Toxicity • Clearly a critical issue – Small currents should not produce toxicity • Currently testing for silver toxicity – Utilizing toxicity specialist at NCSU’s College of Veterinary Medicine – Device has been design and fabricated – Testing will begin shortly

Transcript of Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing...

Page 1: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

Toxicity

• Clearly a critical issue– Small currents should not produce toxicity

• Currently testing for silver toxicity– Utilizing toxicity specialist at NCSU’s College of

Veterinary Medicine– Device has been design and fabricated– Testing will begin shortly

Page 2: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

Test of Other Bacteria

• Currently testing super bugs– MRSA– VRE– C. Dif– Cipro-resistant e. coli– Asperillus– Acinetobacter baumannii– Pseudomonas aeruginosa

Page 3: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

Long lasting and controlled release

• The hip implant would last for as much as 10 years running constantly• Ion release can be regulated by increasing or decreasing current• Device can be turned ON and OFF as required

Page 4: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

An almost ready for market product

ACE treatment on a polycarbonate hard surface. Hospital cafeteria trays could be made very inexpensively using this concept almost immediately.

Page 5: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

Other highly problematic targets

Keyboard Mouse

Page 6: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

Many Touch Surface Opportunities

Page 7: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

Technology Competitors

Safer, longer lasting, and more effective *BioguardCB

NANOVER TM

Page 8: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

Animal Testing

Page 9: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

Early Rat Testing Results

• 37 animals were used• 29 survived surgery and 1-2 weeks of recovery• 9 - 16 rats were used as control – all had

osteomyelitis• Developed the required protocol to infect a rat

tibia, and then install a device that was bactericidal

Page 10: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

Animal Test Results

• Established a protocol for inserting our device into the tibia of Sprague Dawley rats

• Surgeries were performed without any antibiotic!

Page 11: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

ACE Devices Used – First round of surgeries

+

Battery

Silver Cathode(Ø 1mm)

Silver Anode

Flexible Wire (Insulated)Resistance

Epoxy Bubble

80 mm(approx.)

+

Battery

Copper Cathode(35 gage wire)

Silver Anode

Flexible Wire (Insulated)Resistance

Epoxy Bubble

80 mm(approx.)

+

BatteryFlexible Wires

(Insulated)Silver Wire

(Anode)

Silver Paint (Cathode)

InsulationEpoxy Bubble 80 mm

(approx.)

Results2/3 rats were staph-free with this device. Problem, the device would frequently pierce through skin and organs.

0/6 rats were staph-free Problem, device floated uncontrollably.

1/3 rats were staph-free This device showed promise.

Page 12: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

ACE Devices used – Second round of surgeriesResults

0/3 rats were staph-free Problem, the device fell a part.

1/3 rats were staph-free This design concept looked promising.

2/3 rats were staph-free This was a good device for future testing

+

BatteryTeflon coated Silver wires

Anode

Cathode

ANODE(BARE SILVER WIRE)

CATHODE(35 GAGE COPPER WIRE)

1.4 V BATTERY

SILICONE ENCASING

20 mm

45 mm

4 mm

ANODE(BARE SILVER WIRE)

CATHODE(COATED SILVER WIRE)

1.4 V BATTERY

SILICONE ENCASING

Page 13: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

Our Device in a living rat – First set of surgeries third device

Page 14: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

Device designs

Page 15: Toxicity Clearly a critical issue – Small currents should not produce toxicity Currently testing for silver toxicity – Utilizing toxicity specialist at.

Some early rat testing

Rat tibia in agar with pseudomonas