NASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer Measurements Cindy W. Albertson and Willard A. Bauserman, Jr. MARCH 1993

Transcript of Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical...

Page 1: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

NASA Technical Memorandum 4407

Total Temperature Probes forHigh-Temperature HypersonicBoundary-Layer Measurements

Cindy W. Albertson and Willard A. Bauserman, Jr.

MARCH 1993

Page 2: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

NASA Technical Memorandum 4407

Total Temperature Probes forHigh-Temperature HypersonicBoundary-Layer Measurements

Cindy W. Albertson and Willard A. Bauserman, Jr.Langley Research CenterHampton, Virginia

Page 3: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

The use of trademarks or names of manufacturers in this

report is for accurate reporting and does not constitute an

o�cial endorsement, either expressed or implied, of such

products or manufacturers by the National Aeronautics and

Space Administration.

Page 4: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

Summary

This paper presents design details and test resultsof two types of total temperature probes that wereused for hypersonic boundary-layer measurements inthe Langley 8-Foot High-Temperature Tunnel. Theintent of each probe design was to minimize the totalerror and maintain a minimal size for measurementsin boundary layers 1.0 in. thick and less. A singleplatinum{20-percent-rhodium shield, used in bothdesigns, minimized radiation heat transfer losses dur-ing exposure to the high-temperature test stream.The main design di�erence was the vent-to-inlet arearatio. To reduce vertical averaging e�ects on thedata, the shield of the initial design (probe A) was attened at the ow entrance to an interior heightof 0.03 in., which resulted in a vent-to-inlet area ra-tio of 50 percent; to reduce vertical averaging e�ectsnear the wall even further, the shield of the laterdesign (probe B) was attened to 0.02 in., which re-sulted in an area ratio of 60 percent. A stainlesssteel structural support sleeve that was installed onprobe A was excluded from probe B, which resultedin a probe B outer diameter of 0.059 in., to allowcloser placement of the probes to each other and tothe wall. These small design changes to improve theresolution did not a�ect probe performance. Testswere conducted at boundary-layer-edge Mach num-bers of 5.0 and 6.2. The nominal free-stream to-tal temperatures were 2600� and 3200�R. The localReynolds number NRe;x, based on boundary-layer-edge conditions and distance from the leading edge,ranged from 3:5 � 106 to 28:6� 106. The reliabilityof the probes was extremely good. After 11 runs, notype A probes and only 1 of 10 type B probes failed.Temperature measurements in fully turbulent bound-ary layers compared well with Crocco-Busemann pre-dictions. The best performance in terms of recoveryfactor occurred when the wire-based Nusselt num-ber NNu;w was at least 0.04. Suggestions for futureprobe designs are included in \Recommendations."

Introduction

Total temperature measurements are required formany compressible ow experiments, including thosein relatively thin boundary layers. Many total tem-perature probes that are designed for boundary-layer measurements are unshielded to reduce sizeand to minimize spatial averaging, time constant,and disturbances to the boundary layer. (See refs. 1through 4.) Unshielded probes incur radiation errorsthat can be corrected in the data reduction process;however, the errors become larger and more di�cultto correct with increasing temperature. To minimize

the radiation error, single-shielded probes are oftenused. (See refs. 5 through 10.) Winkler (ref. 5) de-signed one of the earliest shielded probes which wasvented to allow the high-temperature gas to ow con-tinuously through the probe. Replenishing the high-temperature gas within the probe minimized errors.Winkler's probe shield was constructed from silica tominimize conduction and was coated with platinumto minimize radiation. Later designs incorporatedadditional features to improve accuracy. East andPerry (ref. 6) designed a probe with a heating ele-ment that was wrapped around the radiation shieldto minimize radiation losses from the thermocouplejunction. Weinstein (ref. 8) mounted a thermocoupleat the base of the probe to aid in computing correc-tions for conduction losses. Many of these probes aretoo large to resolve distributions in thin boundarylayers. Resolution of the boundary layer is particu-larly di�cult with probe installation in a �xed rake;however, �xed rakes are often necessary when highwall temperatures and the associated thermal expan-sion problems complicate the design of a traversingrake.

The total temperature probes described hereinwere developed for hypersonic boundary-layer mea-surements on a at plate (ref. 11) in the Langley8-Foot High-Temperature Tunnel (80HTT). Theprobe design concerns were the high total temper-ature of the test stream, the response time, thevibration attributed to ow unsteadiness, and thesmall size required to de�ne the total temperaturein boundary layers 1.0 in. thick and less. Two probedesigns were constructed with a vented, single shieldto minimize radiation heat transfer losses and main-tain a reasonably small size. To minimize verticalaveraging e�ects, both probe shields were attenedat the ow entrance. The smaller probe had a max-imum outside body diameter of 0.059 in. and an in-terior sampling height of 0.02 in.; these probes aresmaller than other shielded probes that are found inthe literature. Platinum{13-percent-rhodium versusplatinum (type R) thermocouple wire was requiredin both probe types for high total temperature mea-surement. A wire diameter of 0.010 in. was chosento withstand ow-induced vibration.

The probes were mounted in �xed rakes thatwere installed at various locations on the at-platemodel. The tests were conducted at a nominalfree-stream Mach number of 7, and the model waspitched at nominal angles of attack of 5� and 13�.The corresponding boundary-layer-edge Mach num-bers were 6.2 and 5.0, respectively. Near the wall, theMach number ahead of the probes was as low as 1.7.The local Reynolds number NRe;x, based on both the

Page 5: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

conditions at the boundary-layer edge and the dis-tance from the leading edge, ranged from 3:5 � 106

to 28:6 � 106. The local Reynolds number NRe;p,based on thermocouple junction diameter and esti-mated probe internal ow conditions, ranged from 63to 2470.

Symbols

A area, in2

C circumference, in.

c speci�c heat, Btu/lbm-�R

cp speci�c heat at constantpressure, Btu/lbm-�R

D1 through D5 constants used in eqs. (A5)and (A11)

diam diameter, in.

dJ junction diameter, in.

dw wire diameter, in.

E measurement error, �R

F con�guration factor

H enthalpy, Btu/lbm

H distance from tunnelcenterline, in.

h heat transfer coe�cient,

Btu/sec-ft2-�R

I cross-sectional moment of

inertia, in4

i.d. inner diameter

k thermal conductivity,Btu/sec-ft-�R

L cantilevered length, in.

L1; L2 thermocouple wire lengths,in. (see �g. 13(a))

M Mach number

M2 Mach number behindnormal shock (see �g. 9)

m bending moment, in-lbf

N velocity power law expo-nent, equation (C2)

NBi Biot number, (hpdw=4kw)1=2

NNu;p Nusselt number based onthermocouple junctiondiameter and local owconditions inside probe,upstream of junction,hpdJ=kp

NNu;w Nusselt number based onthermocouple junctiondiameter, thermocouplewire conductivity, andlocal ow conditions insideprobe, upstream of junction,hpdJ=kw

NPr Prandtl number, �cp=k

NPr;p Prandtl number based onlocal ow conditions insideprobe, (�cp=k)p

NRe;p Reynolds number basedon thermocouple junctiondiameter and local owconditions inside probe,upstream of junction,�pUpdJ=�p

NRe;x local Reynolds numberbased on distance fromleading edge and evaluatedat boundary-layer-edgetemperature, �eUex=�e

NRe;1 free-stream unit Reynoldsnumber, �1U1=�1,1 per foot

n distance along thermocouplewires, in. (see �g. 13(a))

o.d. outer diameter

p pressure, psia

_q heat transfer rate,

Btu/sec-ft2

r radius, in.

T temperature, �R

T2 static temperature behindnormal shock, �R (see �g. 9)

t time, sec

U velocity, ft/sec

V volume, in3

w distributed load per unitlength, lbf/in.

2

Page 6: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

x; y plate surface coordinates, ft(see �g. 4)

Y modulus of elasticity, psi

ymax maximum de ection, in.(eq. (D3))

z distance measured normalfrom plate surface, in. (see�g. 4)

� angle of attack, deg

�T

thermal boundary-layerthickness, in.

�U

velocity boundary-layerthickness, in.

" emittance

� viscosity, lbm/sec-ft

� density, lbm/ft3

�B

Stefan-Boltzmann constant,

Btu/sec-ft2-�R4

�max maximum bending stress,psi (eq. (D1))

� thermal response time, sec

Subscripts:

aw adiabatic wall

c convection

e boundary-layer edge, justoutside boundary layer

J thermocouple probejunction

k conduction

l local condition

p ow condition inside probe,just upstream of thermo-couple junction

r radiation

si condition at interior ofradiation shield

so condition at exterior ofradiation shield

t total condition

tc total condition in combustor

te total condition at boundary-layer edge

tk known total condition

t1 total free-stream condition

t2 total condition behindnormal shock

U velocity

w thermocouple wire

1 free stream

Description of Experiment

Total Temperature Probes and Rakes

Design Considerations

The goal of the total temperature probe is to pro-duce a ow environment at the thermocouple junc-tion which will enable the thermocouple to measurethe total temperature accurately. The accuracy ofthe probe depends on the balance of the convectiveheat transfer between the gas and the junction, theradiation from the junction to the shield, and theconduction from the junction along the thermocou-ple wires to the cooler probe support. The di�erencebetween the total temperature of the gas Tt and thejunction temperature T

Jis referred to as the total

error Et. The equation for the total error is given as

Et = EU+E

k+Er (1)

In equation (1), EU, E

k, and Er represent the decou-

pled errors due to the velocity of the ow inside theprobe, conduction along the thermocouple wires, andradiation to the junction surroundings, respectively.A discussion of these errors and how to reduce themis given below.

Error due to ow velocity. The error due to ow velocity E

Uis the di�erence between the to-

tal temperature of the gas to be measured Tt andthe adiabatic wall temperature of the thermocouplewires Taw. For all Prandtl numbers, Taw approachesTt with decreasing velocity. Therefore, E

Ucan be

decreased by reducing the ow velocity inside theprobe. However, the convective heat transfer to thethermocouple bead is also reduced when the ow ve-locity decreases, which leads to an increase in theconduction and radiation errors. Therefore, the ve-locity must not be reduced to the extent that thetotal error increases for the design conditions. Be-cause this velocity cannot be measured easily, ex-periments to reduce E

Uemphasize the probe vent-

to-inlet area ratio. (See refs. 5 and 10.) These

3

Page 7: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

experiments indicate that the optimum area ratiovaries with probe size.

Conduction error. The conduction error is afunction of the length and diameter of the thermo-couple wires, the thermophysical properties of thewires, convective heat transfer to the wires, and theprobe support temperature. (See eq. (A11).) Themost practical way to reduce the conduction error isto maximize the length and minimize the diameterof the wires, because selection of the thermophysicalproperties of the wire is limited. The selection ofthermocouple wire type is usually determined by thetemperature of the gas to be measured.

Radiation error. The radiation error is a functionof the emittance of both the shield and the thermo-couple wires, the radiation view factor between theshield and the wires, and the temperature of the ra-diation shield. Selection of a shield material witha low emittance to reduce heat loss to the probesurroundings can reduce this error. However, thestrength of the shield material at high temperaturesmust also be considered, particularly in an environ-ment with large ow-induced vibrations. The junc-tion and wires should be enclosed as much as possiblewithin the shield. However, the shield length and di-ameter must be selected such that the junction is notimmersed in the shield boundary layer. The radia-tion error can be reduced further by increasing thetemperature of the radiation shield, particularly nearthe thermocouple junction and wires. The tempera-ture of the shield can be increased by increasing theinternal ow velocity, which increases the heat trans-fer to the shield, and by decreasing the heat con-ducted along the shield to the cooler support. Theheat conducted along the shield can be reduced byincreasing the shield length and reducing the shieldthickness. An increase in the number of concentricshields increases the temperature of the inner shieldand reduces the radiation error; however, the diame-ter of the probe is increased, which is undesirable fordetailed boundary-layer measurements.

Probe and Rake Description

The present probe design is a compromise be-tween a reduction of errors, size minimization, andthe ability to withstand ow-induced vibrations. Forthe chosen design, the total calculated error rangesfrom 0.4 to 6.2 percent. Conduction losses that occurat the low probe Reynolds number conditions, partic-ularly deep within the boundary layer at a relativelylow total temperature, cause most of the total er-ror. Details of these errors are given in appendix A,and methods to reduce errors in future designs aredescribed in \Recommendations."

Two probe types were used to measure the to-tal temperature distributions through the boundarylayer. These probes are illustrated in �gure 1 andare shown schematically in �gure 2. Both probescontained a platinum{13-percent rhodium versusplatinum thermocouple (type R) that was mountedinside a single 0.005-in-thick radiation shield. Athermocouple wire diameter of 0.010 in. was chosento withstand the ow-induced vibration. Platinum{20-percent rhodium was selected as the shield mate-rial because of its relatively low emittance and rela-tively high strength at high temperatures. The maindi�erence between the two probes was the vent-to-inlet area ratio. To minimize vertical averaging ef-fects on the data, the shield of the initial probe(probe A) was attened at the ow entrance to aninterior height of 0.03 in. To further reduce verti-cal averaging e�ects near the wall, the shield of thelater design (probe B) was attened to 0.02 in. atthe ow entrance. Two vent holes with diameters of0.022 in. were drilled in both probe shields to allowthe probes to vent. The resulting vent-to-inlet arearatios were 50 and 60 percent. Area ratios in thisrange were shown by Bartlett, Edwards, and Hillier(ref. 10) to provide optimal temperature recovery, al-though Bartlett's probes had interior heights at the ow entrance of 0.035 in., which is larger than theprobes described herein. Another di�erence was thatprobe A incorporated a 308 stainless steel structuralsupport sleeve to withstand the ow-induced vibra-tions and to provide a bearing surface to tighten theset screw that holds the probe in the rake. The out-side body diameter of probe A with this sleeve was0.094 in. (See �g. 2(a).) The sleeve was excludedfrom probe B to allow closer placement of the probesto each other and to the wall of the plate. (See�g. 2(b).) The outside body diameter of probe Bwas 0.059 in.

To assemble the probes, the thermocouple wireswere threaded through the two-holed ceramic insu-lation and were then joined to form the thermo-couple junction. For probe A, the stainless steelstructural support sleeve was then installed andbonded to the �berglass-covered thermocouple wirewith epoxy. (See �g. 2(a).) The platinum radiationshield was then installed and spot welded to the sup-port sleeve. Because probe B lacked a support sleeve,the assembly process was simpler. For this probe, theplatinum shield was bonded to the ceramic insulationand �berglass-covered thermocouple wire with epoxy.(See �g. 2(b).)

The probes were mounted in two di�erent �xed-rake struts, rake A and rake B. (See �gs. 3(a)and 3(b).) Rake A, the initial design, contained

4

Page 8: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

11 type A probes. Rake B contained 15 tempera-ture probes; 8 of the 15 were type B probes within0.54 in. of the wall. The remaining probes in rake Bwere type A with the exception of one type B probeoutside the boundary layer at z = 2:200 in. Theprobe types and their locations in each rake are givenin table I. Both rakes were machined to facilitatereplacement of damaged temperature probes. Castbronze alloy 5 was selected as the rake materialbecause of its heat-sink characteristics , which en-able the rake to survive during the relatively shortexposure times to the high-temperature ow.

Test Con�guration

The test model was a large, heavily instrumented at plate that was installed on the panel holder,a generalized test apparatus for the 80HTT. (See�g. 4.) A sharp leading edge was attached to thefront of the panel holder. The model was 9.7 ftlong and 4.3 ft wide. Boundary-layer rake assemblies,which consisted of total pressure, static pressure, andtotal temperature probes, were used to survey theboundary layer at various locations on the plate.(See �gs. 4 and 5.) Model details and additionalinstrumentation are presented in reference 11.

Data Acquisition and Reduction

All data channels, including thermocouple out-puts, were sampled and recorded 20 times per secondper channel with a 12-bit digital data acquisition sys-tem. All signals were �ltered with 10-Hz low-pass �l-ters and digitized before being recorded on magnetictape. The total estimated accuracy of the data ac-quisition equipment is within 1 percent. Additionaldetails of this equipment and its accuracy are givenby Nowak. (See ref. 12.)

Thermocouple outputs from the total tempera-ture probes were converted to temperatures fromthe thermocouple tables given in reference 13 for atype R thermocouple. Pretest calibrations of thethermocouple wire used in the probes were con-ducted up to 2300�R. These calibrations indicatedsmall, consistent variations from the tables. Addi-tional details regarding the data reduction are givenin reference 11.

Test Facility

The tests were conducted in the 80HTT, whichis shown schematically in �gure 6. This facility isa large blowdown wind tunnel that provides true-temperature ight simulation at a nominal free-stream Mach number of 7 and at pressure altitudes

from 80 000 to 126000 ft. The high-energy testmedium consists of a mixture of methane and airburned under high pressure in the combustor. Thetest medium then expands through an axisymmetric,conical contoured nozzle to achieve the test cham-ber Mach number. The ow in the test cham-ber is a free jet that enters a straight-tube super-sonic di�user, where it is pumped to the atmosphereby a single-stage annular air ejector. The tunnelcan be operated at total temperatures from 2400�

to 3600�R and at free-stream dynamic pressuresfrom 240 to 1800 psf. The corresponding free-streamReynolds number NRe;1, based on a model length of

10 ft, ranges from 3 � 106 to 30� 106.

Models are positioned below the test chamber(�g. 7) during tunnel startup and shutdown to min-imize aerodynamic loads. When ow conditions areestablished, the model is inserted into the ow by ahydraulically actuated elevator. Before tunnel shut-down, the model is withdrawn from the ow. Themaximum test time for the facility is 120 sec. Thepresent model was positioned 1.0 ft from the nozzleexit (�g. 7) and was exposed to the ow for no morethan 10 sec. Additional details of the 80HTT can befound in reference 14.

Test Conditions

The tests were conducted at nominal total tem-peratures of 2600� and 3200�R. The nominal free-stream Mach number was 7, and the free-streamReynolds number NRe;1, based on plate length,

ranged from 5:5� 106 to 18:0� 106. To obtain rel-atively high local Reynolds numbers, the model waspitched at angles of attack of 5� and 13�, which pro-duced boundary-layer-edge Mach numbers of approx-imately 5.0 and 6.2, respectively. Most tests wereconducted at an angle of attack of 13�. This angle ofattack produced local Reynolds numbers high enoughfor large areas of equilibrium turbulent boundary-layer ows over the plate, as discussed in refer-ence 11. The local Reynolds number NRe;x, basedon boundary-layer-edge conditions and distance fromthe leading edge, ranged from 3:5�106 to 28:6�106;the local Reynolds number NRe;p, based on internalprobe diameter and ow conditions, ranged from 63to 2470. The local ow conditions and those up-stream in the combustor are given in table II. De-tails of how these conditions were determined andtheir accuracy are given in reference 11.

Results and Discussion

Selected sets of data obtained from the totaltemperature probes are discussed in the following

5

Page 9: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

sections to aid in the assessment of probe perfor-mance. The measurements shown herein were ob-tained along the plate centerline at y = 0 in. (Thecoordinate system is de�ned in �g. 4.) A completedata tabulation for each run is given in appendix Hof reference 11.

The probe reliability was extremely good. After11 runs, no type A probes failed and only 1 of10 type B probes failed.

Temperature Histories

A typical temperature history TJ from a type Atotal temperature probe, placed outside the bound-ary layer at z = 2:500 in. and 72 in. from the nozzleexit plane, is shown in �gure 8(a). The data areshown for an intermediate Reynolds number NRe;p

of 1016; NRe;p corresponds to the thermocouple junc-tion diameter and conditions inside the probe thatwere determined from total temperature measure-ments, total pressure measurements, and the vent-to-inlet area ratio. The total temperature measuredupstream in the combustor Ttc is also shown in �g-ure 8(a). In �gure 8(b), the approximate distanceof the probe from the test section centerline H isshown. (The oscillations shown in the probe positionnearH = 0 in. are probably caused by a problem thatresults from the potentiometer used as a model posi-tion indicator. No evidence exists to indicate that thetest panel de ected this amount during a run.) Be-fore the model entered the test stream, the tempera-ture indicated by the probe was approximately 500�R(�g. 8(a)). The probe began to respond to the tem-perature of the test core approximately 0.65 sec frominitial model movement, or approximately 0.2 sec be-fore the probe crossed the nozzle edge. This responseindicates that the jet boundary expanded beyondthe nozzle edge. After about 2.1 sec, or approxi-mately 1.45 sec after the probe �rst began to reactto the ow, the probe temperature nearly matchedthat of Ttc.

The thermal response times of the probes couldnot be accurately determined from the present exper-iment because of numerous complications. The maincomplication was that the probes were not exposedto a step change in temperature. Total temperaturesurveys of the test core (ref. 11) indicate that thefree-stream total temperature Tt;1 (�g. 6) is lowerthan Ttc near the edge of the nozzle and gradually in-creases to approximately the level of Ttc near the testsection centerline. Also, the exact temperature pro-�le to which the probes are exposed during entry intothe test stream is not precisely known, because free-stream total temperature surveys are taken while the

model is below the test core. However, if a step in-put in temperature is assumed, the thermal responsetime � , corresponding to the time required for thejunction to reach 63.2 percent of the surrounding gastemperature, is estimated between 0.3 and 1.5 secfor the two extremes in ow conditions. (See appen-dix B.) Likewise, the predicted value for 5� , whichcorresponds to the time required for the junctiontemperature to reach 99.3 percent of the gas tem-perature, should range from 1.5 to 7.5 sec, which iswithin the desired model exposure time.

Correlation of Temperature Recovery

Factor

The di�erence between the junction temperatureand the true stagnation temperature is a function ofthe convective heat transfer to the junction and thelosses due to the conduction, radiation, and velocity.(See Winkler, ref. 5.) As convection increases, therelative losses decrease, which results in a higher tem-perature recovery. Winkler assumed that the lossesare mostly caused by conduction and showed thatthe recovery factor TJ � T

1;l=Ttk � T1;l is propor-

tional to the Nusselt number NNu;w at the probejunction that is based on wire conductivity, whereNNu;w = hpdJ=kw. Here, TJ is the total tempera-ture indicated by the probe thermocouple junction,Ttk is a known total temperature, and T

1;l is thelocal static temperature just upstream of the probe.(See �g. 9.) Also, hp is the convective heat transferto the junction, dJ is the junction diameter, and kwis the conductivity of the thermocouple wire. Win-kler then showed that the recovery factor correlatesto a simpler parameter that is based on wire Nus-selt number, given as �J � T�1:75

J, where �J is the

density of the gas at the junction. A simpler parame-ter, pt2�T�1:75

J, can be derived from the equation of

state. Here, pt2 is the total pressure behind a normalshock and the assumed pressure of the gas ahead ofthe junction.

To assess the performance of the total tempera-ture probes, the recovery factor TJ �T

1;l=Ttk�T1;l

is plotted as a function of wire-based Nusselt num-ber at the probe junction NNu;w in �gure 10 andas a function of the modi�ed Winkler parameterpt2� T�1:75

Jin �gure 11. The wire-based Nusselt

number was determined from equation (A1) in ap-pendix A, which relates NNu;p to the probe Reynoldsnumber NRe;p. Here, NNu;p is based on the conduc-tivity of the gas within the probe kp and di�ers fromNNu;w by a factor of kp=kw. The probe Reynoldsnumber was calculated from the measured total tem-perature, the total pressure measured by a nearbytotal pressure probe, and the probe internal Mach

6

Page 10: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

number M2 (�g. 9). The probe internal Mach num-ber was calculated, assuming inviscid ow, from theprobe inlet-to-vent area ratio. The data were takenfrom measurements outside the boundary layer alongthe plate centerline. The combustor total tempera-ture Ttc was used as the known total temperature tocompute the recovery factor. The data show scat-ter in both plots; however, there is no discernibledi�erence in the performance of probes A and B atthe high- and low-temperature conditions. As ex-pected, the recovery factors increase with NNu;w and

pt2 � T�1:75J

. The recovery factors range from 0.89to 1.02 compared with the uncorrected combustor to-tal temperature. Although some of these values werehigher than 1.0, most of the factors compare reason-ably well with the results from reference 15. Thebest performance occurs when NNu;w � 0:04. Themodi�ed Winkler parameter, a more easily calculatedvariable, should be at least 25 � 10�6 psia-�R�1:75

for best performance, at least for the nominal free-stream local Mach numbers of 5.0 and 6.3. Becausethe correlations are based on conditions inside theprobe, these results are expected to apply to otherfree-stream local Mach numbers.

Distributions of Boundary-Layer Total

Temperature

To evaluate the total temperature distributionsmeasured in the boundary layer, the linear Crocco-Busemann enthalpy-velocity relationship (ref. 16)was applied. Additional assumptions and details ofthis relationship are given in appendix C.

The total temperature data and Crocco-Busemann predictions are shown for Reynolds num-bers NRe;x of 28:6 � 106, 13:6 � 106, and 4:5 � 106

in �gures 12(a), 12(b), and 12(c), respectively. Theboundary-layer thickness �U is shown on the plotsfor reference purposes. The rake B data and predic-tions show good agreement for the two high Reynoldsnumber cases (�gs. 12(a) and 12(b)) in which thethermal and velocity boundary layers were fully de-veloped. (See ref. 11.) The data obtained at the lowReynolds number condition (�g. 12(c)) were transi-tional (ref. 11). The lack of agreement between thedata and prediction at the low Reynolds number con-dition may result from diminished probe accuracy atsuch conditions or from a breakdown in the Crocco-Busemann prediction. The Crocco-Busemann pre-diction is not applicable to transitional boundarylayers when the thermal and velocity boundary lay-ers (ref. 17) are nonsimilar. The irregularity inthe total temperature pro�le at z = 0:8 in. for thelow Reynolds number condition (�g. 12(c)) is alsopresent in the Mach number distribution obtained at

this condition (ref. 11). These irregularities indicatethat a weak shock is present in the ow outside theboundary layer.

Recommendations

For future probe designs, both the radiation andconduction errors discussed in appendix A can bereduced by increasing the length of the shield andthermocouple wires, particularly the portion exposedto the ow inside the probe. This increase in lengthwill increase the area available for convection andwill reduce the heat conducted down the wires andthe shield to the relatively cool rake. This changewill also reduce radiation losses from the junction,because the temperature of the radiation shield willbe higher. The structural analysis given in appen-dix D indicates that the length of the shield and wirescan be increased to some extent without signi�cantlyweakening the probe. However, the material proper-ties used in the analysis are uncertain at 3300�R be-cause they must be extrapolated from lower tempera-ture data. The thermal response time of the junctioncan be reduced in future probes by forming the junc-tion with a spot weld rather than with a conventionalbead weld to reduce the junction size.

The vent-to-inlet area ratio has not been opti-mized for the probe designs described in this paper.Although the area ratios were within the values rec-ommended in reference 10, the present probes aresmaller than those in reference 10, and the ow insidethe probe is dominated more by viscosity. Therefore,experiments should be conducted to determine if alarger vent-to-inlet area ratio is required for optimalprobe performance.

Concluding Remarks

This paper presents the design details and test re-sults of two types of total temperature probes thatwere used for hypersonic boundary-layer measure-ments in the Langley 8-Foot High-Temperature Tun-nel. The intent of each probe design was to minimizethe total error and maintain a minimal size for mea-surements in boundary layers 1.0 in. thick and less.A single platinum{20-percent-rhodium shield, usedin both designs, minimized radiation heat transferlosses during exposure to the high-temperature teststream. The main design di�erence was the vent-to-inlet area ratio. To reduce vertical averaging ef-fects on the data, the shield of the initial design(probe A) was attened at the ow entrance to aninterior height of 0.03 in., which resulted in a vent-to-inlet area ratio of 50 percent; to reduce vertical av-eraging e�ects near the wall even further, the shieldof the later design (probe B) was attened to 0.02 in.,

7

Page 11: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

which resulted in an area ratio of 60 percent. A stain-

less steel structural support sleeve that was installed

in probe A was excluded from probe B, which re-

sulted in a probe B outer diameter of 0.059 in., to al-

low closer placement of the probes to each other and

to the wall. These small design changes to improve

the resolution did not a�ect probe performance.

Tests were conducted at boundary-layer-edge

Mach numbers of 5.0 and 6.2. The nominal free-

stream total temperatures were 2600� and 3200�R.

The local Reynolds number NRe;x, based on

boundary-layer-edge conditions and distance from

the leading edge, ranged from 3:5�106 to 28:6�106.

The local Reynolds number inside the probe NRe;p

ranged from 63 to 2470. The reliability of the probes

was extremely good. After 11 runs, no type A probes

and only 1 of 10 type B probes failed. Predictions

of probe thermal response times ranged from 0.3

to 1.5 sec. Temperature recovery factors ranged

from 0.89 to 1.02 compared with the uncorrected

upstream combustor temperature measurement. The

best performance in terms of recovery factor occurred

when the wire-based Nusselt number NNu;w was

at least 0.04. The modi�ed Winkler parameter, a

more easily calculated variable, should be at least

25 � 10�6 psia-�R�1:75 for optimum performance.

Temperature measurements in fully turbulent bound-

ary layers compared well with Crocco-Busemann

predictions.

The total calculated error of the probes in the

test range varies from 0.4 to 6.2 percent. Most of the

errors that occur at the low probe Reynolds number

conditions are estimated to be caused by conduction

losses, particularly deep within the boundary layer

where the total temperature is relatively low.

NASA Langley Research Center

Hampton, VA 23681-0001

January 4, 1993

8

Page 12: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

Appendix A

Calculations of Theoretical Probe Errors

To calculate the theoretical measurement errorsof the total temperature probes, an error analysiswas performed for the two extremes in ow condi-tions that produced the highest and lowest convec-tive heat loads to the thermocouple junction. The�rst extreme (case 1) corresponded to a probe outsidethe boundary layer at the highest probe Reynoldsnumber and at a relatively high total temperaturetest condition; the second extreme (case 2) corre-sponded to a probe close to the wall at the low-est probe Reynolds number and at a relatively lowtotal temperature test condition. The probe in-ternal Reynolds numbers NRe;p, based on junctiondiameter, were 2470 and 63 for cases 1 and 2,respectively. The total temperatures used in the theo-retical calculations were 3140� and 2340�R for cases 1and 2, respectively. These temperatures were ob-tained from the upstream combustor temperaturemeasurement. Because case 2 corresponds to aprobe within the boundary layer, the total temper-ature of 2340�R used in the calculations was de-rived from the combustor temperature measurementbased on the Crocco-Busemann relationship given inappendix C.

Before calculating the errors, NRe;p was calcu-lated to determine the convective heat load to thethermocouple wires. The value of NRe;p was calcu-lated with the measured total temperature TJ , thetotal pressure behind a normal shock (measured froma nearby total pressure probe) pt2, the probe internalMach number, and the uid and transport propertiesof methane-air combustion products given in refer-ence 18. (See �g. 9.) For simplicity, pt2 was assumedto be the pressure of the gas ahead of the thermo-couple junction. The probe internal Mach num-ber was calculated, assuming inviscid ow, from theprobe inlet-to-vent area ratio. The probe Nusseltnumber NNu;p was then calculated from the corre-lation given in reference 15 for wires parallel to the ow as follows:

NNu;p= 0:095N0:31Pr;pN

0:674Re;p (A1)

From equation (A1), the heat transfer coe�cient hpwas then calculated.

Finally, the total probe error Et was assumed tobe equal to the sum of the individual decoupled errorsas follows:

Et = EU +Ek +Er (A2)

In equation (A2), EU , Ek, and Er represent the er-rors from the internal velocity of the ow, conductionalong the thermocouple wires, and radiation to thejunction surroundings, respectively. These errors arediscussed in the following sections.

Error Due to Flow Velocity

The error due to the ow velocity, as follows, isthe di�erence between the total temperature Tt2 andthe adiabatic wall temperature Taw, where Taw isgiven by the standard de�nition and laminar owover the junction is assumed:

Taw = T2+N1=2Pr;p(Tt2 � T2) (A3)

In equation (A3), T2 is the static temperature in-side the shield behind the normal shock. (See �g. 9.)As mentioned in \Probe and Rake Description," thevent-to-inlet area ratios were 50 and 60 percent forprobes A and B, respectively; these ratios produceoptimum temperature recovery (ref. 10) and corre-spond to internal Mach numbers of 0.3 and 0.4 forthe type A and B probes, respectively. The corre-sponding errors due to velocity are 6:3� and 4:7�Rfor cases 1 and 2, respectively. (See table III.)

Conduction Error

For simplicity, conduction errors were computedneglecting radiation. (See �g. 13(a).) First, theconvective heat transfer to the thermocouple wireswas assumed equal to the conduction along the wiresfrom n = 0 to n = L1. The corresponding steady-state, one-dimensional heat balance between the con-vective heat transfer _qc and the conduction heattransfer _qk is given as

hpC [Taw� T (n)] = �kwA

d2T

dn2

!(A4)

In equation (A4), C is the circumference of a thermo-couple wire, A is the cross-sectional area, kw is thewire conductivity, and hp is the convection coe�-cient calculated from equation (A1). Substituting�(n) = T (n)� Taw into equation (A4) and solvingyields

� (n)�D1 cosh

�n

�4hp

dwkw

�1=2�+D2sinh

�n

�4hp

dwkw

�1=2�(A5)

Equation (A5) was then solved with the followingboundary conditions for n = 0:

kw

�d�

dn

�= hp� (n = 0) (A6)

9

Page 13: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

and

� (n = 0) = (TJ � Taw) (A7)

In equation (A7), TJ is the temperature of thethermocouple junction. The solution of equa-tion (A5) is

T (n)� Taw = (TJ � Taw)

(cosh

"n

�4hpdwkw

�1=2#

+NBi sinh

"n

�4hpdwkw

�1=2#)

(A8)

In equation (A8), NBi is the Biot number and is equal

to (hpdw=4kw)1=2. The conduction error Taw � TJ

could be determined if, for example, the tempera-ture at n = L1 was known. Because space was lim-ited, no additional thermocouples could be locatedinside the probes; therefore, the heat transfer wasalso computed along the insulated portion of thethermocouple wire from n = L1 to n = L2, wherethe assumed wire temperature was 540�R. The corre-sponding steady-state, one-dimensional heat transferequation is given as

_q = �kwA

�d�

dn

�(A9)

In equation (A9), the heat transfer rate _q is assumedto be constant along the insulated portion of thethermocouple wire. The solution of equation (A9)is given as

_q = kwAT (n = L1)� T (n = L2)

L1� L2

(A10)

The derivative of equation (A8) was taken at n = L1

to obtain the heat entering the insulated portion ofthe thermocouple wire. When this expression is com-bined with equation (A10), T (n = L1) can be elim-inated. The resulting equation for the conductionerror is

Ek = (Taw� TJ) =Taw� T (n = L2)

D3

(A11)

where D3 = D4+D5(L2�L1) and

D4 = cosh

"L1

�4hp

dwkw

�1=2

#

+NBi sinh

"L1

�4hp

dwkw

�1=2

#

D5 =

�4hp

dwkw

�1=2(sinh

"L1

�4hp

dwkw

�1=2#

+NBi cosh

"L1

�4hp

dwkw

�1=2

#)

The conduction errors Taw� TJ are 2:2�R for case 1and 85:7�R for case 2, which correspond to 18.0 and58.9 percent of the total error, respectively. (Seetable III.) A constant conductivity of 1:20 � 10�3

Btu/sec-in-�R was used in the calculations (ref. 19).This value corresponds to pure platinum at 3200�Rand represents the highest conductivity over the tem-perature range of the analysis (540� to 3200�R);therefore, the result should be a conservative esti-mate of conduction error.

Radiation Error

For simplicity, radiation errors were computed ne-glecting conduction, as illustrated in �gure 13(b).The steady-state heat balance between the convec-tion to the junction _qc;J and radiation from the junc-tion to the shield _qr;J to si is given as

hpAJ (Taw� TJ) = AJ�BF�T 4

J � T 4

si

�(A12)

In equation (A12), F is the con�guration factorfor one gray surface that encloses a second surface(ref. 20); thus,

F =1

1

"J+ AJ

Asi

�1

"si� 1

� (A13)

The steady-state heat balance between the convec-tion to the shield _qc;si, the radiation from the junc-tion to the shield _qr;J to si, and the radiation fromthe shield to the surroundings _qr;so to 1, is given by

hsiAsi(Taw� Tsi) = Aso�B"so

�T4

so� T4

1

��AJ�BF

�T4

J � T4

si

�(A14)

The convective heat transfer coe�cient for the shieldnear the junction hsi was determined from the lami-nar thermal entry-length solution for ow inside cir-cular tubes (ref. 21). Assuming Tsi = Tso, equa-tions (A12) and (A14) were solved iteratively untilconvergence within 0:5�R. A single value of 0.18(ref. 19), which corresponds to polished platinumat 3200�R, was used for "J , "si, and "so. Becausethe emittance of platinum increases with temper-ature, a conservative estimate of the radiation er-rors should result. For cases 1 and 2, the radi-ation errors Taw� TJ are 3:7� and 55:0�R, which

10

Page 14: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

represent 30.3 and 37.8 percent of the total error,

respectively.

The total error from this analysis ranged

from 12:2� to 145:4�R for the two extremes. These

values represent a total error of 0.4 and 6.2 percent,

respectively, in the total temperature measurements.

Most of the errors that occur at the low probe

Reynolds number conditions are estimated to be

caused by conduction losses, particularly deep within

the boundary layer where the total temperature also

is relatively low.

11

Page 15: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

Appendix B

Calculations of Theoretical Probe

Thermal Response Time

To simplify the calculations for theoretical ther-mal response time of the total temperature probes,conduction and radiation losses were neglected. Also,the temperature of the thermocouple junction wasassumed to be uniform. The resulting heat transferequation is

hpA [Tt � TJ (t)] = �cV

�dTJdt

�(B1)

In equation (B1), A is the projected frontal areaof the thermocouple bead, and �; c, and V are itsdensity, speci�c heat, and volume. The values ofdensity and speci�c heat used in the calculations were0.775 lbm/in3 and 0.0314 Btu/lbm�-R; these valuescorrespond to platinum at 540�R (ref. 22). The heattransfer coe�cient hp was determined by calculatingNRe;p for a given ow condition and then usingequation (A1). The probe Reynolds number canbe calculated from the measured total temperature

and pressure behind a normal shock and the probeinternal Mach number, computed from the vent-to-inlet area ratio. Substituting �(t) = TJ (t)� Tt and� = �cV=hpA into equation (B1) and solving yields

� = C1 exp (�t=�) (B2)

In equation (B2), C1 is an unknown constant.Application of the initial condition of�(t = 0) = TJ (t = 0)�Tt and rearrangment of termsyields

TJ (t)� TJ (t = 0)

Tt � TJ (t = 0)= 1� exp (�t=�) (B3)

From equation (B3), the thermal response time �corresponds to the time required for the junction toreach 63.2 percent of the surrounding temperatureof the gas after a step change in the gas temper-ature. For the two extremes in ow conditions, thepredicted � ranged from 0.3 to 1.5 sec. The predictedrange for 5� , which corresponds to the time requiredfor the junction temperature to reach 99.3 percent ofthe gas temperature, was 1.5 to 7.5 sec. These valuesare within the model exposure times.

12

Page 16: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

Appendix C

Crocco-Busemann Predictions

To evaluate the total temperature distributionsmeasured in the boundary layer, the linear Crocco-Busemann enthalpy-velocity relationship (ref. 16)was applied. The linear relationship is derivedfrom the energy equation for a two-dimensional,zero pressure gradient boundary layer and from theassumption that H = H(U). This relationship isgiven as

Ht�Hw

Hte �Hw=

U

Ue(C1)

The linear relationship assumes similarity betweenthe thermal and velocity boundary layers and is validfor boundary layers along isothermal walls with aPrandtl number of unity. However, equation (C1) isconsidered a reasonable approximation for an air testmedium with NPr = 0:7 (ref. 17). For the present

methane-air combustion products test medium, thePrandtl number is nominally 0.76.

To relate total enthalpy to location in the bound-ary layer, the following power law expression wasused:

U

Ue=

�z

�U

�1=N

(C2)

In equation (C2), the exponent N , the velocityboundary-layer thickness �U , and the boundary-layer-edge velocity Ue were derived from the pres-sure and temperature measurements that were ob-tained in the boundary layer (ref. 11). Combiningequations (C1) and (C2) yields

Ht =

�z

�U

�1=N

(Hte�Hw) +Hw (C3)

Temperature was inferred from enthalpy with theMollier charts in reference 18 for methane-aircombustion products.

13

Page 17: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

Appendix D

Probe Stress Analysis

The stresses on the shields were analyzed by �rstassuming that the dominant load was caused by the uctuating pressures beneath the turbulent bound-ary layer. Data compiled in reference 23 indicate thatthe uctuating pressure beneath a turbulent bound-ary layer at the wall is 5 percent of the local staticpressure at Mach 5. Measurements obtained in the80HTT during the present tests (ref. 24) at the sameMach number indicate lower uctuating pressures,which were 3 percent of the local static pressure.This percentage was used to analyze the stresses onthe shields of probes A and B at the highest pres-sure condition. The loads on the thermocouple wireswere determined by treating the wires as circularcylinders that were inclined to the ow at approx-imately 6�. The ow in the normal direction to thewires was then analyzed. According to the experi-ments of Gerrard (ref. 25), the dominant uctuatingload for cylinders at subcritical Reynolds numbers isthe uctuating lift. Farell (ref. 26) found that thepeak uctuating lift is of the same order of magni-tude as the steady drag. The uctuating load on thethermocouple wires was estimated at 2:35�10�4 lbmwith the drag coe�cient data compiled by Schlichting(ref. 27). According to the results of Richter andNaudascher (ref. 28), the estimated ow con�nemente�ect of the radiation shield on the steady and uc-tuating loads on the thermocouple wires is low. Thehighest vortex-shedding frequency is estimated at ap-proximately one-half the resonant frequency of thewires.

Both the shield and thermocouple wires weretreated as cantilevered beams that were simply sup-ported, and the maximum bending stresses werecalculated with the standard equations given inreference 29 as follows:

�max = mr=I (D1)

In equation (D1), I = �=64(d4so� d4

si) for the radia-

tion shield, where I is the cross-sectional moment of

inertia. Treating the thermocouple as a single wirefor simplicity, I = �=64(d4

w). Also, r is the outside

radius of the radiation shield or the radius of thethermocouple wire. The bending moment is given as

m = wL2=2 (D2)

In equation (D2), w is the distributed load per canti-levered length L and is given as p0A=L, where p0 is the uctuating pressure and A is the projected area. Themaximum bending stress for the shield was calculatedfrom these equations to be 15.7 and 25.5 psi forprobes A and B, respectively. The maximum bendingstress for the thermocouple wires was calculated tobe 0.122 psi.

The factors of safety for the shield and wires werecomputed by extrapolating the ultimate strengthdata given in reference 22 for pure platinum atvarious temperatures to the design temperatureof 3300�R. The extrapolated ultimate strength is ap-proximately 1000 psi. For cyclic loading, the fatiguestrength is a percentage of the ultimate strength. Al-though these data are not speci�cally available forplatinum, reference 29 gives a value of 30 percentfor nonferrous metals. The resulting factors of safetyare 19.1 and 11.8 for the shields of probes A and B,respectively. The factor of safety for the thermo-couple wires is even higher at 2450. Becauseplatinum-rhodium alloys have a higher yield strengththan pure platinum, at least at room temperature,the factors of safety may actually be higher.

The thermocouple wire should not de ect andcontact the radiation shield. The maximum de- ection was calculated, assuming that the wire issupported as a cantilevered beam (ref. 29):

ymax=wL4

8Y I(D3)

In equation (D3), Y is the modulus of elasticity andis given in reference 30 as 15� 106 psi for platinum{6-percent rhodium at 3200�R. The maximum de- ection of the thermocouple wire was calculated tobe 1:18� 10�8 in.

14

Page 18: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

References

1. Peecher, D. W.; and Shreeve, R. P.: Stagnation Tem-

perature Measurement at High Mach Number Using Very

Small Probes. D1-82-0945, Boeing Scienti�c ResearchLabs., Jan. 1970. (Available from DTIC as AD 704 962.)

2. Beckwith, Ivan E.; Harvey, William D.; and Clark,

Frank L. (appendix A by Ivan E. Beckwith, William D.

Harvey, and Christine M. Darden and appendix B byWilliam D. Harvey, Lemuel E. Forrest, and Frank L.

Clark): Comparisons of Turbulent-Boundary-Layer Mea-

surements at Mach Number 19.5 With Theory and an As-

sessment of Probe Errors. NASA TN D-6192, 1971.

3. Weinstein, Leonard M.: Hot-Wire Coil Probe for High-

Speed Flows. AIAA J., vol. 11, no. 12, Dec. 1973,pp. 1772{1773.

4. Danberg, James E.: The EquilibriumTemperature Probe,

A Device for Measuring Temperatures in Hypersonic

BoundaryLayers. NOLTR 61-2, U.S. Navy, Dec. 4, 1961.

5. Winkler, E. M.: Stagnation Temperature Probes for Use

at High Supersonic Speeds and Elevated Temperatures.

NAVORD Rep. 3834, U.S. Navy, Oct. 12, 1954.

6. East, R. A.; and Perry, J. H.: A Short Time Response

Stagnation Temperature Probe. C.P. No. 909, British

Aeronautical Research Council, 1967.

7. Rom, Josef; and Kronzon, Yigal: Small ShieldedThermo-

couple Total Temperature Probes. TAE-79, Israel Inst.

of Technology, Dec. 1967. (Available from DTIC asAD 668 466.)

8 Weinstein, Leonard M.: A Shielded Fine-Wire Probe for

Rapid Measurement of Total Temperature in High-Speed

Flows. J. Spacecr. & Rockets, vol. 8, no. 4, Apr. 1971,pp. 425{428.

9. Keener, Earl R.; and Hopkins, Edward J.: Turbulent

Boundary-LayerVelocity Pro�les on a NonadiabaticFlat

Plate at Mach Number 6.5. NASA TN D-6907, 1972.

10. Bartlett, R. P.; Edwards, A. J.; and Hillier, R.: Devel-

opment and Calibration of a Total Temperature Probe for

the Imperial College Aeronautics Department Gun Tun-

nel. IC-AERO-79-02, Imperial College of Science and

Technology (London, England), June 1979.

11. Albertson, Cindy W.: Evaluation of Equilibrium Turbu-lence for a Hypersonic Boundary Layer at Nonadiabatic

Wall Conditions. M.S. Thesis, Old DominionUniv., 1989.

(Available as NASA TM-101663, 1989.)

12. Nowak, Robert J.; Albertson, Cindy W.; and Hunt,

L. Roane: AerothermalTests of a 12.5� Cone at Mach 6.7

for Various Reynolds Numbers, Angles of Attack, and

Nose Shapes. NASA TP-2345, 1985.

13. Powell, Robert L.; Hall, William J.; Hyink, Clyde H., Jr.;

Sparks, Larry L.; Burns, George W.; Scroger, Mar-garet G.; and Plumb, Harmon H.: Thermocouple Refer-

ence Tables Based on the IPTS-68. NBS Monogr. 125,

U.S. Dep. of Commerce,Mar. 1974.

14. Venkateswaran, S.; Hunt, L. Roane; and Prabhu,Ramadas K.: ComputationalMethod To Predict Thermo-

dynamic, Transport, and Flow Properties for the Mod-

i�ed Langley 8-Foot High-Temperature Tunnel. NASA

TM-4374, 1992.

15. Mo�at, Robert J.: Gas Temperature Measurement.Temperature|Its Measurement and Control in Science

and Industry, Volume 3, Part 2, Reinhold Publ. Corp.,

c.1962, pp. 553{571.

16. White, FrankM.: ViscousFluid Flow. McGraw-Hill,Inc.,

c.1974.

17. Fernholz,H. H.; and Finley,P. J.: A Critical Commentary

on Mean Flow Data for Two-Dimensional Compressible

TurbulentBoundaryLayers. AGARD-AG-253,May 1980.

18. Leyhe, E. W.; and Howell, R. R.: Calculation Proce-

dure for Thermodynamic, Transport, and Flow Proper-

ties of the Combustion Products of a Hydrocarbon Fuel

Mixture Burned in Air With Results for Ethylene-Airand

Methane-AirMixtures. NASA TN D-914, 1962.

19. Incropera, Frank P.; and DeWitt,DavidP.: Fundamentals

of Heat andMass Transfer,Third ed. JohnWiley& Sons,Inc., c.1990.

20. Pitts, DonaldR.; and Sissom, LeightonE.: Schaum's Out-

line of Theory and Problems of Heat Transfer. McGraw-Hill, Inc., c.1977.

21. Kays, W. M.; and Crawford, M. E.: ConvectiveHeat and

Mass Transfer, Second ed. McGraw-Hill, Inc., c.1980.

22. Lyman, Taylor, ed.: Metals Handbook. Volume I|

Propertiesand Selectionof Metals, 8th ed. AmericanSoc.

for Metals, c.1961.

23. Bushnell, Dennis M.; Cary, Aubrey M., Jr.; and Harris,

Julius E.: Calculation Methods for Compressible Turbu-

lent Boundary Layers|1976.NASA SP-422, 1977.

24. Parrott, Tony L.; Jones, Michael G.; and Albertson,

Cindy W.: Fluctuating Pressures Measured Beneath a

High-Temperature,Turbulent Boundary Layer on a Flat

Plate at a Mach Number of 5. NASA TP-2947, 1989.

25. Gerrard, J. H.: An Experimental Investigation of theOscillating Lift and Drag of a Circular Cylinder Shed-

ding Turbulent Vortices. J. Fluid Mech., vol. 11, pt. 2,

Sept. 1961, pp. 241{256.

26. Farell, C�esar: Part I|Flow Around Circular Cylinders:

UnsteadyLoads. Proj. Rep. No. 188, Univ. of Minnesota,

Nov. 1979.

27. Schlichting, Hermann (J. Kestin, transl.): Boundary-

Layer Theory, Seventh ed. McGraw-Hill Book Co.,

c.1979.

15

Page 19: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

28. Richter, A.; and Naudascher, E.: Fluctuating Forces on a

Rigid CircularCylinder in Con�ned Flow. J. Fluid Mech.,

vol. 78, pt. 3, 1976, pp. 561{576.

29. Shigley, Joseph Edward: Mechanical Engineering Design,

Third ed. McGraw-Hill, Inc., c.1977.

30. Papadakis, E. P.; Fowler, K. A.; Lynnworth, L. C.;

Robertson, A.; and Zysk, E. D.: UltrasonicMeasurements

of Young's Modulus and ExtensionalWave Attenuation in

Refractory Metal Wires at Elevated Temperatures With

Application to Ultrasonic Thermometry. J. Appl. Phys.,

vol. 45, no. 6, June 1974, pp. 2409{2420.

16

Page 20: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

Table I. Total Temperature Probe Locations

z, in., for|

Designationa Rake A Rake B

1 b0.060 c0.035

2 b.180 c.100

3 b.300 c.165

4 b.420 c.230

5 b.540 c.295

6 b.660 c.360

7 b.800 c.425

8 b1.000 c.540

9 b1.400 c.600

10 b1.900 b.800

11 b2.500 b1.000

12 b1.400

13 b1.900

14 c2.200

15 b2.500

aSee �gure 3.bProbe type A. See �gure 2(a).cProbe type B. See �gure 2(b).

Table II. Combustor and Local Flow Conditions

[Surface pressure and heat transfer measurements were taken during some runs with the rakes removed]

Ttc, ptc, �, Rake location pwall, Twall, Ue, �e, Te, Tte, �T, �U,

Run �R psia deg x, in. y, in. psia �R Me ft/sec lbm/ft3 �R �R in. in. NRe;x

10 3250 1500 13.05 49.00 0.00 0.92 627 4.99 6210 3:79� 10�3 636 3210 0.48 0.47 7:45� 106

16 3340 1970 12.85 60.00 .00 1.22 624 4.94 6300 4.91 651 3270 .72 .68 11.70

18 3120 2480 12.92 60.00 .00 1.56 604 4.96 6040 6.89 593 3030 .76 .72 16.77

19 3280 1490 12.83 60.00 .00 .91 610 4.98 6080 3.82 623 3110 .69 .66 9.15

22 2590 2320 10.82 60.00 .00 1.04 591 5.05 5390 5.89 462 2530 .72 .55 15.68

24 2610 1740 10.85 60.00 .00 .76 597 5.16 5480 4.33 460 2600 .61 .52 11.81

25 3330 1000 4.97 60.00 .00 .21 573 6.33 6300 1.33 414 3090 .42 .49 4.47

30 3050 1500 12.98 86.12 .00 .93 584 4.95 5920 3.98 612 2990 1.03 .92 13.76

31 3180 2500 13.03 86.12 .00 1.66 632 4.88 6200 6.20 701 3210 1.13 1.03 19.94

32 3230 1500 12.72 86.12 .00 .94 597 4.96 6080 3.93 626 3100 1.00 .97 13.62

33 3140 3280 12.70 86.12 .00 2.12 630 4.92 6080 8.51 652 3140 1.15 1.00 28.56

17

Page 21: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

Table III. Summary of Theoretical Error Calculations

Case 1

[NRe;p = 2470; Tt = 3140�R]

Error, �R Percent of Et Percent of TtEU 6.3 51.7 0.2

Ec 2.2 18.0 .1

Er 3.7 30.3 .1

Et 12.2 100.0 .4

Case 2

[NRe;p = 63;Tt = 2340�R]

Error, �R Percent of Et Percent of TtEU 4.7 3.3 0.2

Ec 85.7 58.9 3.7

Er 55.0 37.8 2.3

Et 145.4 100.0 6.2

18

Page 22: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

Supportsleeve

CeramicType R

thermocoupleRadiation

shield

Vent hole (2)

Flow

(a) Probe A.

CeramicType R

thermocoupleRadiation

shield

Vent hole (2)

Flow

(b) Probe B.

Figure 1. Total temperature probes.

19

Page 23: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

A

A

Top view

15°

Section A-A

.03

Section C-CSection B-B

.005

B

B

C

C

Platinum–20-percent-rhodium tubing,

.0585 o.d.,.049 i.d.

Epoxy

Two ventholes,

180° apart,.022 diam

Type R, .010diam wire 308 stainless steel

support sleeve,.0937 o.d., .0595 i.d.

Two-holeceramic,.048 o.d.,.012 i.d.

Fiberglass-covered

two-conductorwire

1.22.21.19

.09

.005

(a) Probe A.

A

A

Top view

Section A-A.020

Section B-B

B

B

Platinum–20-percent-rhodium tubing,

.0595 o.d.,.049 i.d.

EpoxyType R, .010

diam wire

Two-hole ceramic,.048 o.d.,.012 i.d.

Fiberglass-coveredtwo-conductor wire

1.22.21.19

.09Two vent holes,

180° apart,.022 diam

.005

(b) Probe B.

Figure 2. Schematic of total temperature probes. All linear dimensions are in inches.

20

Page 24: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

Top view

Set screw Wiringchannel

Side view

2.70

11

10

9

8

76

54

32

1

.48

2.42

.48.2515°

.015 rAccess

door

Probe designation

(a) Rake A.

Figure 3. Schematic of total temperature rake without probes. All linear dimensions are in inches.

21

Page 25: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

Top view

Set screw Wiringchannel

Side view

2.70

15

14

98 7

65

43

21

.48

2.42

.48.25

15°

.015 r

13

12

11

10

Probe designation

(b) Rake B.

Figure 3. Concluded.

22

Page 26: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

275

30 52

Pod

Air ejector

DiffuserMixing tube

Supersonic diffuser

Test chamber(26-ft diam)Fuel (methane)

Air

Nozzle

Combustor

Ttc

Tt,∞

Figure 6. Langley 8-Foot High-Temperature Tunnel. All dimensions are in feet.

.5

Diffuser

Model positionduring tunnelstartup andshutdown

Flow

Nozzle(8-ft exit diam)

6.0

12.014.0

1.0Pitchcenter

Figure 7. Cross-sectional view of test chamber of Langley 8-Foot High-Temperature Tunnel. All dimensionsare in feet.

25

Page 27: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

Ttc

TJ (outside boundarylayer at z = 2.500 in.)

5000

4000

3000

2000

1000

0-.5 0 .5 1.0 1.5 2.0 2.5 3.0 3.5 4.0

t, sec

T, °R

(a) Boundary-layer probe and combustor total temperature histories.

Test sectioncenterline

-100

-80

-20

20

60

100-.5 0 .5 1.0 1.5 2.0 2.5 3.0 3.5 4.0

t, sec

H,in.

Jet boundary

Nozzleedge

-60

-40

0

40

80

(b) Boundary-layer probe position in test stream.

Figure 8. Typical total temperature and probe position histories. Run 16; NRe;p = 1016 at a distance of 72 in.from nozzle exit plane.

26

Page 28: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

Shock

Thermocouple wire

TJ

Tt2T2pt2M2

Radiation shield

Vent hole

Ceramic

Flow

Ttk, pt,∞,lT∞,l, M∞,l

Figure 9. Top view of ow through a total temperature probe.

.07.06.05.04.03.02.01.8

.9

1.0

1.1

Type A probe; Ttc = 3200°R

Type B probe; Ttc = 3200°R

Type A probe; Ttc = 2600°R

Type B probe; Ttc = 2600°R

TJ – T∞,lTtc – T∞,l

NNu,w

Figure 10. Probe recovery factor as a function of Nusselt number based on wire conductivity.

27

Page 29: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

50403020100.8

.9

1.0

1.1 Type A probe; Ttc = 3200°R

Type B probe; Ttc = 3200°R

Type A probe; Ttc = 2600°R

Type B probe; Ttc = 2600°R

TJ – T∞,lTtc – T∞,l

pt 2 TJ

–1.75, psia-°R–1.75

× 10–6

Figure 11. Probe recovery factor as a function of modi�ed Winkler parameter.

28

Page 30: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

DataPrediction

.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0TJ

, °R

z,in.

1.2

1.1

1.0

.9

.8

.7

.6

.5

.4

.3

.2

.1

0 × 103

δU

(a) NRe;x = 28:6� 106 (run 33).

DataPrediction

.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0TJ

, °R

z,in.

1.2

1.1

1.0

.9

.8

.7

.6

.5

.4

.3

.2

.1

0 × 103

δU

(b) NRe;x = 13:6� 106 (run 32).

Figure 12. Comparison of total temperature data with Crocco-Busemann predictions.

29

Page 31: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

DataPrediction

.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0TJ

, °R

z,in.

1.2

1.1

1.0

.9

.8

.7

.6

.5

.4

.3

.2

.1

0 × 103

δU

(c) NRe;x = 4:5� 106 (run 25).

Figure 12. Concluded.

30

Page 32: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

qc•

qk• qk

• +dqk•

qc•

dn

.010 in.

n = 0 n = L1 n = L2

Insulation

qc•

(a) Conduction.

qc,si•

qc,J•

qr,J to si•

qr,si to J• qr,so to ∞

Radiationshield

Thermocouplejunction

(b) Radiation.

Figure 13. Decoupled heat transfer analysis for estimation of probe measurement errors.

31

Page 33: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

L-88-3993

Figure 4. Panel holder in test chamber of Langley 8-Foot High-Temperature Tunnel.

L-88-3994

Figure 5. Boundary-layer rake assembly.

Page 34: Total Temperature Probes for High-Temperature Hypersonic ...mln/ltrs-pdfs/tm4407.pdfNASA Technical Memorandum 4407 Total Temperature Probes for High-Temperature Hypersonic Boundary-Layer

REPORT DOCUMENTATION PAGEForm Approved

OMB No. 0704-0188

Public reporting burden for this collection of information is estimated to average 1 hour per response, including the time for reviewing instructions, searching existing data sources,gathering and maintaining the data needed, and completing and reviewing the collection of information. Send comments regarding this burden estimate or any other aspect of thiscollection of information, including suggestions for reducing this burden, toWashington Headquarters Services, Directorate for Information Operations and Reports, 1215 Je�ersonDavis Highway, Suite 1204, Arlington, VA 22202-4302, and to the O�ce of Management and Budget, Paperwork Reduction Project (0704-0188), Washington, DC 20503.

1. AGENCY USE ONLY(Leave blank) 2. REPORT DATE 3. REPORT TYPE AND DATES COVERED

March 1993 Technical Memorandum

4. TITLE AND SUBTITLE

Total Temperature Probes for High-Temperature Hypersonic

Boundary-Layer Measurements

6. AUTHOR(S)

Cindy W. Albertson and Willard A. Bauserman, Jr.

7. PERFORMING ORGANIZATION NAME(S) AND ADDRESS(ES)

NASA Langley Research Center

Hampton, VA 23681-0001

9. SPONSORING/MONITORING AGENCY NAME(S) AND ADDRESS(ES)

National Aeronautics and Space Administration

Washington, DC 20546-0001

5. FUNDING NUMBERS

WU 506-43-21-01

8. PERFORMING ORGANIZATION

REPORT NUMBER

L-17124

10. SPONSORING/MONITORING

AGENCY REPORT NUMBER

NASA TM-4407

11. SUPPLEMENTARY NOTES

12a. DISTRIBUTION/AVAILABILITY STATEMENT 12b. DISTRIBUTION CODE

Unclassi�ed{Unlimited

Subject Category 35

13. ABSTRACT (Maximum 200 words)

The design and test results of two types of total temperature probes that were used for hypersonic boundary-layer measurements are presented. The intent of each design was to minimize the total error and to maintainminimal size for measurements in boundary layers 1.0 in. thick and less. A single platinum{20-percent-rhodium shield was used in both designs to minimize radiation heat transfer losses during exposure to thehigh-temperature test stream. The shield of the smaller design was attened at the ow entrance to an interiorheight of 0.02 in., compared with 0.03 in. for the larger design. The resulting vent-to-inlet area ratios were60 and 50 percent. A stainless steel structural support sleeve that was used in the larger design was excludedfrom the smaller design, which resulted in an outer diameter of 0.059 in., to allow closer placement of theprobes to each other and to the wall. These small design changes to improve resolution did not a�ect probeperformance. Tests were conducted at boundary-layer-edge Mach numbers of 5.0 and 6.2. The nominal free-stream total temperatures were 2600� and 3200�R. The probes demonstrated extremely good reliability. Thebest performance in terms of recovery factor occurred when the wire-based Nusselt number was at least 0.04.Recommendations for future probe designs are included.

14. SUBJECT TERMS 15. NUMBER OF PAGES

Total temperature probe; High-temperature measurement;

Hypersonic boundary layer32

16. PRICE CODE

A0317. SECURITY CLASSIFICATION 18. SECURITY CLASSIFICATION 19. SECURITY CLASSIFICATION 20. LIMITATION

OF REPORT OF THIS PAGE OF ABSTRACT OF ABSTRACT

Unclassi�ed Unclassi�ed

NSN 7540-01-280-5500 Standard Form 298(Rev. 2-89)Prescribed by ANSI Std. Z39-18298-102

NASA-Langley, 1993