Top 10 considerations when validating an autoclave.

download Top 10 considerations when validating an autoclave.

of 12

Transcript of Top 10 considerations when validating an autoclave.

  • 7/24/2019 Top 10 considerations when validating an autoclave.

    1/12

    Top 10 considerations when validating an autoclave

    Validating an autoclave is a daunting and time-consuming task. This white paper details the

    tricks, tips and traps to such a validation project from how to choose your Control to which load

    configuration to use for your validation runs.

    Consideration 1

    Choosing the right sterilisation cycle to implement

    There are three basic types of sterilisation cycles. Choose the right one according to the type of

    goods to be sterilised:

    ard !oods "Vacuum#$uitable for items that are easy to sterilise, because air removal and

    steam penetration are highly effective on these items. e.g., open glassware and large diameter

    piping % typical hard goods cycle may draw one vacuum prior to introducing steam to reach the

    desired sterilisation temperature.

    &rapped !oods "Vacuum#'tili(ed for items that are difficult to sterilise, because air

    removal and steam penetration are harder to achieve on these items than on hard goods. e.g.,

    empty bottles "glass or plastic# with lids, gowns, long hoses)tubes, vent filters, portable vessels

    with small inlet)outlet ports % typical wrapped goods cycle may draw three or more vacuumsprior to reaching sterilisation. % post-sterilisation vacuum draws the steam from the load items.

    *i+uids "on-vacuum#tems that contain li+uids generally cannot have a deep vacuum

    pulled or the li+uid will be drawn out of them. %utoclave cycles for li+uids generally heat up and

    cool down without a vacuum. $team, introduced into the top of the chamber, displaces the air.

    The air is pushed to the bottom of the chamber and is removed.

    Consideration 2

    Which load configurations to use

    % var iable to consider is whether to use fied load or variable load configurations. There/s a

    trade-off here between validation effort and operational fleibility 0 do you want to validate a

  • 7/24/2019 Top 10 considerations when validating an autoclave.

    2/12

    wide range of load configurations to increase 1roduction/s fleibility in loading the autoclave2

    ere are some typical load configurations to consider:

    % fied load)fied positionconfiguration means that any load to be sterilised will be placed

    inside the chamber in eactly the same way for every processing run. % diagram of the load

    configuration should appear in the $tandard 3perating 1rocedure "$31# so that operators can

    reproduce the load for every processing run. This situation re+uires the fewest validation runs

    "4#, but offers no fleibility in load configurations.

    % fied load)variable position configuration means that the location of the load items in the

    autoclave can vary. 3nly a list of the items that can be in a load is re+uired for the $31. The

    validation runs must demonstrate positional e+uivalence by rotating the items from location to

    location during the test runs. f positional e+uivalence is proven after three validation runs, then

    you can stop. % fied load)variable position configuration gives operators fleibility in loading the

    autoclave. This saves time when loading large loads of numerous items of different types.

    % variable load configurationmeans that different combinations of items and)or numbers of

    any item"s# can be placed into the chamber. The validation runs must demonstrate that the

    cycle is ade+uate for both a maimum and minimum load configuration. The minimum load tests

    are done with only one item in the autoclave, that item being the load item demonstrated as

    being the most difficult to sterilise .

    Consideration 3

    Choosing the right Control for liquid cycles

    The choice of the Control used when sterilising li+uids determines whether the load you are

    sterilising will pass all the acceptance criteria. 5ore than one li+uid Control may be needed to

    validate all the different types of bottles and li+uids re+uiring sterilisation. Consider the following

    when choosing the Control for li+uid cycles:

    The si(e of the bottle and its fill volume 0 the larger the bottle and the greater its

    volume, the harder it is to sterilise.

    The thickness of the glass 0thicker glass is more difficult to sterilise than thinner

    glass.

    The viscosity of the li+uid 0 the greater the viscosity, the slower the heat-up time, and

    the harder it is to sterilise.

  • 7/24/2019 Top 10 considerations when validating an autoclave.

    3/12

    The li+uid Control will ideally be the one that is the most difficult to sterilise "worst-case# and

    will be located at the coldest spot in the chamber "lower level near the front door or directly

    above the drain#.

    6on/t be tempted to use a Control that is dramatically different from the composition of the load.

    f the li+uid Control takes too long to reach the sterilisation temperature, then the protein

    composition of the media in the rest of the load "which may have eceeded the desired

    temperature by the time the Control reaches the sterilisation temperature# may be denatured.

    f in doubt, perform preliminary studies using different li+uid Controls to obtain information on

    the load7s heat-up times and 89-values. The number of validation runs re+uired for different

    types of li+uids and bottles can be reduced by grouping li+uids with similar viscosities, bottle

    si(es and fill volumes. ach li+uid Control will have a uni+ue maimum and minimum load

    configuration associated with it.

    'se procedural controls to ensure that the choice of li+uid, bottle si(e and fill volume used for

    each Control, and its location in the chamber, are maintained during the validation runs and

    subse+uent processing runs.

    Case study of load configurations and choice of Control:

    *et/s assume that a laboratory prepares four different media types in three different bottle si(es

    in the following configurations:

    Table 1 !edia types and bottles si"es for loads in the autoclave

    MediaMax/Min

    Load

    100 mL

    bottle

    500 mL

    bottle600 mL bottle

    Media

    AFixed at 50 400ml fill N/A

    MediaA

    Max:160Min:40

    50ml fill N/A N/A

    Media

    B

    Max: 144 Min:

    9N/A

    400ml &

    500ml fill

    500ml &

    600ml fill

    Media

    C

    Max: 144 Min:

    9N/A

    400ml &

    500ml fillN/A

  • 7/24/2019 Top 10 considerations when validating an autoclave.

    4/12

    Media

    D

    Max: 144 Min:

    9N/A

    400ml &

    500ml fillN/A

    5edia % is the most viscous fluid of those listed. t is prepared in two very different bottle si(es:

    ;99 m* "with a fied load of ;9 bottles in the autoclave# and 99 m*. This means that Control 4 is a >99

    m* bottle filled with 5edia A ">99 m*# for the variable load covering the three different Aroths,

    i.e., si studies: 4 99 m* bottles.

    ote: ach validation load consisted of the relevant media type only in the locations where

    thermocouples, Aiological ndicators "A# and Chemical ndicators "C# were to be placed. The

    rest of the load was the same si(ed bottles filled with the re+uired volume of water.

    Consideration #

    $etermining which load items are the most difficult to sterilise and whichlocation%s& within the items represents the worst'case conditions

    &ith a large load containing a wide variety of different types of items, the number of possible

    test locations within items seems to approach infinity. t also can be difficult to get the

    thermocouple and indicators "A C# into the item without affecting the item/s ability to be

    sterilised and)or ruining the item "a concern with epensive items#.

    &e must evaluate each item on a case-by-case basis and determine how to best challenge the

    item. 3ften the item must be sealed somehow to return it to a state that represents e+uivalency

    with respect to steam penetration.

    $ome eamples:

    D.&hat is the most difficult point to sterilise in a hose of uniform diameter2

    %.n the centre of the length of hose.

  • 7/24/2019 Top 10 considerations when validating an autoclave.

    5/12

    D.ow do you get a 4 m length of thermocouple into the middle of a @9 m hose2

    %.Ay cutting a slot in the middle of the hose and inserting the thermocouple through the slot,

    making sure to seal the slot with silicon. f you don/t seal it, you will not be challenging the hose

    properly. %lternatively, get two

  • 7/24/2019 Top 10 considerations when validating an autoclave.

    6/12

    &ires can get caught "and be damaged# under the autoclave7s wheels when moving

    loads into and out of the chamber

    t is difficult to place wires inside sealed bottles without "i# touching the inside wall,

    and "ii# compromising the bottle/s ability to be sterilised.

    Eou may be limited by the number of wires you can place through the autoclave7s

    inlet.

    The resistance of the thermocouple in some locations in the chamber can change,

    leading to inaccurate and)or unreliable data even though the pre)post calibration

    verifications meet specifications.

    Case *tudy

    This eample looks at nine thermocouples placed into a loaded chamber. They were evenly

    spaced from one another at the top, middle and bottom levels and at the front, centre and rear

    of the chamber. %ll were in the chamber and subject to 1hase heat distribution re+uirements.

    The study used wired thermocouples in a loaded chamber for a =9-minute cycle at

  • 7/24/2019 Top 10 considerations when validating an autoclave.

    7/12

    t/s important to understand the aim of the autoclave cycle and what its parameters are. 8or

    eample, is it for sterilisation or decontamination2 s the load heat sensitive, or can it be

    subjected to an overkill cycle2 s it a porous load "hard)wrapped goods#, or is it a li+uid2

    5ost Validation departments have a $tandard 3perating 1rocedure "$31# detailing the

    validation re+uirements for sterilisation processes. ncluded in that is a complete list of all the

    acceptance criteria.

    ach phase of the autoclave cycle is likely to have different acceptance criteria:

    1hase 0 eat distribution "empty chamber#

    1hase 0 eat distribution "loaded#

    1hase 0 eat distribution "loaded# cold spot determination within

    1hase V 0 eat penetration

    There may also be different re+uirements for 1hases and V if you are sterilising li+uids "non-

    vacuum# vs. porous items "vacuum#, e.g., 89 I

  • 7/24/2019 Top 10 considerations when validating an autoclave.

    8/12

    6ocumenting what was done during the validation test runs is all about knowing what needs to

    be documented and how to present it. This documentation must be clear, consistent between

    runs and transparent, and must conform to all !51 re+uirements. t must be complete and must

    include the following items:

    a diagram showing the location of all load items within the autoclave chamber

    the precise location)number of each thermocouple, A and C within each item

    the printout from the data recorder

    the printout or chart from the autoclave

    the time the sterilisation period began and finished "per the data recorder time#

    the time difference between the autoclave controller and the validation temperature

    monitoring device

    the results of each A and C

    *abel each document with the e+uipment 6, load description, date, test run number and cycle

    start)end time. f you fail to generate good documentation while conducting the validation test

    runs, you will not be able to analyse the data when putting together the report. nade+uate or

    poor +uality data to support the validation process will not survive the scrutiny of an auditor.

    Tip: Ae cautious about the acceptance criteria you employ to verify the accuracy

    of thermocouples. f the criterion is too tight "e.g., all thermocouples must meet

    the acceptance criteria#, you may lose a lot of runs if one or two thermocouples

    cease functioning or are outside the temperature tolerance after the runs.

    Consideration .

    /aving adequate time and access to the autoclave to complete thevalidation

    f you are performing a large number of test runs "e.g., over the course of several weeks#, you

    need to think about the points at which you will verify the thermocouples7 accuracy. This could

    be done after every run and)or at the end of the entire testing period. f you wait until the end of

    the testing period, you run the risk that all of the runs are of no value due to their failure to meet

    the verification acceptance criteria. Verifying after every run, however, adds considerably to the

    length of time re+uired to complete the testing. 1erforming the verification every three runs or

    every few days is a reasonable compromise.

  • 7/24/2019 Top 10 considerations when validating an autoclave.

    9/12

    %s noted in Consideration >, the acceptance criterion you employ to verify the thermocouples7

    accuracy should allow at least one or two thermocouples to fail.

    Consideration

    /aving adequate time and access to the autoclave to complete thevalidation

    t/s easy to under estimate the length of time it takes to validate an autoclave, and how much

    access you need to it during the process. 8or eample, it can take up to four hours to set up

    your run, i.e., prepare the load, place probes, As and Cs into the load, etc. f 1roduction needs

    to use the autoclave and you need to remove your probes, As and Cs, then you need to start

    all over again, effectively losing a day. &ork with the 1roduction department when planning the

    Validation project to ensure that you have ade+uate access to the autoclave. %nother approach

    is to combine 1hase and "eat 6istribution# with 1hase V "eat 1enetration# studies to

    save time. Combining these three phases could reduce the time it takes to complete the

    validation project? however, you need to consider the following when doing this:

  • 7/24/2019 Top 10 considerations when validating an autoclave.

    10/12

    f you are validating a new autoclave, then you need to allow enough time for: "i# writing a

    validation plan "ii# writing the commissioning and D protocols "iii# preparing the 3D)1D

    protocols "iv# performing 3D)1D studies "v# writing the 3D)1D reports, preparing folders, etc.

    f you are performing a large number of test runs "e.g., over the course of several weeks#, then

    you need to ensure that enough time has been allocated to prepare the folders and write the

    reports. %llow one day to do a run and another day to analyse the data, i.e., two days per study.

    %lso allow time to write protocols and reports, and to have them reviewed and approved by

    other people "if appropriate#. f you are developing the validation cycles, then this will also take

    time

    Consideration 10

    /ave the right procedural controls in place to ensure ongoing consistencyand correct operation

    Congratulations, you have just finished validating a new autoclave for a number of different

    cycles and load configurations. ow, what controls need to be in place to ensure that the

    validated loads are used consistently2

    % $tandard 3perat ing 1rocedure "$31# for the new autoclave should be prepared. t

    must include clear guidelines for each of the validated cycles, including diagrams of the

    load configurations. Test the procedure7s clarity by asking a typical operator to follow the

    instructions with a dummy load.

    ach operator who uses the autoclave should be trained and tested on the $31.

    *ogbooks should be in place for each cycle.

    'se a risk-based approach to determine the troubleshooting guidelines to include in

    the $31. The manufacturer/s documentation and website may detail things that commonly

    go wrong.

    %n ongoing re+uali fication program for the autoclave and the loads is re+uired. The

    fre+uency can be >,

  • 7/24/2019 Top 10 considerations when validating an autoclave.

    11/12

    !allen$e te a#to!lave%s performan!e

    Chemical

    Indicator (CI)

    ommer!iall' availa"le indi!ators tat

    indi!ate expos#re to steam

    Decontamination

    Arti!les are free of mi!ro"ial !ontaminants "#t

    are not s#ita"le for #se in a (rade A sterilearea)

    0 !al"e

    *is is te n#m"er of min#tes re+#ired to ,ill

    a spe!ified n#m"er of mi!roor$anisms wit a

    -.val#e of 10 at a temperat#re of 11)1 )

    Load*e items pla!ed into te a#to!lave to "e

    sterilised

    #!er$illA 1.lo$ red#!tion in te n#m"er of

    mi!roor$anisms wit an SA of 10.6

    %AL

    Sterilit' Ass#ran!e evel) An SA of 10.6 is a

    one in one million pro"a"ilit' of a sin$le

    via"le mi!roor$anism "ein$ present)

    %&T tem'erat"re

    Sterilit' Ass#ran!e evel) *e temperat#re at

    wi! sterilisation o!!#rs) *is is te

    temperat#re te a#to!lave is eld at for a

    spe!ified period of time to rea! te re+#iredsterilisation level)

    %teriliation

    *e destr#!tion of all livin$ mi!roor$anisms

    s#! as pato$eni! or saprop'ti! "a!teria2ve$etative forms and spores) Arti!les tat ave"een sterilised !an "e #sed for asepti!

    p#rposes in all areas2 in!l#din$ areas !lassified

    as (rade A sterile areas)

    Thermoco"'le A temperat#re meas#rin$ devi!e

    -bout harmut

    1harm3ut is a bouti+ue consultancy to the 1harmaceutical, 5edical 6evice and Veterinary drug

    industries.

    1harm3ut specialises in !51 compliance, validation and continuous improvement consulting

    and training. $ome of the company7s customers include %bbott *aboratories, %$TJ%LC%,

    Aernafon, C$*, 8onterra, !$M, ospira, nvetech, ntertek Caleb Arett, 5ayne 1harma,

    1harmatel 8resenius Mabi and 1robe %nalytical *aboratories.

  • 7/24/2019 Top 10 considerations when validating an autoclave.

    12/12

    /ow harmut can help

    &e offer a range of services to help manufacturers achieve regulatory compliance.

    4alidation services

    1harm3ut offers both consulting and contracting services to support the validation of

    manufacturing e+uipment and computer systems. 3ur consulting services can help you with

    high-level decisions, such as taking a risk-based approach to determine which of your computer

    systems and e+uipment to validate, and determining the documentation needed to support

    decisions about what to validate.

    &e can also perform an audit to find out if your systems are compliant with global !1

    regulations. f they aren/t, we can give you practical solutions for achieving compliance and

    passing an audit by a regulatory body. 3ur contract offering supplies people with the knowledge

    and eperience to do the hands-on validation. This includes:

    &riting a Validation 5aster 1lan

    6etermining what to test

    Creating the documentation needed for testing

    &riting a re+uirements traceability matri

    1lus, we can do the testing Nwhich you probably don/t have the resources to do.