Tono en Gonio

download Tono en Gonio

of 100

Transcript of Tono en Gonio

  • 8/6/2019 Tono en Gonio

    1/100

    About your upcoming exam

    50 questions Multiple choice, four possible answers

    All questions taken directly from lecture Lecture notes are available in the

    optometrie section in the library and online

    Similar in style to the last exam No embryology

    No slides

  • 8/6/2019 Tono en Gonio

    2/100

    Lecture notes online

    For those who dont have it, the URL is: http://virtmed.fg.hvu.nl/domeinen/opto

    metrie/optometrie.html

    The password is redeye

  • 8/6/2019 Tono en Gonio

    3/100

    About your upcoming examCOSM3

    Questions per topic covered: Methods (direct and indirect examination

    of the fundus, tonometry, gonioscopy): 5

    Fundus Landmarks: 2

    Congenital variations: 8-9

    Optic Neuropathy: 13 Fundus Spots: 21-22

    Total: 50

  • 8/6/2019 Tono en Gonio

    4/100

    About your upcoming examCOSMB

    Questions per topic covered: DPAs: 15-16

    Methods (direct and indirect examination

    of the fundus, tonometry, gonioscopy): 5

    Fundus Landmarks: 2

    Normal fundus: 7-8 Congenital variations: 8

    Optic Neuropathy: 12

    Total: 50

  • 8/6/2019 Tono en Gonio

    5/100

    Tonometry

  • 8/6/2019 Tono en Gonio

    6/100

    Tonometry

    One abbreviation that will be usedthroughout your career is IOP, whichstands for intraocular pressure

    Get used to it

  • 8/6/2019 Tono en Gonio

    7/100

    Tonometry

    Measures tension in eye The combined resistance of het oog (IOP)

    and the tear film

    The main reason we measure this isbecause high pressure is a key indicator

    of glaucoma Pressure is elevated in most types ofglaucoma

  • 8/6/2019 Tono en Gonio

    8/100

    Tonometry

    Pressure can also be high in cases ofocular inflammation

    Uveitis

    Perforating injuries to the globe mayresult in an abnormally low pressure

    This is called hypotony

  • 8/6/2019 Tono en Gonio

    9/100

    Tonometry

    Measured in mm Hg There are three ways to measure IOP

    Manometry

    Applanation

    Indentation

  • 8/6/2019 Tono en Gonio

    10/100

    Tonometry

    Manometry The most accurate method of determining

    IOP

    A needle connected to a column ofmercury is inserted into the eye

    Not generally performed in clinical practice

  • 8/6/2019 Tono en Gonio

    11/100

    Tonometry

    Applanation Based on the Imbert-Fick formula

    IOP = force/area

    This is the same formula used in physics

    If one measure is kept constant, pressure

    is directly related to the other variable The Goldmann tonometer is the one we

    use, and the worldwide standard

    Also using this theory is the Maklolov unit

  • 8/6/2019 Tono en Gonio

    12/100

    Tonometry

    The Goldmann tonometer has aconstant area

    The force required to flatten the cornea is

    directly related to the IOP

    The Maklokov tonometer applies a

    constant force The area flattened is directly related to theIOP

  • 8/6/2019 Tono en Gonio

    13/100

    Tonometry

    IndentationA constant force applied to the cornea

    The cornea is pushed posteriorly

    The IOP is related to the distance thecornea is pushed

    Influenced by corneal rigidity Tends to be lower in high myopes, who havelarger eyes and therefore slightly softer corneas

    Aqueous humor is forced from the eye

    Subsequent readings will be lower

  • 8/6/2019 Tono en Gonio

    14/100

    Tonometry

    Normal IOP The average IOP is 16 mm Hg

    Normal IOP is anywhere from 6 to 21 mm

    Hg

    An eye with an IOP under 6 is consideredhypotonous

    Look for a wound leak (do the Seidel test)

  • 8/6/2019 Tono en Gonio

    15/100

    Tonometry

    Normal IOPAn eye with an IOP over 21 is referred to

    as hypertensive

    Most people with ocular hypertension do notdevelop glaucoma

    What this means is that some people have

    pressure which is higher than most others, butfor them this is normal

    You should make sure of your diagnosis ofocular hypertension with your anamnese and

    other tests

  • 8/6/2019 Tono en Gonio

    16/100

    Tonometry

    Physical factors affecting IOP Pressure on Globe

    Holding lids--some people just wont keep their

    eyes open no matter what you say, so youhave to force them open

    By doing this, you can inadvertently press onthe globe, giving a false high reading

    Blepharospasm, or forceful blinking, can do thesame thing

  • 8/6/2019 Tono en Gonio

    17/100

    Tonometry

    Physical factors affecting IOP Trauma/Inflammation

    Damage to the trabecular meshwork will

    prevent the proper drainage of aqueous fromthe eye

    This happens in angle recession glaucoma

    Uveitis

    In the beginning of the condition, IOP is slightlylowered due to decreased aqueous production

    In the later stages, IOP increases as a result of cells

    and flare, clogging the trabecular meshwork

  • 8/6/2019 Tono en Gonio

    18/100

    Tonometry

    Physical factors affecting IOP Medications

    Steroids

    Use of these over a long period of time causesin increase in IOP in about 1/3 of all people

    Blood pressure medication

    Beta blockers, alpha agonists, and carbonicanhydrase inhibitors can lower IOP

    Marijuana/alcohol

    The effect is gone when it leaves the body

  • 8/6/2019 Tono en Gonio

    19/100

    Tonometry

    Physiological factors affecting IOP Diurnal variation

    IOP is highest at around 10AM, and lowest in

    the late avond Normal variation is 4mm Hg per day, but may

    be twice that in glaucoma

    If you suspect glaucoma in a patient with highpressure, get a morning reading

    If still unsure, you can do a diurnal curve

    Start in the morgen, and read IOP every 1-2uur

  • 8/6/2019 Tono en Gonio

    20/100

    Tonometry

    Physiological factors affecting IOP Vascular integrity

    In cases of hypoxia, there will be low IOP from

    decreased aqueous production Since this occurs throughout the eye, there

    may be nerve damage despite low IOP

    Patient position for measurement IOP is 2-3 mm higher lying down

    This is not a problem with current techniques

  • 8/6/2019 Tono en Gonio

    21/100

    Tonometry

    Techniques Finger tensions (digital IOP)

    Goldmann

    Perkins

    Non-contact

    Pulsair Tonopen

    Pneumotonometer

    Maklakow Glaucotest

  • 8/6/2019 Tono en Gonio

    22/100

    Tonometry

    Finger tensions (digital IOP) Not accurate

    Should only be done when all other

    methods are not possible

    Lightly push on the closedeye

    Compare each eye to the other

    Rate whether soft, normal, or hard

  • 8/6/2019 Tono en Gonio

    23/100

    Tonometry

    Goldmann applanation tonometry Is the gold standard because of its

    accuracy and repeatability

    The only method used to diagnose andtreat glaucoma

    Used with the spleetlamp

    There is also a hand-held version made bymany companies, including Perkins

  • 8/6/2019 Tono en Gonio

    24/100

    Tonometry

    Goldmann applanation tonometry Uses a tip which has prisms inside

    The tip is 7mm in diameter, but only 3.06mm

    of the cornea is applanated (flattened) by thetip

    The prisms split the image horizontally topermit a proper measurement

    Each half of the image contains a semi-circle,which is called a mire

    When the mires are aligned properly, you have

    determined the IOP

  • 8/6/2019 Tono en Gonio

    25/100

    Tonometry

    This is what the mires look like whenthe unit is properly aligned and the IOPreading is correct

  • 8/6/2019 Tono en Gonio

    26/100

    Tonometry

    Procedure for Goldmann tonometry Disinfect tonometer tip

    Soak in 3% peroxide solution (contact lens

    solutions work well) for tien minutes and thenrinse with saline solution

    Place tip in holder

    Align 180 mark of tip with horizontal line If the patient has more than drie diopters of

    corneal astigmatism, place the minus cylinder

    axis along the red line of the tonometer

  • 8/6/2019 Tono en Gonio

    27/100

    Tonometry

    Procedure for Goldmann tonometry Place a drop of anesthetic in het oog

    Instill fluorescein

    Switch to diffuse illumination of maximumintensity with the cobalt blauw filter

    Observe cornea for staining

    Start with the right oog

    Pull spleetlamp back

    Place illumination system at 60 temporally

  • 8/6/2019 Tono en Gonio

    28/100

    Tonometry

    Procedure for Goldmann tonometry Set measuring drum at 10

    Note: the drum has numbers 0-8 on it. These

    stand for 0 through 80. There are four linesbetween each number. These correspond totwee mm Hg each.

    The line by the number 1 is 10 mm Hg. Theline after that is 12mm Hg. If the reading isbetween the first and second lines after thenumber 1, this is a reading of 13 mm Hg.

  • 8/6/2019 Tono en Gonio

    29/100

    Tonometry

    Procedure for Goldmann tonometry Place tip in position, directly facing the

    patient

    Have the patient look straight ahead Bring tip toward het oog, looking from

    outside the spleetlamp

    When the limbus has a blauw glow, lookthrough the eyepieces

    Move spleetlamp forward until you see themires

  • 8/6/2019 Tono en Gonio

    30/100

    Tonometry

    Procedure for Goldmann tonometryAdjust the mires until they are centered

    Turn the drum until the mires are properly

    aligned Remove tip

    Check for staining again

  • 8/6/2019 Tono en Gonio

    31/100

    Tonometry

    To properly align the mires vertically, movethe spleetlamp in the direction of the largermire. In the picture on the left, you would

    lower the spleetlamp, and this would alignthe mires. In the picture on the right, youwould have to raise the instrument. To alignthe mires horizontally, move the spleetlampin the direction of the mires. So in thepicture on the right, you would have to movethe instrument to the right as well as up.

  • 8/6/2019 Tono en Gonio

    32/100

    Tonometry

    To have a proper reading, the inside edges ofthe mires should touch. The picture on theleft requires you to increase the reading on

    the power drum. The picture on the rightrequires a decrease in the reading. Thepicture in the center shows a tonometerwhere the power is correct.

  • 8/6/2019 Tono en Gonio

    33/100

    Tonometry

    15

    16

    1 Oxybuprocaine 16 30

    To properly record, you must document thetest used, the drop used, the IOP reading,the time it was done, the fact that you told

    the patient not to rub their eyes, and if thelids were held. If multiple readings weretaken, write them all down and averagethem. The method we use here is below.

    lidsheld

  • 8/6/2019 Tono en Gonio

    34/100

    Tonometry

    If there is too much fluorescein, as in thepicture on the left, the mires will be too thickand you will get a false high reading

    If there is not enough fluorescein, as in thepicture on the right, the mires will be too thinand you will get a false low reading

  • 8/6/2019 Tono en Gonio

    35/100

    Tonometry

    If you are pushing too hard on the eye, as inthe picture on the right, you will havedistorted mires which will not move when youchange the reading drum

    If you are not pushing hard enough, as in thepicture on the left, the mires will go in and

    out of view

  • 8/6/2019 Tono en Gonio

    36/100

    Tonometry

    Things to remember Alignment for excessive corneal cylinder

    If a patient has an ocular pulse (also called

    a hippus) then the reading will fluctuate Set the reading so that the center of this

    movement will be where proper alignment is

    If you must hold the lids, you must movethe spleetlamp and the power drum withyour other hand

    Scarred corneas give distorted mires

  • 8/6/2019 Tono en Gonio

    37/100

    Tonometry

    The hand-heldtonometer works thesame way as theGoldmann tonometer

    The power wheel iscontrolled by yourthumb

    You should also recordthe patients position

  • 8/6/2019 Tono en Gonio

    38/100

    Tonometry

    Non-contact tonometry (NCT) Works by applanation (by air)

    A puff of air flattens a fixed area of cornea

    The air stops when the cornea is flattenedenough to allow a light to be reflected to areceiver

    The amount of air needed is related to theIOP, and the machine displays the nummer

    Must average drie readings per oog

    OK for screening purposes

    T

  • 8/6/2019 Tono en Gonio

    39/100

    Tonometry

    air pulse

    Monitoring

    mirror system

    Light source

    Air pulse

    Non-Contact Tonometer (NCT)

    T t

  • 8/6/2019 Tono en Gonio

    40/100

    Tonometry

    Non-contact tonometry procedure Disinfect chin & headrest

    Demonstrate puff

    Switch instrument to regular mode

    Align patient comfortably

    Patient closes eyes

    Set and test safety lock

    Pull instrument slightly back

    Have patient open their eyes

    T t

  • 8/6/2019 Tono en Gonio

    41/100

    Tonometry

    Non-contact tonometry procedure Direct fixation

    Align targets

    Instrument may shoot automatically ormanually

    Take drie readings and average

    Documentation is the same as with theGoldmann

    T t

  • 8/6/2019 Tono en Gonio

    42/100

    Tonometry

    Keeler Pulsair non-contact tonometer We have one of these in the preclinic

    Hand held

    Portable

    Can be used in any position

    Requires a steady hand

    Machine averages vier readings

    T t

  • 8/6/2019 Tono en Gonio

    43/100

    Tonometry

    Tonopen Uses a combination of applanation &

    indentation techniques

    Hand-held, portable, digital, quick, welltolerated

    Excellent back-up for Goldmann

    Tonometry

  • 8/6/2019 Tono en Gonio

    44/100

    Tonometry

    Tonopen procedure Calibrate instrument

    Place sterile disposable cover over tip

    Anesthetize eyes

    Direct fixation

    Press reading button listen for beep

    Hold instrument perpendicular to cornea

    Tonometry

  • 8/6/2019 Tono en Gonio

    45/100

    Tonometry

    Tonopen procedure Gently tap the cornea 3-5 times

    The instrument makes a sound every time

    it gets a reading If it is not reliable, then you should repeat

    the procedure

    It tends to overestimate low IOP andunderestimate high IOP

    Tonometry

  • 8/6/2019 Tono en Gonio

    46/100

    Tonometry

    Im still trying to figure out why the Schiotz,Pneumotonometer, and Heine-Maklakowinstruments are in the module.

    None of these have been used in clinicalpractice for years

    Read them over once, so you have a clue if

    some old-timer starts talking about them Obviously, they will not be on the exam

  • 8/6/2019 Tono en Gonio

    47/100

    Gonioscopy

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    48/100

    Gonioscopy

    Technique used to view the anteriorchamber angle

    It is impossible to directly view the

    angle with the spleetlamp alone We need special lenses/prisms

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    49/100

    Gonioscopy

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    50/100

    Gonioscopy

    Indications: Grade 2 or less on Van Herrick technique

    Suspicion of abnormality in angle

    trauma, maldevelopment, neoplasm,neovascularization

    Pigment dispersion syndrome

    Pseudoexfoliation, exfoliation Glaucoma has been diagnosed or is

    suspected

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    51/100

    Gonioscopy

    Contraindications: Traumatic hyphema

    Corneal abrasion

    Laceration or perforation Post-surgery

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    52/100

    Gonioscopy

    Direct gonioscopy High plus lens, light source, and magnifier

    You can look directly into the angle

    You can view the angle 360 around It provides an erect, virtual image

    The most common is the Koeppe lens,which provides 24X magnification

    Patient must be lying down

    This is rarely used

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    53/100

    Gonioscopy

    Koeppe lens

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    54/100

    Gonioscopy

    Indirect gonioscopy Mirrors and prisms are used in combinationwith the spleetlamp

    The mirrors and prisms do not magnify The mirror/prism should be 180 away

    from the angle you want to observe

    Inverted, virtual images

    Goldmann, Sussman, Posner, Zeiss makelenses for this purpose

    Scleral and corneal lenses are used

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    55/100

    Gonioscopy

    Scleral lenses There are many types available The most common is the Goldmann 3-mirror

    They are larger in size than the cornea,therefore they rest on the sclera

    Require a solution to keep a tight seal

    It should be relatively thick

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    56/100

    Gonioscopy

    Corneal lenses They have a smaller diameter than thecornea

    They rest on the tear film They do not require a solution to stick to

    het oog

    There is limited lens manipulation

    These are less stable than scleral lenses

    Some have a handle for better control

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    57/100

    Gonioscopy

    There are vijf major structures thatneed to be evaluated in every angle

    From posterior to anterior, they are:

    Iris root (IR)

    Ciliary body (CB)

    Scleral spur (SS)

    Trabecular meshwork (TM)

    Schwalbes line (SL)

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    58/100

    Gonioscopy

    I ris root (IR) C iliary body (CB)

    Scleral spur (SS)

    Trabecularmeshwork (TM)

    Schwalbes line (SL)

    I Cant

    See

    This

    S%@#

    The easiest way to remember the structures is to

    remember this statement, in which the words all startwith the same letters as the anatomical structures

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    59/100

    Gonioscopy

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    60/100

    Go oscopy

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    61/100

    py

    There are other things that we need tolook for as well

    Iris processes

    Iris/angle neovascularization Peripheral anterior synechiae (PAS)

    Pigment

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    62/100

    py

    Schwalbes line The termination of Descemets membrane Where the cornea ends

    Color varies from clear to light brown If anterior chamber pigment is deposited

    here, it is called Sampaolesis line

    There is a specific method used to see thisstructure

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    63/100

    py

    Schwalbes line Use an optic section rather than aparallelpiped to see where it is

    The light will shine on both the cornea andangle

    Where they meet is Schwalbes Line

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    64/100

    py

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    65/100

    py

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    66/100

    Trabecular meshwork Is composed of fenestrated sheets ofepithelium

    Its function is to remove aqueous from hetoog

    It can be subdivided into anterior and

    posterior

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    67/100

    Trabecular meshwork

    The anterior portion is less pigmented

    The posterior part does more of the work

    Therefore more pigment will accumulate hier The actual color depends on the

    pigmentation of the patient and the

    amount of free pigment available If you push hard on the lens, you can

    observe Schlemms canal, it will be red

    from blood

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    68/100

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    69/100

    Blood in Schlemms canal

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    70/100

    Scleral Spur

    Ring of collagen and elastic tissue

    Is the end of the trabecular meshwork

    The longitudinal muscles of the ciliary bodyinsert here

    Appears as a white line

    May not be easily found in light eyes, asthere is no pigment to contrast with

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    71/100

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    72/100

    Ciliary body

    It has many functions

    Aqueous production

    Accommodation Point of attachment for iris

    Most of it lies behind the iris

    We can only see the ciliary body band The color varies depending on the patients

    overall pigmentation

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    73/100

    Iris root

    This is the end of the iris

    You must observe how it attaches to the

    ciliary body In cases of trauma:

    The iris may be torn, which is called

    iridodialysis The ciliary body may be torn, which is known

    as an angle recession

    9% of all patients with angle recessionsdevelo laucoma as a result

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    74/100

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    75/100

    Iris processes

    A normal finding

    Are fine strands of iris that attach to the

    trabecular meshwork You must distinguish from these from

    anterior synechiae and neovascularization

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    76/100

    Iris processes

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    77/100

    Iris Processes

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    78/100

    Iris/Angle Neovascularization

    Also known as rubeosis irides

    New bloedvat growth

    These vessels are fragile and leaky Fibrin leaks out of the bloedvat and

    attaches itself to the angle structures,

    stopping the flow of aqueous This happens in response to ischemia

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    79/100

    Peripheral anterior synechiae (PAS)

    Are areas of the iris that have becomeattached to angle structures

    These are irregular, thick, and oftenobscure your view of the angle

    In comparison, iris processes are thin and lacy

    They usually attach to the trabecularmeshwork or Schwalbes line, but mayattach to the cornea or ciliary body as well

    These can close the angle completely

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    80/100

    Peripheral anterior synechiae (PAS)

    Can be caused by angle closure,inflammation, trauma, neovascularization,and after laser procedures

    You can also see tumors in the angle

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    81/100

    Peripheral anterior synechiae

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    82/100

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    83/100

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    84/100

    Indirect Corneal Lens

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    85/100

    Indirect corneal lens procedure Educate and anesthetize the patient Disinfect and rinse lens

    Patient is placed behind the spleetlamp Observation and illumination systems at 0

    2-3 mm parallelpiped

    Pull spleetlamp all the way back

    The patient should look straight ahead

    Place lens on het oog

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    86/100

    Indirect corneal lens procedure

    The lens has four mirrors, which should beat 3, 6, 9, and 12 oclock

    Stabilize your hand on the head rest or onthe patients cheek

    Gently press against corneal surface

    Look in a different mirror to observe adifferent angle

    Pull the lens away to remove it

    It comes off very easily, unlike the scleral lens

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    87/100

    Indirect corneal lens procedure

    You should use the parallelpiped toobserve the entire angle

    The beam should be horizontal for the superiorand inferior angles and vertical for the nasaland temporal angles

    To look for Schwalbes line, use an optic

    section The beam should be vertical for the superior

    and inferior angles, and horizontal for the nasal

    and temporal angles

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    88/100

    Indirect corneal lens procedure

    More difficult to use than the scleral lens

    The examiner must be very steady

    You must maintain good contact with thetear film without wrinkling Descemets orcausing blood to back up into Schlemms

    canal If the patient has a bowed iris, have the

    patient look slightly toward the mirror

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    89/100

    Wrinkling of Descemets membrane

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    90/100

    Indentation gonioscopy

    Only done with corneal lenses

    Used to determine if a closed angle is that

    way because of synechiae or apposition You will see structures if the closure is a result

    of apposition

    This can be treated with a peripheral iridectomy

    If none are seen, then the closure is fromsynechiae

    This is not easy to treat

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    91/100

    Indentationgonioscopy

    This angle isclosed due to

    apposition

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    92/100

    Angle classification

    There are many methods, some contradictothers

    The most common and clinically usefulmethod is to:

    Record the most posterior structure seen

    Grade pigmenting of the trabecular meshwork Indicate any abnormalities Evaluate in all vier quadrants and record

    on an X

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    93/100

    Angle classification

    Read over the Schaffer and Scheie systems

    They are actually graded opposite of the

    van Herrick technique and can createconfusion for examiners

    This is why the previously described

    method is widely used However, a new system has emerged and

    it is what we use in our clinic

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    94/100

    Spaeth angle classification We must determine vier things The site of iris insertion

    The iris approach The angle of insertion

    The amount of pigment

    This system allows for recording of irisprocesses and results of indentationgonioscopy

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    95/100

    Spaeth angle classification

    Site of insertion

    Anterior to Schwalbes line

    Behind Schwalbes line, trabecularmeshwork is not visible

    sC leral spurDeep angle, with visible ciliary body

    Extremely deep

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    96/100

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    97/100

    Spaeth angle classification

    Iris approach

    queer--concave iris

    regular configuration

    steeply convex (bowed) iris

    Gonioscopy

  • 8/6/2019 Tono en Gonio

    98/100

    Gonioscopy

    Spaeth angle classification

  • 8/6/2019 Tono en Gonio

    99/100

    Spaeth angle classificationInsertion angle

    (in degrees)

    Gonioscopy

    Recording

  • 8/6/2019 Tono en Gonio

    100/100

    Recording

    OD OS

    E 40 r0 pigment

    B 10 s1 pigment

    E 40 q

    iris neorecessionC 25 r

    4+ pigment

    A 10 s

    4+ synechiae

    Normal Abnormal