its dose max at dmax, the correction to the depth dose measured at an SSD of fo is:
Opposed Coaxial Fields
Can be used to irradiate volume of tissue in which the tumour cannot be accurately defined
Or because the intention is to give a relatively low dose of palliation
For isocentric opposed field: separations of 12-24 cm would produce an error of 0.5% or less
Example for opposed coaxial fields
Prescribed dose : 3000 cGy No. Of fractions : 10 Separation : 18 cm Field size : 14 cm x 8 cm SSD : 100 cm Equivalent square : 10.2 cm Central depth dose : 70.2 * 2 = 140.4% Maximum percentage dose : 100 + 46.5 = 146.5% Output Factor : 99.8 mu/100cGy
Daily monitor units = (3000/10)*(99.8/100)*(100/140.4)= 213 mu
Maximum dose = 3000*(146.5/140.4) = 3130 cGy
Output and depth dose data have been interpolated
MULTIPLE FIELDS
-
Curative treatments generally require a higher dose
> 3 fields are used (except head & neck, 2 fields)
TP is concerned with the selection of the parameters required to produce the optimum dose distribution: Number of fields Orientation of fields Field sizes Wedges Weights
Beam Weighting A method of achieving uniformity in the
target volume to adjust the contribution from each beam
A
B D
C
O
I
IIIII
Point I II IIIA 70 48 44
B 58 62 40C 52 56 50D 58 42 56
O 60 52 48
Depth dose data for the field arrangement
Make sure that the doses to B and D from field I are equal. If not use a wedge on this field to make them equal
Dose to B = 62 w2 + 40 w3 and Dose to D = 42 w2 + 56 w3
20 w2 = 16 w3 (if w2 = 1, w3 = 1.25) Determine the weight for field I so that the total doses to point A and C
are equal Dose to A = 70 w1+ 48 w2 + 44 w3 and Dose to C = 52 w1+ 56 w2
+ 50 w3 18 w1 = 8w2 + 6w3 w1 = 0.86 Use the weights to calculate the doses to each point: Dose A = dose C = 163, dose B = dose D = 162, dose O = 164
A three field treatment technique. The dose to 5 points within the target volume is to be balanced by altering the relative weights of the three fields
Dose calculation Dose calculation for multiple fields involves the sumation of the dose
distributions of the individual fields. The total relative dose Dp at a point P for an irradiation with n fields
is given by
DDi,p = the depth dose at point P from the ith fieldwi = the weight for that field
Dose calculation within the patient
In the previous section, in the calculation of the dose at a point it was assumed that the radiation beam was normally incident on a unit density medium in practise, a patient differs from this homogeneous situation both in shape and composition these differences must be taken into account when calculating the depth dose.
For routine treatment planning by computer it is not necessary to know how these algorithms work.
However, the physicist responsible for the purchase of new planning software should understand the limitations and accuracy of the methods used
Algorithms for dose calculation in an inhomogeneous medium can be divided into 4 types that are essentially a compromise between speed and accuracy.
Type Algorithm Account for
1 [1] manual methods Effective depth2 [2] power law TAR Distance between inhomogeneity and
point of calculation3 [3] equivalent TAR Scatter correction Volume
integration of differential scatter-air ratiosPosition and shape of inhomogeneity
4 [3] convolution methods Monte Carlo As type 3 but includes electron equilibrium at interfaces
[1] manual methods Can be used manually but occasionally
are used in computer calculations
Only correct for the effective depth of a point, considering only changes in the primary component of dose
-
[2] power law TAR Takes account of the position of the
calculation point with respect to the inhomogeneity, but considers the inhomogeneity to be of uniform thickness over the beam width
This is known as the tissue-air-ratio (TAR) power law method
[3] equivalent TAR Scatter correction Volume integration of differential
scatter-air ratios
Takes into account the [3] shape of the inhomogeneity
These algorithms are the most complex used on commercial planning systems
[3] convolution methods Monte Carlo
Not considered to be practical owing to the lengthy calculation times involves
Monte carlo calculations are not practical but the technique has been used to provide the input data for convolution methods.